Professional Documents
Culture Documents
6
Page : 1 of 8
BREWING MANUAL 1st Issue : SEP 1996
Vol. X - Analytical Methods Rev. Date : OCT 1999
1. INTRODUCTION
a) Many analyses require that the sample's concentration is reduced (diluted) to a known
quantifiable level. Samples are furthermore treated and or reacted with chemical
preparation (reagents) of which the exact concentration must be known. In order to
achieve this with a high degree of precision/accuracy, volumetric equipment (flasks) are
used, as is discussed in this section.
b) In order to ensure an absolute sound volumetric preparation, it is equally important that
all the other "Techniques and Practices" in this manual (Vol. X - Section 1.0) are sound
and in place, for one technique is frequently used together with one or more of the others.
c) Temperature has a major impact on volume and therefore concentration of components in
a prepared solution.
Note: All preparations are made at 20 ± 0,2°C in our laboratories.
d) Volumetric flasks are used for:
i) Quantitative preparation of reagents, or
ii) Quantitative dilution of reagents and or samples.
e) These flasks are designed to either:
i) contain a specified volume, or
ii) deliver a specified volume.
Note: i) Delivery flasks (ID: ‘D’ or ‘TD’) are not used in our laboratories.
ii) Flasks intended to contain (ID: ‘C’, ‘TC’ or ‘IN’) are used for
preparations required in our laboratories.
f) Volumetric flasks are manufactured from high quality glass and graduated to contain a
specific volume under specified conditions.
Note: i) For all work in SAB laboratories, grade A glass volumetric flasks are
specified.
ii) DO NOT use grade B or plastic for quantitative preparations.
g) Polyethylene stoppers are used with these flasks.
h) References
i) SAB, Analytical MDT, 1995.
ii) Vogel’s, Textbook of Quantitative Inorganic Analysis, 4th Edition, 1978.
The South African Breweries Section : 1.6
Page : 2 of 8
BREWING MANUAL 1st Issue : SEP 1996
Vol. X - Analytical Methods Rev. Date : OCT 1999
2.3 PRECAUTIONS
Wear protective gloves when clearing away broken glass.
3. APPARATUS
3.1 EQUIPMENT REQUIRED
a) Volumetric flask with stopper.
Flasks are available in the following general sizes 5, 10, 20, 25 50, 100, 250, 500, 1
000 and 2 000 ml. Odd sizes such as 1, 2, 3, 15, 30, 3 000 and 5 000 ml are
available.
Note: Only use grade A flasks graduated at 20°C (Temperature of graduation
is always shown on the flask).
b) Water bath set at 200,2°C.
- Endothermic reaction absorbs heat, the preparation cools down and can
become very cold.
3.4 CALIBRATION
3.5 STORAGE
Store in a cupboard when not in use.
4. CHEMICALS
The South African Breweries Section : 1.6
Page : 4 of 8
BREWING MANUAL 1st Issue : SEP 1996
Vol. X - Analytical Methods Rev. Date : OCT 1999
5. METHOD
5.1 PROCEDURE
a) Pre-attemperate sufficient purified water to 20 0,2°C.
b) Note specific requirements of the relevant method.
c) Quantitatively transfer the sample* to the flask.
*Can be :
- A fixed volume of sample (product being analysed).
- Fixed volume of a reagent or liquid chemical.
- Fixed mass of solid chemical or sample.
d) Fill the volumetric flask to about 85% of its capacity with pre-attemperated, purified
water and insert the stopper.
e) Mix to dissolve or to blend together.
f) Place in water bath and attemperate to 20 ± 0,2°C.
g) Remove the stopper and wash down into the flask.
Note: i) Avoid losing droplets.
ii) Place the stopper upside down on a clean surface.
h) Fill the flask to a level about 5-10 mm below the graduation mark with attemperated
purified water. Stopper, mix and stand in water-bath.
Note: i) Re-attemperate to 20 °± 0,2°C.
ii) Allow liquid to drain to below the graduation mark.
i) Using a pasteur-pipette add attemperated purified water and make-up to volume.
Note: Avoid parallax errors.
j) Re-stopper and mix well.
6. VERIFICATION
Fill a cleaned volumetric flask with purified water and drain. Check for absolute cleanliness,
that is, even drainage leaving no dry spots or water droplets on the inner wall of the flask, if
present, it is indicative of a dirty unclean flask.
The South African Breweries Section : 1.6
Page : 6 of 8
BREWING MANUAL 1st Issue : SEP 1996
Vol. X - Analytical Methods Rev. Date : OCT 1999
7. COMPETENCY TEST
Assessment 1 2 3
1.0 PRODUCT
I’d know that the end product is right and as
manualised if:-
1. Dedicated flasks are clearly identified and
traceable.
2. Only grade ‘A’ volumetric glassware is used in
the laboratory.
3. A 'Change Management' record is in place and
current.
2.0 PROCESS
If the candidate did the job the correct way as
manualised he/she would:-
1. Adhere and comply to Health and Safety i.r.o.:
1.1 Cracked and chipped glassware are
hazardous.
1.2 Chemicals are hazardous.
3. Clean the volumetric flask:
3.1 With detergent.
3.2 With a 10% acid solution.
3.3 With a solvent.
3.4 Rinse with purified water (3x).
3.5 Inspect for damages.
4. Verify the cleanliness of the flask by observing
the drainage of purified water from the flask.
5. Attemperate purified water to 20 ± 0,2°C to use
in the preparation.
The South African Breweries Section : 1.6
Page : 7 of 8
BREWING MANUAL 1st Issue : SEP 1996
Vol. X - Analytical Methods Rev. Date : OCT 1999
3.0 KNOWLEDGE
A candidate who does this job should be able to:
1. Describe and discuss the differences between
grade 'A and 'B' volumetric flasks.
2. Discuss the effect of temperature on volumetric
preparations.
3. Discuss the term 'Error of parallax'.
4. Discuss the term 'Quantitative'.
5. Explain the terms 'Exothermic' and 'Endothermic'.
6. Name five 'Golden Rules' you will observe in
order to protect and maintain the precision of a
volumetric flask.
7. Name two other volumetric apparatus used in a
laboratory
Assessor 1 Candidate
Sign. Sign.
Assessor 2
Sign.