Professional Documents
Culture Documents
3, MARCH 2021
Abstract— Radar-based microwave imaging (MWI) systems feasibility of MWI in early breast cancer detection. As pre-
have shown encouraging results in early detection of breast sented there, two main approaches in MWI are microwave
cancer; however, there exist remaining challenges. Resolution is tomography and radar-based imaging. The basic principle
one of the most important issues not being addressed with clear
theoretical relations in medical MWI systems. In this analytical behind these approaches is the difference between the elec-
study, we thoroughly clarify the relation of resolution to the most trical properties of malignant and normal breast tissues.
fundamental aspects of an MWI system, namely, limited-view Microwave tomography methods involve solving a nonlinear
versus full-view array geometry, monostatic versus multistatic and ill-posed inverse scattering problem in order to retrieve
configuration, single-frequency versus wideband operation, and the electrical properties of the breast. Meaney et al. [11] and
near-field versus far-field imaging. This goal is achieved by
expressing the inverse scattering problem in the spatial frequency Halter et al. [12] from Dartmouth College, Hanover, NH, USA
domain and utilizing K-space representation. Besides, to analyze have studied microwave tomography breast imaging since the
the MWI system more accurately, the sidelobe level (SLL) of 1990s and conducted extensive patient studies with prototype
its point-spread function (PSF) is taken into account. We show systems [11], [12].
that generally, the resolution limits of a multistatic configuration In radar-based MWI, the goal is not to reconstruct the
are the same as its monostatic counterpart and, hence, have
no superiority in terms of resolution. In addition, in imaging electrical profile of the tissues but to detect and determine
with a full-view array, the signal bandwidth has no effect on the position of highly scattering objects (tumors). Radar-
the resolution. Thus, the single-frequency operation provides the based imaging algorithm for breast cancer detection was first
same resolution as the wideband operation. We also demonstrate proposed by Hagness et al. [8] in 1998. A prototype system
that the effect of using multistatic configuration or wideband for mono-static radar-based imaging has been implemented
operation is to reduce the SLL of the PSF and, hence, improve
the image quality. at the University of Calgary, Calgary, AB, Canada [13], and
Index Terms— Breast cancer detection, microwave imaging the resultant images of scanning eight patients are reported
(MWI), radar-based image reconstruction, resolution, synthetic in [6]. Klemm et al. [7] and Byrne et al. [14] have developed
aperture radar (SAR). multistatic radar-based imaging systems at the University of
Bristol, Bristol, U.K., and presented encouraging results and
I. I NTRODUCTION also challenges of patient studies in [15]. Although pioneering
Authorized licensed use limited to: INDIAN INSTITUTE OF TECHNOLOGY ROORKEE. Downloaded on September 12,2021 at 09:06:52 UTC from IEEE Xplore. Restrictions apply.
NAGHIBI AND ATTARI: NEAR-FIELD RADAR-BASED MWI FOR BREAST CANCER DETECTION 1671
Authorized licensed use limited to: INDIAN INSTITUTE OF TECHNOLOGY ROORKEE. Downloaded on September 12,2021 at 09:06:52 UTC from IEEE Xplore. Restrictions apply.
1672 IEEE TRANSACTIONS ON ANTENNAS AND PROPAGATION, VOL. 69, NO. 3, MARCH 2021
where ρ, ρ t , and ρ r are the position vectors of the target, appendix of [28], one can derive
transmitter, and receiver. O(ρ) is the contrast function of the
target defined as O(ρ) = k 2 (εr (ρ) − 1), where k is the F1D {ex p(− j krt )} = ex p − j k yt D − j k yt y − j k xt x
wavenumber in the propagation medium, εr is the relative F1D {ex p(− j krr )} = ex p − j k yr D − j k yr y − j k xr x . (9)
complex permittivity of the object, E is the total electric
Substituting (9) into (4) and after some algebraic arrange-
field inside the target, and G is the Green’s function for the
ments, the spatial Fourier transform of the received scattered
background medium.
field E s can be expressed as
Under the first-order Born approximation, the total electric
field inside the target E can be approximated by the incident
E s (k xt , k xr , k) = ex p − j k yt + k yr D · O(x, y)
field, E i . Assume that the transmitters are 2-D line sources so x y
that the incident field and the Green’s function are of the form · ex p − j (k xt +k xr )x − j k yt +k yr y · d x d y.
