Professional Documents
Culture Documents
CONTENTS
2 Introduction ________________________________________________________________________ 12
3 Safety _______________________________________________________________________________ 20
Rev 1.00 2
CX9Plus Installation, Operation & Service Manual
4.3 Vertical movement control switch & Collimator control switch ____________________________________ 28
Rev 1.00 3
CX9Plus Installation, Operation & Service Manual
5.11 Assembly of the X-ray tube, the Collimator and DAP meter _____________________________________ 47
6 Calibration __________________________________________________________________________ 86
Rev 1.00 4
CX9Plus Installation, Operation & Service Manual
8.2 Starting and Shutting down the Main unit ________________________________________________________ 141
8.3 Starting and Shutting down the Monitor Cart ____________________________________________________ 142
11 Function __________________________________________________________________________149
12 Xvision ____________________________________________________________________________157
Rev 1.00 5
CX9Plus Installation, Operation & Service Manual
13 Maintenance _____________________________________________________________________166
15 Specification ______________________________________________________________________175
Rev 1.00 6
CX9Plus Installation, Operation & Service Manual
Chapter 1
[CX9Plus]
Rev 1.00 7
CX9Plus Installation, Operation & Service Manual
This manual may not be reproduced in whole or in part without the written permission of ECORAY CO.,
LTD.
All illustrations in this manual are exemplary only, and may differ from the actual situation.
The present Operating Instructions describe a system with maximum configuration. The system
configuration chosen by you may not contain all options and functions described here.
For this system options, separate operating instructions may be available. They are supplied with the
system, provided that the system configuration includes the respective option. You will find a
corresponding reference to those operating instructions in the relevant sections of this document.
Rev 1.00 8
CX9Plus Installation, Operation & Service Manual
Each page of the document has revision letter at the bottom of the page. The revision letter identifies
Revision Date
Revision No. Change Description
MM DD, YYYY
1.2 Warranty
This warranty shall apply only to defects that arise within 12months of purchase.
This product could have a free of charge service if troubles are occurred during the warranty
period.
The warranty does not cover defects caused by abuse, mishandling, accidents or improper
installation.
- Improper use
In case of repaired during the warranted period, User shall advise the result of test to
manufacture.
Rev 1.00 9
CX9Plus Installation, Operation & Service Manual
The present document does not constitute a complete catalog of all safety measures necessary for the
operation of the respective medical equipment, since special operating conditions may require further
measures. However, it does contain instructions which must be observed in order to ensure the
personal safety of operating staff and patients as well as to avoid damage to property. These
▶ DANGER
DANGER INDICATES A HAZARDOUS SITUATION WHICH, IF NOT AVOIDED, WILL RESULT IN DEATH OR
SERIOUS INJURY.
▶ WARNING
▶ CAUTION
CAUTION INDICATES A POTENTIALLY HAZARDOUS SITUATION WHICH, IF NOT AVOIDED, MAY RESULT
IN MINOR OR MODERATE INJURY. IT MAY ALSO BE USED TO ALERT AGAINST UNSAFE PRACTICES.
▶ NOTE
NOTES ARE MERELY INFORMATIVE. ADDITIONAL USEFUL INFORMATION AND HINTS ARE PROVIDED
Rev 1.00 10
CX9Plus Installation, Operation & Service Manual
Chapter 2
Introduction
[CX9Plus]
Rev 1.00 11
CX9Plus Installation, Operation & Service Manual
2 Introduction
This equipment is image diagnosis system with radiology and it was controlled by high frequency type.
This equipment consists of Image Intensifier, CCD camera, Collimator, X-ray tube, X-ray generator (high
voltage transformer, generator) and “C” type of C-arm structure supporting image intensifier and
workstation (Computer & Monitor) for display unit with acquisition images, processing saving and
transmission. Also, it has accessories with hand switch and foot switch for X-ray exposure. Image
intensifier have a 9 inch with standard for purpose and image size and it has a 6 inch image intensifier
for optional and if image intensifier is bigger, could get more bigger image and could diagnosis.
High voltage which is generated from the inverter X-ray high voltage generator provides a positive
into the X-ray tube. At this time, after this generated X-ray irradiates and passes an object, residual
radiation reaches the image intensifier and makes X-ray image by chemical action of emitted light
photo. These are done imaging with CCD camera and what can transmit human anatomical structure
through the monitor in real time and simultaneously image is X-ray fluoroscopic or radiographic
equipment.
The CX9 Plus is intended for use in mobile applications, by a qualified/trained doctor or technician on
both adult and pediatric subjects for taking diagnostic radiographic exposures of body parts.
Rev 1.00 12
CX9Plus Installation, Operation & Service Manual
2.3 Features
This equipment is used in operation room, emergency room and other clinical purpose and it has
features as follows.
High frequency inverter type x-ray generator, it could get stability x-ray output and low
voltage ripple.
Dual focal X-ray Tube with high level heating unit to get more good image and high value x-
ray output.
The exposure of humans to ionizing radiation must always be medically justified. Especially on
pregnant women, children and adolescents this procedure should be used with caution or be avoided
altogether. However, the final decision lies with the attending physician or attending surgeon.
2.5 Options
The present Operating Instructions describe a system with maximum configuration. The system
configuration chosen by you may not contain all options and functions described here.
The following options can be integrated into the system upon request:
Wireless LAN
PACS Software
2.6 Training
After installation, ECORAY engineer or ECORAY’s agent will carry our training of its use and if you have
any further questions on the system, please contact address as shown in this manual.
Rev 1.00 13
CX9Plus Installation, Operation & Service Manual
This X-ray system must be used by regulations after reading and understanding relative
radiation regulations.
Regulations on its use cover the area of only diagnostic radiography applications.
This system is classified as follows, based on safety standards for electrical medical equipment.
Class I equipment
No applied part
Operation Mode
Rev 1.00 14
CX9Plus Installation, Operation & Service Manual
ISO 13485
IEC 60601-1
IEC 60601-2
IEC 60601-1-3
IEC 60601-1-6
IEC 60601-2-28
IEC 60601-2-54
Imaging performance
Item Details
Age No limitation
Weight No limitation
Nationality No limitation
Rev 1.00 15
CX9Plus Installation, Operation & Service Manual
Item Details
Sex No limitation
Nationality No limitation
Necessary.
Others Person who can confirm the safety by looking ahead of the
equipment
Rev 1.00 16
CX9Plus Installation, Operation & Service Manual
Symbol Description
included.
Made Date : The made date of the product shows with this
symbol.
Rev 1.00 17
CX9Plus Installation, Operation & Service Manual
Labeling: Product Name, Model, Main power, manufactured date, Manufacturer & Address, Serial No.
