You are on page 1of 8

Materials Science and Engineering C 63 (2016) 164–171

Contents lists available at ScienceDirect

Materials Science and Engineering C

journal homepage: www.elsevier.com/locate/msec

Perbandingan ketahanan fraktur implan gigi dengan sudut lancip abutmen yang berbeda
Comparison of the fracture resistance of dental implants with different
abutment taper angles
Kun Wang a, Jianping Geng b, David Jones c, Wei Xu d,⁎
a
Department of Dentistry, Southwest Hospital, Third Military Medical University, Chongqing 400038, China
b
Institute of Biomedical Engineering, Nanjing Tech University, Nanjing 211800, China
c
MicroStructural Studies Unit, Department of Mechanical Engineering Sciences, Faculty of Engineering and Physical Sciences, University of Surrey, Guildford, Surrey GU2 7XH, UK
d
Department of Mechanical Engineering Sciences, Faculty of Engineering and Physical Sciences, University of Surrey, Guildford, Surrey GU2 7XH, UK

a r t i c l e i n f o a b s t r a c t

Article history: To investigate the effects of abutment taper angles on the fracture strength of dental implants with TIS (taper in-
Received 25 October 2015 tegrated screwed-in) connection. Thirty prototype cylindrical titanium alloy 5.0 mm-diameter dental implants
Received in revised form 28 January 2016 with different TIS-connection designs were divided into six groups and tested for their fracture strength, using
Accepted 3 February 2016
a universal testing machine. These groups consisted of combinations of 3.5 and 4.0 mm abutment diameter,
Available online 4 March 2016
each with taper angles of 6°, 8° or 10°. 3-Dimensional finite element analysis (FEA) was also used to analyze stress
Keywords:
states at implant–abutment connection areas. In general, the mechanical tests found an increasing trend of im-
Dental implants plant fracture forces as the taper angle enlarged. When the abutment diameter was 3.5 mm, the mean fracture
Static compression test forces for 8° and 10° taper groups were 1638.9 N ± 20.3 and 1577.1 N ± 103.2, respectively, both larger than
Fracture resistance that for the 6° taper group of 1475.0 N ± 24.4, with the largest increasing rate of 11.1%. Furthermore, the differ-
Implant–abutment connection ence between 8° and 6° taper groups was significant, based on Tamhane's multiple comparison test (P b 0.05). In
Taper angle 4.0 mm-diameter abutment groups, as the taper angle was enlarged from 6° to 8° and 10°, the mean fracture
Finite element analysis value was increased from 1066.7 N ± 56.1 to 1241.4 N ± 6.4 and 1419.3 N ± 20.0, with the largest increasing
rate of 33.1%, and the differences among the three groups were significant (P b 0.05). The FEA results showed
that stress values varied in implants with different abutment taper angles and supported the findings of the static
tests. In conclusion, increases of the abutment taper angle could significantly increase implant fracture resistance
in most cases established in the study, which is due to the increased implant wall thickness in the connection part
resulting from the taper angle enlargement. The increasing effects were notable when a thin implant wall was
present to accommodate wide abutments.
© 2016 Elsevier B.V. All rights reserved.

1. Introduction implant–abutment interface and abutment screw to greater external


loads and bending moments, which can lead to screw joint opening
Most osseointegrated dental implant systems are composed of an and screw loosening [1,2]. Zarb and Schmitt reported the clinical out-
endosteal fixture, a component inserted into the jaw bone, and an come of 274 Brånemark implants with the external connection, and
abutment, connecting the fixture to support or retain the prosthetic su- they noted 9 abutment fractures and 53 gold screw fractures over a 4-
perstructure. The abutment is secured to the fixture with a mechanical to 9-year period [3]. In their one-year follow-up study, Jemt et al. report-
attachment method and is named the implant–abutment connection. ed that the overall success rate was 98.6% for the Brånemark implants,
At present, there are a number of implant–abutment connection with the most common complications related to loosening gold screws
designs offered by implant companies. They may be classified as either and esthetic complaints [4]. Moreover, problems of screw loosening or
externally or internally connected. fracture are more likely to occur when external connection implants
The external hexagonal interface of the original design of the are used to support single-unit restorations, where implants are not
Brånemark system, which is a typical external design, has been in use splinted and are subjected to multidirectional loading that challenges
the longest and has functioned well over the years. Recently, it the external connection components and restoration structural integri-
has been incorporated in a number of competing systems. However, ty. In a multicenter prospective study on external connection implants
the connection has the mechanical disadvantages of exposing the for single tooth replacement, the most obvious problem experienced
during the first year was related to loosening abutment screws with
⁎ Corresponding author.
an incidence of 26% [5]. Becker and colleagues found that retaining
E-mail addresses: dentalimplants@163.com (K. Wang), jpgeng2005@163.com screws loosened in 8 of 24 implants restoring single molars with
(J. Geng), d.g.jones@surrey.ac.uk (D. Jones), w.xu@surrey.ac.uk (W. Xu). follow-up of 24 months [6].

