Figure 3.1 Optical Bench
WBC Measurement
Overview
The Optical Channel is used for the determination of WBC data. During sample
aspiration, 20 µL of sample is segmented in the Shear Valve for WBC
measurement. The WBC Syringe dispenses 0.973 mL of WBC lyse to the Shear
Valve. The sample and lyse are then transferred to the WBC Mixing Chamber/
WOC Heater where the dilution is mixed, resulting in a 1:50 dilution ratio.
The Sample Transfer Pump transfers the WBC dilution from the mixing chamber
to the sample feed nozzle in the Optical Flow Cell. At the same time, sheath
reagent, under constant pressure in the Sheath Reservoir, is transferred to the sheath
feed nozzle in the Optical Flow Cell and injected into the cell. At the same time,
the Sample Metering Syringe injects 46.5 µL of the WBC dilution into a sheath
stream. The sample stream is then hydrodynamically focused to align the cells in
single file as they pass through the Optical Flow Cell, which is an optically clear
quartz chamber. A vertically polarized Helium Neon Laser is the light source.
The instrument measures:
• Both types of forward angle light scatter (1° to 3°, referred to as 0°, and 7° to
11°, referred to as 10° or narrow angle)
• Both types of orthogonal (side) light scatter (70° to 110°, referred to as 90°,
and 70° to 110° depolarized, referred to as 90°D).
This is referred to as MAPSS (for Multi-Angle Polarized Scatter Separation)
technology. Various combinations of these four measurements are used to classify
the WBC subpopulations and provide morphological flagging.
3 5
The previous figure illustrates the measurement of light scattered during the WBC
optical measurement process.
The WBC count is determined by enumerating the number of occurrences above a
hardware threshold in the 0° channel. The information from all four measurements
is used to differentiate the WBC into five subpopulations:
Neutrophils
Lymphocytes
Monocytes
Eosinophils
Basophils
The WBC data is presented graphically as scatterplots or histograms.
WBC Reagent
The WBC reagent used with the CELL-DYN Ruby instrument is the CELL-DYN
WBC Lyse. It is an integral part of the WBC analysis. White blood cells diluted in
the reagent maintain cellular integrity close to their native state. The structure of
the basophils changes slightly due to the hygroscopic nature of the basophilic
granules.
The RBC are also altered by the reagent. The osmotic pressure of the RBC is higher
than that of the reagent. Therefore, the hemoglobin in the RBC diffuses out of the
cell and water from the reagent diffuses into the cell. The cell membrane remains
intact but the RBC now has the same refractive index as the sheath, thereby
rendering it invisible to the laser.
WBC Differential
The light scatter information is graphically presented in the form of scatterplots.
(The data can also be presented in histograms.) Each cell analyzed is represented
by a dot on the scatterplot. The dots are plotted at a point determined by the
intersection of the channel information designated on the X and Y axes. For
example, if a cell falls in channel 50 on the X axis and channel 50 on the Y axis, it
is plotted at the intersecting point of the two channels.
The scatter information may be plotted in various combinations to yield different
information. The CELL-DYN Ruby uses the scatterplots to differentiate the WBC
into five subpopulations:
Neutrophils
Eosinophils
Lymphocytes
Basophils
Monocytes
90° Lobularity
Mononuclear-Polymorphonuclear Separation
The scatter information is plotted with the 90° scatter on the Y axis and the 10°
scatter on the X axis. (The 90°/10° scatterplot is shown in the previous figure.) Two
populations of cells are clearly seen on the display. The mononuclear cells fall in
the cluster in the lower left corner of the scatterplot and the polymorphonuclear
cells fall in the cluster above and to the right of them.
The instrument uses a dynamic threshold to determine the best separation between
the two populations. Each cell is then identified as a MONO or a POLY. Once each
cell is identified, it retains this classification no matter where it appears on other
scatterplots.
Neutrophil-Eosinophil Separation
The scatter information is plotted with the 90°D scatter on the Y axis and the 90°
scatter on the X axis. (The 90°D/90° scatterplot is shown in the previous figure.)
Only the polymorphonuclear cells are plotted on this scatterplot. The mononuclear
cells have been identified and therefore do not interfere in the further classification
of the polymorphonuclear cells.
Two populations of polymorphonuclear cells are clearly seen on the display. The
neutrophils fall in the lower of the two clusters. The eosinophils fall in the upper
cluster. The instrument uses a dynamic threshold to determine the best separation
between the two populations. Each cell is then classified as a NEUT or an EOS.
