LOGIN
Firstly, switch on the main power of tester then switch on the power of the computer.
After logging into Windows, select the Testing software display from the desktop.(If
UTM is equipped with UPS system, please also power it on)

The “User Login” window shows as below:

Click on the down arrow to view all user names. After selecting User, enter login password,
and press ‘enter’. Once password entered, the screen will display the main screen.

2
Universal Testing Machine ManualV2011FEB

Main Screen

Tool Bar Figure display

Data display Curve display Control


Button

Data item Display

3
Universal Testing Machine ManualV2011FEB

Tool Bars

NEW: Open a new test file


Editing the “fixed
Read: Access previous saved testing test file form”, please see” report
form” for reference
Save: Save current test result

Read setting: Access setting files, it would should the last using record file name. If opening other setting file,
the window title would also change.
Setting: Going to the setting function

Data display area Test Serial Number


Curve

Mark Product Name

Click left button of the mouse at


the this area will show up “file
tool list”

4
Universal Testing Machine ManualV2011FEB

Figures、Curve、Data display area

Test Figures display area -it could show 10 figures and divided into 2 pages, one page shows 5 testing
figures. Choose Page1 or two to see different figures

If select Cursor”Cursor figure will show on


below chart

Select to ener the changing Axis Graphical display when the various forces,Power
display range default to automatic, the power will
automatically adjust in accordance with the actual
test upper and lower limits. To modify the displayed
Y axis scale, please press the Up button in the
Y-axis or Dn "change window: Auto to automatically
adjust the To modify the display X axis scaleLeft in
the X-axis press the key or Right "change window:
Auto For the automatic adjustment
Change the setting to leave after Keyi
figures
Decimal point shows the median curve zoom, users
can view the demand, change the decimal display
digits
5
Click the arrow, there speed setting.Testing Machine ManualV2011FEB
Universal
Manually change the speed of rise and Manual testing immediately
fall of the control region。 and take some。
Control area

To zero before starting


Click on the arrow
the test, press the stop
to adjust, move up
button to terminate the
and down to do the
test
speed change; or
direct input speed,
Print Form click the Set button
This version does not open to confirm.

Here the "Preview" for the


fixed print preview report

Can use "Excel Report”


button to test data
Can be decreased by using up button
output to Excel file.
will move the fixture to the desired
(must be set the Excel
location
report format)

After testing and remove the specimen, press this


Setting Curve
button to return the grips to initial position
Color

6
Universal Testing Machine ManualV2011FEB

Setting Figures

NEW-Open a new test profile


OPEN-Open a saved configuration file in the test folder
The test file path is: C:\Porgram Files\TensionPC\Stf ,sub-file name is .stf
SAVE-Setting is complete, press the "Save” button can store the test conditions。
SAVE AS-The test condition "Save As ", you can use "open" this file to call out the use of key。

If set to end, press the "OK " button to save it to the test conditions *. Press the "Exit"button will not
be saved。

* "OK " button will only be stored in the settings, re-open the program will not be displayed again

。To open the program the next time when you press the "Save " button.
7
Universal Testing Machine ManualV2011FEB

Basic Information

Test content In the text box to enter the test content


Operator Default login user name
Reviewer Enter the name of the competent review
Name Enter Customer Name
Memo In the text box to enter information on the testing Remarks

8
Universal Testing Machine ManualV2011FEB

Test Condition

Test conditions set into test conditions, the unit set, spacing settings, stop setting, automatic back,
the speed setting。

9
Universal Testing Machine ManualV2011FEB

Test Condition

Load cell Please select correct load cell capacity,(Caution: If the wrong choice of load cell
may cause damage)。
Test method Please select the damage testing or fatigue testing
Test type Please Select Fixed(Volume for the conditions to test the power)Or length test
(The length of the test for the condition)。
Test direction Please select tensile or compress or tensile and compress(tensile first then compress)
Test times Fatigue tests done, you can set the number of times coming back, to the
automatic shutdown of a number of times。
Gross weight This setting is reached only when the above calculation of displacement, the gap
is too large to resolve the error caused.
Test force Please enter the maximum pull force value, reaches the set value will stop the
test (destructive test) or reverse (fatigue test).
Buffer force Reaches more than this setting, the motor will do a slow speed. The default is
80% of test requirements, test requirements change, it will automatically change.
This setting and testing the fatigue test the settings. .)

10
Universal Testing Machine ManualV2011FEB

Stop time When the pull force to reach the maximum pull force set value, the stay time.
Buffer elong Length test done to reach this length, the motor will be replaced by slow speed
forward. This setting and testing the fatigue test the settings.

