Chapter 6

Replacements
Contents

Replacement matrix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6:2


Actions to be done - description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6:4
AK 96 Fluid unit - flow path . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6:6

HCEN12211 Revision.05.2008 Replacements 6:1


Replacement matrix
The actions to carry out after a component replacement are brought
together in a replacement matrix.

The numbers in the matrix show which actions and in which order they
will be done. It is important that the actions are done and in the correct
order.

In the matrix, BM stands for Blood Monitor and FM stands for Fluid
Monitor.

The replacement matrix is only a guideline for service technicians which


actions to perform when a component has been replaced.

Note
For technical error handling, (information regarding error codes) refer
to AK 96 Error Code List - HCEN12371.
Note

6:2 Replacements HCEN12211 Revision.05.2008


HCEN12211 Revision.05.2008
Replaced FM components Replaced BM components Electronics

BPM
INPS
SAGS

Heater
pH cell
UF-cell B

Relay board

FM I/O board
BM I/O board

Suction pump
Syringe pump

Flow restrictor
PD Transducer

Feeding pump A
Feeding pump B
Priming detector

Operator's panel

Degassing pump
HPG Transducer
Blood pump rotor
Control CPU board

Conductivity Cell A
Conductivity Cell B
Conductivity Cell P

Blood leak detector


Control driver board

Clamp unit complete


AK 96

Protective CPU board

Blood pump motor unit


Protective driver board

Pressure regulator, PR1


Pressure regulator, PR2
Variable flow unit (or parts)
Level detector (Air detector)
Pressure Transducer (V & A)
Replacement matrix

Degassing Pressure Transducer


Clean the machine from calcium
1 1 1
and then from proteins and fats
1 2 2 2 1 1 1 1 I2C scan
1 1 Total default
2 2 Load preset file
A
3 Calibrate Conductivity Cell AA
3A Calibrate Conductivity Cell BA
3A Calibrate Conductivity Cell PA

Replacements
1 Calibrate pH
2A 2A 2A Calibrate Pressure TransducerA
2 Calibrate Bloodleak
2 1 Calibrate pressure regulator, PR2
3 2 4 4 4 3 1 1 1 1 1 1 3 1 3 2 3 2 1 1 1 1 1 2 1 2 1 1 1 3 2 7 5 1 2 Fch
4 6 6 6 4 2 2 2 3 2 2 4 4 3 3 2 2 2 3 Heat Disinfection
5 5 5 Check conductivity value
Actions to be done

2 2 Simulated treatment
1 Calibrate Pressure Regulator, PR1
1 Check BPM
1 1 4 Calibrate Priming detector
2A Calibrate Pressure transducerA
1 Occlusion
3 3 Calibrate A/D conv.
4 1 Calibrate clock
2 2 6 Check fluid flow
1 5 Calibrate variable flow

Remarks:
A
Not necassary to calibrate if new spare part is installed (factory calibrated)
B
Important to follow the order of the calibrations/actions for machine performance

6:3
Actions to be done - description
Following is a list of the suggested actions including explanations how to
execute them or references to other instructions.

Clean the machine from Perform actions according to Operators manual.


calcium and then from
proteins and fats
Perform actions according to AK 96 Service Manual
I2C scan / Maintenance support description /
Diagnostics mode - System.

Perform actions according to instructions in


Total default AK 96 Service Manual / Maintenance support
description.

Load preset file Perform actions according to clinic demands.

Perform actions according to calibration instructions


Calibrate Conductivity Cell A
in the AK 96 Service Manual / Calibrations.

Calibrate according to calibration instructions in the


Calibrate Conductivity Cell B
AK 96 Service Manual / Calibrations.

Calibrate according to calibration instructions in the


Calibrate Conductivity Cell P
AK 96 Service Manual / Calibrations.

Calibrate according to calibration instructions in the


Calibrate pH
AK 96 Service Manual / Calibrations.

Calibrate Pressure Calibrate according to calibration instructions in the


Transducer AK 96 Service Manual / Calibrations.

Calibrate according to calibration instructions in the


Calibrate Blood leak
AK 96 Service Manual / Calibrations.

Calibrate Pressure Regulator, Calibrate according to calibration instructions in the


PR2 AK 96 Service Manual / Calibrations.

Start the machine and let it perform a Function


Fch
Check.

Perform a heating according to the AK 96 Operators


Heat disinfection
Manual.

Check conductivity value according to hospital


Check conductivity value
laboratory.

Simulate a treatment according to check/calibration


Simulated Treatment instructions in the AK 96 Service Manual /
Maintenance Manual.

Calibrate according to calibration instructions in the


Calibrate Pressure Regulator,
AK 96 Service Manual, / Calibrations and Technical
PR1
Data and Specifications.

Check according to BPM test instructions in the


Check BPM
AK 96 Service Manual / Calibrations.

Calibrate according to calibration instructions in the


Calibrate Priming Detector
AK 96 Service Manual / Calibrations.

Calibrate according to calibration instructions in the


Occlusion
AK 96 Service Manual, / Calibrations.

6:4 Replacements HCEN12211 Revision.05.2008


Calibrate Pressure Calibrate according to calibration instructions in the
transducer AK 96 Service Manual, / Calibrations.

Calibrate according to calibration instructions in the


Calibrate A/D conv.
AK 96 Service Manual / Calibrations.

Calibrate according to calibration instructions in the


Calibrate clock
AK 96 Service Manual / Calibrations.

Start the machine and let it go to green path. Press


Check fluid flow
twice on the select button and check the fluid flow.

Calibrate according to calibration instructions in the


Calibrate variable flow
AK 96 Service Manual / Calibrations.

HCEN12211 Revision.05.2008 Replacements 6:5


6:6
CHVA

C P BiCart
CBVA holder*

BiCart
Mixing chamber Mixing chamber
B
A pump

RIVA
Cond. T FLVA Cond. T
A B pump cell A cell B

Flow Expansion Degassing


switch chamber chamber
Heater DRVA* P Flow
pH Cell
(upgrade kit) M pump Cond.
Cell P
AIVA Degassing T
P
PR 1 INPS PR 2 pressure
Pure water INVA Flow restrictor
T
inlet
Blood leak P HPG
Outlet restrictor detector ZEVA
Drain
AK 96 Fluid unit - flow path

T
EVVA Cond.
Draining
cell C*
restrictor

Replacements
Heat
exchangers REVA Suction UF-measuring
pump Deairating 2 1 cell
PD P
chamber

Dialyzer

TAVA BYVA

SAGS FIVA* DIVA

Pure water
A concentrate Fluid connector
B concentrate detection
Dialysis fluid Dialysis fluid
Dialysate
filter (U9000S)*
*=optional

HCEN12211 Revision.05.2008