{Modification/Type1}{Production}{Carestream Health}{Confidential}

Publication No. 8G7613


16APR10
Supersedes 8G7613
27JUL09

Modification and Recovery Instructions


for the
KODAK DIRECTVIEW CR 825/850/950/975 Systems
and
KODAK DIRECTVIEW Classic/Elite CR Systems
Service Codes: 1883, 4825, 4826, 5258, 5259, and 5634
Modification No. 68/13
Type 1 Selective
Purpose: To install, upgrade to, or restore the Software Version 5.2.x for the KODAK DIRECTVIEW CR 825/
850/950/975 Systems and the KODAK DIRECTVIEW Classic/Elite CR Systems.

Important
• Qualified service personnel must do these procedures.
• See the Carestream Health Service Portal for possible updates before doing this modification. Locate the
Software Version 5.2 documents at: Digital Capture>CR and DR Software Version 5.2 and higher. Locate
CR-specific documents at: Digital Capture>CR and DR Software Version 5.2 and higher>Click here for CR
specific SW documents.
• Additional support documents (for example, RMS documents) can be located by using the portal search function
to look for the document part number.
• When doing the procedures in this document, you must use safe work practices and wear the correct personal
protective equipment (for example, safety eyewear) according to your company’s standard operating
procedures.
Service Effects: None
Special See Page 4.
Requirements:
Serial Numbers: All CR 825/850/950/975 Systems and Classic/Elite CR Systems.
Installation Time: • 1. 5 hours - new installation
• 2.5 hours - upgrade from Software Version 5.1
• 2 hours - recovery
Special Tools: None
Parts Status: Available from Service Parts Management.
Parts See Page 25.
Requirements:

Confidential

© CARESTREAM HEALTH, INC. 2010


PLEASE NOTE The information contained herein is based on the experience and knowledge relating to the
subject matter gained by Carestream Health, Inc. prior to publication.
No patent license is granted by this information.
Carestream Health, Inc. reserves the right to change this information without notice, and makes
no warranty, express or implied, with respect to this information. Carestream Health shall not be
liable for any loss or damage, including consequential or special damages, resulting from any
use of this information, even if loss or damage is caused by Carestream Health’s negligence or
other fault.
This equipment includes parts and assemblies sensitive to damage from electrostatic
discharge. Use caution to prevent damage during all service procedures.

Table of Contents
Description Page

Preparing for the Modification or Recovery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4


Special Requirements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Obtaining the Support Documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Using this Document. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Doing MOD 68 from the Software Version 4.x . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Doing MOD 68/13 for a System with Software Version 5.1.x . . . . . . . . . . . . . . . . . . . . . . 10
Doing the Configuration for the New CR 825/850/975 Systems . . . . . . . . . . . . . . . . . . . . 15
Doing the Configuration for the New Classic/Elite CR Systems . . . . . . . . . . . . . . . . . . . . 17
Doing a Software Recovery for All CR Systems with Software Version 5.2.x . . . . . . . . . . 20
Detailed Procedures for All CR Systems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Checking the Modification and Software Installation Kit . . . . . . . . . . . . . . . . . . . . . . 25
Preparing for the Software Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Checking for the Necessary Hardware . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Making a Backup Diskette for the Software Version 4.x . . . . . . . . . . . . . . . . . . . . . . . 27
Checking the Connections to the Computer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Installing the New License Sticker . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Recording the Network Configuration Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Storing the MCB Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Storing the User Preferences and MIM Destinations . . . . . . . . . . . . . . . . . . . . . . . . . 33
Extracting the Options from a System with Software Prior to Software Version 5.2. . 35
Doing the System Shutdown. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Checking and Setting the System BIOS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Setting the Boot Order . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Installing the Software Version 5.2.x. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
For Model 8810 Computers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
For Model 8212 Computers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
For HP 5700 Computers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Energize and Log in to the CR System. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Doing the Calibration for the Touch Screen Monitor. . . . . . . . . . . . . . . . . . . . . . . . . . 42
Entering the System Information for the Option Registration . . . . . . . . . . . . . . . . . . . 43
Enabling the Options for an Upgrade from Legacy Software . . . . . . . . . . . . . . . . . . . 45
Enabling the Options from the Certificates . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Setting the Network Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Setting the Configuration for the Scanner on the Classic/Elite CR Systems . . . . . . . 49
Setting the Regional Setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Restoring the User Preferences . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Setting the Print Destinations to the MIM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Setting the Store Destinations to the MIM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57

2 16APR10 – 8G7613
Setting the HIS/RIS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Setting the Profile Destination Configuration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Downloading the Software to the Boards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Downloading the Software for CR 825M, 850M or 975 Systems . . . . . . . . . . . . . 60
Downloading the Software for CR 825/850/950 Systems . . . . . . . . . . . . . . . . . . 62
Downloading the Software for Classic/Elite CR Systems . . . . . . . . . . . . . . . . . . 64
Downgrading to Software Version 5.1 for the Classic/Elite CR Systems . . . . . . . . . . 66
Checking the Configuration for the Bar-Code Scanner . . . . . . . . . . . . . . . . . . . . . . . . 68
Setting the Other Preferences for the Bar-Code Scanner. . . . . . . . . . . . . . . . . . . . . . 68
Country Code . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
Code 39 Full ASCII Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Automatic Trigger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Beeper Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
Reread Delay . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
Doing the Setup for the 11-Digit Cassette ID . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
Doing the Calibration for the Luminance of the Monitor . . . . . . . . . . . . . . . . . . . . . . . 73
Setting the Look Code . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
Doing the Setup for Remote Management Services (RMS) . . . . . . . . . . . . . . . . . . . . 75
Installing the KODAK DIRECTVIEW Remote Operations Panel (ROP) . . . . . . . . . . . 75
Installing the Other Remote Devices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
Setting the Slow-Scan Correction Profile for the EHR Cassettes . . . . . . . . . . . . . . . . 76
Completing the Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
Doing the Feedback for a Customer-Purchased Modification
and the Software Refresh Option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
Publication History . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79

8G7613 – 16APR10 3
MODIFICATION INSTRUCTIONS, TYPE 1

Section 1: Preparing for the Modification or Recovery

Special Requirements
• You must have KODAK Secure Link Client Software Version 2.8 installed on the laptop or other PC that you use
for Secure Link.
• Review the Capture Link Compatibility Matrix on the Carestream Health Service Portal at: Digital
Capture>Capture Link for the latest version of Capture Link that is compatible with Software Version 5.2. This
latest Capture Link must be acquired as a separate kit.
• Remote Access Software (RAS) Version 2.2.6.x and Remote Operations Panel (ROP) Version 2.3.3.5 are the
RAS and ROP versions that are compatible with Software Version 5.2. These versions are included with the
Software Version 5.2 software kits.
• You must have access to the support publications listed in Obtaining the Support Documents.

Obtaining the Support Documents

Important
• Check the Carestream Health Service Portal for possible updates to the support documents listed below before
doing this modification. Locate the Software Version 5.2 documents at: Digital Capture>CR and DR Software
Version 5.2 and higher. Locate CR-specific documents at: Digital Capture>CR and DR Software Version 5.2
and higher>Click here for CR specific SW documents.
• Additional support documents (for example, RMS documents) can be located by using the portal search function
to look for the document part number.

You must have the following support documents available to perform this modification:
• Customer Site Survey for the KODAK Remote Management Services, 7F0966, Rev. H
• Site Information Sheet for the KODAK DIRECTVIEW CR Systems, 6H6825
• FE Checklist, 6F5825
• Modification Instructions and Software Installation Instructions for the Remote Management Services for the
Software Version 5.1, 7H2373
• Service Bulletin, SB8H6204
• Service Guide for the Barcode Reader for all CR/DR Systems and CAD, 7H7713
• Remote Access Software Installation Instructions, 6H6955
• Software Installation Instructions for the 19 in. KODAK DIRECTVIEW Remote Operations Panel (ROP) used on
the KODAK DIRECTVIEW CR 825/850/950/975 Systems and the KODAK DIRECTVIEW Classic/Elite CR
Systems with Software Version 5.2, 8G7963
• Modification Instructions and Software Installation Instructions for Software Version 4.60.17 CP4 for all KODAK
DIRECTVIEW CR Systems 7H2690
• Modification Instructions for the KODAK DIRECTVIEW CR 825/850/950/975 Systems, KODAK DIRECTVIEW
Classic/Elite CR Systems, and the KODAK DIRECTVIEW WAIV Console PC with Software Version 5.2.00.13/
.16/.20/.25/.30, 8J2555

4 16APR10 – 8G7613
Preparing for the Modification or Recovery

Using this Document

Important
Do the procedures in this document in the order listed.

Sections 2–8 of this document are divided into the following:


• Section 2 - Doing MOD 68 from the Software Version 4.x on Page 7.
• Section 3 - Doing MOD 68/13 for a System with Software Version 5.1.x on Page 10.
• Section 4 - Doing the Configuration for the New Classic/Elite CR Systems on Page 17.
• Section 5 - Doing the Configuration for the New CR 825/850/975 Systems on Page 15.
• Section 6 - Doing a Software Recovery for All CR Systems with Software Version 5.2.x on Page 20.
• Section 7 - Detailed Procedures for All CR Systems on Page 25.
• Section 8 - Publication History on Page 79.
The steps in Sections 2–6 are listed with the name of the required procedure in the order the procedure is to be
performed. Links are provided from each of these sections to the detailed procedures required in Section 7 together
with instructions for returning to the page where you left off in the applicable section.
Section 8 identifies the publication history for this document.

Important
If you are using your laptop to read these instructions, when you complete a procedure in Section 7, you will be
instructed to click the Previous View button on the screen until you return to the page in the Section 2–6 where you
left off.

Toolbars

Page Navigation

The Previous View button is not a standard back-arrow button. To display the Previous View button, select
View>Toolbars and click a check mark next to Page Navigation.

8G7613 – 16APR10 5
MODIFICATION INSTRUCTIONS, TYPE 1

Previous View button - older ADOBE Readers

Previous View button - newer ADOBE Readers

The Previous View button is green in older versions of ADOBE Reader software, and blue in newer versions.

Page Navigation

If the Page Navigation toolbar is enabled and the Previous View button is not displayed, right-click on the Page
Navigation toolbar and click Previous View.

6 16APR10 – 8G7613
Doing MOD 68 from the Software Version 4.x

Section 2: Doing MOD 68 from the Software Version 4.x

Important
• You must do the procedures in this section for MOD 68 in the order listed.
• For detailed instructions on how to perform the procedures referenced here, see Section 7.

[1] Do Checking the Modification and Software Installation Kit on Page 25 for the corresponding CR System.
[2] Do Preparing for the Software Installation on Page 25 to:
• Disconnect the bar-code scanner.
• Connect:
– keyboard
– mouse
[3] Do Checking for the Necessary Hardware on Page 26.

Important
The backup diskette for Software Version 4.x is only used for reference during the installation of this modification.
[4] Do Making a Backup Diskette for the Software Version 4.x on Page 27.
[5] Record the settings from Step 4 on the Site Information Sheet.
[6] Do Extracting the Options from a System with Software Prior to Software Version 5.2 on Page 35.

Caution
Dangerous Voltage
[7] De-energize the CR System.
[8] Do Checking the Connections to the Computer on Page 27.
[9] Do Installing the New License Sticker on Page 29.
[10] Energize the CR System.

Important
Do NOT de-energize the system when you complete Step 11 because you will need to set additional BIOS settings
after this step.
[11] Do Checking and Setting the System BIOS on Page 36.
[12] Do Installing the Software Version 5.2.x on Page 38.
[13] Do Energize and Log in to the CR System on Page 41.
[14] Do Doing the Calibration for the Touch Screen Monitor on Page 42.
[15] If the network cable is not connected, connect it now.
[16] Do Setting the Network Configuration on Page 48.

Note
• If you change any of the network settings, you must do an immediate reboot of the CR System.
• After the system reboots. you must log on to the Service Menu again.
[17] Do Entering the System Information for the Option Registration on Page 43.

8G7613 – 16APR10 7
MODIFICATION INSTRUCTIONS, TYPE 1

[18] Do you have an Options Conversion file from Step 6?