of Hankel function and with far-field approximation, we have
(10)
E i ρ t , ρ, k ∼= ex p − j k|ρ t − ρ| / |ρ t − ρ| Now, by defining k x = k xt + k xr and k y = k yt + k yr , (10)
G ρ, ρ , k ∼
r = ex p − j k|ρ − ρ| / |ρ − ρ|.
r (2) r represents a 2-D Fourier transform of the contrast function of
the object
For image reconstruction via beamforming techniques that
synthetically focus the backscattered signals, the phase infor-
E s (k xt , k xr , k) = ex p − j k y D · O(x, y)
mation plays the major role. Therefore, for the sake of x y
simplicity and without loss of generality, the magnitude factors · ex p − j k x x − j k y y · d x d y
are dropped hereinafter. Hence, considering that (x t , yt ) and
= ex p − j k y D · F2D {O(x, y)}. (11)
(x r , yr ) are transmitters and receivers positions, respectively,
and yt = yr = −D, (1) in the Cartesian coordinate system is The above equation shows that the spatial Fourier transform
now rewritten as of the received scattered field can be used to determine the
spatial spectrum of O(x, y). This spatial spectrum determines
E s (x t , xr , k) = O(x, y) also the achievable resolution of the image. Although Fourier
x y
imaging is not utilized in this study, the K-space representation
· ex p − j k (x t − x)2 + (−D−y)2 of E s is beneficial to find out the possible image resolution
of the system and its dependence to the bandwidth, array
geometry and array configuration.
· ex p − j k (xr − x)2 + (−D−y)2
· d x d y. (3) B. K-Space Representation
By taking the 2-D Fourier transform of both sides of (3) and To better understand the effect of system bandwidth and
gaining insight into the K-space or spatial frequency domain configuration on the K-space sampling and spatial resolution,
representation of the image, we would have a better under- we start with a multistatic and single-frequency linear array,
standing of imaging properties like the image resolution. Let then move on to the multiple-frequency (or wideband) array
E s be the spatial Fourier transform of the received scattered and also its monostatic counterpart.
field E s , then we have 1) Imaging With a Multistatic Linear Array: First, consider
an infinite multistatic linear array that works at a single
2D{E s (x t , xr , k)}
E s (k xt , k xr , k) = F frequency. We know that
= O(x, y) · F1D {ex p(− j krt )} 2
k xt + k 2yt = k 2 , k xr
2
+ k 2yr = k 2 . (12)
x y
· F1D {ex p(− j krr )} · d x d y (4) Hence, using the previously mentioned definitions of
k x and k y , we have
in which
2
(k x − k xt )2 + k y − k yt = k 2 . (13)
rt = (x t − x)2 + (−D−y)2 (5)
Equation (13) shows the locus of circles in K-space with a
rr = (xr − x)2 + (−D−y)2 (6) constant radius of k and variable centers of (k xt , k yt ). Also,
according to (12), the center points of (k xt , k yt ) locate on a
F1D {ex p(− j krt )} = ex p(− j krt ) · ex p(− j k xt x t ) · d x t
xt circle of the same radius k around the origin. Geometry of the
(7) problem in Fig. 1 shows that k yt and k yr , and consequently
k y , are always positive in this system, so only semicircles
F1D {ex p(− j krr )} = ex p(− j krr ) · ex p(− j k xr xr ) · d xr . with positive k y can be allocated. Fig. 2 depicts the area that
xr
(8) an infinite multistatic linear array supports in K-space. Each
transmitter takes a sample on the central semicircle and then
The Fourier integrals in (7) and (8) can be derived with receivers, sensing the backscattered signal of each transmitter,
the stationary phase method. Similar to the procedure in the collect samples on a new semicircle.
Authorized licensed use limited to: INDIAN INSTITUTE OF TECHNOLOGY ROORKEE. Downloaded on September 12,2021 at 09:06:52 UTC from IEEE Xplore. Restrictions apply.
NAGHIBI AND ATTARI: NEAR-FIELD RADAR-BASED MWI FOR BREAST CANCER DETECTION 1673
Authorized licensed use limited to: INDIAN INSTITUTE OF TECHNOLOGY ROORKEE. Downloaded on September 12,2021 at 09:06:52 UTC from IEEE Xplore. Restrictions apply.