All cells scatter a certain amount of 90°D light. The eosinophils scatter more 90°D
light than any of the other cells because of the unique nature of granules they
contain. This property of the eosinophils is used to positively identify them and
thus clearly differentiate them from the neutrophil population.
Mononuclear Mononuclear
Separation Identification
0° Size
0° Size
Mononuclear Separation
The scatter information is plotted with the 0° scatter on the Y axis and the 10°
scatter on the X axis. (The 0°/10° scatterplot is shown in the previous figure.) The
mononuclear cells are plotted on this scatterplot. The algorithm also uses the
orientation of the neutrophil cluster to aid in classifying the mononuclears. Three
populations of mononuclear cells are clearly seen on the display.
There are three populations of mononuclears because basophils are included in the
mononuclear cluster. Typically, basophils are granulated cells and therefore more
complex than the mononuclear cells. However, the basophilic granules are water
soluble and dissolve in the WBC Lyse reagent. Consequently, the degranulated
basophils becomes a less complex cell that falls into the mononuclear cluster.
The lymphocytes fall in the lowest large cluster. (The small population of cells
below the lymphocytes contains particles that are unlikely to be WBC.) The
basophils fall in the cluster above and slightly to the right of the lymphocytes. The
monocytes fall in the cluster above the lymphocytes and basophils. The instrument
uses dynamic thresholds to determine the best separation between the three main
populations. Each cell is then classified as a LYMPH, a MONO or a BASO.
Finally, the instrument evaluates the area below the lymphocyte cluster but above
the hardware threshold (channel 23). Any particles that fall in this area are
separated from the lymphocytes by a dynamic threshold. The following cell types
may be present in this region:
NRBC
Unlysed RBC
Giant PLT
PLT clumps
All particles in this region are excluded from the WBC count and the Differential.
Other Scatterplots
90°/0°
The scatter information is plotted with the 90° scatter on the Y axis and the 0°
scatter on the X axis.
90°D/0°
The scatter information is plotted with the 90°D scatter on the Y axis and the 0°
scatter on the X axis.
90°D/10°
The scatter information is plotted with the 90°D scatter on the Y axis and the 10°
scatter on the X axis.
All scatterplots may be displayed and printed at operator request.
Resistant RBC
When a specimen containing resistant RBC is run in the CBC test selection, the
lytic agent in the WBC lyse reagent may be insufficient to lyse the “resistant” cells
in the time allotted for the WBC count. Consequently, unlysed RBC can be
erroneously included in the WBC count resulting in a falsely elevated value. When
this occurs, a significant amount of cellular debris will be present in the region
below the WBC dynamic threshold on the 0º/10º scatterplot.
When these types of specimens are rerun in the CBC+RRBC test selection, the
diluted WBC sample is held in the mixing chamber 15 seconds longer than in the
routine patient mode. This additional lysing time is used to break down (lyse) the
resistant RBC cells and prevent them from interfering with the WBC count and
differential.
NOTE: A higher incidence of false positive band flags may be evident on
specimens run under the Resistant RBC test selection.
WBC Histograms
The CELL-DYN Ruby can present the WBC scatter information as two
histograms: NWBC-LYM-MONO (N-L-M) and Mono-Poly (M-P). The NOC
(Nuclear Optical Count) data can also be presented as a histogram. (Refer to the
previous figure.) These histograms may be displayed and printed at the operator’s
request.
NWBC-LYM-MONO Histogram
The scatter information is plotted in a histogram format with the relative number
of cells on the Y axis and the NWBC, Lymphocyte and Monocyte size distribution
data on the X axis.
MONO-POLY Histogram
The scatter information is plotted in a histogram format with the relative number
of cells on the Y axis and the mononuclear and polymorphonuclear size
distribution data on the X axis.
NOC Histogram
The NOC data is plotted in a histogram format with the relative number of nuclei
on the Y axis and the size distribution data on the X axis.
WBC Parameters
The WBC data is generally displayed as depicted in Figure 3.8. All numeric and
graphic data are automatically displayed in the Run View Chartable, Lab, and
Graphics tabs in the format selected in Customizing Run View. See
Section 2: Installation Procedures and Special Requirements,
Subsection: Customize Run View…. After the WBC scatter information has been
plotted and the cells have been classified into the five subpopulations, the
algorithms then determine the WBC and the percent of cells in each subpopulation.