Yes No
a. Do Enabling the Options for an Upgrade from Legacy a. Do Enabling the Options from the
Software on Page 45. Certificates on Page 46.
Note: In addition to the converter file for a purchased b. Continue with Step 19.
upgrade from Software Version 4.x to 5.2.x, you will
need the Capture Console Software and Option
Version 5.2 option certificates that came with the
Software Version 4.x to 5.2.x upgrade kit.
b. Enable the new feature option. Use the new feature
certificates and Enabling the Options from the
Certificates on Page 46.
c. Continue with Step 19.

[19] Do Setting the Regional Setting on Page 51.


[20] Do Setting the Print Destinations to the MIM on Page 55.
[21] Do Setting the Store Destinations to the MIM on Page 57.
[22] Do Setting the HIS/RIS on Page 58.
[23] Do Setting the Profile Destination Configuration on Page 59.

Important
• The customer is responsible for providing a Security Administrator (SA) and creating all the Key Operator and
User Accounts.
• If the user names were restored from a backup, the:
– user name is restored.
– passwords are reset to the defaults.
• You must contact the SA to have the operator reset the passwords.
[24] Do Downloading the Software to the Boards on Page 60.
[25] Disconnect the keyboard.
[26] Connect the bar-code reader.

Important
To do the configuration for the bar-code reader, see the Service Guide for the Barcode Reader for all CR/DR Systems
and CAD, 7H7713.
[27] Do the configuration for the bar-code reader.
[28] Do Checking the Configuration for the Bar-Code Scanner on Page 68 for the correct operation.
[29] Is the operation of the bar-code scanner correct?
Yes No
Continue with Step 30. a. Do Step 27–Step 29 again.
b. If the operation of the bar-code scanner is not correct, call the TSC.
c. Continue with Step 30.

[30] Do Setting the Other Preferences for the Bar-Code Scanner on Page 68.
[31] If necessary, do Doing the Setup for the 11-Digit Cassette ID on Page 71.
[32] Do Doing the Calibration for the Luminance of the Monitor on Page 73.
[33] Do Setting the Look Code on Page 74.

8 16APR10 – 8G7613
Doing MOD 68 from the Software Version 4.x

[34] Do Doing the Setup for Remote Management Services (RMS) on Page 75.
[35] If necessary, do Installing the KODAK DIRECTVIEW Remote Operations Panel (ROP) on Page 75. See the
Software Installation Instructions for the 19 in. KODAK DIRECTVIEW Remote Operations Panel (ROP) used on
the KODAK DIRECTVIEW CR 825/850/950/975 Systems and the KODAK DIRECTVIEW Classic/Elite CR
Systems with Software Version 5.2, 8G7963.

Important
The customer is responsible for installing the Remote Access software. See the Remote Access Software Installation
Instructions, 6H6955.
[36] Do Installing the Other Remote Devices on Page 75.
[37] Do Setting the Slow-Scan Correction Profile for the EHR Cassettes on Page 76.
[38] Do Completing the Installation on Page 77:
• Do the FE Checklist located on the Carestream Health Service Portal.
• Check all the cassettes.
• Check the image quality with the DisplayPro test pattern.
[39] Do Storing the MCB Parameters on Page 32.
[40] Do Storing the User Preferences and MIM Destinations on Page 33.

Important
Before leaving the site, you must:
• Check that the System Lockdown check box on the Service Functions Menu is selected.
• Log off from the Service Menu.
[41] Check that the System Lockdown check box is selected.
[42] Log off from the Service Menu.

Important
You must follow the feedback instructions as described on Page 78.
[43] Do Doing the Feedback for a Customer-Purchased Modification and the Software Refresh Option on Page 78.
[44] Do the “Install the Modification” instructions for Software Version 5.2.00.40 in the publication Modification
Instructions for the KODAK DIRECTVIEW CR 825/850/950/975 Systems, KODAK DIRECTVIEW Classic/Elite
CR Systems, and the KODAK DIRECTVIEW WAIV Console PC with Software Version 5.2.00.13/.16/.20/.25/
.30, 8J2555, as required.

8G7613 – 16APR10 9
MODIFICATION INSTRUCTIONS, TYPE 1

Section 3: Doing MOD 68/13 for a System with Software Version 5.1.x

Important
• You must do the procedures in this section for MOD 68/13 in the order listed.
• For detailed instructions on how to perform the procedures referenced here, see Section 7.
• This procedure is used to upgrade:
– Maintenance content to the current maintenance release.
– To the latest content that can be purchased. The latest feature option must be enabled.
• The same DVDs are used for both maintenance content and purchased content. The Software Version 5.2 option
feature certificate enables the Software Version 5.2 purchased content.

To do an upgrade for a system with Software Version 5.1.0x.50 or 5.1.0x.51 for the HP 5700:
[1] Do Checking the Modification and Software Installation Kit on Page 25 for the corresponding CR System.

Important
If you have a USB bar-code scanner, you do NOT have to disconnect the keyboard to connect the bar-code scanner.
[2] Do the Preparing for the Software Installation on Page 25 to:
• Disconnect:
– bar-code scanner
– KODAK DIRECTVIEW Capture Link System - if the CR System is part of the Capture Link System
• Connect:
– keyboard
– mouse

Important
• Your system operation might not be correct if a connection is not correct.
• Checking your connections might prevent the need for a software recovery.
[3] Do Checking the Connections to the Computer on Page 27.

Important
You must record the Network Configuration Data manually. The data is not stored in any backup location.
[4] Do Recording the Network Configuration Data on Page 30 on the Site Information Sheet.

Important
For all CR Systems with software versions prior to Software Version 5.2, you must run the
“AddOnOptionConverter.exe” program to extract the existing options.
[5] Do:
• Storing the MCB Parameters on Page 32.
• Storing the User Preferences and MIM Destinations on Page 33.
• Extracting the Options from a System with Software Prior to Software Version 5.2 on Page 35.

Caution
Dangerous Voltage
[6] Do Doing the System Shutdown on Page 36.

10 16APR10 – 8G7613
Doing MOD 68/13 for a System with Software Version 5.1.x

[7] Do Installing the Software Version 5.2.x on Page 38.


[8] Do Energize and Log in to the CR System on Page 41.
[9] Do Doing the Calibration for the Touch Screen Monitor on Page 42.
[10] If the network cable is not connected, connect it now.
[11] Do Setting the Network Configuration on Page 48.

Note
If you change any network settings, an immediate reboot is required. After the system reboots, you must log on to
the Service Menu again.
[12] Do you have a Classic/Elite CR System?
Yes No
a. Do Setting the Configuration for the Scanner on the Classic/Elite Continue with Step 13.
CR Systems on Page 49.
b. Continue with Step 13.

[13] Do Entering the System Information for the Option Registration on Page 43.

Important
The scanner must communicate with the software before the options will function correctly.
[14] Check the model number in Weblink. Is it correct?
Yes No
Continue with Step 15. a. Do Downloading the Software to the Boards on Page 60.
b. Continue with Step 15.

Important
• If you are doing a maintenance release upgrade, you must have run the “AddOnOptionConverter.exe” program
in Step 5.
• If this is a purchased upgrade, you must also enable the new feature option. Use the Enabling the Options from
the Certificates on Page 46 and the new feature certificate.
[15] Do you have an “legacyoptions.opt” file from Step 5?
Yes No
a. Do Enabling the Options for an Upgrade from Legacy Software on a. Do Enabling the Options from
Page 45. the Certificates on Page 46.
b. For a purchased upgrade, enable the new feature option. Use the b. Continue with Step 16.
new feature certificates and Enabling the Options from the
Certificates on Page 46.
c. Continue with Step 16.

8G7613 – 16APR10 11
MODIFICATION INSTRUCTIONS, TYPE 1

[16] Do you have a current backup file of the User Preferences on the backup diskette?
Yes
For an upgrade from For an upgrade from
Software Version 5.1.0x.50 <Software Version 5.1.0x.50 No
a. Do:
Important Important • Setting the Regional Setting on
Do not restore: Only restore: Page 51.
• Display Configuration • Procedure Mapping • Setting the Print Destinations to
• Remote Device • Destinations the MIM on Page 55.
Configuration a. Do Restoring the User • Setting the Store Destinations to
the MIM on Page 57.
• Network Settings Preferences on Page 52.
• Setting the HIS/RIS on Page 58.
a. Do Restoring the User b. Continue with Step 17.
Preferences on Page 52. b. Continue with Step 17.
b. Continue with Step 17.

Important
• The customer is responsible for providing a Security Administrator (SA) and creating all the Key Operator and
User Accounts.
• If the user names were restored from a backup, the:
– user name is restored.
– passwords are reset to the defaults.
• You must contact the SA to have the operator reset the passwords.
• If you installed a new circuit board, you must download the software to the board.
[17] Do Downloading the Software to the Boards on Page 60.
[18] Disconnect the keyboard.
[19] Connect the bar-code reader.

Important
To do the configuration for the bar-code reader, you must use the Service Guide for the Barcode Reader for all CR/
DR Systems and CAD, 7H7713.
[20] Do the configuration for the bar-code reader.
[21] Do Checking the Configuration for the Bar-Code Scanner for the correct operation.
[22] Is the operation of the bar-code scanner correct?
Yes No
Continue with Step 23. a. Do Step 20–Step 22 again.
b. If the operation of the bar-code scanner is not correct, call the TSC.
c. Continue with Step 23.

[23] Do Setting the Other Preferences for the Bar-Code Scanner on Page 68.
[24] If necessary, do Doing the Setup for the 11-Digit Cassette ID on Page 71.

12 16APR10 – 8G7613
Doing MOD 68/13 for a System with Software Version 5.1.x

Important
• The Look Code is installed during the installation of MOD 50 and when the system is installed for the first time.
• You only have to do the installation for the Look Code if there is not a current backup on the user preference
diskette.
[25] If necessary, do Setting the Look Code on Page 74.
[26] Do Doing the Calibration for the Luminance of the Monitor on Page 73.
[27] Do Doing the Setup for Remote Management Services (RMS) on Page 75 to restore the settings for the Remote
Management System.
[28] If necessary, do Installing the KODAK DIRECTVIEW Remote Operations Panel (ROP) on Page 75. Do the
Software Installation Instructions for the 19 in. KODAK DIRECTVIEW Remote Operations Panel (ROP) used on
the KODAK DIRECTVIEW CR 825/850/950/975 Systems and the KODAK DIRECTVIEW Classic/Elite CR
Systems with Software Version 5.2, 8G7963.

Important
The customer is responsible for installing the Remote Access software. See the Remote Access Software Installation
Instructions, 6H6955.
[29] Do Installing the Other Remote Devices on Page 75.

Important
If you installed a new slow scan board, you must set the Slow Scan Corrections Profile.
[30] vSetting the Slow-Scan Correction Profile for the EHR Cassettes on Page 76, if you installed a new slow scan
board.
[31] Do Completing the Installation on Page 77 for the procedures to:
• Do the FE Checklist located on the Carestream Health Service Portal.
• Check all the cassettes.
• Check the image quality with the DisplayPro test pattern.

Important
If the data has changed since the last restore, you must store the new data.
[32] Do:
• Storing the MCB Parameters on Page 32.
• Storing the User Preferences and MIM Destinations on Page 33.

Important
Before leaving the site, you must:
• Check that the System Lockdown check box on the Service Functions Menu is selected.
• Log off from the Service Menu.
[33] Check that the System Lockdown check box is selected.
[34] Log off from the Service Functions Menu.

8G7613 – 16APR10 13
MODIFICATION INSTRUCTIONS, TYPE 1

[35] Record the information for feedback:


(a) For a Maintenance Release, use the table below to SCAN or CRM.
SCAN
Service Activity Installation Install
K-number System Code Serial No. Code CRM PSL Time (actual) Part No. Feedback
CR 825 5634 M68 SC5634M68 --- No
CR 850 4825 M68 SC4825M68 --- No
CR 950 4826 M68 SC4826M68 --- No
CR 975 1883 M68 SC1883M68 --- No
Classic CR 5258 M13 SC5258M13 --- No
Elite CR 5259 M13 SC5259M13 --- No

(b) For a Customer-Purchased Modification, use the procedure Doing the Feedback for a Customer-
Purchased Modification and the Software Refresh Option on Page 78.
[36] Do the “Install the Modification” instructions for Software Version 5.2.00.40 in the publication Modification
Instructions for the KODAK DIRECTVIEW CR 825/850/950/975 Systems, KODAK DIRECTVIEW Classic/Elite
CR Systems, and the KODAK DIRECTVIEW WAIV Console PC with Software Version 5.2.00.13/.16/.20/.25/
.30, 8J2555, as required.