1674 IEEE TRANSACTIONS ON ANTENNAS AND PROPAGATION, VOL. 69, NO. 3, MARCH 2021
Authorized licensed use limited to: INDIAN INSTITUTE OF TECHNOLOGY ROORKEE. Downloaded on September 12,2021 at 09:06:52 UTC from IEEE Xplore. Restrictions apply.
NAGHIBI AND ATTARI: NEAR-FIELD RADAR-BASED MWI FOR BREAST CANCER DETECTION 1675
Fig. 6. Point target detection by a linear array of length L = 10 cm. (a) Multistatic with a bandwidth of 1–10 GHz. (b) Multistatic at a single
frequency of 10 GHz. (c) Monostatic at a single frequency of 10 GHz. (d) Monostatic with a bandwidth of 1–10 GHz. The point target position is
(0, 4 cm), (0, 0), and (0, −10 cm) in order from top to bottom.
TABLE I
R ANGE R ESOLUTION OF I MAGING W ITH A L IMITED -V IEW A RRAY
Authorized licensed use limited to: INDIAN INSTITUTE OF TECHNOLOGY ROORKEE. Downloaded on September 12,2021 at 09:06:52 UTC from IEEE Xplore. Restrictions apply.
1676 IEEE TRANSACTIONS ON ANTENNAS AND PROPAGATION, VOL. 69, NO. 3, MARCH 2021
Fig. 8. Schematic of imaging problem with a square array of side length 2a B. K-Space Representation
and an object of area S.
It would be interesting to see the effect of imaging with
a full-view array on the K-space sampling and resolution.
We show that the possible achievable resolution in both
the linear array discussed before, we consider a square array
directions is approximately constant, neither depending on the
consisting of four linear arrays, as shown in Fig. 8. This choice
bandwidth nor relating to the distance (near field/far field).
does not question the generality of the result, and it can be
Thus, equation δ = v/2 f is not valid here, even at the far
easily extended to circular or other encircling-structure arrays.
field.
1) Imaging With a Multistatic Square Array: For the single-
A. Theory frequency operation, similar to the procedure of the linear
The most significant difference of a square and a linear array array, we reach to the equation of (k x − k xt )2 + (k y −
is that the former provides a full view of the target, while the k yt )2 = k 2 . However, having a full-view of the target,
latter limits the viewing angles. Consequently, a square array i.e., allowing for both positive and negative values of k x and
takes samples in K-space more widely rather than a linear k y , we have the ability to sample on the whole circles. Fig. 9(a)
array. Similar to the linear array, the received scattered field represents the K-space support of a multistatic square array in
in the Cartesian coordinate system can be written as single-frequency operation. For a point target inside the array,
each transmitter corresponds a point on the central circle,
E s (x t , yt , xr , yr , k) = O(x, y) shown with green color, and then, receivers collect samples
x y
on circles which their centers are placed on the central circle
· ex p − j k (x t − x)2 + (yt −y)2 corresponding to the transmitter.
Note that while adding a bandwidth to the multistatic linear
· ex p − j k (xr − x) + (yr −y)
2 2 array could expand the K-space sampling area, in this scenario,
it does not affect the K-space support significantly, if k max
· dx dy (22) in wideband operation is not higher than the value of k in
single frequency operation. It means that adding a bandwidth
in which (x t , yt ) and (xr , yr ) are transmitters’ and receivers’
to a multistatic full-view array imaging system for the sake of
positions, respectively. Taking 2-D Fourier transform of both
resolution enhancement might not be logical.
sides of (22) and applying stationary phase method, we have
2) Imaging With a Monostatic Square Array: Again, (15)
E s (k xt , k xr , k) describes the locus of sampling points in K-space with a
= F2D {E s (x t , yt , xr , yr , k)} single-frequency and monostatic square array, which is a fixed
circle with a constant radius of 2k centered at the origin,
= O(x, y) as shown in Fig. 9(b). For a point target inside the array,
x y
each transmitter corresponds a point on the inner central circle,
· F1D {ex p(− j krt )} · F1D {ex p(− j krr )} · d x d y
shown with green color, and then, the receiver antenna placed
= 4 exp −jk yt a + exp(− j k xt a) at the same direction of the transmitter takes a sample on the
· exp −jk yr a + exp(− j k xr a) outer circle in the K-space.