http://dx.doi.org/10.1016/j.msec.2016.02.015
0928-4931/© 2016 Elsevier B.V. All rights reserved.
K. Wang et al. / Materials Science and Engineering C 63 (2016) 164–171 165

Since the 1990s, several modifications of the design of the external 0.3 mm thread depth and 0.6 mm thread pitch. The superstructure
abutment screw, the material of the screw itself and the coefficient of part started from the TIS abutment and gradually widened to be the cor-
friction between the mated surfaces, have been made to reduce the con- onal restorative part, which was simplified into a combination of a cyl-
nection complications. However, mechanical complications of external inder (4.5 mm-diameter, 5.5 mm-height) and a hemispherical dome
connections are not eliminated and still remain a concern in the implant (Fig. 1). Two different abutment diameters, 3.5 mm and 4.0 mm, at
community. In order to overcome the connection problems, a new con- the implant platform level were designed. The abutment connection
cept of internal connection was developed. Contrary to the external part had a depth of 3.0 mm and taper angles of 6°, 8° and 10°. For the
connection, the internal connection design has a feature that extends purpose of brevity, each specimen was named by two hyphenated num-
from the inferior to the coronal portion of the implant and is located in- bers, representing the abutment diameter and taper angle, respectively.
side the implant body. The internal connection has a mechanical advan- Thus 5.0 mm-diameter implants had names of 3.5-6, 3.5-8, 3.5-10, 4-6,
tage of dramatically reducing screw failures by distributing occlusal 4-8, and 4-10, respectively.
forces deep into the implant and shielding the abutment retention
screw from excess loading. Further, deep joints in internal connections 2.3. Overview of the static test set-up
are more likely to resist bending forces than shallow joints in external
connections. Therefore, internal connections have superior joint The implant specimens were investigated in a test setup fabricated
strength than that of external counterparts [7]. according to the ISO 14801 static testing standard (Fig. 2). The implants
Of various internal connections, the taper integrated screwed-in were embedded and secured in a custom jig, which was made up of an
(TIS) abutment is becoming more popular, which uses simultaneously aluminum alloy cylinder and a stainless steel block. An internal
a screw and a tapered fit to provide mechanical stability. The TIS-type threaded hole in the depth of 10 mm was cut in the center of the alumi-
connection offers high resistance to loosening torques, and it has been num alloy cylinder to accommodate test implants, and the stainless
reported that loosening of the abutment is prevented [8]. Bozkaya and steel block functioned as a holder for the aluminum alloy cylinder. Im-
Müftü analyzed the mechanical properties of the TIS-type connection, plants were inserted into the threaded hole to a depth of 10 mm in a
with the focus on connection stability parameters of tightening and manner simulating 3.0 mm of the crestal bone loss. The jig carrying im-
loosening torques [9]. They developed analytical formulas to predict plant specimens was fixed onto the universal testing machine (Model
tightening and loosening torque values by combining the equations 6025; Instron, Canton, MA, USA) in such a way that specimens were
related to the tapered interface with screw mechanics equations. loaded with a 30° oblique force recommended by the ISO 14801 stan-
They found that the value of the coefficient of frictions, taper angles, dard. Off-axis loading was applied to the hemispherical cap of each
connection depth and outer radius of the implant were the factors implant by a flat indenter, ensuring the distance from the center of
affecting implant–abutment connection stability. the hemisphere cap to the cylinder surface (clamping plane) was
It has been well documented that the implant–abutment connection 11.0 mm. Therefore the moment arm was defined as 11.0 mm × sin
is the weakest part in terms of the whole implant mechanical strength, 30° (5.5 mm). The ISO 14801 was followed by ensuring unconstrained
especially for the internal connection designs which have a thin fixture movement of the loading member transverse to the loading direction.
wall at the connecting parts [10]. As for the TIS-type connection, those This was achieved by a socket fit joint between the loading member
abovementioned connection parameters can also affect its mechanical and the test machine structure. The joint was close to the load cell and
strength. Nowadays, different taper angles have been used by different was approximately 200 mm away from the lower end of the flat indent-
manufacturers in their TIS-type implants. However, there is only limited er. The abutments were tightened to the implants with a torque value of
information that can be found in the literature about the relationship 35 N cm using the BTG60CN-S torque gauge (Tohnichi, Tokyo, Japan).
between taper angles and the mechanical strength of implants. The pur- Ten minutes after the torque tightening, the test was carried out with
pose of the present in vitro research was to compare the compressive a crosshead speed of 0.5 mm/min until the implant fractured or exhib-
fracture strength of dental implants with different abutment taper an- ited a significant amount of plastic deformation. This kind of irreversible
gles. In addition to the experimental tests, 3-dimensional finite element deformation is determined by fitting the load–displacement curves
analysis (3D-FEA) was carried out to evaluate the stress state of im- with the regression lines, and the force at which the load–displacement
plant–abutment connection areas as a function of different abutment curve first deviates by 10% from the regression line will be recorded as
taper angles. an indicator for initiation of significant plastic deformation [12].