14 16APR10 – 8G7613
Doing the Configuration for the New CR 825/850/975 Systems

Section 4: Doing the Configuration for the New CR 825/850/975 Systems

Important
• You must do the procedures in this section for new CR 825/850/875 Systems in the order listed.
• For detailed instructions on how to perform the procedures referenced here, see Section 7.

To do the configuration for the new CR System:


[1] Do Preparing for the Software Installation on Page 25 to connect the:
• keyboard
• mouse

Important
• Some connections to the computer might loosen during shipment from the factory.
• You must check the computer connections.
[2] Do Checking the Connections to the Computer on Page 27.
[3] Do Energize and Log in to the CR System on Page 41.
[4] Do Doing the Calibration for the Touch Screen Monitor on Page 42.
[5] If the network cable is not connected, connect it now.
[6] Do Setting the Network Configuration on Page 48.

Note
If you change any network settings, you must do an immediate reboot. After the system reboots, you must log on to
the Service Menu again.
[7] Do Entering the System Information for the Option Registration on Page 43.
[8] Do Enabling the Options from the Certificates on Page 46.
[9] Do Setting the Regional Setting on Page 51.
[10] Do Setting the Print Destinations to the MIM on Page 55.
[11] Do Setting the Store Destinations to the MIM on Page 57.
[12] Do Setting the HIS/RIS on Page 58.

Important
• The customer is responsible for providing a Security Administrator (SA) and creating all the Key Operator and
User Accounts.
• For the new CR Systems, you must check for new software to download to the board.
[13] Do Downloading the Software to the Boards on Page 60.
[14] Do Checking the Configuration for the Bar-Code Scanner on Page 68 to check for the correct operation.
[15] Is the operation of the bar-code scanner correct?
Yes No
Advance to Step 19. Continue with Step 16.

Important
To do the configuration for the bar-code reader, you must use the Service Guide for the Barcode Reader for all CR/
DR Systems and CAD, 7H7713.
[16] Do the configuration for the bar-code reader.
[17] Do Checking the Configuration for the Bar-Code Scanner to check for the correct operation.

8G7613 – 16APR10 15
MODIFICATION INSTRUCTIONS, TYPE 1

[18] Is the operation of the bar-code scanner correct?


Yes No
Continue with Step 19. a. Do Step 16–Step 18 again.
b. If the operation of the bar-code scanner is not correct, call the TSC.
c. Continue with Step 19.

[19] Do Setting the Other Preferences for the Bar-Code Scanner on Page 68.
[20] If necessary, do Doing the Setup for the 11-Digit Cassette ID on Page 71.
[21] Do Doing the Calibration for the Luminance of the Monitor on Page 73.
[22] Do Setting the Look Code on Page 74.
[23] Do Doing the Setup for Remote Management Services (RMS) on Page 75.
[24] Do Installing the KODAK DIRECTVIEW Remote Operations Panel (ROP) on Page 75.

Important
The customer is responsible for installing the Remote Access software. See the Remote Access Software Installation
Instructions, 6H6955.
[25] Do Installing the Other Remote Devices on Page 75.
[26] Do Completing the Installation on Page 77 for the procedures to:
• Do the FE Checklist located on the Carestream Health Service Portal.
• Check all the cassettes.
• Check the image quality with the DisplayPro test pattern.
[27] Do:
• Storing the MCB Parameters on Page 32.
• Storing the User Preferences and MIM Destinations on Page 33.

Important
Before leaving the site, you must:
• Check that the System Lockdown check box on the Service Functions Menu is selected.
• Log off from the Service Menu.
[28] Check that the System Lockdown check box is selected.
[29] Log off from the Service Menu.
[30] The software setup for your system is complete. Continue with the installation procedures for your CR System.
[31] Do the “Install the Modification” instructions for Software Version 5.2.00.40 in the publication Modification
Instructions for the KODAK DIRECTVIEW CR 825/850/950/975 Systems, KODAK DIRECTVIEW Classic/Elite
CR Systems, and the KODAK DIRECTVIEW WAIV Console PC with Software Version 5.2.00.13/.16/.20/.25/
.30, 8J2555, as required.

16 16APR10 – 8G7613
Doing the Configuration for the New Classic/Elite CR Systems

Section 5: Doing the Configuration for the New Classic/Elite CR Systems

Important
• You must do the procedures in this section for the new Classic/Elite CR Systems in the order listed.
• For detailed instructions on how to perform the procedures referenced here, see Section 7.

To do the configuration for the new Classic/Elite CR Systems:


[1] If necessary, do Preparing for the Software Installation on Page 25 to connect the:
• keyboard
• mouse

Important
You must check the connections to the computer.
[2] Do Checking the Connections to the Computer on Page 27.
[3] Do Energize and Log in to the CR System on Page 41.
[4] Do Doing the Calibration for the Touch Screen Monitor on Page 42.
[5] Check that the network cable is connected.
[6] Do Setting the Network Configuration on Page 48.

Note
If you change any network settings, an immediate reboot is required. After the system reboots. you must log on to
the Service Menu again.
[7] Do Setting the Configuration for the Scanner on the Classic/Elite CR Systems on Page 49.
[8] Wait for the software to begin communicating with the scanner.
[9] Reboot the computer again.
[10] Do Entering the System Information for the Option Registration on Page 43.
[11] Do Enabling the Options from the Certificates on Page 46.
[12] Do Setting the Regional Setting on Page 51.
[13] Do Setting the Print Destinations to the MIM on Page 55.
[14] Do Setting the Store Destinations to the MIM on Page 57.
[15] Do Setting the HIS/RIS on Page 58.

Important
• The customer is responsible for providing a Security Administrator (SA) and creating all the Key Operator and
User Accounts.
• For the new Classic/Elite CR Systems, you must check for new software to download to the board.
[16] Do Downloading the Software to the Boards on Page 60.
[17] Do Checking the Configuration for the Bar-Code Scanner on Page 68 for the correct operation.
[18] Is the operation of the bar-code scanner correct?
Yes No
Advance to Step 22. Continue with Step 19.

8G7613 – 16APR10 17
MODIFICATION INSTRUCTIONS, TYPE 1

Important
To do the configuration for the bar-code reader, you must use the Service Guide for the Barcode Reader for all CR/
DR Systems and CAD, 7H7713.
[19] Do the configuration for the bar-code reader.
[20] Do Checking the Configuration for the Bar-Code Scanner for the correct operation.
[21] Is the operation of the bar-code scanner correct?
Yes No
Continue with Step 22. a. Do Step 19–Step 21 again.
b. If the operation of the bar-code scanner is not correct, call the TSC.
c. Continue with Step 22.

[22] Do Setting the Other Preferences for the Bar-Code Scanner on Page 68.
[23] If necessary, do Doing the Setup for the 11-Digit Cassette ID on Page 71.
[24] Do Doing the Calibration for the Luminance of the Monitor on Page 73.
[25] Do Setting the Look Code on Page 74.
[26] Do Doing the Setup for Remote Management Services (RMS) on Page 75.
[27] Do Installing the KODAK DIRECTVIEW Remote Operations Panel (ROP) on Page 75.

Important
The customer is responsible for installing the Remote Access software. See the Remote Access Software Installation
Instructions, 6H6955.
[28] Do Installing the Other Remote Devices on Page 75.
[29] Do Completing the Installation on Page 77 for the procedures to:
• Do the FE Checklist located on the Carestream Health Service Portal.
• Check all the cassettes.
• Check the image quality with the DisplayPro test pattern.
[30] Do:
• Storing the MCB Parameters on Page 32.
• Storing the User Preferences and MIM Destinations on Page 33.

Important
Before leaving the site, you must:
• Check that the System Lockdown check box on the Service Functions Menu is selected.
• Log off from the Service Menu.
[31] Check that the System Lockdown check box is selected.
[32] Log off from the Service Menu.

18 16APR10 – 8G7613
Doing the Configuration for the New Classic/Elite CR Systems

[33] For new Classic/Elite CR Systems, feed back:


• the accessory service code 5941 for the WAIV Console PC
• the serial number on the WAIV Console S/N Sticker
• the time spent on the software and PC portions of the install to XN for service code 5941
• installation time, as indicated in the product installation instruction for XN, in relation to the primary Classic/
Elite CR service codes

[34] The software setup for your system is complete. Continue with the installation procedures for your Classic/Elite
CR System.

8G7613 – 16APR10 19
MODIFICATION INSTRUCTIONS, TYPE 1

Section 6: Doing a Software Recovery for All CR Systems with


Software Version 5.2.x

Important
• You must do the procedures in this section for the recovery of all CR Systems with Software Version 5.2.x in the
order listed.
• For detailed instructions on how to perform the procedures referenced here, see Section 7.

To do a software recovery:
[1] Do Checking the Modification and Software Installation Kit on Page 25 for the corresponding CR System.

Important
If you have a USB bar-code scanner, you do NOT have to disconnect the keyboard to connect the bar-code scanner.
[2] Do Preparing for the Software Installation on Page 25 to:
• Disconnect:
– bar-code scanner
– KODAK DIRECTVIEW Capture Link System - if the CR System is part of the Capture Link System
• Connect:
– keyboard
– mouse

Important
• Your CR System may not operate properly if the connections are not correct.
• Check your connections to minimize the need for a software recovery.
[3] Do Checking the Connections to the Computer on Page 27.

Important
If you are doing a software recovery for a new computer, you must move the software license sticker from the old
computer to the new computer.
[4] Do Installing the New License Sticker on Page 29, if you are doing a software recovery for a new computer.

Important
• In Software Version 5.2.x, the network configuration data is restored with the backup\restore procedures.
• If you are doing a software recovery for a system that is not operational, you must complete the Site Information
Sheet.
[5] Is your CR System operational?
Yes No
a. Do Storing the MCB Parameters on Page 32. a. Complete the Site Information Sheet.
b. Do Storing the User Preferences and MIM Destinations on b. Continue with Step 6.
Page 33
c. Continue with Step 6.

20 16APR10 – 8G7613
Doing a Software Recovery for All CR Systems with Software Version 5.2.x

Caution
Dangerous Voltage
[6] Do Doing the System Shutdown on Page 36.
[7] Energize the CR System.

Important
Do NOT de-energize the system when you complete Step 8 because you will need to set additional BIOS settings
after this step.
[8] Do Checking and Setting the System BIOS on Page 36.
[9] Do Installing the Software Version 5.2.x on Page 38.
[10] Do Energize and Log in to the CR System on Page 41.
[11] Do Doing the Calibration for the Touch Screen Monitor on Page 42.

Important
• In Software Version 5.2.x, the network configuration data is restored with the backup\restore procedures.
• If network settings are changed either manually or as a result of the restore, an immediate reboot is required.
After the system reboots, you must log on to the Service Menu again.
[12] If the network cable is not connected, connect it now.
[13] Restore only the network settings. Do Restoring the User Preferences on Page 52. Select only the Network
Settings check box.

Note
The system reboots.
[14] If the Classic/Elite CR System has been energized the entire time, de-energize and energize the
Classic/Elite CR System again.

Important
You must check scan status to determine if the Classic/Elite CR System is communicating with the console computer
[15] Is the Classic/Elite CR System communicating with the console computer?
Yes No
Continue with Step 16. a. De-energize and energize the Classic/Elite CR System again.
b. Do Setting the Configuration for the Scanner on the Classic/Elite CR
Systems on Page 49.
c. Continue with Step 16

8G7613 – 16APR10 21
MODIFICATION INSTRUCTIONS, TYPE 1

[16] Do you have a current backup file of user preferences on the backup diskette?
Yes No
a. Do Setting the Network Configuration on Page 48.
Important b. Do Setting the Regional Setting on Page 51.
• If you are using a backup created in Software c. Do Setting the Print Destinations to the MIM on
Version 5.2.x, you can restore all data except the Page 55.
network settings.
d. Do Setting the Store Destinations to the MIM on
• You already restored the network settings in Page 57.
Step 13.
e. Do Setting the HIS/RIS on Page 58.
• Restoring the network settings again would not
f. Continue with Step 17.
cause a problem, but another reboot would be
required.
a. Do Restoring the User Preferences on Page 52.
b. Continue with Step 17.