Rev 1.00 18
CX9Plus Installation, Operation & Service Manual
Chapter 3
Safety
[CX9Plus]
Rev 1.00 19
CX9Plus Installation, Operation & Service Manual
3 Safety
3.1 General safety
▶ WARNING
YOU MUST BE FAMILIAR WITH THE CONTENTS OF THE PRESENT OPERATING INSTRUCTIONS IN
ORDER TO BE ABLE TO OPERATE THE SYSTEM AS INTENDED. READ AND UNDERSTAND THE PRESENT
▶ WARNING
THE RESPONSIBILITY FOR MANAGEMENT OF USE AND MAINTENANCE OF MEDICAL EQUIPMENT LIES
IN A USER.
▶ WARNING
▶ WARNING
EXTERNAL INTERFACES OF THE UNIT MUST FULFILL THE REQUIREMENTS OF STANDARD IEC 60601-1.
Operation
Only authorized personnel are allowed to assemble the system and to provide technical service.
▶ WARNING
Rev 1.00 20
CX9Plus Installation, Operation & Service Manual
▶ WARNING
THE SYSTEM IS NOT AN AP DEVICE. ITS USE IN AN ATMOSPHERE CONTAINING COMBUSTIBLE GAS
▶ WARNING
DO NOT USE DISINFECTANT SPRAYS THAT ARE FLAMMABLE OR COULD CAUSE AN EXPLOSION. THE
▶ WARNING
The system produces X-rays. If you do not observe the safety measures and
precautions required by your local radiation protection regulatory body, these X-rays
can be hazardous both to operating staff and other persons within the radiation zone
of occupancy.
Many kinds of researches on effects or influences about X-radiation provide basics about maximum
permissible dose. These research results were used for ICRP to recommend Maximum Permissible
Dose but it is hard to define exactly and it is updated from time to time with new research result. For
occupational exposures, the ICRP 60 recommends that the MPD accumulated shall not exceed the
following limits:
▶ WARNING
THE SYSTEM IS INTENDED FOR PROCEDURES WHERE THE SKIN DOSE MAY BE SO HIGH THAT THERE
Rev 1.00 21
CX9Plus Installation, Operation & Service Manual
▶ WARNING
THE SYSTEM MAY ONLY BE OPERATED BY PERSONNEL WHO HAS UNDERGONE RADIOLOGICAL
TRAINING.
▶ WARNING
BE OBSERVED.
▶ WARNING
IN ORDER TO AVOID UNINTENTIONAL RADIATION, THE FOOT SWITCH MUST BE HUNG UP ON THE
FOOT SWITCH SUPPORT WHEN THE SYSTEM IS SWITCHED ON, BUT NOT IN USE.
Protection of staff
Staff members who stay within the radiation controlled area must wear X- Protection of staff ray
protective clothing.
To minimize the radiation burden of the patient, you must keep the source/skin distance as large as
Thyroid collar
Medical electrical equipment requires special precautionary measures with respect to EMC and must
be installed and put into service in accordance with the EMC guidelines contained in the
accompanying documents.
Portable and mobile RF communications equipment may interfere with medical electrical equipment.
Rev 1.00 22
CX9Plus Installation, Operation & Service Manual
All operating modes of the system have been considered in the EMC tests. There are no exceptions to
the rules.
Only conductors, connecting cables and accessories that are specified by the manufacturer may be
used.
▶ WARNING
The meaning of the symbol on the product, its accessory or packaging indicates
that this product shall not be treated as household waste. Please, dispose of this
equipment at your applicable collection point for the recycling of electrical &
electronic equipment waste. In the European Union and Other European countries
which there are separate collection systems for used electrical and electronic
product. By ensuring the correct disposal of this product, you will help prevent
potentially hazardous to the environment and to human health, which could otherwise be caused by
unsuitable waste handling of this product. The recycling of materials will help conserve natural
resources, Please do not therefore dispose of your old electrical and electronic equipment with your
household waste. For more detailed information about recycling of this product, please contact your
local city office, your household waste disposal service or the shop where you purchased the product.
ECORAY
TEL (+82)-62-952-3251
FAX (+82)-62-951-5261
e-mail x-ray@eco-ray.co.kr
Homepage www.eco-ray.co.kr
Rev 1.00 23
CX9Plus Installation, Operation & Service Manual
The system's service life is 7 years, and is limited to systems that receive the prescribed maintenance
checks.
ECORAY is only responsible for the safety of its products when their maintenance, repair
and/or modification has been carried out by ECORAY or by personnel expressly authorized to
do so by ECORAY itself.
ECORAY shall not be held responsible in any way for malfunction, damage and/or danger due
The user of the installation where the system is connected is responsible for making sure that
the installation itself is only used by suitably trained and qualified operators.
Rev 1.00 24
CX9Plus Installation, Operation & Service Manual
Chapter 4
System Description
[CX9Plus]
Rev 1.00 25
CX9Plus Installation, Operation & Service Manual
4 System Description
4.1 C-arm stand
CCD Camera PC. (Sends the acquired images from Image Intensifier
to the PC)
Image
X-ray image receptor
Intensifier
Generator Boosts the input voltage at the x-ray generator and then
Rev 1.00 26
CX9Plus Installation, Operation & Service Manual
Emergency
Immediately stops operation of the equipment in time of
Stop
emergency.
Switch
Connection for The connection between the Main Unit and the Hand
Power supply
Power input of the equipment.
Connection
Connection for The connection between the Main Unit and the Foot
Coupling cable The connection between the Generator and the X-ray
Connection Tube.
Connection for
The connection between the Generator and the Monitor
Generator
Cart
power
Rev 1.00 27
CX9Plus Installation, Operation & Service Manual
Vertical device
Adjusts the vertical movement of the C-arm
Up/Down
Collimator
Adjusts the horizontal movement of the Collimator
horizontal S/w
Collimator
Adjusts the vertical movement of the Collimator
vertical S/w
Collimator
Adjusts the rotational movement of the Collimator
rotation S/w
Rev 1.00 28
CX9Plus Installation, Operation & Service Manual
4.4 Console
Radiography mode
1 Lights on when it is selected
Button
Fluoroscopy mode
2 Lights on when it is selected
Button
Radiation Timer
4 Makes Radiation Timer reset
Reset Button
Power on/off
7 On and off the power input
Button
IMAGE INTENSIFIER
9 Selects the field of IMAGE INTENSIFIER
select Button
X-ray exposure
11 Performs the X-ray exposure
Button
12 kV, mA, mAs Adjusts the value of loading factor: kV, mA,
Rev 1.00 29
CX9Plus Installation, Operation & Service Manual
Rev 1.00 30
CX9Plus Installation, Operation & Service Manual
4.6 Workstation
Power supply
Connects the power cord.