Once the WBC count is determined, the absolute number of cells in each
subpopulation is calculated by multiplying that WBC count by the percentage. The
results are expressed as follows:
WBC # x 10e3/µL
NEU # x 10e3/µL and %
LYM # x 10e3/µL and %
MONO # x 10e3/µL and %
EOS # x 10e3/µL and %
BASO # x 10e3/µL and %
The decimal point moves to display up to three decimal places for the absolute
number and percent.
WBC Flagging
Refer to the “Operational Messages and Data Flagging” subsection of this section
for WBC flagging information.
RBC/PLT Measurement
Overview
The Optical Channel is used for the determination of RBC and PLT data. During
sample aspiration, 1.67 µL of sample is segmented in the Shear Valve for RBC/PLT
measurement.
The Diluent/Sheath Syringe dispenses 2.79 mL of diluent to the Shear Valve. The
sample and diluent are then transferred to the RBC/PLT Mixing Chamber where
the dilution is mixed, resulting in a 1:1675 dilution ratio.
The Sample Transfer Pump transfers the RBC/PLT dilution from the mixing
chamber to the sheath feed nozzle in the Optical Flow Cell. The Sample Metering
Syringe injects 24 µL of RBC/PLT dilution into the sheath stream. The sample
stream is then hydrodynamically focused to align the cells in single file as they pass
through the Optical Flow Cell, which is an optically clear quartz chamber. A
vertically polarized Helium Neon Laser is the light source.
There are 256 size channels for each of the parameters, each RBC size channel
being equivalent to 1 fL and each PLT size channel being equivalent to 0.137 fL.
The RBC parameters are calculated using 0°, 10°, and 90° sensor data, while the
PLT parameters are calculated using 0° and 10° sensor data.
RBC Parameters
All numeric and frequency size distribution data are automatically displayed on the
Run View in the format selected. The size distribution data for the red cells is
displayed graphically as a histogram using 0° data. The size distribution data is
plotted on the X axis. The relative number of cells is normalized and plotted on the
Y axis. The RBC data are shown in the previous figure.
RBC Count
The Red Blood Cell Count is directly measured, and is expressed as follows:
RBC = # x 10e6/µL
Counts below 1.0 x 10e6/µL are displayed to three decimal places. The RBC count
is corrected for coincidence and WBC interference.
MCV
The Mean Cell Volume is the average volume of the individual red blood cells.
The MCV is derived from the RBC size distribution data on the 0°, 10°, and 90°
histograms, and is expressed in femtoliters.
HCT
The Hematocrit is the ratio of red blood cells to plasma and is expressed as a
percentage of the whole blood volume. The HCT is calculated from the red blood
cell count and the mean cell volume as follows:
HCT = (RBC x MCV)/10
MCH
The Mean Corpuscular Hemoglobin is the average amount of hemoglobin
contained in the red blood cell expressed in picograms. The MCH is calculated
from the RBC and the HGB as follows:
MCH = (HGB/RBC) x 10
MCHC
The Mean Corpuscular Hemoglobin Concentration is the ratio of the weight of
hemoglobin to the volume of the average red blood cell expressed in grams per
deciliter. MCHC is calculated from the HGB and the HCT as follows:
MCHC = (HGB/HCT) x 100
RDW
Red Cell Distribution Width is a measure of the heterogeneity of the RBC
population. The CELL-DYN Ruby reports a relative RDW equivalent to a CV in
grams per deciliter. The RDW is derived from the RBC histogram using the 20th
and 80th percentiles.
RBC Flagging
Refer to Subsection: Operational Messages and Data Flagging for RBC Flagging
information.
Platelet Parameters
Events counted in the RBC/PLT dilution between floating thresholds are included
in the platelet (PLT) data, which is collected using the 0° and 10° sensors. The
lower threshold floats between 1 and 3 fL and the upper threshold floats between
15 and 35 fL. If there are not enough data to determine the PLT count, the lower
and upper thresholds are set at 2 and 35 fL respectively. Once the thresholds have
been determined, the PLT count is derived from the 10° data.
Data can be displayed in two formats. Data can be displayed as a scatterplot (0°/
10°) including the RBC. Data can also be displayed as one of the following three
histograms:
PLT only using 10° data
PLT and RBC using 0° data
PLT and RBC using 10° data
PLT data are shown as a histogram of the 10° data in the following figure.
