Professional Documents
Culture Documents
https://doi.org/10.1007/s10916-017-0878-1
Received: 7 June 2017 / Accepted: 13 December 2017 / Published online: 27 December 2017
# Springer Science+Business Media, LLC, part of Springer Nature 2017
Abstract
The aim of the study was to estimate improvement of work efficiency in the laboratory after implementation of total laboratory
automation (TLA) by Abbott Accelerator a3600 in the laboratory with measuring different key performance indicators (KPIs)
before and after TLA implementation. The objective was also to recommend steps for defining KPIs in other laboratories. For
evaluation of improvement 10 organizational and/or technical KPIs were defined for all phases of laboratory work and measured
before (November 2013) and after (from 2015 to 2017) TLA implementation. Out of 10 defined KPIs, 9 were successfully
measured and significantly improved. Waiting time for registration of samples in the LIS was significantly reduced from 16 (9–
28) to 9 (6–16) minutes after TLA (P < 0.001). After TLA all tests were performed at core biochemistry analyzers which
significantly reduced walking distance for sample management (for more than 800 m per worker) and number of tube touches
(for almost 50%). Analyzers downtime and engagement time for analyzers maintenance was reduced for 50 h and 28 h per
month, respectively. TLA eliminated manual dilution of samples with extreme results with sigma values increment from 3.4 to >6
after TLA. Although median turnaround time TAT for potassium and troponin was higher (for approximately 20 min), number of
outliers with TAT >60 min expressed as sigma values were satisfying (>3). Implementation of the TLA improved the most of the
processes in our laboratory with 9 out of 10 properly defined and measured KPIs. With proper planning and defining of KPIs,
every laboratory could measure changes in daily workflow.
Keywords Key performance indicators (KPI) . Productivity . Total laboratory automation (TLA) . Quality
Before the TLA implementation, in the core laboratory automation. All KPIs were measured in November 2013
for the medical biochemistry and analytical toxicology and afterwards in different periods from 2015 to 2017.
most of the sample management processes were manually
performed, e.g. assessment of the sample quality, Key performance indicators
aliquoting and sample distribution. Besides that, laborato-
ry personnel had to walk a lot between rooms in the lab- 1. Workspace area (WSA)
oratory to distribute samples on different analyzers.
Manual steps in the procedures were not standardized, Workspace area was defined as a space directly associ-
were subjective, time consuming and prone to errors [6]. ated with preanalytical, analytical and postanalytical man-
In order to standardize processes, optimize workflow, in- agement of the samples in the core biochemistry laborato-
crease the productivity of personnel and to reduce the ry. Width and length of workspace area were measured
possibility for errors, Abbott Accelerator a3600 (Abbott, with the laser rangefinder (Bosch DLE 40, Bosch, UK)
Abbott Park, IL, USA) TLA with complete solution for and area of each room was expressed in square meters.
preanalytical, analytical and postanalytical sample han- Area where samples were processed, with all analytical
dling was implemented. systems, centrifuges and space for sample quality assess-
The main goal of this study was to evaluate accom- ment was designated as (pre)analytical area, while area for
plishment of the TLA implementation by measuring lab reports releasing was designated as postanalytical area
changes of 10 preanalytical, analytical and postanalytical (computer rooms). Considering constant increase of labo-
KPIs, before and after reorganization of the laboratory ratory requests, our goal was to be able to process as many
work. Additional objective was to provide a flowchart of samples by occupying lower amount of space. Remaining
steps required for evaluating implementation of TLA. (saved) space can be allocated for laboratory staff (reports
releasing, teaching, communication with clinicians).
i2000; Abbott Laboratories, Illinois, USA). Cobas e411 (Roche management of samples, several stages were recognized: ad-
Diagnostics, Indianapolis, USA), Axsym (Abbott Laboratories, mission of samples, registration to LIS, distribution to
Abbott Park, Illinois, USA) and Vitros 250 (Ortho Clinical preanalytical and analytical systems and to working stations
Diagnostics, Buckinghamshire, UK) were available as analyzers distant from the core biochemistry analyzers.
