You are on page 1of 2

Do not apply more than the rated voltage, as marked Rotary Function

on the Meter, between the termials or between any Switch


terminal and grounding. If the value to be measured is Position
unknown, use the maximum range and reduce the range OFF Power is turned off.
step by step until a satisfactory reading is obtained. AC/DC voltage measurement.
When measurement has been completed, disconnect : Diode test.
red black
test leads from the circuits under test, remove the testing : Continuity test.
leads away from the input terminals of the Meter and ( Figure3)
turn the Meter power off. : Resistance measurement.
The rotary switch should be placed in the right o o Temperature measurement
CF (UT202 only)
position and no any changeover of range shall be made
during measurement to prevent damage of the Meter. Warning
To avoid electric shock, do not carry out the AC current measurement range from 0.001A
A To avoid harms to you or damages to the Meter from
to 400.0A
measurement when the Meter,s back case and battery eletric shock, do not attempt to measure voltages higher
compartment are not closed. Functional Buttons than 600V AC/DC, although readings may be obtained.
Do not input higher than 600V between the Meter,s Below table indicated for information about the functional The DC Voltage ranges are:
terminals and the grounding to avoid electric shock button operations.
200.0mV, 2.000V, 20.00V, 200.0V and 600V.
and damages to the Meter.
Press HOLD to enter and exit the
 To measure DC voltage, connect the Meter as follows:
When the Meter working at an effective voltage over HOLD
Hold mode in any mode, the Meter beeps. 1. Insert the red test lead into the V terminal and
60V in DC or 30V rms in AC, special care should be
 Press and hold HOLD button the black test lead into the COM terminal.
taken for there is danger of electric shock.
while turning on the Meter to display 2. Set the rotary switch to V .
Use the proper terminals, function, and range for full icons. 3. Connect the test leads across with the object being measured.
your measurements.
MAX Press MAX to start recording and updating The measured value shows on the display.
Do not use or store the Meter in an environment of of maximum values.
high temperature, humidity, explosive, inflammable and Note
SELECT Press SELECT button to switch between In each range, the Meter has an input impedance of 10 KΩ
strong magnetic field. The performance of the Meter and C F .
o o