Connection
Rev 1.00 31
CX9Plus Installation, Operation & Service Manual
The CX9PLUS is equipped with the hand switch and foot switch.
Hand Switch
Foot Switch
Radiation is initiated either with the hand switch or the foot switch.
Rev 1.00 32
CX9Plus Installation, Operation & Service Manual
The external connection LAN connector is used for sending obtained image data to external
computers
Insert the cable into the LAN connector, connect the cable to the external device, and transfer data in
DICOM
▶ CAUTION
DO NOT CONNECT THE LAN CONNECTOR TO EXTERNAL PERIPHERALS NOT SPECIFIED BY ECORAY
DOING SO COULD CAUSE UNIT FAILURE, MALFUNCTION, SMOKE, OVERHEAT, ELECTRIC SHOCK, OR
On the monitor, the live screen function has been factory-assigned to the left screen, and the
Live screen: Live images and saved images, both at full size.
Rev 1.00 33
CX9Plus Installation, Operation & Service Manual
Chapter 5
Installation of CX9PLUS
[CX9Plus]
Rev 1.00 34
CX9Plus Installation, Operation & Service Manual
5 Installation of CX9PLUS
5.1 Conditions for installation
Use environment
(The upper bound of the temperature is allowed up to 50℃ only for the period of less than one
month.)
Power Supply
AC power
Frequency : 50/60Hz
Rev 1.00 35
CX9Plus Installation, Operation & Service Manual
Screw drivers
Nipper
Pliers
Tester
Cutter
Waste cloth
Alligator clip
Oscilloscope
Rev 1.00 36
CX9Plus Installation, Operation & Service Manual
5.4 Unpacking
• Inspect the pack for evidence of shipping damage. If there is evidence of shipping damage,
▶ NOTE
OPEN THE PACK CAREFULLY. SHARP TOOLS MAY DAMAGE THE CONTENS.
▶ CAUTION
ONE PERSON SHOULD NOT ATTEMPT TO LIFT OR MOVE THIS ASSEMBLY WITHOUT PROPER
EQUIPMENT OR ASSISTANCE.
Rev 1.00 37
CX9Plus Installation, Operation & Service Manual
• Unpack the manuals and any other paperwork that may be packed with the units.
• Keep the shipping containers. In case of shipping damage, place the unit back in its
shipping pack and notify the carrier and the customer support department.
Image Intensifier
Camera
Collimator
Workstation
Monitor
Monitor cart
Hand switch
Foot switch
Rev 1.00 38
CX9Plus Installation, Operation & Service Manual
Rev 1.00 39
CX9Plus Installation, Operation & Service Manual
Rev 1.00 40
CX9Plus Installation, Operation & Service Manual
• Install the LAN Card for CCD Camera to the LAN Connector as below.
Rev 1.00 41
CX9Plus Installation, Operation & Service Manual
X-ray image intensifier are operated with 24±1Vdc input voltage applied to the power supply. Wiring
Rev 1.00 42
CX9Plus Installation, Operation & Service Manual
NO
1 DC24V Input
2 NC
3 0V
4 M1
5 M2
Glass Implosion
The X-ray image intensifiers (IMAGE INTENSIFIER.) are a high-vacuumed electron tube containing glass.
Do not scratch or strain the glass of the IMAGE INTENSIFIER. for the prevention of the implosion.
Because of the risk of implosion, a protection mask shall be put on and extreme care shall be
The aluminum plate (for light-shielding) on the entrance plane of housing also plays as a
protection against the injury due to implosion. When removing the plate, the input side shall
not face a person. The protection plate such as anti-scattering grid is recommended when
The distance between the surface of output window and lens attaching plate shall be confirm
due to the data sheet or the specification when the object lens is mounted for the prevention
Rev 1.00 43
CX9Plus Installation, Operation & Service Manual
Rev 1.00 44
CX9Plus Installation, Operation & Service Manual
NO NO
1 O TXVNA 2 O TXVNB
3 O TXVNC 4 O TXVND
5 O Shielding 6 O TXVPA
7 O TXVPB 8 O TXVPC
9 O TXVPD 10 O Shielding
Pin Pin
I/O Signal I/O Signal
NO NO
3 O GND 4 I GND
NO NO
1 I LIH 2 I Fod
5 O Reserved 6 O Reserved
7 O XR-Low 8 O XR-High
9 I GND-EXT 10 I GND-EXT
Pin Pin
I/O Signal I/O Signal
NO NO
1 O VCC 2 O SCL
3 O GND 4 Reserve
5 IO SDA 6 Shielding
Rev 1.00 45
CX9Plus Installation, Operation & Service Manual
Phase
C2 Blue 2.4A 0.4A 62Ω (C-S) 30uF
shift
Rev 1.00 46
CX9Plus Installation, Operation & Service Manual
5.11 Assembly of the X-ray tube, the Collimator and DAP meter
• Fix the X-ray tube and the Collimator to the Support of the C-Frame.
Rev 1.00 47
CX9Plus Installation, Operation & Service Manual
Rev 1.00 48
CX9Plus Installation, Operation & Service Manual
5.12 Wiring
Rev 1.00 49
CX9Plus Installation, Operation & Service Manual
• Select and open the <PRO Win32.exe>file at the [Intel Pro 1000 PCI LAN Driver] folder.
Rev 1.00 50
CX9Plus Installation, Operation & Service Manual
Rev 1.00 51
CX9Plus Installation, Operation & Service Manual
• Select the 《Advanced Network Service》 at the Setup Options, Click【Next >】.
Rev 1.00 52
CX9Plus Installation, Operation & Service Manual
• Connect the Communication Cable between the Camera and the PC.
Rev 1.00 53
CX9Plus Installation, Operation & Service Manual
▷ Solution
• Check the status of the connection between the LAN port of the PC and the LAN port of the
Camera.
The PC does not recognize the Camera signal - Connected normally, but if it does not recognize the
signal.
Rev 1.00 54
CX9Plus Installation, Operation & Service Manual
▷ Solution
• Click icon
• Click 【Properties】.