2. Material and methods 2.4. Statistical analysis & fracture analysis

2.1. Sample preparation for mechanical tests Throughout the loading, the raw data of force–displacement values
were recorded by the computer. Data were subsequently used to deter-
For this in vitro investigation, thirty prototype cylindrical titanium mine the maximum load levels and create force–displacement curves.
alloy (Ti–6Al–4V) implants divided into six groups (n = 5) were The mean and standard deviations of the fracture forces or the maxi-
fabricated using a BUMOTEC S-191 V (Bumotec SA, Switzerland) CNC mum deformation forces were determined, and Tamhane's multiple
(computer numerical control) machining center. The manufacturing di- comparison test was used to assess differences between groups. The
mensional tolerances were set to 10 μm. To the best of the authors' level of significance was set as P b 0.05.
knowledge, this accuracy tolerance level should be sufficient to satisfy After the mechanical testing, macrofracture mode analysis was per-
the testing requirements for the present study [11]. formed to identify different fracture modes for all the specimens. Fur-
ther fractographic analysis was performed using a scanning electron
2.2. Overview of implant specimen designs microscope (SEM) (JEOL, JSM-7100F, Japan). For SEM evaluations
implants were cleaned and dried. Digital images of the specimens
The TIS-type dental implant specimens used in this study can be di- were recorded at various magnifications to evaluate the fracture
vided into three parts: implant body, abutment, and the restorative part. surfaces.
The latter two parts were simplified into one section of the superstruc-
ture. The implant body had a diameter of 5.0 mm and length of 13.0 mm. 2.5. Finite element analysis
It consisted of two parts: the 2.0 mm-height of the non-threaded highly
polished cylindrical neck and an 11.0 mm-height of the threaded part. Numerical simulations were carried out to evaluate the mechanical
The threaded part featured a triangular thread design with a uniform properties of the implants with different abutment taper angles with
166 K. Wang et al. / Materials Science and Engineering C 63 (2016) 164–171

Fig. 1. The engineering drawing of the implant specimens tested in this study.

particular reference to the implant–abutment connection area where and displayed as two obvious downward turning points in the force–
the fracture was expected to occur. Three dimensional models of displacement curve (Fig. 4). The first turning point was at the top of
the jig (an aluminum alloy cylinder and a stainless steel block) and im- the curve, which was defined as the failure force of implant specimens.
plants were generated using the SolidWorks 2008 software (Dassault And the second one was shortly after the first one and corresponded to
Systèmes SolidWorks Corp., Massachusetts, USA), which were the abutment screw fracture.
same as described in previous sections. The models were then trans-
ferred into the ANSYS 13.0 software (ANSYS, Inc., Pennsylvania, USA) 3.2. The maximum loads
to generate 3D-FEA models, including six models of 3.5-6, 3.5-8, 3.5-
10, 4-6, 4-8, and 4-10. Ten-node tetrahedral elements were used to gen- The maximum loads for all the six groups were summarized in Fig. 5,
erate the model mesh. Since this study was designed to evaluate the ranging from 1023.9 N (in group 4-6) to 1698.5 N (in group 3.5-10). In
stress patterns, with particular reference to the connection region, general, specimens of group 4-6 provided lower maximum loads than
simplifications were made to define the contacts (implant–abutment the other 5 groups did. The value of every specimen in groups 3.5-6,
interface and implant–jig interface) as a fully bonded interface. The 3.5-8, 3.5-10, and 4-10 was larger than 1400 N, and all the five speci-
boundary condition of total fixation on the nodes of the bottom face of mens in group 3.5-8 had a value of more than 1600 N.
the jig was chosen. The material properties of the models were assumed
to be homogeneous, isotropic and linearly elastic, and the specific values 3.3. Statistical and SEM analysis
were listed in Table 1. A 200-N load, which was in the range of normal
bite forces, was applied to the hemispherical dome of each implant The mean maximum load, standard deviation, coefficient of varia-
with an inclination of 30° from the implant long axis [13]. Distribution tion, and bending moment values of all the specimens are shown in
of von-Mises stresses in the implant neck area was observed. Table 1. Implants in group 3.5-8 had the highest mean maximum load
level of 1638.9 N ± 20.3, and the lowest value of 1066.7 N ± 56.1 was
3. Results found in group 4-6. Groups 3.5-6, 3.5-10, 4-8, and 4-10 provided values
of 1475.0 N ± 24.4, 1577.1 N ± 103.2, 1241.4 N ± 6.4, and 1419.3 N ±
3.1. The failure modes & force–displacement curve 20.0, respectively.
The results of the mean bending moment listed in Table 2 were ob-
The failure mode was almost identical in all specimens, including tained by multiplying the mean maximum load and moment arm. In the
large deformations of the framework at the implant neck area, and loading set of this study, the moment arm was fixed as 5.5 mm. There-
two fracture surfaces: implant neck fracture and abutment screw frac- fore, it was easier to get the mean bending moment values. Group 4-6
ture (Fig. 3). The implant neck fracture took place in the neck region be- had the lowest mean maximum load value, which was significantly
tween the first and second threads which was flush with the surface of lower than that for the other five groups, by Tamhane's multiple com-
the aluminum cylinder. The fracture was large and accompanied by a parison test (P b 0.05). The mean maximum load differences between
strong bricking sound during the test. The abutment fracture was at group 4-8 and any other groups were statistically significant, with
the head of the screw just below the base of the abutment cone. In P b 0.05. There were no significant differences between the result of
contrast, it was accompanied with a weak breaking sound. These two 3.5-10 and that of 3.5-6, and 3.5-8. Moreover, the result of 4-10 did
fracture events were recorded as sudden drops of loads by computer, not significantly differ from that of 3.5-6 and 3.5-10. The mean
K. Wang et al. / Materials Science and Engineering C 63 (2016) 164–171 167

