You are on page 1of 4

Calibration technology

Temperature dry-well calibrator


Model CTD9100-375
WIKA data sheet CT 41.32

Applications

■■ Testing and calibration of temperature measuring


instruments
■■ Reference instrument for on-site laboratories for the
calibration of thermometers
■■ Also suitable for on-site calibration

Special features

■■ High accuracy and stability


■■ Possibility to check temperature switches
■■ Low weight and compact design
Temperature dry-well calibrator model CTD9100-375
■■ Simple operation

Description

Versatile in application Easy to use


Nowadays, fast and simple testing of thermometers is a The temperature dry-well calibrators of the CTD9100 series
“must” when it comes to the operational safety of machines work with temperature-controlled metal blocks and inter-
and plants. changeable inserts.
The portable calibrators of the CTD9100 family are particularly
suited to local calibration tasks and extremely user-friendly. The calibration temperature, adjusted simply using two keys
Due to their compact design and their low weight, the on the controller, can be very quickly controlled. The set
instruments can be taken and used almost anywhere. The temperature of the heating block is displayed on a large,
CTD9100-375 in particular is characterised by its robust case 4-line, high-contrast LED display. Thus reading errors are
design and compact dimensions. virtually eliminated.

The new instrument concept brings together a stable heat Thermometers with different diameters can be fitted into the
source with precision Pt100 temperature measurement. This calibrator using inserts, drilled to suit. A new block design,
enables industrial temperature probes to be calibrated even with improved temperature homogeneity at the calibrator's
more efficiently. Regular monitoring of temperature probes lower range, leads to smaller measurement uncertainties.
helps to recognise failures promptly and shorten downtimes.

WIKA data sheet CT 41.32 ∙ 08/2015 Page 1 of 4


Specifications Model CTD9100-375
Display
Temperature range tamb … 375 °C (tamb ... 707 °F)
Accuracy 1) ± 0.5 ... 0.8 K
Stability 2) ± 0.05 K to 100 °C (212 °F)
Resolution 0.1 °C
Temperature control
Heating time 6 min from 20 °C to 300 °C (from 68 °F to 572 °F)
Cooling time 14 min from 300 °C to 60 °C (from 572 °F to 140 °F)
Stabilisation time 3) 5 min
Insert
Immersion depth 100 mm (3.94 in)
Insert dimensions Ø 13 x 100 mm (0.51 x 3.94 in)
Insert material Brass
Voltage supply
Power supply AC 230 V, 50/60 Hz or AC 110 V, 50/60 Hz
Power consumption 250 VA
Power cord for Europe, 230 V
Communication
Interface RS-232
Case
Dimensions (W x H x D) 149 x 74 x 155 mm (5.9 x 2.9 x 6.1 in)
Weight 1.7 kg (3.8 lbs)
1) Is defined as the measuring deviation between the measured value and the reference value.
2) Maximum temperature difference at a stable temperature over 30 minutes.
3) Time before reaching a stable measuring value.

The measurement uncertainty is defined as the total measurement uncertainty (k = 2), which contains the following shares:
accuracy, measurement uncertainty of reference, stability and homogeneity.

CE conformity, approvals, certificates


EC declaration of conformity
EMC directive 2004/108/EC, EN 61326 emission (group 1, class B) and interference immunity (industrial
application)
Low voltage directive 2006/95/EC, EN 61010-1 and EN 61010-2-10, safety requirements for electrical equipment for
measurement, control and laboratory use
Approvals
EAC Import certificate, Eurasian Economic Community
GOST Metrology/measurement technology, Russia
Certificate
Calibration Standard: 3.1 calibration certificate per DIN EN 10204
Option: DKD/DAkkS calibration certificate
Recommended recalibration interval 1 year (dependent on conditions of use)

Approvals and certificates, see website

Page 2 of 4 WIKA data sheet CT 41.32 ∙ 08/2015


Dimensions in mm (inch)

Front view Side view (left)

151 (5.94)
142 (5.59) 160 (6.30)

Top view
76 (2.99)

Rear
1 Selection of the mains voltage
1 2 3 2 ON/OFF switch

3 Fuse

4 Fan

5 Mains connection socket

6 RS-232 interface

6 5 4

WIKA data sheet CT 41.32 ∙ 08/2015 Page 3 of 4


Design and controls

1 1 Carrying handle

2 Display

3 Controls

4 Sockets for temperature switch test

5 Insert with opening for test item

5 4

Scope of delivery Accessories


■■ Temperature dry-well calibrator model CTD9100-375 ■■ Inserts, undrilled and drilled to specification
■■ Power cord 1.5 m (5 ft) with safety plug ■■ Transport case
■■ Insert with 6.2 mm bore (0.24 in) ■■ Power cord for Switzerland
■■ Replacement tools ■■ Power cord for USA/Canada
■■ Operating instructions ■■ Power cord for UK
■■ 3.1 calibration certificate per DIN EN 10204

Options
■■ DKD/DAkkS calibration certificate

Ordering information
Model / Certificate type / Transport case / Power cord / Additional ordering information

© 2009 WIKA Alexander Wiegand SE & Co. KG, all rights reserved.
The specifications given in this document represent the state of engineering at the time of publishing.
We reserve the right to make modifications to the specifications and materials.

Page 4 of 4 WIKA data sheet CT 41.32 ∙ 08/2015


08/2015 EN

WIKA Alexander Wiegand SE & Co. KG


Alexander-Wiegand-Straße 30
63911 Klingenberg/Germany
Tel. +49 9372 132-0
Fax +49 9372 132-406
info@wika.de
www.wika.de