Important
Refer to the Software Version 5.2.x training material or call the TSC for any problems with the Option Registration
for Software Version 5.2.x.
[17] Did the options restore correctly?
Yes No
a. Continue with Step 18. a. Do Enabling the Options from the Certificates on
Page 46.
b. Continue with Step 18.

Important
• The customer is responsible for providing a Security Administrator (SA) and creating all the Key Operator and
User Accounts.
• If the user names were restored from a backup, the:
– user name is restored.
– passwords are reset to the defaults.
• You must contact the SA to have the operator reset the passwords.
• If you installed a new circuit board, you must download the software to the board.
[18] Do Downloading the Software to the Boards on Page 60.
[19] Disconnect the keyboard.
[20] Connect the bar-code reader.
[21] Do Checking the Configuration for the Bar-Code Scanner on Page 68 for the correct operation. See the Service
Guide for the Barcode Reader for all CR/DR Systems and CAD, 7H7713 for the procedures.
[22] Is the operation of the bar-code scanner correct?
Yes No
Advance to Step 26. Continue with Step 23.

[23] Do the configuration for the bar-code reader. See the Service Guide for the Barcode Reader for all CR/DR
Systems and CAD, 7H7713 for the procedures.
[24] Do Checking the Configuration for the Bar-Code Scanner on Page 68 for the correct operation.

22 16APR10 – 8G7613
Doing a Software Recovery for All CR Systems with Software Version 5.2.x

[25] Is the operation of the bar-code scanner correct?


Yes No
Continue with Step 26. a. Do Step 23–Step 25 again.
b. If the operation of the bar-code scanner is not correct, call the TSC.
c. Continue with Step 26.

[26] Do Setting the Other Preferences for the Bar-Code Scanner on Page 68.

Important
• The Look Code is installed during the installation of MOD 68 as well as when the system is installed for the first
time.
• You only have to do the installation for the Look Code if there is not a current backup on the user preference
diskette.
[27] If necessary, do Setting the Look Code on Page 74.
[28] Do Doing the Calibration for the Luminance of the Monitor on Page 73.
[29] Do Doing the Setup for Remote Management Services (RMS) on Page 75 to restore the settings for the KODAK
Remote Management System.
[30] If necessary, do Installing the KODAK DIRECTVIEW Remote Operations Panel (ROP) on Page 75. See the
Software Installation Instructions for the 19 in. KODAK DIRECTVIEW Remote Operations Panel (ROP) used on
the KODAK DIRECTVIEW CR 825/850/950/975 Systems and the KODAK DIRECTVIEW Classic/Elite CR
Systems with Software Version 5.1, 8G7963.

Important
The customer is responsible for installing the Remote Access software. See the Remote Access Software Installation
Instructions, 6H6955.
[31] Do Installing the Other Remote Devices on Page 75.

Important
If you installed a new slow-scan board, you must set the slow-scan corrections profile.
[32] Do Setting the Slow-Scan Correction Profile for the EHR Cassettes on Page 76 if you installed a new slow-scan
board.
[33] Do Completing the Installation on Page 77 to:
• Do the FE Checklist located on the Carestream Health Service Portal.
• Check all the cassettes.
• Check the image quality with the DisplayPro test pattern.

Important
If the data has changed since the last restore, you must store the new data.
[34] Do Storing the MCB Parameters on Page 32.
[35] Do Storing the User Preferences and MIM Destinations on Page 33.

Important
Before leaving the site, you must:
• Check that the System Lockdown check box on the Service Functions Menu is selected.
• Log off from the Service Menu.
[36] Check that the System Lockdown check box is selected.
[37] Log off from the Service Menu.

8G7613 – 16APR10 23
MODIFICATION INSTRUCTIONS, TYPE 1

[38] Record the information for feedback:

Note
Feedback recovery of software against RAM for Classic/Elite against the WAIV Console PC service code 5941.
Service Activity Installation Install
K-number System Code Serial No. Code Time (actual) Part No. Feedback
CR 825 5634 RAM --- No
CR 850 4825 RAM --- No
CR 950 4826 RAM --- No
CR 975 1883 RAM --- No
Classic 5258 RAM --- No
Elite 5259 RAM --- No
WAIV Console PC 5941 RAM --- No

[39] Do the “Install the Modification” instructions for Software Version 5.2.00.40 in the publication Modification
Instructions for the KODAK DIRECTVIEW CR 825/850/950/975 Systems, KODAK DIRECTVIEW Classic/Elite
CR Systems, and the KODAK DIRECTVIEW WAIV Console PC with Software Version 5.2.00.13/.16/.20/.25/
.30, 8J2555, as required.

24 16APR10 – 8G7613
Detailed Procedures for All CR Systems

Section 7: Detailed Procedures for All CR Systems

Checking the Modification and Software Installation Kit


Check the components of the Modification Kit SP8G8098:

Part No. Description Quantity


8G8755 DVD - Software Version 5.2.x for Model 8212 computer 1
8G8756 DVD - Software Version 5.2.x for Model 8810 computer 1
8G8757 DVD - Software Version 5.2.x for HP 5700 computer 1
8G8754 CD - Version 5.x software option converter tool 1
8G7613 Modification and Software Installation Instructions 1

Note
When you complete a procedure in this section, you will be instructed to click the Previous View button on the screen
until you return to the page in the applicable Section 2–6 where you left off. For information on how to access this
button, see Page 5.

Preparing for the Software Installation

Caution
• Dangerous Voltage
• If you have a USB bar-code scanner, you do NOT have to disconnect the keyboard to connect the bar-code
scanner.

[1] Disconnect the bar-code scanner, if applicable.


[2] Connect:
• keyboard
• mouse
[3] Is your CR System part of the Capture Link System?
Yes No
a. Remove your CR System from the Capture Link System, click: Continue with Step 4.
• Key Operator Functions
• Capture Link Configuration
b. Select Disconnect from a Capture Link Server.
c. Click:
• Save Changes
• OK - the CR System restarts

[4] This procedure is complete. Click Previous View until you return to the applicable page in Section 2–6.

8G7613 – 16APR10 25
MODIFICATION INSTRUCTIONS, TYPE 1

Checking for the Necessary Hardware

Important
• The Classic/Elite CR Systems only use the Model 8810 computer.
• The Software Version 5.2.x only operates with a 19 in. KODAK DIRECTVIEW Remote Operations Panel (ROP).
• If you have a 19 in. ROP connected to a CR System with Software Version 5.2.x, you cannot connect the 19 in.
ROP to a CR System with an earlier software version at the same time.
• Software Version 5.2.x includes the software refresh feature, which is only available to customers who have
purchased it. See Service Bulletin, 7H9511.
• The types of software refresh include:
– Basic Software Refresh: The customer must purchase the necessary hardware separately.
– Premium Software Refresh: You must order the necessary hardware components that are indicated in a
spreadsheet sent to your e-mail. For an example of the spreadsheet, see the graphic below.
• If you are doing an installation of necessary hardware for a CR System that has Software Refresh, you must feed
back at the end of these procedures:
– Modification 51 for the installation of the new computer.
– Modification 52 for the installation of the 2 GB of Random Access Memory (RAM).

[1] Check the CR System for the necessary hardware:


• Model 8212, Model 8810, or HP 5700 computer
• Minimum of 2 GB of RAM
• Monitor is:
CR 825/850/950/975 Systems Classic CR System Elite CR System
17 in. monitor 17 in. non-touch screen monitor 19 in. touch screen monitor

[2] This procedure is complete. Click Previous View until you return to the applicable page in Section 2–6.

26 16APR10 – 8G7613
Detailed Procedures for All CR Systems

Making a Backup Diskette for the Software Version 4.x

Important
• For Software Version 5.1, the backup diskette for the Software Version 4.x is only necessary for:
– Going back to a previous version of Software Version 4.x.
– Visually checking the previously used values.
• For Step 1, you must use the procedure “Storing and Recording the Data” in Modification Instructions and
Software Installation Instructions for Software Version 4.60.17 CP4 for all KODAK DIRECTVIEW CR Systems
7H2690.

[1] Make a backup diskette of:


• Medical Imaging Manager (MIM)
• User Preferences
[2] Complete the Site Information Sheet. See the Site Information Sheet for the KODAK DIRECTVIEW CR
Systems, 6H6825.
[3] This procedure is complete. Click Previous View until you return to the applicable page in Section 2–6.

Checking the Connections to the Computer


Model 8212 Computer
CUSTOMER
NETWORK KEYBOARD
COM 1
MOUSE

INTERNAL
COM 3 NETWORK
MONITOR
COM 4

POWER

H206_0167GCA
H206_0167GC

8G7613 – 16APR10 27
MODIFICATION INSTRUCTIONS, TYPE 1

Model 8810 for the Classic/Elite CR Systems Model 8810 for Other CR Systems
INTERNAL NETWORK

CUSTOMER CUSTOMER
NETWORK NETWORK
COM 1 COM 1
COM 3 COM 3

COM 4 COM 4
MONITOR MONITOR

BAR CODE BAR CODE


READER READER
TOUCHSCREEN TOUCHSCREEN
MOUSE MOUSE
KEYBOARD KEYBOARD

POWER POWER

H219_0006GCA
H219_0006GC H219_0007GCA
H219_0007GC

Model HP 5700 for the Classic/Elite CR Systems Model HP 5700 for Other CR Systems

unused
unused PS/2 PORTS
POWER
PS/2 PORTS POWER

COM3/unused
(ignore the COM3/UPS
COM1/
COM1/unused (ignore the
painted PC label TOUCHSCREEN
that says COM2) painted PC label
MONITOR that says COM2)
MONITOR
USB
KEYBOARD USB
USB MOUSE
USB MOUSE KEYBOARD

unused
extra NIC BAR CODE BAR CODE
READER INTERNAL NIC/ READER
SCANNER NIC
TOUCHSCREEN CUSTOMER CUSTOMER
NETWORK unused USB
NETWORK
H226_1027GCA
H226_1027GC H226_1027GCB
H226_1027GC

28 16APR10 – 8G7613
Detailed Procedures for All CR Systems

Important
After the installation of Software Version 5.1, all COM connections may be changed from your last installation.
[1] Check for the correct connections to the computer:
Touch Screen
Computer CR System Monitor UPS
8212 - 3.0 GHz CR 825/850/950/975 Systems COM 1 COM 3
8810 - 2.13 GHz, Dual Core COM 1 COM 3
HP 5700 - 2.13 GHz, Dual Core COM 1 COM 3
8810 - 2.13 GHz, Dual Core Classic/Elite CR Systems USB ---
HP 5700 - 2.13 GHz, Dual Core Classic/Elite CR Systems USB ---

[2] Check that your CR System is connected to the site network.


[3] This procedure is complete. Click Previous View until you return to the applicable page in Section 2–6.

Installing the New License Sticker


[1] Are you doing?
A Software Recovery on a CR System
MOD 68 from Software Version 4.1 with Software Version 5.2.x

Important Important
If you are doing this installation for MOD 68, you must If you are doing a software installation for a new
do Making a Backup Diskette for the Software Version replacement computer, you must transfer the license
4.x on Page 27 to record the data from your previous sticker from the old computer to the new computer.
version of software. a. If you are doing a software installation for a new
a. Install the new license sticker, which is included in replacement computer, transfer the license sticker
the customer-ordered upgrade kit, on top of the from the old computer to the new computer.
old license sticker to cover the old license sticker.

[2] This procedure is complete. Click Previous View until you return to the applicable page in Section 2–6.

8G7613 – 16APR10 29
MODIFICATION INSTRUCTIONS, TYPE 1

Recording the Network Configuration Data

Important
You must record the Network Configuration Data manually. The data is not stored in any backup location.
[1] If necessary, energize the CR System.
[2] Insert your USB service key jump drive into the USB port.
[3] Click Service Login.

Note
Your USB service key jump drive is selected.

[4] Enter your password.


[5] Click USB Service Login.
[6] From the Main Menu, click:
• Service Functions
• Network Configuration

30 16APR10 – 8G7613
Detailed Procedures for All CR Systems

[7] Record the values from the screen onto the Site Information Sheet:
• Local IP Address
• Subnet Mask
• Default Gateway
• RAS Start Address
• Device Serial Number
• Computer Name
• Network Speed/Duplex
• Firewall Enabled
• RAS End Address
[8] Click Main Menu.
[9] This procedure is complete. Click Previous View until you return to the applicable page in Section 2–6.

