Edition 1, Revision 2
06-2008 printed in Germany
Document Node ID: 16099429
Agfa Company Confidential Copyright © 2008 Agfa HealthCare N.V.
ReadMeFirst
IMPORTANT:
The “order list for documentation” is not part of the service manual for download.
Purpose of the “order list for documentation”:
To verify the latest level and completeness of your Service Manual.
(1) In the MEDNET GSO library select the product of your interest.
(2) Select “Order list and Front page → PRODUCT - Chapter 00 - Order List for
Documentation”.
(3) Download the “order list for documentation” to your computer.
(4) When creating a paper manual:
Print the order list and put it behind the cover sheet.
IMPORTANT:
Preferably print this manual double-sided:
This PDF manual contains empty pages at the end of several chapters, to have the
next chapter starting with an uneven page number when printed doubles-sided.
If printed one-sided, dispose these empty pages.
NOTE:
“Shrink to printable area” may be named on other Adobe Reader versions
“reduce to printer margins” or “shrink oversized pages to paper size”.
3 Adding Comments
• If you open this file in an Adobe Reader version ≥ 7, the comment toolbar will
show-up.
• This allows adding comments, to highlight or underline text and many more text
manipulations.
NOTE:
Exporting your comments allows importing them again in a next version of the manual.
(1) In the drop down menu “Comment & Markup” select "Show comments List".
(2) Select the desired comments: Press the CTRL-key for multiple selections.
(3) Select “Options - Export Selected Comments”.
(4) Save the file with any name.
(1) In the drop down menu “Comment & Markup” select "Show comments List".
(2) Select “Options - Import Comments”.
(3) Browse for the comments file and press “select”.
NOTE:
The imported comments possibly appear on different pages, if the file where the
comments have been imported has a different number of pages.
CR 75.0
*1EOSTJ1* Type 5146 / 100 and 101
1 Piece EOSTJ MA1
Centricity CR MP3510
Type 5146 / 200 and 201
1st Edition
To order spare parts and Tel.: xx49 - (0)89 - 6207 – 3760
spare part assortments: Fax.: xx49 - (0)89 - 6207 – 7388
Caution:
This system uses mains voltage. Please consider the respective safety
regulations.
Note:
CE Declaration:
CR 75.0 The CE Declaration (CE Conformity) becomes invalid if the product is changed
Digitizer without explicit consent of the manufacturer!
Type 5146 This applies to all parts, not only to safety elements.
Base: cPCI
6. Accessories
7. Field Modifications
Required text input via the vips Enter vips and click on
12. Glossary and Key Word Index keyboard <Continue>
DD+DIS002.04E General Information
Overview of types:
The main changes between ADC Compact Plus and CR 75.0 are:
• 1 µs scan mode for all cassette formats to allow a higher throughput.
This option can be activated as of software version ACP_3101.
• a new photo multiplier tube to prevent from saturation artifacts
• Mammo capability (after its release)
this is achieved by a modified optic module that provides 50 µm scanning for
mammo.
The documentation of all ADC Compact Plus devices installed in the field will
be continued in the CR 75.0 service documentation as of March 2004.
This means:
• that all Supplements and Field Service Bulletins published for
CR 75.0 do also apply for ADC Compact Plus.
• that all Field Service Bulletins published for ADC Compact Plus stay valid and can
be found in chapter 10 of the CR 75.0 documentation.
• that the service documentation for ADC Compact Plus (DD+DIS024.01E) becomes
obsolete by end of March 2004, so please scrap this Service Manual and replace it
by DD+DIS002.04E.
Required Documentation
for UNIX based ADC Systems (VIPS)
ADC
Compact
Digitizer
Type 5145
ADC Processing
Station
Type 4406/700
ADC
Solo
Digitizer
Type 5155
ADC Preview-
and ID Station
Type 4406/841
ADC Software
Installation
Type 4406/...
CR 75.0 /
ADC Compact Plus
Digitizer
Type 5146
Required Documentation
for Windows NT based ADC Systems (ADC QS)
ADC
Compact
Digitizer
Type 5145
ADC
Solo
Digitizer
Type 5155
CR 75.0 /
ADC Compact Plus
Digitizer
Type 5146
This Generic Safety Directions document comprises the general safety relevant information including
relevant environmental and occupational safety instructions for the Service Engineer.
It is valid for all Agfa HealthCare Imaging Products and part of each Service Documentation as well as
Installation Planning document.
The latest version is available via MedNet, GSO Library path:
General Info => Agfa HealthCare => Publications => Service Manual
► Document History
► Referenced Documents
Document Title
Not applicable Not applicable
Edition 1, Revision 3
07-2009 printed in Germany Document Node ID: 11849633
eq_generic_safety_directions_e_template_v01
Agfa Company Confidential Copyright © 2009 Agfa HealthCare N.V.
DD+DIS238.06E Generic Safety Directions
► Manufacturer
Agfa HealthCare N.V.
Publisher
Agfa-Gevaert HealthCare GmbH
Tegernseer Landstraße 161
D - 81539 München
Germany
Copyright © 2009 Agfa HealthCare N.V.
All rights reserved.
Technical modifications reserved.
AGFA and the Agfa-Rhombus are trademarks of Agfa HealthCare N.V.
WARNING:
Improper operation or service activities may cause damage or injuries.
INSTRUCTION:
(1) Read the "Generic Safety Directions" document
(see MEDNET GSO => General Info => Agfa HealthCare => Publications =>
Service Manual) prior to attempting any operation, repair or maintenance task on the
equipment.
(2) Strictly observe all safety directions within the "Generic Safety Directions" and on
the product.
Edition 1, Revision 3 Generic Safety Directions for HealthCare Imaging Products Page 2 of 28
07-2009 Agfa Company Confidential
DD+DIS238.06E Generic Safety Directions
LIST OF CONTENTS
1 DISCLAIMER......................................................................................................... 5
3 LABELS ................................................................................................................. 7
3.1 CE Mark................................................................................................... 7
3.2 System Labels ......................................................................................... 7
3.3 System Labels concerning Laser Radiation ............................................ 8
4 PRODUCT COMPLAINTS .................................................................................. 10
5 REFERENCES .................................................................................................... 10
6 INTENDED USE.................................................................................................. 11
12 COMPLIANCE..................................................................................................... 14
Edition 1, Revision 3 Generic Safety Directions for HealthCare Imaging Products Page 3 of 28
07-2009 Agfa Company Confidential
DD+DIS238.06E Generic Safety Directions
24 RECYCLING ........................................................................................................27
Edition 1, Revision 3 Generic Safety Directions for HealthCare Imaging Products Page 4 of 28
07-2009 Agfa Company Confidential
DD+DIS238.06E Generic Safety Directions
1 Disclaimer
Fitters, engineers and other persons who are not employed by or otherwise directly
affiliated with or authorized by Agfa HealthCare or one of its affiliates are directed to
contact one of the local offices of Agfa HealthCare or one of its affiliates before attempting
installation or service procedures.
No part of this document may be reproduced, copied, adapted or transmitted in any form
or by any means without the written permission of Agfa HealthCare.
Agfa HealthCare shall under no circumstances be liable for any damage arising from the
use or inability to use any information, apparatus, method or process disclosed in this
document.
Agfa HealthCare is not liable for resulting consequences, damages or injuries if you don’t
operate the product correctly or if you don’t have it serviced correctly.
Agfa HealthCare reserves the right to change the product, the characteristics and its
documentation without further notice to improve reliability, function or design.
NOTE:
In the United States, Federal Law stipulates that medical devices should only be sold to,
distributed and used by or by order of a licensed physician.
Edition 1, Revision 3 Generic Safety Directions for HealthCare Imaging Products Page 5 of 28
07-2009 Agfa Company Confidential
DD+DIS238.06E Generic Safety Directions
2 Used Icons
INSTRUCTION:
If used in combination with the warning or caution sign: Indicates
a specific instruction, which if followed exactly, avoids the
subject of the warning or caution.
If used without warning or caution sign: Indicates an instruction
where it is important to follow literally as described.
IMPORTANT:
Highlights very important actions which have to be carried out to
prevent malfunction.
NOTE:
Indicates advice to facilitate the following step or action.
Highlights unusual points.
Indicates background information.
Can be used to explain or highlight displays of the
graphical user interface.
Is additional information without influence on the action or step!
Edition 1, Revision 3 Generic Safety Directions for HealthCare Imaging Products Page 6 of 28
07-2009 Agfa Company Confidential
DD+DIS238.06E Generic Safety Directions
3 Labels
3.1 CE Mark
CE Mark This product carries the CE Mark. The CE Declaration (CE Conformity)
becomes invalid if the product is changed without explicit consent of the
manufacturer! This applies to all parts, not only to safety elements.
All system labels and software version number locations are referred to within this service
document in the appropriate section.
Enclosed an overview of common labels, according to ISO 3864.
This list is not complete.
Edition 1, Revision 3 Generic Safety Directions for HealthCare Imaging Products Page 7 of 28
07-2009 Agfa Company Confidential
DD+DIS238.06E Generic Safety Directions
According to its classification, laser radiation can lead to eye and skin injuries.
Each laser source is classified from class 1 to class 4, based on standard
DIN EN 60825-1:2007.
The table below lists the meaning of the different laser classes. Note the detailed
instructions in the user manual and technical documentation.
conditions.
Edition 1, Revision 3 Generic Safety Directions for HealthCare Imaging Products Page 8 of 28
07-2009 Agfa Company Confidential
DD+DIS238.06E Generic Safety Directions
Edition 1, Revision 3 Generic Safety Directions for HealthCare Imaging Products Page 9 of 28
07-2009 Agfa Company Confidential
DD+DIS238.06E Generic Safety Directions
4 Product Complaints
Any service person who has any complaints or has experienced any dissatisfaction in the
quality, durability, reliability, safety, effectiveness or performance of this product must
notify Agfa HealthCare by the Agfa HealthCare complaint procedure.
If the product malfunctions and may have caused or contributed to a serious injury of a
patient or an accident or if there are any hazards which may cause an accident
Agfa HealthCare must be notified immediately by telephone, fax or written
correspondence to the following address:
Agfa Service Support - local support addresses and phone numbers listed on:
www.agfa.com
5 References
Technical Documentation is available via MedNet (PDF) and your local Agfa HealthCare
support organisation (Paper).
Access to MedNet:
IntraNet: http://docs.agfanet/bu/mi/mednet/mednetcso.nsf
ExtraNet: http://extranet.agfa.com/bu/mi/mednet/mednetcso.nsf
Edition 1, Revision 3 Generic Safety Directions for HealthCare Imaging Products Page 10 of 28
07-2009 Agfa Company Confidential
DD+DIS238.06E Generic Safety Directions
6 Intended Use
This Agfa HealthCare product should only be operated in a hospital or clinical radiological
environment by qualified staff.
It must only be operated according to its specifications and its intended use. Any
operation not corresponding to the specifications or intended use may result in hazards,
which in turn may lead to serious injuries or fatal accidents (for example electric shocks).
AGFA will not assume any liability whatsoever in these cases.
Make sure that the product is constantly monitored in order to avoid inappropriate
handling, especially by children.
The product must only be installed and put into operation under the specified conditions.
7 Intended User
This manual is written for Agfa trained Field Service Engineers and Clinical Application
Specialists, trained users of Agfa HealthCare products and trained diagnostic X–Ray
clinical personnel who have received proper training. Users are considered as the persons
who handle the equipment as well as the persons having authority over the equipment.
This Technical Documentation describes adjustments and routines which must only to be
performed by qualified technical personnel.
The Agfa (trained) Field Service Engineers and Clinical Application Specialists must have
received adequate Agfa HealthCare training on the safe and effective use of the product
and applicable environmental and occupational safety matters before attempting to work
with it. Training requirements may vary from country to country.
Agfa (trained) Field Service Engineers and Clinical Application Specialists must make
sure that training is received in accordance with local laws or regulations that have the
force of law.
Your local Agfa HealthCare representative can provide further information on training.
Edition 1, Revision 3 Generic Safety Directions for HealthCare Imaging Products Page 11 of 28
07-2009 Agfa Company Confidential
DD+DIS238.06E Generic Safety Directions
Each Agfa (trained) Field Service Engineer and Clinical Application Specialist:
• Must make his or her personal contribution to improve safety and protect the
environment.
• When working on a customers site, has a duty to take reasonable care to avoid injury
to himself or herself or to others who may be affected by their acts or omissions.
• Is obligated to adhere strictly to regulations and instructions.
• Shall familiarise himself or herself with the provisions of the Agfa Healthcare
Health, Safety and Environment Policy and any specific rules or procedures relating to
occupational safety at work and the protection of the environment.
• Shall promptly report any near misses, accidents, incidents or dangerous occurrences
to their line manager and co-operate fully in any investigation.
• Shall co-operate with company management on matters relating to
health, safety and environment and, where appropriate, discuss with and / or assist
their manager in resolving matters relating to health, safety and environment.
• Shall ensure that any company equipment issued to them, or, for which they are
responsible, is correctly used and properly maintained.
• Shall wear protective equipment whenever instructed or if it is recommended to do so.
• Shall be responsible for good housekeeping in the area in which he or she is working.
• Shall report situations, which could put them at risk, on either company or
customers' premises, to their manager or supervisor; and, if warranted, directly and in
confidence, to the Health and Safety Co-ordinator, Global HSE Manager, or ultimately
to the Managing Director.
• Shall report any injuries, diseases or dangerous occurrences to his or her line
manager.
• Shall report any accidents, incidents or near misses to his or her line manager.
• Shall report any situation of which he or she is aware that is potentially dangerous.
• Shall comply with any health surveillance procedure instituted for his or her benefit or
for compliance with regulations.
Edition 1, Revision 3 Generic Safety Directions for HealthCare Imaging Products Page 12 of 28
07-2009 Agfa Company Confidential
DD+DIS238.06E Generic Safety Directions
Agfa HealthCare equipment must only be used in combination with other Agfa HealthCare
equipment or components if these are expressly recognized by Agfa HealthCare as
compatible. A list of such equipment and components is available from Agfa HealthCare
service on request.
Changes or additions to the equipment must only be carried out by persons authorized to
do so by Agfa HealthCare. Such changes must comply with best engineering practice and
all applicable laws and regulations that have the force of law within the jurisdiction of the
hospital.
The Agfa HealthCare products are designed to communicate with other devices in the
hospital network using DICOM protocols.
INSTRUCTION:
Consult the Technical Documentation before making any connections to other equipment.
Consideration relating to the choice of accessory equipment shall include:
• Use of the accessory equipment in the patient vicinity.
• Evidence that the safety certification of the accessory equipment has been performed
in accordance with the appropriate IEC 60601-1 and IEC 60601-1-1 harmonized
national standard.
In addition all configurations must comply with the medical electrical systems standard IEC
60601-1-1. The party that makes the connections acts as system Configurer and is
responsible for complying with the systems standard.
If required, contact your local service organization.
Edition 1, Revision 3 Generic Safety Directions for HealthCare Imaging Products Page 13 of 28
07-2009 Agfa Company Confidential
DD+DIS238.06E Generic Safety Directions
INSTRUCTION:
• Replace defective parts with Agfa HealthCare original spare parts.
• Use only tools and measuring instruments which are suitable for the procedure.
• Only approved Agfa HealthCare accessories must be used. For a list of compatible
accessories contact your local Agfa HealthCare organization or www.agfa.com.
12 Compliance
Edition 1, Revision 3 Generic Safety Directions for HealthCare Imaging Products Page 14 of 28
07-2009 Agfa Company Confidential
DD+DIS238.06E Generic Safety Directions
• IEC 60601-1, Ed. 3: Medical electrical equipment - Part 1: General requirements for
basic safety and essential performance
Harmonization:
This document has been prepared to comply with Study Group 1 guidance document
of the Global Harmonization Task Force (GHTF) www.ghtf.org/ to assist development
of a consistent, harmonized definition for a medical device that could be used within a
global regulatory model and would offer significant benefits to the manufacturer, user,
patient or consumer, and to Regulatory Authorities and support global convergence of
regulatory systems.
IECEE CB SCHEME
The IECEE CB (Certification Body) Scheme is the world's first truly international system
for acceptance of test reports dealing with the safety of electrical and electronic products.
It is a multilateral agreement among participating countries and certification organizations.
Agfa has produced a CB test report and claims national certification in all other member
countries of the CB Scheme.
Edition 1, Revision 3 Generic Safety Directions for HealthCare Imaging Products Page 15 of 28
07-2009 Agfa Company Confidential
DD+DIS238.06E Generic Safety Directions
Note:
This product has been tested and found to comply with the limits for a Class A
computing device pursuant to Subpart B of Part 15 of FCC Rules, which are designed to
provide reasonable protection against such interference when operated in a commercial
environment.
Operation of this equipment in a residential area is likely to cause interference.
The user will be required to take all necessary measures to correct the interference at
his own expense.
Edition 1, Revision 3 Generic Safety Directions for HealthCare Imaging Products Page 16 of 28
07-2009 Agfa Company Confidential
DD+DIS238.06E Generic Safety Directions
• Under certain conditions the Agfa HealthCare product will show a display containing a
message. This message will show that either a problem or action has occurred or that
a requested action is required or cannot be performed. The user must read these
messages carefully they will provide information on what to do. This will be either
performing an action to resolve the problem or to contact the Agfa HealthCare
service organization. Details on the contents of messages can be found in this
Technical Documentation.
• All images created using any image technology can show artifacts which could be
confused with diagnostic information. If there is any doubt that the diagnostic
information could be corrupted, additional investigations must be performed to get
clear diagnostic information.
• Ventilation openings must not be covered.
• If you notice conspicuous noise or smoke, disconnect the product immediately from
the mains.
• Do not pour water or any other liquid over the device.
• If a system malfunction causes an emergency situation involving the patient, operating
personnel or any system component, activate the emergency stop for the system
concerned. All motor driven system movements will be stopped.
• Do not store any magnetic media near or on devices, which produce magnetic fields,
since stored data may be lost.
Explosive environment:
DANGER:
Risk of explosion.
Never operate this device in zones where there are flammable anesthetics or oxygen which
may cause an explosion.
Edition 1, Revision 3 Generic Safety Directions for HealthCare Imaging Products Page 17 of 28
07-2009 Agfa Company Confidential
DD+DIS238.06E Generic Safety Directions
14 Radiation Protection
Only qualified and authorized personnel shall operate any X-Ray system. In this context
qualified means those legally permitted to operate this X-Ray equipment in the jurisdiction
in which the X-Ray equipment is being used, and authorized means those authorized by
the authority controlling the use of the X-Ray equipment. Full use must be made of all
radiation protection features, devices, systems, procedures and accessories.
Ionizing radiation can lead to radiation injuries if handled incorrectly. When radiation is
applied, the required protective measures must be complied with.
• Details about cleaning and disinfection or sterilization methods that may be used on
SYSTEM parts or ACCESSORIES that can become contaminated through contact
with the PATIENT or with body fluids, are referred to within the individual service
documents.
• Disconnect the power supply from the equipment prior to cleaning the equipment.
Edition 1, Revision 3 Generic Safety Directions for HealthCare Imaging Products Page 18 of 28
07-2009 Agfa Company Confidential
DD+DIS238.06E Generic Safety Directions
• This system uses high voltage. Please consider the respective safety regulations.
• Electrical repairs and connections must only be performed by a qualified electrician.
• Mechanical repairs and connections must only be performed by a qualified technician.
• The safety directions for operation (see section 13) are also valid for all service
activities.
• During all service activities observe prescribed local and country-specific
requirements (e.g. occupational safety and accident prevention regulations).
• All existing screw connections must be tightened sufficiently firmly, but they may not
be overstressed when tightening. There must always be compliance with stated
torque values!
• Damaged or missing screws may be replaced only with the same screw types that
have the specified hardness rating. Unless a different value is listed in the
instructions, all Allen screws used must be hardness rated 8.8.
• All screws must be secured in accordance with the corresponding data.
If "Loctite" has to be used to secure screws, this is stated in the text.
• Any Agfa service PC or tool which is to be connected via RS232, RJ45, USB or other
interface to an Agfa device must not be connected to the mains but must be operated
on its internal battery or indirect supply (low voltage).
• When handling printed circuit boards (abbr.: PCBs) the following points must be
observed:
o Always switch off the equipment and unplug the power cord, before you
disconnect or connect cables on printed circuit boards.
o When working on PCBs, always wear an anti-static wrist strap. Never touch any
parts or components on PCBs with your bare fingers.
o PCBs have to be kept or transported in their protection bags. Never carry a
PCB without protection bag and walk on carpet or plastic floor covering
(electrostatic charge).
o Once the PCB is taken out of its protection bag, it has to be protected from
electrostatic charge by a grounded mat.
Edition 1, Revision 3 Generic Safety Directions for HealthCare Imaging Products Page 19 of 28
07-2009 Agfa Company Confidential
DD+DIS238.06E Generic Safety Directions
The digitizer and the cassette storage shall be protected against X-ray radiation this way,
that the annual dose equivalent at the installation place will not exceed 1 mSv.
The film-screen system shall be protected against X-ray radiation this way, that the annual
dose equivalent at the installation place will not exceed 1 mSv.
INSTRUCTION:
Edition 1, Revision 3 Generic Safety Directions for HealthCare Imaging Products Page 20 of 28
07-2009 Agfa Company Confidential
DD+DIS238.06E Generic Safety Directions
• The device complies with the EN 60601-1, 2006 standard for Information Technology.
This means that, although it is absolutely safe, patients may not come in direct contact
with the equipment. Therefore the operator console must be placed outside a radius
of 1.5 m around the patient.
R = 1.5m
1.5m
1.5m
Patient environment
• This device should be installed behind the institution firewall for network security and
anti-virus protection. No ongoing computer virus protection or network security for this
medical device is provided (e.g. a computer firewall). Network security and anti-virus
provisions are the ongoing responsibility of the user or institution.
Floor load:
CAUTION:
Heavy device may damage the floor covering.
Make sure that the floor covering is solid enough to stand the weight of the device.
Edition 1, Revision 3 Generic Safety Directions for HealthCare Imaging Products Page 21 of 28
07-2009 Agfa Company Confidential
DD+DIS238.06E Generic Safety Directions
INSTRUCTION:
Prior to connecting the device to the mains:
• Compare the power requirements indicated on the type label with the available
power supply in the installation room.
• Check the service manual for the type of input voltage selection, manual or
automatic: If manual, select the appropriate voltage and fuses.
• Confirm to use the correct socket and plug for the required power supply.
• Check the equipment will work with the power supply available.
Edition 1, Revision 3 Generic Safety Directions for HealthCare Imaging Products Page 22 of 28
07-2009 Agfa Company Confidential
DD+DIS238.06E Generic Safety Directions
INSTRUCTION:
• Always connect the associated monitor to the same Uninterruptible Power Source as
the PC.
• When different combinations of equipment are used in various medical environments
a potential difference (V) can exist between the protective earths in different localities.