Rev 1.00 55
CX9Plus Installation, Operation & Service Manual
• Click 【OK】
Rev 1.00 56
CX9Plus Installation, Operation & Service Manual
• Reset the “Local Area Connection Status” if the monitor is displayed as follows
▷ Solution
Rev 1.00 57
CX9Plus Installation, Operation & Service Manual
Rev 1.00 58
CX9Plus Installation, Operation & Service Manual
• Check the set-up status of ““Intel(R) Pro/1000 GT Desktop Adapter” of the “Network adapters”
Rev 1.00 59
CX9Plus Installation, Operation & Service Manual
Rev 1.00 60
CX9Plus Installation, Operation & Service Manual
• Select the 《Intel(R) PRO/1000 GT Desktop Adapter》 at the “Korba Installer”, Click 【Apply】
Rev 1.00 61
CX9Plus Installation, Operation & Service Manual
Rev 1.00 62
CX9Plus Installation, Operation & Service Manual
Rev 1.00 63
CX9Plus Installation, Operation & Service Manual
• Select the “Jumbo Frame” at the “Property”, select the “9014 Bytes” at the “Value”
• Click 【OK】
Rev 1.00 64
CX9Plus Installation, Operation & Service Manual
Rev 1.00 65
CX9Plus Installation, Operation & Service Manual
Rev 1.00 66
CX9Plus Installation, Operation & Service Manual
• Click 【Finish】.
Rev 1.00 67
CX9Plus Installation, Operation & Service Manual
Rev 1.00 68
CX9Plus Installation, Operation & Service Manual
• Click【 Exit】.
Rev 1.00 69
CX9Plus Installation, Operation & Service Manual
• Click【Next>】.
Rev 1.00 70
CX9Plus Installation, Operation & Service Manual
• Click【Finish】.
Rev 1.00 71
CX9Plus Installation, Operation & Service Manual
• Click【Next>】.
Rev 1.00 72
CX9Plus Installation, Operation & Service Manual
• Click【OK】.
desktop.
Rev 1.00 73
CX9Plus Installation, Operation & Service Manual
Run “XVision”
Login
Rev 1.00 74
CX9Plus Installation, Operation & Service Manual
• “Configuration” is displayed.
• Click 【WORKSTATION】
• Click 【New】
Rev 1.00 75
CX9Plus Installation, Operation & Service Manual
• “ID” is case-sensitive
• Click 【OK】
• Click 【Apply】
• Click 【Exit】
Rev 1.00 76
CX9Plus Installation, Operation & Service Manual
• Click 【Config.】
• Click 【DICOM】
• 【DICOM】 is configured “Local” Tab, “Worklist Server” Tab, “DICOM Printer” Tab and “PACS
Server” Tab.
Rev 1.00 77
CX9Plus Installation, Operation & Service Manual
• Click 【DICOM】
• Click 【APPLY】
• Click 【EXIT】
Rev 1.00 78
CX9Plus Installation, Operation & Service Manual
- “Work list Server" Tab is function to check the status of the connection between the product and the
Work list.
• Click 【DICOM】
• “Work list Server” Tab is configured “Remote AE”, “Remote IP”, “Remote Port” and 【Test】
• Remote AE(Application Entity) : Enter the name of the Work list Server.
• Remote Port : Enter the port number of the Work list Server.
• Click 【Test】
• When it is abnormal, check the AE, IP, Port of the “Work list Server” and then enter correctly.
Rev 1.00 79
CX9Plus Installation, Operation & Service Manual
• Click 【APPLY】
• Click 【EXIT】
Rev 1.00 80
CX9Plus Installation, Operation & Service Manual
• Click 【DICOM】
• 【Test】 : Connection status between the product and the Printer is confirmed.
• Click 【New】
• “Setting” is displayed.
Rev 1.00 81
CX9Plus Installation, Operation & Service Manual
• To specify the priority printer,, select the check box at the “DICOM Print Priority”. (If multiple
• Click 【OK】
Rev 1.00 82
CX9Plus Installation, Operation & Service Manual
• Click 【DICOM】
• 【Test】 : Connection status between the product and the PACS Server is confirmed.
Rev 1.00 83
CX9Plus Installation, Operation & Service Manual
• Click 【New】
• To specify the priority PACS Server, select the check box at the “Priority”.
• Click 【OK】
Rev 1.00 84
CX9Plus Installation, Operation & Service Manual
Chapter 6
Calibration
[CX9Plus]
Rev 1.00 85
CX9Plus Installation, Operation & Service Manual
6 Calibration
6.1 Generator Calibration
6.1.1 Overview
This Generator calibration must be performed only by the calibration software program of ECORAY.
• It has functions for changing X-ray exposure parameters (e.g. kV, mA and selection of X-ray
X-ray mA meter
Rev 1.00 86
CX9Plus Installation, Operation & Service Manual
After product
Calibration for X-ray
installation, including Refer to the chapter 5
exposure parameters
relocation
required)
Replacement or repair of critical components requires a calibration for X-ray exposure parameters.
ECO-XR5-MiO Board
ECO-XR5-CnD Board
ECO-XR5-FnD Board
Rev 1.00 87
CX9Plus Installation, Operation & Service Manual
6.2.1 Specifications
• Confirmation of PC specifications
• The software for RS232 Driver is available on the website and the installation method
6.2.2 Installation
• Turn on PC.
• Click “Properties”.
Rev 1.00 88
CX9Plus Installation, Operation & Service Manual
Rev 1.00 89
CX9Plus Installation, Operation & Service Manual
• On the displayed “Advanced Setting for COM1”, select “Use FIFO buffers(requires 16550
compatible UART)”.
▶ NOTE
If you do not select a one from COM1 to COM5, this program would not operate.
Rev 1.00 90
CX9Plus Installation, Operation & Service Manual
6.3.1 Specifications
• Identification : ECO-GEN-CALIBRATION-PC-M
• Confirmation of PC specifications
6.3.2 Installation
• Turn on PC.
• On the “Devices with Removable Storage” pane, double click “ECORAY Calibrate”.
Rev 1.00 91
CX9Plus Installation, Operation & Service Manual
Rev 1.00 92
CX9Plus Installation, Operation & Service Manual
Rev 1.00 93
CX9Plus Installation, Operation & Service Manual
Rev 1.00 94
CX9Plus Installation, Operation & Service Manual
6.4.1 PC
Rev 1.00 95
CX9Plus Installation, Operation & Service Manual
Rev 1.00 96
CX9Plus Installation, Operation & Service Manual
Rev 1.00 97
CX9Plus Installation, Operation & Service Manual
Rev 1.00 98
CX9Plus Installation, Operation & Service Manual
• Click “Port Check”, and confirm that the selected communication port is a suitable one from
COM1 to COM5.
▶ NOTE
If the communication port is not selected in normal range, “Power ON” button will not operate.
Port Check
Rev 1.00 99
CX9Plus Installation, Operation & Service Manual
• If the communication port is not selected in normal range, change the selection of “USB to RS232
Converter”.
