You are on page 1of 2

position and no any changeover of range shall be 1.

Insert the red test lead into


UT206A: Switches between AC
made when measurement is conducted to prevent voltage and DC Voltage measure the V terminal and the black
damage of the Meter. -ment, and continuity and diode test lead into the COM
When the Meter is working at an effective voltage measurement; the Meter beeps.
UT205A: Switches between continuity
terminal.
over 60V in DC or 42V rms in AC, special care and diode measurement; the Meter 2. Set the rotary switch to
should be taken for there is danger of electric shock. SELECT beeps. DC voltage measurement
Use the proper terminals, function, and range for When it is under sleep mode, press mode.
your measurements. to activate the Meter at the effective
measurement range,e.g. continuity 3. Connect the test leads Red Black

When using the test leads, keep your fingers behind buzzer. Auto power off feature will be across with the object
the finger guards. disabled until Switching off and turning being measured.
on the Meter again.
Disconnect circuit power and discharge all high- The measured value shows Figure 4
voltage capacitors before testing. Press HOLD to enter and exit the on the display.
To avoid harm to you or damage to the Meter from HOLD Hold mode in any mode, the Meter Note
electric shock, please do not attempt to apply higher beeps. In each range, the Meter has an input impedance of
than 600V between the terminals and grounding. 10M .This loading effect can cause measurement
Replace the battery as soon as the battery indicator Display Symbols (See Figure 2) errors in high impedance circuits. If the circuit imped
appears. With a low battery, the Meter might -ance is less than or equal to 10k , the error is
produce false readings that can lead to electric negligible (0.1% or less).
shock and personal injury. When DC voltage measurement has been completed,
When servicing the Meter, use only the replacement disconnect the connection between the testing leads
parts with the same model or identical electrical and the circuit under test, and remove the testing leads
specifications. away from the input terminals of the Meter.
Overview The internal circuit of the Meter shall not be altered C. Measuring AC Voltage (See Figure 4)
This Operating Manual covers information on safety and at will to avoid damage of the Meter and any accident. AC Voltage measurement is an auto ranging measure
cautions. Please read the relevant information carefully Soft cloth and mild detergent should be used to -ment mode.
and observe all the Warnings and Notes strictly. clean the surface of the Meter when servicing. No Note:
Warning abrasive and solvent should be used to prevent Follow the same procedure as the DC Voltage Meas
the surface of the Meter from corrosion, damage Figure 2
To avoid electric shock or personal injury, read the -urement mode. When you using UT206A to carry
“Safety Information”carefully before using the Meter. and accident. out the measurement, use SELECT button to switch
Turn the Meter off when it is not in use and take Number Symbol Description between ACV and DCV.
Model UT205A and UT206A are AC clamp multimeters
out the battery when not using for a long time. 1 Data hold is active.
(hereafter referred to as "the Meter") with remarkable D. Measuring Resistance (See Figure 5)
features: versatile functions, stable operation, low power Constantly check the battery as it may leak when 2 AC Indicator for AC Voltage.
consumption and ergonomic structural design. They it has been using for some time,replace the battery The battery is low. Warning
can measure AC/DC voltage, AC current, resistance, as soon as leaking appears. A leaking battery will 3
Warning: To avoid false readings, which
To avoid damages to the Meter or to the devices
damage the Meter. could lead to possible electric shock or
diode, continuity, frequency and temperature(UT206A),etc. personal injury, replace the battery as soon under test, disconnect circuit power and discharge
Do not use or store the Meter in an environment as the battery indicator appears. all the high-voltage capacitors before measuring
Except where noted, the descriptions and instructions in of high temperat ure, humidity,explosive, inflamm
4 % Percent: Used for duty cycle measurements. resistance.
this Operating Manual apply to both the Model UT205A -able and strong magnetic field. The performance
5 A Indicator for AC current To measure resistance, connect the Meter as follows:
/UT206A of the Meter may deteriorate after dampened
Hertz. The unit of frequency. 1. Insert the red test lead into
International Electrical Symbols Hz, kHz, Kilohertz. 1x 10 3 or 1000 hertz. the terminal andthe
Unpacking Inspection 6
MHz 6
Megahertz. 1x10 or 1,000,000 hertz. black test lead into the
Open the package case and take out the Meter. Check the AC (Alternating Current) COM terminal.
7 ,k , Ohm. The unit of resistance.
following items carefully for any missing or damaged part: 2. Set the rotary switch to .
DC (Direct Current) kilohm. 1 x 10 3 or 1000 ohms.
M 3. Connect the test leads
Item Description Qty Megaohm. 1x10 6 or 1,000,000 ohms.
AC or DC across with the object
8 The continuity buzzer is on.
1 Operating Manual 1 pc Grounding being measured. Red Black