Test elong Length test done to reach this length, the motor will take this setting is condition

Units Setting
This is the decimal point setting

Force units:Kgf、gf、KN、N、Tonf、lbf、ozf、daN、cN、MN、etc。
Length units:m、cm、mm、ft、etc。
Stress unit:Kgf/mm2、Kgf/cm2、Kgf/m2、gf/mm2、gf/cm2、KN/m2、N/m2、N/cm2、N/mm2、lbf/ft2、
lbf/in2、Mpa、Kpa、pa、kpsi、psi、Force/Areaetc。
Time units:Sec、Min ,etc。
Speed units:cm/min、cm/sec、mm/min、mm/sec、in/min、in/sec etc。
Constant Stress units:Stress/min、Stress/sec, etc。
Area units:mm2,cm2,m2,ft2,in2 等。
Responding units:mm/mm、cm/cm、m/m、ft/ft、in/in,etc。
Percentage units:%

11
Universal Testing Machine ManualV2011FEB

Spacing Setting(Set grip)


J i g d i s t a n c e Specimen elongation when measured as
an extension of set value rate
Gauge distance Specimen elongation when measured as
an extension of set value rate.
( extensometer is sold separately)

Stop Setting
S t a r t p o i n t Get over this setting to determine
whether to shut down only when, (as
long as one arrives, have effect before
the end of the test.) The default is 10%
of test requirements, test requirements
change, it will change automatically。
M a x f o r c e When the power reaches the set value
over volume, that is shut down
(destructive test) or reverse (fatigue
test).
M a x When the elongation reaches the set
E l o n g a t i o n value or more, the machine will stop (the
specimen may have been cracking
down.)

Auto retuen
Please select the "Start test point"(end of the test back to
the original location) or "no use "。Some special test
cases must be set not to use. (Possibly back to the bit
will bump into each specimen and direct damage to the
force sensor)

12
Universal Testing Machine ManualV2011FEB

Speed set

Test speed of the shift on the fixture.


Different test conditions in accordance
Dis.speed with the "speed shift"and"power
measured speed"and the two choices.

Press forward or reverse shift lever when


Jog Speed
the speed of cross head.

Return End of the test fixture on the speed of the


Speed shift。

Buffer force to reach the set value of


Low-speed
volume shift on the fixture speed。

Thickness Buffer :Can be first specimen after measuring its


Specimen
thickness, according to numbers entered to save operating
time.

Can open the sample files


have been saved

Can set the conditions for filing a


good specimen to facilitate the
use in the future。

13
Universal Testing Machine ManualV2011FEB

Sample size window Keyi your company needs, all the specifications of the product, one by one
input, in order to print the form, applied directly, without having to input, end save you valuable time
Fields for each cell corresponds to the print form of the project, set on the print form, see chapter 』
『 statements.
Testing requirements can be selected according to the shape of specimen

Hollow circular

Dumbbell
Circle

Rectangle
Irregular

Tear

14
Universal Testing Machine ManualV2011FEB

Display Area

Display settings set by the red box selected block, from top to bottom, respectively, numerical
display area, the curve display area, the data display area。

15
Universal Testing Machine ManualV2011FEB

Numerical display settings

Showing items can be set


to specify the content

Can set the maximum value (the value of retention


test) is the value (Current value)

16
Universal Testing Machine ManualV2011FEB

Curves display settings

Check to be used

This setting appears as


a special use (Such as
force / width of the
specimen)

Fixed scale - need display range limit curve can check the "fixed scale” in the X, Y-axis input to limit
the range of values.

17
Universal Testing Machine ManualV2011FEB

Information display area setting

Data display settings - to set the information item,By Step 1. → 2. → 3. sequence settings, please
refer to the following cases:

First in "a choice of projects", click"007 breaking point, “the right "2 Select sub" option appears.
Then select "2 Select the sub, " click the "breaking point 161 - power "(quick press 2, the left mouse
button), enter the setup screen breaking point.
Finally, "3 display items " will appear just select the "breaking point 161 - force ", the user can click
on the blue right down arrow to "data were tested zone " display item moves up and down sort
position Adjustment.

18
Universal Testing Machine ManualV2011FEB

Display the project settings

Can modify the display name display area, the name change,
in the information display area and fixed report shows

Enter the test curve display text


Check the display tag, the test curve
markers
appears markers

Choose whether the test is completed the value


of the specimen divided by the length ... etc.

Display Name-Information display area "and" fixed Report "shows the name, display name can be
modified.
Numerical upper and lower limit-If the test is completed the value exceeds the set range, the test
values shown in red.
Test specifications are subject to limited range of values, the user can check this feature, enter the
upper and lower limit values.

19
Universal Testing Machine ManualV2011FEB

Report output, such as selection of Excel, in Excel part → "cell"Please enter the Excel seat
placement
Placement according to user settings, select the output only once, choose the right place, put down,
extending the values are automatically selected by the user or down to the right place. If required,
the display unit, check out units. Simply select the items, such as Excel report shows, check Only
Excel Use; the project will not be displayed in the data display area.

Statistics show setting

Select to enter setting

Add the displayed value, click on the left of the text, then click the blue arrow to the right to do the
new action;To remove click the displayed value, click on the right to delete the displayed value, the
red arrow at the click of the left to do to delete the selected action.

