CAUTION!
*3-5600.090-1* • Remove power to unit before wiring input and
3-5600.090-1 (Rev. E-8/04) English output connections.
• Follow instructions carefully to avoid personal
injury.
1. Wiring terminals 1. 2. 3.
1 2 3 4 5 6 1 2 3 4 1 2 3 4 5 6 1 2 3 4
Sen. Pwr.
+ +
Sen. Pwr.
Gnd Gnd 1 2
Freq. IN
Freq. IN
2
Flow
Gnd 1
Iso. Gnd
Gnd
Freq. IN
Freq. IN
Flow
Iso. Gnd
PLS AUX Total PLS AUX Total
UL LISTED
Sensor
UL LISTED
77CJ
Sensor
77CJ
12-24 V 12-24 V
10 W 1 2 10 W
+ - + -
1 2 1 2
2. Power Connection
5600
CAUTION!
Never connect 115 VAC or 230 VAC to rear power 12-24 V Terminals
10 W
terminals. High voltage AC will damage instrument and External power
void warranty. supply
+ -
Technical Notes: - 12 - 24 VDC
• Maximum 4-20 mA loop impedance (sec. 8A) is affected by the OR
= Double Insulated 12 - 24 VAC
supply voltage. +
• To reduce the possibility of noise interference, isolate AC power = DC or AC power
lines from signal lines.
Freq. IN
Vortex* power
5600
Std. Sensor flow
Open Collector Terminals
Sensor valve
Technical Notes: Technical Notes:
• To reduce the possibility of noise interference, route sensor • Maximum alarm contact ratings: 5 A @ 30 VDC, 5 A @ 125
cable )away from AC power lines. VAC, or 3 A @ 250 VAC
• *Vortex sensor or system frequency output • To reduce the possibility of noise interference, isolate AC power
lines from signal lines.
5. Remote Control Wiring 6. End of Batch/Counter Pulse Output Wiring
See section 9 for simple and advanced mode configuration op-
Rear START, STOP, and RESUME Momentary Switch Contacts tions.
(Rsm) terminals can provide remote (customer supplied)
batch control from a distance using
Start A. External counter connection B. Daisy chaining two 5600
one of four methods: batch controllers together
Stop
• Mechanical switch closures (shown) + 5 - 30
Rsm
RTN
Start
Stop
Gnd
Rsm
RTN
Start
Stop
second 5600 (sec. 7C). Terminals +
Gnd
CNT/EOB
Remote output Remote
• End of batch current pulse from a CNT/EOB
second 5600 (sec. 8C).
output 5600
Terminals First 5600 Second 5600
Fuse** 4 - 20 mA
5600
1/8A
+
Terminals *
Option
NO C NC NO
Batch
C NC
-
+ - 5600
AC or DC Current
Terminals
power output
Bypass line
B. Missing Signal Alarm (simple or advanced mode, sec. 9) B. Valve Control (advanced mode only, sec. 9.2)
OR Overrun Alarm (advanced mode only, sec. 9.2)
#5 ( + )
4 - 20 mA
5600
Terminals #6 ( - )
Option Batch
NO C NC NO C NC
+ - 5600
Terminals George Fischer electronic
Current
Alarm AC or DC output actuator with PE 22P
device power positioner required
C. End of Batch Pulse (advanced mode only, sec. 9.2) C. End of Batch Pulse (advanced mode only, sec. 9.2)
Daisy chaining two 5600 batch controllers together for a second Daisy chaining two 5600 batch controllers together for a second
batching stage. batching stage.
5600
Option Batch
Terminals
NO C NC NO C NC
Rsm
RTN
Start
Stop
Terminals
Current
Rsm
RTN
Start
Stop
output Remote
Second
Remote
5600
First 5600 Second 5600
Technical Notes: Technical Notes:
• Use 2-conductor shielded twisted-pair cable for contact lines • Wiring MUST be connected exactly as shown.
up to 30 m (100 ft.) max. • Use 2-conductor shielded twisted-pair cable for output lines
up to 30 m (100 ft) max.
• Counter/End of Batch Pulse Output (sec. 6): In simple mode, the output is configured as a counter pulse output. The output emits a
130 millisecond pulse for each engineering (total) unit measured. The output is a open collector type which requires an external pull up
resistor and power supply for external counter use (sec. 6).
