Professional Documents
Culture Documents
Noise Characteristics
Application Note
Introduction
Random noise, sometimes referred of each scope of equal bandwidth the level of noise at 10-mV/div. For the
to as white noise, is theoretically should be compared at each scope’s Agilent MSO7054B, this relative RMS
unbounded and exhibits a Gaussian most sensitive V/div setting that noise component is approximately
distribution. Unbounded simply means provides full bandwidth. 2% of the V/div setting. The relative
that because of the random nature component of RMS noise on the
of noise, the more data you collect in Many oscilloscope evaluators make Tektronix 500-MHz and LeCroy 600-
noise characterization measurements, the mistake of only testing the base- MHz bandwidth scopes appears to be
the higher the peak-to-peak excursions line noise floor characteristic at the in the range of 3% to 4%.
will grow. For this reason, random scope’s most sensitive V/div setting,
phenomenon such as vertical noise and then assuming this amplitude After determining the fixed component
and random jitter should be measured of noise applies to all V/div settings. of noise (approximate base-line noise
and specified as an RMS (one standard There are actually two components floor) and the relative component of
deviation) value. Table 1 and Table 2 of noise inherent in oscilloscopes. noise, you can estimate the amount of
shows RMS noise floor measurements One component of noise is a fixed noise at intermediate V/div settings
of six competitively priced level of noise contributed primarily by by using a square-root-of-the-sum-
oscilloscopes with bandwidths ranging the scope’s front-end attenuator and of-the-squares formula. And as you
from 500 MHz to 1 GHz. Each scope amplifier. The base-line noise floor can see from the noise measurement
was terminated into 50-Ω and was set at the scope’s most sensitive full- results documented in Table 1 and
up to acquire waveforms with no input bandwidth V/div setting is Table 2, Agilent’s InfiniiVision Series
signal connected, using each scope’s a good approximation of this MSOs provide the lowest overall noise
maximum specified sample rate. component of noise. This component characteristics when used on most
of noise dominates at the most full-bandwidth V/div settings.
What is often considered the “base- sensitive settings, but it is negligible
line noise floor” of an oscilloscope is when the scope is used on the least
the level of noise when the scope is sensitive settings (higher V/div).
set at its most sensitive V/div setting
(lowest V/div). But many scopes on The second component of noise is
the market today have reduced a relative level of noise based on
bandwidth characteristics when they the scope’s dynamic range, which
are used at the most sensitive V/div is determined by the specific V/div
settings. As we mentioned earlier, setting. This component of noise is
scopes are broadband instruments negligible when the scope is used
and the higher the bandwidth, the on the most sensitive settings, but
higher the noise floor – typically. So it dominates on the least sensitive
if you compare base-line noise floor settings. Even though the waveform
characteristics at each scope’s most may appear to be less noisy when the
sensitive V/div setting, you may be scope is set at high V/div settings, the
comparing a lower-bandwidth scope actual amplitude of noise
against a higher-bandwidth scope, can be quite high, as you can
which is not an apples-to-apples see by comparing the noise
comparison. The base-line noise floor measurements in Table 1 at 1-V/div to
2
Understanding Noise and How it Should Be Measured (continued)
1. Bandwidth limited to 200 MHz at 1 mV/div and limited to 350 MHz at 2 mV/div
2. Bandwidth limited to 150 MHz at 2 mV/divv
3. Waveform expansion used below 5 mV/div
3
Measuring Peak-to-Peak Noise
Although for best results you should As these tables show, the Agilent will update waveforms at a rate of
evaluate and compare noise as an InfiniiVision series MSOs oscilloscope approximately 100,000 waveforms per
RMS value because of the random and produced the lowest overall level of second. This means that if you collect
unbounded nature of noise, it is often peak-to-peak noise (based on 1-M infinite persistence waveforms for
desirable to measure and compare points of data) at most full-bandwidth 10 seconds, the Agilent scope will
peak-to-peak noise. After all, it is the V/div settings. Again, both the collect approximately 3000 times more
peak excursions of noise that are Tektronix and LeCroy oscilloscopes data for these peak-to-peak noise
viewed on the oscilloscope’s screen produced significantly higher levels measurements. And as we mentioned
and induce the highest amplitude of peak-to-peak noise on most V/div earlier, the more data you collect, the
errors in real-time/non-averaged settings. more the peak-to-peak measurements
measurements. For this reason, many will grow because of the random and
oscilloscope users prefer to compare Although it may be tempting to just set Gaussian nature of random vertical
and measure noise as a peak-to-peak each scope at equal time/div settings noise.
