Professional Documents
Culture Documents
Figure 6.1 Viewing a reticle with dust, smears, and fungus in the instrument. Note
the small particles of dust. Fungus appears as an irregular, fuzzy, light grey shape
with a dark center; smears are the dark lines in the upper right side of the image.
Due to eye safety reasons, lasers and laser pointers should never be used as
a light source to view instruments or contaminants—the reflected bright beam
can strike the eye. See Section 1.1.
Figure 6.2 Viewing an instrument using an LED light source. Use a notebook to
record and map the areas to be cleaned. Be sure to position cleaning tools in the
work area.
Figure 6.3 Viewing an instrument using an LED light source and an applicator.
Figure 6.4 Position marks on the mounting cell and surfaces of the lenses. Note
the slight movement of the lens to the cell reference mark. The lens was loose in
the cell and caused a small rotation of the lenses.
Figure 6.5 Marking the edge of a lens with a permanent ink pen to help with the
realignment after cleaning.
Figure 6.6 Marking the lens and edge of the mounting cell for reference, using a
china marker.
Figure 6.7 Marking the edge of a lens using a pencil with an upside down “V”.
• An arrow, arrow tip, caret, or “V” are simple marks that can be
used to show the direction of the light (Fig. 6.7). Dots or lines
are good for clocking the elements. Mark the optical cell with
the location of the optic relative to the cell.
Figure 6.9 A spanner wrench is used in the removal of retainer rings from a lens
cell (image courtesy of FitTools, Inc.).
Figure 6.10 Spanner wrenches are made in various sizes. They adjust to fit a
range of retainer rings. A spanner wrench can be custom made for special
applications (image courtesy of FitTools, Inc.).
is cause for concern. Start with any surfaces exposed to the environment
and carefully look at each surface or optic under normal room lighting.
If the contaminant is found, begin the cleaning process; if it is not easily
observed, it will require closer inspection of all surfaces with a better
light source and optical aids. The unit should be powered down for
safety and to prevent overexposure to the detector or imager. Always
check the instrument manual for directions: if the process of cleaning is
limited due to possible damage to the detector or imager or if the optics
cannot be removed, contact the manufacturer for further instructions.
A small or portable instrument can be examined on a worktable.
A diffused light source should be used to observe most contaminants on
the outer exposed surfaces. Tilt and move the instrument to examine
each surface under the light. If any surface is not observable, use a
handheld, portable light source instead. To view the optical system
correctly, look at each optic one surface at a time. Make notes about
what is observed.
The position of a contaminant can be found by comparing its size or
shape with the image on a TV, computer screen, or detector. Look at how
well focused the contaminant appears: outside of an image plane, a
contaminant will be out of focus, whereas a contaminant at the image
plane, such as a reticle or the surface of a CCD, will appear to be in focus.
Starting with the exposed surface, use compressed dry air to remove
dust and fibers, and then re-examine. If the contaminant has not been
removed, solvent on an applicator may be required. Use caution during
this process—solvent use should be limited to ensure that it does not
leak into the system. Afterwards, test the system performance: if the
contaminant is still visible, investigate further. Restart the process with
a different light source, such as fiber optic or focused light.
it is safe (i.e., it does not warm the surfaces or interfere with viewing due to
brightness). Review the instrument manual for safety concerns.
Note that lasers in the system should never be turned on or used as the light
source.
Prepare the workstation with the proper tools and lighting. The
instrument should be disassembled in a clean area, such as a laboratory or
under a flow bench. Small containers are handy to store various screws,
nuts, washers, springs, and other parts. The optics should be handled with
the utmost care and stored either on a tray or in a labeled, covered box; a
tray can be sectioned off to identify each optic as it is removed from the
instrument. Place alignment marks on the edge of each optic.
If the optics are easily accessible and the surfaces are exposed,
cleaning in place is an option; however, this option depends on what needs
to be cleaned and how it is mounted. Dust and fibers can be blown off
using a hand puffer or dry, low-pressure air (high pressure will blow the
dust and fibers into the instrument). Cleaning oils, stains, smears, or
fingerprints will require the use of solvents and will be difficult to do—a
strippable coating or the CO2 cleaning process may be viable alternatives.
Figure 6.12 A typical work area arranged for cleaning a refractor lens.
washer, or set screws). If the lens is glued, in the cell only the outer
surfaces can be cleaned.
Before removing the lens cell from the tube, check the alignment
of the lens to the tube. If needed, test the optical performance before
disassembly. Note the position and mark its location to the tube.