• Option Contact - Missing Signal Alarm (sec. 7B): This function provides alarm capability if flow is not detected in 30 seconds after
a batch cycle starts. If the flow sensor signal is missing, the option contact energizes and the front panel Batch started...
contact indicator turns on, signaling the missing sensor signal. The batch contact also Sensor input
de-energizes, closing the flow control valve and stopping the batch. The alarm condition will remain until the
front panel ENTER key is pressed or a remote contact closure is made to the rear STOP terminal (sec. 5). option contact
batch contact
• Current Output - Batch Completion (sec. 8A): This function is offered in the simple or advanced opera-
tion mode. The output is a linear increase from 4 mA at batch start to 20 mA at batch end.
Batch running... 100%
0%
20 mA
9.2 Advanced Operation Mode
• Batch Contact (sec. 4): The batch contact is dedicated for on/off control and is either manually or auto- 4 mA
matically
compensated for batch overrun. Manual overrun compensation allows the operator to compensate for valve closure time in a batching
system. The operator is prompted to enter a batch volume that de-energizes the batch contact early, compensating
Overrun Comp:
automatic manual for valve closing time and eliminating batch overrun. Automatic overrun compensation counts sensor pulses during
the batch cycle, and any excess pulses after the batch stops. The instrument calculates the estimated batch overrun
based on the additional sensor pulses, then automatically reduces the next batch size for the next batch. During the next batch cycle, the
batch contact de-energizes early, thus closing the flow control valve early and eliminating batch overrun.
Option Contact: • Option Contact - Two Stage Shutdown (sec. 7A): This function is designed to pre- Overrun
Batch running...
Overrun Alarm vent over filling or to minimize water hammer. Both the batch and the option contact 0%
100%
are energized when the batch starts. The option contact then de-energizes at Option contact
a programmed batch percentage, forcing flow through a smaller bypass line to reduce the fill rate. After
the entire batch is measured, the batch contact de-energizes, completing the sequence. Front keypad batch contact
and remote START, STOP, and RESUME inputs (sec. 5) also control the option contact. The front panel
contact indicator turns on when the option contact is energized.
• Option Contact - Overrun Alarm (sec. 7B): This function provides overrun alarm
Option Contact: Batch running... 100%
End of Batch capability for detecting a leaky or stuck valve. If the programmed batch overrun
0% 500ms
volume is measured after the batch ends, the option contact is energized and the front Option contact
panel contact indicator turns on, alarming the operator that the flow shut off valve is leaking or stuck open.
To cancel alarm, press the ENTER key (or remote STOP switch) once. The overrun alarm will retrigger if batch contact
the condition persists.
Option Contact: • Option Contact - End of Batch Pulse (sec. 7C): This function is designed to trigger an external batch counter or
Missing Signal second 5600 Batch Controller at the end of a batch. The option contact is energized for 500 milliseconds at the end
of every batch cycle. An external power supply switched through the option contact's common (COM) and normally
open (NO) terminal provides a pulse for triggering these devices.
Option Contact: • Option Contact - Missing Signal Alarm (sec. 7B): The missing signal alarm represents a missing sensor signal
Off after batch start-up. An adjustable time delay entry (in seconds) is offered which specifies how long the instrument
will wait after batch start-up before initiating the missing signal alarm. See simple operation mode (sec. 9.1) for alarm
operation details.
Missing Signal • Option Contact On/Off (sec. 10.4H): This function enables or disables the option contact and front panel option
Alarm: On LED. When the option contact is configured "Off", the option contact and front panel LED are disabled.
Batch running...
Current Output:
• Current Output - Batch Completion (sec. 8A): In advanced mode, the current 0% 100%
valve
Valve Control output functions identical to the simple operation mode. See the simple operation mode
20 mA valve
explanation above.
4 mA
Current Output:
• Current Output - Valve Control (sec. 8B): This function is designed for use with a George Fischer electric actua-
End of Batch tor with PE 22P positioner for preventing overflow or minimizing water hammer. When a batch starts, the current
output is held at 20 mA, forcing the PE 22P positioner to the full open valve position. Batch running...
100%
Near batch end, the output gradually ramps downward to slow the flowrate, then drops to 4 mA closing the
valve and ending the batch. Front keypad and remote START, STOP, and RESUME inputs also control 0% 20 mA
the output (sec. 5). A STOP command any time during a batch sequence will force the output to 4 mA and 500ms
0 mA
close the PE 22P positioner. (Note: for optimum performance, the batch process should last at least 1
minute!)
