You are on page 1of 2

ADVIA 560 Reagent Measurement Flow Diagram

START BLANK START Meas.

Needle to SR Sample in

Needle piercing

Cleaning processes Sampling Preparing for


making dilutions
Generating Needle washing
Vacuums (outside) HGB blank

SV to CP

Generating Dilutions in parallel Needle Pre-washing


Vacuums WBC, MIX, 4diff (Inside)

SV to NP

Making RBC dilution

SV = Shear Valve
SV to CP SR = Sample Rotor
CP = Chamber Position
NP = Needle Position
TCU = Temperature Control Unit

DCHT-A06.844.02.01.02
Cleaning Needle Making 4diff RBC
(Inside) dilution + measurement
refilling sheath
puffer

Generating
vacuum
SV to NP HGB
measurement
+ reset

4diff Cleaning WBC


measurement measurement
MIX, RBC chambers,
TCU loops 1st time

Taking BASO
Regenerate sample
vacuum

SV to CP

BASO Cleaning
measurement
WBC, RBC chambers
TCU loops 2nd time

RESULT SV to NP

START NEXT
DCHT-A06.844.02.01.02