• Possible select X-ray exposure parameters (e.g. kV, mA and selection of X-ray tube).
manufacturing process)
functions spot.
• Possible select the capacity of generator and the filament preheat voltage.
function
failure description.
window part
tube, focal spot ray tube which is selected on the setup part.
program.
exposure
parameters
6.7.1 Power ON
• Then, it indicates 52 of data signal (input/output) address number (see No 2 of the above fig).
• Select min. and max. kV on the “Small Focal (S.kV)” (see No 1 of the above fig).
• Select min. and max. kV on the “Large Focal (L.KV)” (see No 1 of the above fig).
• Select min. and max. mA on the “Small Focal (S.mA)” (see No 2 of the above fig).
• Select min. and max. mA on the “Large Focal (L.mA)” (see No 2 of the above fig).
• Select filament preheat voltage for small and large focal spot (see No 3 of the above fig).
▶ NOTE
It is recommended that the filament preheat voltage selected by the manufacture does not change.
• Click “Store” for saving input data (see No 4 of the above fig).
▶ NOTE
Using the stored data of MiO Board (Main Input & Output Board) of generator, it can execute
“Calibration Window Part” directly on “Show Calibration Window” without any selection of capacity of
generator.
• The “S/N” indicates year, month and identification for calibration data (see No 1 of the above fig).
• The indication of “No 2” is only used in the manufacturing process (see No 2 of the above fig).
• The “Slice” indicates the number of X-ray exposures (see No 3 of the above fig).
• The “HV, BC” indicates the DAC values of kV and mA (see No 4 of the above fig). HV means kV of
• The “kV, mA” indicates the detected values of kV and mA on a high voltage transformer (see No 5
• The indication of “No 6” indicates the fixed 4.1 for calibration (see No 6 of the above fig).
• The “Reset” is only used in the manufacturing process (see No 7 of the above fig).
• The indication of “No 8” indicates the firmware version of MiO board (see No 8 of the above fig).
▶ NOTE
• The “Store” indicates the function for saving data (see No 9 of the above fig).
• The indication of “No 1” indicates the state of product (see No 1 of the above fig).
• The indication of “No 2” indicates the state of other components and functions (see No 2 of the
above fig). When it is connected in normal, green indicator is lighting. When it is not connected in
• The “System State” indicates “Now Data Access” (see No 5 of the above fig).
• When it indicates “Now Data Access”, the calibration data of MiO board move to a calibration
• Note : MiO board means a “Main Input & Output Board” of generator
• When this calibration data has been moved, data table indicates the calibration data (see No 7 of
• Select a suitable X-ray tube and focal spot (see No 2 of the above fig).
• Check the selected X-ray tube and focal spot (see No 6 of the above fig).
• The ““Data Clear” and “Current Focal Load” are only used in the manufacturing process (see No 3
• Click “Store” for saving input data (see No 4 of the above fig).
• Click the No 5 related to No 4 for min. kV and No 3 for min. mA (see No 3, 4, 5 of the above fig).
• When click the No 5, the colume changes to yellow colour. (see No 5 of the above fig).
• Note : Under pushing “Ctrl” on keyboard for “Ready”, pushing “X” on keyboard for “X-ray
exposure”
• When the detected kV and mA are not incorrect, execute “X-ray exposure” again and change data.
• Select 40 kV → Select 10mA → Execute “X-ray exposure” → Check the measured dose on the
• Select 40 kV → Select 25mA → Execute “X-ray exposure” → Check the measured dose on the
• Select 40 kV → Select 50mA → Execute “X-ray exposure” → Check the measured dose on the
• Select 40 kV → Select 75mA → Execute “X-ray exposure” → Check the measured dose on the
• Select 40 kV → Select 100mA → Execute “X-ray exposure” → Check the measured dose on the
- For sequence of calibration At 60, 80, 100, 125, 150 kV, perform the same sequence at 40 kV.
- For sequence of calibration at 60, 80, 100, 125, 150 kV, perform the same sequence at 40 kV.
• Select 40 kV → Select 150mA → Execute “X-ray exposure” → Check the measured dose on the
- For sequence of calibration at 60, 80, 100, 125, 150 kV, perform the same sequence at 40 kV.
• The “No 1” indicates X-ray exposure parameters (kV, mA and exposure time) for radiography
• The “No 5” indicates selectable function for msec and mAs (see No 5 of the above fig).
▶ NOTE
Under pushing “Ctrl” on keyboard for “Ready”, pushing “X” on keyboard for “X-ray exposure”
6.9.1 Power ON
• Then, it indicates 52 of data signal (input/output) address number (see No 1 of the above fig).
• Select min. and max. kV on the “Small Focal (S.kV)” (see No 1 of the above fig).
• Select min. and max. kV on the “Large Focal (L.KV)” (see No 1 of the above fig).
• Select min. and max. mA on the “Small Focal (S.mA)” (see No 2 of the above fig).
• Select min. and max. mA on the “Large Focal (L.mA)” (see No 2 of the above fig).
• Select filament preheat voltage for small and large focal spot (see No 3 of the above fig).
▶ NOTE
It is recommended that the filament preheat voltage selected by the manufacture does not
change.
• Click “Store” for saving input data (see No 4 of the above fig).
▶ NOTE
Using the stored data of MiO Board (Main Input & Output Board) of generator, it can execute
“Calibration Window Part” directly on “Show Calibration Window” without any selection of
capacity of generator.
9
1
8
2
3 7
4
6
5
• The “S/N” indicates year, month and identification for calibration data (see No 1 of the above
• The indication of “No 2” is only used in the manufacturing process (see No 2 of the above
fig).
• The “Slice” indicates the number of X-ray exposures (see No 3 of the above fig).
• The “HV, BC” indicates the DAC values of kV and mA (see No 4 of the above fig). HV means
• The “kV, mA” indicates the detected values of kV and mA on a high voltage transformer (see
• The indication of “No 6” indicates the fixed 4.1 for calibration (see No 6 of the above fig).
• The “Reset” is only used in the manufacturing process (see No 7 of the above fig).
• The indication of “No 8” indicates the firmware version of MiO board (see No 8 of the above
fig).
• Note : MiO board means a “Main Input & Output Board” of generator
• The “Store” indicates the function for saving data (see No 9 of the above fig).
• The indication of “No 1” indicates the state of product (see No 1 of the above fig).
• The indication of “No 2” indicates the state of other components and functions (see No 2 of
the above fig). When it is connected in normal, green indicator is lighting. When it is not
• The “System State” indicates “Now Data Access” (see No 5 of the above fig).
