Professional Documents
Culture Documents
ABSTRACT
A precise and rugged high performance size exclusion
chromatography (HPSEC) method for determination of
peak maximum molecular mass (q ), weight average
molecular mass (Y,), number average molecular mass
(ry,), and molecular mass distribution (MMD) in Low
Molecular Weight heparin (LMW heparin) has been deve-
loped and validated on two UltrahydrogelR 250 (ID 7,8
mm, length 30 cm) columns in series at three labora-
tories using different equipment. The calibration is
based on four local laboratory heparin standards with
relatively narrow molecular mass distribution and Mp
covering the range from 3,000 to 17,000 Da. The cali-
bration curve describing the logarithm of the mole-
cular mass versus retention time is linear in the
range 3,000 to 17,000 Da. The precision (relative
standard deviation) within-run and between-run is
better than 2%. The between-laboratory variation is
below 5%, 3% and 8% for , & and M,,, respectively.
This method is reproducib Y e, rugged, and fast and is
suited for the determination of average molecular
masses and molecular mass distribution of LMW heparins
on a routine basis.
131
132 MW CALIBRATION FOR LMW HEPARIN Vol. 64, No. 2
INTRODUCTION
TABLE 1
15 20 25 15 20
;- L
25
I 8.
Retention
time(min) Retention time (min)
Fig. la Fig. lb
Overlayed chromatograms. Chromatogram of LMW heparin
Standard Set 2. 20.14 min - q
25.90 min - Sucrose Retention Time
26.73 min - Salt Peak Retention Time
Vol. 64, No. 2 MW CALIBRATION FOR LMW HEPARIN 135
RESULTS
30r
26- /
20- /'
/*
lo-
16- /
6 _ /'
/"
OL* / 1 / I / 1
0 0.2 1.2 1.4
'.'AmounoP~mg/i~j:ction)'
log. Mp
-/
5.00 -
4.00 -
3.00 -
2,oo J , , , , , , , , ,
20 30 40
TABLE 2
The influence of the pump flow on%, bh, and % of WS3.
a) The numerical deviation in per cent from 0.7 ml/min.
b) The numerical deviation in per cent from the value
determined at 0.7 ml/min.
Mp 8 b’ M, % b’ % % b’
Da Da Da
TABLE 3
RT.SUC(min)
Standards Mean % RSD
TABLE 4
Within-run reproducibility. Each sample is repeated six times
and calculated against a cubic (C) and a linear (R) calibration
curve (Fig. 3). RSD is stated in the brackets.
TABLE 5
Between-run reproducibility. Each sample is examined once in
six independent days and calculated against a cubic (C) and a
linear (R) calibration curve (Fig. 3). The relative standard
deviation is stated in the brackets.
TABLE 6
Reproducibility between laboratories. Each standard is deter-
mined six times at each laboratory. The results are mean of
eighteen determinations. The relative standard deviation is
stated in the brackets.
Mp Y-4 Y Polydispersity
Da (Mw/Mn)
DISCUSSION
When this method is used for quality control analysis the use
of a marker is important since even small differences in pump
flow result in considerable variation in the results (Table 2).
ACKNOWLEDGEMENT
REFERENCES