If the protective earthing fails this potential difference can cause a HAZARD for the
OPERATOR or for the PATIENT.
• This Technical Documentation identifies the parts on which preventive inspection and
maintenance shall be performed by Agfa HealthCare service personnel, including the
periods to be applied.
• In general the device has to be switched off during service activities. Exception: If the
device is switched on to perform tests pay particular attention to any hazards due to
moving and rotating parts. Avoid lose clothing or finger traps. Switch off the device
immediately after the tests.
• Do not turn motors manually. If required, first disconnect the motor from the motor
control board.
• Make sure that the power cord does not show any signs of damage.
• After repair work always check that the integrated safety features are not overridden
or disconnected.
• If there is any visible damage to the machine casing do not hand-over the product to
the customer. First repair the machine casing.
Edition 1, Revision 3 Generic Safety Directions for HealthCare Imaging Products Page 23 of 28
07-2009 Agfa Company Confidential
DD+DIS238.06E Generic Safety Directions
Replacing batteries:
WARNING:
Battery can explode, causing chemical burns.
INSTRUCTION:
NOTE:
On MedNet, GSO Library path "General Info => Agfa HealthCare => Publications =>
Service Manual" a form “IEC 62353:2007 Test Documentation” for CR Digitizers is
available, to be handed out to the responsible organization (Intranet Link / Extranet Link).
1
Responsible Organization:
Entity accountable for the use and maintenance of a medical equipment or a medical equipment system. The
accountable entity can be, for example, a hospital or an individual clinician.
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 3 Generic Safety Directions for HealthCare Imaging Products Page 24 of 28
07-2009 Agfa Company Confidential
DD+DIS238.06E Generic Safety Directions
INSTRUCTION:
• Strictly observe the warning notes in the service manual of devices emitting laser
radiation (See service manual chapter describing Safety Guidelines / General
Repair Instructions) and at the corresponding steps of instructions.
• Strictly observe the warning labels at the modules emitting laser light. For the
meaning of the labels refer to section 3.3 in this document.
• Do not look into the laser beam.
• Do not open modules containing a laser. Only open modules containing a laser if
explicitly instructed to do so.
• Do not keep tools in the laser beam unless explicitly instructed to do so.
• Make yourself familiar with the path of the laser light and the conditions, when the
laser beam is switched on. Refer to the Functional Description in the
corresponding service manual.
• Do not operate modules with laser outside the device.
Sharp edges:
CAUTION:
Sharp edges inside the device: Cut or abrasion possible.
Be careful at maintenance and replacement of parts.
Edition 1, Revision 3 Generic Safety Directions for HealthCare Imaging Products Page 25 of 28
07-2009 Agfa Company Confidential
DD+DIS238.06E Generic Safety Directions
Secured screws:
CAUTION:
Opening screws secured by red lacquer may misalign important device
adjustments:
Do not open screws that are secured by red lacquer.
Replacing fuses:
Warning:
Replacing fuses by wrong type may lead to fire hazard!
Use only fuses of the exact value and characteristics stated in the service manual or on
the device.
Edition 1, Revision 3 Generic Safety Directions for HealthCare Imaging Products Page 26 of 28
07-2009 Agfa Company Confidential
DD+DIS238.06E Generic Safety Directions
'Hazardous materials' is the designation for substances which can ignite or explode or
which are toxic, injurious to health, corrosive or irritating. The “Hazardous Material”
instructions must be read and the required protective measures must be complied with
when performing work to avoid health risks.
Their properties together with the hazards and protective measures connected with them
are identified clearly by symbols and described by the instructions appertaining to the
hazardous substances.
24 Recycling
Agfa HealthCare has Recycling Passports available for all equipment. The Recycling
Passport explains whether hazardous materials, special components and batteries are
present, where they are located and how they can be removed at the end of the life cycle.
The Recycling Passports are meant to be used as information for waste treatment
partners and companies that want to recycle end-of-life Agfa equipment.
To get a copy of the required Agfa HealthCare Recycling Passport please contact your
local Sales organization.
Edition 1, Revision 3 Generic Safety Directions for HealthCare Imaging Products Page 27 of 28
07-2009 Agfa Company Confidential
DD+DIS238.06E Generic Safety Directions
25 Waste Disposal
AGFA HealthCare Field Service Personnel or its authorized affiliates are responsible for
the removal of Protected Health Information (PHI) patient data from devices, modules or
parts that are removed from the customer’s site. This also applies to the exchange of
spare parts, especially to parts that are returned to central warehouses for repair or
refurbishing. Examples for parts or modules that may contain Protected Health
Information (PHI) are: Computer hard disks, CD-ROMS, backup tapes, archive tapes.
Edition 1, Revision 3 Generic Safety Directions for HealthCare Imaging Products Page 28 of 28
07-2009 Agfa Company Confidential
DD+DIS226.05E Controls, Connections and Setup Procedures
Chapter 1
List of Contents
1 Safety ......................................................................1
1.1 General Safety Instructions ............................................... 1
1.2 Safety Instructions for Laser Products............................. 4
1.3 Safety Instructions for Cassette Unit ................................ 4
1.4 Safety Regulations ............................................................. 4
1 Safety
• The CR 75.0 has been designed for scanning medical X-ray image plates
and should only be used for these purposes.
• The CR 75.0 must only be operated by qualified staff trained on the
machine.
• Make sure that the CR 75.0 is constantly monitored in order to avoid
inappropriate handling, especially by children.
• Only trained service personnel must make repairs. Only authorized service
personnel must make changes to the CR 75.0.
• If there is any visible damage to the machine casing, do not start nor use
the CR 75.0.
• If you want to connect the CR 75.0 with other devices, components or
assemblies and if the technical data do not permit determining whether the
combination with these devices, components or assemblies involves
hazards, you must consult the respective manufacturers to avoid danger for
operating personnel or the environment.
• Do not override or disconnect the integrated safety features.
• Switch off the CR 75.0 before performing any maintenance work or repairs.
Disconnect the CR 75.0 from the mains before making repairs or
performing any maintenance activities.
• As is the case for all technical devices, the CR 75.0 must be operated,
cared for and serviced correctly.
• If you don't operate the CR 75.0 correctly or if you don't have it serviced
correctly, Agfa-Gevaert is not liable for resulting disturbances, damages or
injuries.
• When installing the CR 75.0, care must be taken to ensure that there is
either a mains plug or an all-cable disconnecting device in the internal
installation fitted near the CR 75.0 and that it is easily accessible.
• If you notice conspicuous noise or smoke, disconnect the CR 75.0
immediately.
• Check that the mains voltage is within the specified range of the self
adapting power supply of the machine.
• You can hurt your fingers if they are caught between the ADC Cassette
and the edge of the input slot. Insert the cassette in the input buffer as
described in the User Manual. At all times, keep your fingers clear of the
input slot. As soon as the CR 75.0 takes in the cassette, release it.
In order to reduce the risk of electric shock, do not remove any covers.
Caution hot:
Type B equipment:
Indicates that the CR 75.0 complies with the limits for type B equipment.
Intergrounding connector:
Provides a connection between the Digitizer and other equipment which might
exhibit minor ground potential differences. These differences may degrade
the quality of communication between different equipment. Never remove
connections to this terminal.
Provides a connection between the Digitizer and the protective earth of the
mains. Do not remove this connection, because this will have a negative
influence on the leakage current.
Power on
Power off:
Note that the power cord has to be disconnected from the wall outlet in order
to disconnect the unit entirely from the mains.
Installation Regulations
Electrical installations in the installation room must be in compliance with:
• IEC 60364 (e.g. for Germany: VDE 0100) regulations for Europe,
• NEC for USA / Canada.
Mains Connection
Connection via plug and socket.
Certificates
• The CR 75.0 Digitizer is in compliance with EG regulation 93/42/EEC
(Medical device directive).
• The CR 75.0 Digitizer is tested in compliance with:
- IEC 601-1 1988 + A1 : 1991 + A2 : 1995
- EN 60601-1 : 1990 + A1 : 1993 + A2 : 1995
- EN 60601-1-2 : 1993
- EN 60825-1:1994 + A11:1996 : 2001
- UL2601-1 Second Edition 22 No. 601.1 – M909
- The CR 75.0 Digitizer is UL / CUL classified.
- E166741 (PIDF)
In case of damage visible from the outside turn to your local AGFA
representative.
oben - top
white field, if o.k.
450034BM.cdr
Figure 1
• Taking out the accessories or boxes and storing them in a safe place.
• The CR 75.0 must be taken off the pallet and transported to the installation
site.
• Disposal of the packing material.
Pieces Description
1 CR 75.0 Digitizer
1 base plate
1 installation instructions
1 installation report
1 CR 75.0 user manual information (on CD)
1 certificates
6 floppies
1 UTP network cable (2 x RJ45 connection)
1 Allen key, size 8 with extension
1 connection cable (Europe: CEE(7) VII 250 V/16 A)
1 connection cable (USA, Japan: Nema 6-15 P)
2 adjustable bases, riveted
Remove all cushioning parts and shipping brackets prior to first operation.
B
514601eg_ab.cdr
Figure 2
(1) Unscrew the locking screws at the front of the digitizer (A)
(8 mm Allen key, extension included in accessory kit).
(2) Remove the blind flange protecting the RS232 connector. (B)
(3) Then open the doors of the digitizer.
7 mm
C
B B
8 mm A
7 mm
514601EC.cdr
Figure 3
B C
514601EB.CDR
Figure 4
514601EE.cdr
Figure 5
A
514601ej.cdr
Figure 6
(14) Remove the cushioning block between IP transport unit postscan and
scan unit (hardly visible) (A).
514601EL.cdr
Figure 7
(15) Cut the polystyrene blocks jammed at the feed- and exit rollers of the
scan unit (2 pieces).
B
514601EA.cdr
Figure 8
B 514601ek.cdr
Figure 9
(20) Remove the two allen screws at the rear of the machine securing the
scan unit (A).
(21) Mount the light tight tin plate (included in accessory kit) by fastening six
Phillips screws (B).
514601AC.cdr
Figure 10
(22) Mount the base plate with four screws on the floor (recommended, but
not mandatory), see arrows.
514604ae.cdr
Figure 11
(23) Move the digitizer onto the base plate. Take care that the rollers are
guided by the rail!
(24) Move the rollers of the digitizer gently to the stop (see circle).
514601af.cdr
Figure 12
(25) Turn screws (on both sides, left and right) by hand to their stop, then
screw it in one more turn. Do not give more turns!
514601ag.cdr
Figure 13
514601ah.cdr
Figure 14
3 2
6
5 Twisted Pair
VDE
UL 514601ef.cdr
Figure 15
1 Status indicator
2 Line switch Digitizer
3 Port (RS 232 - 25-pin D connector, female D) for the Service PC
4 Network connection (Twisted Pair)
5 Mains cable VDE/UL
6 Supplementary Grounding cable connection
Potential Equalization
515502aa.cdr
Figure 16
Display Functions
The key operator level is intended to be used by the normal operator. He uses
the system only to produce diagnostic usable images without the requirement
for special technical skill or background.
Key Function Key Function
4.1 General
Preconditions
• finished pre-installation
• the digitizer is at its installation site
4.2.1 CCM-Tool
The latest version of the CCM - Tool can be downloaded from MedNet.
The CCM - Tool is needed to create and modify the configuration file (adc.cpf)
for the CR 75.0 and the other ADC System Components.
Installation:
• install by executing "setup.exe"
The CCM Tool will be
installed in directory
"IMOS" or any
directory of your
choice.
GS100-MULTI-
Supply-BD
5146_reg02_008.cdr
Figure 17
Mechanical Start-Up
Start-Up:
1. Switch on the digitizer.
2. Check the operation terminal during the initialization.
While booting for the first time, the message
<Installation not yet confirmed> appears.
3. Confirm with OK.
4. After a successful start-up <ready> appears.
5. Check/adjust date and time in key operator mode.
6. Choose terminal language in the service menu
<Configure>
<User terminal language>
7. Store the modified machine-specific data.
Check:
8. Open the service menu
<Maintenance>
<Handling Cycle>
and check the function of cassette- and plate operation with the machine
open. Each size should run at least one time.
9. Erase all image plates.
End of Mechanical Start-Up (approx. 30 min.)
Figure 18
Precondition
The destinations, i.e. Processing Stations, intended for the digitizer, have to
be put into operation and must be obtainable via Ethernet.
System Check
1. Check whether the Processing Station(s) is/are obtainable.
Service menu:
<Checks>
<Check destinations>
2. Send a test image to the Processing Station(s).
Key operator menu:
<Send test image>.
End of System Check.
Backup
1. Insert a new formatted floppy.
2. Start the service program.
3. Select:
<save on floppy>
<Machine specific data>
4. Start the backup procedure.
5. Label the floppy as follows:
S/N: <XXX>
Date: <date>
Software- ACP_xxxx
version:
End of Backup.
The hard disk of the digitizer contains three samples showing the minimal
required image quality (2x flatfield, 1x testsheet).
One testsheet for quality comparison and two flatfields (Banding, Calibration)
can be sent to a printer via Service-Menu
<CHECKS>
<SEND FLATFIELD>
<SEND TESTSHEET>
• Expose a gray field (flatfield) and evaluate the image on the Processing
Station and the laser printer.
Criteria: homogenous field, stripes etc.
• Have a new plate of every format exposed twice as follows:
Rotating
Anode x-ray tube
Cathode
514003pm.cdr
Figure 19
For exposure place the cassette in length direction to the X-ray tube see figure
above.
1. 1st exposure: parameters:
12 mAs-77kV-1,3 m (51.22 In) distance,
1.5 mm Cu filter
2. Turn cassette through 180°.
3. 2nd exposure: parameters as in 1st exposure.
• Insert the cassette into the Digitizer. Print the image on a printer with a
window setting of 0.6 without changing the level setting on the Processing
Station.
• Check plates for running askew with a test sheet.
Chapter 2
List of Contents
Figure 1
2 ADC Cassettes
The principle function of the digitizer is close connected to the structure of the
ADC Cassettes. The digitizer accepts only ADC Compact and Solo Cassettes.
All other cassettes, like ADC 70, are refused.
The image plate lying in the cassette consists of five different layers:
The phosphor layer is “storing” the Image Plate Structure
picture information as energy . 1
The EBC top coat provides a smooth 2
5146_reg02_003.CDR
surface of the image plate to reach a 3
good signal-to-noise ratio. 4
Specification:
The Embrace Mammo Cassette is a dedicated cassette for use in combination
with Embrace Mammo Plates 1.00 for dedicated mammography applications.
It features a narrow hinge to allow scanning right up to the chest wall.
Identification:
The cassette is labeled with “Mammo” on both sides.
The cassette is pre-initialized as a Mammo Cassette, but if the identification is
lost, it can be given via the ID Station.
Figure 4
1 positioning mechanism for IP (2 x, left and right cassette side
2 passive suction cup, keeping the IP in place inside the cassette
Formats:
The ADC Mammo Cassette is available in two formats:
• 18 cm x 24 cm
• 24 cm x 30 cm
Identification:
5146_reg02_004.cdr
Figure 5
Formats:
The ADC Asymmetric Cassette is available in:
• 21 cm x 43 cm (asymmetric partial scan of dedicated 35 x 43
cassettes)
Specification:
The Asymmetric Cassette is dedicated for X-ray exposures where no centered
positioning of the part of the body is possible (e.g. side view of a spinal
column).
Input Buffer
Power Unit
Output Buffer
cPCI-Rack
Cassette Unit 1 / 2
IP Transport Unit Prescan
Rotation Unit
Optic Module
Erasure Unit
Figure 6
Figure 7
1 Transport belt
The Input Buffer has a capacity of 10 cassettes which can be put at the same
time before the door. The cassettes are detected and the machine cycle is
started with the first one. To transport the cassette inside the digitizer the
transport belt shifts the cassette, the feed gate opens and the RF-Tag reader
and the antenna read out the ID- and image specific data.
After the data is read out the transport roller of the cassette module, which is
at the input position, takes over the movement of the cassette.
4
3
Figure 8
The power supply of the Power Unit is a single phase System, means the
cable connection is a common inlet connector for non-heating apparatus.
The supply voltage for the Power Unit is automatically detected. Therefore for
installation it is not necessary to change any setting manually, but has to be
checked before the first use.
If the interlock switch is active, all the power supplies of mechanical parts and
the laser are interrupted, for safety during service interventions. The power
supply of the cPCI-Rack is placed in the rack itself, so e.g. software updates
can be done although the right door of the power unit is open and the interlock
switch is active.
The Multi Supply Board (MS-Board) controls the mechanical periphery of the
digitizer and protects the stepper motor cards. The supply voltage is detected
by the board and errors are displayed by LEDs.
The following components are controlled by the IO-Bus of the board: The two
vacuum pumps with two magnetic valves on the IP transport units, the parking
position function by using solenoids at the IP transport unit postscan, the fan
of the erasure unit and of the power unit and the voltage selection between
200V, 208V and 230/240V.
LEDs on the inner side of right door at the power unit show the status of the
supply voltages of the power unit:
GS100-MULTI-
Supply-BD
5146_reg02_008.cdr
Figure 9
Additional LEDs are visible if right door and power unit housing is open:
D19 – D32 red LEDs are corresponding to stepper motor fuses where they are
placed next to. The stepper motor supply is protected with a fuse on every
stepper motor board itself (12 in total). The status is shown with a green LED
on the side of the power unit (see figure 8).
D 33 green LED is corresponding to the fuse (Si13) of power supply of simple
switcher, which is placed next to the fuse.
Normal status:
The status of the LEDs, after switching on the digitizer with active interlock
switch and without any error, is:
Example:
5146_reg02_009.cdr
Sup
Figure 10
The failure of a lamp is detected by the current sense board, which is also part
of the power unit and is connected to the Erasure-Control-Board.
Figure 11
1 Lever
The Output Buffer has a capacity of 10 cassettes, like the Input Buffer. The
transport rollers at the gate move the cassette outside. To avoid blocking
when one cassette is following after another a lever pushes the cassettes
which already lay outside in the buffer aside.
The cassettes which are going out are detected and after the last one the
machine cycle stops.
3.5 cPCI-Rack
7
6
5
4
3 2
Figure 12
3.5.1 CPU-Ariel
Figure 13
The Scan Master Board controls the main functions of the digitizer which are
related to the image data acquisition. The board has to support two operating
levels, low resolution and high resolution. Therefore timing, frequency and
converters need to be controlled.
Polygon Control:
Connector (female) Controls via the Polygon the deflection of
to Polygon the laser beam on the image plate, controls
the rotational speed and detects the facets.
5146_reg02_013.cdr
Figure 14
Green LEDs to
monitor power supply
5146_reg02_014.cdr
Figure 15
During normal operation the LEDs monitor 5,1V and 3,3V voltage levels.
The power supply of the cPCI Rack is not interrupted by the interlock switch.
This independent supply makes it possible to update software or work out
other actions at the rack while both doors of the digitizer are open.
The gray rectangle in the partition map marks the directories which are only
necessary for remote control via http access.
Directories and
Partitions Files Annotations
subdirectories
*.BAT batch and
C: *.SYS
STARTUP.COM
system files
*.PRG program files
ERROR.MSG error message files
*.ESX tables
*.GIF
IMAGES *.JPG
*.TBL tables
*.SHD calibration files
*.SWP sweep curve
*.TEMP
temporary files
HTTPTEMP.OUT
E:
FLATFLD CONTROL.DAT
IMAGE.DAT
FLATFLD2 CONTROL.DAT
IMAGE.DAT
temporary directory
SCCTEMP 5146_reg02_015.cdr
Figure 16
4
2
3
Figure 17
The Cassette Unit 1 or 2, depends on which one is at the input buffer, fixes
and holds the cassette during the IP is taken out and scanned. An opener
mechanism for the cassette is also part of each unit. The Cassette Units 1and
2 together are supported and turned by the rotation unit.
On the inner side of each of the Cassette Units is a 5fold stepper motor board
placed, which controls the motors of the corresponding unit.
1
2
Figure 18
1 Vacuum unit
2 Suction arms
3 Guide plate
The IP transport robot takes the IP out of the cassette and leads it along the
guide plate to the optic module. The transport robot is guided in the frame of
the unit.
51
46
_re
g0
2_
01
8.C
D R
2
Figure 19
1 Rotational drive
2 Stepper motor board
Rotates the Cassette Units by 180° to bring the empty cassette of the input
side to the output side, for putting back the scanned and erased image plate
inside the cassette.
2
1
Figure 20
1 Photomultiplier
2 Optic module
3 Slow scan motor
4 5fold st scan
5 Holder for a second Photomultiplier
The stimulation of the image plate is done pixel by pixel, therefore the laser
beam is moved. The laser power on the focal plane is 50 mW + 5%. The
movement in the x-axis is achieved by a rotating polygon mirror (fast scan
direction) which is part of the optical path. The movement in the y-axis is done
mechanically by transporting the image plate (slow scan direction) with two
pairs of rolls.
Fa
s ts
ca
n dir
1 ec
tio
n
Figure 21
3.9.2 Polygon
The Polygon has 6 facets, each with a mirror (see 1 in figure 20). It is turned
by a DC motor which is clock pulsed fromthe scanning hardware. Pixel times
between 2.5 and 0.6 µs can be reached, but for scanning are two times used,
1 and 2 µs.
The amplification of the signal is set by the high tension voltage (HT) of the
photomultiplier. The high tension is calculated with two mathematical
constants MFA and MFB (machine factor A and B). The constants are part of
the settings of the digitizer and do not have to be changed in the field. The
high tension supply is integrated in the photomultiplier box.
The photomultiplier is equipped with a blue filter glass for separating the
emitted blue light from the stimulating red light. For self tests a blue LED is
mounted at the back side of the PM.
3
4
Figure 22
The lamps are controlled by the erasure control board and the CS board. The
lamps are controlled during work in pairs and also the power supply of 12 V is
generated via 5 single transformer. By initialization of the digitizer every lamp
is controlled separately. For erasing the image plate a power input of in total
1000 Watt with 230V / 50Hz is present.
The erasure unit fan (placed in the cPCI Rack) is switched on from the MS
board at the same time the lamps are switched on. After the last scanned and
erased image plate the fan is running a certain time for cooling down the unit.
The lamps are already switched on for 0.5 s before the IP reaches the erasure
window. The speed the IP is transported past the erasure unit is varied
between image plate begin and end for optimized homogeneity – dependent
on speed class and format (Erasure energy per time is determined by means
of the demographic data). An image plate format of 35 x 43 cm is transported
centered past the erasure front glass. The maximal distance from image plate
to erasure front glass is 20mm + 5mm.
2
Figure 23
1 Guide plate
2 Rocker for parking position
3 Suction arms
4 Vacuum unit
The IP transport robot leads the IP along the guide plate and puts the IP back
to the cassette in the cassette unit.
The transport robot is guided in the frame of the unit. To move the suction
arms back in working position without a conflict with the next IP coming out of
the scanner, the robot is moved in a parking position.