Fig. 3. Photograph of a tested specimen showing fractures of the implant neck and
abutment screw.

the six models. In all the models, the biggest von-Mises stress values
Fig. 2. The compressive loading set-up with an implant specimen mounted on. at the abutment connection part were remarkably lower than those at
the implant neck area, and the values at the abutment screw were less
than 50 MPa.
displacement values of the crosshead corresponding to the first load
drop in Fig. 4 recorded by the computer were shown in Table 2. As the 4. Discussion
vertical displacements were from 1.3 mm to 2.5 mm, there were only
smaller lateral displacements of the specimens. Fig. 6 is a typical SEM The experimental test is a reliable and useful method to determine
image of the fractured surface of one specimen, including the implant the mechanical integrity for implant research and development pur-
neck wall fracture and failure surface of the abutment screw. poses. It is repeatable and allows a comparative analysis of the mechan-
ical response of different designs of implants undergoing the same
loading and boundary conditions. Previous in vitro studies have evaluat-
3.4. 3D-FEA results ed effects of various parameters on implant static failure strength,

Distribution of von-Mises stresses at the implant–abutment connec-


tion area was compared for implants with different abutment taper an-
gles. For all the implants, the von-Mises stresses were concentrated at
the implant neck area around the first and second threads and the abut-
ment connection part (Fig. 7). The greatest stress values were at the lin-
gual (L) side of the implant necks, and the values at the buccal (B) side
were a little bit smaller. Fig. 8 shows the greatest stress values at the
buccal and lingual sides of the implant necks and the 4-6 model had
the biggest stress values of 322.7 MPa (B) and 368.4 MPa (L) among

Table 1
Mechanical properties of the finite element models.

Materials Young's modulus (GPa) Poisson's ratio

Ti–6Al–4V 110 0.32


Aluminum alloy 79 0.33
Stainless steel 195 0.25
Fig. 4. A representative load–displacement curve for implant specimens tested.
168 K. Wang et al. / Materials Science and Engineering C 63 (2016) 164–171

