You are on page 1of 6

Diafiltration for Desalting

or Buffer Exchange
Larry Schwartz

D
iafiltration is an
ultrafiltration
membrane technique
for completely
removing, replacing, or
lowering the concentration of salts
or solvents from solutions
containing proteins, peptides,
nucleic acids, and other
biomolecules. The process
selectively uses permeable (porous)
membrane filters to separate the
components of solutions and
suspensions based on their
molecular size. Smaller molecules
such as salts, solvents, and water
pass freely through the
ultrafiltration membrane, which
retains the larger molecules. A typical development laboratory system comprising a Centramate holder with a
peristaltic pump and pressure gauges, valves, and tubing. PALL CORPORATION
CONCENTRATION (WWW.PALL.COM)
The solution retained by a
membrane is known as concentrate characteristics for the mixture that is pressure. Therefore, when selecting
or retentate. The solution that to be concentrated. As a general a membrane for concentration or
passes through a membrane is rule, the molecular weight cut-off diafiltration, consider the time factor
known as the filtrate or permeate. A (MWCO) of a membrane should be versus product recovery. In most
membrane for concentration is a third to a sixth the molecular biological applications, recovery
selected based on its rejection weight of the molecule to be outweighs the time consideration.
retained. This is known as the 3–6 Increasing the amount of membrane
Rule, which ensures complete area used can also reduce the
PRODUCT FOCUS: ALL BIOLOGICALS retention. The closer the MWCO is process time.
to that of the sample, the greater Figure 1 provides an example of
PROCESS FOCUS: DOWNSTREAM the risk for some product loss concentration. The sample is placed
PROCESSING occurring during concentration. in a device containing a suitable
WHO SHOULD READ: PROCESS That risk increases if diafiltration ultrafiltration membrane that will
DEVELOPMENT, MANUFACTURING
will also be used because the relative retain the large molecules. Pressure
loss depends on the total volume of is applied until half the volume has
KEYWORDS: ULTRAFILTRATION, filtrate generated. passed through that membrane.
DIAFILTRATION, BUFFER EXCHANGE, Membrane flux rate (defined as the Large molecules are retained in half
DESALTING, PROTEIN CONCENTRATION filtrate flow rate per unit area of the original volume (concentrate),
membrane) is related to pore size. which also contains half of the salt
LEVEL: INTERMEDIATE The smaller the pores, the lower the molecules. The filtrate contains the
flux rate for the same applied other half of the salt molecules but
MAY 2003 BioProcess International 43

Figure 1: 2 concentration of a sample mixture by ultrafiltration. Large circles represent
molecules that are bigger than the pores in the membrane, and small circles represent
molecules (such as salts or solvent) that are smaller than those pores.

Another
ADVANTAGE
of using
diafiltration is
concentration of
samples on one
system, minimizing
the risk of sample
loss or
contamination.

none of the large molecules.


Therefore, the large molecules are
concentrated as liquid and salt are
removed. The salt-molecule-to-
volume ratio in the concentrate
remains constant, so the ionic
strength of the concentrated Table 1: Continuous (constant volume) diafiltration
solution remains relatively constant.
Diafiltration, the process of Removal of Small Molecules
“washing” the remaining salt out Diafiltration Permeability 100% Permeability 75%
with water, can subsequently reduce Volumes Rejection Coefficient  0 Rejection Coefficient  0.25
the ionic strength of a concentrate 1 63% 53%
(retentate) solution. This is 2 86% 77%
essentially a dilution process 3 95% 89%
performed in conjunction with 4 98.2% 95%
concentration. Water is added while 5 99.3% 97.6%
filtrate is removed. If the washing 6 99.7% 98.9%
solution is another buffer instead of 7 99.9% 99.4%
water, the new buffer salt will 8 99.7%
9 99.9%
replace the initial salt in the sample.
For simplicity, the examples herein Note: 0% rejection salts, solvents, buffers, etc.; 25% rejection molecules lower in molecular weight than
the molecular weight cutoff of the membrane, but bigger than salts
use a direct-flow filtration device
such as a centrifugal concentrator.
But the same principles apply to
cross-flow filtration devices such as
cassettes and hollow fibers. a process can lead to the possibility and volume required to complete
of sample loss or contamination. the process varies considerably. It is
BENEFITS OF DIAFILTRATION With diafiltration, salt or solvent important to understand the
Other techniques used for salt removal and buffer exchange can be different methods used and when to
removal or buffer exchange (such as performed quickly and conveniently. choose one over the other.
membrane dialysis and column- Another advantage of using
based gel filtration) can be effective diafiltration is concentration of CONTINUOUS DIAFILTRATION
but have certain limitations. Dialysis samples on one system, minimizing The technique of continuous
procedures can take several days, the risk of sample loss or diafiltration (also referred to as
require large volumes of water for contamination. Diafiltration is constant volume diafiltration)
equilibration, and risk product loss performed in three main ways: by involves washing out the original
through manual manipulation of continuous diafiltration and by buffer salts (or other low–molecular-
dialysis bags. Gel filtration dilutes discontinuous diafiltration through weight species) in the retentate
the sample and often requires an sequential dolution or volume (sample) by adding water or a new
additional ultrafiltration step to reduction. Although the end result buffer to it at the same rate filtrate
concentrate it back. Adding steps to may be the same overall, the time is generated. As a result, the
44 BioProcess International MAY 2003
Table 2: Salt reduction from sample using volume reduction or constant volume diafiltration