8G7613 – 16APR10 31
MODIFICATION INSTRUCTIONS, TYPE 1

Storing the MCB Parameters

Important
You will not use the backup diskette created in this procedure; however, if necessary, you can use the backup
diskette at a later time when installing a new board.
[1] From the Main Menu, click:
• Service Functions
• Diagnostics
• Configuration Backup/Restore

Note
The Configuration Backup/Restore screen appears.
[2] For:
HP 5700 Computers Other Computers
a. Click Copy to PC Backup. a. Insert a blank diskette into the diskette drive.
b. Wait for the message “Status: Copied to... b. Click COPY TO FLOPPY.
complete.” to display. c. Wait for the message “Status: Copied to...
c. Copy the complete.” to display.
D:\CRDiagnosticsBackup\MCB\CRxxxx_mcb d. Remove the diskette.
_configuration.txt file to your USB service key
jump drive. e. Label and keep the diskette.

d. Continue with Step 3. f. Continue with Step 3.

[3] Click:
• Done
• Back
• Main Menu
[4] This procedure is complete. Click Previous View until you return to the applicable page in Section 2–6.

32 16APR10 – 8G7613
Detailed Procedures for All CR Systems

Storing the User Preferences and MIM Destinations


[1] Click:
• Main Menu
• Key Operator Functions
• Backup Restore
[2] Click Select All.

Note
A mark displays in each check box.

Backup / Restore
Directory Field

[3] Select the Backup / Restore Directory field.

8G7613 – 16APR10 33
MODIFICATION INSTRUCTIONS, TYPE 1

[4] If necessary, insert:


• USB jump drive,
• USB service key jump drive, or
• Diskette
[5] From the drop-down menu, select the correct drive:
Drive Media
A:\3 1/2 Inch Floppy Drive diskette
F:\Removable Disk USB jump drive

[6] Select the backup folder or make a new backup folder.


[7] Click:
• Select
• Backup

Important
The status of the backup is indicated by a blue progress indicator.
[8] Wait for the backup to complete.
[9] Click OK.
[10] If necessary, remove the diskette from the diskette
drive.
[11] Click Main Menu.

Note
You might also need to manually record these settings:

If doing MOD 68/13 from Software Version <5.1.0x.50 If doing MOD 68/13 from Software Version 5.1.0x.50
Manually record all settings except: Manually record:
• Procedure Mapping • Display Configuration
• Destinations • Remote Device Configuration
• Network Settings
[12] This procedure is complete. Click Previous View until you return to the applicable page in Section 2–6.

34 16APR10 – 8G7613
Detailed Procedures for All CR Systems

Extracting the Options from a System with Software Prior to Software Version 5.2

Important
• The AddOnOptionConverter.exe utility must be run on a system prior to upgrading the system to Software
Version 5.2.x.
• The AddOnOptionConverter.exe utility extracts the options from a system with Software Version 5.1.x or 4.x.
• The AddOnOptionConverter utility and the related files can be obtained from the Carestream Health Service
Portal. It is also on a CD provided with the Software Version 5.2 kit.
[1] Obtain:
• AddOnOptionConverter.exe utility
• msvcr80.dll file
• msvcp80.dll file
[2] For:
Software Version 5.1.x Software Version 4.x
Copy the AddOnOptionConverter.exe utility to the Copy to the C:\Temp folder:
C:\Temp folder. • AddOnOptionConverter.exe utility
• msvcr80.dll file
• msvcp80.dll file

[3] Open the C:\Temp folder and run the AddOnOptionConverter.exe utility.

Note
• The legacyoptions.opt file is created in the C:\Temp folder.
• The legacyoptions.opt file exported from the AddOnOptionConverter.exe utility will be used after upgrading the
CR System to Software Version 5.2.
[4] Copy the legacyoptions.opt to your USB service key jump drive.
[5] This procedure is complete. Click Previous View until you return to the applicable page in Section 2–6.

8G7613 – 16APR10 35
MODIFICATION INSTRUCTIONS, TYPE 1

Doing the System Shutdown

Quick Menu
Icon

[1] From the Main Menu, click:


• Quick Menu icon
• Shutdown
• Shutdown/Power Off
• OK
• Yes
[2] This procedure is complete. Click Previous View until you return to the applicable page in Section 2–6.

Checking and Setting the System BIOS

Important
• You only have to set the system BIOS for the computer one time.
• Hyper-threading is added in Software Versions 5.1.x and 5.2.x.

[1] Are you doing a software recovery on a system with:


• Software Version 5.2.x, and
• The same computer installed with the installation of MOD 68/13?
Yes No
Advance to Step 10. Continue with Step 2.

Caution
Dangerous Voltage
[2] If necessary, de-energize the CR System.
[3] Energize the CR System.
[4] Observe the screen when the CR System energizes.

36 16APR10 – 8G7613
Detailed Procedures for All CR Systems

[5] When the data first appears on the screen, press:


For Model 8212 and Model 8810 Computers For HP 5700 Computers
• F1 • F10
• Enter 518699 if prompted for a password. • Enter 518699 if prompted for a password.
• Select the appropriate language.

[6] Use the table below to set the system BIOS for your CR System:

CR 825/850/950/975 Systems Classic/Elite CR Systems


Model 8212 Computers Model 8810 Computers Model 8810 Computers
Advanced>Hyperthreading: Enabled Advanced>Core Multi-Processing: Enabled
Security>TCG Feature Setup>TCG Feature: Disabled
Startup>Startup Sequence>Primary Startup Sequence:
1. IDE HD
2. IDE CD
3. Diskette Drive A
Startup>Power On Self Test: Quick
Startup>Power On Logo: Disabled
N/A Startup>Keyboard Numlock State: Off
Power>After Power Loss: Power On Power>After Power Loss: Last State
Security>Set Passwords>Set Administrator Password: 518699

HP 5700 Computers
Storage>Boot Order:
Hard drive
ATAPI CD-ROM Drive
For CR825/850/950/975: Advanced>Power on Options>After Power Loss: On
For Classic/Elite: Advanced>Power on Options>After Power Loss: Previous State
Security>Set Passwords>Set Administrator Password: 518699
[7] Press F10.
[8] At the Setup Confirmation screen, select Yes.
[9] Press Enter.
[10] This procedure is complete. Click Previous View until you return to the applicable page in Section 2–6.

8G7613 – 16APR10 37
MODIFICATION INSTRUCTIONS, TYPE 1

Setting the Boot Order

Important
It is no longer necessary to use this procedure for setting the boot order. Using the appropriate function key during
boot-up allows you to select the boot order at boot time. The appropriate function key is noted in Installing the
Software Version 5.2.x on Page 38.

Caution
• Dangerous Voltage
• To install the software for the CR System, you must set the system BIOS to boot from the CD-ROM.
• After the software installation, you must set the system BIOS to boot from the hard drive.
[1] For:
Model 8810 Computers Model 8212 Computers
a. Press: a. Press [I] or [J] to select the Boot Menu.
• [I] or [J] to select the Startup Sequence b. Continue with Step 2.
Menu
• [Enter]
b. Continue with Step 2.

[2] To set the Boot Order.


To boot to CD-ROM To boot to Hard Drive
a. Press: a. Press:
• [ K ] or [ L ] to select the CD-ROM • [ K ] or [ L ] to select the hard drive
• [ + ] or [ - ] to move the CD-ROM above the • [ + ] or [ - ] to move the hard drive above the
hard drive CD-ROM
• F10 to save settings • F10 to save settings
• [I] or [J] to select Yes • [I] or [J] to select Yes
b. Continue with Step 3. b. Continue with Step 3.

[3] When data first appears on the screen, press the power button to de-energize the CR System.
[4] This procedure is complete. Click Previous View until you return to the applicable page in Section 2–6.

Installing the Software Version 5.2.x


Preparation
[1] Explain to the customer that all images will be deleted during the Software Version 5.2.x install.

Important
Before continuing:
• You must de-energize the CR System.
• You must remove all USB jump drives.
[2] If necessary:
• De-energize the CR System.
• Remove all USB jump drives.
[3] For:
Model 8810 Computers Model 8212 Computers HP 5700 Computers
Continue with For Model 8810 Advance to For Model 8212 Advance to For HP 5700
Computers. Computers on Page 39. Computers on Page 40.
38 16APR10 – 8G7613
Detailed Procedures for All CR Systems

For Model 8810 Computers

Important
• You must insert the DVD into the DVD drive immediately after energizing the CR System in order for the CR
System to boot from the DVD.
• If the CR System does not boot from the DVD, you must de-energize the CR System and do Step 1 and Step 2
again.
[1] Energize the CR System.
[2] Insert the Software Version 5.2 DVD into the DVD drive.
[3] Reboot the PC and press F12 at startup to get the Boot Menu.
[4] When prompted for a password, enter 518699.
[5] At the message “Press Enter to Select Boot Device, ESC to Exit.”, select CD-ROM.

[6] At the PC DOS 7.1 Startup Menu screen, check that 1. Restore the entire drive and overwrite all data is
selected.
[7] For Enter a Choice, press Enter.
[8] For the message “Press any key to continue.”, press Enter.
[9] Wait approximately 15 minutes for the software installation to complete.

Note
When the software installation is complete, the A:\ prompt displays.

[10] At the A:\ prompt, remove the DVD from the DVD drive.
[11] Press Ctrl + Alt + Del.

Note
• The computer boots.
• The software installation is complete.
[12] If the message “New Hardware Found” displays, click No.
[13] If the message “System Settings have changed, you must reboot.” displays, click No.
[14] If the warning “Firewall OFF” displays, close the window.
[15] This procedure is complete. Click Previous View until you return to the applicable page in Section 2–6.

For Model 8212 Computers

Important
• You must insert the DVD into the DVD drive immediately after energizing the CR System in order for the CR
System to boot from the DVD.
• If the CR System does not boot from the DVD, you must de-energize the CR System and do Step 1 and Step 2
again.
[1] Energize the CR System.
[2] Insert the Software Version 5.2 DVD into the DVD drive.

Note
If the system boots to the WINDOWS Operating System, the system BIOS values will not be correct. Do Checking
and Setting the System BIOS on Page 36.
[3] Reboot the PC and press F12 at start-up to get the Boot Menu.
[4] When prompted for a password, enter 518699.
[5] At Press Enter to Select Boot Device, ESC to Exit, select CD-ROM.

8G7613 – 16APR10 39
MODIFICATION INSTRUCTIONS, TYPE 1

[6] At the PC DOS 7.1 Startup Menu screen, check that 1. Restore the drive and overwrite all data is selected.
[7] For Enter a Choice, press Enter.
[8] For the message “Press any key to continue.”, press Enter.
[9] Wait approximately 15 minutes for the software installation to complete.

Note
When the software installation is complete, the A:\ prompt displays.
[10] At the A:\ prompt, remove the DVD from the DVD drive.
[11] Press Ctrl + Alt + Del.

Note
• The computer boots.
• The software installation is complete.

[12] If the message “New Hardware Found” displays, click No.


[13] If the message “System Settings have changed, you must reboot.” displays, click No.
[14] If the warning “Firewall OFF” displays, close the window.
[15] This procedure is complete. Click Previous View until you return to the applicable page in Section 2–6.

For HP 5700 Computers

Important
• You must insert the DVD into the DVD drive immediately after energizing the CR System in order for the CR
System to boot from the DVD.
• If the CR System does not boot from the DVD, you must de-energize the CR System and do Step 1 and Step 2
again.
[1] Energize the CR System.
[2] Insert the Software Version 5.2 DVD into the DVD drive.
[3] Reboot the PC and press F9 at start-up to get the Boot Menu.
[4] When prompted for a password, enter 518699.
[5] At Press Enter to Select Boot Device, ESC to Exit, select CD-ROM.
[6] At the PC DOS 7.1 Startup Menu screen, check that 1. Restore the entire drive and overwrite all data is
selected.
[7] For Enter a Choice, press Enter.
[8] For the message “Press any key to continue.”, press Enter.
[9] Wait approximately 15 minutes for the software installation to complete.

Note
When the software installation is complete, the A:\ prompt displays.
[10] At the A:\ prompt, remove the DVD from the DVD drive.
[11] Press Ctrl + Alt + Del.