Carefully remove the lens cell from the tube, then rotate or loosen the
set screws. Place the lens cell on the workstation (Fig. 6.13), and
inspect the lens again.
Clean the outer surfaces—first with soap and DS water on an
applicator, then with a solvent if needed—and re-inspect. If the inside
surfaces need to be cleaned, use a china marker or pencil to mark the
clock position of the lenses in relation to the cell with tick marks on
the outer surface and cell and the inner surface and cell.
Remove the retainer ring with a spanner wrench. (If other methods
are used, choose the corresponding tool.) Lay the lens assembly on the
covered post, then slowly remove the lens from the cell by lowering the
cell (Fig. 6.14). Mark the edges with a pencil and clean the lens using
the normal procedure (Fig. 6.15), but avoid removing the tick marks.
Re-inspect the lens and repeat as needed (Fig. 6.16).
Figure 6.13 After the lens cell is removed from the tube assembly, position the
cell on a foam mat and remove the retaining ring.
Figure 6.14 Removing the lens from the cell by lowering the cell using the
methods described in this section.
Figure 6.15 Cleaning the individual lenses using a wipe and a hemostat.
Figure 6.16 Inspecting the lens using a light-colored background (note the air
gap spacers).
Reassemble the lens, using the edge and tick marks to realign. Place
the lens on the cover post. Carefully lift the cell to fit the lens inside, and
then remove the lens assembly from the post. Secure the lens with the
retainer ring or set screws. Check the tick mark positions—if the lens
rotated, realign it. Reassemble the telescope and realign the optics to
telescope performance specifications.
Figure 6.18 Looking down the tube of a typical small-size reflector (152 mm). The
primary mirror is located at the bottom. The mount for the folding mirror or
secondary is supported by the 4 vanes. The eyepiece focuser is visible in the upper
left corner of the image.
Figure 6.19 The primary mirror removed from the mounting cell and placed in a
tray ready to be cleaned.
should be covered with a foam or rubber mat. If needed, raise the mirror
up on wooden or plastic blocks. Make sure that the mirror is properly
secured before tilting it at a slight angle to drain the cleaning liquid.
Pour warm (20–30 °C) distilled water over the surface to flush away
any large particles. With a gloved hand and a large wipe (dampened
with warm distilled water), using only the weight of the wipe, slowly
wipe the surface in a random pattern. Rinse with clean DS water, then
blow with clean, compressed air and inspect the mirror.
A small mirror (<250 mm) can be soaked in a tub of warm soap and water; use
a large wipe in a random pattern to clean the surface. See Figs. 6.20 and 6.21.
A large mirror (>250 mm) lying flat or tilted on the workstation, or in a large
sink, can be washed in sections with soap and water.
Rinse the mirror with distilled water only. Dry by patting with a wipe
(Fig. 6.22) or use clean, dry air. Finish cleaning with a solvent and
applicator if needed. Be sure to dry the sides and bottom of the optic; it
may be necessary to dry under a flow bench. Inspect the surface, and
Figure 6.20 The pouring of distilled water and soap directly onto the surface.
Figure 6.21 With the mirror submerged in the water, using a wipe to clean the
mirror.
record any defects on a map. Reassemble the optics into their cells using
the reference marks. Review the procedure on how best to secure the
optics with the mechanical parts. Secondary mirror cleaning (mounted
and surface-exposed) is shown in Figs. 6.23 and 6.24.
The mechanical parts should not affect the performance of the
optic. Test the optic in the cell to look for pinch points, if needed.
Reassemble the telescope and realign the optics to telescope
specifications.
Blow away any dust or fibers with clean, dry, compressed air and
inspect the surface. If needed, clean with soap and distilled water or a
liquid lens cleaner.
Dampen the applicator or wipe and shake to remove any excess
liquid. Do not allow any liquid to leak into the cell. Using an outward
spiral pattern, stroke the applicator or wipe from the center out to the
Figure 6.26 Using an applicator to clean a small area on the correcting lens of a
catadioptric telescope.
edge (Fig. 6.26); inspect and repeat if necessary. Solvent use should be
limited to alcohol—some correctors have a painted spot in the center to
cover the back of the secondary mirror, and acetone or another strong
solvent can remove it.
Decide if the internal surface of the corrector needs to be cleaned. If
so, determine how the corrector is secured to the tube assembly (in most
cases, there is a threaded or flat retaining ring secured with screws).
Note the position of the corrector and mark its location relative to the
tube assembly; this is important during reassembly, as the optical
performance can be degraded if it is not positioned in the same location.
Test the telescope system prior to disassembly for reference.