Norton used the 3-point bending test method which applied loads at an
angle of 90°. At this angle, implants were prone to be fractured with
lower load levels. Furthermore, all implant components in Norton's
study were manufactured from commercially pure titanium, which
had inferior mechanical strength to the Ti–6Al–4V alloy used in
this study. Moreover, implants tested in our study had a diameter
of 5.0 mm, which was larger than that of Astra (4.5 mm) and ITI
(4.1 mm). Norton demonstrated Astra implant systems had a better re-
sistance to 3-point bending loadings than ITI ones. However, there was
less comparability in terms of implant shapes, dimensions, and im-
plant–abutment connections of the two tested systems, which are the
factors determining implant mechanical strengths [20,21]. Therefore,
it is not possible to find how these variables affect implant fracture
strengths, much less the exact effect of connection design or taper an-
gles on the fracture resistance. The present study for the first time stud-
Fig. 5. Values of maximum loads for the six test groups. ied relationships between taper angles and implant static strength, with
good intergroup comparability by using identical implant designs,
which included implant or abutment dimensions, shapes, test protocol diameters, and materials.
variables, materials, and implant–abutment interfaces. When it came In this study, all implants were statically compressed under 30° off-
to investigating the effects of dimensions, materials and test protocol axis loadings until fracture occurred. The failure mode was almost
variables, usually the same implant system was chosen for the purpose identical in all specimens, including two fracture surfaces: implant
of comparability [14–16]. As for comparisons of failure strength of neck fracture and abutment screw fracture. And this failure mode was
implants with different implant–abutment interfaces, researchers had similar to that of ITI implants with 8° TIS-connection designs reported
to use different implant systems, which varied not only in implant– in one in vitro study [22]. The SEM analysis showed that the mode of
abutment interfaces but also in shapes, dimensions, surface characteris- fracture for abutment screws was ductile, characterized by rough and
tics, and material properties [17]. Therefore, in most of the cases, the dull surfaces with numerous large dimples (Fig. 6b). Since the circular
comparability was compromised and it was of an insufficient level to abutment screw fracture had two facets which met at an angle, their
isolate influences of implant–abutment interfaces on fracture strength junction appeared in a rung shape in the SEM images. The abutment
values precisely. For example, with the objective of comparing fracture screw was in a small diameter of 2.0 mm and it fractured in much
strength of the implant–abutment connection of six established implant lower loads than the implant wall, therefore its failure corresponded
systems, investigators tested specimens from five different manufac- to the second sharp drop in the load–displacement curve (Fig. 4). The
turers, which were different not only in implant–abutment interface implant wall fracture developed at the thread bottom of the buccal
geometries but also different in diameter and morphological designs side of the implant neck (indicated by a small black arrow) and ad-
in their neck, body and threads [18]. As reported by Möllersten, these vanced along the thread spiral line to the lingual side (indicated by a
studies could not find the exact influences of implant–abutment con- large black arrow), with the fracture direction being marked by the
nections on implant fracture strengths, but rather compare the mechan- white arrow in Fig. 6. The implant wall fracture started with a slow duc-
ical behavior of the whole systems [7]. However, in this study good tile fracture with dimples and microvoids (Fig. 6c & d) and ended with
comparability between different groups was guaranteed by custom- rapid fractures as shown by the shiny surfaces (Fig. 6e & f). These SEM
manufacturing implants with identical parameters of implant diameter, findings are consistent with the results reported by Apicella and Chan
implant shapes, material, and abutment connection depth. et al. [23,24].
In the present study, the mean values of maximum load were from The implant neck fracture took place in the neck region, and was re-
1066.7 N ± 56.1 to 1638.9 N ± 20.3. And the mean maximum bending corded as the first turning point at the top of the force–displacement
moments for the six groups ranged from 5866.6 N mm to 9014.1 N mm. curve, which was defined as the failure force. Therefore, the fracture re-
The present results of the maximum forces are in the similar ranges of sistance of the whole specimen was determined by strength in the area
previously published studies of static tests using implants with a diam- of the implant neck under present loading settings. More specifically,
eter of 5.0 mm but different implant structures. Shemtov-Yona et al. re- implant fracture strength was closely correlated to wall thickness in
ported the maximum fracture forces of their implant system in a range the implant neck area, which varied among groups with different abut-
of 1400–1700 N, with a mean of 1584 N ± 115 [15]. Steinebrunner and ment taper angles in this study (Fig. 1). This agrees with the experimen-
colleagues tested four commercial implant systems with the median tal results of one static test, which confirmed that the triangular
fracture strengths varying from 782 N to 1542 N [18]. However, our re- connection design of the NobelSpeedy Replace 3.5 mm implant system
sults compare favorably with those previously reported by Norton [19], compromised its neck wall thickness, and made it more predisposed to
who found the mean maximum bending moments of 5507 N mm fracture than the thicker continuous wall in the other systems [21]. In
(load = 220 N) and 3269 N mm (load = 131 N) which were recorded addition, it is consistent with numerical studies. van Staden and
for the Astra (11° taper angle) and ITI (8° taper angle) TIS-type implants, colleagues investigated stresses of dental implants with different wall
respectively. This is likely to be the result of three factors. One was that thicknesses using the FEA method, which demonstrated that the

Table 2
Descriptive statistics of results for the six test groups.

Specimen group 3.5-6a 3.5-8b 3.5-10a,b 4-6c 4-8d 4-10a

Mean (N) 1475.0 1638.9 1577.1 1066.7 1241.4 1419.3


SD 24.4 20.3 103.2 56.1 6.4 20.0
CV 0.0166 0.0124 0.0655 0.0526 0.0051 0.0141
MBM (N mm) 8112.3 9014.1 8674.1 5866.6 6827.5 7806.4
MDC (mm) 1.5 2.0 2.5 1.3 1.5 1.6

SD = standard deviation, CV = coefficient of variation, MBM = mean bending moments, MDC = mean displacement of crosshead. Values denoted by the same superscripts do not differ
with statistical significance.
K. Wang et al. / Materials Science and Engineering C 63 (2016) 164–171 169