Note
• The computer boots.
• The software installation is complete.
[12] If the message “New Hardware Found” displays, click No.
[13] If the message “System Settings have changed, you must reboot.” displays, click No.
[14] If the warning “Firewall OFF” displays, close the window.
[15] This procedure is complete. Click Previous View until you return to the applicable page in Section 2–6.

40 16APR10 – 8G7613
Detailed Procedures for All CR Systems

Energize and Log in to the CR System

Caution
Dangerous Voltage
[1] Energize the CR System.
[2] If necessary, insert your USB service key jump drive into the USB port.
[3] Click Service Login.

Important
• Your USB service key jump drive is selected.
• If your USB service key jump drive is not recognized, you must restart your system with the USB service key
jump drive installed.

[4] Enter your password.


[5] Click USB Service Login.

Important
If the message “Call Service to schedule Preventive Maintenance.” displays, do NOT select the Do not remind me
again check box.

[6] If the message “Call Service to schedule Preventive Maintenance.” displays, click OK.
[7] If the message “System Settings have changed, you must reboot.” displays, click No.
[8] If the Removable Disk window appears:
a. Select:
• Take No Action
• Always do this
b. Click OK.
[9] If the message “System has found new Hardware.” displays, click No.
[10] This procedure is complete. Click Previous View until you return to the applicable page in Section 2–6.

8G7613 – 16APR10 41
MODIFICATION INSTRUCTIONS, TYPE 1

Doing the Calibration for the Touch Screen Monitor


[1] Does your CR System have a touch screen monitor?
Yes No
Continue with Step 2. It is not necessary to do this procedure.

[2] From the Main Menu, click:


• Key Operator Functions
• Monitor Configuration
• Touch Screen Calibration
[3] Touch the target when it moves to the 3 areas on the screen.

Check Mark Button Back Button

[4] Does the cursor position change when you touch the target?
Yes No
a. Click the check mark button. a. Click the back button.
b. Continue with Step 5. b. Do Step 3 and Step 4 again.
c. If necessary, install a new monitor.
d. Do Step 2–Step 4 again.

[5] Click Main Menu.


[6] This procedure is complete. Click Previous View until you return to the applicable page in Section 2–6.

42 16APR10 – 8G7613
Detailed Procedures for All CR Systems

Entering the System Information for the Option Registration


[1] If Software Patch 5.2.00.40 is already installed on this CR System, see “Changes to KeyOp Screen: New Option
Added” in the publication Modification Instructions for the KODAK DIRECTVIEW CR 825/850/950/975
Systems, KODAK DIRECTVIEW Classic/Elite CR Systems, and the KODAK DIRECTVIEW WAIV Console PC
with Software Version 5.2.00.13/.16/.20/.25/.30, 8J2555, for instructions on how to add the site information to
the Site Information tab on the Option Registration screen.
[2] From the Main Menu, click:
• Key Operator Functions
• System Configuration
• Delivery Preferences

[3] Enter the information for:


• Hospital name
• Hospital address
[4] From the Main Menu, click:
• Save Changes
• Main Menu
• Service Functions
• Weblink
[5] On the left side of the screen, select System Information>Identification.
[6] Click Modify.

8G7613 – 16APR10 43
MODIFICATION INSTRUCTIONS, TYPE 1

[7] Check that the serial number is correct.

Important
When entering the K-number, enter only the number. Do NOT type the K-.
[8] Enter the customer information:
• K-number - enter only the number
• Service Phone Number - if necessary
• Country Code
• Region
• Hospital Name
• Department Name
• Device Location
• Language
[9] Click Save.
[10] This procedure is complete. Click Previous View until you return to the applicable page in Section 2–6.

44 16APR10 – 8G7613
Detailed Procedures for All CR Systems

Enabling the Options for an Upgrade from Legacy Software

Important
• Legacy refers to any system with software versions prior to Software Version 5.2.
• You must have created a legacyoptions.opt file in Extracting the Options from a System with Software Prior to
Software Version 5.2 on Page 35.

[1] From the Main Menu, click:


• Key Operator Functions
• Option Registration

[2] Click Import Legacy Option File.

Note
• If this is a purchased upgrade, the procedure is complete until you add the Software Version 5.2 certificate.
• If this is a maintenance upgrade and only the option converter file is used, continue with the remainder of the
procedure.
• The options from the previous software version display on the screen as Legacy Option Code.
• The Status for each of these options is displayed as Grace Period Active.
[3] For:
Purchased Upgrade Maintenance Upgrade
Advance to Step 11, Continue with Step 4,

[4] Click Export Request File.


[5] Select the USB jump drive.
[6] Click OK.

Note
• A file named OptionActivationRequest.xml is created on the USB jump drive.
• If you do not have immediate access to the Internet, the option remains active for several days.
[7] Send the new OptionActivationRequest.xml file to the activation@carestreamhealth.com.

Note
The activation file OptionActivationResponse.xml is returned to you.
[8] Click Import Activation File.
[9] Select the new activation file.

8G7613 – 16APR10 45
MODIFICATION INSTRUCTIONS, TYPE 1

[10] Click OK.

Note
The status changes from Grace Period to Accepted.
[11] This procedure is complete. Click Previous View until you return to the applicable page in Section 2–6.

Enabling the Options from the Certificates

Important
• Some options might be enabled during manufacturing.
• If this is an installation of a new CR System with the computer serial number recorded on the options certificate
from manufacturing, the options on this certificate do not need to be enabled.

[1] Locate the Options Certificates for your system.


[2] From the Main Menu, click:
• Key Operator Functions
• Option Registration

46 16APR10 – 8G7613
Detailed Procedures for All CR Systems

Important
• If the bar-code reader is not functioning yet, you can do this procedure after it is functioning.
• Only type the bar code if you cannot enter the bar code with the reader.
• If you type the Activation Code incorrectly, an error message displays.
• You only have to click Add Code if you type the bar code.
[3] Type or scan the bar code from the first Option Certificate.
[4] If you typed the bar code, click Add Code.

Note
• Valid codes display under Codes.
• Invalid codes display an error message.
[5] For all Option Certificates, do Step 3 and Step 4 again.
[6] Click Export Request File.
[7] Select the USB drive.
[8] Create a new folder with a new name to identify this CR System, and click Save.
[9] Click Select to save your option request file.

Note
• A file named OptionActivationRequest.xml is created on the USB drive.
• If you do not have immediate access to the Internet, the option remains active for several days.
[10] E-mail the new OptionActivationRequest.xml file to the activation@carestreamhealth.com.

Note
• The activation file OptionActivationResponse.xml is returned to you.
• The activation file should be returned to you within 5 minutes.
• If the activation file is not returned to you, do Step 10 again. If necessary, call the TSC.
• If you do not have access to e-mail at this time, you can complete Step 11–Step 13 at a later time.
[11] Click Import Activation File.
[12] Select the new activation file.
[13] Click OK.

Note
The Status changes from Grace Period to Accepted.
[14] Store all option certificates at the site.
[15] This procedure is complete. Click Previous View until you return to the applicable page in Section 2–6.

8G7613 – 16APR10 47
MODIFICATION INSTRUCTIONS, TYPE 1

Setting the Network Configuration

Important
• The software does not automatically restore the network configuration values.
• You must enter the values in manually.
[1] From the Main Menu, click:
• Service Functions
• Network Configuration

Important
The device serial number must contain 6 digits. If the device serial number is less than 6 digits, add the number zero
to the left of the number until the number has a total of 6 digits. For example, enter number 4444 as 004444.
[2] Enter the correct information from the Site Information Sheet:
• Local IP Address
• Subnet Mask
• Default Gateway
• RAS Start Address
• Device Serial Number - must contain 6 digits. If necessary, add zeros to the left of the number.
• K-number - must contain 8 numeric digits. Do NOT type the K-.
• Computer Name
• Network Speed/Duplex
• Firewall Enabled
• RAS End Address

48 16APR10 – 8G7613
Detailed Procedures for All CR Systems

[3] Click:
• Save Changes
• Yes at the message “System must reboot.”

Note
The CR System reboots.
[4] Is the hospital network cable connected?
Yes No
Continue with Step 5. a. Connect the cable.
b. Enter the K-number again - the number must contain 8 numeric
digits. Do NOT type the K-.
c. Continue with Step 5.

[5] This procedure is complete. Click Previous View until you return to the applicable page in Section 2–6.

Setting the Configuration for the Scanner on the Classic/Elite CR Systems


[1] Have you already downloaded the Software Version 5.2.x to the Classic/Elite CR System boards?
Yes No
At the Local User Interface (LUI), set the IP Address of the computer: Continue with Step 2.
a. At the LUI, touch:
• Network Configuration
• Console TCP/IP Settings
b. Enter the IP Address of the computer.
c. Touch the green arrow to exit the menu of the LUI.
d. Advance to Step 10.

[2] Check that your USB service key jump drive is inserted into the USB port.
[3] From the Main Menu, click Service Login.

Note
The USB service key jump drive is automatically selected.
[4] Enter your password.
[5] Click:
• USB Service Login
• Service Functions
• Diagnostics

8G7613 – 16APR10 49
MODIFICATION INSTRUCTIONS, TYPE 1

[6] For the Classic/Elite CR Systems, enter the Scanner IP Address.


[7] Click Save Changes.
[8] Click OK at the message “A reboot is needed.”

Note
The computer reboots.
[9] De-energize, and then energize the scanner.
[10] This procedure is complete. Click Previous View until you return to the applicable page in Section 2–6.

50 16APR10 – 8G7613
Detailed Procedures for All CR Systems

Setting the Regional Setting


[1] Check that your USB service key jump drive is inserted into the USB port.
[2] From the Main Menu, click Service Login.

Note
The USB service key jump drive is automatically selected.

[3] Enter your password.


[4] Click USB Service Login.
[5] At the Main Menu of the CR System, click:
• Key Operator Functions
• System Configuration
• Regional Configuration

[6] Set the Regional Settings to the correct settings.


[7] Click:
• Save Changes
• Main Menu
[8] This procedure is complete. Click Previous View until you return to the applicable page in Section 2–6.

8G7613 – 16APR10 51
MODIFICATION INSTRUCTIONS, TYPE 1

Restoring the User Preferences


[1] Click:
• Main Menu
• Key Operator Functions
• Backup Restore

[2] Click Select All.

Note
• A check mark displays in each check box.
• You may have to remove some check marks.
[3] For:
If doing MOD 68/13 from If doing MOD 68/13 from If doing Recovery from
Software Version <5.1.0x.50 Software Version 5.1.0x.50 Software Version 5.2.x
Remove the check mark from all Remove the check mark from: Leave all items selected.
items except: • Display Configuration
• Procedure Mapping and Views • Remote Device Configuration
• Destinations • Network Settings

Backup / Restore
Directory field

[4] Select the Backup / Restore Directory field.

52 16APR10 – 8G7613
Detailed Procedures for All CR Systems

[5] Insert the correct media:


• User preference diskette, or
• USB jump drive
[6] From the drop-down menu, select the correct drive:
Drive Media
A:\ 3 1/2 Inch Floppy Drive Diskette
F:\Removable Disk USB jump drive

[7] Select the correct folder to restore.


[8] Click:
• Select
• Restore
[9] Wait for the restore to complete.

Note
The status is indicated by a blue progress indicator.

8G7613 – 16APR10 53
MODIFICATION INSTRUCTIONS, TYPE 1

[10] Click OK.

Important
Do NOT remove the USB jump drive if you are using it for service login.
[11] Remove:
• Diskette, or
• USB jump drive
[12] Click Main Menu.
[13] This procedure is complete. Click Previous View until you return to the applicable page in Section 2–6.

54 16APR10 – 8G7613
Detailed Procedures for All CR Systems

Setting the Print Destinations to the MIM


[1] From the Main Menu, click:
• Service Functions
• Weblink

Important
You might have to obtain information from Carestream Health V&V for specific devices.

[2] Select:
• Configuration>Destinations>Install DICOM Print
• correct printer at the Print Destination Defaults drop-down menu
[3] Click Modify.
[4] If necessary, select the Identification tab.
[5] Enter the Device Name of the Print Destination.

[6] Select the DICOM tab.


[7] Use the Site Information Sheet to enter:
• AE Title
• IP Address
• Port

8G7613 – 16APR10 55
MODIFICATION INSTRUCTIONS, TYPE 1

[8] Select:
• Media tab
• correct Media Size
• correct Media Base
[9] Set Configured to:
• True - for the selected media
• False - for the media that is not selected
[10] Click Save.
[11] To edit any print destinations, select Configuration>Destinations>DICOM Print.
[12] Click Main Menu.
[13] This procedure is complete. Click Previous View until you return to the applicable page in Section 2–6.