Adjust the display position of numerical order, click on the blue, and the next position up and down
arrows to adjust.

20
Universal Testing Machine ManualV2011FEB

Result

Automatic

The test is completed automatically archived - When testing is completed, according to "automatic
archive file name used by the" automatic archiving.
Automatically after the archive file name plus 1 Automatically print the test is completed-
Form No.-Please enter the form has code for use when printing a form。

Testing Report

21
Universal Testing Machine ManualV2011FEB

Fixed report

Click Or in the "main screen toolbar” Click ,You can modify the fixed contents of the
report into the。

Fixed report – Title

Users can enter the title 1,2,3, respectively, the rise of text, such as: company name, company
phone and address.
Click the "Font" to modify the font

22
Universal Testing Machine ManualV2011FEB

Fixed report - on the part of the content

Click on the left of "Name" under the title text,Then click on the blue right arrow, the selected text
will be added to the right of "Item" below,If the user wants to modify the title name, click on
"Rename To" below to modify the text, the contents of the text changes from blue to。
Modify the order of the title text, click the blue up or down the right side of the blue arrow to adjust
the order to do。

Fixed report-Other, users can check their own use。


Curve - Test curve
A single test - the test data display area, a check of the first data。
Form - Test Information Form。

Fixed Report - Remarks


Fixed statements – footer
Users can enter the footer text, such as: Review Officer, to sign the nuclear ... etc。

23
Universal Testing Machine ManualV2011FEB

Preview of the default form

Title

Upper content

FORM

CURVE

END OF SHEET

24
Universal Testing Machine ManualV2011FEB

Excel form
Automatically print the test is completed, the user
can choose whether this feature is checked。

Excel Filename-Please click browse ,Select the Excel report has been produced well kept

paths.

To test the values into Excel, to be in the "set parameters → information → display area " to do
settings.
To apply the test data has been entered into the Excel report, as shown above Excel test report as
an example. The setting parameter - display - data display.

25
Universal Testing Machine ManualV2011FEB

Example: the name of the operator into the Excel

report.

In the "Select an item 1," select the test method → "2 Select sub"in the"011 users " Come into the set 2

Enter the cell location


of Excel reports

As long as the selected


item in the Excel report
using, please remember
to check the Only Excel
Use

26
Universal Testing Machine ManualV2011FEB

Test types Prescription


Customer setup
Users could set up certain testing type by themselves. Activating by sequences, click “add” to add an extra
test condition.
Action in order to add a test condition, select the test conditions to increase the number then click
"Insert. " To cancel the test conditions, please select the test condition number then click "Remove.
"

Click the drop-down


menu will appear select the test
direction

疲勞測試。設定的動作次數。

The test conditions are as follows:

Zero power - check, perform this step, the force will be zero current value

27
Universal Testing Machine ManualV2011FEB

Displacement of zero - tick, perform this step, the displacement value of zero will now
Shutdown conditions - tick, perform this step, start the stop condition detection.
Test direction - Stretch: action area is in the drawing area.
Compression: Motion area is in the compressed region
Speed Mode-Set the pace: speed mode mobile number per minute crosshead displacement。
Given stress: speed mode is the number per minute stress of moving crosshead。
Testing Seed-According to the selected speed mode enter a custom test speed。
Test Condition-
Force: The machine will control the strength of the specimen reaches the set
Displacement: machine will control the displacement of the specimen reaches the set.
Relative Force(+):To the current force, coupled with the power to set the value of
Relative displacement(+):To the current displacement, coupled with the set-bit
migration.
Damage: to set the direction, directly to do damage (based on the power limit.)
Bigger than the force: the power of control over the power to set, if the power is already
greater than the set value, it will skip this step。
Less than the force: the power of the power control to set the following, if the power is
already less than the set value, will skip this step。
Bigger than the displacement: the displacement control to set the displacement of more
than if we had set the value of the displacement, will skip this step。
Less than the displacement: the displacement of the displacement control to set the
following, if we have less than the set value of the displacement, will skip this step。
Yield strength: according to given stress action, check whether the power to reach the
yield difference。
Relative power (-): The Power of Now, and then lose the power to set the value

Test value - input according to the selected test conditions the custom test value.
Hold time - lag time pressure, set the machine will retain the power until the end of stay。

28
Universal Testing Machine ManualV2011FEB

Steel Chain Static state dynamic

Please setting by sequence:


1.RTS (max tensile force of the steel chain)
2.Load initially:When reaching RTS load initially,force and displacement will become to zero
3.1st test force:Testing force is RTS X setting percentage
4.1st test hold time:Reaching setting force, hold time
5. Please follow above regulations to set 2nd and 3rd item
6. Speed:Testing speed
7. After NO.4 test is completed, the machine will break the steel chain, please pay attention to this

29
Universal Testing Machine ManualV2011FEB

Three points bending

According to Three points bending, in


put “L” length

Four Point bending testing

According to four point bending grips


input the distance “L” and Span
‘L2”length