OR Permanent Total
Batch Volume:
200.500 Gallons
To exit without
changes:
Overrun: Manual
2. Enter volume E. 0.00000 Gallons E. Shown only when "advanced" mode is chosen
quick (sec. 10.4E)
press Overrun compensation
3. ENTER Save
To step C → ← To step A)
1 2 3
Choose: Change: Save: ←Back to step C
G↓
Batch Units: Set batch units: Batch K-Factor: Set batch
Batch Units: Aa - Zz, 0 - 9, Batch K-Factor:
A. Gallons >
Liter D. 60.000 >
3456.7 K-Factor:
/, _ 0.0001 - 99999.
Batch units Batch K-Factor (see note*)
Auto Calibrate:
ENTER
<ENTER> to Stop Cancel ENTER Totalizer K-Factor (see note*)
Auto Calibrate Start
Procedure ENTER
Auto Cal Volume:
XX.XXX Gallons OR
Stop "SAVING" Last CAL: Last CAL: Sets user
Accurate reference Last CAL: defined date
G.
ENTER
briefly 01-01-98
container required. Enter volume Save displays 01-01-98
See procedure
(sec. 10.3) Last calibration
Batch K-Factor:
A. 1234.5
←To step A
Reenter ENTER
Batch K-Factor To return To restore original value:
volume? 1234.3 to VIEW:
B.
Accept quick quick
K-Factor? press press
C.
Modify
Display shown only Overrun Volume: K-Factor?
when advanced mode XX.XXX Gallons
(sec.10.4E) and manual
overrun comp. (sec.10.4G) ENTER
*Note: Batch and totalizer K-Factor settings represent the number
is selected. D. Overrun comp. Save of pulses generated by the +GF+ SIGNET flow sensor for each
To step D →
engineering unit of fluid measured (published in the flow sensor
To return To restore original value: manual).
to VIEW:
quick quick
press press
5600
Requirements:
40 60 Batch contact terminals (sec. 4) OR
A. The flow sensor must be installed in the actual batch pipeline. Batch Option
B. The flow control valve can be connected to the common (C) 0 100
100
80
60
40
20
0
Reference Container
WARNING!
The auto calibration procedure generates a custom batch K-Factor. The batch and Totalizer K-Factors are independent from
one another. ALWAYS reenter your totalizer K-Factor after performing the auto calibration procedure.
6 of 12 Signet 5600 Batch Controller
10.4 OPTIONS Menu (factory defaults shown in menu column 1)
1 2 3 ←Back to step F
Menu Navigation
NOT shown if Missing Signal Alarm is Off (step H)
• Advanced mode chosen (step E): menu continues, steps F - I.
• Simple mode chosen (step E): down arrow key advances to Missing Signal Set missing
I. Missing Signal signal time
top of menu (step A). Delay: 30.0 sec > Delay: 25.0 sec
delay: 0.10 -
Option Contact 999. sec.
←To step A
To return To restore original value:
to VIEW:
quick quick
press press
• Power supply, 115 VAC - 24 VAC, #3-5000.075 (code 159 000 321)
• Front snap-on bezel, #3-5000.525-4 (code 159 001 221)
• 5600 instruction sheet #3-5600.090-1 (code 198 869 929)
12. Specifications
General Agency Approvals
Sensor compatibility: All Signet flow sensors • CE, CSA, UL listed
Accuracy: ±0.5% of batch reading • Manufactured under ISO 9001 and ISO 14001
Input: Optically isolated
Enclosure: Electrical
• Rating: NEMA 4X/IP65 front Power:
• Dimensions: 1/4 DIN, 96 x 96 x 88 mm (3.8 x 3.8 x 3.5 in.) • 12 to 24 VDC or 12 to 24 VAC, unregulated, 50-60 Hz,
• Case materials: ABS plastic 10 W max.
• Keypad material: Sealed 4-key silicone rubber
• Weight: Approximately 500 g (18 oz.)