outside core analyzers. Since samples with tests for other ana- Number of steps for management of emergency and
lyzers needed to be manually aliquoted and distributed, those routine biochemistry samples was measured by pedome-
tests were identified as having potential for errors. t e r ( O m r o n p e d om e t e r H J - 7 2 0 I T C , H o o f d d o r p ,
In November 2013 and February 2017 (before and after The Netherlands) fixed at the belt of personnel during
TLA, respectively) all biochemistry tests requests were the weekly working hours (7:30–15:30). Steps were
counted from LIS. Afterwards, proportions of tests per turned into meters. Firstly, 7 chosen representative staff
analyzers outside the core biochemistry analyzers were members crossed 20 steps. Afterwards, distance of those
calculated. As a measure of high quality performance of 20 steps were measured with laser rangefinder (Bosch
processes, sigma value >4 was set. DLE 40, Bosch, UK). All measured distances were sum-
Our goal was to decrease proportion of tests measured out- marized and divided with 7 to obtain average distance of
side of core biochemistry analyzers. 20 steps. Walking distance was measured in two periods:
in December 2013 (before TLA) and June 2016 (after
4. Walking distance for sample management TLA implementation). Distance walked per test was cal-
culated for 30 the most frequently requested tests in LIS
Minimum number of laboratory staff required for daily according to the equation:
management of biochemistry samples was assessed. For daily
Meters walked per test ¼ ðsum of steps walked by all included workers during the day average
distance o f 20 stepsÞ=average number of 30 most common test requested per day
For comparison of daily distance walked per worker, aver- distribution) of emergency samples was done manually.
age number of tests requests after TLA was used: After TLA implementation all samples (routine and
emergency) were managed automatically, on TLA.
Proportions of samples with different combinations
Daily distance walked per worker ðmÞ
of tests (A, B or C) were exported from LIS for two
¼ ðtotal number of walked steps in day=number of workersÞ days in November 2013 and February 2017. Proportions
average number of tests after TLA of A, B and C combinations of tests were multiplied
with number of tube touches according to the following
average distance of 20 steps
equation:
Tube touch moment (TTM) was defined as an aver- A samples with tests only for the core biochemistry
age number of touches necessary to process the sample analyzers
(tube) from the sample registration to LIS to the sample B samples with tests for the core biochemistry analyzers
archiving. TTM was counted for three most common and aliquots for other analyzers
combinations of routine laboratory tests: samples with C samples that only needed aliquots for other analyzers
tests for the core biochemistry analyzers only (A), sam- (C)
ples with tests for the core biochemistry analyzers and TTM tube touch moment
aliquots for other analyzers (B) and samples that only
needed an aliquots for other analyzers (C) (Table 1). Our goal was to reduce number of tube touches.
Before TLA, preanalytical system aliquoted only rou-
tine samples, while management (aliquoting and 6. Manipulation tube time (MTT)
28 Page 4 of 12 J Med Syst (2018) 42: 28
Legend: *only for routine laboratory samples, A – samples with tests for the core biochemistry analyzers only, B –
samples with tests for the core biochemistry analyzers and aliquots for other analyzers, C – samples that only
needed an aliquots for other analyzers
Manipulation tube time (MTT) per sample was defined before TLA were: core biochemistry analyzers (AU2700/
as duration (in seconds) of every stage counted as tube AU680 (Beckman Coulter, Brea, USA), Cobas e411 (Roche
touch moment (TTM). Duration for all procedures was Diagnostics, Indianapolis, USA), Vitros 250 (Ortho Clinical
measured with stop-watch. To reduce variability of mea- Diagnostics, Buckinghamshire, UK), preanalytical system
surement (due to different conditions) duration of every Beckman OLA 2500 (Beckman Coulter, Brea, USA), Axsym
stage was measured 20 times and at the end, average (Abbott Laboratories, Abbott Park, Illinois, USA) and centri-
duration was calculated. Three different MTTs were cal- fuges. After TLA, measurement was done for 3 Architect ana-
culated in emergency and routine biochemistry laboratory lyzers: two c8000 and i2000, as well as for preanalytical system
according to the following equations: Accelerator 3600 (Abbott Laboratories, Abbott Park, Illinois,
USA). DT was calculated for one month period by multiplying
1. MTT per sample = tube touch moment (TTM) x dura- number of maintenance operations (for daily, weekly and month-
tion of all procedures x proportion of samples with the ly maintenance, reagents filling, calibration and quality controls)
most common combinations (A, B, C) with time in minutes.
2. MTT per 500 samples (average daily number of sam- Our goal was to reduce total downtime of analyzers in one
ples) = MTT per sample × 500 month period.
3. MTT per day (in hours) = MTT per 500 samples / 3600
8. Engagement time (ET) per month
Our goal was to reduce duration of all tube touch moments.