Fig. 6. (a) is an overall view (×25) of the fractured surfaces of one specimen, with the white arrow showing the fracture direction of the abutment screw. (b) is a higher magnitude (×500)
of fractured area in the middle of the abutment screw surface shown in (a), with dimples being the characteristic of ductile failure. (c) shows the implant wall fracture profiles shown in
(a) and indicated by the small black arrow, with dimples and microvoids indicating a slow ductile fracture. (e) shows the end area of the implant wall fracture shown in (a) and indicated
by the large black arrow with rapid fractures as shown by the shiny surfaces. (d) and (f) are the original magnification ×500 of (c) and (e), respectively.

implant wall thickness had a significant influence on the stress magni- diameter, reducing the abutment diameter could decrease the bone
tude and distribution pattern within the implant and reduced wall stress significantly under loading, which might be due to the fact that
thickness would result in higher stress magnitudes [25]. In another a small abutment diameter provides a thicker fixture wall for transmit-
FEA study, the authors found that, for the implants with the same ting stresses than a larger abutment diameter. The higher thickness of
the fixture wall seems to provide the benefit of reducing the stress in
the implant, leading to lower stresses in the bone [26].
According to the engineering principle on fracture strength of cylin-
ders interpreted by Misch, increasing the wall thickness of two-piece
implants will significantly increase implant fracture resistance [20].
This fact is exemplified in the present study by the findings that groups
with abutments in 3.5 mm-diameter showed higher mean fracture
strengths compared with the groups with abutments in 4.0 mm-diam-
eter. Obviously, the wall thicknesses in the former groups were larger
than those of the latter ones. Furthermore, the abovementioned me-
chanical principle could be verified by comparing the mean failure
strength of groups with 4.0 mm-diameter abutments. The minimum
wall thickness in the implant neck area increased from 0.34 mm (4-6)
to 0.41 mm (4-8) and 0.48 mm (4-10), as the abutment cone taper in-
creased from 6° to 8° and 10°. The mean fracture strength resulting
from these minor changes of wall thickness increased gradually from
1066.7 N to 1241.4 N and 1419.3 N, respectively. Compared with the

Fig. 7. The top is one 3D-FEA model with enlargement of the implant deformation. The
bottom shows the distribution of von-Mises stresses (MPa) in the implant–abutment
connection area under the 200-N 30° off-axis loading. The stress concentration was
located at the implant neck region between the first and second threads from the buccal
(B) to the lingual (L) side (bottom left), and at the abutment connection part (bottom
right). Fig. 8. The greatest von-Mises stress values at the implant neck areas of the six models.
170 K. Wang et al. / Materials Science and Engineering C 63 (2016) 164–171