Remove the retainer ring. Use care when removing the corrector
from the tube assembly (suction cups may be needed). Place the
corrector on the workstation. Under diffused light, inspect the
corrector. Follow the lens-cleaning procedure, and avoid removing
the reference marks. Inspect and, if clean, reassemble the corrector to
the tube assembly. The primary mirror can be cleaned using the
reflective telescope procedure. Test the telescope system to confirm
that its performance is within specifications.
During the optic cleaning process there is an opportunity to remove old grease
from the mechanical threads and gears, and to apply new grease. Use
wooden sticks and a foam applicator (no solvent) first to remove most of the
gummed grease. Use a lightly dampened applicator with solvent for final
cleaning; be cautious of plastic parts. Apply new grease per the manufac-
turer’s instructions.
6.2.8 Eyepiece
Eyepieces used in microscopes, telescopes, spotting scopes, and
binoculars need special care when cleaning the exposed optical surfaces
of the field lens and eye lens (Fig. 6.31). The field lens is the closest element
to the image plane. Contaminants such as dust, stains, and fingerprints on
the field lens can be magnified by the eye lens. Contaminants on the eye
lens surfaces will interfere with viewing; small specks, moisture, or dots or
stains from eye tears degrade the appearance of images.
To inspect the eyepiece, point the instrument or eyepiece at a light-
colored screen or wall (Fig. 6.32). Look through the eyepiece in both
directions: the background should appear unfocused and flat without
structure. Slowly rotate the eyepiece a quarter of a turn and observe; note
the position of dust or other contaminants. In some instruments, other
optical surfaces can be in or near focus, such as surfaces on a prism,
mirror, or reticle, which can be mistaken for the eyepiece surfaces. If they
rotate with the eyepiece, remove the eyepiece from the instrument.
Inspect the field lens and eye lens in reflection under a diffused light,
and then blow each surface with dry air. Inspect the surfaces, and then
clean them with a solvent-dampened applicator by stroking from the
center out to the edge (Fig. 6.33). Solvent may not be necessary—moisture
or tear stains may require using soap and DS water or just water alone
to remove. Repeat, and then inspect the eyepiece using the
background. It may be difficult to remove all of the contaminants,
Figure 6.33 Cleaning the inside surface of a field lens of an eyepiece assembly.
but it is important: even the smallest dust can appear 10–40× larger
due to magnification. Install the eyepiece back into the instrument to
verify that the dust and contaminant has been removed.
Outer surface cleaning should reduce the amount of dust and
contaminants and improve the performance of the eyepiece. It is a
personnel or functional decision whether further cleaning is required;
if so, the eyepiece will have to be taken apart.
Figure 6.34 Cleaning the front surface of a camera lens using a wipe.
Figure 6.35 Cleaning the front surface of a camera lens using an applicator.
Figure 6.36 A typical CMOS digital eyepiece for a microscope with a USB
connector for computer imaging. The sensor is the small rectangle inside the tube
(image courtesy of Dino-Lite Company).
Figure 6.37 Image of “Dust Donuts” caused by dust located on the internal
surfaces of a digital camera.
Figure 6.38 Cleaning the cover plate of a CCD imager using a pointed applicator.
Vehicle and building windows may not be precision optics, but they are part of
visual optics. Many technicians in the optics field have been asked how to
clean these surfaces.
For uncoated surfaces and glass types such as plate glass (float
glass), extra-hardened contaminants on the glass surface can be
removed by using a single-edge razor blade or a plastic or wooden
scraper. Start by pouring the cleaning liquid on a paper or cloth
wipe, place the wipe over the area, and then soak the contaminant
and allow the liquid to hydrate the dried substance. After a few
minutes, rinse away any particles; this will reduce the chance of
scratching. Starting at one edge of the contaminant, tilt the razor or
scraper to an angle of 10–25 deg, using light pressure and a
semicircular hand motion. Scrape the contaminant by working the
tool from one edge across the contaminated area. Soak and repeat,
scrape as needed. Clean around the area with a wipe to remove any
remaining particles. A sponge covered with a mesh or textured
surface on its backside can also be used to lightly scrub the offending
contaminant.
One of the best mixtures for cleaning off road grime, persistent dust,
insect parts, vinyl film stains, and unknown contaminants is a mixture
of basic white toothpaste, soap and water, or diluted (water) and/or
ammonia-based glass cleaner. Place a dab of toothpaste on a dampened
paper towel or squeeze a dab from the tube onto the glass surface.