Engagement time (ET) was defined as time needed for
7. Analyzers downtime (DT) per month active engagement of personnel for all maintenance proce-
dures with analyzers (filling with reagents, calibration, quality
Analyzers downtime (DT) was defined as time while ana- control, daily, weekly and monthly maintenance). ET was
lyzers were out of function due to the maintenance procedures. measured for analyzers from 7th KPI (DT) using stop-watch
DTwas measured with stop-watch from the moment of analyzers in one month period in 2013 and 2017 and expressed in hours.
set in off-line (or shut down) until they were in the fully function- Our goal was to reduce the total time of personnel spent for
ing mode (on-line). Analyzers included in the measurement maintenance procedures.
J Med Syst (2018) 42: 28 Page 5 of 12 28
Measurement ranges (MR) and proportion of samples with When all 10 KPI’s were measured and recorded in the old
values outside of measurement range were recorded during setting, implementation of TLA began. Laboratory space
one year period for enzymes: amylase (AMY) in serum and predefined for new system was cleared out and reconstructed.
urine, alanine aminotransferase (ALT), aspartate aminotrans- Firstly, new analyzers were delivered in the laboratory. Since
ferase (AST), creatine kinase (CK) and lactate dehydrogenase Department of Clinical Chemistry is accredited to ISO 15189
(LD) in serum at core biochemistry analyzers used before standard, all new tests or systems should be verified before
(AU2700 and AU640) and after (Architect c8000) TLA im- introduction in the routine work. Therefore, verification of
plementation. All analyzers used in 2013 and 2015 had auto- analytical performances (precision, linearity, accuracy, inter-
matic dilution protocols, with different type of sample man- ferences) for all tests was done. Also, patient’s results were
agement. On AU2700/AU680, only one automatic dilution compared with results from previously used analyzers. If any
for samples with enzyme activities above upper measurement bias in reported results were noticed, clinicians were notified
range was available. Samples had to be manually handled and about the possible bias regarding previous results.
put on the analyzer for rerun in dilution. Also, automated When all construction work was finished, analyzers were
secondary dilution protocol was not available and all samples connected to the track system. About two months later, when
with values higher than primary dilution needed to be manu- system was stable and all personnel properly educated, all
ally diluted and afterwards put on analyzers. With TLA, sam- defined KPIs were measured again.
ples were automatically rerun (for primary and secondary di-
lution) from track-system and sent to analyzers. Statistical analysis
Automatic dilution of samples with extreme values in-
creases turnaround-time (TAT), while manual dilution in- Data were presented as count and percentages and median
creases TAT and is more vulnerable to errors. Therefore, with interquartile range (IQR). Normality of data distribution
manual management of samples with extremely high en- was checked with Kolomogorov-Smirnov test. Mann-
zyme activities was defined as possible error.
Whitney test was used for testing the differences for waiting
The rate of errors due to manual management was evaluat- time median and turnaround time. Comparison of proportions
ed and sigma values were calculated. For calculation of sigma was used for percentages of measurement ranges and TAT
value, total number of tests requested for each enzyme in one
outliers before and after implementation of TLA.
year period was defined as the opportunities, and number of Sigma values were calculated using web calculator [7] ac-
results above MR for defined enzymes was defined as defects. cording to the following eqs. [8]:
Sigma values >3 were defined as acceptable.
Our goal was to reduce number of samples that requited
automatic or manual dilution. Defects per million opportunities ðDPMOÞ
¼ ðNumber of defined errors=Total number of opportunities
10. Turnaround time (TAT) ðor samplesÞÞ 1000000
laboratory space was configured in only two rooms with This indicator was evaluated as successful.
total surface of 113 m2 which is almost 20% less space for
preanalytical and analytical sample management and 35% 4. Walking distance for sample management
more space for postanalytical validation, education and
communication with clinicians (Fig. 1). This indicator Laboratory personnel members (N = 7) walked approx-
was evaluated as successful. imately 14.3 m per 20 steps (average length of step
0.7 m). Sum of 30 the most common tests performed
2. LIS registration waiting time in laboratory per day increased for 16% after TLA im-
plementation (4008 before vs. 4662 tests after, respec-
Together with implementation of TLA, the organization tively). In spite of greater number of tests performed,
of admission ward was changed and hospital information number of laboratory workers needed for daily sample
system (HIS) was introduced. With HIS, entry of patients management was lower (9 vs. 8 before and after TLA,
in the LIS was simplified and all data could be transferred respectively). Moreover, all workers included in the sam-
from HIS to LIS using barcode reader. ple management walked 2 m less per test (5.1 vs. 3.1 m
Waiting time for routine samples was significantly reduced before and after TLA, respectively), while number of
after TLA implementation and reorganization of admission steps walked by each worker decreased for 31% per
with median time 16 (IQR 9–28) before TLA and 9 (IQR 6– day. When numbers of steps walked per person were
16) minutes after TLA (P < 0.001). Also, 90% of samples calculated for average of 4662 tests performed daily after
were registered to LIS 13 min earlier after TLA implementa- TLA, walking distance decreased for more than 800 m
tion (43 before vs. 30 min after TLA, respectively). This indi- per person per one day (Fig. 3).
cator was evaluated as successful. This indicator was evaluated as successful.