56 16APR10 – 8G7613
Detailed Procedures for All CR Systems

Setting the Store Destinations to the MIM


[1] From the Main Menu, select:
• Configuration>Destinations>Install DICOM Store
• correct store at the Store Destination Defaults drop-down menu
[2] Click Modify.

[3] Select the DICOM tab.


[4] Use the Site Information Sheet to enter:
• AE Title
• IP Address
• Port
[5] Set Send Private Tags to True.
[6] Click Save.
[7] To edit any store destinations, select Configuration>Destinations>DICOM Store.
[8] Click Main Menu.
[9] This procedure is complete. Click Previous View until you return to the applicable page in Section 2–6.

8G7613 – 16APR10 57
MODIFICATION INSTRUCTIONS, TYPE 1

Setting the HIS/RIS


[1] At the Main Menu, click:
• Key Operator Functions
• System Configuration
• HIS/RIS Configuration

[2] Select the Remote Query tab.


[3] If necessary, enter the values recorded on the Site Information Sheet.
[4] Select the Polling tab.

Important
• You must set the polling parameters for optimum function:
– of the CR System
– for the site
• It is recommended that a maximum of 40 records be polled per minute. For example, if the Minutes Between
Search Updates is 3 minutes, then a maximum of 120 records should be polled at one time.
• The recommended value for Hours Before Current Time is 1. This allows for small differences in time.
• The Hours After Current Time value should be set to the time that the site might schedule records in advance.
• After setting the values for polling, you must check the polling values at the busiest time of the day.
• To make adjustments to obtain the optimum value of a maximum of 40 records be polled per minute, you can:
– Increase the value for Minutes Between Search Updates
– Decrease the value for Hours After Current Time
• You must check with the customer to determine which parameter is more important:
– Hours After Current Time
– Minutes Between Search Updates

[5] If necessary, enter the values recorded on the Site Information Sheet.

58 16APR10 – 8G7613
Detailed Procedures for All CR Systems

[6] Select the Options tab.


[7] If necessary, enter the values recorded on the Site Information Sheet.
[8] If you entered any values on the HIS/RIS screens, click Save Changes.
[9] Click Main Menu.
[10] This procedure is complete. Click Previous View until you return to the applicable page in Section 2–6.

Setting the Profile Destination Configuration


[1] At the Main Menu, click:
• Key Operator Functions
• System Configuration
• Destination Profile Configuration

8G7613 – 16APR10 59
MODIFICATION INSTRUCTIONS, TYPE 1

[2] If necessary, enter the values recorded on the Site Information Sheet.
[3] If you entered any values on the Destination Profile Configuration screen, click Save Changes.
[4] Click Main Menu.
[5] This procedure is complete. Click Previous View until you return to the applicable page in Section 2–6.

Downloading the Software to the Boards


For:
CR 825M, 850M, or 975 Systems CR 825/850/950 Systems Classic/Elite CR Systems
Continue with Downloading the Advance to Downloading the Advance to Downloading the
Software for CR 825M, 850M or Software for CR 825/850/950 Software for Classic/Elite CR
975 Systems. Systems on Page 62. Systems on Page 64.

Downloading the Software for CR 825M, 850M or 975 Systems


[1] At the Main Menu, click:
• Service Functions
• Diagnostics
• Software Download

Note
• The graphic above is an example only.
• See the table in Step 14 for the correct values.
[2] Are any of the Available Files on your screen a newer version than the CURRENT VERSION?
Yes No
Continue with Step 3. Advance to Step 16

60 16APR10 – 8G7613
Detailed Procedures for All CR Systems

Important
• You must click the DOWNLOAD MCPU APPLICATION last to prevent additional shutdowns.
• You must do a shutdown at the prompt after downloading a new MCPU application.
• If you are unable to complete the software download to the board in Step 3–Step 9, try to download the MCPU
and shutdown, first.
• If the System Initializing screen displays during this procedure, you can move it to the side and continue with the
procedure.
[3] If necessary, to install a newer version of software for MCB BOOT, click:
• DOWNLOAD MCB BOOT
• CONFIRM MCB DOWNLOAD
[4] If necessary, to install a newer version of software for MCB PLD APPLICATION, click:
• DOWNLOAD MCB PLD
• CONFIRM MCB PLD DOWNLOAD
[5] If necessary, to install a newer version of software for MCB APPLICATIONS, click:
• DOWNLOAD MCB APPLICATION
• CONFIRM MCB DOWNLOAD
[6] If necessary, to install a newer version of software for MCB PARAMETERS, click DOWNLOAD MCB
PARAMETERS.
[7] At the MCB Parameters screen, click Current
MCB Values.
[8] Click CONFIRM MCB PARAMETERS
DOWNLOAD.
[9] If necessary, to install a newer version of software
for MCPU APPLICATION, click:
• DOWNLOAD MCPU APPLICATION
• CONFIRM MCPU DOWNLOAD
[10] Do a shutdown of the CR System.

Caution
Dangerous Voltage

[11] Energize the CR System.


[12] Use your USB service key jump drive to log on to the Service Menu.
[13] From the Main Menu, click:
• Service Function
• Diagnostics
• Software Versions
[14] Use the table to check that the software versions on the screen are correct.
Platform CR 825M System CR 850M System CR 975 System
MCPU 4.7.11 4.7.11 5.3.11
MCB Board Boot 1.62 1.62 1.62
MCB Application 1.0.0.7 1.0.0.7 1.0.2.6
MCB PLD 1 1 3
MCB Parameters 5.0 5.0 12.0
Diagnostics 5.2.00.10 5.2.00.10 5.2.00.10

8G7613 – 16APR10 61
MODIFICATION INSTRUCTIONS, TYPE 1

[15] Click:
• Back
• Main Menu
[16] This procedure is complete. Click Previous View until you return to the applicable page in Section 2–6.

Downloading the Software for CR 825/850/950 Systems


[1] From the Main Menu, click:
• Service Functions
• Diagnostics
• Software Download

Important
• The graphic above is an example only.
• See the table below for the correct values.
• For systems with the mammography feature MOD 13 or MOD 14 installed:
– The Available File for all boards must be CR825850950A.SSParams.txt.
– You must only install software with an “A” in the Available File field.
[2] Are any of the Available Files on the screen a newer version than the CURRENT VERSION?
Yes No
Continue with Step 3. a. Click:
• Done
• Back
• Main Menu
b. Advance to Step 12.

[3] Click:
• DOWNLOAD MSC APPLICATION
• CONFIRM MSC DOWNLOAD
• DOWNLOAD SSC BOOT and APPLICATIONS

62 16APR10 – 8G7613
Detailed Procedures for All CR Systems

[4] If necessary, at the Slow Scan Home Position screen, select the correct home position value.
[5] Click:
• DOWNLOAD SSC PARAMETERS
• DOWNLOAD MCPU APPLICATION
[6] Do a shutdown of the CR System.
[7] Energize the CR System.
[8] Use your USB service key jump drive to log on to the Service Menu.
[9] At the Main Menu, click:
• Service Functions
• Diagnostics
• Software Versions
[10] Use the table to check that the software versions on the screen are correct.
Platform CR 825 System CR 850A System CR 850 System CR 950 System
MCPU 4.7.11 4.7.11 3.8.15 3.8.15
MSC Board Boot 3.01 3.01 3.01 3.01
MSC Board 4.24 4.24 4.24 4.24
SSC Boot 5.89 5.89 5.89 5.89
SSC Product Application 1.13 1.13 1.13 1.13
SSC Board Fixed Application 5.89.18 5.89.18 5.89.18 5.89.18
Diagnostics 5.2.00.10 5.2.00.10 5.2.00.10 5.2.00.10

[11] Click:
• Back
• Main Menu
[12] This procedure is complete. Click Previous View until you return to the applicable page in Section 2–6.

8G7613 – 16APR10 63
MODIFICATION INSTRUCTIONS, TYPE 1

Downloading the Software for Classic/Elite CR Systems

Important
If it is ever necessary to downgrade back to Software Version 5.1, see Downgrading to Software Version 5.1 for the
Classic/Elite CR Systems on Page 66.
[1] From the Main Menu, click:
• Service Functions
• Diagnostics
• Configuration Backup/Restore

[2] Click:
• COPY TO PC BACKUP
• DONE
• Software Download

Important
• If the current IEB software is V2.0.00.xx, the Display Boot Loader and the IEB Boot Loader button are disabled.
• The graphic above is an example only.
• See the table below for the correct values.
• When upgrading from IEB V2.xx to IEB V3.xx, the IEB software automatically downloads first.

64 16APR10 – 8G7613
Detailed Procedures for All CR Systems

[3] Are any of the Available Versions on the screen a newer version than the Current Version?
Yes No
Continue with Step 4. Advance to Step 15.

[4] Click:
• SELECT ALL Applications
• START DOWNLOAD
[5] Does the message “Use the current MCB Board Values or use Default Values?” appear?
Yes No
Click: Click OK at the message “Scanner is now initializing.”
• Current MCB Values
• OK at the message “Scanner is now initializing.”

Note
• When upgrading from Software Version 5.1 to 5.2, the file format of the display application is changed and the
Boot software can now be loaded.
• The Software Version 5.2 display boot and display application software are downloaded.
• The display board automatically restarts when the software download is complete.

[6] Select IEB Boot Loader.


[7] Click START DOWNLOAD.
[8] Wait for the status to display as Done.
[9] Do a shutdown of the computer and the scanner.
[10] Energize the CR System.
[11] Use your USB service key jump drive to log on to the Service Menu.
[12] At the Main Menu, click:
• Service Functions
• Diagnostics
• Software Download

8G7613 – 16APR10 65
MODIFICATION INSTRUCTIONS, TYPE 1

[13] Use the table to check that the software versions on the screen are correct.
Platform Classic/Elite CR Systems
MCB Boot 1.62
MCB PLD 1
MCB Application 1.0.3.8
MCB Parameters 13.0
Display Application 2.0.00.b06
IEB FPGA 19
IEB Kernel 3.11
IEB Application 3.0.10.b8
Diagnostics 5.2.00.10

[14] Click:
• Done
• Back
• Main Menu
[15] This procedure is complete. Click Previous View until you return to the applicable page in Section 2–6.

Downgrading to Software Version 5.1 for the Classic/Elite CR Systems

Important
• To downgrade to Software Version 5.1, you must downgrade the embedded software version while Software
Version 5.2 is still installed on the computer.
• The Software Version 5.1 cannot download the display boot embedded software.
• The necessary Software Version 5.1 embedded software files are in the
C:\CRDiagnostics\ClassicElite V5.1.01.xx GoBack directory.

[1] At the Service Menu, clear the System Lockdown check box.
[2] At the desktop, select Start>Run.

66 16APR10 – 8G7613
Detailed Procedures for All CR Systems

[3] Browse to and click the C:\CRDiagnostics\ClassicElite V5.1.01.xx GoBack\GoBack.bat file.


[4] Click:
• Open
• OK

Note
• The Software Version 5.1 files are copied to the software download directory.
• The Software Version 5.2 files are backed up to the C:\CRDiagnostics\ClassicElite V5.2 GoForward directory.
• Do not restart the computer.
[5] At the Service Menu, select the System Lockdown check box.
[6] Click:
• Diagnostics
• Software Download

[7] At the SOFTWARE DOWNLOAD screen, select the check boxes for:
• Display Boot Loader
• Display Application
• IEB Boot Loader
• IEB FPGA
• IEB Kernel
• IEB Application
[8] Click START DOWNLOAD.
[9] Install the drive image for Software Version 5.1 to the computer.
[10] Download the MCB, display, and IEB software as required for that software version.

Important
• For instructions to restore the embedded software files to Software Version 5.2, do Step 11–Step 15.
• The embedded software files for Software Version 5.2 are in the
C:\CRDiagnostics\ClassicElite V5.2 GoForward directory.
[11] At the Service Menu, clear the System Lockdown check box.

8G7613 – 16APR10 67
MODIFICATION INSTRUCTIONS, TYPE 1

[12] At the desktop, select Start>Run.