• When it indicates “Now Data Access”, the calibration data of MiO board move to a calibration
Note : MiO board means a “Main Input & Output Board” of generator
• When this calibration data has been moved, data table indicates the calibration data (see No
• Select a suitable X-ray tube and focal spot (see No 2 of the above fig).
• Check the selected X-ray tube and focal spot (see No 6 of the above fig).
• The ““Data Clear” and “Current Focal Load” are only used in the manufacturing process (see
• Click “Store” for saving input data (see No 4 of the above fig).
3 2
6 1
• Click the No 5 related to No 4 for min. kV and No 3 for min. mA (see No 3, 4, 5 of the above
fig).
• When click the No 5, the colume changes to yellow colour. (see No 5 of the above fig).
▶ NOTE
Under pushing “Ctrl” on keyboard for “Ready”, pushing “X” on keyboard for “X-ray exposure”
• When the detected kV and mA are not incorrect, execute “X-ray exposure” again and change
data.
- Select 40 kV → Select 6mA → Execute “X-ray exposure” → Check the measured dose
• For sequence of calibration at 100 kV, perform the same sequence at 40 kV.
• For sequence of calibration at 120 kV, perform the same sequence at 40 kV.
• For sequence of calibration at 100 kV, perform the same sequence at 40 kV.
• For sequence of calibration at 125 kV, perform the same sequence at 40 kV.
• For sequence of calibration at 150 kV, perform the same sequence at 40 kV.
• The bolts of the displayed portion are removed, and then remove the cap of the Camera
Camera Part
Iris Part
• Release the fixed state by turning the Iris Adjustment knob to the left.
• Execute the X-ray exposure at the Fluoroscopy mode by pressing on the foot switch.
• Release the fixed state by turning the Focus Adjustment knob to the left.
• Adjust to fit the focus while watching the image displayed on the Live Screen.
• “Configuration” is displayed.
• Click 【DSP】
• Click 【CONTROL】
▶ NOTE
Shading Correction
Circular Blanking
Look up table
Enhancement
Overlay
Memory
• Image Store : Used to save the image of the Last Image Hold
• Auto Level : Adjusting automatically the gain value of the Live image.
• Threshold
Signal analysis
• Display Measure Frame : To activate the area of the detector at ABC, select this option.
- V Stretch : Adjusting the image smoothness of the vertical axis of the Circle detector.
Test
• Pattern : Selecting the image output signal of the Camera during the Camera test.
Chapter 7
Mechanical Handling
[CX9Plus]
7 Mechanical Handling
- The Orbital rotation fixing bar used to lock or release the movement the C-arm.
• Stop the movement of the C-arm in the best point by locking orbital rotation fixing bar.
• Function : Rotates the C-arm with the left and right side.
- The Arm rotation fixing bar used to lock or release the movement the C-arm.
• Stop the movement of the C-arm in the best point by locking Arm rotation fixing bar.
• Panning motion: The C-arm move to the left and right side per 12.5⁰.
- The Panning motion fixing bar used to lock or release the movement of the FR arm.
• Stop the movement of the C-arm in the best point by locking Panning motion fixing bar.
- The FR Arm fixing bar used to lock or release the movement of the C-arm.
• Stop the movement of the C-arm in the best position by locking FR Arm fixing bar.
• Levers in upper position : the wheels are locked, the brake is locked.
• Levers in lower position : the wheels can move freely, the brake is released.
7.8 Collimator
• Do not radiate as the X-ray coverage larger than the exposure area.
Chapter 8
[CX9Plus]
Before switching on the system, e.g. after a transport, you must connect the two system components
to each other and also connect the entire system to the power supply.
• Plug the coupling cable connector into the socket located at the side of the C-arm stand and
• Make sure that a suitable supply voltage is available and that the socket-outlet is properly
• Check the power plug on the monitor cart power cable and the socket-outlet for
compatibility.
• Make sure that the inclination of the system does not exceed 5° from the level in operating
position.
The main unit has power switches for turning the system on or off.
The system turns off. Any unsaved patient data and/or images are lost.
There is an EMERGENCY STOP button on the main unit, enabling you to turn off all electrical functions
The EMERGENCY STOP button is locked in this position. The system turns off immediately. All electrical
• To unlock the EMERGENCY STOP button, press the red knob with a slight twist in clockwise direction.
The unit remains switched off. Press the On switch on the main unit, you can turn on the power again.
Chapter 9
Fluoroscopy Operation
[CX9Plus]
9 Fluoroscopy Operation
9.1 Steps
To perform a fluoroscopy, you must always make the following steps and settings:
Chapter 10
Radiography Operation
[CX9Plus]
10 Radiography Operation
10.1 Steps
To perform a radiography, you must always make the following steps and settings:
Chapter 11
Function
[CX9Plus]
11 Function
11.1 Selecting the ABC
• If the switch is pressed continuously, the icon is selected in the following order.
• The Radiation Time display shows the accumulated radiation time in minutes and seconds.
• If the radiation time is sustained for more than 5 minutes, the X-ray radiation is stopped.
• The ‘beep’ sounds from the console during X-ray radiation, the ‘beep’ is shown below.
Five minutes elapsed : The ‘beep’ stop, the X-ray radiation is stopped.
• If you want to radiate the stopped x-ray, press the "Time reset" switch.
If the Mute Button is pressed again, the ‘beep’ sounds from the console.
While radiation is active, the current fluoroscopic image is displayed on the live screen.
When you terminate radiation (by releasing the hand or foot switch), the last fluoroscopic image is
displayed on the live screen and as a thumbnail on the control panel (Last Image Hold).
During the exposure (fluoroscopy or direct radiography), the yellow radiation warning lamp on the
The dose area product meter attaches to the front of the collimator and displays the dose-area
After turning the power ON, wait 3 minutes for the dose area product meter to stabilize.
• Perform exposure.
Display Unit
Chapter 12
Xvision
[CX9Plus]
12 Xvision
12.1 Main Screen
The Xvision software offers unified user interface for operating CX9PLUS.
Main menu and information window are located on the right side of the software.
Main image window is located on the center side and the Thumbnail window is located on lower right
side.
Function Description
Monitor2 Image
Thumbnail, Total, Shutdown, Exit Program
Information
• If you want to display the other image without saving, Press the Foot Switch.
• Click 【Capture】
• The folder is created within the [c: \ Program files \ XVision \ Image] folder.
• The LIH Image is saved there with patient information as DICOM file
• Click 【Review】
• Enter the search object, Can retrieve information for the item(s) you selected.
• Click 【Search】
• The patient information and image information are displayed at the top of screen.