value. But since random vertical noise and then collect data using the infinite
is theoretically unbounded, you must persistence display mode for a
first establish a criterion of how much set amount of time, such as 10
data to collect, and then realize that seconds, you should be cautioned not
peak-to-peak measurement results of to use this more intuitive method of
noise will be qualified on this criterion. peak-to-peak noise testing. Not only
Table 3 and Table 4 shows peak-to- can memory depths be significantly
peak noise measurements on the six different when the scopes are set up at
500-MHz to 1-GHz scopes tested based the same timebase setting, but update
on collecting 1-M points of digitized rates also can be significantly different.
data. For instance, if you initially begin
with a default setup condition and
then set the WaveRunner 104Xi and
Since it is possible that one particular Agilent MSO7104B to 20-ns/div, the
acquisition of 1-M points of data could LeCroy scope will acquire and update
produce either high or low peak-to- waveforms at a rate of approximately
peak measurements, we repeated 30 waveforms per second. Because of
these 1-M point peak-to-peak noise Agilent’s 7000 Series’ extremely fast
measurements ten times at each waveform update rates, the Agilent
V/div setting. These measurement scope with MegaZoom III technology,
results were then averaged to produce
a “typical” peak-to-peak noise figure
based on 1-M points of acquired data.
4
Measuring Peak-to-Peak Noise (continued)
50 mV/div 12 mV 15 mV 8.9 mV
100 mV/div 26 mV 37 mV 16 mV
200 mV/div 50 mV 60 mV 30 mV
Table 3: Typical peak-to-peak noise1 comparisons of 500-MHz and 600-MHz bandwidth MSOs
50 mV/div 13 mV 17 mV 12 mV
100 mV/div 30 mV 52 mV 19 mV
200 mV/div 55 mV 75 mV 38 mV
5
Noise Measurements with Probes
Most oscilloscopes come supplied will actually be set at 2-mV/div. This
with 10:1 passive probes, which means that you will probably observe
can provide up to 600-MHz system a fairly high level of noise relative to
bandwidth (for 600-MHz oscilloscopes the screen height since the base-line
or higher). In addition, active probes noise floor is effectively multiplied by
also can be used to achieve higher a factor of ten. If you need to perform
bandwidth in higher bandwidth scopes. critical low-level signal measurements,
Whether you are using a passive such as measuring the ripple of a
or active probe, the probe itself will power supply, you might consider
add an additional component of using a 1:1 passive probe. In addition,
random noise. Today’s digital scopes if the scope employs bandwidth
will automatically detect the probe limiting on its more sensitive V/div
attenuation factor and re-adjust the ranges, just be aware that bandwidth
scope’s V/div setting to reflect the limiting may now apply to higher V/div
input signal’s attenuation induced by settings, based on the particular probe
the probe. So if you are using a 10:1 attenuation factor.
probe, the scope will indicate a V/div
setting that is ten times the actual
setting inside the scope. In other
words, if the scope is set at 20-mV/
div with a 10:1 probe attached, the
scope's input attenuator and amplifier
6
Making Measurements in the Presence of Noise
When you use an oscilloscope on its
most sensitive V/div settings, inherent
random oscilloscope noise sometimes
can mask real signal measurements.
However, there are measurement
techniques you can use to minimize
the effects of the scope’s noise. If you
are measuring the level of noise and
ripple on your power supply, you may
need to use a scope at or near its more
sensitive V/div setting. First, try using
a 1:1 probe, as discussed above, rather
than using the standard 10:1 passive
probe that was probably shipped with
your instrument. Secondly, if you are
attempting to measure the RMS noise
of your power supply, your Figure 1. Noise measurement of power supply and scope/probe
measurements will also include noise measurement system
contributed by your scope/probe
system, which may be significant. But
with careful characterization of both
your signal (power supply) and your
measurement system, you can back
out the measurement system noise
component to provide a more accurate
estimate of actual power supply noise
(RMS).