Contacts (2 sets):
• Mechanical SPDT contacts
Display: • Maximum voltage rating: 5 A @ 30 VDC, 5 A @ 125 VAC, or
• Type: Microprocessor controlled air-core meter movement and 3 A @ 250 VAC, (power factor = 1.0)
alphanumeric 2 x 16 LCD
• Update rate: <200 ms
• Contrast: User selected Current output:
• Alarm annunciators: green batch LED (left), red option LED • 4 to 20 mA, non-isolated, internally powered
(right) • Update rate: <200 ms
• Max loop impedance: 350 Ω max with a 12 V instrument supply
voltage, 950 Ω max with a 24 V instrument supply voltage
Totalizers: • Accuracy: ±0.1% of max range
• 8-digit resettable with security option
• 8-digit non-resettable
Totalizer (CNT)/End of batch (EOB) pulse outputs:
• Open-collector transistor, optically isolated, 5 mA max. sink,
28 VDC max. pull-up voltage, 130 ms (CNT) pulse width
• End of batch pulse output 500 ms (fixed)
GND
Environmental
Operating temp.: -10 to 55 °C (14 to 131 °F), 50 °C (122 °F)
max. with optional rear cover
Storage temp.: -15 to 80 °C (5 to 176 °F) Noise immunity: EN50082-2
Relative humidity: 0 to 95%, non-condensing Noise emissions: EN55011
Altitude: 4000 m max. Safety: EN61010-1
Pollution degree: 2
Panel
96 mm (3.8 in.) Optional Splashproof Rear Cover (sec. 11)
Gasket
+GF+
40 60 Mounting Panel
Batch % Option
Batch 96 mm
20 80
(3.8 in.)
Mounting Clamp (2 ea.)
0 100
96 mm 88 mm
(3.8 in.) (3.5 in.)
91 mm
(3.6 in.)
SIDE VIEW
ENTER
Front
Signet Controller
View
76 mm (3 in.)
96 mm (3.8 in.)
88 mm (3.5 in.)
88 mm (3.5 in.)
Batch
Iso. Gnd
+
Sen. Pwr.
Freq. IN
Freq. IN
Gnd
+
Rsm
RTN
Start
Stop
CNT/EOB Current
Remote
Patent No. D 376,328
output output
Std. Sensor
Open Collector 88 mm 92 x 9
Sensor
(3.5 in.) (3.62 2 mm
12-24 V x 3.62
10 W in.)
Option Batch
Signet
Model #3-5600 + - NO C NC NO C NC
Rear
View Panel cutout
15. Troubleshooting
Time Left - - - min Remaining batch time greater than Display will show actual remaining minutes after the remaining
<ENTER> to Stop 999 minutes (overrange) batch time is less than 999 minutes
-0001.8 Gallons Flow rate detected after batch end Repair leaky or stuck flow shuttoff valve.
<ENTER> to Start
Flow K-Factor Flow (batch) display K-Factor cannot Enter K-Factor greater than zero (sec. 10.2D)
cannot be zero be zero
Total K-Factor Totalizer K-Factor cannot be Enter K-Factor greater than zero (sec. 9.2E)
cannot be zero zero
Must be between Missing signal time delay must be set Enter missing signal time delay within 0.10 - 999 seconds
0.10 - 999 for 0.10 - 999 seconds (sec. 10.4I)
Overrun Volume Overrun alarm volumetric setting Enter overrun alarm volumetric setpoint within 0.00100 -
cannot be zero must be within 0.00100 - 999999. 999999.) engineering units (sec. 10.4H)
engineering units
Power Failure Instrument power interrupted during Use RIGHT arrow key to select “Resume” or “Clear”.
Resume Clear active batch process Selecting resume will continue the batch from the point where
the power interruption occurred. Selecting clear will clear the
power failure. After your selection has been
made, press ENTER to continue.
Setup Read Error Power fault occurred while saving Press any key to reload factory defaults then reprogram flow
Press any Key setup menu entry system setup parameters. Note: Totalizer displays remain
undistrubed and will resume after instrument reset.
Dial Selection
• The analog display shows percentage of batch complete.
• The reversible dial scales are 0 to100%, or 100 to 0%. 0 -
100 ALAR
M 1
%
• The instrument is shipped with the 0 to 100% scale installed. Ba
• Reverse the face plate for either scale. 100 tch ALAR
D
M 2
C
A B
ALAR
M1
%
Ba
0 tch ALAR
M 2
100
ENTE
R
3-5600.090-1 (E-8/04) English © George Fischer Signet, Inc. 2001 Printed in U.S.A. on recycled paper