Before TLA, 13 out of 73 biochemistry tests were mea- 6. Manipulation tube time (MTT)
sured on analyzers outside of the core biochemistry analyzer
while after TLA all tests were measured on core analyzers In 2013, before TLA implementation, 20 stages for sample
(78 test; 5 new tests were introduced). Proportion of tests management from registration to LIS to storage of samples to
needed to be manually transferred to other analyzers de- refrigerator were possible. The most of the emergency and
creased from 1.11% before TLA to 0% after TLA (1301 routine samples had only tests performed on core biochemis-
out of total 117,278 tests performed versus 0 out of try analyzers (91% vs. 98% for emergency and 75% vs 74 for
146,234, respectively, P < 0.001). Total number of tests routine samples before and after TLA, respectively).
per month and number of tests per other analyzers before Emergency samples before TLA had combination (B) of tests
TLA was expressed as 100%, while values after TLA were for aliquoting and core analyzers in 8%. On the other hand,
calculated against values before implementation (Fig. 2). proportion of routine samples with B combination remained
Calculated process sigma value increased from 3.79 to >6. the same (24%). The smallest proportion (1–2%) of samples
Fig. 1 Ground plan of the laboratory before and after implementation of total laboratory automation (TLA)
J Med Syst (2018) 42: 28 Page 7 of 12 28
had C combination (only aliquoting tests) before TLA. After 9. Measurement ranges (MR) and dilution factors (DF)
TLA none of emergency sample needed aliquoting for other
analyzers (none with B and C combination). Total tube touch Values for upper measurement ranges with number of
moment decreased for almost 50% after TLA for emergency samples with values above MR during one year are sum-
and routine samples. Duration of every step in TTM fell for marized in the Table 2. Sigma values for samples with
approximately 35–40% for routine and emergency samples. possible errors due to manual management (manual load-
Duration expressed in hours per approximately 500 samples ing of samples to the analyzers) was calculated for period
was lower for more than 4four h per day after TLA introduc- before and after TLA implementation. TLA eliminated
tion (Fig. 4). Those indicators were evaluated as successful. manual manipulation of samples and real sigma values
go to infinity.
7. Analyzers downtime (DT) per month This indicator was evaluated as successful.
DT before TLA was 8875 min (148 h) and after implemen- Turnaround time for potassium and troponin was longer
tation of TLA 5880 min (98 h). Before TLA analyzers were after implementation of TLA for approximately 20 min.
out of function even during setting the calibrators and quality Number of samples that exceeded TAT of 1 h was larger 7%
controls on the analyzers. ET decreased after TLA for approx- for potassium, and almost 4 times larger for troponin
imately 28 h per month (118.8 vs. 90.8 h, respectively). (Table 3).
These indicators were evaluated as successful. This indicator was evaluated as unsuccessful.
Due to the different management of the samples before and laboratory, reduce the waste of personnel time and energy
after TLA, new KPI for TAT was defined as sigma values in for manual handling of samples and improve entire workflow.
which were included number of emergency samples that All increment in sample processes efficiency is possible in
exceeded TAT of 1 h per all received samples. As acceptable even smaller workspace area. Possibilities for errors are sig-
criteria sigma value >3 was defined. Sigma values were lower nificantly reduced and personnel are less exposed to possibly
in the two months period while TLA was introduced. After contagious biological material.
