modulation) and an r.f. coil (frequency (see fig. below).
Markem_Imaje_45 Sound Holes MIT 12 AD_GT3L_Man- 10.1.1.619.9544 22120e Cellule Ningbo 211640385-Erbe-
HW EN 12-07… Ant Electronics… VIO300S-Servic…
Topics
– Resonance absorption of an r.f. tuned
circuit
– Electron spin resonance (resonant
frequency as a function of magnetic field
strength, determination of the g -factor)
Equipment list:
1 ESR basic unit 514 550
1 ESR control unit 514 570
1 Pair of Helmholtz coils 555 06 0
2 Multimeters M 2012 531 560
1 Two-channel oscilloscope 575 211
3 Saddle bases 300 110
1 Insulated stand rod 590 130
2 Screened cables BNC, 4 mm 575 2 40
3 Conn. leads, 25 cm 501 230
Resonance frequency f as a function of magnetic Oscilloscope display 5 Conn. leads, 50 cm 501 280
resonant field strength B
428 B
Scope of delivery:
1 ESR probe head (variable high-frequency
transmitter, frequency divider and low-
frequency signal amplifier)
3 Plug-in coils for different frequency
514 57 514 55 on 300 11
ranges
1 Measuring cable, for employing the unit
as a resonance meter
1 Passive electric resonant circuit for inves- Está leyendo una previsualización
tigating the dependence of the resonant
frequency on the magnetic field
1 DPPH-probe (diphenyl-picryl-hydrazyl) Cargue sus documentos para descargar
514 57 ESR control unit 514 56 ESR adapter 55506 Pair of Helmholtz coils
Power supply unit for the RF transmitter of For connection of the ESR basic unit The pair of coils are used to produce a
the ESR basic unit (514 55) and the pair of (514 55) to other power supply units with homogeneous magnetic field of variable
Helmholtz coils (555 06) for experiments on 4 mm sockets and meters with BNC strength in which the deflection of electrons
electron spin resonance; integrated high-fre- sockets; required when the ESR control unit within tubes 555 07–17 can be shown and
quency counter with digital display and (514 57) is not used. investigated, also used with the apparatus
phase shifter between the oscilloscope- A built-in frequency output allows high fre- for electron spin resonance (514 55). For
compatible output voltages, which are quency measurements with conventional use with stand (555 05).
proportional to the RF amplitude or the coil counters (e.g. 575 40 or 575 45). Number of turns per coil: 320
current. Output: 5-pole multiple socket for ESR basic unit Max. continuous current: 1.5 A (U = 10 V)
Power supply for magnetic field: Signal output: BNC socket Max. short-time current: 2 A (U = 15 V)
0 to 10 V DC, Frequency output: BNC socket Average diameter: 13.5 cm
0 to 5 V AC, continuously adjustable Connection: + 1 2 V/0 V/–12 V, via 4 mm sockets Stand rod: 14.5 cm x 8 mm dia.
Current: max 3 A (no overload protection) Dimensions: 9.5 cm x 7.5 cm x 2.5 cm Connection two 4-mm sockets on each coil
Adjustable phase difference: 0 to 90°
Frequency display: 4 decades (MHz) Additionally required:
Connection: DC power supply
115/230 V, 50/60 Hz, via mains cable 0 to 15 V........................................ 521 45
Fuses:
Variable extra low voltage
for 230 V: T 0.8 B
for 115 V; T 1.6 D transformer S.................................... 521 35
Dimensions: 30 cm x 21 cm x 23 cm Counter P ..........................................57545
Weight: approx. 6.2 kg Stopclock...........................................313 07