2 Volume Reduction Constant Volume


Diafiltration 100% Permeability 75% 100% Permeability 75%
Volumes 0% Retention 25% 0% Retention 25%

1 50% 41% 63% 53%


2 75% 65% 86% 77%
3 88% 79% 95% 89%
4 94% 88% 98.2% 95%
5 96.9% 93% 99.3% 97.6%
6 98.4% 95.6% 99.7% 98.7%
7 99.2% 97.4% 99.9% 99.4%
8 99.6% 98.4% 99.7%
9 99.9% 99.4% 99.9%
10 99.9% 99.4%
Note: 0% rejection salts, solvents, buffers, etc.; 25% rejection molecules lower in molecular weight than
the molecular weight cutoff of the membrane, but bigger than salts
A Minimate TFF Capsule device for
development and processing
applications using small sample
volumes. Samples can be concentrated Liquid is added at the same rate as Permeability is often described in
down to about 5 mL and diafiltered. filtrate is generated, and when the terms of the rejection coefficient of
PALL CORPORATION (WWW.PALL.COM) volume of filtrate collected equals the membrane (its ability to hold
the starting retentate volume, 1 DV back or reject a given molecule from
has been processed. Using passing through). Thus, see


continuous diafiltration, over 99.5% Equation 2.
of a 100% permeable solute can be A rejection coefficient of 1 equals
removed by washing through six 0% permeability. A rejection
Typically, the larger volumes (6 DV) with the buffer of coefficient of 0 equals 100%
the M O L E C U L E , choice. Molecules that are larger permeability. Permeability is affected
than salts and solvents but smaller by such factors as transmembrane
the lower its than the pores in the membrane also pressure (TMP), crossflow rate,
permeability and can be washed out. The retentate concentration, pH, ionic
the greater the permeability of such molecules, strength, and gel layer formation
wash volume however, may be less than 100%. It (concentration polarization).
required. takes more liquid (more DVs) to Therefore, the permeability may
completely wash a partially change over the course of the
permeable molecule through the diafiltration process.
membrane than it does to remove a Table 1 shows the relationship
100% permeable molecule from a between permeability through a
retentate volume and product mixture. Typically, the larger the membrane and the number of
concentration do not change during molecule, the lower its permeability diafiltration volumes required for
the diafiltration process. If water is and the greater the wash volume removal of permeating species. As
used for diafiltering, salts will be required. The permeability of a noted above, greater volumes of
washed out and the retentate particular molecule through a buffer are required to remove
conductivity lowered. specific membrane can be molecules that are partially retained.
If a buffer is used for diafiltering, determined by measuring the To remove 99.9% of a molecule that
the new buffer’s salt concentration concentration of that molecule in is 25% permeable to the membrane
will increase at a rate inversely the filtrate compared to its requires 9 DVs, whereas for a 100%
proportional to that of the molecular concentration in the retentate under permeable species, only 7 DVs are
species being removed. The amount specified conditions (Equation 1). required to achieve the same
of salt removed is related to the
filtrate volume generated relative to
the retentate volume. The filtrate Equation 1
volume generated is usually referred % Permeability  (ConcentrationFILTRATE  ConcentrationRETENTATE )  100
to in terms of diafiltration volumes
(DVs). Equation 2
A single DV is the volume of
retentate when diafiltration begins. Rejection Coefficient  1 – (ConcentrationFILTRATE  ConcentrationRETENTATE)
46 BioProcess International MAY 2003
Figure 2: Discontinuous diafiltration (sequential dilution). Large circles represent molecules
that are bigger than the pores in the membrane, and small circles represent molecules (such DIAFILTRATION DEFINITIONS
as salts or solvent) that are smaller than those pores.
Diafiltration: Diafiltration is a
technique that uses ultrafiltration
membranes to completely remove
or to lower the concentration of
salt or solvent — or to replace
buffer salts from solutions
containing proteins and other
large molecules.
Diafiltration Volume: One
diafiltration volume equals the
initial volume in which the
molecule of interest is suspended.
The number of diafiltration
volumes required depends on
whether the permeating molecular
species is freely passing (salts,
buffers, solvents) or partially
retained.