that period, sigma values raised to values reached before im- If any of the following changes are introduced in the
plementation of the new system (Fig. 5). laboratory: new analytical system, implementation of in-
Overall, 9 of 10 defined KPIs were successfully measured formational system, reorganization of the workflow, TLA
and improved after TLA. Only one indicator (median TAT for or accreditation process, it is crucial to asses if those
potassium and troponin) was not properly defined and could changes improved efficiency of the processes. KPIs
not be evaluated after implementation of TLA, therefore new should be objective and measureable indicators of effi-
KPIs was established for monitoring TAT. ciency in all stages of total testing process (preanalytical,
analytical and postanalytical) [9, 10]. There are several
published studies on using KPIs as a measure of success
Discussion in clinical setting, laboratories and clinics. In emergency
departments in Ireland KPIs were defined to monitor clin-
This study showed that total laboratory automation can opti- ical outcomes and measure time needed to therapy and to
mize and diminish many processes in the biochemistry measure economical cost of collecting data [11]. To
Table 2 Measurement ranges and calculated sigma values for proportion of samples above MR
Enzyme Measurement ranges Percentages of samples above Percentages of samples above Sigma values for Sigma values for
(U/L) MR with automatic dilution MR with manual dilution samples with samples with
automatic dilution manual dilution
Before After Before After P* Before After P* Before After Before After
TLA TLA TLA TLA TLA TLA TLA TLA TLA TLA
Amy (S) 1500 3010 0.33% 0.06% <0.001 0.31% 0.00% <0.001 4.22 4.76↑ 5.34 >6↑↑↑
Amy (U) 4800 3010 3.24% 4.16% 0.022 0.04% 0.00% <0.001 3.35 3.23↓ 4.85 >6↑↑↑
ALT 500 942 0.76% 0.35% <0.001 0.01% 0.00% <0.001 3.93 4.20↑ 5.29 >6↑↑↑
AST 1000 913 0.32% 0.41% 0.003 0.00% 0.00% <0.001 4.23 4.15↓ 5.46 >6↑↑↑
CK 2000 4267 1.70% 0.68% <0.001 0.06% 0.00% <0.001 3.62 3.97↑ 4.73 >6↑↑↑
LD 1200 2000 0.92% 0.39% <0.001 0.01% 0.00% <0.001 3.86 4.18↑ 5.32 >6↑↑↑
Abbreviations: Amy (S), serum amylase; Amy (U), urine amylase; ALT, alanine aminotranspherase; AST, aspartate aminotranspherase; CK, creatine
kinase; LD, lactate dehydrogenase; MR, measurement range. *Statistically significant values are bold. Arrows represent decrease or increase of sigma
values.
J Med Syst (2018) 42: 28 Page 9 of 12 28
Abbreviations: IQR, interquartile range; TLA, total laboratory automation. *Statistically significant values are
bold.
estimate efficiency in public laboratory network in Brazil, procedures. The worldwide request for better space utilization
group of researchers defined several indicators which in- of laboratory was previously recognized. Automation brings
cluded characteristics of tests produced, usage of tests per cost reduction as well as quality improvement [13].
patient and cost of personnel, tests and equipment used in Implementation of TLA in our laboratory significantly re-
the laboratory network [12]. Although recently different duced area needed for analytical and preanalytical sample
laboratories measured miscellaneous indicators, none of processing, leaving larger area for postanalytical procedures.
them was completely applicable to our setting. Every lab- In the past years, analytical systems are highly sophisticated
oratory has its own specificities in the workflow, sample with very low error rate [14]. Also, reducing errors in
management and organization of other processes. preanalytical phase has been put in great focus [15].
Therefore, it is necessary to define own specific and mea- However, great attention should be paid on postanalytical
surable criteria of improvement. KPIs in our laboratory phase, especially on interpretation of results, reporting critical
were defined before TLA implementation for the most results and improving patients healthcare with reducing pos-
erroneous and possible vulnerable processes. Our KPIs sible errors in diagnosis [16–19]. With new organization of
were not only technical (e.g. acquisition of new equipment our workspace, specialists in laboratory medicine have larger
and adaptation of workspace area), but also organizational. area for postanalytical analysis of results, communication with
Changes in organization of processes were performed dur- clinicians and scientific work.
ing and after implementation in order to optimize as much Although modification of registration of samples to
processes as possible. LIS was purely organizational and not dependent on
Dimension of the workspace depends on assigned area for TLA implementation, KPI showed significant reduction
the laboratory in the hospital and usually it is impossible to (50%) of wasted time for samples. With new organization,
enlarge it. However, reorganization of the workflow, new and samples are processed sooner, stability of tests in speci-
modern analyzers that are usually smaller than the old ones, mens is not compromised and possibility of errors in sam-
make possible to reduce space for analytical and preanalytical ple management is reduced [20].
