Professional Documents
Culture Documents
熱工實驗
Viscosity measurement via falling ball
method
Edlin Variste 方以良 4080H114
Table 1
Bigger
ball
1.68 7.99 0.562 0.977 0.83
Smaller
ball 1
7.65 7.65 0.067 0.741
smaller
ball 2
7.65 8.13 8.91 7.26 7.99 0.061 0.774
Table 2
Viscosity Error
Actual viscosity ave (Pa.sec) 0.95 0.00%
o
o A completely worked-out sample calculation is
required for repetitive calculations.
o Use MS Excel (or any other spreadsheet program) for
tabulation and plotting graphs.
Error
0.14%
0.12%
0.10%
0.08%
0.06%
0.04%
0.02%
0.00%
0.95 0.83
Actual viscosity ave (Pa.sec) Theoretical viscosity (Pa.sec)
6. Error analysis
We first explore the suitable range of viscosity for the ball drop technique,
using a 2 mm and 4.9 mm diameter steel ball, which can be conveniently
dropped in a thin glass capillary. Starting from the low end of viscosity by
filling a capillary with fluid, we soon discover that the bigger ball drops too
fast (<2 sec) to measure its speed reliably using a telephone stopwatch,
given variable human reaction time on the order of 1.68 sec. A simple
solution to slow down the ball drop speed is to let it fall in a tilt capillary. The
error in recording time can be reduced by orders of magnitude using a
camera with a computer interface, but the experimental condition under
which the ball falls faster than a few centimeters per second needs to be
prevented as it also violates the low Reynolds number criterion. The simplest
design of telephone recording of ball drop time proves to be adequate, since
other sources of error are often more significant than that caused by the time
recording method.
We proceed with experiments using a smaller ball. Table 1 lists typical results
from a set of measurements of the time traveling the smaller ball. That is for
retarding the ball travel.
7. Problems, suggestions and conclusions
The ball drop viscosity measurement described in this report is conveniently
low-tech, but practically useful. It requires no computer interface, nor any
electric instrument other than a telephone. In addition to plotting averages in
ball drop speeds from multiple repeats with standard errors, students were
asked to calculate the theoretical viscosity and plot it versus the actual
viscosity. Preliminary findings from those lab sessions prompted the author
to repeat the experiment in a more systematic manner. These additional
results confirm the reliability of the technique to yield viscosity values in the
error range of 0.14-1%, dependent on how meticulous the experimenter is,
how many standard samples are measured for calibration, and how many
repeated measurements are made on a sample of interest. If performed with
care, as elaborated in this paper, the technique is reliable for practical
applications
8. Appendices