Figure 3: Discontinuous diafiltration (sequential dilution). Large circles represent molecules Continuous Diafiltration: The
that are bigger than the pores in the membrane, and small circles represent molecules (such technique of continuous
as salts or solvent) that are smaller than those pores. diafiltration (also referred to as
constant volume diafiltration)
involves washing out the original
buffer salts (or other
low–molecular-weight species) in
the retentate (sample) by adding
water or a new buffer to the
retentate at the same rate as
filtrate is being generated.
Discontinuous Diafiltration by
Sequential Dilution involves first
diluting the sample to a
predetermined volume, and then
concentrating that sample back to
its original volume with water or
replacement buffer. This is
repeated until the unwanted salts,
solvents, or smaller molecules are
removed. Each subsequent
dilution removes more of them.
removal. small molecules. As shown in
Discontinuous Diafiltration by
Figure 2, the sample is generally
Volume Reduction involves first
DISCONTINUOUS DIAFILTRATION diluted with an equal volume of
concentrating the sample to a
Sequential Dilution. Discontinuous buffer (1 DV). Adding volumes 1
predetermined volume, then
diafiltration by sequential dilution DV will be less effective. For
diluting that sample back to its
involves first diluting a sample with example, diluting the sample with
original volume with water or
water or a replacement buffer to a 5 DVs and concentrating back will
replacement buffer. This is
predetermined volume. The diluted reduce the salt by only 80%
repeated until the unwanted salts,
sample is then concentrated back to compared with 88% for 2 volume
solvents, or smaller molecules are
its original volume by ultrafiltration. reduction or sequential dilution.
removed. Each subsequent
The process is repeated until the Diluting the sample usually lowers
concentration and dilution
unwanted salts, solvents, or smaller viscosity, which may increase the
removes more of them.
molecules are removed. Each filtrate flux rate.
subsequent dilution and Volume Reduction. Discontinuous
concentration step removes more diafiltration by volume reduction reverses the sequential dilution
MAY 2003 BioProcess International 47
Figure 4: Determination of the CG value for a product.
find optimum conditions that
maximize productivity.

WHICH TECHNIQUE
SHOULD BE USED?
When deciding which technique to
use and where in the downstream
process to perform diafiltration,
consider the following factors:
•Initial sample volume,
concentration, and viscosity
•Required final sample
concentration
•Stability of the molecule of
equal to the volume where dilution interest at various concentrations


occurs. Therefore, half the volume •Volume of buffer required for
was required. diafiltration
That being the case, it would •Total processing time
Although it might seem that concentrating before •Reservoir size available
take less volume to diafiltration — by either •Economics.
diafilter a discontinuous sequential dilution or The choice of which method to
constant volume diafiltration — use must be based on several
concentrated
should reduce the required criteria. Scale is an important
sample, it could diafiltration buffer volume and save consideration. What we do at
take more time time. And in most cases this is true. laboratory scale may be very
compared to a less The factor we have not accounted different than at process scale,
concentrated for is filtrate flux rate, which equates especially if the process is
to process time. As the product automated. At lab scale,
sample.
becomes concentrated, viscosity discontinuous diafiltration is often
increases, and the filtrate flux rate used for its simplicity. Continuous
decreases. The filtrate flux rate varies diafiltration requires a pump or
inversely as the log of the special equipment to add the
procedure. A mixture is first concentration factor: diafiltration solution at a constant
concentrated to a predetermined rate. Both techniques can be
volume and then diluted back to its J  k ln (CGCB) automated for process applications.
original volume with water or a If we eliminate the equipment
replacement buffer. This is repeated where J  filtrate flux rate, issue and focus on the process itself,
until the unwanted salts, solvents, or k  constant, ln  natural log, we can compare the differences.
smaller molecules are removed. CG  gel layer concentration, and Ionic strength, buffer composition,
Each subsequent concentration and CB  retentate (bulk flow) and stabilizer concentration can
dilution removes more small concentration. This becomes very affect stability of the sample.
molecules (Figure 3). significant as the CB increases above Diafiltration may remove salts or
After the last buffer addition, the a few percent and is dependent on stabilizing molecules, resulting in
sample is generally concentrated the characteristics of the specific protein product denaturation and
before analysis or before the next molecules making up the sample aggregation. The process of
purification step is performed. After mixture. Although it might take concentrating and diluting a protein
diafiltration by either method significantly less volume to diafilter solution can affect molecular
(discontinuous 2 volume a concentrated sample, it could take interactions to cause denaturation or
reduction or sequential dilution), considerably more time compared aggregation, subsequent
the final product is at the same to a less concentrated sample. precipitation, and product loss. It is
volume and concentration as when Simple protocols are available to necessary to evaluate the effect of
diafiltration started. The salt
concentration has been equally
reduced in both examples. Equation 3
However, the volume of diafiltration Process Time  Filtrate Flow Rate  Volume
buffer used by the volume reduction
method reduced by the initial Equation 4
concentration step was half that
used in sequential dilution. A DV is ln (CG  CR)  1 or CROPTIMUM  CG  e  0.37  CG
48 BioProcess International MAY 2003
performing (continuous) Continuous diafiltration is also