Now, by adding a bandwidth to the system, various circles
· O(x, y) · ex p − j (k xt +k xr )x − j k yt +k yr y ·d x d y on the K-space can be sampled, allowing for sampling in an
x y
area. Fig. 9(c) depicts the K-space support of a wideband
(23)
monostatic square array that now can be comparable with
length (Fig. 8), rt =
in which 2a is the square array side Fig. 9(a). It clearly explains why the performance of a single-
(x t − x)2 + (yt −y)2 , and rr = (xr − x)2 + (yr − y)2 . frequency monostatic imaging system cannot compete with
Authorized licensed use limited to: INDIAN INSTITUTE OF TECHNOLOGY ROORKEE. Downloaded on September 12,2021 at 09:06:52 UTC from IEEE Xplore. Restrictions apply.
NAGHIBI AND ATTARI: NEAR-FIELD RADAR-BASED MWI FOR BREAST CANCER DETECTION 1677
Fig. 9. K-space support of (a) multistatic square array with single-frequency operation and (b) and (c) monostatic square array with single-frequency operation
and wideband operation, respectively.
C. Spatial Resolution
For any point inside the array, and according to (16), we can
approximately define the resolutions of both the multistatic and
monostatic along the x- and y-axes as
k xmin = k xt
min
+ k xr
min
= −2k max
(25)
k xmax = k xt
max
+ k xr
max
= 2k max
k min
y = k min
yt + k yr = −2k
min max
Fig. 10(c) shows that a monostatic full-view array operating
(26)
k max
y = k max
yt + k yr = 2k
max max
at a single frequency can effectively detect a target anywhere
2π v in between. It means that the system resolution in any direction
δx = δ y = max = . (27) is not restricted to having a bandwidth, while it was restricted
4k 4 f max
for the far-field imaging with a limited-view array [Fig. 6(c)].
According to the symmetry of K-space in Fig. 9, one can Adding a bandwidth of 1–10 GHz to the signal [Fig. 10(d)]
conclude that (27) is valid for the resolution of any radial or turning to a multistatic configuration [Fig. 10(b)] does not
direction. This equation proves that—regardless of monostatic improve the resolution; however, it results in an image with
or multistatic configurations, far-field or near-field imaging, lower sidelobes. In Section IV, we discuss the reason behind
and single-frequency or wideband operation, the possible this behavior.
resolution is about λ/4, and it does not show any dependency Comparing Fig. 10(a) and (b) implies that adding a band-
on bandwidth or distance. width to a multistatic full-view array does not necessarily
Equation (27) confirms the possibility of having a reason- improve the resolution and image quality (as expected from
able resolution in full-view radar imaging with zero band- K-space representation). Thus, wideband operation may not
width. It should be emphasized that this capability does not be preferable to a multistatic full-view array, especially that it
owe to the multistatic configuration or even near-field imaging, adds complexity to the imaging system.
but it is due to using the full-view array that enables us to Fig. 11 shows the 1-D amplitude PSF of the point target
sample symmetrically along the axes. positioned at the center of the imaging system, working at a
single frequency of 10 GHz and with a monostatic configu-
D. Simulation Results and Discussion ration. Note that 1-D PSFs at x = 0 or y = 0 cross sections
are the same. The FWHM of PSF in Fig. 11 is approximately
Back-projection algorithm is applied on the scattered signals
equal to the estimate made by (27)
received by the square array shown in Fig. 8. The square array
v
has a side length of 2a = 10 cm and is surrounding a point δx = δ y = = 2.5 mm. (28)
target placed at different positions. The propagation medium 4 f max
has a dielectric constant of 9, resembling a homogeneous To summarize this section, Table II reports the relation of
fatty breast tissue. Fig. 10 depicts the squared reconstructed resolution to different parameters, for a full-view array, which
images using wideband multistatic, single-frequency multi- is valid for both multistatic and monostatic configurations.
static, single-frequency monostatic, and wideband monostatic In contrast to Table I, neither bandwidth nor near-field imaging
arrays. improves the resolution. Note that these results are valid for
Authorized licensed use limited to: INDIAN INSTITUTE OF TECHNOLOGY ROORKEE. Downloaded on September 12,2021 at 09:06:52 UTC from IEEE Xplore. Restrictions apply.
1678 IEEE TRANSACTIONS ON ANTENNAS AND PROPAGATION, VOL. 69, NO. 3, MARCH 2021
Fig. 10. Point target detection by a square array of side length 2a = 10 cm. (a) Multistatic with a bandwidth of 1–10 GHz. (b) Multistatic at a single frequency
of 10 GHz. (c) Monostatic at a single frequency of 10 GHz. (d) Monostatic with a bandwidth of 1–10 GHz. The point target position is (0, 0), (0, −2 cm),
and (2 cm, 2 cm) in order from top to bottom.