Figure 25
Figure 28
Figure 29
Figure 30
Figure 31
5146_REG02_030.CDR
Figure 32
Figure 33
Figure 34
Figure 35
Figure 36
Figure 37
Figure 38
Figure 39
Figure 40
Figure 43
Figure 44
5 Software Architecture
5.1 Diagnostics Software
The Diagnostics Software is a tool which helps to find defective parts in the
digitizer. A dialog with requests and questions via the display guides you
through the test of a module.
Input buffer and R/F-TAG Opens input door and moves cassette
to position 'silver dot detection'; tests
R/F-Tag reading and writing.
Identified cassette 35x43 cm is for
this test needed!
Cassette module 1 Turns cassette module 1 to right side;
opens, clamps and closes cassette;
turns cassette module 1 to left side
and shifts cassette out.
A cassette 35x43 cm is for this test
needed!
Cassette module 2 Turns cassette module 2 to right side;
opens, clamps and closes cassette;
turns cassette module 2 to left side
and shifts cassette out.
A cassette 35x43 cm is for this test
needed!
Rotation unit Turns cassette module 1 to right side
and then cassette module 2 to right
side.
Prescan transport+vac. Opens cassette on right side; tests
prescan valves, vacuum pump and
vacuum switch; moves robot up to
cassette module; sucks image plate
and releases it again; moves robot
down to scanner; closes cassette
again.
A cassette 35x43 cm with an image
plate inside is for this test needed!
Scanner Turns scanner rollers; moves pre-
/post-alignment drives and rollers-lift
home.
The Software of the digitizer allows access from the Internet to the service
level of the machine. The access via a graphic User Interface (ADC web
pages) is part of the service concept of MI CSO. The web pages follow the
structure of the well-known User Terminal.
Access via web pages guarantees a convenient and fast service by making
use of standard tools. The usage of different ports for these ways allows
working simultaneously with FTP, HTTP, and TelNet.
Spare parts which are sent with a floppy, need to be included in the system via
software. The floppy e.g. of the optic module contains specific settings of
production calibration and data of the spare part. This information need to be
installed after the spare part is mounted in the digitizer. For other spare parts
sent with a floppy only the information for the info counter is included.
The floppy is put into the floppy drive of the cPCI-Rack and the installation
starts automatically. The information concerning the spare part is integrated in
the system.
IP data
Cassette data Conversion of
Exposure emitted light
Destination into current
Patient
5146_reg02_056.cdr
Figure 46
7 Calibration Principle
There are two different calibrations to work out with the CR 75.0 in the field.
The IP Center Calibration, which is necessary after replacement of the optic
module or a change on the mechanical adjustment of the begin of scan
sensor, and the shading calibration which is responsible for the image quality
and follows always after an IP Center Calibration.
The IP center calibration adjusts two different scan signal settings in one
process, the begin of line (BOL) signal for the fast scan direction and the
begin of scan (BOS) signal for slow scan direction. Both signals in
combination are responsible for the area which is scanned and where pixels
are created.
7.1.1 BOL
The BOL sensor is a pin diode placed in the optic module. It controls the
switching off and on of the laser. The sensor is beside the IP, therefore the
laser beam hits the sensor before the IP. Out of the defined distance between
the sensor and the border of the IP the moment of reading out the pixel lines
on the IP is set.
7.1.2 BOS
The BOS sensor is a long distance light barrier which is placed at the bottom
part of the scan unit. The light barrier gives the signal to start the scan process
in slow scan direction. When the IP reaches the sensor it is detected and then
passes a fixed distance to the begin of scan line, then the scan process starts.
A number of lines of a flat field image are averaged and scaled to the
maximum pixel value. All the lines need to be in a certain range to make the
calibration successful, if not the error is displayed at the user terminal.
The y-axis of the coordinate system shows the Scan Average Level (SAL)
which is the digital value (gray scale) of a pixel and is one of the criteria for a
successful calibration. The x-axis shows the number of pixels.
CR85-X
Type 5148 / 100
CR75.0
Type 5146 / 100 (Up to SN < 6000)
Type 5146 / 105 (As of SN ≥ 6000)
► Please note
* Access the document by entering the document ID in the Agfa HealthCare Library search window,
or follow the direct link.
Edition 1, Revision 2
05-2011 printed in Germany Document Node ID: 27652068
service_bulletin_cover_e_template_v02
Agfa Company Confidential Copyright © 2011 Agfa HealthCare N.V.
DD+DIS099.09E Software / Hardware Compatibility
This page is intentionally left blank to enable print out of even and odd pages on duplex printers.
Chapter 3.1
List of Contents
1 Safety Precautions.................................................1
1.1 General Safety Instructions ............................................... 1
1.2 Safety Instructions for Laser Products............................. 4
1.3 Safety Instructions for Cassette Unit................................ 4
1.4 Cassettes............................................................................. 4
1 Safety Precautions
• The CR 75.0 has been designed for scanning medical X-ray image plates
and should only be used for these purposes.
• The CR 75.0 must only be operated by qualified staff trained on the
machine.
• Make sure that the CR 75.0 is constantly monitored in order to avoid
inappropriate handling, especially by children.
• Only trained service personnel must make repairs. Only authorized
service personnel must make changes to the CR 75.0.
• If there is any visible damage to the machine casing, do not start nor use
the CR 75.0.
• If you want to connect the CR 75.0 with other devices, components or
assemblies and if the technical data do not permit determining whether
the combination with these devices, components or assemblies involves
hazards, you must consult the respective manufacturers to avoid danger
for operating personnel or the environment.
• Do not override or disconnect the integrated safety features.
• Switch off the CR 75.0 before performing any maintenance work or
repairs. Disconnect the CR 75.0 from the mains before making repairs or
performing any maintenance activities.
• As is the case for all technical devices, the CR 75.0 must be operated,
cared for and serviced correctly.
• If you don't operate the CR 75.0 correctly or if you don't have it serviced
correctly, Agfa-Gevaert is not liable for resulting disturbances, damages or
injuries.
• When installing the CR 75.0, care must be taken to ensure that there is
either a mains plug or an all-cable disconnecting device in the internal
installation fitted near the CR 75.0 and that it is easily accessible.
• If you notice conspicuous noise or smoke, disconnect the CR 75.0
immediately.
• Check that the mains voltage is within the specified range of the self
adapting power supply of the machine.
• You can hurt your fingers if they are caught between the ADC Cassette
and the edge of the input slot. Insert the cassette in the input buffer as
described in the User Manual. At all times, keep your fingers clear of the
input slot. As soon as the CR 75.0 takes in the cassette, release it.
In order to reduce the risk of electric shock, do not remove any covers.
Caution hot:
Type B equipment:
Indicates that the CR 75.0 complies with the limits for type B equipment.
Intergrounding connector:
Provides a connection between the Digitizer and the protective earth of the
mains. Do not remove this connection, because this will have a negative
influence on the leakage current.
Power on
Power off:
Note that the power cord has to be disconnected from the wall outlet in order
to disconnect the unit entirely from the mains.
1.4 Cassettes
Observe great care whenever removing the image plate from the ADC
Compact cassette. Refer to the cleaning procedure described in the user
manual.
Make sure that the automatic exposure control device is placed above the
cassette, to prevent patients from receiving an overdose of X-rays. When it is
located underneath the cassette, the backscatter protection (lead) contained
in the red side of the cassette, retains a certain amount of X-rays. The dose
measured by the cell will then be much lower than the dose actually given to
the patient.
The image plate causes a specific X-ray scattering. This influences the
response of the exposure control device. To compensate for this, re-
calibration of the device for the use with ADC Compact cassettes could be
necessary.
2.1 General
Do not
Repairs in progress!
activate!
Location:......................
Repairs in Sign must only be removed by:
progress. ..................
Figure 1
The machine has one safety switch (interlock switch) to ensure the
customer's safety.
Nevertheless, it might be possible that in case of software errors the laser
beam is still active.
Figure 2
The power supply for all electronic components, except the PCI-Rack, is
interrupted when the Digitizer is opened. The PCI-Rack remains energized in
order to allow file uploads/downloads.
Consider that the service key must be removed again before the doors are
closed. After use of the service key, the function of the safety switch has to be
checked.
Do not perform any service interventions in the optical path of the digitizer.
3mm
5mm
Feed Gate
The Erasure Unit gets extremely hot during operation. Avoid contact, as there
is a risk of burning. Watch the corresponding label, see Figure 5.
200 c
Figure 5
Chapter 3.2
List of Contents
In addition to the common tools every service engineer is carrying in his tool
box, the following auxiliary equipment is required for the service on the
CR 75.0 components:
Test Images CD- CM+9 5145 3055 0 contains four test images:
ROM for CR 75.0 flatfld, flatfld2, testimg,
tsheet
1) Local keypad
local keypad Service PC The local keypad or operation terminal is part of the
digitizer and allows on-site access to the machine. All
terminal emulation commands are directly entered via 10 keys on the keypad.
(xt, hyperterminal) Easy to handle due to online-instructions.
ADC_C:\>Dir 2) Service PC
Digitizer The Service PC can be connected to the digitizer via the
serial interface RS232. Commands are entered in
command lines (CUI).
3) PC in LAN
Ethernet
The digitizer can be called from any PC connected to LAN.
TCP/IP TCP/IP Three different ways of access are offered:
PC in LAN Internet
Processing Station Input via TelNet: only one connection at the same time
PPP
or
ID-Station
adc_C:\> dir
Plus
TelNet LAN
Remote PC Local Area Network
CUI GUI Several PCs are connected to
each other in a limited area.
Plus
Plus
ftp> adc_C:\>Dir
figure 1
* The optical fiber has been omitted, but the presence of the laser beam can
still be validated by visual inspection through the opened input rollers.
Mind the safety instructions in chapter 3.1.
(3) Enter the IP address of the digitizer in the URL address field of your
browser.
(4) You will be asked to enter a user name and a password. If you do not
know the proper password, please contact GSC Munich.
(5) HTML pages will open with the Welcome Page on your screen.
Plus
Plus
figure 2
Navigation Instructions:
HTML pages are structured in a three-frame layout. The top frame is static
and displays information about the digitizer and its location. The lower left
frame is the navigation frame and displays a set of topics. Information about a
selected topic will open in the main frame.
In the main frame most of the information pages are equipped with a
"Home" link and a "Back" link. "Home" leads back to the Welcome Page.
"Back" returns to the primarily selected page.
NOTE:
When selecting files for downloads, these files are first zipped and copied to
the download area (adc_c:\dl\). From the download area, the zip-files can be
downloaded to a directory of your HDD.
User Terminal Language The standard language for normal user and
key operator of the system can be changed.
NOTE:
Changes are activated immediately.
NOTE:
The following item ‘Advanced Access’ allows access via the command line. It
is for advanced users only. Protective mechanisms are no longer in force.
Chapter 3.3
List of Contents
Checkpoint 2
Cassette unit
Checkpoint 3
Rotation unit
Checkpoint 7
Postscan transp. Checkpoint 1
Input buffer
Checkpoint 4
Prescan transp.
Checkpoint 6 & Vacuum
Erasure unit
Checkpoint 5
Scanner
The error numbers of the internal reports are not identical to the error codes
shown on the display.
Ok
Test if X positioning
lightbarrier of motor
M513 is interrupted Error xxxx ocurred when
testing X positioning light
Ok barrier of motor M513
Test if 0 positioning
lightbarrier of motor
Ok M514 is clear Error xxxx ocurred when
testing 0 positioning light
barrier of motor M514
Ok
Test if X positioning
lightbarrier of motor
M514 is interrupted Error xxxx ocurred when
testing X positioning light
barrier of motor M514
Waking up RF-tab reader
Wake up RF-tag reader
failed
Ok
Order RF-tag reader to Writing data to tag failed Ok
write data back to tag
Ok
CASMODCON
ROTCON
Ok
Ok
Ok
Ok
Ok
Ok
Ok
Stopping input rolls of Stopping input rolls of
cassette module x cassette module x failed
Ok
Lift of cassette module x Moving down lift of
is being moved down cassette module x failed
Ok
Unclamping cassette of Unclamping cassette of
cassette module x cassette module x failed
Ok
Ok
Opening cassette in Opening cassette in
module x module x failed
Ok
Closing cassette in Closing cassette in
module x module x failed
Ok
Unclamping cassette in Unclamping cassette in
module x module x failed
Ok
Moving up lift of cassette Moving up lift of cassette
module x module x failed
Ok
Lift of cassette module x
is moved up
Checkpoint
Turning cassette module 2 Turning cassette module
1/2 to input 1/2 to input failed
Ok
Test connections of
rotation module
Test if clear and interrupt
Move rotation unit home Moving rotation unit functionality of 0
home failed positioning lightbarrier
Ok of motor Mxxx
Ok
Initialize rotation module Initialization of rotation
module failed
Ok
Ok
Test connections of
prescan module
Initializing prescan
module
Test connections of
cassette module Ok
Ok
Unclamp clamper of Unclamping clamper of
cassette module x cassette module x failed
Ok
Move up lift of Moving up lift of
cassette module x cassette module x failed
Ok
Move home belt of Moving home belt of
cassette module x cassette module x failed
Ok
Stop input rolls of Stopping input rolls of
cassette module x cassette module x failed
Ok
Start input rolls of Starting input rolls of
cassette module x cassette module x failed
Ok
Stop input rolls of Stopping input rolls of
cassette module x cassette module x failed
Ok
Ok
Clamp cassette of Clamping cassette of
cassette module x cassette module x failed
Ok
Open cassette of Opening cassette of
cassette module x cassette module x failed
Ok
Ok
Start of moving up
prescan robot failed
Ok
Prescan robot started
to move up
Ok
Ok
Ok
Finishing cycle of Finishing cycle of
prescan module prescan module failed
Ok
Ok
Unclamp clamper of Could not unclamp
cassette module x cassette of cassette
module x
Ok
Move up lift of Could not move up lift of
cassette module x cassette module x
Subprogram VACUUM:
Ok
Ok
Ok
Switch on vacuum
pump prescan/postscan
module
Ok
Ok
Ok
Ok
Ok
Ok
User interference:
remove tube
Ok
Test connections of
slow scan module
Ok
Stopping slowscan
rollers failed
Ok
Ok
Move scanner motors Moving scanner motors Moving home pre-align Moving home pre-align
to home position into to home position slow scan drive slow scan drive failed
failed
Test connections of
erasure unit
Ok
Test transformer 1
Ok
Test transformer 2
Ok
Test transformer 3
Ok
Test transformer 4
Ok
Test transformer 5
Ok
LAMPS
Subprogram LAMPS :
Ok
Test connections of
postscan module
Initializing postscan
module
Timeout
Timeoutwhen
whentesting
testingXifif
Timeout when testing X if positioning lightbarrier
puls lightbarrier of of
positioning lightbarrier of motor M403
subnode is interrupted
2 of GS414 is
motor M400 is clear Test if lightbarrier GS414 interrupted
is interrupted
Error xxxx ocurred when Error xxxx ocurred when
testing X positioning light testing
testingXpulspositioning light
lightbarrier
barrier of motor M400 Ok ofbarrier
subnode2of motor M403
of GS414
is clear isisinterrupted
interrupted
Move home suction
motor of postscan
Ok module
Ok
Ok
Test connections of
rotation module
Test connections of
cassette module x
Ok
Unclamp clamper of Unclamping clamper of
cassette module x cassette module x failed
Ok
Move up lift of Moving up lift of
cassette module x cassette module x failed
Ok
Move home belt of Moving home belt of
cassette module x cassette module x failed
Ok
Stop input rolls of Stopping input rolls of
cassette module x cassette module x failed
Ok
Start input rolls of Starting input rolls of
cassette module x cassette module x failed
Ok
Stop input rolls of Stopping input rolls of
cassette module x cassette module x failed
Ok
Move down lift of Moving down lift of
cassette module x cassette module x failed
Ok
Open cassette of Opening cassette of
cassette module x cassette module x failed
Ok
Ok
Close opener of Could not close cassette
cassette module x of cassette module x
Ok
Unclamp clamper of Could not unclamp
cassette module x cassette of cassette
module x
Ok
Move up lift of Could not move up lift of
cassette module x cassette module x
Ok
Stop input rolls of Stopping input rolls of
cassette module x cassette module x failed
Ok
Start input rolls of Starting input rolls of
cassette module x cassette module x failed
Ok
Stop input rolls of Stopping input rolls of
cassette module x cassette module x failed
Ok
Move down lift of Moving down lift of
cassette module x cassette module x failed
Ok
Unclamping cassette of Unclamping cassette of
cassette module x cassette module x failed
Ok
Clamp cassette of Clamping cassette of
cassette module x cassette module x failed
Ok
Open cassette of Opening cassette of
cassette module x cassette module x failed
Ok
VACUUM
Ok
Waiting for postscan Postscan robot did not
robot to reach upper reach upper position
position
Ok
Ok
Ok
Moving postscan robot
down to interrupt suction
X-POS lightbarrier
failed
Ok
Ok
Ok
Ok
Test connections of
cassette module x
Ok
Unclamp clamper of Could not unclamp
cassette module x cassette of cassette
module x
Ok
Move up lift of Could not move up lift of
cassette module x cassette module x
Ok
Test connections of
cassette module x
Chapter 3.4
List of Contents
GS IOB-SIN-5Step Boards:
CM+9.9499.8120.3
1 8
LJGS210sn2.cdr
JL 2 JL2
1 8
LJGS210sn1.cdr
1 8
LJGS260sn2.cdr
JL2 JL2
1 8
LJGS260sn1.cdr
1 8
LJGS610sn2.cdr
JL2 JL2
1 8
LJGS610sn1.cdr
GS IOB-SIN-Step Boards
CM+9.9499.8140.4
1 8
LJGS202.cdr
1 8
LJGS412.cdr
1 8
LJGS418.cdr
1 8
LJGS428.cdr
Board 1 8
LJGS434.cdr
1 8
JLGS510.cdr
1 8
LJGS532.cdr
1 8
LJGS552.cdr
1 8
LJGS556.cdr
2 Fuses Overview
Chapter 3.5
List of Contents
7 cPCI-Rack .............................................................25
7.1 Replacing a defective Hard Disk (HDD) ...........................25
7.2 Exchange of the Hard Disk / Procedure to
install ACP_3101 on a virgin Hard Disk ...........................28
1 Safety Note
Switch off the digitizer before performing any service interventions at the
digitizer.
Figure 1
Figure 2
Figure 3
Figure 4
Figure 6
Figure 7
Figure 8
Figure 9
Figure 10
Figure 11
Figure 12
Figure 13
Figure 14
By repositioning the transport unit take care of the guiding bar on the bottom
of the digitizer.
Figure 15
Figure 16
Figure 17
Figure 18
Figure 19
Figure 20
Figure 21
5.3 Replacing the 5fold Stepper Motor Board at the Scan Unit
(1) Pull out the scan unit to its
stop.
Figure 22
Make sure that the digitizer is switched off, before you remove the optic
module.
Figure 23
Figure 24
Locking clips
prevent the plug from working
loose.
Figure 25
Figure 26
Figure 27
Figure 28
Figure 29
Figure 30
Figure 31
(3) An adjustment of
BOL/BOS settings and a
shading calibration must
be done after installation
(see chapter 3.6).
• Check the image quality with test sheet and flatfield exposures, see
chapter 3.6.
• Update the backup to save the calibration data.
• Select <3 Save on floppy>
<2 Machine specific data>
• Follow the on-line instructions.
• Enter the S/N number of the Digitizer on the label of the photomultiplier
floppy disk.
Therefore slide out the scan unit (see 5.1) and remove first this modules and
parts as described before:
• Photomultiplier Tube (PMT) (see 5.2)
• 5fold stepper motor board at Scan Unit (see 5.3)
• Optic Module (see 5.4)
Figure 32
Figure 33
(3) Take out all cables of the
clamps
(4) All connections are
disconnected now and the
scan unit can be moved out
of the digitizer completely
In order to remove the scan unit safely, use the ramp which has come with
the digitizer.
514601ei.cdr
1. The ramp for the scan unit is hidden behind the left side panel.
2. It is needed to remove the scan unit from the digitizer.
• Hook the ramp into the gaps between the two feet of the digitizer.
• Let the scan unit slide onto the ramp and then smoothly to the floor.
Mind to fill out the Defect Label in a correct and detailed way for the old
scan module!
Figure 34
(4) Slide back the scan unit onto
the ramp and into the
digitizer.
6.1 Preparations
514635AA.CDR
Figure 35
Figure 36
Figure 37
Figure 38
Overview: Cassette
module 2 with
complete cabling.
Figure 39
Figure 40
Figure 41
Figure 42
Figure 43
Figure 44
12
Figure 45
Figure 46
7 cPCI-Rack
Switch off the digitizer and disconnect the machine from the mains.
Safety instructions for electronical parts see Chapter 3.1
In order to avoid any damage at the storage board, replace the HDD very
cautiously.
Figure 47
Figure 48
Figure 49
Figure 50
1 Storage board
2 Connector
3 Terminal rail
4 Ribbon cable for hard disk
5 Ribbon cable for floppy disk
6 Hard disk
7 Floppy disk drive
8 Connector
Figure 51
Figure 52
Hold the hard disk firmly in one hand (see Figure 52), while you remove the
four round-head screws with a mid-size screw driver.
Figure 53
• Take the Hard disk formatter floppy of ACP_3101 and insert it into the
floppy disk drive of the digitizer.
• Disconnect the digitizer from the network (since a default IP-Address &
hostname is configured)
• Connect a PC with terminal program or a Terminal to service RS232 port
• Press the Reset button on Ariel-CPU or turn power off/on of the digitizer
• Confirm partitioning/formatting of the hard disk with ‘y’.
• Wait until you get the final message on a terminal connected to the
service RS232 port or wait until control LED of floppy disk drive is off.
• Insert the floppy ACP_3101 Portex Installer 2/2 into the disk drive of
the digitizer
• Press the Reset on Ariel-CPU or turn Power Off/On of the digitizer
• Wait until you get the final message on a terminal connected to the
service RS232 port or wait until control LED of floppy disk drive is off.
• Remove floppy (don't insert any other floppy)
• Press the Reset on CPU-Board or turn Power Off/On the digitizer
• Wait until you get the final message on a Terminal connected to the
service RS232 port, or wait approximately 2 minutes.
• Insert the floppy ACP_3101 Release Disk 1/3 into the disk drive of the
digitizer
• Press the Reset button on Ariel-CPU or turn power off/on of the digitizer
• Wait until you get the final message on a terminal connected to the
service RS232 port, or wait until control LED of floppy disk drive is off.
• Proceed the same way with the ACP_3101 Release Disk 2/3 and 3/3
• Press the Reset button on Ariel-CPU or turn power off/on of the digitizer
and wait until system is up
Then parse the CPF-File using the guidelines of the Service Menu Topic
Install from floppy, CPF-File, displayed on the user interface.
• If the previous software had a lower version than ACP_1201, then close
the machine doors and scan or erase one cassette.
