Professional Documents
Culture Documents
CEN/TC 88
Date: 2011-03-01/ CPR draft 2013-07-01
EN 13172:2012
CEN/TC 88
The accuracy of the test equipment shall be ensured by periodic calibration. All calibrations and calibration
checks shall be traceable to relevant internationally or, failing that, nationally recognised reference test
specimens (standards). Where no such reference test specimens exist, the basis used for calibration shall
be documented.
Calibration is the right term for the first time “calibration” and in the case of a calibration check leads to
adjustment of the equipment. The calibration check, e. g. annually for HFM (see Table 1) is the right term
to use in case of no adjustment of the equipment. To simplify the text here the term “calibration” is used
both for calibration and calibration check.
The manufacturer shall define compliance criteria for each piece of equipment.
periodically at a frequency described by the manufacturer respecting the minimum frequencies specified in
Table 1;
to verify the test results obtained before being taken out of service. If internal checks are sufficient (e.g.
Heat Flow Meter (HFM) equipment) then this calibration is not needed.
Equipment not listed in Table 1 shall be calibrated in accordance with the manufacturer's documented
procedures.
1/2
NB SG 19 N 132
Calibration of thermal conductivity equipment SDG-5 EXPERT GROUP N 42 draft
The manufacturer shall assess the calibration results and document the results of such assessments. The
calibration records shall be maintained for a period of 10 years.
In addition to the traceable calibration, the manufacturer shall carry out internal checks to verify the stability
of the equipment. The manufacturer shall define compliance criteria for results of internal checks. Internal
checks shall be carried out at frequencies respecting the minimum frequencies given in Table 1.
Equipment not listed in Table 1 shall be checked in accordance with the manufacturer's documented
procedures. Records of internal checks shall be maintained for a period of 10 years.
The reference test specimen to be used for calibration and calibration checks is the IRMM 440 defining the
European thermal conductivity level.
2/2