Authorized licensed use limited to: INDIAN INSTITUTE OF TECHNOLOGY ROORKEE. Downloaded on September 12,2021 at 09:06:52 UTC from IEEE Xplore. Restrictions apply.
NAGHIBI AND ATTARI: NEAR-FIELD RADAR-BASED MWI FOR BREAST CANCER DETECTION 1679
V. C ONCLUSION
In this article, we clarified the theoretical relations of
image resolution and quality in two possible array geometry
of MWI systems, namely, limited-view and full-view array.
The former was explained with a linear array and the latter
Fig. 12. 1-D amplitude PSFs of the monostatic square array at (a) two with a square array that surrounds the target. Both of these
different frequencies (dotted and dashed lines) versus multiple-frequency
operation (solid line) and (b) single-frequency (dashed line) versus wideband
geometries can be used in breast cancer detection. K-space
operation (solid line). representation was utilized to support the theory and better
demonstrate the effects of three essential features of imaging
systems, which are monostatic versus multistatic configuration,
single-frequency versus wideband operation, and near-field
versus far-field imaging. PSFs of the imaging systems were
evaluated through 2-D numerical simulations, to both confirm
the theoretical results and also better reveal the effects of
studied factors. It is worthy of note that the term “near-field”
in this study relates to the Fresnel zone, not the reactive near-
field; thus, the results presented here hold true as long as
evanescent waves are insignificant.
We demonstrated that in a limited-view array, near-field
imaging enables us to achieve range resolution in single-
frequency operation and enhances the range resolution in
wideband operation. In contrast, a full-view array allows for
having resolution at both the near-field and far-field, regardless
of its bandwidth.
We also concluded that the resolution limits of multista-
tic and monostatic configurations are the same; however,
a single-frequency multistatic configuration outperforms its
monostatic counterpart, as it can sample an area of K-space
instead of a ring or an arc, and thus it smooths the PSF.
On the other hand, wideband operation has the same effect
of smoothing the PSF. Therefore, a monostatic configuration
must uptake a bandwidth to have a desired PSF with low
SLL and to be comparable with a single-frequency multistatic
array.
R EFERENCES
[1] R. Chandra, H. Zhou, I. Balasingham, and R. M. Narayanan, “On
the opportunities and challenges in microwave medical sensing and
Fig. 13. Point target imaging with a full-view array. (a) and (b) Point target at imaging,” IEEE Trans. Biomed. Eng., vol. 62, no. 7, pp. 1667–1683,
the real space, and its Fourier transform. (c) and (d) Ring of K-space samples Jul. 2015.
and its IFT in the real space. (e) and (f) Area of K-space samples and its IFT [2] A. T. Mobashsher, A. Mahmoud, and A. M. Abbosh, “Portable wide-
in the real space. band microwave imaging system for intracranial hemorrhage detection
using improved back-projection algorithm with model of effective head
Fig. 13 demonstrates and compares the 2-D PSFs. As shown permittivity,” Sci. Rep., vol. 6, no. 1, p. 20459, Apr. 2016.
[3] S. Y. Semenov and D. R. Corfield, “Microwave tomography for brain
in Fig. 13(c) and (d), with a single-frequency monostatic imaging: Feasibility assessment for stroke detection,” Int. J. Antennas
full-view array, samples on a ring are collected, and the Propag., vol. 2008, pp. 1–8, May 2008.
Authorized licensed use limited to: INDIAN INSTITUTE OF TECHNOLOGY ROORKEE. Downloaded on September 12,2021 at 09:06:52 UTC from IEEE Xplore. Restrictions apply.