[13] Browse to and click the C:\CRDiagnostics\ClassicElite V5.1.01.xx GoForward\GoForward.bat file.
[14] Click:
• Open
• OK

Note
Do NOT restart the computer at this time.
[15] Follow the instructions to upgrade from Software Version 5.1 to 5.2.
[16] This procedure is complete. Click Previous View until you return to the applicable page in Section 2–6.

Checking the Configuration for the Bar-Code Scanner


[1] From the Main Menu, click:
• Key Operator Functions
• System Configuration
• Bar Code Configuration
• Bar Code Configuration
[2] Use the bar-code scanner to scan the bar code of any cassette.

Note
If the bar code is configured correctly:
• Test field displays Plate ID or Cassette ID
• Test Data field displays the ID number of the cassette
[3] Click Main Menu.
[4] This procedure is complete. Click Previous View until you return to the applicable page in Section 2–6.

Setting the Other Preferences for the Bar-Code Scanner


Country Code

Important
• The default country is USA.
• If the country is not USA, you must change the country code at the Bar Code Configuration screen.
[1] Is the country USA?
Yes No
Advance to Step 1. Continue with Step 2.

[2] At the Main Menu, touch:


• Key Operator Functions
• System Configuration
• Bar Code Configuration
• Bar Code Configuration
[3] Select the Country Code field.

Note
The countries display.
[4] Select the correct country.

68 16APR10 – 8G7613
Detailed Procedures for All CR Systems

[5] Touch:
• Save Changes
• Back
• Bar Code Programming
• Country Code
[6] Use the bar-code scanner to scan the 3 bar codes for your country.
[7] Touch Main Menu.

Code 39 Full ASCII Mode


[1] Touch:
• Key Operator Functions
• System Configuration
• Bar Code Configuration
• Bar Code Programming
• Code 39 Full ASCII Mode

Code 39 Full ASCII Mode

Important
• Some sites might want you to scan the Off bar code for Code 39 Full ASCII Mode.
• You must check with the Site Administrator for the correct bar code to scan.
[2] Use the bar-code scanner to scan the On bar code.
[3] Touch Back.

Automatic Trigger
[1] From the Main Menu, click:
• Key Operator Functions
• System Configuration
• Bar Code Configuration
• Bar Code Programming
• Automatic Trigger

Automatic Trigger
Manual Trigger

8G7613 – 16APR10 69
MODIFICATION INSTRUCTIONS, TYPE 1

Note
• The bar codes for Automatic Trigger and Manual Trigger display:
– Automatic Trigger: The bar-code scanner operates automatically.
– Manual Trigger: The bar-code scanner operates when the trigger is actuated.
• You must check with the Site Administrator for the correct bar code to scan.
[2] Use the bar-code scanner to scan the bar code corresponding with the choice of the SA.
[3] Click Back.

Beeper Mode
[1] Touch Beeper Mode.

Beeper Volume

Beeper Tone

Important
You must check with the Site Administrator for the correct bar code to scan.
[2] Use the bar-code scanner to scan the bar code corresponding with the choice of the SA.
[3] Touch Back.

70 16APR10 – 8G7613
Detailed Procedures for All CR Systems

Reread Delay

Important
• Reread delay is only functional in the Automatic Trigger mode.
• You must check with the Site Administrator for the correct bar code to scan.
[1] Touch Reread Delay.

Reread Delay

[2] Use the bar-code scanner to scan the bar code corresponding with the choice of the SA.
[3] Touch:
• Back
• Main Menu
[4] This procedure is complete. Click Previous View until you return to the applicable page in Section 2–6.

Doing the Setup for the 11-Digit Cassette ID

Important
The 11-digit cassette ID configuration adds an extra character to all bar codes to match the Carestream Health
cassette IDs
[1] From the Main Menu, click:
• Key Operator Functions
• System Configuration
• Bar Code Configuration
• Bar Code Configuration
[2] Select Add # to Cassette IDs to make them 11 digits.
[3] Click:
• Save Changes
• Bar Code Programming
• Prefix & Suffix
[4] Click Next until the corresponding bar codes are displayed.

8G7613 – 16APR10 71
MODIFICATION INSTRUCTIONS, TYPE 1

For Bar-Code Reader Set


Up as USB as Serial in
Bar-Code Configuration

For Bar-Code Reader


Set Up as PS2 to
USB Keyboard

For Bar-Code Reader


Set Up as PS2 Keyboard

Important
If you scan the bar codes on the monitor, you must scan the second Suffix screen for:
• PS2 to USB keyboard
• PS2 keyboard
[5] Use the bar-code scanner to scan the bar code that corresponds to your system configuration.
[6] Click Back.
[7] This procedure is complete. Click Previous View until you return to the applicable page in Section 2–6.

72 16APR10 – 8G7613
Detailed Procedures for All CR Systems

Doing the Calibration for the Luminance of the Monitor


[1] At the Main Menu, click:
• Key Operator Functions
• Monitor Configuration
[2] Connect the calibration puck into any of the USB connections on the front of the computer.
[3] Click Photometer Calibration.

Note
The Verilum screen appears.
[4] Click Measure Luminance Response for Gamma Correction.

Important
You must hold the calibration puck against the screen:
• lightly to prevent pressing on the touch screen.
• until the test is complete.
• for approximately 3 minutes.
[5] Hold the calibration puck at the center of the screen.
[6] Click Continue.
[7] Wait for the test to complete.
[8] Remove the calibration puck from the screen.
[9] Click:
• Close
• Exit
• Main Menu
[10] This procedure is complete. Click Previous View until you return to the applicable page in Section 2–6.

8G7613 – 16APR10 73
MODIFICATION INSTRUCTIONS, TYPE 1

Setting the Look Code

Important
• The Look Code is only available if the DIRECTVIEW EVP Plus Software option is enabled.
• The customer can obtain the Look Code by running the Look Selection Preference Tool.
• The Look Code is installed for the first time during the installation of this modification.
• You only have to do the installation for the Look Code if there is not a current backup on the user preference
diskette.
[1] Obtain the Look Code from the customer.
[2] Click:
• Main Menu
• Service Functions
• Image Processing Preferences

[3] Click:
• Quick Menu icon
• Enter Look Code
[4] Enter the Look Code.
[5] Click:
• Save Changes
• Main Menu
[6] This procedure is complete. Click Previous View until you return to the applicable page in Section 2–6.

74 16APR10 – 8G7613
Detailed Procedures for All CR Systems

Doing the Setup for Remote Management Services (RMS)

Important
• Some sites do not allow the operation of the RMS. If your CR System is an approved site for the RMS, you must
use this procedure to set it up.
• The software for the RMS is included with the Software Version 5.2.x.

[1] Do you need to set up the RMS on your system?


Yes No
a. Do the Modification Instructions and Software Continue with Step 2.
Installation Instructions for the Remote
Management Services for the Software Version
5.1, 7H2373.
b. Continue with Step 2.

[2] This procedure is complete. Click Previous View until you return to the applicable page in Section 2–6.

Installing the KODAK DIRECTVIEW Remote Operations Panel (ROP)


[1] Do the Software Installation Instructions for the 19 in. KODAK DIRECTVIEW Remote Operations Panel (ROP)
used on the KODAK DIRECTVIEW CR 825/850/950/975 Systems and the KODAK DIRECTVIEW Classic/Elite
CR Systems with Software Version 5.2, 8G7963.
[2] This procedure is complete. Click Previous View until you return to the applicable page in Section 2–6.

Installing the Other Remote Devices

Important
The customer is responsible for installing the remote access software. See the Remote Access Software Installation
Instructions, 6H6955.

[1] If necessary, use your USB service key jump drive to log on to the Service Menu.
[2] Are you installing other remote devices?
Yes No
Continue with Step 3. Advance to Step 8.

[3] At the Main Menu, click:


• Key Operator Functions
• System Configuration
• Remote Device Configuration

8G7613 – 16APR10 75
MODIFICATION INSTRUCTIONS, TYPE 1

[4] Click New Device.


[5] Enter the IP address for the new remote device.
[6] Click ROP or Remote Access Software.
[7] Click:
• Save Changes
• Main Menu
[8] This procedure is complete. Click Previous View until you return to the applicable page in Section 2–6.

Setting the Slow-Scan Correction Profile for the EHR Cassettes

Important
The slow-scan correction profile:
• increases the uniformity for EHR cassettes.
• must be enabled on CR Systems using the EHR cassettes.
[1] From the Main Menu, click:
• Service Functions
• Diagnostics
• Slow Scan Profiles
[2] Click:
• DISABLE SLOW SCAN PROFILES to
disable the slow-scan correction profile, or
• DOWNLOAD SLOW SCAN PROFILES to
enable the slow-scan correction profile
• Done
• Main Menu

[3] This procedure is complete. Click Previous View until you return to the applicable page in Section 2–6.

76 16APR10 – 8G7613
Detailed Procedures for All CR Systems

Completing the Installation

Important
If your system has an MCB board, the SSC software is not available on your system
[1] Is your system a CR 825M, 850M, 975 System, or Classic/Elite CR System?
Yes No
Advance to Step 6. Continue with Step 2.

[2] Did you download a new version of SSC software in Step 3 of Downloading the Software to the Boards on
Page 60?
Yes No
Continue with Step 3. Advance to Step 6.

[3] At the Main Menu, click:


• Service Functions
• Diagnostics
• Slow Scan Current Calibration

[4] Click START.


[5] Is PASSED displayed for the calibration results?
Yes No
Continue with a. Use the diagnostics for
Step 6. your CR System to correct
the problem.
b. Continue with Step 6.

[6] If desired, do the optional FE Checklist, 6F5825.


[7] Scan all cassette sizes.

8G7613 – 16APR10 77
MODIFICATION INSTRUCTIONS, TYPE 1

[8] Send a DisplayPro to all destinations. Click:


• Key Operator Functions
• Monitor Configuration
• Select Test Pattern
[9] Select Display Pro.
[10] Click:
• OK
• Print Image
[11] Visually inspect the top 2 rows of the DisplayPro pattern. You should see targets in both halves of every circle
pattern.
[12] If you do not see targets in both halves of every circle pattern at the CR System, do Doing the Calibration for
the Luminance of the Monitor on Page 73.
[13] Send an image to all destinations.
[14] If you do not see targets in both halves of every circle pattern at the destination:
• Calibration of the destination is necessary, or
• Image space at the destination does not match the CR System. For example, P-values are sent to a
destination expecting luminance values.
[15] At the PEC Information screen, check that the bar-code reader scans the information correctly.
[16] Check for correct operation:
• All equipment
• HIS/RIS
• Interlock switch
[17] This procedure is complete. Click Previous View until you return to the applicable page in Section 2–6.

Doing the Feedback for a Customer-Purchased Modification and the Software Refresh Option

Important
• If the part number for this kit was ordered from Service Parts Management and displays on your inventory, you
must feed back the part number to SCAN to have the part removed from your inventory.
• If you used the part number of this kit for feedback during an earlier installation, do NOT record the part number
of this kit for this installation.
[1] Record the following information for feedback to SCAN or CRM:
SCAN
Service Activity Installation Install
K-number System Code Serial No. Code CRM PSL Time (actual) Part No. Feedback
CR 825 5634 M68 SC5634M68 --- No
CR 850 4825 M68 SC4825M68 --- No
CR 950 4826 M68 SC4826M68 --- No
CR 975 1883 M68 SC1883M68 --- No
Classic CR 5258 M13 SC5258M13 --- No
Elite CR 5259 M13 SC5259M13 --- No

78 16APR10 – 8G7613
Publication History

[2] If you performed an installation of necessary hardware for a CR System that has software refresh (see Service
Bulletin, 7H9511), you must also feed back:
• MOD 51 for the installation of the new computer.
• MOD 52 for the installation of the 2 GB of RAM.
[3] This procedure is complete. Click Previous View until you return to the applicable page in Section 2–6.

Section 8: Publication History

Publication Publication
Date No. ECO No. Changed Pages File Name Notes
16APR10 8G7613 --- All 8g7613.fm Revision
27JUL09 8G7613 --- All 8g7613.fm Revision
23MAR09 8G7613 --- --- 8g7613.fm New Publication

8G7613 – 16APR10 79
DIRECTVIEW is a trademark of Carestream Health, Inc.
The Kodak trade dress and trademark are used under license from Kodak.

Printed in U.S.A. • 8g7613.fm

Carestream Health, Inc.


150 Verona Street
Rochester, NY 14608