Using fingers or the towel, slowly work the toothpaste in a circular
pattern over a wide area. The scrubbing action of the chalk in the
toothpaste will lightly scrub the glass. The effect of the light scrubbing
will clean so that water sheets off the glass easily. See Figs. 6.39–6.42 for
an illustration. This method is also used on a scuba diver’s uncoated
mask window to reduce the effects of fogging during a dive.
Figure 6.40 Wetting the toothpaste and wipe with the cleaning solution.
Figure 6.42 Cleaning the surface, reapply cleaning solution if needed. Dry the
surface using a fresh wipe.
wipe, pat the surface to collect particles, and repeat as needed. Dry
and inspect. Oily contaminants should be cleaned with solvents such
as acetone or methanol; pat the surface during application. Patting
reduces the chance of scratching or damaging during this step.
If the metal optic is not mounted, flush the surface with a mixture
of soap and DS water. Dampen a cloth or paper wipe with soap and
water, and use just the weight of the wipe while stroking across the
surface—no pressure should be applied to the applicator. Rinse the
surface with clean water.
For a difficult-to-clean contaminant, place a dampened wipe or
applicator with solvent over the area, and hold it in place for a few
seconds so that the contaminant dissolves into the wipe (see Figs. 6.43
and 6.44). Lift slowly while rotating the wipe up and away from the
surface. Use clean, dry, compressed air to remove any fibers or
particles. Inspect and repeat as needed. An ammonia-based liquid
cleaner will also work on these metal substrates (do not use on
aluminum). Test on a small area first and look for any reaction before
continuing to clean the whole surface.
Figure 6.43 Cleaning the polished surface of a metal optic using an applicator.
Figure 6.44 Cleaning the polished surface of a metal optic using a small folded
wipe.
A commercial metal cleaner can be tried, but dilute the liquid by 50% with
distilled water. Test the solution on a small area of the surface or use the tip of
an applicator pressed against the surface, then re-inspect. Some cleaning
metal liquids contain hard particles used for scrubbing—do not use these.
Figure 6.45 Cleaning microscope slides with a wipe after being soaked in a warm
bath of distilled water and soap.
bowl for a few minutes. Using a gloved hand or tweezers, remove a slide
and rub it with a wipe dampened in solvent. Inspect the slide under a
diffused or fiber light and repeat if needed. Clean the slide with soap
and water after solvent cleaning. Alternatively, lay a cotton or fiber
wipe on a clean, flat table, board of plastic, or sheet of glass. Pour a
mixture of water and soap on the wipe to dampen. Holding the slide
with a gloved hand, place it on the wipe, and then slowly move the slide
back and forth with mild pressure. Flip to the other side and repeat.
Place the slide in a bowl of warm water to soak for a few minutes, hand
dry using a wipe, and then air dry in a covered tray. Use the chalk
method for heavily soiled slides.
Removing a mounting media with a resin base (Canada balsam)
or synthetic resin may require the use of toluene or xylene solvents.
Clear nail polish can be removed with acetone. Check with the
manufacturer of the mounting media for which solvent is best.
Observe and follow all safety requirements for the handling and use
of any solvent.
Eyeglasses may not be precision optics, but they are part of visual optics.
Many technicians in the optics field have been asked how to clean these
surfaces.
Rinse for 15–30 s, and slowly move the lenses back and forth
under the water to cover the surfaces. The warmth of the water
and its motion across the lens will remove most of the oils, dirt,
and dust. After the rinse is complete, dry the lenses by hand using
a soft cotton (clean) or microfiber cloth or wipe (Fig. 6.47). Some
paper tissue products can be too rough and will scratch an AR
coating on the lenses. Inspect the eyeglasses under a light source
to check for cleanliness. A clear dishwashing liquid can be used if
necessary: wet the surface with water and place a few drops
directly on each lens, rub the liquid over the surfaces using a light
touch with a finger or cloth, rinse completely, and then dry with a
soft cloth.
There are times when an eyeglass liquid cleaner may be required.
Modern lenses are made from various materials, and the safest liquid
cleaner contains only a few drops of soap and DI or distilled water
with alcohol. These cleaners come in a spray pump bottle. Spray the
liquid sparingly on the lenses (see Fig. 6.48), and, using a soft cloth
and light finger pressure, gently wipe both surfaces of each lens
Figure 6.47 Using a wipe to clean both surfaces with the thumb and index
fingers.
between the thumb and index fingers until dry. Inspect and repeat as
needed. Special eyeglass cleaner wipes that are pretreated with
solution are also available; follow the manufacturer’s usage
instructions.