This cassette needs not be exposed.
For an upgrade of a lower version than ACP_1201 to ACP_3101 a new IP
center calibration is necessary, because of a changed IP stop position before
start of scan. (Calibration procedure, see chapter 3.6 of your Technical
Documentation)
• Make a new backup (therefore you find an empty floppy within the set of
delivered floppies)
Chapter 3.6
List of Contents
2 Mechanical Adjustments.........................................2
2.1 Adjustment of Rotation Unit ................................................ 2
2.2 Adjustment of the Guiding Plate in Postscan Unit ............ 7
2.3 Adjustment of Suction Cups...............................................10
2.4 Adjustment Cycle of Transport Robot ...............................11
3 Optical Adjustments..............................................14
3.1 Possible Reasons for a Calibration....................................14
3.2 Preparation of Calibration Procedures ..............................15
3.3 IP-Center Calibration (BOL/BOS Adjustment) ...................17
3.4 Shading Calibration .............................................................18
5 Backup....................................................................22
Adjustments
mechanical optical
see 2 see 3
Rotation Unit
see 2.1 IP Center
Calibration
see 3.1
Guiding Plate
see 2.2
Shading
Calibration
Suction Cups see 3.2
see 2.3
Adjustment Cycle
see 2.4
2 Mechanical Adjustments
2.1.1 General
For orientation the two modules of the Cassette Unit are marked with stickers
“Module 1” and “Module 2” on the front side. Adjustment needs to be done if
the light barrier flag is misadjusted and as a result the suction cups do not
suck the IP properly.
Carry out this adjustment only if it is sure that the rotation unit is misadjusted,
see check at 2.1.3. As normally the light barrier flag is preadjusted in
production and secured by a positioning bracket.
The rotation unit turns around very fast! Watch your head and hands!
OK
OK
(2) Select
<7 DIAGNOSTICS>
(3) Select
<2 Mechanical modules>
(6) Check and press the confirm key Check that nothing blocks Service menu
DIAGNOSTICS
rotation of cassette
modules (cassettes, pre- Cassette mod.
/postscan transport...) : ok
: exit
514536ba.cdr
514536bb.cdr
(8) Press right arrow key → to exit Note: Rotation motor is Service menu
currentless now. Thus you DIAGNOSTICS
can turn cassette modules Cassette mod.
by hand and open them to : ok
watch mechanics if : exit
necessary.
514536bc.cdr
514536bd.cdr
1
Cassette Unit 2
3
514535AA.CDR
Figure 1
5145_109.02_aa.CDR
Figure 2
Figure 3
Make sure that the position of both cassette modules are correct before you
perform an Adjustment Cyle (see 2.4 of this Chapter).
2.2.1 General
If a guiding plate is not in perfect
position the IP may be misguided. The
robot carriage cannot insert the IP in
the cassette correctly.
Figure 4
1 2 3
Guiding plate too far left: Guiding plate too far right: Guiding plate correct:
IP will miss the cassette IP will miss the cassette IP will enter the cassette
Cassette
IP
Guiding plate
Figure 5
Figure 6
(5) During the last 10 cm of the
upward movement the space
between the white plastic
gearwheel and the guiding plate
has to be examined
Figure 7
(8) If the guiding plate touches the
plastic gear wheel, touch the
bottom ledge of the guiding plate
and move the guiding plate away
from the plastic gear wheel
Figure 8
Figure 9
Before an adjustment cycle is carried out, make sure that both cassette
modules of the rotation unit are adjusted correctly (see 2.1 of this Chapter).
(Shown display menus are based on Software Version ACP_1104 and may
vary with further Software updates.)
(4) Select
<5 Adjustment cycle>
2 mm
(7) Check the horizontal position of the ADC Cassette
IP inside the Cassette by moving
the IP carefully by hand while the
suction cups are still holding the IP. Image Plate
2 mm
tendency to the right side of the
Cassette (towards the front of the
digitizer). Tolerance is 2 mm
Figure 10
Figure 11
(9) Select
<2 MAINTENANCE>
514666ca.cdr
3 Optical Adjustments
The GenRad X-Ray source can be used for both, the GenRad calibrations
and the Mammo calibration.
For each calibration procedure you need a copper filter and an erased image
plate, which is dry, clean and immaculate. The cassette with the image plate
inside should not be identified. If it is identified, an additional confirmation on
the display is needed.
For the IP center calibration an exposed cassette with a size smaller than 35
x 43 cm is needed.
For the shading calibration three different steps are necessary:
a) GenRad Calibration High Resolution
an exposed cassette only with 35 x 43 cm size format must be used (for
detecting the borders of the IP).
b) GenRad Calibration Standard Resolution*
an exposed cassette only with 35 x 43 cm size format must be used (for
detecting the borders of the IP).
c) Mammo Calibration
an exposed cassette with 35 x 43 cm size format can be used, but it is
also possible to use a min. 24 x 30 cm cassette (Mammo cassette also
possible).
* The Standard Resolution calibration is also scanned in HiRes mode, only the
applied pixel time is different.
• 7.5 mAs
• 75 kV
• 1.3 m distance
• large focus
• 1.5 mm copper filter
Note:
The entire image plate
must be fully exposed!
No collimation is needed!
Figure 13
(3) Select
<2 Calibration>
(4) Select
<2 IP center calibration>
• Calibration successful
• Failure – underexposed!
In this case, increase exposure dose and redo IP center calibration.
• Failure – overexposed!
In this case, decrease exposure dose and redo IP center calibration.
• Calibration failed!
• Big IP used!
In this case, use a format with a width smaller than 35 cm.
• No flat field?!
• Damaged IP edges?!
• Begin of IP not found!
• IP skewed!
• IP-center out of tolerance
(2) Select
<2 MAINTENANCE>
(3) Select
<2 Calibration>
(4) Select
<1 Shading Calibration>
5 Backup
• Update the backup to save the new machine specific data (calibration,
PD-factor).
• Select <3 Save on floppy>
<2 Machine specific data>
• Follow the on-line instructions.
CR 85-X
Type 5148/100
CR 75.0
Type 5146/100 (Up to SN < 6000)
Type 5146/105 (As of SN ≥ 6000)
* Previously supplied software versions (prior to version ACP_4005) are no more listed in this document.
► Document History
► Referenced Documents
Document Title
DD+DIS233.08E CR Mammography Solution, Service Bulletin No. 37
Edition 1, Revision 0
07-2010 printed in Germany Document Node ID: 12099766
eq_standard_e_template_v07
Agfa Company Confidential Copyright © 2010 Agfa HealthCare N.V.
DD+DIS207.10E Repair and Service
► Manufacturer
Agfa HealthCare N.V.
Publisher
Agfa-Gevaert HealthCare GmbH
Tegernseer Landstraße 161
D - 81539 München
Germany
Copyright 2010 Agfa HealthCare N.V.
All rights reserved.
Technical modifications reserved.
AGFA and the Agfa-Rhombus are trademarks of Agfa HealthCare N.V.
WARNING:
Improper operation or service activities may cause damage or injuries.
INSTRUCTION:
(1) Read the "Generic Safety Directions" document
(see MEDNET GSO => General Info => Agfa HealthCare => Publications =>
Service Manual) prior to attempting any operation, repair or maintenance task on
the equipment.
(2) Strictly observe all safety directions within the "Generic Safety Directions" and
on the product.
NOTE:
To verify the latest version of single documents and of Service Manuals refer to the
Document Type ‘Order List’ in the GSO library.
LIST OF CONTENTS
1.2 Changes in Service and Customer functionality since Software Version ACP_5102 ..............4
2.2 Changes in Service and Customer functionality since Software Version ACP_5007 ..............5
Software was adapted in order to avoid issues during the sensitivity adjustment done in
the production line.
Mammo cassettes are not closed while turning the cassette module with the
empty cassette. This prevents possible damage of the fleece of the new type
of Mammography cassettes.
In the screen which is displayed after a shading calibration failed, was no option
to skip this menu. Now the <Chancel> bottom is available to skip this screen.
Support of the new revive-racks for CR 85-X with USB interface instead of
floppy disk drive.
HQ_0511110001: Improved handling of RF-Tags. The software will not write
anymore data in protected sectors of the RF-Tag.
New screens of the shading calibration are translated in the different languages
(Version ACPL5007 of the language floppy).
New version of operating system “portex 2006.02”
Modified menu:
o Specific terms like “floppy“ are replaced by e.g. “storage medium”.
o New version of language files
Support of the new revive-racks for CR 85-X with USB interface instead
of floppy disk drive.
Modified calibration procedure:
o IP-Center and Shading calibration is also possible with cassettes
exposed on mammography X-ray modalities.
Modified user interface (Service Menu):
o Service menus for calibrations are adapted to the menus of
CR 25.0 / CR 35-X.
o Improved shading calibration screens
o New version of language files
o Specific terms like “floppy” are replaced by e.g. “storage medium” or “data”.
Chapter 4
List of Contents
A Identification Diagrams
100
500
100
400
600
2.1 Diagram
MC101
GS204
GS314
ST319
1 8
M212
JL1
GS202
GS206
3.1 Diagram
Si1 GS100
Si2
Si3
Si4
Si6
Si7
TR106 Si8
Si9
S103 Si10
Si11
M101 Si12
Si13
GS102
TR105
GS104
S102
TR104
TR103
TR102 514607af.cdr
TR101
4.1 Diagram
8
JL1
GS210
JL2
1
1
GS220
GS230 GS216
GS214 M205
M202
M201
M203
M204
GS222
GS226
GS224
514607ak.cdr
GS228
5.1 Diagram
8
JL1
GS260
JL2
1
1
GS270
GS280 GS266
GS265 M210
M207
M206
M208
M209
GS278
GS276
GS274
514607al.cdr
GS277
6.1 Diagram
M701
M300
GS302
GS304
GS300
GS310
GS312 GS306
7.1 Diagram
MG422
JL1 GS420
1
514607aj.cdr
GS434
GS432
M403 GS426
GS424 M402
GS428
1 JL1 8
8.1 Diagram
MG411
GS402
MG402
GS406
MG401
GS404
JL1 1
GS412
M400 GS410
GS414 GS408 M401
GS416
GS418
8 JL1 1
9.1 Diagram
JL1
1 8
GS514 GS510
GS518
M514
GS516
M513 GS512
GS532
JL1
GS536
1
GS538
GS540
10.1 Diagram
JL1
1 8
GS552
M516
GS554
GS560
GS562
GS556 GS558 M515 GS564
514607ao.cdr
1 JL1 8
11.1 Diagram
GS570 GS568
GS566
12.1 Diagram
8
8
JL2
JL1
GS610
1
1
GS622 GS620
GS616
M612
GS612
M622
M702
13.1 Diagram
M621
GS628
GS618
GS614 GS626
14.1 Diagram
LA7
LA8
LA9
LA10
LA1
LA2
LA3
LA4
LA5
LA6
F1.5146.4005.0 (sheet 1 – 7)
CR75.0
Type 5146/101/105, 5148/100
CR 85-X CR 75.0
Type 5148/100 Type 5146/101/105
Internal update: 4
CONFIDENTIALITY NOTE:
Use, dissemination, distribution or reproduction of this document by unauthorized personnel is not permitted and may be unlawful.
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Manufacturer
Agfa HealthCare N.V
Published by
Agfa-Gevaert HealthCare GmbH
Tegernseer Landstraße 161
D - 81539 München
Germany
WARNING:
INSTRUCTION:
(1) Read the "Generic Safety Directions" document
(see MEDNET GSO => General Info => Agfa HealthCare => Publications =>
Service Manual) prior to attempting any operation, repair or maintenance task on
the equipment.
(2) Strictly observe all safety directions within the "Generic Safety Directions" and on
the product.
WARNING:
INSTRUCTION:
• Replace defective parts with Agfa HealthCare original spare parts.
• Use only tools and measuring instruments which are suitable for the procedure.
• Only approved Agfa HealthCare accessories must be used. For a list of
compatible accessories contact your local Agfa HealthCare organization or
www.agfa.com.
NOTE:
To verify the latest version of single documents and of Service Manuals refer to the
Document Type ‘Order List’ in the GSO Library.
Document History
NOTE:
Agfa reserves the right to deliver compatible parts or alternative parts.
Prices to the original order may differ.
Contact
Spare Parts ordering
Europe orderprocessing-europe.matrium@eads.com
Overseas orderprocessing-overseas.matrium@eads.com
Worldwide returns.matrium@eads.com
NOTE:
For Recycling Information please refer to:
http://intra.agfanet/cd/ep/ehs.nsf
Contents
AUSGABEPUFFER
OUTPUT BUFFER
RESERVE DE SORTIE EINGABEPUFFER
PAGE 10 - 15 INPUT BUFFER
RESERVE D'ENTREE
PAGE 16 - 21
TÜR LINKS
DOOR LHS
PORTE A GAUCHE
PAGE 36 - 37
CPCI-RACK, REVIVE RACK
CPCI-RACK, REVIVE RACK TÜR RECHTS, LEISTUNGSEINHEIT
CPCI-RACK, REVIVE RACK DOOR RHS, POWER UNIT
PAGE 38 - 41 PORTE A DROITE, BLOC DE PUISSANCE
PAGE 42 - 45
KASSETTENTEIL
CASSETTE PART
PARTIE CASSETTE
PAGE 22 - 35 ROTATIONSANTRIEB
ROTATION DRIVE
COMMANDE ROTATIVE
LÖSCHEINHEIT
PAGE 76 - 77
ERASURE UNIT
UNITE D'EFFACEMENT
PAGE 46 - 47
KASSETTEN
CASSETTES
CASSETTES
PAGE 80 - 101
FOLIENTRANSPORT, EINGABE
IP TRANSPORT, INPUT
TRANSPORT DE L'ECRAN, ENTREE
FOLIENTRANSPORT, AUSGABE PAGE 54 - 59
IP TRANSPORT, OUTPUT
TRANSPORT DE L'ECRAN, SORTIE SCANMODUL
PAGE 48 - 53 SCAN UNIT
MODIFIKATIONSBAUGRUPPEN / ZUBEHÖR MODULE SCAN
MODIFICATION KITS / ACCESSORIES PAGE 60 - 75
KIT DE MODIFICATION / ACCESSOIRE
PAGE 78 - 79 5148_chap05_INHALT.cdr
02
19*
17*
01*
18
D2853
D635
D6426
D3088
16*
02 03
07*
X
15*
05
08*
13 09*
14*
Detail X D6408
D6426
06
04 11*
12* 10*
12*
5148_chap05_01_m.CDR
= Assembly
D = Standard part. For order please refer to the separate spare parts list for standard parts "DD+DIS011.93M".
* = Spare part standardly is not kept in stock, extended delivery time is possible.
99 = Spare Part is not shown in Exploded Views.
01
(GS 314)
06
05 (GS 204) 02
07
(St319)
04
D708
08
D117
D5682 03
03
D5682
D117
5148_chap05_03.CDR
= Assembly
D = Standard part. For order please refer to the separate spare parts list for standard parts "DD+DIS011.93M".
* = Spare part standardly is not kept in stock, extended delivery time is possible.
99 = Spare Part is not shown in Exploded Views.
01 02
D217
D2514
D217
D2853
07
02
06
(GS 564)
04
(M515)
05 03
5146_chap05_04.CDR
= Assembly
D = Standard part. For order please refer to the separate spare parts list for standard parts "DD+DIS011.93M".
* = Spare part standardly is not kept in stock, extended delivery time is possible.
99 = Spare Part is not shown in Exploded Views.
01
(M516)
02
D579
D2514
02
D2861
03
(GS562) (GS560)
04
06 (GS554)
05
D6522
D6522
05 5148_chap05_05.CDR
= Assembly
D = Standard part. For order please refer to the separate spare parts list for standard parts "DD+DIS011.93M".
* = Spare part standardly is not kept in stock, extended delivery time is possible.
99 = Spare Part is not shown in Exploded Views.
08
01
(GS 552)
02
10
05 04
03
01
(GS 556)
02
09
07 08
03
(GS 558)
06
04
5148_chap05_06.CDR
= Assembly
D = Standard part. For order please refer to the separate spare parts list for standard parts "DD+DIS011.93M".
* = Spare part standardly is not kept in stock, extended delivery time is possible.
99 = Spare Part is not shown in Exploded Views.
01
02
05 D2513
(GS518)
D2763
D2861
D417
D2514
D2861
D2514
D579
D6522
03
04
(M514)
D372
5148_chap05_07.CDR
= Assembly
D = Standard part. For order please refer to the separate spare parts list for standard parts "DD+DIS011.93M".
* = Spare part standardly is not kept in stock, extended delivery time is possible.
99 = Spare Part is not shown in Exploded Views.
06
01
04
03
(GS512) 10
09
05 07 08
(GS510)
(GS516)
02 04
D6522 (GS514)
04
5148_chap05_08.CDR
= Assembly
D = Standard part. For order please refer to the separate spare parts list for standard parts "DD+DIS011.93M".
* = Spare part standardly is not kept in stock, extended delivery time is possible.
99 = Spare Part is not shown in Exploded Views.
16 01
(GS530) 02
(GS534) 03
15 04
(M513)
D333
12
D579 D369
D2514
D542
14
(GS532)
13 10
12
08 09
11
04
07
05
(GS540)
06
(GS536, GS538) D2513
D369
5148_chap05_09.CDR
= Assembly
D = Standard part. For order please refer to the separate spare parts list for standard parts "DD+DIS011.93M".
* = Spare part standardly is not kept in stock, extended delivery time is possible.
99 = Spare Part is not shown in Exploded Views.
05
D6522 D6522 (M202)
(M207)
D579
5141
D2681
04
01
D6522
D3154
03 D333
(GS214)
(GS265)
02
D6522
5148_chap05_10.CDR
= Assembly
D = Standard part. For order please refer to the separate spare parts list for standard parts "DD+DIS011.93M".
* = Spare part standardly is not kept in stock, extended delivery time is possible.
99 = Spare Part is not shown in Exploded Views.
06
04
05*
03
02
(GS220, GS270)
01*
(M205, M210)
5148_chap05_11.CDR
= Assembly
D = Standard part. For order please refer to the separate spare parts list for standard parts "DD+DIS011.93M".
* = Spare part standardly is not kept in stock, extended delivery time is possible.
99 = Spare Part is not shown in Exploded Views.
06
(M204)
(M209)
01
05
03
02
(GS228)
(GS277)
04
5148_0100_12.CDR
= Assembly
D = Standard part. For order please refer to the separate spare parts list for standard parts "DD+DIS011.93M".
* = Spare part standardly is not kept in stock, extended delivery time is possible.
99 = Spare Part is not shown in Exploded Views.
01 (M201)
(M206)
02
05
(GS230)
(GS280) 04
03
D3083
D2995
D320
07
D377
06
D6522
(M203)
(M208)
5148_chap05_13.CDR
= Assembly
D = Standard part. For order please refer to the separate spare parts list for standard parts "DD+DIS011.93M".
* = Spare part standardly is not kept in stock, extended delivery time is possible.
99 = Spare Part is not shown in Exploded Views.
01
02 (GS216)
(GS266)
04
05
06
07 09
07
08
(GS226, GS276)
07
5148_chap05_14.CDR
= Assembly
D = Standard part. For order please refer to the separate spare parts list for standard parts "DD+DIS011.93M".
* = Spare part standardly is not kept in stock, extended delivery time is possible.
99 = Spare Part is not shown in Exploded Views.
01
02 03 04 05 03 06 07 08
12 11 10 09
(GS210, GS260)
5148_chap05_15.CDR
= Assembly
D = Standard part. For order please refer to the separate spare parts list for standard parts "DD+DIS011.93M".
* = Spare part standardly is not kept in stock, extended delivery time is possible.
99 = Spare Part is not shown in Exploded Views.
05
04
06
03* 09*
02
01
08
07*
5148_chap05_16.CDR
= Assembly
D = Standard part. For order please refer to the separate spare parts list for standard parts "DD+DIS011.93M".
* = Spare part standardly is not kept in stock, extended delivery time is possible.
99 = Spare Part is not shown in Exploded Views.
D5682
D117 D706
02 04
(M701)
D5682
D117
D2681
D335
D6522
03
01
D706
5148_chap05_17.CDR
DOOR LHS
= Assembly
D = Standard part. For order please refer to the separate spare parts list for standard parts "DD+DIS011.93M".
* = Spare part standardly is not kept in stock, extended delivery time is possible.
99 = Spare Part is not shown in Exploded Views.
DOOR LHS
01
(M300)
02
04
03
(GS302)
(GS304)
05
09
(GS300)
08
(GS312)
06
07 (GS306)
(GS310)
CR85-X
Type 5148/100 (up to SN <4000)
CR75.0
Type 5146/101
Type 5144/105 (up to SN <6500) 5148_chap05_18.CDR
CPCI-RACK
= Assembly
D = Standard part. For order please refer to the separate spare parts list for standard parts "DD+DIS011.93M".
* = Spare part standardly is not kept in stock, extended delivery time is possible.
99 = Spare Part is not shown in Exploded Views.
CPCI-RACK
01
02
07**
03
06
05 04
REVIVE RACK
= Assembly
D = Standard part. For order please refer to the separate spare parts list for standard parts "DD+DIS011.93M".
* = Spare part standardly is not kept in stock, extended delivery time is possible.
99 = Spare Part is not shown in Exploded Views.
REVIVE RACK
05 D5682
D117
D703 07
D706
06
02
(M101)
03
(TR106) 04*
01
D321 D854
D855 D865
D539
5148_chap05_20.CDR
= Assembly
D = Standard part. For order please refer to the separate spare parts list for standard parts "DD+DIS011.93M".
* = Spare part standardly is not kept in stock, extended delivery time is possible.
99 = Spare Part is not shown in Exploded Views.
04
(S103)
05 09
12
(GS104)
01
(S102)
07
07 08
(GS100)
02
06 10
02, 03
11
(GS102)
10 06 5148_chap05_21.CDR
= Assembly
D = Standard part. For order please refer to the separate spare parts list for standard parts "DD+DIS011.93M".
* = Spare part standardly is not kept in stock, extended delivery time is possible.
99 = Spare Part is not shown in Exploded Views.
01 05
06
08
D221 07
D353
D81 D221 07
02
03
04
09
10
11
5148_chap05_22.CDR
ERASURE UNIT
= Assembly
D = Standard part. For order please refer to the separate spare parts list for standard parts "DD+DIS011.93M".
* = Spare part standardly is not kept in stock, extended delivery time is possible.
99 = Spare Part is not shown in Exploded Views.
ERASURE UNIT
01*
02
(GS401, GS402)
10
D542
12 03
D2514
04
D706
D540
D6426
D2514
D2861
11 (GS401, GS402)
ONLY FOR DEVICES
WHICH ARE LOCATED
HIGHER THAN 2000M
ABOVE SEE LEVEL
(SEE SB DD+DIS276.07E)
08
(MG402)
08
(MG401)
D6426
D2514
09
05
07 D542
(GS406)
07 06
(GS404)
5148_chap05_23.CDR
12 CM+9042421640 MAGNETVENTIL
SOLENOID VALVE
ELECTROVANNE
= Assembly
D = Standard part. For order please refer to the separate spare parts list for standard parts "DD+DIS011.93M".