1680 IEEE TRANSACTIONS ON ANTENNAS AND PROPAGATION, VOL. 69, NO. 3, MARCH 2021
[4] S. A. Rezaeieh, A. Abbosh, and Y. Wang, “Wideband unidirectional [24] W. Zhi and F. Chin, “Entropy-based time window for artifact removal in
antenna of folded structure in microwave system for early detection of UWB imaging of breast cancer detection,” IEEE Signal Process. Lett.,
congestive heart failure,” IEEE Trans. Antennas Propag., vol. 62, no. 10, vol. 13, no. 10, pp. 585–588, Oct. 2006.
pp. 5375–5381, Oct. 2014. [25] Y. Chen, E. Gunawan, K. S. Low, S.-C. Wang, Y. Kim, and C. B. Soh,
[5] P. M. Meaney et al., “Clinical microwave tomographic imaging of the “Pulse design for time reversal method as applied to ultrawideband
calcaneus: A first-in-human case study of two subjects,” IEEE Trans. microwave breast cancer detection: A two-dimensional analysis,” IEEE
Biomed. Eng., vol. 59, no. 12, pp. 3304–3313, Dec. 2012. Trans. Antennas Propag., vol. 55, no. 1, pp. 194–204, Jan. 2007.
[6] E. C. Fear, J. Bourqui, C. Curtis, D. Mew, B. Docktor, and C. Romano, [26] W. Zhang and A. Hoorfar, “Two-dimensional through-the-wall radar
“Microwave breast imaging with a monostatic radar-based system: imaging with diffraction tomographic algorithm,” in Proc. IEEE Int.
A study of application to patients,” IEEE Trans. Microw. Theory Techn., Conf. Microw. Technol. Comput. Electromagn., May 2011, pp. 96–99.
vol. 61, no. 5, pp. 2119–2128, May 2013. [27] J. F. Clodfelter and M. W. Hibbard, “Leading and lagging aper-
[7] M. Klemm, I. J. Craddock, J. A. Leendertz, A. Preece, and R. Benjamin, ture for linear ground penetrating radar array,” U.S. Patent 8 040 272,
“Radar-based breast cancer detection using a hemispherical antenna Oct. 18, 2011.
array—Experimental results,” IEEE Trans. Antennas Propag., vol. 57, [28] S. S. A. Ahmed, “Electronic microwave imaging with planar multi-
no. 6, pp. 1692–1704, Jun. 2009. static arrays,” Doctoral Degree, Fac. Eng., Friedrich-Alexander Univ.,
[8] S. C. Hagness, A. Taflove, and J. E. Bridges, “Two-dimensional FDTD Erlangen, Germany, 2013.
analysis of a pulsed microwave confocal system for breast cancer [29] M. Klemm, J. A. Leendertz, D. Gibbins, I. J. Craddock, A. Preece, and
detection: Fixed-focus and antenna-array sensors,” IEEE Trans. Biomed. R. Benjamin, “Microwave radar-based differential breast cancer imaging:
Eng., vol. 45, no. 12, pp. 1470–1479, Dec. 1998. Imaging in homogeneous breast phantoms and low contrast scenarios,”
[9] E. C. Fear, S. C. Hagness, P. M. Meaney, M. Okoniewski, and IEEE Trans. Antennas Propag., vol. 58, no. 7, pp. 2337–2344, Jul. 2010.
M. A. Stuchly, “Enhancing breast tumor detection with near-field [30] A. Naghibi, M. S. Hoseini-Rad, V. Khorashadi-Zadeh, and A. R. Attari,
imaging,” IEEE Microw. Mag., vol. 3, no. 1, pp. 48–56, Mar. 2002. “Skin artifact removal algorithms for radar-based microwave breast
[10] N. Nikolova, “Microwave imaging for breast cancer,” IEEE Microw. cancer detection,” presented at the 26th Iranian Conf. Elect. Eng. (ICEE),
Mag., vol. 12, no. 7, pp. 78–94, Dec. 2011. 2018.
[11] P. M. Meaney, K. D. Paulsen, A. Hartov, and R. K. Crane, [31] M. Yousefnia, A. Ebrahimzadeh, M. Dehmollaian, and A. Madannejad,
“An active microwave imaging system for reconstruction of 2-D electri- “A time-reversal imaging system for breast screening: Theory and
cal property distributions,” IEEE Trans. Biomed. Eng., vol. 42, no. 10, initial phantom results,” IEEE Trans. Biomed. Eng., vol. 65, no. 11,
pp. 1017–1026, Oct. 1995. pp. 2542–2551, Nov. 2018.
[12] R. J. Halter et al., “The correlation of in vivo and ex vivo tissue dielectric [32] R. Solimene, A. Cuccaro, G. Ruvio, D. F. Tapia, and M. O’Halloran,
properties to validate electromagnetic breast imaging: Initial clinical “Beamforming and holography image formation methods: An analytic
experience,” Physiol. Meas., vol. 30, no. 6, pp. S121–S136, Jun. 2009. study,” Opt. Express, vol. 24, no. 8, pp. 9077–9093, 2016.