may deleriorate after dampened. This loading effect can cause measurement errors in high
Overview When using the test leads, keep your fingers behind The Effectiveness of Functional Buttons impedance circuits. If the circuit impedance is less than or
This Operating Manual covers information on safety and the finger guards. equal to 10kΩ, the error is negligible (0.1 or less).
Not every functional buttons can be use on every rotary
cautions. Please read the relevant information carefully and Disconnect circuit power and discharge all high When DC voltage measurement has been completed,
switch positions.Below two tables describe on which rotary
observe all the Warnings and Notes strictly. behind the finger guards. disconnect the connection between the testing leads and
position functional buttons can be valid accordingly.
Disconnect circuit power and discharge all high Model: UT201 the circuit under test and remove testing leads from the
Warning
To avoid electric shock or personal injury , read the -voltage capacitors before testing resistance, continuity input terminals.
Rotary Switch
“Safety Information” and carefully before using the Meter. and diode. Positions Functional Buttons B. Measuring AC Voltage (See Figure 4)
Replace the battery as soon as the battery indicator
Model UT201/202(hereafter referred to as "the Meter") is SELECT MAX HOLD
appears. Whith a low battery, the Meter might produce
2000-count stable, safe and reliable digital clamp multimeter. V N/A
false readings that can lead to electric shock and
It is designed with large-scale integrated circuits and A/D
personal injury V N/A
converter as the core as well as the overload protection and
When servicing the Meter, use the replacement parts N/A N/A
novel structure, which make it a superb tool for electricians.
with the same model or identical electrical specifications. N/A
The Meter can measure AC/DC voltage, AC current,
To avoid any damage to the meter or any accident, red black
resistance, temperature(℉/℃), diode, continuity and so on. A 2/20A N/A
do not alter the internal circuit of the Meter randomly.
Unpacking Inspection Soft cloth and mild detergent should be used to clean A 200/400A N/A
the surface of the Meter when servicing. No abrasive Model: UT202
Open the package case and take out the Meter. Check
and solvent should be used to prevent the surface of Rotary Switch
the following items carefully for any missing or damaged patr: ( Figure 4)
the Meter from corrosion, damage and accident. Positions Functional Buttons Warning
Item Description Qty The Meter is suitable for indoor use. SELECT MAX HOLD To avoid harms to you or damages to the Meter from
1 English Operating Manual 1 pc Turn the Meter off when it is not in use and take out eletric shock, do not attempt to measure voltages higher
V N/A
2 Test Lead 1 pair the battery when not using for a long time. than 600V AC/DC, although readings may be obtained.
3 Point Contact Temperature Probe 1 pc Constantly check the battery as it may leak when it V N/A
The AC Voltage ranges are:
(UT202 only) has been used for some time, replace the battery as N/A 2.000V, 20.00V, 200.0V and 600V.
4 1.5V Battery (AAA) 2 pcs soon as leaking appears. A leaking battery will damage F C(K-Type)
o o
N/A To measure AC voltage, connect the Meter as follows:
the Meter. A 2/20A N/A 1. Insert the red test lead into the V terminal and
In the event you find any missing or damaged part, please
Probe assemblies to be used for MAINS measurements the black test lead into the COM terminal.
contact your dealer immediately. A 200/400A N/A
shall be RATED as appropriate for MEASUREMENT 2. Set the rotary switch to V .
Safety Information CATEGORY III according to IEC 61010-031 and shall have Display Symbols (See Figure 2) 3. Connect the test leads across with the object being measured.
This Meter complies with the standard EN 61010-1, a voltage RATING of at least the voltage of the circuit to The measured value shows on the display.
61010-2-032,61010-2-033,Pollution Degree 2, Measurement be measured.
Note
category. (CAT II 600V,CAT III 300V) and Double Insulation.
International Electrical Symbols In each range, the Meter has and input impedance
ETL/ cETL: Conforms to UL STD 61010-1, 61010-2-032
of 10 MΩ. This loading effect can cause measurement
and 61010-2-033 AC (Alternating Current).
errors in high impedance circuits. If the circuit impedance
Certified to CSA STD C22.2 NO.61010-1, IEC STD DC (Direct Current). is less than or equal to 10 KΩ, the error is negligible
61010-2-032, IEC STD 61010-2-033 AC or DC. (0.1 or less).
No. Symbol Description
CAT II (MEASUREMENT CATEGORY II): Applicable to Grounding. When AC voltage measurement has been completed,
test and measuring circuits connected directly to utilization 1 AC Indicator for AC voltage or current disconnect the connection between the testing leads
points (socket outlets and similar points) of the low-voltage Double Insulated. 2 DC Indicator for DC voltage and the circuit under test and remove testing leads
MAINS installation. Deficiency of Built-In Battery The battery is low. from the input terminals.
Examples are measurements on MAINS CIRCUITS of Continuity Test. Warning: To avoid false readings, (See Figure 5)
household appliances, portable tools and similar equipment. which could lead to possible electric
Diode. 3
CAT III (MEASUREMENT CATEGORY): Applicable to shock or personal injury, replace
test and measuring circuits connected to the distribution Capacitance Test the battery as soon as the battery
part of the building’s low-voltage MAINS installation. Fuse. indicator appears.
Examples are measurements on distribution boards The Meter is in the auto range mode
Warning. Refer to the Operating Manual.
(including secondary electricity meters), circuitbreakers, 4 in which the Meter automatically selects
wiring, including cables, bus-bars, junction boxes, switches, Application around and removal from the range with the best resolution.
UNINSULATED HAZARDOUS LIVE 5 Test of diode.
socket-outlets in the fixed installation, and equipment for
conductors is permitted.
industrial use and some other equipment such as stationary 6 The continuity buzzer is on.
motors with permanent connection to the fixed installation. Conforms to Standards of European Union. 7 MAX Maximum reading displayed
Use the Meter only as specified in this operating manual, 8 Date hold is active.
This symbol signify the product comply
otherwise the protection provided by the Meter may be impaired. 9 o
CF
o
The unit of temperature:
with both USA and Canada requirement. o
In this manual, a Warning identifies conditions and actions C: Centigrade temperature To avoid harms to you, do not attempt to input voltages
4007682