diafiltration can be determined as in perceived as a “kinder and gentler”
Equation 4, where CG  gel layer process on active biomolecules.
concentration, CR  retentate On the other hand,
Stirred cells and concentration, and CROPTIMUM  discontinuous diafiltration with
centrifugal devices highest retentate concentration volume reduction requires less
are best suited for where diafiltration should be volume than continuous
performed (1). diafiltration. Concentrating a sample
discontinuous The CG value for a sample mixture before diafiltration usually
diafiltration; mixture can be determined from reduces the required filtrate volume
tangential flow experimentation by concentrating a and saves time. However, if sample
devices have the sample on a membrane while viscosity becomes too great, the
advantage of being recording and plotting data for filtrate flux rate will decrease, which
filtrate flux rate compared with log can increase process times
useful for either concentration (concentration substantially. Determining the CG
diafiltration factor). Then a curve can be for your sample can help answer the
technique. extrapolated to filtrate flux rate  0. question: At what concentration
The CG value will be the same for should I perform diafiltration?
this particular product regardless of Highly selective polyethersulfone
the starting concentration or filtrate (PES) microfiltration and
concentration on the product itself flux rate. ultrafiltration membranes with
to determine where diafiltration is In Figure 4’s example, the CG narrow pore size distribution and
best performed relative to value is a concentration factor of high flow rates provide fast
concentration effects. Continuous approximately 33. Therefore, the processing time and efficient
diafiltration offers an advantage over optimal concentration factor to separation for diafiltration. Initial
discontinuous diafiltration in that perform diafiltration would be volumes from a few milliliters up to
the retentate concentration remains 0.37  CG  12.2. If the starting thousands of liters can be processed
constant. It is often seen as a more product concentration is 5 mg/mL, using tangential flow filtration. TFF
gentle process. then diafiltration should be systems allow easy addition of
performed when that concentration membrane surface area, providing
BEFORE OR AFTER CONCENTRATION? reaches 61 mg/mL. If the final the flexibility to reduce processing
We have already seen that concentration will be less than 61 times or scale up your process. The
concentrating a sample first can mg/mL, then diafiltration should range of devices and systems
significantly reduce the volume of be performed after concentration available provides scalability as well
diafiltration solution required. We unless it is necessary to remove a as the flexibility to work with almost
have also seen that continuous specific molecule first. any sample volume.
diafiltration takes less volume than The ultrafiltration product
discontinuous diafiltration with selected may dictate your choice of REFERENCE
sequential dilution. Therefore, if the continuous or discontinuous 1 Beaton, NV; Klinkowski, PR.
sample is first concentrated to its Industrial Ultrafiltration Design and
diafiltration. Stirred cells and
Application of Diafiltration Processes.
required final concentration and centrifugal devices are best suited J. Separ. Proc. Technol. 1983, 4(2) 1–10. 
then continuous diafiltration for discontinuous diafiltration
performed, acceptable results should because of their mode of operation.
be obtained. Tangential flow devices have the Larry Schwartz is Senior Technical
However, above a certain advantage of being useful for either Manager at Pall Life Sciences,
concentration, filtrate flux rates may diafiltration technique. 516-801-9076, fax 516-801-9548;
become prohibitively slow. It may larry_schwartz@pall.com,
actually take longer to diafilter a A USEFUL STEP www.pall.com.
concentrated sample than it would if IN DOWNSTREAM PROCESSING
the sample were first diluted to Diafiltration is a fast and effective
reduce its concentration. In this technique for desalting or buffer
situation, even though continuous exchange of solutions. It can be
diafiltration of the diluted sample performed in a continuous or
requires a greater diafiltration discontinuous mode. Continuous
volume, the total processing time diafiltration usually takes less
would be lower because of the faster volume to achieve the same degree
filtrate flux rate (Equation 3). of salt reduction as discontinuous
In general, the optimum diafiltration with sequential dilution
retentate concentration for and can be easier to perform.
MAY 2003 BioProcess International 49

You might also like