VISUAL INSTRUCTION

ULTRAPROBE 10,000SD
Bearing Inspection
10,000SD BEARING INSPECTION
Ultraprobe 10,000SD - Bearing Inspection: How To Guide
NOTE: UE DMS Software and Spectralyzer Software are available for download at www.uesystems.com

FUNDAMENTALS
Ultrasonic Bearing Inspection is a Non Intrusive Way
to Assess the Condition of Bearings. The UltraProbe
10,000SD Allows for Analysis of Audible Quality
Identifying Inefficiencies. Recording Capabilities to
Visually Identify Faults Through Spectralyzer
Software, Compare Sounds and Conduct Fault
Frequency Analysis.

Using the UE Systems Ultraprobe 10,000SD


Detector, the Dynamic Changes Occurring in
Bearings Can Be Addressed Using Historical
Trending with Decibel (dB) Values Guiding
Maintenance Practices.

SAFETY: WRIST STRAP WORN ALL TIMES


10,000SD BEARING INSPECTION
Ultraprobe 10,000SD - Bearing Inspection: How To Guide
NOTE: UE DMS Software and Spectralyzer Software are available for download at www.uesystems.com

INITIAL SETUP
To Power On the UltraProbe 10,000SD, Pull and
Hold the Instrument Trigger. Keep Trigger
Pulled for Instrument to Remain ON. The Unit
Will Power On in The Main Display.

The UltraProbe 10,000SD Powers On in


Default Mode of 40KHz with the Sensitivity
at Maximum (70).

NOTE: The Ultraprobe Can Also Be Set to Remain Powered On


for a Determined Amount of Time.

SAFETY: WRIST STRAP WORN ALL TIMES


10,000SD BEARING INSPECTION
Ultraprobe 10,000SD - Bearing Inspection: How To Guide
NOTE: UE DMS Software and Spectralyzer Software are available for download at www.uesystems.com

INITIAL SETUP 30 KHz


To Set Up the Ultraprobe 10,000SD For Bearing
Inspections, Set the Frequency To 30KHz Using
the Sensitivity Dial. “Click” Dial Until “KHZ” is
Flashing, Then Spin to the Desired Frequency.
Once 30KHz is Reached, Click the Sensitivity
Dial Again to Set That Frequency.

Stethoscope Module

Insert the Stethoscope Module (STM) or the


Remote Magnetic Transducer (RMT) into the
Module Port. Modules Are Removed By
Pulling Module Straight Out of Port Without
Remote Magnetic Twisting.

Transducer

SAFETY: WRIST STRAP WORN ALL TIMES


10,000SD BEARING INSPECTION
Ultraprobe 10,000SD - Bearing Inspection: How To Guide
NOTE: UE DMS Software and Spectralyzer Software are available for download at www.uesystems.com

INITIAL SETUP
Select the Bearing Application from the Setup
Menu Located in the Function Bar. Press the
Sensitivity Dial Until the Setup Menu is Flashing,
Then Press the ENTER Button. Spin the Dial to find
Menu #3, Application Select. To Enter the Menu,
Click the Dial Again.

From Menu #3, Application Select, Spin the


Sensitivity Dial to Navigate to BEARINGS,
Then Click the Dial Again to Save That
Selection. Once BEARINGS is ENABLED, Press
the ENTER Button to Return to the Main
Display.

SAFETY: WRIST STRAP WORN ALL TIMES


10,000SD BEARING INSPECTION
Ultraprobe 10,000SD - Bearing Inspection: How To Guide
NOTE: UE DMS Software and Spectalyzer Software are available for download at www.uesystems.com

INITIAL SETUP
Downloading the DMS Software From
UESystems.com Will Allow Inspectors to Build a
Database For All Bearings Within a System.
Each Route Can Be Downloaded to the Ultraprobe
10,000SD For Organized Data Acquisition and
Report Generation.

Each Route is Stored Within the Plant Created


Within the DMS Software and is Available For
Upload at Anytime. Each Route Can House Up
to 400 Individual Bearings.