o
that pose hazards to the user, or may damage the Meter or The Meter Structure (See Figure 1) F: Fahrenheit temperature higher than 60V DC or 30V rms AC.
the equipment under test. : Ohm. The unit of resistance. To avoid damages to the Meter or to the devices under
A Note identifies the information that user should pay 1. Input Terminals 10 ,k ,M k : kilohm.1 x 103 or 1000 ohms. test, disconnect circuit power and discharge all the
attention to. 2. LCD Display M :Megaohm. 1 x 106 or high-voltage capacitors before measuring resistance.
3. Functional Buttons 1,000,000 ohms.
The resistance ranges are:
4. Rotary Switch 11 A Amperes (amps). The unit of current.
To avoid pssible electric shock or personal injury, 200.0Ω,2.000KΩ,20.00KΩ,200KΩ,2.000MΩ and 20.00MΩ
5. Trigger: Push/release it 12 mV, V Volts. The unit of voltage.mV: Millivolt.
andto avoid possible damage to the Meter or to the To measure resistance,connect the Meter as follows:
to open/close the clamp jaw. 1x10-3 or 0.001 volts
equipment under test, adhere to the following rules: 1. Insert the red test lead into the terminal and
6. Hand Guards: to protect user’s Indicates negative reading
13 the black test lead into the COM terminal.
Before using the Meter inspect the case. Do not use hand from touching the
14 OL The input value is too large for the 2. Set the rotary switch to it, defaults at resistae(Ω)
dangerous area. ( Figure 1)
selected range
the Meter if it is damaged or the case (or part of the case) -nce mod or press SELECT button to select Ω mode.
7. Transformer Jaws: designed to pick up the AC current
is removed. Look for missing plastic. Pay attention to 3. Connect the test leads across with the object being measured.
the insulation around the connector.
flowing through the conductor. It could transfer current Measurement Operation
The measured value shows on the display.
to voltage. A. Measuring DC Voltage (See Figure 3)
Inspect the test leads for damaged insulation or
exposed metal. Check the test leads for continuity. Rotary Switch
The test leads can add 0.1Ω to 0.3Ω of error to resistance
Replace damaged test leads with identical model number Below tabel indicated for information about the rotary measurement.
or electrical specifications before using the Meter. switch positions. For high-resistance measurement(>1MΩ), it is normal
taking several seconds to obtain a stable reading.
REV.1
DATE:2015-12-04
P/N:110401105587X

If Ω reading with shorted test leads is not≤0.5Ω, check the connection between the testing leads and the circuit under
test and remove testing leads from the input terminals. Opening
for loose test leads, wrong function selected, or enabled
Tested
data hold function. F. Measuring Temperature
The LCD displays OL indicating open-circuit or the teste (See Figure 8)
resistor value is higher than the maximum range of the Meter
Resistance measurement is default to auto range mode.
Measuring the tested object that is already removed sinewave RMS (AVG response)
from the in-line circuit can help to obtain a more accu-rate RMS value
reading.
When resistance measurement has been completed, Maintenance
disconnect the connection between the testing leads and 2. This section provides basic maintenance information including
the circuit under test and remove testing leads from the battery replacement instruction.
input terminals.
D.Testing Diodes (See Figure 6) Warning
Do not attempt to repair or service your Meter unless
you are qualified to do so and have the relevant calibration,
performance test, and service information. To avoid
H electrical shock or damage to the Meter,do not get water
nside the case.
A. General Service
 Periodically wipe the case with a damp cloth and mild
red black
detergent. Do not use abrasives or solvents.
 To clean the terminals with cotton bar with detergent,
3. as dirt or moisture in the terminals can affect readings.
( Figure 6)  Turn the Meter power off when it is not in use.
 Take out the battery when it is not using for a long time.
Warning  Do not use or store the Meter in a place of humidity,
To avoid damages to the Meter or to the devices under high temperature, explosive, inflammable and strong
test, disconnect circuit power and discharge all the magnetic field.
high-voltage capacitors before testing diodes.
B. Replacing the Battery (See Figure 10)
Use the diode test to check diodes, transistors, and other
semiconductor devices. The diode test sends a current
through the semicondutor junction, then measure the
voltage drop across the junction. A good silicon junction G. Measuring AC Current (See Figure 9) Screw
drops between 0.5V and 0.8V.
To test the diode out of a circuit, connect the Meter as
follows:
1. Insert the red test lead into the V terminal and Display sinewave RMS( AVG response)
the black test lead into the COM terminal.
2. Set the rotary switch to and press SELECT button
to select measurement mode. ( Figure10)