* = Spare part standardly is not kept in stock, extended delivery time is possible.
99 = Spare Part is not shown in Exploded Views.
01
02
10
(GS412) 03
D372
D2514
D579
12
(GS414 / 04 (M401)
GS416)
D372
D2514
D579
10
(GS418)
11 06
09
07
05
08
5148_chap05_24.CDR
= Assembly
D = Standard part. For order please refer to the separate spare parts list for standard parts "DD+DIS011.93M".
* = Spare part standardly is not kept in stock, extended delivery time is possible.
99 = Spare Part is not shown in Exploded Views.
03 (GS410)
D372 03
02 D2514
D579
01 (GS408)
04
06
07
09
08
07
05 12 06
11
(M400)
D372
D2514
D579
05
10
10
5148_chap05_25.CDR
= Assembly
D = Standard part. For order please refer to the separate spare parts list for standard parts "DD+DIS011.93M".
* = Spare part standardly is not kept in stock, extended delivery time is possible.
99 = Spare Part is not shown in Exploded Views.
01*
02
D6424
D2514
D2861
03
(GS422)
(GS420) 05
07
D706
D2514
04
D542
D6426 (GS422)
06 (GS420)
ONLY FOR DEVICES WHICH ARE LOCATED
HIGHER THAN 2000M ABOVE SEE LEVEL
(SEE SB DD+DIS276.07E) 5148_chap05_26.CDR
= Assembly
D = Standard part. For order please refer to the separate spare parts list for standard parts "DD+DIS011.93M".
* = Spare part standardly is not kept in stock, extended delivery time is possible.
99 = Spare Part is not shown in Exploded Views.
05
10
D372
D2514
D579
11
04 (GS428)
(GS430) 07
(GS432) (M403)
D372
D2514
D579
11
(GS434)
12 02
09
03
08
06
01
5148_chap05_27.CDR
= Assembly
D = Standard part. For order please refer to the separate spare parts list for standard parts "DD+DIS011.93M".
* = Spare part standardly is not kept in stock, extended delivery time is possible.
99 = Spare Part is not shown in Exploded Views.
05 (GS424)
D372 05
D2514
D579
10 (GS426)
04
09
02
06
01
07
06
08 03 02
11
(M402)
D372
D2514
D579
08
12
12
5148_chap05_28.CDR
= Assembly
D = Standard part. For order please refer to the separate spare parts list for standard parts "DD+DIS011.93M".
* = Spare part standardly is not kept in stock, extended delivery time is possible.
99 = Spare Part is not shown in Exploded Views.
05
02
(GS610)
04
01
(GS620)
03
(GS622)
CR85-X
Type 5148/100
CR75.0
Type 5146/105 (as of SN 6000) 5148_chap05_29.cdr
= Assembly
D = Standard part. For order please refer to the separate spare parts list for standard parts "DD+DIS011.93M".
* = Spare part standardly is not kept in stock, extended delivery time is possible.
99 = Spare Part is not shown in Exploded Views.
05
02
(GS622)
(GS620)
03
04
(GS610)
01
CR75.0
Type 5146/101 (up to SN < 6000) 5148_chap05_30.CDR
1 CM+9047913370 LEITERPLATTENHALTER
PCB HOLDER
CARTE FIXATION
2 CM+9514622504 PMT-MODUL - (GS622) - (TYPE 5146/101 ONLY)
PMT MODULE - (GS622) - (TYPE 5146/101 ONLY)
MODULE PMT - (GS622) - (TYPE 5146/101 ONLY)
3 CM+9514827003 OPTIKBAUGRUPPE - (GS620) CR75.0
OPTIC MODULE - (GS620)
ENSEMBLE OPTIQUE - (GS620)
Type 5146/101 (up to SN <6000)
4 CM+9949981300 GS IOB-SIN-5STEP - (GS610)
5 FOLD MOTOR CONTROLLER PCB - (GS610)
CI IOB-SIN-5STEP - (GS610)
5 CM+9514621070 VERKLEIDUNG
PLASTIC COVER
PLASTIC COVER
= Assembly
D = Standard part. For order please refer to the separate spare parts list for standard parts "DD+DIS011.93M".
* = Spare part standardly is not kept in stock, extended delivery time is possible.
99 = Spare Part is not shown in Exploded Views.
03
06
02
05
D3252 D2763
01
04 D2513
D2851 D2763
01
D369
D3252
CR85-X
04 D2851
Type 5148/100 D2851
CR75.0 D3252
Type 5146/105 (as of SN 6000) 04 5148_chap05_31.CDR
= Assembly
D = Standard part. For order please refer to the separate spare parts list for standard parts "DD+DIS011.93M".
* = Spare part standardly is not kept in stock, extended delivery time is possible.
99 = Spare Part is not shown in Exploded Views.
01
03
02
D3252 D2763
05
04 D2513
D2851 D2763
05
D369
D3252
04 D2851
D2851
D3252
04
CR75.0
Type 5146/101 (up to SN < 6000)
5148_chap05_32.CDR
= Assembly
D = Standard part. For order please refer to the separate spare parts list for standard parts "DD+DIS011.93M".
* = Spare part standardly is not kept in stock, extended delivery time is possible.
99 = Spare Part is not shown in Exploded Views.
D5681
03 D2778
04
02
D444
D2765 01
(M702)
06
(GS612)
05
CR85-X (M612)
Type 5148/100
CR75.0
Type 5146/105 (as of SN 6000)
= Assembly
D = Standard part. For order please refer to the separate spare parts list for standard parts "DD+DIS011.93M".
* = Spare part standardly is not kept in stock, extended delivery time is possible.
99 = Spare Part is not shown in Exploded Views.
01
04
D5681
06 D2778
(GS612)
09
02
D444
D2765
05 03
(M612) (M702)
08
D2443
01
D3008
07
D579
D2514
CR75.0 D542
Type 5146/101 (up to SN < 6000)
= Assembly
D = Standard part. For order please refer to the separate spare parts list for standard parts "DD+DIS011.93M".
* = Spare part standardly is not kept in stock, extended delivery time is possible.
99 = Spare Part is not shown in Exploded Views.
02
05
02
04
(GS616)
06
01
03
(M622)
5148_chap05_35.CDR
= Assembly
D = Standard part. For order please refer to the separate spare parts list for standard parts "DD+DIS011.93M".
* = Spare part standardly is not kept in stock, extended delivery time is possible.
99 = Spare Part is not shown in Exploded Views.
04
(M621)
05
03
(GS618)
06 09
06
07
(GS626)
(GS628)
D6522
D2514
D2853
02
03 01
(GS614)
08
5148_chap05_36.CDR
= Assembly
D = Standard part. For order please refer to the separate spare parts list for standard parts "DD+DIS011.93M".
* = Spare part standardly is not kept in stock, extended delivery time is possible.
99 = Spare Part is not shown in Exploded Views.
06
06
D2873
D2514
D579 05
(GS202)
04
02
(GS206)
D706
D4338
07
01
(M212)
03
5148_chap05_37.CDR
ROTATION DRIVE
= Assembly
D = Standard part. For order please refer to the separate spare parts list for standard parts "DD+DIS011.93M".
* = Spare part standardly is not kept in stock, extended delivery time is possible.
99 = Spare Part is not shown in Exploded Views.
ROTATION DRIVE
01 02
04
03
06
07 CM+9514565502 (1x)
05
CM+9514571252 (1x)
5148_chap05_38_M.cdr
= Assembly
D = Standard part. For order please refer to the separate spare parts list for standard parts "DD+DIS011.93M".
* = Spare part standardly is not kept in stock, extended delivery time is possible.
99 = Spare Part is not shown in Exploded Views.
01*
02
03*
12 04*
11 10
13
15 05
14
16
07
08
09
10 09
06 08
07
5148_chap05_50.CDR
= Assembly
D = Standard part. For order please refer to the separate spare parts list for standard parts "DD+DIS011.93M".
* = Spare part standardly is not kept in stock, extended delivery time is possible.
99 = Spare Part is not shown in Exploded Views.
02*
03* 01*
04
05*
14
06*
13 12
15
17 07
16
18
09
10
11
12 11
08 10
09
5146_chap05_54.CDR
= Assembly
D = Standard part. For order please refer to the separate spare parts list for standard parts "DD+DIS011.93M".
* = Spare part standardly is not kept in stock, extended delivery time is possible.
99 = Spare Part is not shown in Exploded Views.
02*
03* 01*
04
05*
14
06*
13 12
15
17 07
16
18
09
10
11
12 11
08 10
09
5146_chap05_54.CDR
= Assembly
D = Standard part. For order please refer to the separate spare parts list for standard parts "DD+DIS011.93M".
* = Spare part standardly is not kept in stock, extended delivery time is possible.
99 = Spare Part is not shown in Exploded Views.
01*
02
03*
12 04*
11 10
13
15 05
14
16
07
08
09
10 09
06 08
07
5148_chap05_50.CDR
= Assembly
D = Standard part. For order please refer to the separate spare parts list for standard parts "DD+DIS011.93M".
* = Spare part standardly is not kept in stock, extended delivery time is possible.
99 = Spare Part is not shown in Exploded Views.
01*
02
03*
12 04*
11 10
13
15 05
14
16
07
08
09
10 09
06 08
07
5148_chap05_50.CDR
= Assembly
D = Standard part. For order please refer to the separate spare parts list for standard parts "DD+DIS011.93M".
* = Spare part standardly is not kept in stock, extended delivery time is possible.
99 = Spare Part is not shown in Exploded Views.
01
02*
11 03*
10 09
12
14 04
13
15
06
07
08
09 08
05 07
06
5148_chap05_53.CDR
= Assembly
D = Standard part. For order please refer to the separate spare parts list for standard parts "DD+DIS011.93M".
* = Spare part standardly is not kept in stock, extended delivery time is possible.
99 = Spare Part is not shown in Exploded Views.
02*
03* 01*
04
05*
14
06*
13 12
15
17 07
16
18
09
10
11
12 11
08 10
09
5146_chap05_54.CDR
= Assembly
D = Standard part. For order please refer to the separate spare parts list for standard parts "DD+DIS011.93M".
* = Spare part standardly is not kept in stock, extended delivery time is possible.
99 = Spare Part is not shown in Exploded Views.
01*
02
03*
12 04*
11 10
13
15 05
14
16
07
08
09
10 09
06 08
07
5148_chap05_50.CDR
= Assembly
D = Standard part. For order please refer to the separate spare parts list for standard parts "DD+DIS011.93M".
* = Spare part standardly is not kept in stock, extended delivery time is possible.
99 = Spare Part is not shown in Exploded Views.
01*
02
03*
12 04*
11 10
13
15 05
14
16
07
08
09
10 09
06 08
07
5148_chap05_50.CDR
= Assembly
D = Standard part. For order please refer to the separate spare parts list for standard parts "DD+DIS011.93M".
* = Spare part standardly is not kept in stock, extended delivery time is possible.
99 = Spare Part is not shown in Exploded Views.
03*
02* 01* 04
05*
12
06
13 07
14
09
10
11
06
11
08 10
09
5148_chap05_042.CDR
= Assembly
D = Standard part. For order please refer to the separate spare parts list for standard parts "DD+DIS011.93M".
* = Spare part standardly is not kept in stock, extended delivery time is possible.
99 = Spare Part is not shown in Exploded Views.
03*
02* 01* 04
05*
12
06
13 07
14
09
10
11
06
11
08 10
09
5148_chap05_042.CDR
= Assembly
D = Standard part. For order please refer to the separate spare parts list for standard parts "DD+DIS011.93M".
* = Spare part standardly is not kept in stock, extended delivery time is possible.
99 = Spare Part is not shown in Exploded Views.
• Number of machines
• Extension of the service area
• Local service structure (centralized / decentralized)
• Stockpilling
Type Overview
This spare parts list is valid for the following machine type(s):
Accessory Overview
Following accesssories are separately available:
► Document History
► Referenced Documents
Document Title
Not applicable
Edition 4, Revision 1
12-2010 printed in Germany Document Node ID: 32047827
so_standard_e_template_v07
Agfa Company Confidential Copyright © 2010 Agfa HealthCare N.V.
DD+DIS025.05E CR User Station
► Manufacturer
Agfa HealthCare N.V.
Publisher
Agfa-Gevaert HealthCare GmbH
Tegernseer Landstraße 161
D – 81539 München
Germany
Copyright © 2010 Agfa HealthCare N.V.
All rights reserved.
Technical modifications reserved.
Agfa and the Agfa rhombus are trademarks of Agfa-Gevaert N.V., Belgium,
or its affiliates.
WARNING:
Improper operation or service activities may cause damage or injuries.
INSTRUCTION:
(1) Read the “Generic Safety Directions” document
(see MEDNET GSO => General Info => Agfa HealthCare => Publications =>
Service Manual) prior to attempting any operation, repair or maintenance task on
the equipment.
(2) Strictly observe all safety directions within the “Generic Safety Directions” and on
the product.
LIST OF CONTENTS
1 INTRODUCTION ......................................................................................................................5
2 SAFETY INSTRUCTIONS........................................................................................................7
2.1 General Safety Instructions.......................................................................................................7
5 TROUBLESHOOTING............................................................................................................26
5.1 Isolating Transformer ..............................................................................................................26
5.2 ID Tablet..................................................................................................................................27
8 INSTALLATION PLANNING...................................................................................................28
8.1 Dimensions .............................................................................................................................28
8.4.2 ID Tablet..................................................................................................................................31
1 Introduction
The CR User Station (CRUS) is an operating console that facilitates the work of the
radiographer. It is suitable for all CR environments: centralized, decentralized or
personal CR solutions.
It has been designed in line with the look and feel of the range of the new generation
digitizers, combining with the digitizers to form a “whole”. It provides a professional way
of delivering an integrated solution for computer equipment in Agfa’s CR offering.
Figure 1
NOTE:
• For all required information concerning digitizers, where the CRUS is attached
to, please refer to the relevant Service Documentation:
• CR 25.0: DD+DIS071.04E
• CR 35-X: DD+DIS219.06E
• CR 75.0: DD+DIS002.04E
• CR 85-X DD+DIS378.05E
2 Safety Instructions
Figure 2
• Make sure that the two safety caps are mounted on top of the right side rollers in
case of stand-alone CRUS (this means that the CRUS is not connected to a
CR 25.0/CR 35-X Digitizer), as shown below.
Figure 3
Pay attention to the warning label at the rear side of the CRUS:
Check power selection before connecting to mains
Figure 4
In combination with a digitizer the isolating transformer assures that the digitizer still
complies with EN 60601-1-1 medical, electrical systems.
The CR User Station bears the CE mark and fully complies with the
CE Directive 89/336/EEC and with the federal code of the United States, bearing on:
• Emission and immunity according to EN 60601-1-2, for emissions the
equipment complies with EN 55011 class A (CISPR 11).
This is a Class A product. In a domestic environment this product may
cause radio interference in which case the user may be required to take
adequate measures.
• Emissions according to 47 CFR part 15 subpart B, Class A.
This equipment has been tested and found to comply with the limits for a
Class A digital device, pursuant to part 15 of the FCC Rules. These limits
are designed to provide reasonable protection against harmful
interference when the equipment is operated in a commercial environment. This
equipment generates, uses, and can radiate radio frequency energy and, if not
installed and used in accordance with the instruction manual, may cause harmful
interference to radio communications. Operation of this equipment in a residential
area is likely to cause harmful interference in which case the user will be required to
correct the interference at his own expense.
• Radio-parameters of the ID Tablet according to ETS 300330.
3 Installation Instructions
NOTE:
The actions described in this section apply only, if the CRUS is installed as stand-
alone device.
Skip this section if you want to dock it to a CR 25.0 / CR 35-X digitizer.
Figure 5
Figure 6
Figure 7
Figure 8
Figure 11
Figure 12
Figure 13
NOTE:
By default, a rack for the Barco monitor is installed.
If you have a Barco monitor, go directly to step (2).
If you have another monitor, exchange the rack before proceeding
with step (2). The exchange of the rack for the NEC monitor is
described in section 3.9.
Figure 14
(3) Adjust the monitor according to
section 3.10.
Figure 15
(1) Put the ID-Tablet back to its place and fix the Phillips screw
(see Figure 8, page 11).
(2) Connect the USB cable coming from the ID-Tablet to a free USB port of the PC.
Figure 17
100 V: 6.3 AT
120 V 5.0 AT
230 V 2.5 AT
240 V 2.5 AT
Figure 18
Figure 19
Figure 20
20
100
2 01
240
Figure 21
Figure 22
(1) Connect the network cable of the digitizer to the hub of the CRUS.
NOTE:
If you want to dock the CRUS to a CR 25.0 digitizer, continue with section 3.8.
Result The installation of the CRUS as a stand-alone device is finished. Connect it to the
mains.
Figure 23
Figure 24
Figure 25
Figure 26
Figure 27
Figure 29
Figure 30
Figure 31
NOTE:
The monitor rack, the bearing and the
washers can easily fall down!
Be careful when removing the axis.
Figure 32
Figure 33
NOTE:
If the monitor rack is not fixed properly, it
may tilt unwanted.
Check that the hole of the retainer (see
circle) and the form of the axis fit smoothly
together.
Figure 35
Figure 34
Figure 36
Figure 37
Figure 38
Figure 39
Adjust the inclination of the monitor by using the lever at the rear side:
NOTE:
The lever is a quick-fix system: Its fixed or loose position can be changed when it is
pulled (lever free, see Figure 40).
There is no need to open the Allen screw at the axis for the adjustment of the monitor.
lever lever
(2) Loosen lever by turning counter-
free locked
clockwise.
Figure 40
Result The individual components of the CR User Station are ready for use now.
4 Functional Description
Monitor (*)
Keyboard (*)
ID Tablet
Cassette storage
Inside:
- Isolating transformer
- PC (*)
- UPS (*)
- Network Switch (*)
Figure 41
3
store 2
exposed
cassette
store
scanned
and erased
cassette
4
Figure 42
• Up to five exposed cassettes are stored in the upper side storage racks (1).
• The exposed and identified cassettes are inserted directly into the digitizer or, if the
digitizer is busy, stored in the top storage rack (3).
• The scanned and erased cassettes waiting to be used again are stored in the lower
side storage rack (4).
In order to save time it is also possible to use the CRUS for Direct ID.
• For this an exposed but unidentified cassette is inserted into the digitizer.
• By entering the cassette, the digitizer reads the cassette information of the RF Tag
and a communication to the CRUS is opened.
• The digitizer starts scanning the IP and sends an ID Data Request Message to the
CRUS. On the monitor of the CRUS the ID window pops up and the ID data have to
be entered.
• After the user completed the information at the CRUS, it is sent back to the digitizer.
The communication channel is closed afterwards.
5 Troubleshooting
WARNING:
Fire hazard:
Only use fuses of the exact value and characteristics stated above.
5.2 ID Tablet
Figure 44
For the current Spare Parts List refer to MedNet GSO Library.
7 Maintenance
8 Installation Planning
8.1 Dimensions
193 (75”)
5 (2”)
43 (17”) 45 (18”) 100 (40”)
5 (2”)
Digitizer
64 (25”)
75 (30”)
200 (77”)
120 (47”)
Figure 45
64 (25”)
Digitizer
200 (77”)
127 (48”)
Figure 46
Weight • 53 kg (empty)
• Approx. 100 kg (fully equipped)
Operating Conditions
Temperature 10°C – 30°C
Change of temperature 0.5°C / min.
Humidity 15% RH to 80% RH
Type Label
The type label is at the rear side of the CRUS.
Figure 48
Figure 47
8.4 Connections
CRUS
Primary:
100 V 6,3 A, slow blow
120 V 5,0 A, slow blow
230 – 240 V 2,5 A, slow blow
CR 25.0
Operating voltage:
230 V / 240 V + 10%
120 V + 10%
100 V + 10%
Mains Fuse Protection
Europe 16 A, slow blow
USA & Japan 15 A, slow blow
8.4.2 ID Tablet
Local and International wiring regulations must be observed. Check all supplies and
voltages, currents, trips and fuses with the hospital facilities department or their
engineers.
CR User Station
Type 5157/100
Internal update: 4
CONFIDENTIALITY NOTE:
Use, dissemination, distribution or reproduction of this document by unauthorized personnel is not permitted and may be unlawful.
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Manufacturer
Agfa HealthCare N.V
Published by
Agfa-Gevaert HealthCare GmbH
Tegernseer Landstraße 161
D - 81539 München
Germany
Copyright © 2011 Agfa HealthCare N.V.
All rights reserved.
Technical modifications reserved.
AGFA and the Agfa-Rhombus are trademarks of Agfa HealthCare N.V., Belgium,
or its affiliates.
WARNING:
INSTRUCTION:
(1) Read the "Generic Safety Directions" document
(see MEDNET GSO => General Info => Agfa HealthCare => Publications =>
Service Manual) prior to attempting any operation, repair or maintenance task on
the equipment.
(2) Strictly observe all safety directions within the "Generic Safety Directions" and on
the product.
WARNING:
INSTRUCTION:
• Replace defective parts with Agfa HealthCare original spare parts.
• Use only tools and measuring instruments which are suitable for the procedure.
• Only approved Agfa HealthCare accessories must be used. For a list of
compatible accessories contact your local Agfa HealthCare organization or
www.agfa.com.
NOTE:
To verify the latest version of single documents and of Service Manuals refer to the
Document Type ‘Order List’ in the GSO Library.
Document History
NOTE:
Agfa reserves the right to deliver compatible parts or alternative parts.
Prices to the original order may differ.
Contact
Spare Parts ordering
Spare Parts ordering
Europe
Europe Orderprocessing-europe@matrium.net
orderprocessing-europe.matrium@eads.com
Overseas
Overseas Orderprocessing-overseas@matrium.net
orderprocessing-overseas.matrium@eads.com
Worldwide returns.matrium@eads.com
Worldwide Returns@matrium.net
NOTE:
For Recycling
Recycling of theInformation
electronicplease refer to:waste equipment will ensure safety
and electrical
of human health and the environment.
For information about electronic and electrical waste equipment disposal,
http://intra.agfanet/cd/ep/ehs.nsf
recovery and collection points, please contact your local waste disposal service
or producer / distributor of this equipment.
If your equipment contains removable batteries or accumulators please dispose
of these separately according to local regulations.
Contents
LEISTUNGSEINHEIT
POWER UNIT
MODULE DE PUISSANCE
PAGE 12 - 13
GESTELL
FRAME
CHÂSSIS
PAGE 6 - 7
ID STATION - AB FN 10.000
ID STATION - AS OF SN 10.000
ID STATION - A PARTIR DE NS 10.000
PAGE 10 - 11
08
06
05
09
07
04
03*
02
01
5157_chap05_01.CDR
FRAME
= Assembly
D = Standard part. For order please refer to the separate spare parts list for standard parts "DD+DIS011.93M".