[13] J. Bourqui, J. M. Sill, and E. C. Fear, “A prototype system for measuring [33] M. Lazebnik et al., “A large-scale study of the ultrawideband microwave
microwave frequency reflections from the breast,” Int. J. Biomed. Imag., dielectric properties of normal breast tissue obtained from reduction
vol. 2012, pp. 1–12, Apr. 2012. surgeries,” Phys. Med. Biol., vol. 52, no. 10, p. 2637, 2007.
[14] D. Byrne, M. Sarafianou, and I. J. Craddock, “Compound radar approach [34] M. Pastorino, Microwave Imaging. Hoboken, NJ, USA: Wiley, 2010.
for breast imaging,” IEEE Trans. Biomed. Eng., vol. 64, no. 1, pp. 40–51,
Jan. 2017.
[15] A. W. Preece, I. Craddock, M. Shere, L. Jones, and H. L. Winton,
“MARIA M4: Clinical evaluation of a prototype ultrawideband radar
scanner for breast cancer detection,” J. Med. Imag., vol. 3, no. 3, Atefeh Naghibi received the B.Sc. and M.Sc.
Jul. 2016, Art. no. 033502. degrees in electrical engineering from the Ferdowsi
[16] A. J. Devaney, “Geophysical diffraction tomography,” IEEE Trans. University of Mashhad, Mashhad, Iran, in 2012 and
Geosci. Remote Sens., vol. GE-22, no. 1, pp. 3–13, Jan. 1984. 2015, respectively, where she is currently pursuing
[17] G. Gennarelli, I. Catapano, F. Soldovieri, and R. Persico, “On the achiev- the Ph.D. degree.
able imaging performance in full 3-D linear inverse scattering,” IEEE Her current research interests include microwave
Trans. Antennas Propag., vol. 63, no. 3, pp. 1150–1155, Mar. 2015. imaging and transformation electromagnetics.
[18] E. C. Fear and M. A. Stuchly, “Microwave detection of breast cancer,”
IEEE Trans. Microw. Theory Techn., vol. 48, no. 11, pp. 1854–1863,
Nov. 2000.
[19] T. Sleasman, M. Boyarsky, M. F. Imani, T. Fromenteze, J. N. Gollub,
and D. R. Smith, “Single-frequency microwave imaging with dynamic
metasurface apertures,” J. Opt. Soc. Amer. B, Opt. Phys., vol. 34, no. 8,
p. 1713, 2017.
[20] A. J. D. Dekker and A. V. D. Bos, “Resolution: A survey,” J. Opt. Soc. Amir Reza Attari was born in Mashhad, Iran,
Amer. A, Opt. Image Sci., vol. 14, no. 3, pp. 547–557, 1997. in 1971. He received the B.S. and M.S. degrees
[21] C. Curtis, “Factors affecting image quality in near-field ultra-wideband in electrical engineering from the Sharif University
radar imaging for biomedical applications,” Ph.D. thesis, Dept. Elect. of Technology, Tehran, Iran, in 1994 and 1996,
Comput. Eng., Univ. Calgary, Calgary, AB, Canada, 2015. respectively, and the joint Ph.D. degree in electrical
[22] E. J. Bond, X. Li, S. C. Hagness, and B. D. Van Veen, “Microwave engineering from the Sharif University of Tech-
imaging via space-time beamforming for early detection of breast nology, and University Joseph Fourier, Grenoble,
cancer,” IEEE Trans. Antennas Propag., vol. 51, no. 8, pp. 1690–1705, France, in 2002.
Aug. 2003. In 2004, he joined the Department of Electri-
[23] E. C. Fear, X. Li, S. C. Hagness, and M. A. Stuchly, “Confocal cal Engineering, Ferdowsi University of Mashhad,
microwave imaging for breast cancer detection: Localization of tumors Mashhad, Iran, where he is currently a Professor.
in three dimensions,” IEEE Trans. Biomed. Eng., vol. 49, no. 8, His research interests are antennas, passive microwave devices, and microwave
pp. 812–822, Aug. 2002. imaging.
Authorized licensed use limited to: INDIAN INSTITUTE OF TECHNOLOGY ROORKEE. Downloaded on September 12,2021 at 09:06:52 UTC from IEEE Xplore. Restrictions apply.