7
Making Measurements in the Presence of Noise (continued)
Using a 1:1 passive probe on the non-correlated to the scope's auto accurate measurement of the average
Agilent MSO/DSO7000 series triggering, using this technique to back DC component of the power supply
oscilloscope, we measured out the measurement system error requires sufficient oscilloscope DC
approximately 1.5-mV RMS of noise component will provide a very close offset range (only available in the
on this noisy 5-V power supply. Figure approximation of the total RMS noise Agilent oscilloscopes documented in
2 shows a noise characterization of your power supply. this application note).
measurement of just the measurement
system using the same 1:1 passive Individual deterministic/systematic Using this averaged measurement
probe. With the ground lead of the components of interference, such technique on the same noisy
probe connected to the probe’s tip, as power supply switching or digital power supply signal, we measured
we measured approximately 480-µV system clock interference, also can be approximately 4.9-mVp-p of
RMS of system measurement noise measured accurately in the presence interference induced by the system's
at 10-mV/div. You will see that this of fairly large random measurement 10-MHz clock (lower/green waveform).
measurement of oscilloscope/probe system noise. By triggering on suspect To find all sources of deterministic
noise is significantly higher than the sources of interference using a (non-random) interference/ripple, you
figure shown in Table 1 (250-µV RMS) separate channel of the oscilloscope, need to perform multiple averaged
because we used a 1:1 probe, which you can repetitively acquire input measurements using various
adds an additional component of signals and then average-out suspected sources of interference as
noise. Plus, we used a 1-MΩ input all random and non-correlated your oscilloscope trigger source.
termination rather than the original components of noise and interference
50-Ω termination (used for the contributed by both the scope/probe
base-line RMS noise measurements and input signal. The result will be
documented in Table 1). Now, using a high-resolution measurement of a
a square-root-of-the-sum-of-the- particular component of interference
squares formula, we can back out this of your power supply, even when you
component of measurement system are using the scope on a very sensitive
noise, which indicates approximately V/div setting such as 2-mV/div, as
1.4-mV RMS of power supply noise. shown in Figure-3. Again, making an
8
Viewing the “Fat” Waveform
Some oscilloscope users believe that
digital storage oscilloscopes (DSOs)
induce a higher level of random
vertical noise than older analog
oscilloscopes do. They reach this
conclusion because a trace on a DSO
typically appears wider than a trace
on an older analog oscilloscope. But
the actual noise level of DSOs is no
higher than noise levels in older analog
oscilloscopes of equal bandwidth.
With analog oscilloscope technology,
random extremes of vertical noise
are either displayed dimly or not at all
because of the infrequent occurrence
of signal extremes. Although engineers
often think of oscilloscopes as Figure 4. 100% intensity display with no intensity gradation
simple two-dimensional instruments
that display volts versus time, older
analog oscilloscopes actually show
a third dimension as a result of the
swept electron beam technology they
use. The third dimension shows the
frequency-of-occurrence of signals
using trace intensity modulation,
which means that older analog
oscilloscopes actually hide or visually
suppress extremes of random vertical
noise.
9
Viewing the “Fat” Waveform (continued)
low-level input signal (approximately positive peak of the sine wave, that This technique can be used on single-
50 mVp-p) measured at 10 mV/div is were previously masked when viewed shot acquisitions to filter out high-
primarily due to inherent oscilloscope with a constant level of intensity frequency components of noise and
noise – not due to input signal noise. (100%) due to the relatively high level interference using DSP/digital filtering
Figure-5 shows the same 10-MHz of scope noise. to increase vertical resolution up to
signal, but now with the intensity 12-bits, but at the expense of
adjusted to Alternatively, you can eliminate measurement system bandwidth.
20% to more closely emulate the measurement system and random
display of analog oscilloscopes that signal noise using waveform averaging
naturally suppress extremes of noise. if you are acquiring a repetitive
We can now view a more “crisp” input signal, as illustrated in the
waveform without viewing the effects example shown in Figure 3. For real-
of the scope’s inherent noise at this time/single-shot applications (when
relatively sensitive V/div setting. In repetitive averaging cannot be used),
addition, we now can see waveform some scopes also have a high-
details, such as “wiggles” on the resolution acquisition mode.