* = Spare part standardly is not kept in stock, extended delivery time is possible.
99 = Spare Part is not shown in Exploded Views.
FRAME
01*
07
08
04 06
02
05 05
03
5157_chap05_02.CDR
ID STATION - UP TO SN 10.000
= Assembly
D = Standard part. For order please refer to the separate spare parts list for standard parts "DD+DIS011.93M".
* = Spare part standardly is not kept in stock, extended delivery time is possible.
99 = Spare Part is not shown in Exploded Views.
ID STATION - UP TO SN 10.000
01*
05
04
03
02
5157_chap05_03.CDR
ID STATION - AS OF SN 10.000
= Assembly
D = Standard part. For order please refer to the separate spare parts list for standard parts "DD+DIS011.93M".
* = Spare part standardly is not kept in stock, extended delivery time is possible.
99 = Spare Part is not shown in Exploded Views.
ID STATION - AS OF SN 10.000
02
01
03, 04, 05
5157_chap05_04.CDR
POWER UNIT
= Assembly
D = Standard part. For order please refer to the separate spare parts list for standard parts "DD+DIS011.93M".
* = Spare part standardly is not kept in stock, extended delivery time is possible.
99 = Spare Part is not shown in Exploded Views.
POWER UNIT
• Number of machines
• Extension of the service area
• Local service structure (centralized / decentralized)
• Stockpilling
• Not available
Type Overview
This spare parts list is valid for the following machine type(s):
Accessory Overview
Following accesssories are separately available:
• Not available
CR 75.0
Type 5146/100, 101
Order-No.: DD+DIS075.04E
CR 75.0
*1WRN1B1*
1 Piece WRN1B MA1
Type 5146 /100 and 101
Centricity CR MP3510
Type 5146 / 200 and 201
List of contents
1 General ....................................................................... 3
1.1 Main changes in ACP_1203 .................................................... 3
1.2 Hardware Requirements ......................................................... 3
1.3 Software Requirements .......................................................... 3
1.4 Order number .......................................................................... 4
1.5 Scope of Delivery .................................................................... 4
2 Installation ................................................................. 5
2.1 Upgrade to ACP_1203 ............................................................. 5
2.2 Exchange of the Hard Disk / Procedure to install
ACP_1203 on a virgin Hard Disk ............................................ 5
1 General
cPCI-Rack:
• 200 MHz Ariel Board
• EPROM Mon 4.26
• MSB, Master-Scan-Board, Revision: 0
with Scan IO-Board Boot EPLD, Version: 0
and Scan IO-Board serial PROM, Version: 0
• Floppy (no SCSI)
• SCSI Hard disk
CM+9514610505
2 Installation
2.1 Upgrade to ACP_1203
• Take the Hard disk formatter floppy of ACP_1203 and insert it into the
floppy disk drive of ADC Compact Plus.
• Disconnect ADC Compact Plus from the network (since a default IP-
Address & hostname is configured)
• Connect a PC with terminal program or a Terminal to service RS232 port
• Press the Reset button on Ariel-CPU or turn power off/on of ADC
Compact Plus
• Confirm partitioning/formatting of the hard disk with ‘y’.
• Wait until you get the final message on a terminal connected to the
service RS232 port or wait until control LED of floppy disk drive is off.
• Insert the floppy ACP_1203 Portex Installer 2/2 into the disk drive of
ADC Compact Plus
• Press the Reset on Ariel-CPU or turn Power Off/On of ADC Compact
Plus
• Wait until you get the final message on a terminal connected to the
service RS232 port or wait until control LED of floppy disk drive is off.
• Insert the floppy ACP_1203 Release Disk 1/3 into the disk drive of ADC
Compact Plus
• Press the Reset button on Ariel-CPU or turn power off/on of ADC
Compact Plus
• Wait until you get the final message on a terminal connected to the
service RS232 port, or wait until control LED of floppy disk drive is off.
• Proceed the same way with the ACP_1203 Release Disk 2/3 and 3/3
• If the previous software was < ACP_1104, then close the machine doors
and scan or erase one cassette. This cassette does not need to be
exposed. Then make a new backup.
For explanation see "What's new in ACP_1203 compared with
ACP_1104"
Possible problems:
(1) Backup immediately after software installation. In this case the
default 0 of ht_pm_test is backed-up and after a restore, the
calibration is performed again. ÆNo documentation of
Photomultiplier ageing.
(2) Scan or erasure of first cassette with open machine doors and bright
daylight. In this case the wrong first calibration is backed-up. This
might lead to info counter entries, when this back-up is used later.
Remedy: Make an additional back-up at the next visit.
• Limit for polygon jitter entries to info counters was lowered from 0.3 to
0.25 [pixels]. This influences the generation of the jitter warning code 20C89
within the info counters.
• Now avoiding warning 20C46. A change in the software now ensures the
required settling time for the high tension.
• Service function ‚ Check Laser‘ now works also at polygon standby.
• SAL display now consistent 12 Bit. The functions “shading calibration”
and “SAL inspection” returned 14 bit SAL values, which was confusing.
Now all screen outputs are consistent 12 bit.
• New service feature for ‘blue LED‘ image generation. A new menu entry
has been made: /Service/Checks/Send Flatfield/Blue LED, which
creates a 18CMx24CM test image using the blue LED that is built into
the PMT.
The Digitizer is beeping (and showing the code 20500) while performing
the scan. An error during this procedure requires that the digitizer is
restarted.
• Under very special circumstances the info counter contents can be lost
or might be scrambled in case of switching off the device during
operation
• Jitter measurement (EVA) delivers not yet the correct results
• No indication of folders in status 'error'
• If 10 folders are in the status ‚error‘. These files have to be manually
erased via Key-Operator mode before the system accepts new
cassettes
• Photomultiplier test may fail (20E0D) due to large tolerances
• The serial numbers of parts are stored in configuration files on hard disk.
These files are deleted with prepship.bat. In releases > ACP_1104 they
will be stored in adjustment files
• Invalid cpf-files are not always rejected after configuration; there is no
hint that a reset is necessary to apply the changes
So, after a cpf-file upload, you have to do a reset.
Order-No.: DD+DIS069.04E
CR 75.0
*1WRGYG1*
1 Piece WRGYG MA1
Type 5146 /100 and 101
Centricity CR MP3510
Type 5146 / 200 and 201
List of contents
1 General ....................................................................... 3
1.1 Hardware Requirements ......................................................... 3
1.2 Software Requirements .......................................................... 3
1.3 Order number .......................................................................... 3
1.4 Scope of Delivery .................................................................... 3
2 Installation ................................................................. 4
2.1 Upgrade to ACP_1204 ............................................................. 4
2.2 Exchange of the Hard Disk / Procedure to
install ACP_1204 on a virgin Hard Disk................................. 4
1 General
The software package ACP_1204 also contains ten pieces of "Philips
CAPSULEline Pro" erasure lamps which show an improved behavior
regarding the problem of gray bulbs. These lamps must be installed in the
erasure unit prior to the Sw - installation. The respective sticker on the erasure
unit has to be changed, too.
cPCI-Rack:
• 200 MHz Ariel Board
• EPROM Mon 4.26
• MSB, Master-Scan-Board, Revision: 0
with Scan IO-Board Boot EPLD, Version: 0
and Scan IO-Board serial PROM, Version: 0
• Floppy (no SCSI)
• SCSI Hard disk
2 Installation
2.1 Upgrade to ACP_1204
• Take the Hard disk formatter floppy of ACP_1204 and insert it into the
floppy disk drive of ADC Compact Plus.
• Disconnect ADC Compact Plus from the network (since a default IP-
Address & hostname is configured)
• Connect a PC with terminal program or a Terminal to service RS232 port
• Press the Reset button on Ariel-CPU or turn power off/on of ADC
Compact Plus
• Confirm partitioning/formatting of the hard disk with ‘y’.
• Wait until you get the final message on a terminal connected to the
service RS232 port or wait until control LED of floppy disk drive is off.
• Insert the floppy ACP_1204 Portex Installer 2/2 into the disk drive of
ADC Compact Plus
• Press the Reset on Ariel-CPU or turn Power Off/On of ADC Compact
Plus
• Wait until you get the final message on a terminal connected to the
service RS232 port or wait until control LED of floppy disk drive is off.
• Insert the floppy ACP_1204 Release Disk 1/3 into the disk drive of ADC
Compact Plus
• Press the Reset button on Ariel-CPU or turn power off/on of ADC
Compact Plus
• Wait until you get the final message on a terminal connected to the
service RS232 port, or wait until control LED of floppy disk drive is off.
• Proceed the same way with the ACP_1204 Release Disk 2/3 and 3/3
Then parse the CPF-File using the guidelines of the Service Menu Topic
Install from floppy, CPF-File, displayed on the user interface.
• If the previous software was < ACP_1104, then close the machine doors
and scan or erase one cassette. This cassette does not need to be
exposed. Then make a new backup.
• Under very special circumstances the info counter contents can be lost
or might be scrambled in case of switching off the device during
operation.
• No indication of folders in status 'error'.
• If 10 folders are in the status ‚error‘. These files have to be manually
erased via Key-Operator mode before the system accepts new
cassettes.
• Photomultiplier test may fail (20E0D) due to large tolerances.
• The serial numbers of parts are stored in configuration files on hard disk.
These files are deleted with prepship.bat. In releases > ACP_1104 they
will be stored in adjustment files.
• Invalid cpf-files are not always rejected after configuration; there is no
hint that a reset is necessary to apply the changes
So, after a cpf-file upload, you have to do a reset.
• Stepper motor board communication error with belt node input buffer.
Occurred sporadically only at one digitizer, no problem at all others.
Cause is unknown.
• Erasure unit fan speed too low error message. Occurred only once and
is not reproducible.
• SW crash to monitor with cpf file configuration.
• Install of SW by upload procedure from Internet Explorer fails.
Order-No.: DD+DIS070.04E
CR 75.0
*1WRPKL1*
1 Piece WRPKL MA1
Type 5146 / 100 and 101
Centricity CR MP3510
Type 5146 / 200 and 201
List of contents
1 General ....................................................................... 3
1.1 Hardware Requirements ......................................................... 3
1.2 Software Requirements .......................................................... 3
1.3 Order number .......................................................................... 3
1.4 Scope of Delivery .................................................................... 3
2 Installation ................................................................. 4
2.1 Upgrade of versions < ACP_1204 with ARIEL-CPU to
ACP_3101................................................................................. 4
2.2 Exchange of the Hard Disk / Procedure to install
ACP_3101 on a virgin Hard Disk ............................................ 6
5 Important Remarks.................................................... 9
1 General
1.1 Hardware Requirements
cPCI-Rack:
• 200 MHz Ariel Board or Oberon Board
• EEPROM Mon 4.26
• MSB, Master-Scan-Board, Revision: 0
with Scan IO-Board Boot EPLD, Version: 0
and Scan IO-Board serial PROM, Version: 0
• Floppy (no SCSI)
• SCSI Hard disk
Optic Module
• For usage as a mammography system, an optic module (identification
number: F8.5146.2700.0) with switchable second laser power of 30 mW
and laser control version code 1 is required (visible in info counters: “
Optics module version ...”).
Currently this mammography function is disabled, as it is not released up
to now.
2 Installation
2.1 Upgrade of versions < ACP_1204 with ARIEL-CPU to ACP_3101
Portex Installation
• Insert the floppy ACP_3101 Portex Installer 1/2 into disk drive of the
digitizer
• Press the Reset on CPU-Board or turn Power Off/On of the digitizer
• Wait until you get the final message on a Terminal connected to the
service RS232 port, or wait until control LED of floppy disk drive is off.
• Insert the floppy ACP_3101 Portex Installer 2/2 into disk drive of the
digitizer
• Press the Reset on CPU-Board or turn Power Off/On of the digitizer
• Wait until you get the final message on a Terminal connected to the
service RS232 port, or wait until control LED of floppy disk drive is off.
• Remove floppy (do not insert any other floppy)
• Press the Reset on CPU-Board or turn Power Off/On of the digitizer
• Wait until you get the final message on a Terminal connected to the
service RS232 port, or wait approximately 2 minutes.
Installation of Application
• Insert the floppy ACP_3101 Release Disk 1/3 into disk drive of the
digitizer
• Press the Reset on CPU-Board or turn Power Off/On of the digitizer
• Wait until you get the final message on a Terminal connected to the
service RS232 port, or wait until control LED of floppy disk drive is off.
• Insert the floppy ACP_3101 Release Disk 2/3 into disk drive of the
digitizer
• Press the Reset on CPU-Board or turn Power Off/On of the digitizer
• Wait until you get the final message on a Terminal connected to the
service RS232 port, or wait until control LED of floppy disk drive is off.
• Insert the floppy ACP_3101 Release Disk 3/3 into disk drive of the
digitizer
• Press the Reset on CPU-Board or turn Power Off/On of the digitizer
• Wait until you get the final message on a Terminal connected to the
service RS232 port, or wait until control LED of floppy disk drive is off,
plus some additional time.
• Remove the floppy, press the Reset on CPU-Board or turn Power Off/On
of the digitizer and wait approx. 2 minutes (system extracts SW-release)
• Press the Reset on CPU-Board or turn Power Off/On of CR 75.0 and
wait until system is up
Finishing Installation
• If the previous software had a lower version than ACP_1201, then close
the machine doors and scan or erase one cassette.
This cassette needs not be exposed.
For an upgrade of a lower version than ACP_1201 to ACP_3101 a new IP
center calibration is necessary, because of a changed IP stop position before
start of scan. (Calibration procedure, see chapter 3.6 of your Technical
Documentation)
• Make a new backup (therefore you find an empty floppy within the set of
delivered floppies)
• Take the Hard disk formatter floppy of ACP_3101 and insert it into the
floppy disk drive of the digitizer.
• Disconnect the digitizer from the network (since a default IP-Address &
hostname is configured)
• Connect a PC with terminal program or a Terminal to service RS232 port
• Press the Reset button on Ariel-CPU or turn power off/on of the digitizer
• Confirm partitioning/formatting of the hard disk with ‘y’.
• Wait until you get the final message on a terminal connected to the
service RS232 port or wait until control LED of floppy disk drive is off.
• Insert the floppy ACP_3101 Portex Installer 2/2 into the disk drive of
the digitizer
• Press the Reset on Ariel-CPU or turn Power Off/On of the digitizer
• Wait until you get the final message on a terminal connected to the
service RS232 port or wait until control LED of floppy disk drive is off.
• Remove floppy (don't insert any other floppy)
• Press the Reset on CPU-Board or turn Power Off/On the digitizer
• Wait until you get the final message on a Terminal connected to the
service RS232 port, or wait approximately 2 minutes.
• Insert the floppy ACP_3101 Release Disk 1/3 into the disk drive of the
digitizer
• Press the Reset button on Ariel-CPU or turn power off/on of the digitizer
• Wait until you get the final message on a terminal connected to the
service RS232 port, or wait until control LED of floppy disk drive is off.
• Proceed the same way with the ACP_3101 Release Disk 2/3 and 3/3
• Press the Reset button on Ariel-CPU or turn power off/on of the digitizer
and wait until system is up
Then parse the CPF-File using the guidelines of the Service Menu Topic
Install from floppy, CPF-File, displayed on the user interface.
• If the previous software had a lower version than ACP_1201, then close
the machine doors and scan or erase one cassette.
This cassette needs not be exposed.
For an upgrade of a lower version than ACP_1201 to ACP_3101 a new IP
center calibration is necessary, because of a changed IP stop position before
start of scan. (Calibration procedure, see chapter 3.6 of your Technical
Documentation)
• Make a new backup (therefore you find an empty floppy within the set of
delivered floppies)
This version:
• A small white border may be visible at mammo images, if they contain
areas with very low signal level.
• CR 75.0 crashes after display of several calibration curves
5 Important Remarks
When upgrading from a version < ACP_1007 to ACP_3101, a new BOL/BOS
Calibration is necessary because of a changed IP stop position before start of
scan.
CR 75.0
*1YEB111*
1 Piece YEB11 MA1
Type 5146 / 100 and 101
Centricity CR MP3510
Type 5146 / 200 and 201
List of Contents
1 General ....................................................................... 3
1.1 Hardware Requirements ......................................................... 3
1.2 Software Requirements .......................................................... 3
1.3 Order number ACP_3303 ........................................................ 4
1.4 Scope of Delivery .................................................................... 4
2 Important Remarks.................................................... 4
3 Installation ................................................................. 5
3.1 Upgrade of version ACP_3101 to ACP_3303 ........................ 5
3.2 Upgrade of versions < ACP_1204 with
ARIEL-CPU to ACP_3303 ........................................................ 5
3.3 Exchange of the Hard Disk / Procedure to
install ACP_3303 on a virgin Hard Disk................................. 7
1 General
cPCI-Rack:
• 200 MHz Ariel Board or Oberon Board
• EEPROM Mon 4.26
• MSB, Master-Scan-Board, Revision: 0
with Scan IO-Board Boot EPLD, Version: 0
and Scan IO-Board serial PROM, Version: 0
• Floppy (no SCSI)
• SCSI Hard disk
Optic Module
• For usage as a mammography system, an optic module (identification
number: F8 5146 2700 X) with switchable second laser power of
30 mW and laser control version code 1 is required (visible in info
counters: “Optics module version ...”).
2 Important Remarks
3 Installation
• Select in Service Menu Install from floppy, Software and follow the
guidelines on CR 75.0 display (use ACP_3303 Release Disk 1/3 to 3/3).
• Reboot the system
• Make a new backup (therefore you find an empty floppy within the set of
delivered floppies)
Portex Installation
• Insert the floppy ACP_3303 Portex Installer 1/2 into disk drive of the
digitizer
• Press the Reset on CPU-Board or turn Power Off/On of the digitizer
• Wait until you get the final message on a Terminal connected to the
service RS232 port, or wait until control LED of floppy disk drive is off.
• Insert the floppy ACP_3303 Portex Installer 2/2 into disk drive of the
digitizer
• Press the Reset on CPU-Board or turn Power Off/On of the digitizer
• Wait until you get the final message on a Terminal connected to the
service RS232 port, or wait until control LED of floppy disk drive is off.
• Remove floppy (do not insert any other floppy)
• Press the Reset on CPU-Board or turn Power Off/On of the digitizer
• Wait until you get the final message on a Terminal connected to the
service RS232 port, or wait approximately 2 minutes.
Installation of Application
• Insert the floppy ACP_3303 Release Disk 1/3 into disk drive of the
digitizer
• Press the Reset on CPU-Board or turn Power Off/On of the digitizer
• Wait until you get the final message on a Terminal connected to the
service RS232 port, or wait until control LED of floppy disk drive is off.
• Insert the floppy ACP_3303 Release Disk 2/3 into disk drive of the
digitizer
• Press the Reset on CPU-Board or turn Power Off/On of the digitizer
• Wait until you get the final message on a Terminal connected to the
service RS232 port, or wait until control LED of floppy disk drive is off.
• Insert the floppy ACP_3303 Release Disk 3/3 into disk drive of the
digitizer
Finishing Installation
• If the previous software had a lower version than ACP_1201, then close
the machine doors and scan or erase one cassette.
This cassette needs not be exposed.
For an upgrade of a lower version than ACP_1201 to ACP_3303 a new IP
center calibration is necessary, because of a changed IP stop position before
start of scan. (Calibration procedure, see chapter 3.6 of your Technical
Documentation)
• Make a new backup (therefore you find an empty floppy within the set of
delivered floppies)
• Take the Hard disk formatter floppy of ACP_3303 and insert it into the
floppy disk drive of the digitizer.
• Disconnect the digitizer from the network (since a default IP-Address &
hostname is configured)
• Connect a PC with terminal program or a Terminal to service RS232 port
• Press the Reset button on Ariel-CPU or turn power off/on of the digitizer
• Confirm partitioning/formatting of the hard disk with ‘y’.
• Wait until you get the final message on a terminal connected to the
service RS232 port or wait until control LED of floppy disk drive is off.
• Insert the floppy ACP_3303 Portex Installer 2/2 into the disk drive of
the digitizer
• Press the Reset on Ariel-CPU or turn Power Off/On of the digitizer
• Wait until you get the final message on a terminal connected to the
service RS232 port or wait until control LED of floppy disk drive is off.
• Remove floppy (do not insert any other floppy)
• Press the Reset on CPU-Board or turn Power Off/On the digitizer
• Wait until you get the final message on a Terminal connected to the
service RS232 port, or wait approximately 2 minutes.
• Insert the floppy ACP_3303 Release Disk 1/3 into the disk drive of the
digitizer
• Press the Reset button on Ariel-CPU or turn power off/on of the digitizer
• Wait until you get the final message on a terminal connected to the
service RS232 port, or wait until control LED of floppy disk drive is off.
• Proceed the same way with the ACP_3303 Release Disk 2/3 and 3/3
• Press the Reset button on Ariel-CPU or turn power off/on of the digitizer
and wait until system is up.
Then parse the CPF-File using the guidelines of the Service Menu Topic
Install from floppy, CPF-File, displayed on the user interface.
• If the previous software had a lower version than ACP_1201, then close
the machine doors and scan or erase one cassette.
This cassette needs not be exposed.
For an upgrade of a lower version than ACP_1201 to ACP_3303 a new IP
center calibration is necessary, because of a changed IP stop position before
start of scan. (Calibration procedure, see chapter 3.6 of your Technical
Documentation)
• Make a new backup (therefore you find an empty floppy within the set of
delivered floppies)
• under very special circumstances the info counter contents can be lost or
might be scrambled in case of switching off the device during operation.
• Jitter measurement (EVA) delivers not yet the correct results.
• If 10 folders are in the status ‚error‘. These files have to be manually erased
via Key -Operator mode before the system accepts new cassettes.
• Photomultiplier test may fail (20E0D) due to large tolerances.
• Invalid CPFs are not always rejected, after configuration, there is no hint
that a reset is necessary to apply the changes.
• SW crash to monitor with CPF file configuration, in case of corrupted CPF
file.
*1YX6ZU1*
CR 75.0
Type 5146 / 100 and 101
1 Piece YX6ZU MA1
Centricity CR MP3510
Type 5146 / 200 and 201
List of Contents
1 General.......................................................................................... 3
1.1 Hardware Requirements...................................................................... 3
1.2 Software Requirements ....................................................................... 4
2 Installation .................................................................................... 4
2.1 Upgrade of version ACP_3303 to ACP_3304 ..................................... 4
2.2 Upgrade of versions < ACP_1204 with
ARIEL-CPU to ACP_3304..................................................................... 4
2.3 Exchange of the Hard Disk / Procedure to install
ACP_3304 on a virgin Hard Disk......................................................... 6
3 What is new in ACP_3304 compared with ACP_3303?............ 8
Scope of delivery:
• 1 x Hard disk formatter floppy disk
• 2 x Portex installer floppy disks
• 3 x Application floppy disks
• 1 x Empty floppy
• 1 x Installation Instructions
1 General
1.1 Hardware Requirements
cPCI-Rack:
• 200 MHz Ariel Board or Oberon Board
• EEPROM Mon 4.26
• MSB, Master-Scan-Board, Revision: 0
with Scan IO-Board Boot EPLD, Version: 0
and Scan IO-Board serial PROM, Version: 0
• Floppy (no SCSI)
• SCSI Hard disk
Optic Module:
For usage as a mammography system, an optic module (identification
number: F8 5146 2700 X) with switchable second laser power of
30 mW and laser control version code 1 is required (visible in info
counters: “Optics module version ...”).