6° taper, 8° and 10° taper angles increased the mean maximum force by when superior implant strength is preferred to withstand higher masti-
16.4% and 33.1%, respectively. catory forces in posterior region restorations or in patients with brux-
Dental implants used in this study were not of regular hollow ism. In addition, clinicians can use the data obtained by this research
cylinder design but complicated shapes. In addition, taper enlargement to make a judgment on the mechanical qualities of implants claimed
would make wall thickness larger, and result in dimension and size by manufacturers, who might exaggerate their products' mechanical
changes of the abutment itself as well. Accordingly, the abovementioned properties for the purpose of commercial interests.
engineering principle is not necessarily applicable in all implant speci-
mens, and it should be used with caution, especially when minor chang- 5. Conclusion
es happened within a small dimension of 5.0 mm. It would be reasonable
to expect that as wall thickness became larger and larger, implant frac- This paper presents the statically tested results of the effects of
ture strength could not increase continuously without limitations. In abutment taper angles on fracture strengths of dental implants, and
fact, it would inevitably enter into a plateau, where extra wall thickness the 3D-FEA numerical results of the stress state of implant–abutment
enhancement would not strengthen implants more. Accordingly, as for connection areas as a function of different abutment taper angles.
groups with 3.5 mm-diameter abutments, when the minimum wall Within the limitations of this research, the following conclusions can
thickness in the implant neck area increased from 0.59 mm (3.5-6) to be drawn:
0.66 mm (3.5-8) resulting from taper angle enlargement from 6° to 8°,
the increasing rate of the mean fracture force was 11.1%. Then, as com- 1) The increases of taper angle could significantly increase implant frac-
pared to the 8° taper group, the mean fracture resistance of 10° taper ture resistance in most cases established in the study, which is due to
specimens did not go to a higher level but dropped slightly, indicating the increased implant wall thickness in the taper part resulting from
that a plateau has been reached. the taper angle enlargement.
In all the FEA models, a 30° off-axis load was applied to the implants,
2) The increasing effects are notable when a thin implant wall is
which produced a force moment bending implants to their lingual side.
present to accommodate wide abutments.
Therefore, the stress concentration was at the neck areas between the
3) As for implants with small diameter abutments, the increasing rate
first and second threads, and the greatest stress levels at the lingual
of implant fracture resistances is relatively low or will level off for
side were larger than those at the buccal side. These results agree with
the further increases of the taper angle.
the FEA findings reported by Sannino et al. [27], and verified by the pres-
4) The 3D-FEA stress results confirm the findings of the mechanical
ent static tests in which implants' fractures occurred at their neck areas.
tests by showing that stress values varied in implants with different
The present numerical results showed that stress values varied in im-
abutment taper angles, and the stress variation tendencies of im-
plants with different abutment taper angles. The 4 mm abutment diam-
plants with different abutment taper angles were consistent with
eter implants had higher stress values than their 3.5 mm counterparts
the variation trends of implant fracture strengths obtained from
did in the implant neck areas. For 4.0 mm abutment diameter implants,
the static tests.
stress values of 4-6 were larger than those of 4-8 and 4-10, and there
was no notable difference between the latter two models. And it was
also true of the 3.5 mm counterparts. From the mechanical perspective, Acknowledgments
the thinner implant wall thickness meant less material volume at the
implant wall area, which performed less well in dispersing stresses The authors would like to acknowledge Mr. C.Y. Tang and P. Haynes
and caused greater stress concentration [25]. Accordingly, these stress for their valued technical assistance with the specimen tests, and thank
differences could be accounted for by the variations of the implant Prof. K.F. Wang for his help in statistical data analysis.
neck wall thickness, which changed with the abutment taper angles.
Therefore, these stress results confirmed the findings of the present me- References
chanical tests, where implant specimens with a thinner implant wall
[1] B. Rangert, T. Jemt, L. Jörneus, Forces and moments on Branemark implants, Int. J.
thickness had poorer fracture strength. And moreover, the variation
Oral Maxillofac. Implants 4 (1989) 241–247.
tendencies of stress values of implants with different abutment taper [2] L. Jörnéus, T. Jemt, L. Carlsson, Loads and designs of screw joints for single crowns
angles were consistent with the variation trends of implant fracture supported by osseointegrated implants, Int. J. Oral Maxillofac. Implants 7 (1992)
353–359.
loads obtained from the static tests.
[3] G.A. Zarb, A. Schmitt, The longitudinal clinical effectiveness of osseointegrated den-
In addition to the implant fracture strength, the efficiency of TIS-type tal implants: the Toronto study, part III: problems and complications encountered, J.
connection is another important mechanical configuration for an Prosthet. Dent. 64 (1990) 185–194.
implant system, which is defined as the ratio of the loosening torque [4] T. Jemt, B. Lindén, U. Lekholm, Failures and complications in 127 consecutively
placed fixed partial prostheses supported by Brånemark implants: from prosthetic
(abutment removal torque) to the tightening torque (abutment treatment to first annual checkup, Int. J. Oral Maxillofac. Implants 7 (1992) 40–44.
screwing in torque). High efficiency means the connection performs [5] T. Jemt, W.R. Laney, D. Harris, P.J. Henry, P.H. Krogh Jr., G. Polizzi, G.A. Zarb, I.
well in preventing abutment screw loosening. As calculated by Bozkaya Herrmann, Osseointegrated implants for single tooth replacement: a 1-year report
from a multicenter prospective study, Int. J. Oral Maxillofac. Implants 6 (1991)
et al. [9], the efficiency is negatively correlated to the taper angle, 29–36.
depending on the value of the friction coefficient. Accordingly, when [6] W. Becker, B.E. Becker, Replacement of maxillary and mandibular molars with single
determining the ideal taper angle for developing a new implant system endosseous implant restorations: a retrospective study, J. Prosthet. Dent. 74 (1995)
51–55.
of the TIS-type connection with proper mechanical configurations, the [7] L. Möllersten, P. Lockowandt, L.A. Lindén, Comparison of strength and failure mode
designers should achieve a balance between the increasing effects of of seven implant systems: an in vitro test, J. Prosthet. Dent. 78 (1997) 582–591.
taper angle for implant strength and the decreasing effects of taper [8] G.E. Romanos, G.H. Nentwig, Single molar replacement with a progressive thread
design implant system: a retrospective clinical report, Int. J. Oral Maxillofac.
angle for the torque efficiency.
Implants 15 (2000) 831–836.
It is evident that the results of the present study are of significance to [9] D. Bozkaya, S. Müftü, Mechanics of the taper integrated screwed-in (TIS) abutments
researchers in implant design and manufacturing. Moreover, they are used in dental implants, J. Biomech. 38 (2005) 87–97.
[10] Y. Maeda, T. Satoh, M. Sogo, In vitro differences of stress concentrations for internal
useful for clinicians. According to Allum and co-authors [28], most of
and external hex implant–abutment connections: a short communication, J. Oral
the implant manufacturers keep the implant mechanical strength infor- Rehabil. 33 (2006) 75–78.
mation unpublished, which is largely unavailable to clinical practi- [11] P.P. Binon, Implants and components: entering the new millennium, Int. J. Oral
tioners. Therefore the results of this study break down the barriers, Maxillofac. Implants 15 (2000) 76–94.
[12] S. Dittmer, M.P. Dittmer, P. Kohorst, M. Jendras, L. Borchers, M. Stiesch, Effect of
and provide the foundation for making a choice from TIS-connection implant–abutment connection design on load bearing capacity and failure mode
implant systems with different taper angles by clinical practitioners, of implants, J. Prosthodont. 20 (2011) 510–516.
K. Wang et al. / Materials Science and Engineering C 63 (2016) 164–171 171

