Professional Documents
Culture Documents
NURSING
ASSIGNMENT
ON
SETTING, USE AND MAINTAINANCE
OF CENTRIFUGE MACHINE
DEFINITION
Centrifuge Machine:
Centrifuge machine is a device for separating particles from a solution according to their size,
shape, density, viscosity of the medium.
A device use to separate heavier particles from the lighter one by the action of centrifugal
force
Centrifugation:
Centrifugation is a process which involves the use of the centrifugal force for the
sedimentation of heterogeneous mixtures with a centrifuge.
Centrifugal force:
Centrifugal force is a force, arising from the body's inertia, which to act on a body moving in
a circular path and is directed away from the centre around which the body is moving.
2. Ultracentrifuges :-
It can spin in excess of 70,000 rpm, which makes it well suited for the
isolation of small particles, like DNA or viruses.
2. Tubes
a) Glass centrifuge tubes
Can be used with most solvents, but tend to be more expensive.
They can be cleaned like other laboratory glassware, and can be
sterilized by autoclaving.
TYPES
Centrifuges are generally divided into 3 categories based on their maximum attainable speed:
1. "Low-speed": max ~5 x 10 3 rpm.
2. "High-speed": max ~2 x 10 4 rpm.
3. "Ultracentrifuges": max ~10 5 rpm
APPLICATIONS
Separate solids from solution
Used during urinalysis to separate disease identifying components
Separate blood into plasma and cellular components
PROCEDURE OF USE
1. Identify the speed and duration at which you wish to centrifuge samples prior to using
the instrument
2. Place the instrument on a firm surface
3. Don gloves
4. Check the rotor you intend to use and be certain that the rotor is rated for the speed at
which you like to use it.
5. Enter the centrifugation speed
6. If the rotor is not capable of being operated at the target speed, you will need to
identify the rotor that is capable of being used at the desired speed, and then transfer
your samples to a centrifuge tube that will fit and rebalance the samples, remembering
to include the lids when balancing
7. Place the rotor in the centrifuge with the 2 pins on the underside of the rotor forming a
cross with the 2 pins found on the spindle of the centrifuge. Check the name of the
rotor and confirm the target speed. Locate the correct lid for the selected rotor and
place it beside the centrifuge.
8. Place the samples into the rotor. If the samples do not all have the same mass, place
samples with the same mass on opposite sides of the rotor. Load the tubes opposite
each other in the centrifuge
9. Once the samples are loaded, check the rotor type again, confirm that you have the
correct lid for the selected rotor and then screw the rotor lid onto the spindle.
10. The rotor lid set screw tums opposite to "normal" screws. Check for proper set screw
direction by looking at the diagram on the lid.
11. If this is a swinging bucket rotor, there is no lid.
12. Once the rotor lid is secured, close the centrifuge lid and set the desired temperature,
the desired speed and the desired spin time
13. When everything is set, press the "Start" button and wait for the instrument to ramp
up to the desired speed.
14. It is normal to see a small vibration in the instrument as the speed moves from 100 to
2500 rpm.
15. Any large vibration in the instrument or irregular noise may indicate that something
has gone wrong with the run. Press the "Stop" button and move away from the
instrument.
16. Immediately contact a faculty member who can assist you BEFORE you open the lid
to the centrifuge.
17. If no problems are detected as the instrument reaches the desired speed, you may
leave the area until the run is complete.
18. When the run has completed and the rotor has come to a complete stop, you may
unscrew the rotor lid and carefully remove your samples.
19. Should a sample vial have leaked during the run, you MUST IMMEDIATELY clean
the spill and dry the rotor.
20. It is easy to miss spilled liquids in the bottom of the positions of the rotor; however,
this liquid will cause an imbalance in the subsequent run that may lead to catastrophic
failure.
21. Remove the rotor from the centrifuge and place in the appropriate box adjacent to the
unit.
22. If the rotor is a fixed angle rotor, put the lid on top of the rotor in the box.
23. Should a sample vial have leaked during the run, you MUST IMMEDIATELY clean
the spill and dry the rotor.
24. It is easy to miss spilled liquids in the bottom of the positions of the rotor, however,
this liquid will cause an imbalance in the subsequent run that may lead to catastrophic
failure.
25. Clean the area around the centrifuge with 70% ethanol (denatured) and wipe dry.
26. This MUST be done following EVERY run. Failure to clean up after you may ruin
subsequent users' experiments and/or facilitate the spread of contamination
throughout the department.
CENTRIFUGE SAFETY
Always wear gloves and PPE when type of use necessitates
Before using, read the owner's manual.
Do not operate a centrifuge until having been shown proper use by an experienced
operator.
Check that the centrifuge chamber and drive spindle of the rotor is free of scratches or
burrs.
Decontaminate centrifuge of biological hazards before servicing.
Never repair a cracked or bent rotor.
If rotor has been replaced, tighten knob/screw fastening rotor to the shaft to prevent
hazardous loosening.
Prolonged contact with some disinfectants can cause damage to rotor and other
components (e.g. 10% sodium hypochlorite).
Follow your facility’s safety procedures when cleaning and disinfecting the centrifuge
Never attempt to touch or stop a spinning rotor by hand or with another tool or object.
Always place the centrifuge on a flat surface
Always unplug the power cord before cleaning
Before moving the centrifuge to a new location, the exterior and interior surfaces
should be cleaned and disinfected
Plug in centrifuge only when completely dry
Ensure that centrifuge bowls and tubes are dry.
Ensure that the spindle is clean
Ensure that the rotor is properly seated on the drive shaft.
Make sure that tubes or containers are properly balanced in the rotor.
Do not exceed the rotor’s maximum run speed.
Close the centrifuge lid during operation.
Make sure that the centrifuge is operating normally before leaving the area.
Make sure that the rotor has come to a complete stop before opening the lid.
MAINTENANCE OF CENTRIFUGE
1. Daily maintenance
Wipe the inside of the bowl with disinfectant solution and rinse thoroughly.
The centrifuge must not be used if the interior is hot, if unusual vibrations or
noises occur, or if deterioration (corrosion of parts) is detected.
A qualified service technician should be contacted.
Most vibrations are due to improper balancing and can be corrected by
rebalancing the buckets and tubes.
2. Monthly maintenance
Clean the centrifuge housing, rotor chamber, rotors and rotor accessories with
a neutral cleaning agent.
Clean plastic and non-metal parts with a fresh solution of 0.5% sodium
hypochlorite.
How to clean centrifuge
Centrifuge interiors are made from a range of materials-plastic, ceramic, aluminium and
stainless steel-that manufacturers use in the search for performance in a high-stress
environment. Manufacturers' recommendations on the cleaners and scrubbers to use and
method and frequency of application take this into account and will get the best results.