2 Installation
• Select in Service Menu Install from floppy, Software and follow the
guidelines on CR 75.0 display (use ACP_3304 Release Disk 1/3 to 3/3).
• Reboot the system
• Make a new backup (therefore you find an empty floppy within the set of
delivered floppies)
Portex Installation
• Insert the floppy ACP_3304 Portex Installer 1/2 into disk drive of the
digitizer
• Press the Reset on CPU-Board or turn Power Off/On of the digitizer
• Wait until you get the final message on a Terminal connected to the
service RS232 port, or wait until control LED of floppy disk drive is off.
• Insert the floppy ACP_3304 Portex Installer 2/2 into disk drive of the
digitizer
• Press the Reset on CPU-Board or turn Power Off/On of the digitizer
• Wait until you get the final message on a Terminal connected to the
service RS232 port, or wait until control LED of floppy disk drive is off.
• Remove floppy (do not insert any other floppy)
• Press the Reset on CPU-Board or turn Power Off/On of the digitizer
• Wait until you get the final message on a Terminal connected to the
service RS232 port, or wait approximately 2 minutes.
Installation of Application
• Insert the floppy ACP_3304 Release Disk 1/3 into disk drive of the
digitizer
• Press the Reset on CPU-Board or turn Power Off/On of the digitizer
• Wait until you get the final message on a Terminal connected to the
service RS232 port, or wait until control LED of floppy disk drive is off.
• Insert the floppy ACP_3304 Release Disk 2/3 into disk drive of the
digitizer
• Press the Reset on CPU-Board or turn Power Off/On of the digitizer
• Wait until you get the final message on a Terminal connected to the
service RS232 port, or wait until control LED of floppy disk drive is off.
• Insert the floppy ACP_3304 Release Disk 3/3 into disk drive of the
digitizer
• Press the Reset on CPU-Board or turn Power Off/On of the digitizer
• Wait until you get the final message on a Terminal connected to the
service RS232 port, or wait until control LED of floppy disk drive is off,
plus some additional time.
• Remove the floppy, press the Reset on CPU-Board or turn Power Off/On
of the digitizer and wait approx. 2 minutes (system extracts SW-release)
• Press the Reset on CPU-Board or turn Power Off/On of CR 75.0 and
wait until system is up
Finishing Installation
• If the previous software had a lower version than ACP_1201, then close
the machine doors and scan or erase one cassette.
This cassette needs not be exposed.
For an upgrade of a lower version than ACP_1007 to ACP_3304 a new IP
center calibration is necessary, because of a changed IP stop position before
start of scan. (Calibration procedure, see chapter 3.6 of your Technical
Documentation)
• Make a new backup (therefore you find an empty floppy within the set of
delivered floppies)
• Take the Hard disk formatter floppy of ACP_3304 and insert it into the
floppy disk drive of the digitizer
• Disconnect the digitizer from the network (since a default IP-Address &
hostname is configured)
• Connect a PC with terminal program or a Terminal to service RS232 port
(9600 baud, Xon/Xoff, no parity, 8 bit, 1 stop bit)
• Press the Reset button on Ariel-CPU or turn power off/on of the digitizer
• Confirm partitioning/formatting of the hard disk with ‘y’
• Wait until you get the final message on a terminal connected to the
service RS232 port or wait until control LED of floppy disk drive is off
• Insert the floppy ACP_3304 Portex Installer 2/2 into the disk drive of
the digitizer
• Press the Reset on Ariel-CPU or turn Power Off/On of the digitizer
• Wait until you get the final message on a terminal connected to the
service RS232 port or wait until control LED of floppy disk drive is off.
• Remove floppy (do not insert any other floppy)
• Press the Reset on CPU-Board or turn Power Off/On the digitizer
• Wait until you get the final message on a terminal connected to the
service RS232 port, or wait approximately 2 minutes
• Insert the floppy ACP_3304 Release Disk 1/3 into the disk drive of the
digitizer
• Press the Reset button on Ariel-CPU or turn power off/on of the digitizer
• Wait until you get the final message on a terminal connected to the
service RS232 port, or wait until control LED of floppy disk drive is off
• Proceed the same way with the ACP_3304 Release Disk 2/3 and 3/3
• Press the Reset button on Ariel-CPU or turn power off/on of the digitizer
and wait until system is up
• Make sure that the digitizer is still disconnected from the network (never
connect the digitizer to the network before configuring it with a valid
CPF-file!)
• Generate the CPF-file with the QS, the VIPS or the CCM
• Parse the CPF-File using the guidelines of the Service Menu topic Install
from floppy, CPF-File, displayed on the user interface.
• Reboot the system
• Reconnect CR 75.0 to the network
• If the previous software had a lower version than ACP_1201, then close
the machine doors and scan or erase one cassette.
This cassette needs not be exposed.
For an upgrade of a lower version than ACP_1007 to ACP_3304 a new IP
center calibration is necessary, because of a changed IP stop position before
start of scan. (Calibration procedure, see chapter 3.6 of your Technical
Documentation)
• Make a new backup (therefore you find an empty floppy within the set of
delivered floppies)
Warning 207a3
E_SCC_FLB_OFFSET_LIMIT,
Meaning: White border could not be removed from mammo image.
Reason: Calculated white border offset out of range.
Cure: Check if begin of scan detection works properly.
Warning 207a4
E_SCC_FLB_SLOPE_LIMIT,
Meaning: White border could not be removed from mammo image.
Reason: Calculated white border slope out of range.
Cure: Check slow-scan skew.
Warning 207a5
E_SCC_FLB_BAD_COLUMNS_LIMIT,
Meaning: White border could not be removed from mammo image.
Reason: Too many columns, where IP border could not be detected.
Cure: Check image. Are there metal parts or similar in the image?
Order-No.: DD+DIS337.05E
*1YZ5F21*
CR 75.0
Type 5146/100
1 Piece YZ5F2 MA1
For the HW and SW requirements refer to section 1.1 and 1.2 in this
document.
List of Contents
BEFORE replacing an old optics module by a new one with mammo features
(identification number: F8 5146 2700 X) installation of ACP_3101 or higher is
required.
BEFORE replacing an Ariel CPU Board by an Oberon Board installation of
ACP_3101 or higher is required.
cPCI-Rack:
• 200 MHz Ariel Board or Oberon Board
• EEPROM Mon 4.26
• MSB, Master-Scan-Board, Revision: 0
with Scan IO-Board Boot EPLD, Version: 0
and Scan IO-Board serial PROM, Version: 0
• Floppy (no SCSI)
• SCSI Hard disk
Optic module:
For usage as a Mammography system, an optic module (identification
number: F8 5146 2700 X) with switchable second laser power of
30 mW and laser control version code 1 is required (visible in info
counters: “Optics module version ...”).
Start with the hard disk formatter floppy if you replaced the hard disk
beforehand. Otherwise, start with the Portex Installer 1/2 floppy.
(1) Take the floppy Hard disk formatter / Portex Installer 1/2 and insert it
into the disk drive of the digitizer.
(2) Disconnect the digitizer from the network (since a default IP address &
hostname is configured).
(4) Press the Reset button on CPU-Board or turn power off/on of the
digitizer.
(5) Confirm partitioning/formatting of the hard disk on your terminal with ‘y’.
(6) Wait until you get the final message on the terminal, or wait until control
LED of floppy disk drive is off.
(7) Insert the floppy Portex Installer 2/2 into the disk drive of the digitizer.
(8) Press the Reset button on CPU board or turn power of the digitizer
off/on.
(9) Wait until you get the final message on the terminal, or wait until control
LED of floppy disk drive is off.
(11) Press the Reset button on CPU board or turn power of the digitizer
off/on.
(12) Wait until you get the final message on the terminal or wait
approximately 2 minutes.
(13) Insert the first floppy of the ACP_3403 Release Disks (labeled
ACP3403_1_3) into the disk drive of the digitizer.
(14) Press the Reset button on CPU board or turn power of the digitizer
off/on.
(15) Wait until you get the final message on the terminal or wait until control
LED of floppy disk drive is off.
(16) Insert the second floppy of the ACP_3403 Release Disks (labeled
ACP3403_2_3) into the disk drive of the digitizer.
(17) Press the Reset button on CPU board or turn power of the digitizer
off/on.
(18) Wait until you get the final message on the terminal or wait until control
LED of floppy disk drive is off.
(19) Insert the third floppy of the ACP_3403 Release Disks (labeled
ACP3403_3_3) into the disk drive of the digitizer.
(20) Press the Reset button on CPU-Board or turn power of the digitizer
off/on.
(21) Wait until you get the final message on the terminal or wait until control
LED of floppy disk drive is off.
(22) Remove the floppy, press the Reset button on CPU board or turn power
of the digitizer off/on and wait approx. 2 minutes (system extracts
software release).
(23) Press the Reset button on CPU board or turn power of digitizer off/on
and wait until the system is up.
If you exchanged the hard disk, restore the original device specific
parameters (using the backup floppy) by following the guidelines of the
Service Menu Topic Install from floppy, Machine specific data.
Take care to use the right backup disk for the digitizer in the backup restore
procedure. (A backup disk with the original Machine Specific Data is
available for any machine from production in case of loss or fault).
If the previous software was a lower version than ACP_1201, then close the
machine doors and scan or erase one cassette. This cassette needs not to
be exposed. (Calibration procedure of PM monitoring).
Do not forget to restore the test images from CDROM CM+9 5145 1055 0
after the restore procedure.
(1) Make sure the digitizer is disconnected from the network (never connect
the digitizer to the network before configuring it with a valid CPF file!).
(2) Generate the CPF file with the QS, the VIPS or the CCM-tool.
(3) Parse the CPF file using the guidelines of the Service Menu Topic Install
from floppy, CPF-File follow the instructions displayed on the user
interface.
If the previous software was lower than ACP_1201, then close the machine
doors and scan or erase one cassette (for calibration of PM monitoring). This
cassette needs not to be exposed.
(6) Refresh your backup by following the guidelines of Service Menu Topic
Save on floppy, Machine specific data.
• Some invalid CPF files will be rejected and an error will be displayed.
• Infocounter
Under very special circumstances the infocounter contents can be lost or
might be scrambled in case of switching off the device during operation.
• Jitter measurement
Jitter measurement (EVA) delivers not yet the correct results.
• Folders in status ‘error’
If 10 folders are in the status ‚error‘, these files have to be manually erased
via Key-Operator mode before the system accepts new cassettes.
• Invalid CPFs
Invalid CPFs are not always rejected. After configuration, there is no hint
that a reset is necessary to apply the changes.
• Additional error message shown in decimal mode
The operating system application error.ppc interprets a given error number
hexadecimal as well as decimal (while actually error numbers are only used
in hexadecimal way). This leads to a second (wrong) error message, if a
second error number exists that decimally corresponds to the given error
number.
Only the first error message shows the right corrective action.
Order-No.: DD+DIS206.06
CR 75.0
*16BMWO1*
1 Piece 6BMWO MA1
Type 5146 / 100, 101
CONFIDENTIALITY NOTE:
Use, dissemination, distribution or reproduction of this document by unauthorized personnel is not permitted and may be unlawful.
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
List of contents
1.2 CR Applications
REQUIRED TOOLS:
You need max. 7 empty formatted floppy disks with 1.44 MB capacity each:
• one for the backup of site specific data and
• six for the preparation of the installation set of floppy disks
if you obtained the digitizer software from MedNet.
9 pins 25 pins
figure 1
REQUIRED TIME:
• approximately 30 minutes for the software upgrade or
• approximately 60 minutes for the installation on the blank hard disk.
3 Installation Procedure
3.1 Making Installation Floppy Disks
Execute these actions if you downloaded the digitizer software from MedNet.
Skip them if you obtained the software via the spare parts supply.
(2) Extract all files from ACP_4005.zip into an empty folder on the
service PC.
NOTE:
For each of these folders, a floppy disk with the same name is created
in the following steps.
(4) Select the floppy drive entry (A:) in the Windows Explorer.
(6) Select tab General and enter the corresponding folder name,
e.g. HDFORMAT.
(8) Copy the contents of the folder (e.g. HDFORMAT) from the service PC
to the floppy disk.
(9) Remove the floppy disk from the drive and label it with the folder name,
e.g. HDFORMAT.
(10) Repeat steps (3) to (9) for the other five floppy disks.
Result Now you have the required installation set of floppy disks.
Before you do the software update, create a backup floppy disk with the
machine specific data in case you have to restore it later
in case of failure of the software installation.
(2) Enter the Service menu on the digitizer keypad and select
Save on floppy / Machine specific data.
NOTE:
In the rare case that a backup floppy cannot be generated in this way,
obtain the out-of-factory backup data using this link:
http://intra.agfanet/BU/MI/MEDNET3.NSF/EN/factorybackupftp.htm
(1) Enter the Service menu on the digitizer keypad and select
Install from floppy / Software.
(2) Insert the floppy disk labeled PORTEX1 into the disk drive of the digitizer.
(4) Wait until the control LED of the floppy disk drive is off.
(6) Insert the floppy disk labeled PORTEX2 into the disk drive of the digitizer.
(8) Wait until the control LED of floppy disk drive is off.
NOTE:
Do not insert any other floppy disk!
WAITING TIME:
Wait approx. 2 minutes.
The files of the Portex operating system are installed.
(11) Insert the floppy disk labeled ACP4005_1_3 into the disk drive of the
digitizer.
(13) Wait until the control LED of floppy disk drive is off.
(15) Insert the floppy disk labeled ACP4005_2_3 into the disk drive of the
digitizer.
(17) Wait until the control LED of floppy disk drive is off.
(19) Insert the floppy disk labeled ACP4005_3_3 into the disk drive of the
digitizer.
(21) Wait until the control LED of floppy disk drive is off.
WAITING TIME:
Wait approx. 2 minutes.
The system extracts the ACP_4005 software.
(24) Close the front door and switch off/ on the digitizer.
Result Normally, the digitizer performs its self test and afterwards changes its state
to Ready. Proceed with section 3.5.
If the digitizer does not perform the self test, the software installation was not
successful. The cause is a corruption of the file system on the hard disk.
Perform the installation according to section 3.4.
NOTE:
This procedure must be carried out only if the hard disk of the digitizer is
corrupted or has been replaced by a blank one.
In this case the hard disk must be formatted using the installation disk
HDFORMAT.
(2) Connect your service PC to the RS232 port at the cPCI rack.
(4) Create a new connection and select your serial port (e.g., COM1).
(5) Set terminal settings to 9600 baud, Xon/Xoff, no parity, 8 bit, 1 stop bit.
(7) Insert the floppy disk labeled HDFORMAT into the disk drive of the
cPCI rack.
(9) Confirm partitioning/formatting of the hard disk on your terminal with ‘y’.
(12) Insert the floppy disk labelled PORTEX2 into the disk drive of the
digitizer.
NOTE:
Do not insert any other floppy disk!
(19) Insert the floppy disk labeled ACP4005_1_3 into the disk drive of the
digitizer.
(23) Insert the floppy disk labeled ACP4005_2_3 into the disk drive of the
digitizer.
(27) Insert the floppy disk labeled ACP4005_3_3 into the disk drive of the
digitizer.
(35) Enter the Service menu on the digitizer keypad and select
Install from floppy / Machine specific data.
NOTE:
Without restoring the device-specific parameters, an operation of the
digitizer is impossible.
(37) Restore the test image files from CD (CM+9 5145 3055 0) to the
directory E:\testimg on the digitizer.
(38) Generate the CPF file with the QS, the VIPS or the CCM tool. Please
refer to the service documentation, chapter 1, for details.
(39) Enter the Service menu on the digitizer keypad and select
Install from floppy / CPF-File.
(43) Close the front door and switch digitizer power off/ on.
Result The digitizer performs its self test and afterwards changes its state to Ready.
Proceed with section 3.5.
NOTE:
Skip action (1) if the previous software version was ACP_1201 or higher.
(3) Refresh your backup by following the guidelines of Service Menu topic
Save on floppy / Machine specific data. nochmal???
NOTE:
If the version of the previous software was less than ACP_1007, perform
IP Center and Shading calibrations. Please refer to the service
documentation, chapter 3.6.
(1) Enter the Service menu on the digitizer keypad and select
Info / Device info.
Result If the entry Software is equal to ACP_4005, the software installation was
successful.
Otherwise repeat the installation and closely follow the instructions in
section 3.
1 The language for the Operator and Key Operator menus cannot be
changed since the necessary language files are not available.
This bug does not impact the digitizer operation.
CR 75.0
Type 5146/100, 101
► Document History
► Referenced Documents
Document Title
Preceding Bulletin Release of ACPL1004 Language Files for Digitizer User
Interface of CR 75.0, DD+DIS089.07E
WARNING:
Improper operation or service activities may cause damage or injuries.
INSTRUCTION:
(1) Read the "Generic Safety Directions" document
(see MEDNET GSO => General Info => Agfa HealthCare => Publications =>
Service Manual) prior to attempting any operation, repair or maintenance task on
the equipment.
(2) Strictly observe all safety directions within the "Generic Safety Directions" and
on the product.
LIST OF CONTENTS
This change does not affect CR 75.0 operation and service. Therefore the update of
digitizer software is not necessary if software ACP_4005 is already installed.
See chapter 3.8 of the service manual for the software release information.
1.2 CR Applications
REQUIRED TOOLS:
You need max. 8 empty formatted floppy disks with 1.44 MB capacity each:
• One for the backup of site specific data and
• Seven for the preparation of the installation set of floppy disks
if you obtained the digitizer software from MedNet.
9 pins 25 pins
figure 1
REQUIRED TIME:
• approximately 30 minutes for the software upgrade or
• approximately 60 minutes for the installation on the blank hard disk.
3 Installation Procedure
3.1 Making Installation Floppy Disks
Execute these actions if you downloaded the digitizer software from MedNet.
Skip them if you obtained the software via the spare parts supply.
(2) Extract all files from ACP_4007.zip into an empty folder on the service PC.
NOTE:
For each of these folders, a floppy disk with the same name is created in the following
steps.
(3) Insert an empty floppy disk.
(4) Select the floppy drive entry (A:) in the Windows Explorer.
(6) Select tab General and enter the corresponding folder name,
e.g. HDFORMAT.
(8) Copy the contents of the folder (e.g. HDFORMAT) from the service PC to the
floppy disk.
(9) Remove the floppy disk from the drive and label it with the folder name, e.g.
HDFORMAT.
(10) Repeat steps (3) to (9) for the other five floppy disks.
Result Now you have the required installation set of floppy disks.
(2) Enter the Service menu on the digitizer keypad and select
Save on floppy / Machine specific data.
NOTE:
In the rare case that a backup floppy cannot be generated in this way, obtain the
out-of-factory backup data using this link:
http://intra.agfanet/BU/MI/MEDNET3.NSF/EN/factorybackupftp.htm
(1) Enter the Service menu on the digitizer keypad and select
Install from floppy / Software.
(3) Close the front door and switch off/ on the digitizer.
Wait until the system is ready.
Result The digitizer software is installed.Proceed with section 3.6.
(1) Enter the Service menu on the digitizer keypad and select
Install from floppy / Software.
(2) Insert the floppy disk labeled PORTEX1 into the disk drive of the digitizer.
(4) Wait until the control LED of the floppy disk drive is off.
(6) Insert the floppy disk labeled PORTEX2 into the disk drive of the digitizer.
(8) Wait until the control LED of floppy disk drive is off.
NOTE:
Do not insert any other floppy disk!
WAITING TIME:
Wait approx. 2 minutes.
The files of the Portex operating system are installed.
(11) Insert the floppy disk labeled ACP4007_1_3 into the disk drive of the digitizer.
(13) Wait until the control LED of floppy disk drive is off.
(15) Insert the floppy disk labeled ACP4007_2_3 into the disk drive of the digitizer.
(17) Wait until the control LED of floppy disk drive is off.
(19) Insert the floppy disk labeled ACP4007_3_3 into the disk drive of the digitizer.
(21) Wait until the control LED of floppy disk drive is off.
WAITING TIME:
Wait approx. 2 minutes.
The system extracts the ACP_4007 software.
(24) Close the front door and switch off/ on the digitizer.
Result Normally, the digitizer performs its self test and afterwards changes its state to Ready.
Proceed with section 3.6.
If the digitizer does not perform the self test, the software installation was not successful.
The cause is a corruption of the file system on the hard disk.
Perform the installation according to section 3.5.
NOTE:
This procedure must be carried out only if the hard disk of the digitizer is corrupted or
has been replaced by a blank one.
In this case the hard disk must be formatted using the installation disk HDFORMAT.
(2) Connect your service PC to the RS232 port at the cPCI rack.
(4) Create a new connection and select your serial port (e.g., COM1).
(5) Set terminal settings to 9600 baud, Xon/Xoff, no parity, 8 bit, 1 stop bit.
(7) Insert the floppy disk labeled HDFORMAT into the disk drive of the
cPCI rack.
(9) Confirm partitioning/formatting of the hard disk on your terminal with ‘y’.
(12) Insert the floppy disk labelled PORTEX2 into the disk drive of the digitizer.
NOTE:
Do not insert any other floppy disk!
(19) Insert the floppy disk labeled ACP4007_1_3 into the disk drive of the digitizer.
(23) Insert the floppy disk labeled ACP4007_2_3 into the disk drive of the digitizer.
(27) Insert the floppy disk labeled ACP4007_3_3 into the disk drive of the digitizer.
(35) Enter the Service menu on the digitizer keypad and select
Install from floppy / Machine specific data.
NOTE:
Without restoring the device-specific parameters, an operation of the digitizer is
impossible.
(37) Restore the test image files from CD (CM+9 5145 3055 0) to the directory
E:\testimg on the digitizer.
(38) Generate the CPF file with the QS, the VIPS or the CCM tool. Please refer to the
service documentation, chapter 1, for details.
(39) Enter the Service menu on the digitizer keypad and select
Install from floppy / CPF-File.
(42) Close the front door and switch digitizer power off/ on.
Result The digitizer performs its self test and afterwards changes its state to Ready. Proceed
with section 3.6.
(2) Open Service menu on the local key pad of the digitizer.
(3) Choose
<INSTALL from floppy>
<language files>
(5) Insert floppy disk with ACPL1004 language files at disk drive of cPCI-Rack
and confirm with <;> on the local key pad.