9 Test of diode The measured value shows


2 Test Leads 1 pair Double Insulated
10
o
C Centigrade. The unit of temperature. on the display.
3 Model UT206A: Point Contact T Fuse Note Figure 5
emperature Probe 1 pc
11 The relative value mode is on to display
The test leads can add 0.1 to 0.3 of error to
4 9V Battery (NEDA 1604, 6F22 or 0006P) 1 pc Low Battery Indication the stored value minus the present
resistance measurement.To obtain precision readings
(installed inside the Meter) Continuity Test Volts. The unit of voltage.
12 V, mV -3
Millivolt. 1 x 10 or 0.001 volts.
in low-resistance measurement under the range of
5 Vinyl Bag 1 pc Diode 400.0 , short-circuit the input terminals beforehand,
using the relative measurement function button REL
In the event you find any missing or damaged part, please Conforms to Standards of European Union Measurement Operation to automatically subtract the value measured
contact your dealer immediately. Warning. Refer to the Operating Manual Make sure the Sleep Mode is not on if you found there
when the testing leads are short-circuited from the
is no display on the LCD after turning on the Meter.
Risk of Electric Shock reading.
Safety Information Make sure the Low Battery Display is not on;
This Meter complies with the standard IEC61010: Pollution otherwise false readings may be provided. For high-resistance measurement (>1M ), it is
Degree 2, Overvoltage Category (CAT. II 600V / CAT III
The Meter Structure (See Figure 1) Pay extra attention to the symbol which is located normal to take several seconds to obtain a stable
300V) and Double Insulation. 1. Transformer Jaw designed to pick besides the input terminals of the Meter before carrying reading.
up the AC current flowing through out measeurement. When resistance measurement has been completed,
CAT II: Local level, appliance, PORTABLE EQUIPMENT 1
the conductor. Make sure you are in the proper measurement range disconnect the connection between the testing leads
etc., with smaller transient voltage overvoltages than CAT. III
2. Rotary Switch. before carrying out measurement. and the circuit under test, and remove the testing leads
CAT. III: Distribution level, fixed installation, with smaller
3. LCD Display. 2 away from the input terminals of the Meter.
transient overvoltages than CAT. IV.
Under the influence of Radiated, Radio-Frequency Electro 4. Function Buttons 3 A. Measuring AC Current (See Figure 3) E. Testing for Diodes and Continuity
-magnetic Field phenomenon,the captioned model may 5. Input Terminals Warning (See Figure 6)
6 4
malfunction and can self-recover after the test. 6. Lever. Press the lever to open The measurement value obtained is wrong if testing
Use the Meter only as specified in this operating manual, the transformer jaws. When the 5 two or more current conductors at the same time. Warning
otherwise the protection provided by the Meter may be pressure on the lever is released, To avoid damages to the Meter or to the devices
impaired. the jaws will close. under test, disconnect circuit power and discharge
Figure 1
In this manual, a Warning identifies conditions and acti all the high-voltage capacitors before testing for
-ons that pose hazards to the user, or may damage the Functional Buttons continuity.
Meter or the equipment under test. Below table indicated for information about the functional To test for diodes and continuity, connect the Meter as
A Note identifies the information that user should pay button operations. below:
attention to. 1. Insert the red test lead into
Button Description the terminal and the
Warning black test lead into the
Turn the display backlight on and off.In
To avoid possible electric shock or personal injury, order to avoid the hazard arising from COM terminal.
and to avoid possible damageto the Meter or to the mistaken readings in insufficient light or
poor vision, please use Display Backlight 2. For diode test
equipment under test, adhere to the following rules: Figure 3
function. a. Set the rotary switch to
Check for the lever is in good condition when Red Black

1. At ACV or DCV range .


measuring AC current. measure
Must center the wire (conductor) within the trans
Press the button to switch into Hz or b. For forward voltage drop
duty cycle if under ACV or DCV range. readings on any semicon
-former jaw. The Meters enters the manual ranging Figure 6
Before using the Meter inspect the case. Do not measurement mode after pressing the RMS value -ductor component, place the
Hz% (UT205A
use the Meter if it is damaged or the case (or part only) button at the ACV or DCV measurement red test lead on the component’s anode and place
of the case) is removed. Look for cracks or missing
range. Turn the rotary switch or switch the black test lead on the component’s cathode.
off the Meter to return to auto ranging
plastic. Pay attention to the insulation around the measurement mode.The Meter is at the
The measured value shows on the display.
connectors. DCV – 400mV range or ACV-4V range 3. For continuity test:
Inspect the test leads for damaged insulation or after returning from the Hz% cycle a. Press SELECT to switch between Diode and Contin
measurement mode. B. Measuring DC Voltage (See Figure 4)
exposed metal. Check the test leads for continuity. 2. At Hz range -uity test.
Replace damaged test leads with identical model Press the button to step through Warning b. Connect the test lead to the two end of the circuit
number or electrical specifications before using frequency and duty cycle measurem To avoid harm to you or damage to the Meter from under test.
-ent mode.
the Meter. electric shock, please do not attempt to measure c. The buzzer sounds if the resistance of a circuit under
Do not apply more than the rated voltage, as marked Press REL to enter and exit the REL voltages higher than 600V although readings may test is less than 100 .
on the Meter, between the terminals or between REL (UT206A mode in any measuring mode except in be obtained.
only) Note
any terminal and grounding. frequency and duty cycle,the Meter beeps.
To measure DC voltage, connect the Meter as follows:
The rotary switch should be placed in the right In a circuit, a good diode should still produce a
P/N: 110401104276X
MAY.2018 REV. 6

