Remarks : The Needle Detector operator must verify the calibration and accuracy 3 times per day.
At the beginning middle and end of a working period by using the 1.2 mm sphere ferrous cards to check machine sensitivity.
BOTTOM A1 A2 A3 A4 A5 A6 A7 A8
1. Switch on the machine and adjust sensitivity level to less than 1.2 mm sphere.
2. Position the 1.2 mm sphere ferrous check card onto the conveyor belt at position A1, let the test
card pass through the search head.
3. The needle detection should be activated ( alarming)
4. Repeat the procedure 1-3 with position the 1.2 mm sphere ferrous card on the conveyor belt at position A2, A3,A4…..A8.
5. The needle detection should be activated (alarming)also.
6. Test A1, A2 , A3…….A8 with 1.2 mm ferrous check card only. No garments or paper is allowed to put together during testing.
7. Repeat the procedure 1-3 with position the 1.2 mm sphere ferrous check card on the stand and placed the stand on
to the conveyor belt at the position A9,A10………..A24 end record.
"Y" = The needle detection activate (Alarming) when1.2 mm ferrous check and pass through the search head.
N'' = The needle detection activate (No Alarming) when 1.2 mm ferrous check and pass through the search head.
Auditor Approved By