NOTE: For Full Details on Uploading and Downloading Routes, See


Visual Instruction: Uploading and Downloading

SAFETY: WRIST STRAP WORN ALL TIMES


10,000SD BEARING INSPECTION
Ultraprobe 10,000SD - Bearing Inspection
NOTE: UE DMS Software and Spectralyzer Software are available for download at www.uesystems.com

INSPECTION PROCEDURE
With the Bearing Route Downloaded to the
UltraProbe 10,000SD, and the Frequency Set to
30KHz, Inspectors Can Collect Data. Make Contact
With the Grease Fitting or Bearing Housing Using
the Stethoscope Module (STM) or Remote Magnetic
Transducer (RMT).

Adjust/Optimize the Sensitivity and Tune


Frequency (If Necessary) to Obtain a Good,
Quality Signal. Listen For Any Operational
Deviations. Ensure Inspectors Have Adequate
Operational Knowledge of Each Type of
Bearing Prior to Inspection.

SAFETY: WRIST STRAP WORN ALL TIMES


10,000SD BEARING INSPECTION
Ultraprobe 10,000SD - Bearing Inspection: How To Guide
NOTE: UE DMS Software and Spectalyzer Software are available for download at www.uesystems.com

INSPECTION PROCEDURE 30 KHz

Utilizing the Spectralyzer Software Can Directly


Support Accurate, and Repeatable Analysis
Practices. The UltraProbe 10,000SD Has
Independent WAV File Recording Capacity Providing
Real Time, Condition Based Analysis. Inspectors Can
Visually Determine the Presence of Harmonics to
Conduct Fault Frequency Analysis.

When Recording WAV Files, Ensure Accurate


Recording Parameters, (2 – 4 Bars) Within the
Intensity Meter and Frequency Setting. This
Allows for the Sound to be Recorded Within
Range and Not Clipped for Analysis.

NOTE: For Full Detail on Recording, See


Visual Instruction: Recording WAV Files

SAFETY: WRIST STRAP WORN ALL TIMES


10,000SD BEARING INSPECTION
Ultraprobe 10,000SD - Bearing Inspection: How To Guide
NOTE: UE DMS Software and Spectralyzer Software are available for download at www.uesystems.com

SAVING DATA
Because the Majority of Bearing Data Consists of
Audible Quality and Recording WAV Files,
Permanently Storing the Data to the SD Card Can Be
Achieved Through the Record WAV Option Within
the Function Bar. Ensure Proper Recording
Parameters Prior to Selecting Record WAV.

30 KHz
Once Recording is Complete, Inspectors Can
Playback Recording to Ensure Quality or Click
the Sensitivity Dial to Continue. Pressing the
ENTER Button Will Permanently Save the
WAV File to the SD Card and Automatically
Advance to the Next Test Point. Once
Stored, a Confirmation Prompt Will Be
Displayed.

SAFETY: WRIST STRAP WORN ALL TIMES


10,000SD BEARING INSPECTION
Ultraprobe 10,000SD - Bearing Inspection: How To Guide
NOTE: UE DMS Software and Spectralyzer Software are available for download at www.uesystems.com

SUPPORTIVE SOFTWARE

To Support Communication and Reporting, Download the DMS Software


From UESystems.com. Once Downloaded, Build Routes For Each Bearing
Within the Facility. Designating Specific Inspection Locations and
Processes. Directly Supporting Organization and Repair. All Reports are
Generated in Excel Formatting.

SAFETY: WRIST STRAP WORN ALL TIMES


10,000SD BEARING INSPECTION
Ultraprobe 10,000SD - Bearing Inspection: How To Guide
NOTE: UE DMS Software and Spectralyzer Software are available for download at www.uesystems.com

SUPPORTIVE SOFTWARE

The DMS Software Allows For Data Storage and Report Generation. From
the Reports Tab, Select the Desired Report to Generate, Then Select
“Generate Selected Report”. Next, Naming and Selecting Where the
Report Will Be Located is Required. Once Generated, the Reports Can Be
Shared, Reviewed, and Stored for Future Access. All Reports are in Excel
Formatting.

SAFETY: WRIST STRAP WORN ALL TIMES