[13] J. Aguirrebeitia, M. Abasolo, J. Vallejo, R. Ansola, Dental implants with conical im- [21] A.Y. Imam, A. Moshaverinia, E.A. McGlumphy, Implant–abutment interface: a com-
plant–abutment interface: influence of the conical angle difference on the mechan- parison of the ultimate force to failure among narrow-diameter implant systems,
ical behavior of the implant, Int. J. Oral Maxillofac. Implants 28 (2013) e72–e82. J. Prosthet. Dent. 112 (2014) 136–142.
[14] C.K. Lee, M. Karl, J.R. Kelly, Evaluation of test protocol variables for dental implant [22] J. Perriard, W.A. Wiskott, A. Mellal, S.S. Scherrer, J. Botsis, U.C. Belser, Fatigue resis-
fatigue research, Dent. Mater. 25 (2009) 1419–1425. tance of ITI implant–abutment connectors — a comparison of the standard cone
[15] K. Shemtov-Yona, D. Rittel, L. Levin, E.E. Machtei, Effect of dental implant diameter with a novel internally keyed design, Clin. Oral Implants Res. 13 (2002) 542–549.
on fatigue performance, part I: mechanical behavior, Clin. Implant. Dent. Relat. [23] D. Apicella, M. Veltri, P. Balleri, A. Apicella, M. Ferrari, Influence of abutment material
Res. 16 (2014) 172–177. on the fracture strength and failure modes of abutment-fixture assemblies when
[16] C.F. Wang, H.L. Huang, D.J. Lin, Y.W. Shen, L.J. Fuh, J.T. Hsu, Comparisons of maxi- loaded in a bio-faithful simulation, Clin. Oral Implants Res. 22 (2011) 182–188.
mum deformation and failure forces at the implant–abutment interface of titanium [24] H.L. Chan, W.S. Oh, H.S. Ong, J.H. Fu, M. Steigmann, M. Sierraalta, H.L. Wang, Impact
implants between titanium-alloy and zirconia abutments with two levels of margin- of implantoplasty on strength of the implant–abutment complex, Int. J. Oral
al bone loss, Biomed. Eng. Online 12 (2013) 45. Maxillofac. Implants 28 (2013) 1530–1535.
[17] C.N. Elias, D.J. Fernandes, C.R. Resende, J. Roestel, Mechanical properties, surface [25] R.C. van Staden, H. Guan, Y.C. Loo, N.W. Johnson, N. Meredith, Stress evaluation of
morphology and stability of a modified commercially pure high strength titanium dental implant wall thickness using numerical techniques, Appl. Osseointegration
alloy for dental implants, Dent. Mater. 31 (2015) e1–e13. Res. 6 (2008) 39–47.
[18] L. Steinebrunner, S. Wolfart, K. Ludwig, M. Kern, Implant–abutment interface design [26] C.M. Chu, H.L. Huang, J.T. Hsu, L.J. Fuh, Influences of internal tapered abutment
affects fatigue and fracture strength of implants, Clin. Oral Implants Res. 19 (2008) designs on bone stresses around a dental implant: three-dimensional finite element
1276–1284. method with statistical evaluation, J. Periodontol. 83 (2012) 111–118.
[19] M.R. Norton, In vitro evaluation of the strength of the conical implant-to-abutment [27] G. Sannino, A. Barlattani, Mechanical evaluation of an implant–abutment self-
joint in two commercially available implant systems, J. Prosthet. Dent. 83 (2000) locking taper connection: finite element analysis and experimental tests, Int. J.
567–571. Oral Maxillofac. Implants 28 (2013) e17–e26.
[20] C.E. Misch, H.L. Wang, C.M. Misch, M. Sharawy, J. Lemons, K.W. Judy, Rationale for [28] S.R. Allum, R.A. Tomlinson, R. Joshi, The impact of loads on standard diameter, small
the application of immediate load in implant dentistry, part II, Implant. Dent. 13 diameter and mini implants: a comparative laboratory study, Clin. Oral Implants
(2004) 310–321. Res. 19 (2008) 553–559.

You might also like