• Click 【Capture】 at the current screen, Save the Live Image at the Workstation.
Adjusts window
Initialize all settings
width and level
Chapter 13
Maintenance
[CX9Plus]
13 Maintenance
To ensure satisfactory operation, CX9PLUS requires periodic planned maintenance and routine user
checks. These are essential to keep the equipment operating safely, effectively and reliably. Planned
maintenance may only be carried out by qualified and authorized service technicians. In this context,
qualified means those qualified to work on this type of medical electrical equipment in the jurisdiction
in which the equipment is being used, and authorized means those authorized by the user of the
equipment. CX9PLUS is able to provide a full planned maintenance and repair service on either a call
Check the following functions or perform the following maintenance work at the specified time
intervals:
Yearly
on the console.
Clamping levers
of
Check the Grease thread with extreme pressure lubricant
mechanical
brakes
Motorized
Inspect bearings for damage
vertical travel*
Monthly
Radiation warning
Must be lit during radiation
lamp
C-arm positions
Motorized
Move up and down to respective end stops
vertical travel
The response of the ionization chamber is affected by the radiation quality and the air density. Graphs
The measuring chamber features an automatic function for tube voltage correction (see Instructions
for Use, Interface description). The temperature dependence of the air density is automatically
corrected.
ATTENTION
If the measuring chamber is used at altitudes above 300 m the systematic error caused
by the change of air density depending on the altitude has to be corrected.
After installation the calibration should be tested at installation conditions (refer to chapter 4 of the
DAP meter manual). The obtained calibration factor includes the correction for the altitude. If a
calibration test at installation conditions is not possible the correction can also be done by taking the
13.3 Cleaning
Recommended detergents
For cleaning the system, use only water with mild detergents applied with a damp cloth. Never use
abrasive cleansing agents, organic solvents or detergents which contain solvents (e.g. alcohol,
For cleaning the display screens, use only pure alcohol or a mixture consisting of 1/3 alcohol and 2/3
distilled water. Wipe the screens and surrounding painted surfaces dry with a soft cotton cloth
Chapter 14
Error code
[CX9Plus]
14 Error code
Code Description Solution
MiO Board.
15VDC
Board
FnD Board .
COM 35V
Filament.
The X-ray
radiation
• Check how to use again.
Err switch is
• Check the Hand Switch
SEC stopped faster
• Check the Ready Switch.
than the time
set.
FND.
Err HVT
• Check the voltage between Test point(TP950,
PrO overvoltage
951) : DC 12V
O--- occurs.
• Check the ‘Calibration Data’ values of the
‘MIO Board’.
I-- fault signal is • Check the voltage between the U408 and
When The
• Check the lighting of the LED(D955) on the
power of the
FND Board.
FET Driver on
Err • Check the voltage between the Test point
the Board is
PrO - (TP950, 951) on the FND Board. : normal
unstable or low
-P- 12VDC ±0.5V
or not
• If the voltage is abnormal, adjust to turn the
supplied, is
VR950.
occurred.
X-ray tube is • The pin2 of the P904 and the pin1 of the
too hot and P904 on the MIO Board is a short circuit and
Err
the thermal if an error occurs, the MIO Board is bad.
tE-P
sensor is • Check the tube temperature.
broken.
Chapter 15
Specification
[CX9Plus]
15 Specification
Dimensions and Weight
Power Supply
Item Specification
Item Specification
Max.
12.5kW(120kV, 100mA at 100msec)
Electric Power
kV, 100 mA
100 kV, 6 mA
Item Specification
Rad. Mode-Nominal
130kV
Max Tube Voltage
Max. Continuous
300W
Heat Dissipation Rate
Max. Continuous
140W
Heat Dissipation Rate
Weight 15Kg
Collimator
Item Specification
Item Specification
Ratio 8:1
FD 1000mm
Al equivalent <1.2 mm
Item Specification
AD Converter 14 Bit
Item Specification
Item Specification
SID 1,000mm
Item Specification
Ø (8 ... 70) mm
chamber)
Protection class
IP40
(acc. to IEC 60529)
Radiography
Fluoroscopy
Loading Factors
(cm) 40 60 80 100
Chapter 16
Trouble Shooting
[CX9Plus]
16 Trouble Shooting
16.1 Overview
This generator has a self-diagnosing function and display error code on the console display if system
has an error.
X-ray mA meter
▶ NOTE
We will provide circuit diagrams, component part lists, descriptions, calibration instructions to assist to
▶ WARNING
THOSE PARTS OF THE EQUIPMENT SHALL NOT BE SERVICED OR MAINTAINED WHILE IN USE WITH A
PATIENT.
Turn the equipment off and disconnect the power cord from the receptacle (wall mains outlet).