10
Summary
When you evaluate various 256 levels of intensity grading that can oscilloscopes on the market. Agilent
oscilloscopes for purchase, be sure to be used to visually suppress random is able to achieve lower measurement
carefully consider the inherent noise extremes of inherent oscilloscope noise performance primarily because
characteristic of the oscilloscopes. Not noise. of higher levels of integration using
all oscilloscopes are created equal. lower-power integrated-circuit (IC)
Not only can vertical random noise in a You can use various measurement technologies.
scope degrade measurement accuracy, techniques, such as mathematical
but it also can degrade the viewing calculations, waveform averaging, DSP It should be noted that a very limited
quality of digitized signals. When you filtering, and display intensity gradation sample size of oscilloscopes was
are evaluating oscilloscope noise to minimize or eliminate measurement chosen for this characterization of
characteristics, it is important that you system noise components in order random vertical oscilloscope noise.
carefully setup oscilloscopes to be to make more accurate measurements All scopes selected for testing
tested under the same measurement of low-level random and deterministic were recent production units from
criteria including same bandwidth noise components in your system. each vendor. Only channel 1 was
scope, same V/div setting (with full tested because this is the channel
bandwidth), same sample rate, same Although this document has focused engineers most often use. Although
memory depth, and same number of on noise measurements comparisons we don’t know for sure if the
acquisitions. between just 500-MHz to 1-GHz measurements documented in this
bandwidth oscilloscopes, the principles paper are typical, we have assumed
Agilent’s InfiniiVision Series presented in this document apply to that these measurement results are
oscilloscopes provide the lowest any bandwidth oscilloscope – higher representative of all current units in
overall noise characteristics in the or lower. In fact, Agilent’s higher- production by each scope vendor.
industry compared to other 500-MHz bandwidth 13-GHz DSO90000 Infiniium
to 1-GHz oscilloscopes, as shown in Series oscilloscope has the lowest
this document. In addition, Agilent’s inherent internal measurement system
InfiniiVision Series oscilloscopes with noise of any oscilloscope in this
MegaZoom III technology provide the bandwidth range with noise levels not
highest resolution display quality with that much higher than current 1-GHz
11
Glossary
Base-line Noise Floor the level of RMS noise measured at a scope’s most
sensitive, full bandwidth V/div setting
Noise Floor the level of RMS noise measured at each scope's V/div setting
Dynamic Range the full range of a digital storage oscilloscope’s (DSO’s) analog-
to-digital converter that depends on the scope’s V/div setting, which usually
varies from 8 division peak-to-peak (full screen) to 10 divisions peak-to-peak in
most oscilloscopes
12
Related Literature
13
www.agilent.com
Agilent Email Updates Remove all doubt
www.agilent.com/find/emailupdates Our repair and calibration services
will get your equipment back to you, For more information on Agilent Technologies’
Get the latest information on the products and products, applications or services, please
applications you select. performing like new, when promised. contact your local Agilent office. The
You will get full value out of your Agilent complete list is available at:
equipment throughout its lifetime.
Agilent Direct Your equipment will be serviced by www.agilent.com/find/contactus
Agilent-trained technicians using the
www.agilent.com/find/agilentdirect Americas
latest factory calibration procedures,
Quickly choose and use your test Canada (877) 894-4414
automated repair diagnostics and
equipment solutions with confidence. Latin America 305 269 7500
genuine parts. You will always
United States (800) 829-4444
have the utmost confidence in your
Agilent Channel Partners
measurements. Asia Pacific
www.agilent.com/find/channelpartners Australia 1 800 629 485
Get the best of both worlds: Agilent’s Agilent offers a wide range of additional China 800 810 0189
measurement expertise and product expert test and measurement services Hong Kong 800 938 693
breadth, combined with channel for your equipment, including initial India 1 800 112 929
partner convenience. start-up assistance onsite education
Japan 0120 (421) 345
Korea 080 769 0800
and training, as well as design, system
Malaysia 1 800 888 848
integration, and project management. Singapore 1 800 375 8100
Taiwan 0800 047 866
For more information on repair and Thailand 1 800 226 008
calibration services, go to
Europe & Middle East
www.agilent.com/find/removealldoubt Austria 43 (0) 1 360 277 1571
Belgium 32 (0) 2 404 93 40
Denmark 45 70 13 15 15
Finland 358 (0) 10 855 2100
www.lxistandard.org France 0825 010 700*
LXI is the LAN-based successor to GPIB, *0.125 €/minute