3. For forward voltage drop readings on any semiconductor F


component, place the red test lead on the component,s
anode and the black test lead on the component,s cathode. Warning
Note To avoid false readings, which could lead to possible
In a circuit, a good diode should still produce a forward electric shock or personal injury, replace the battery
voltage drop reading of 0.5V to 0.8; however, the reverse as soon as the battery indicator “ ” appears.
voltage drop reading can vary depending on the resistance Make sure the transformer jaw and the tets leads are
of other pathways between the probe tips. disconected from the circuit being tested before
Connect the test leads to the proper terminals as said opening the case bottom.
above to avoid error display. To replace the battery:.
The LCD will display OL indicating either open circuit or 1.Turn the Meter off and remove all the connections
wrong polarity connection. from the input terminals
The unit of diode is volt(V), displaying the forward voltage 2. Turn the Meter’s case top down.
3. Remove the screw from the battery compartment,
drop readings.
and separate the battery compartment from the case
Measuring the tested object that is already removed from then bottom.
the in-line circuit can help to obtain a more accutate reading. release the Meter slowly until the transformer jaw is completely 4. Remove the old battery from the battery compartment.
When diode testing has been completed, disconnect the closed, Make sure the conductor to be tested is placed at
5. Replace the battery with 2pcs of new 1.5V (AAA)
connection between the testing leads and the circuit under the center of the transfor- mer jaw, otherwise it will casue
battery.
test and remove testing leads from the input terminals. deviation.
6. Rejoin the case bottom and the battery compartment,
(See Figure 7) 4. The measured value shows on the display,it is a effective
and reinstall the screw.
value of sine wave (mean value response).
Note
To obtain accurate reading, measure only one conductor
at each time. ** END **
When current measurement has been completed, disconnect
the connection between the conductor under test and the This operating manual is subject to change without notice.
jaw, and remove the conductor away from the transformer
jaw of the Meter.
Sleep Mode
To preserve battery life, the Meter automatically turns off
if you do not turn the ratary switch or press any button for
around 15 minutes.
To avoid damages to the Meter or to the devices under
The Meter can be activated by turning the rotary switch
test, disconnect circuit power and dicharge all the
or pressing any button with the following conditions:
high-voltage capacitors before measuring continuity.
1) When the Meter enters Sleep Mode at temperature
To test for continuity,connect the Meter as follows:
functions of Model: UT202, the Meter cannot be activated
1. Insert the red lead into the terminal and the
by turning the rotary switch to AC current ranges.
black test lead into the COM terminal.
2) Pressing any button must be according to“ The effectiveness
2. Set the rotary switch to and press SELECT
of Functional Buttons” section.
button to select measurement mode.
3) The Hold function will be cancelled if the Meter is activated
3. The buzzer sounds if the resistance of a circuit under
by pressing the HOLD button.
test is less than 50 Ω.
To disable the Sleep Mode function, Press and hold
4. The buzzer may or may not sounds if the resistance of
HOLD button while turning on the Meter.
a circuit under test is between 50 Ω to 120 Ω.
5. The buzzer does not sound if the resistance of a circuit Specifications
under test is higher than 120Ω. 1. General Specifications:
Note Maximum voltage including transient overvoltage between
any terminals and grounding:500V rms
The buzzer beeps once when pressing any buttons at any
Display: Maximum display 1999
rotary switch positions except at 2/20A positions if the button
Auto Polarity Display
is valid. If the button is not valid, it does not beep. At 2/20A
Overloading : Display OL or - OL
rotary switch position, the buzzer is set not to beep.
Low Battery Indication : Display
The buzzer beep 5 times continuously on around 1 minute
Measurement Speed : Updates 3 times/second
before entering the sleep mode. When it is just before entering
Measuremnet Deviation: When the conductor being meaured
the sleep mode, it will have one long beep to warn you.
is not placed in a correct position
The LCD displays OL indicating the circuit being tested
during AC current measurement,
is open.
current measurement, it will cause
When continuity testing has been completed, disconnect
±1% reading deviation.

You might also like