forward voltage drop reading of 0.5V to 0.8V; however, To enter or exit REL mode: C. AC Voltage
the reverse-voltage drop reading can vary depending Use rotary switch to select the measurement function
on the resistance of other pathways between the probe Overload
before selecting REL . If measurement function Range Resolution Accuracy Remarks
tips. Protection
changes manually after REL is selected, the Meter
Connect the test leads to the proper terminals as said exits the REL mode. 4V 1mV Input impedance around 10M .
above to avoid error display.The LCD will display OL Press REL to enter REL mode, auto ranging Displays RMS value of sine
40V 10mV (1.2%+5) 600V DC
indicating diode being tested is open or polarity is turns off, and the present measurement range is wave (mean value response).
reversed. The unit of diode is Volt (V), displaying the 400V 100mV 600V AC Frequency response:
locked and stored as the stored value. All the measure
forward voltage drop readings. -ments done after will be automatically subtract this 40Hz ~ 400Hz.
600V 1V (1.5%+5)
When diode and continuity testing has been completed, stored value.
disconnect the connection between the testing leads Press REL again to exit REL mode and return D. Resistance
and the circuit under test, and remove the testing leads to common measurement mode.If you want to enter
away from the input terminals of the Meter. the auto ranging measurement mode or other ranges, Range Resolution Accuracy Overload Protection
F. Measuring Frequency and Duty Cycle please turn the rotary switch or power off the Meter
400 0.1 (1.2%+2)
and start again.
(See Figure 7) 500V DC or AC
UT205A is designed with frequency and duty cycle The SELECT Button 4k 1 (1%+2)
functions, but only frequency for UT206A. It uses for selecting the required measurement function 40k 10
To measure frequency and duty cycle, connect the when there is more than one function at one position (1%+2)
of the rotary switch. 400k 100 500V DC or AC
Meter as follows:
1. UT205A Turning on the Display Backlight 4M 1k (1.2%+2)
a. Insert the red test lead 40M 10k (1.5%+2)
Warning
into the Hz% terminal
and the black test lead In order to avoid the hazard arising from mistaken
readings in insufficient light or poor vision, please E. Diode and Continuity Test
into the COM terminal. Hz%
use Display Backlight function.
b. Set the rotary switch to Hz ; Function Range Resolution Overload Remarks
press Hz% button to switch Red Black Press and hold button for over 2 seconds to turn
between Hz and Duty cycle the Display Backlight on. Displays the nearest value
Diode 1mV
measurement mode. Press button again to turn the Display Backlight of forward voltage drop
c. Or set the rotary to ACV or Figure 7 off, otherwise it will stay on continuously. 500V DC or AC
DCV range, then press Hz% button to switch into Hz Buzzer beeps when
Sleep Mode Continuity 0.1 100
and Duty cycle measurement mode. But the accuracy
will have a little difference based on the input frequency To preserve battery life, the Meter automatically turns
off if you do not turn the rotary switch or press any button
and wave. F. Frequency (auto-ranging)
d. Connect the test leads across with the object being for around 15 minutes.
measured.The measured frequency value or duty The Meter can be activated by turning the rotary switch Range Resolution Accuracy Overload Protection Remarks
cycle value shows on the display. or pressing the effective button, it will return to Amplitude a:
working mode. 10Hz- Min (0.1%+3) 500V DC or VAC 1MHz: 300mV rms a 30V rms;
2. UT206A
10MHz 0.001Hz 1MHz: 600mV rms a 5V rms
a. Insert the red test lead into the Hz terminal and the
black test lead into the COM terminal.
General Specifications
b. Set the rotary switch to Hz. Maximum Voltage between G. Duty Cycle (Model UT205A only)
c. Connect the test leads across with the object being any Terminals and Grounding: 600V rms or 600V DC.
Range Resolution Accuracy Overload Protection
measured. The measured frequency value shows on Maximum Current Measurement: 1000A.
the display. of Transformer Jaw: 0.1%-99.9% 0.1% For reference only 500V DC or AC
Maximum Jaw Opening: 40mm.
Note
Maximum Display: Digital: 3999 H. Temperature (Model UT206A only)
When frequency and duty cycle measurement has
been completed, disconnect the connection between Overload Display: OL Range Resolution Accuracy Overload Protection
the testing leads and the circuit under test, and remove Range: Auto
o o

the testing leads away from the input terminals of the Polarity Display: Negative display: “-“ -40 C~0 C (4%+4)
Meter. Measurement Speed: Updates 3 times/second. o
1 C~400 C
o
1C
o
(2%+8) 500V DC or AC
o o o
Temperature: Operating: 0 C to +40 C (32 F to+104 F).
o