(6) After a short while following message appears on the Service menu:
<Please remove the floppy and press ;>
To change the user interface language press ;. To leave the menu without
changing the language press :.
NOTE:
Skip action (1) if the previous software version was ACP_1201 or higher.
(3) Refresh your backup by following the guidelines of Service Menu topic Save on
floppy / Machine specific data.
NOTE:
If the version of the previous software was less than ACP_1007, perform IP Center and
Shading calibrations. Please refer to the
service documentation, chapter 3.6.
NOTE:
If you use the digitizer for Mammography applications, enable “Mammo” manually in the
Service Menu of the digitizer.
(1) Enter the Service menu on the digitizer keypad and select
Info / Device info.
Result If the entry Software is equal to ACP_4007, the software installation was successful.
Otherwise repeat the installation and closely follow the instructions in section 3.
CR 75.0
Type 5146/100,101
► Document History
► Referenced Documents
Document Title
n.a. n.a.
WARNING:
Improper operation or service activities may cause damage or injuries.
INSTRUCTION:
(1) Read the "Generic Safety Directions" document
(see MEDNET GSO => General Info => Agfa HealthCare => Publications =>
Service Manual) prior to attempting any operation, repair or maintenance task on
the equipment.
(2) Strictly observe all safety directions within the "Generic Safety Directions" and
on the product.
TABLE OF CONTENT
The following modifications of CR 75.0 Type 5146 are listed in chronological order, the
latest modification on top of the list.
2006
October 5. Modification of the 5-fold Stepper Motor Board n.a.
July 6. Modification of the Erasure Unit n.a.
May 7. Introduction of Digitizer Software ACP_4007 FSB No. 47:
DD+DIS204.06E,
Enclosure: DD+DIS200.06E,
Chapter 3.8 of the Service
Documentation:
DD+DIS205.06E
April 8. Introduction of Digitizer Software ACP_4005 FSB No. 46:
DD+DIS104.06E,
Enclosure: DD+DIS159.06E,
Chapter 3.8 of the Service
Documentation:
DD+DIS205.06E
February 9. Delivery of Auto QC1 Test Protocol n.a.
2005
October 10. Introduction of Digitizer Software ACP_3403 FSB No. 43:
DD+DIS316.05E,
Enclosure: DD+DIS315.05E,
Chapter 3.8 of the Service
Documentation:
DD+DIS205.06E
July 11. Introduction of new Type Label n.a.
July 12. Modified Cassette Opener Component n.a.
June 13. Introduction of Digitizer Software ACP_3304 FSB No. 40:
DD+DIS200.05E,
Chapter 7 of the service
manual: DD+DIS201.05E
March 14. Increased Dimensioning of cPCI Power Supply FSB No. 36: DD+DIS067.05E
2004
October 15. Upgrade of ADC Compact Plus to CR 75.0 Enclosure: DD+DIS302.04E
(Mammo 1B)
October 16. Repleacement of CPU Ariel by CPU Oberon Enclosure: DD+DIS302.04E
Sep. 17. Introduction of Digitizer Software ACP_3303 FSB No. 34:
DD+DIS326.04E,
Chapter 7 of the service
manual: DD+DIS327.04E
July 18. Modification of the Erasure Unit n.a.
June 19. Introduction of Digitizer Software ACP_3204 and Enclosure: DD+DIS144.04E,
Slow Scan Unit with 17:1 Gear FSB No. 33: DD+DIS234.04E
February 20. Introduction of Digitizer Software ACP_3101 Chapter 7: DD+DIS070.04E,
FSB No. 30: DD+DIS095.04E
SB – Service Bulletin
Reason A modified vacuum pump has been introduced for the use of the
digitizer on sites between 2000 m and 4000 m above sea level.
Product(s)/ The High Altitude Vacuum Pump is not part of series production
Serial and only available as upgrade kit for new installations or already
Number(s) installed digitizers as well as spare part for the replacement of
modified vacuum pumps, which are defective.
Part • Upgrade Kit “High Altitude Vacuum Pump for CR 75.0/CR 85-
Number(s) X”
ABC Code: EZKFF
• Spare part “High Altitude vacuum pump”,
CM+9 5148 6230 0
Referenced • Service Bulletin No. 52, DD+DIS277.07E
Document(s)
• Enclosure - Replacement Instructions: High Altitude Vacuum
Pump for Use on Sites between 2000 m and 4000 m above
Sea Level, DD+DIS151.07E
Reason User interface language files have been released to ensure that
messages in the digitizer user interface are available in several
languages. Eleven languages can be used with the release of
ACPL1004 language files.
Product(s)/ As no CR 75.0 Type 5146/101 and ADC Compact Plus Type
Serial 5146/100 are produced, this modification is only available as
Number(s) spare part.
Part • ACPL1004 language files can be obtained by Download
Number(s) from Mednet GSO Library:
Computed Radiography Æ CR Digitizers Æ CR 75.0 Æ
Freeware
4 Modification of cPCI-Rack
Reason The power supply unit has been modified to enable the use of hard
disks with higher voltage input.
The modified power supply unit is RoHS compliant.
Product(s)/ As no CR 75.0 Type 5146/101 and ADC Compact Plus Type
Serial 5146/100 are produced, this modification is only available as
Number(s) spare part.
Part Following spare parts have been modified:
Number(s)
• Spare part “cPCI-Rack”, CM+9 5146 1400 6
• Spare part “cPCI-Rack assembled”, CM+9 5146 1405 7
• Spare part “PCI power supply”, CM+9 5146 1430 2
Referenced n.a.
Document(s)
Reason The fixation of the 5-fold stepper motor board has been changed by
introducing new clips to prevent bending of the board.
Reason From now on the Auto QC1 test protocol of the digitizer will be in
scope of delivery for all new devices.
Reason The new type label is compliant with the legal directive 2002/96/EC
Reason The power supply was too poor to cope with the increased power
consumption caused by the IO-bus nodes check during boot
sequence.
Product(s)/ Type 5146/101: SN = 3834, 3874, 3875, 3876, 3878, 3879,
Serial SN ≥ 3881, CW 09/2005
Number(s)
Type 5146/100: SN 2812 – 2821, CW 09/2005
Part • Spare part “cPCI rack”: CM+9 5146 1405 6
Number(s)
• Spare part “PCI power supply complete”: CM+9 5146 1430 1
Referenced For more information on the issue refer to FSB No. 36
Document(s) (DD+DIS067.05E)
Reason Release of the Oberon board as successor of the the Ariel board.
Reason • Together with this software slow scan gear 17:1 is introduced.
• Software is updated to allow usage of both old (12:1) and
new (17:1) scanner types.
• Speed of postscan robot for motions into and out of points is
reduced (motor frequency: 3001 Hz -> 1500 Hz).
• New Mammo Factor 1.19
• Software solves various bugs.
Product(s)/ Type 5146/101: SN 3283, > 3307, CW 27/2004
Serial
Type 5146/201: SN > 5501, CW 27/2004
Number(s)
Part • Spare part “slow scan unit”: CM+9 5146 2200 3
Number(s)
• Spare part “software ACP_3204”: CM+9 5146 1050 8
Download from MedNet, GSO Library:
Computed Radiography/CR Digitizers/CR 75.0/Freeware/
CR 75.0 – Software – Device Software ACP_3204
Referenced For the installation instructions of the software refer to the
Document(s) enclosure document: DD+DIS144.04E
For the release Information refer to FSB No. 33: DD+DIS234.04E
Order-No.: DD+DIS238.05E
*1YY3LV1*
CR 75.0
Type 5146 / 100 and 101
1 Piece YY3LV MA1
Centricity CR MP3510
Type 5146 / 200 and 201
Internal update # 3
Caution:
This system uses mains voltage. Please consider the respective safety regulations.
These instructions describe adjustments and routines which must only be performed by qualified
technical personnel.
Note:
CE Declaration:
The CE Declaration (CE Conformity) becomes invalid if the product is changed without explicit
consent of the manufacturer! This applies to all parts, not only to safety elements.
Chapter 9:
List of Contents
1 Safety ......................................................................... 1
1 Safety
• The CR 75.0 has been designed for scanning medical X-ray image plates
and should only be used for these purposes.
• The CR 75.0 must only be operated by qualified staff trained on the
machine.
• Make sure that the CR 75.0 is constantly monitored in order to avoid
inappropriate handling, especially by children.
• Only trained service personnel must make repairs. Only authorized service
personnel must make changes to the CR 75.0.
• If there is any visible damage to the machine casing, do not start nor use
the CR 75.0.
• If you want to connect the CR 75.0 with other devices, components or
assemblies and if the technical data do not permit determining whether the
combination with these devices, components or assemblies involves
hazards, you must consult the respective manufacturers to avoid danger
for operating personnel or the environment.
• Do not override or disconnect the integrated safety features.
• Switch off the CR 75.0 before performing any maintenance work or
repairs. Disconnect the CR 75.0 from the mains before making repairs or
performing any maintenance activities.
• As is the case for all technical devices, the CR 75.0 must be operated,
cared for and serviced correctly.
• If you don't operate the CR 75.0 correctly or if you don't have it serviced
correctly, Agfa-Gevaert is not liable for resulting disturbances, damages or
injuries.
• When installing the CR 75.0, care must be taken to ensure that there is
either a mains plug or an all-cable disconnecting device in the internal
installation fitted near the CR 75.0 and that it is easily accessible.
• If you notice conspicuous noise or smoke, disconnect the CR 75.0
immediately.
• Check that the mains voltage is within the specified range of the self
adapting power supply of the machine.
• You can hurt your fingers if they are caught between the ADC Cassette
and the edge of the input slot. Insert the cassette in the input buffer as
described in the User Manual. At all times, keep your fingers clear of the
input slot. As soon as the CR 75.0 takes in the cassette, release it.
2 General Information
To ensure quality and functional reliability of the system all the points listed
below (minimum maintenance points) must be carried out.
Evaluation of infocounters.txt
What to check in the Comment
infocounter
2.2 Throughput For throughput most important are the cycles per
day. They usually count between 50 and 200.
3.3 Hardware By comparing the status of the device with the
Modification History available “Field Modifications”, chapter 7, the
exact hardware status can be determined.
3.4 Software Modification Check the software modification history to
History determine, whether a recent software upgrade
solved a problem, that occurred quite often in the
error list.
4.6 Laser Power Check that laser power is constant: no more than
1 mW difference in between two entries. If the
value is higher, run diagnostic software and scan
& signals and check laser diode module.
4.7 Polygon Jitter If amplitude exceeded 300 milli-pixel, check
Monitoring accurately the image quality on the workstation.
Do not exchange optic modules without having
contacted the Support Center.
The entries are only indications and can
only be interpreted as one symptom
which is conducted to the optic module.
5.3 Retries Retries lead to less throughput of the device. If
the fraction of retries exceeds 1%, compare with
frequent error codes.
5.7 Error History Check errors since last maintenance, how often
they appeared as well as the CBF (cycles
between failures) of these errors. This gives an
overview of the current status of the machine.
Compare frequently occurring errors with the
error list in chapter 3.3, and take actions.
5.8 Error List Relatives Troubleshoot these errors using the technical
and documentation, chapter 3.3, “Troubleshooting”.
5.9 Error List Total
• Flashlight
(1) Check overall condition of the machine – outside and inside – for obvious
changes or damage.
3.2 Inside
• Vacuum cleaner
• Lint free cloth
• Dust brush
• Roller
(CM+9 5145 9100 0)
• Toothed belt
(CM+9 5145 5195 0)
5146_reg02_000cdr
Figure 2
Belt (3) Exchange the transport belt
once a year.
• Soft cloth
• Air filter
(CM+9 5146 1799 2)
5146_reg09_007.cdr
Figure 4
• Vacuum cleaner
• Soft cloth
• Air filter
(CM+9 5146 1799 2)
Figure 5
5146_reg09_012.cdr
Figure 6
• Soft cloth
• ADC Cleaner
(if not available, use water)
• Discharge brush
(CM+9 5145 2442 1)
The scan rollers have to be cleaned in place and must not be removed.
To move the scan rollers, just turn the drive of the slow scan motor
manually.
5146_reg09_002.cdr
Figure 7
• ADC Cleaner
(if not available, use water)
• Suction cups
(CM+9 5145 6550 2)
Figure 8
• Soft cloth
• Dust brush
• ADC Cleaner
(if not available, use water)
• 1 set of 10 pieces 100 W
halogen erasure lamps:
Philips CAPSULEline Pro
(CM+9 0450 6582 0)
5146_reg09_003.cdr
Figure 9
In case of persistent dirt, you may also use ADC Cleaner for cleaning all
the surfaces with the exception of the inner side of the large glass plate.
This side must not be cleaned with anything wet since a gelatin layer is
attached to it.
3.9 Cassettes
• Aluminum label
(CM+9 8300 1131 0)
- Locking
- Opening leaf springs Figure 11
- Aluminum label
(3) Attach missing aluminum
label.
• Flatfield
Make sure that the outlined areas are filled in as shown in the example.
(4) Insert the cassette into the digitizer and print the image on a printer with a
window setting of 0.6 without changing the level setting.
Figure 13
If there are still unacceptable effects, please contact the Support Center.
Figure 15
If there are still unacceptable effects please contact the Support Center.
If there are still unacceptable effects please contact the Support Center.
5 Completion of Maintenance
(1) Confirm the maintenance by signing the checklist.
(2) Make a backup of the system on floppy.
(3) Inform the customer about what was done during the maintenance and
which repairs need to be done in next future.
Published by
Agfa-Gevaert AG
Fototechnik
Tegernseer Landstraße 161
D - 81539 München
CR 75.0
Type 5146 / 100 and 101
Centricity CR MP3510
Type 5146 / 200 and 201
Transport Units • Check position of suction cups and clean them with
ADC cleaner
Erasure Unit • Clean grid of the fan
• Remove dust from reflector, input and output opening of
air stream, KG2 filter, outer front panel
• Check the KG2 filter
• Exchange lamps every maintenance
Cassettes • Check the following items of the cassette:
outside condition, hinges, locking, opening leaf springs,
aluminum label
• Attach missing aluminum labels
Remarks:
...................................................................................................................................................
...................................................................................................................................................
...................................................................................................................................................
...................................................................................................................................................
...................................................................................................................................................
...................................................................................................................................................
...................................................................................................................................................
................................................................... ...................................................................
Date / Signature Service Technician Customer
Centricity CR MP3510
Type 5146 / 200 and 201
► Please note
Service Bulletins are not part of the Service Manual for Download.
Edition 1, Revision 0
11-2008 printed in Germany Document Node ID: 26646152
service_bulletin_cover_e_template_v01
Agfa Company Confidential Copyright © 2008 Agfa HealthCare N.V.
DD+DIS002.04E Service Bulletins
Internal update # 1
The Installation Planning Instructions are separately available. Order No: DD+DIS227.05E
Chapter 11
List of Contents
1 Safety ............................................................................. 1
1.1 General Safety Instructions ........................................................ 1
1.2 Safety Instructions for Laser Products...................................... 4
1.3 Safety Instructions for Cassette Unit ......................................... 4
1.4 Safety Regulations ...................................................................... 4
7 Specifications.............................................................. 12
7.1 Type Overview ........................................................................... 12
7.2 Electrical Data ............................................................................ 12
7.3 Packing Dimensions and Weights ........................................... 13
7.3.1 Package................................................................................................. 14
7.3.2 Weights.................................................................................................. 14
7.3.3 Transport Conditions ............................................................................. 14
1 Safety
1.1 General Safety Instructions
• The CR 75.0 has been designed for scanning medical X-ray image plates
and should only be used for these purposes.
• The CR 75.0 must only be operated by qualified staff trained on the
machine.
• Make sure that the CR 75.0 is constantly monitored in order to avoid
inappropriate handling, especially by children.
• Only trained service personnel must make repairs. Only authorized
service personnel must make changes to the CR 75.0.
• If there is any visible damage to the machine casing, do not start nor use
the CR 75.0.
• If you want to connect the CR 75.0 with other devices, components or
assemblies and if the technical data do not permit determining whether
the combination with these devices, components or assemblies involves
hazards, you must consult the respective manufacturers to avoid danger
for operating personnel or the environment.
• Do not override or disconnect the integrated safety features.
• Switch off the CR 75.0 before performing any maintenance work or
repairs. Disconnect the CR 75.0 from the mains before making repairs or
performing any maintenance activities.
• As is the case for all technical devices, the CR 75.0 must be operated,
cared for and serviced correctly.
• If you don't operate the CR 75.0 correctly or if you don't have it serviced
correctly, Agfa-Gevaert is not liable for resulting disturbances, damages or
injuries.
• When installing the CR 75.0, care must be taken to ensure that there is
either a mains plug or an all-cable disconnecting device in the internal
installation fitted near the CR 75.0 and that it is easily accessible.
• If you notice conspicuous noise or smoke, disconnect the CR 75.0
immediately.
• Check that the mains voltage is within the specified range of the self
adapting power supply of the machine.
• You can hurt your fingers if they are caught between the ADC Cassette
and the edge of the input slot. Insert the cassette in the input buffer as
described in the User Manual. At all times, keep your fingers clear of the
input slot. As soon as the CR 75.0 takes in the cassette, release it.
In order to reduce the risk of electric shock, do not remove any covers.
Caution hot:
Type B equipment:
Indicates that the CR 75.0 complies with the limits for type B equipment.
Intergrounding connector:
Provides a connection between the Digitizer and other equipment which might
exhibit minor ground potential differences. These differences may degrade
the quality of communication between different equipment. Never remove
connections to this terminal.
Provides a connection between the Digitizer and the protective earth of the
mains. Do not remove this connection, because this will have a negative
influence on the leakage current.
Power on
Power off:
Note that the power cord has to be disconnected from the wall outlet in order
to disconnect the unit entirely from the mains.
Installation Regulations
Electrical installations in the installation room must be in compliance with:
• IEC 60364 (e.g. for Germany: VDE 0100) regulations for Europe,
• NEC for USA / Canada.
Mains Connection
Connection via plug and socket.
Certificates
• The CR 75.0 Digitizer is in compliance with EG regulation 93/42/EEC
(Medical device directive).
• The CR 75.0 Digitizer is tested in compliance with:
– IEC 601-1 1988 + A1 : 1991 + A2 : 1995
– EN 60601-1 : 1990 + A1 : 1993 + A2 : 1995
– EN 60601-1-2 : 1993
– EN 60825-1:1994 + A11:1996 : 2001
– UL2601-1 Second Edition 22 No. 601.1 – M909
– The CR 75.0 Digitizer is UL / CUL classified.
– E166741 (PIDF)
Pieces Description
1 CR 75.0 Digitizer
1 base plate
1 installation instructions
1 installation report
1 CR 75.0 user manual information
1 certificates
6 floppies
1 UTP network cable (2 x RJ45 connection)
1 Allen key, size 8
1 connection cable (Europe: CEE(7) VII 250 V/16 A)
1 connection cable (USA, Japan: Nema 6-15 P)
2 adjustable bases, riveted
3 System Overview
rk
etwo
n
rk
etwo
n
rk
two
ne
on
ly
Figure 1
The above example shows the Digitizer connected to a WIN NT based ADC
Quality System (ADC QS). It may as well be connected to a Unix based ADC
System (VIPS Processing Station).
CR 75.0
Digitizer
B
A
complete modules.
5 cm 61 cm (24.03 In)
(1.97 In)
106 cm (47.76 In)
dimensions in cm
3 cm 1 cm (1.182 In 0.39 In)
138 cm (54.37 In)
84 cm (33.09 In)
68 cm (26.79 In)
93 cm (36.64 In)
115 cm (45.31 In)
514611ac.cdr
Figure 2
Printed copies of this document are not controlled.
The reader should always refer to MedNet to ensure that they are working with the latest copy of this document.
Chapter 11 / 8 CR 75.0 Edition 1, Revision 9
Agfa Company Confidential (Type 5146/100 and 101) 2005-08-08
1 2 3
DD+DIS227.05E Installation Planning
It must be possible to transport the digitizer through all hallways and doors up
to the installation site.
Once the machine is unpacked, it can be moved to the installation site on four
mounted rollers.
To remove the machine off the pallet a minimum space of 100 cm x 500 cm
(39.4 In x 197 In) is required.
Make sure that the floor covering is solid enough to stand the weight of the
digitizer.
Light Tightness
tightness SALMIN still met at 2500 Lux ambient
light
Floor Conditions
base Digitizer stands on a metal plate, 2 mm
(0,08 In) thickness
horizontal alignment Digitizer works without leveling,
horizontal alignment is not required.
Nevertheless, the digitizer should be run
preferably in a horizontal position.
Take care that the digitizer can not tilt at
the installation site.
Magnetic Fields
max. permissible magnetic field in according to EN 61000-4-8: Level 5
the room
Emissions
noise level during scanning max. 65 dB (A)
stand-by mode / idle max. 46 dB (A)
heat dissipation 0.8 kWh / 2730 BTU per hour
at max. throughput with 35x43 cm IP
6 Electrical Connections
7 Specifications
7.1 Type Overview
Digitizer CR 75.0
Type 5146/100 and 101/200 and 201
Example
6
Ampere
0
:3 7
:3 3
:3 9
:3 5
:3 2
:3 8
:3 4
:3 1
:3 7
:3 3
:3 9
:3 6
:3 2
:3 8
:3 4
:3 1
:3 7
13 4:23
:3 9
:3 6
:3 2
:3 8
54
13 2:3
13 2:4
13 2:4
13 2:5
13 3:0
13 3:0
13 3:1
13 3:2
13 3:2
13 3:3
13 3:3
13 3:4
13 3:5
13 3:5
13 4:0
13 4:1
13 4:1
13 4:2
13 4:3
13 4:4
13 4:4
4:
:3
:3
13
Figure 3
7.3.1 Package
Packed on a pallet the size is:
height 159 cm (62.64 In)
width 100 cm (39.4 In)
depth 132 cm (52.01 In)
7.3.2 Weights
Customer: ........................................................................................
.................................
Department: ........................................................................................
.................................
System ........................................................................................
components: .................................
........................................................................................
.................................
........................................................................................
.................................
........................................................................................
.................................
........................................................................................
.................................
Remarks:
8.1 Checklist
= main components
2
or ADC Preview Station
Published by
Agfa-Gevaert AG
Fototechnik
Tegernseer Landstraße 161
D - 81539 München
Chapter 12
List of contents
1 Glossary..................................................................1
1 CR 75.0 Glossary
Centricity CR MP3510
Order Number:
DD+DIS002.04E
Type 5146 / 200 and 201
1st edition
CR 75.0
Type 5146 / 100 and 101
Centricity CR MP3510
Order Number:
DD+DIS002.04E
Type 5146 / 200 and 201
1st edition
CR 75.0
Type 5146 / 100 and 101
Centricity CR MP3510
Order Number:
DD+DIS002.04E
CR 75.0
Type 5146 / 100 and 101
Centricity CR MP3510
Order Number:
DD+DIS002.04E