184
Lan Cable
To. PC
Line Cable 18AWG
To. Monitor CART 18AWG
6 220V 6㎟
1
0V 6㎟ D-Sub, 9P
5 7 2 10㎟
GND 6㎟
16 14 12 10 8 6 4 2
190V 6㎟
4 3 15 13 11 9 7 5 3 1 Dummy
P804
190V 6㎟ To. MIO
0V 6㎟
0V 6㎟
ED-AS-CONT-0003-04 10㎟ 5 4 3 2 1
4 190V
(CND) SMC1 9 8 7 6
TB1 C2
3 0V 1uF 1500V P90
6 7 8 9
T1
L1
2 0V 6
1 190V 5 1 2 3 4 5 1 2 3
10㎟
P810 10㎟ 10㎟
Input Voltage 4
L2
T2
220V 6㎟
190V 6㎟
190V 6㎟
GND 6㎟
10㎟
0V 6㎟
0V 6㎟
0V 6㎟
3
2 D-Sub, 9P
1
NF3
220V 18AWG
NF4
0V 18AWG
Filter Choke
0V 6㎟
GND 268V 537V
0V 6㎟ 1 2 3 10㎟
537V 6㎟
SMC3 190V 6㎟ 10㎟ TC174V
268V 6㎟
0V 6㎟
190V 6㎟
G T2 T1 N2 N1 S L C
TR1 18AWG 18AWG Neutral Filament G
0V 6㎟ 0 VAC 500VA To. SMPS2 To. SMPS2
18AWG G
0 VAC 18AWG 10㎟
18AWG
10㎟ -m A
Fuse4 110 VAC 20 VAC 18AWG
AC Power 220V 30A NF1 22 VAC SMC Power SMC Power 0V 18AWG
6㎟ 18AWG 18AWG Feedback +m A
120 VAC ED-AS-HVT-0003-02
T1
T2
220V 18AWG
0VAC 0V 18AWG
6㎟ N G L 1 2 3 4
-kV
220 VAC 50 VAC 50V 18AWG
ED-AS-CONT-0003-05(FET Block) (High Voltage Tank)
P901 P801 +kV
220V Input Volrage To. FET
230 VAC
Hand SW 230V 6㎟ P709 FAN1
3 SMPS1 240 VAC 220 VAC 220V 18AWG 5
1 2 3 4
9
E 8
4 P401
DC 24V AC 3 Feedback
2 OUT PUT INPUT 7
2
1 6
1
ED-AS-CONT-0003-03 FAN2 L C S A
2 P402
+ + - - F N L 3 2 1 3 2 1 1 Sensor Trans (FND)
Foot SW 3 2 1 2 1
Com R1S R1M
2 0V 50V 220V 2 4 6 8 10
P701 P702 P906 P904 1 3 5 7 9 High Voltage Cable
Rotor_input Voltage To. SMC3 To. Tube Rotor Tube Temp
B-
b+
1 220V 18AWG Flat Cable 2 4 6 8 10
0V 18AWG 0V 18AWG 30P 1 3 5 7 9
24V 18AWG 1 0V P907
2 24V Input Voltage 2930 30 29
10㎟ 10㎟
2728 28 27
2526 26 25
1 XR_Low 2324 24 23 NF2
2 XR_High P903 2122 22 21
6C X 24AWG 3 AGC_Stop To. Camera 1920 20 19
P403
P105 C3 10000uF 350V (8EA)
4 ABC_Stop To. FND 1718 18 17
To. MIO DC IN
5 LIH 1516 16 15
10㎟ 10㎟ 10㎟ VCC+
6 1314 14 13 + 1
GND GND
1112 12 11
Fuse5 + 2
9 10 10 9 220V 30A 268V
7 8 8 7
5 6 6 5
10㎟ VCC- 3
- 537V
1 M2 3 4 4 3
P908
6C X 24AWG 2 M1 To. I.I 1 2 2 1
3
4
24V
0V
ED-AS-CONT-0004-02 P404 1
18AWG
18AWG
10 9 10 9
2
(MIO) 8 7
6 5
8 7
6 5
3
18AWG
1516 4 3 4 3
1 1314
P905 2 1 2 1
To. SMC2 11 12
P106
2 9 10
To. CND E L S C
7 8 18AWG
5 6 18AWG
220V 18AWG 3 4 18AWG
GND 1 2 18AWG
0V 18AWG P901 Flat Cable 16P
COM
MIO Power
220V 18AWG
220V
EMG_SW
18AWG
P706 SMPS2
1 Rotor CAP
2 DC AC
RL1 5V 24V
P101
2
1
8
7
P104 To. Console OUT PUT INPUT
C4
30uF 250V 3432302826242220 1816141210 8 6 4 2 1514 131211 10 9 + - + - F N L
3331292725232119 17151311 9 7 5 3 1 8 7 6 5 4 3 2 1
5V 18AWG
SC
16.4 System Block Diagram
3
4
5
6
24V 18AWG SM
0V 18AWG
C S M
TUBE
5V G2 G1 24V
Flat Cable 220V 0V Flat Cable
34P 12P 1 P112 (To. Collimator)
2 1. Brown - Aperture
3
Connting Board
6C X 24AWG 2. Green - Aperture
4 3. White - Aperture
P111 Collimator
Lift Controler
5 Input Voltage 4. Yellow - Aperture
6
2122
2324
2526
2728
2930
3132
3334
1112
1314
1516
1718
1920
1 2
3 4
5 6
7 8
9 10
LCD Membrane 5. Black - Rotation
7 6. Red - Rotation
1 P103
VSS 8 To. Lift Controler
2
Key LED VDD 9 Membrane
P1 3
To. MIO Vo 10 P908 (To. I.I)
4 1. White - N1
RS 11
5
20 18161412 10 8 6 4 2
1917151311 9 7 5 3 1
R/W 12 2. Green - N2
6
E ED-AS-CONT-0004-04 I.I 3. Red - 24V
P2 7
To. LCD 8 DB0
1 Actuator Data Lift Flat Cable
1
(Sub OP) 4. Black - 0V
ED-AS-CONS-0004-03 9
10
DB1
DB2
DB3
Membrane Lift Controler
12P
2
3
24V 18AWG
(Console) 11 0V 18AWG
Connting Board
DB4 4
12 1
DB5 5
P4 13 P112
Rev 1.00
DB6 Lift Controler 6 2
To. Switch To. Collimator LAN Cable
CX9Plus
14 Membrane
DB7 7 3
15
A(+) 8
P106
4
Camera
COM EXP Ready P3 16 To. Lift Controler
Blight K(-) 9 Membrane 5
1
10 6
3
2
1
11 7
P109
GND 18AWG 12 To. Actuator 8
EXP 18AWG P109 (To. Actuator)
Ready 18AWG 1 2 3 1. Orange
2. Red + Gray
3. Green
3C X 24AWG
CX9Plus Installation, Operation & Service Manual
Stand-by Lamp
• Check the connection between the Monitor Cart and the Generator
Input : DC 5V
Output : DC 15V
When the lamp of the P805 is lights-out, check the cable connection of the P804.
Check the cable connection of the P106 on the Main CPU Board.
- Check the lighting of the lamp D409, D407, D957. (normal : light out)
Check the voltage between the GND and the 72V terminal
- Check the voltage between the Common and Small. (Normal : DC35V)
- Check the voltage between the Common and Large. (Normal : DC35V)
- Check the voltage between the TP950 and TP951. (Normal : DC12V)
- Check the voltage between the 1 and 3 on the P901 ( Normal : AC220V ± 5%)
Check the voltage between both ends of the P702. (Normal : AC220V ± 5%)
- Check the voltage between the 3 and 4 on the P907 ( Normal : DC24V)
- When it is not measured, check the output on the SMPS. (SMPS output : DC24V)
Check the voltage between the GND and TP110. (Normal : DC7.5V)
Check the voltage between the GND and TP111. (Normal : DC15V)
Check the voltage between the 1 and 3 on the P906. (Normal : AC220V)
Check the voltage between the 1 and 2 on the P906. (Normal : AC60V)
Check the voltage between the 1 and 3 on the P701. (Normal : AC220V± 5%)
Check the voltage between the 1 and 2 on the P901. (Normal : AC50V± 5%)
When the above voltage is not confirmed, check the wiring of the X-ray Tube.
Turn off the power, and then check the resistance value of the P906 on the Main CPU Board.
- Normal : Between the pin C and pin M - 24 Ω, Between the pin C and pin S - 61 Ω,