Events counted in the region below the lower threshold are usually either optical
noise or small particulate matter. Events counted in the region above the upper
threshold are counted as RBC. If interference with either threshold region exceeds
a predetermined limit, the PLT parameters are flagged accordingly. The flags are
discussed in the last section of this section.
PLT Count
The PLT count is expressed as thousands per microliter (10e3/µL).
MPV
The Mean Platelet Volume is derived from the PLT histogram after the PLT count
has been determined. The MPV is expressed in femtoliters.
PCT
The Plateletcrit is the product of PLT and MPV and is analogous to the hematocrit.
It is expressed in percent and is calculated as follows:
PCT = (PLT x MPV)/10
PDW
Platelet Distribution Width is a measure of the heterogeneity of the PLT
population. It is expressed as the geometric standard deviation.
NOTE: Clinical significance has not been established for PCT and PDW.
Therefore, they are not reportable in the US.
Platelet Flagging
Refer to Subsection: Operational Messages and Data Flagging for PLT flagging
information.
Hemoglobin Measurement
Overview
The HGB channel is used for the colorimetric determination of hemoglobin.
During sample aspiration, 12 µL of sample is segmented in the Shear Valve for
HGB measurement.
The Diluent/Sheath Syringe dispenses 1.7 mL of Diluent/Sheath to the Shear
Valve, transferring the HGB segment to the HGB Mixing Chamber. The HGB Lyse
Syringe then dispenses 0.9 mL of HGB Lyse into the mixing chamber. The mixture
is mixed, resulting in a 1:218 dilution ratio. The HGB lyse reagent lyses the red
blood cells, converting the hemoglobin that is released by a cyanide-free chemical
process. When the lysing action is completed, a low-energy LED in the HGB Flow
Cell, attached to the mixing chamber, measures the amount of absorbance which is
proportional to the HGB concentration. Five separate HGB readings are made on
the sample. The lowest and highest are eliminated and the remaining three are
averaged to give the final HGB sample reading. After the hemoglobin readings
have been made, the HGB Flow Cell is rinsed with diluent/sheath.
A reference value is then obtained using the diluent/sheath in the HGB Flow Cell.
A zero or blank reading is obtained on the diluent to provide a reference to which
the sample signal is compared. Five separate blank readings are made on the
diluent. The lowest and highest are eliminated and the remaining three are
averaged to give the final HGB reference reading.
A LED with a wavelength of 555 nm is the light source. A photodetector measures
the light that is transmitted.
The sample and reference readings are compared to determine the HGB
concentration of the sample. The HGB result is expressed in grams of hemoglobin
per deciliter of whole blood. Up to two decimal places may be displayed for
hemoglobin results less than 10.0 g/dL.
HGB Parameters
The Hemoglobin is directly measured and is expressed in grams of hemoglobin per
deciliter of whole blood.
HGB Flagging
Refer to Subsection: Operational Messages and Data Flagging for HGB flagging
information.
Lab Page
The Run View Lab Page is provided to assist the laboratory staff in data review and
validation (refer to the following figure). This screen is for laboratory use only. The
lab page displays the 5-Part Differential plus additional parameters. The Run View
Chartable Page displays only the 5-Part Differential (refer to the figure in the WBC
Scatterplots subsection). The difference between the two formats is shown in the
following tables.
NOTE: The parameters MON and LYM have an “e” after the label, indicating that
the values are estimated. MONe represents monos minus blasts. LYMe
represents reported lymphs minus variant lymphs.
All numeric and graphic data are automatically displayed in the Run View Lab tab
in the format selection in Customize Run View. See Section 2: Installation
Procedures and Special Requirements, Subsection: Customize Run View….
The 5-Part Differential separates WBC into 5 components: Neutrophils,
Lymphocytes, Monocytes, Eosinophils, and Basophils. The additional parameters
further separate the Neutrophils, Lymphocytes, and Monocytes into their
constituent components. Eosinophils and Basophils are the same in both tables.
WBC 7.23
1 NEU 4.65
2 LYM 1.67
3 MONO .639
4 EOS .228
5 BASO .045
WBC 7.23
NEU
1a SEG 4.40
1b BAND .208
1c IG .038
MONO
3a BLST .001
3b MONe .638
4 EOS .228
5 BASO .045
LYM
2a LYMe 1.64
2b VARL .030