G. Measuring Temperature (3%+10)


o o
o o o o
Storage: -10 C to +50 C (14 F to +122 F). 401 C~1000 C
(UT206A Only, See Figure 8) Relative Humidity: 75% @ 0 C to 40 C ;
o o

To measure temperature, connect the Meter as follows:


70% @ -10 to 50 C .
o Maintenance
1. Insert the red temperature This section provides basic maintenance information including battery replacement instruction.
Altitude: Operating: 2000 m.
o
probe into the C terminal
and the black temperature Storage: 10000 m. Warning
probe into the COM terminal. Do not attempt to repair or service your Meter unless you are qualified to do so and have the relevant calibr
Battery Type: One piece of 9V (NEDA1604 or 6F22
o
2. Set the rotary switch to C. -ation, performance test, and service information.To avoid electrical shock or damage to the Meter, do you get
or 006P).
3. Place the temperature water inside the case.
Low Battery Indication:Display
probe to the object being A. General Service
measured. Red Black Dimensions (HxWxL): 236mm x 97mm x 40mm
The measured value shows on Weight: Approximate 350g (battery included). Periodically wipe the case with a damp cloth and mild detergent. Do not use abrasives or solvents.
the display. Safety/Compliances: IEC61010 CAT. II 600V / CAT III To clean the terminals with cotton bar with detergent, as dirt or moisture in the terminals can affect readings.
.
Note 300V and Double Insulation. Turn the Meter to OFF position when it is not in use.
Figure 8
When there is no temperature probe Take out the battery when it is not using for a long time.
inserted into the terminals, the LCD displays OL. Certifications: Do not use or store the Meter in a place of humidity, high temperature, explosive, inflammable and strong magnetic field.
The included point contact temperature probe can only
o
be used up to 250 C.For any measurement higher than
Accuracy Specifications B. Replacing the Battery (See Figure 9)
that, it is necessary to use another type of temperature Accuracy: (a% reading + b digits), guarantee for 1 year. Warning
o o
probe. Operating temperature: 23 C 5 C .
Relative humidity: 75%. To avoid false readings, which could lead to possible electric shock or personal injury,replace the battery as
When temperature measurement has been completed,
soon as the battery indicator “ ” appears.
disconnect the connection between the temperature
probe and measured object, and remove the tempera
A. AC Current Make sure the transformer jaw and the test leads are disconnected from
-ture probe from the input terminals of the Meter. Range Resolution Accuracy Remarks the circuit being tested before opening the case bottom. Screw

Make sure the test leads are removed from the input terminals.
Frequency resp
Operation of Hold Mode 400A 0.1A (1.5%+5) -onse 50Hz~ To replace the battery:
800A (2%+5) 60Hz.Display
Warning RMS value of
1. Turn the rotary switch of the Meter to OFF position and remove all the
To avoid possibility of electric shock, do not use 1000A 1A sine wave(mean connections from the terminals.
>800A (3%+5) value response). 2. Remove the screw from the battery compartment, and separate the battery
Hold mode to determine if circuits are without power. Figure 9
The Hold mode will not capture unstable or noisy compartment from the case bottom.
readings. B. DC Voltage 3. Remove the battery from the battery compartment.
4. Replace the battery with a new 9V battery (NEDA1604, 6F22 or 006P) .
The Hold mode is applicable to all measurement Range Resolution Accuracy Overload Protection Remarks
functions. 5. Rejoin the case bottom and battery compartment, and reinstall the screw.
400mV 100 V (0.8%+3)
Press HOLD to enter Hold mode; the Meter beeps.
Press HOLD again or turn the rotary switch to exit Hold 4V 1mV ** END **
mode; the Meter beeps. 600V DC Input impedance:
40V 10mV (0.8%+1) 600V AC
In Hold mode, is displayed. around 10M .
400V 100mV
The Use of Relative Value Mode (UT206A only)
The REL mode applies to all measurement functions 600V 1V (1%+3)
except frequency and duty cycle. It subtracts a stored
value from the present measurement value and displays
the result.
For instance, if the stored value is 20.0V and the
present measurement value is 22.0V, the reading would
be 2.0V. If a new measurement value is equal to the
stored value then display 0.0V.