0.0
28 Page 10 of 12 J Med Syst (2018) 42: 28
The most important benefit of TLA is reduction of the decisions for critically ill patients [27]. Median TAT for
probability for errors which was very high with manual potassium and troponin after TLA implementation in our
handling of samples [21].Especially vulnerable to errors study rose to 45 and 46 min, respectively. Our results are
are processes such as labelling of aliquot tube and man- similar to previously published study of Angeletti et al.
ual dilution of samples with extreme values. Errors in where median TAT for K and troponin in their laboratory
labeling could lead to severe risk for patient’s safety in Italy, was 39 and 42 min, respectively [28]. Likewise,
and healthcare quality [22]. In our laboratory number Lou et al., in Canada, measured 34 min as median TAT
of samples needed to be manually handled was mini- for potassium [29]. However, although their median TAT
mized with TLA, therefore the potential errors were also was slightly lower than ours, those values are not
reduced. Our laboratory is Reference center for completely comparable to ours, especially due to differ-
Emergency laboratory medicine of the Ministry of ent percentages of emergency samples and number and
Health, Republic of Croatia, and has many pathological type of analyzers included in the TLA. Our laboratory
samples, thus is particularly important to have wide mea- has almost 50% of emergency samples while Canadian
surement ranges at analyzers and automatic dilution for and Italian have only 30% and 3%, respectively.
the highest test results. TLA enables double and even Moreover, laboratory in Canada has 3 biochemistry ana-
triple rerun of samples and automatic dilution of extreme lyzers (Architect c16000) and 5 immunochemistry ana-
values. With automated procedures, dilution errors as a lyzers (Architect i2000) connected to TLA while our lab-
result of human mistakes are almost non-existing. Not oratory has two slower Architect c8000 and only one
only that TLA decreases possibility for errors but with Architect i2000. Other laboratory in the Italy, has differ-
properly planned dilution management also TAT can be ent analyzers connected to TLA, so those results are not
significantly reduced. In the study of Ialongo et al. the entirely comparable to ours. Nevertheless, recently pub-
TAT for troponin was reduced for almost 80% for lished extensive research in almost 480 laboratories ob-
prediluted samples of patients with suspected high values tained median TAT for potassium of 43 min which is
of troponin [23]. similar to our results [30].
Furthermore, implementation of TLA significantly re- Our process from sample registration to LIS to putting
duced number of tube touches needed for transferring the the samples on analyzers was reorganized and true value
samples to analyzers other than core biochemistry ana- for TAT was not possible to calculate. Before TLA, the
lyzer. With all those improvements in sample processing, most of the emergency samples were centrifuged before
exposition of personnel to potentially contaminated bio- registration to LIS. In the meantime HIS was introduced
logical material is considerably reduced with TLA. In the which increased speed of sample registration to LIS and
previous study we have confirmed that personnel usually samples are now immediately put on the TLA. This
do not comply with recommended safety measures [24], brings the difference for TAT which was not measurable
therefore we can conclude that TLA may contribute to or comparable. Because of that we have introduced new
safety of personnel and consequently improve quality of KPI, number of outlier samples with TAT exceeding de-
results with positive impact on patient healthcare. sirable 60 min for reporting results [31]. Those outliers
With introduction of automation time used for sample were expressed as sigma values which were lower than
management and analyzer maintenance could be substan- predefined 3 only two months after TLA implementation,
tially saved. Decreased energy waste for unneeded proce- which can be considered as time for adaptation to new
dures and reduction of walking distance for sample man- system and process organization. After initial period, sig-
agement allows higher rate of productivity and efficiency ma values raised and now are constantly above 3.
for staff members without compromising patient safety This study has some limitations. Our defined KPIs are
[25]. Besides that, contact with possible contaminated specific for our workflow and facilities and are probably
biological material is reduced and procedures are stan- not entirely applicable to other laboratories with different
dardized. Our data collected for only two years after im- s a m p l e m a n a g e m e n t . M o r e o v e r, d u e t o r i s e o f
plementation of the TLA are concordant to the results informatization level in hospital and reorganization of
provided in the long term study of Sarkozi et al. They processes some predefined KPIs were not measurable
collected data for 36 years and proved that automation or comparable with values gained after implementation
increase productivity of personnel, while number of em- of the TLA, e.g. TAT. Although it is important to define
ployees and price per reported test could be reduced [26]. KPIs that are measurable, easy to follow, representative
The most of defined and measured KPIs in this study and time independent [5, 11], it is not easy to define
showed significant improvement of procedures after objective measures before new system is introduced.
TLA implementation, excluding the turnaround time. The system is constantly changing and KPIs or other
TAT is highly important for clinicians and clinical measures of improvement should be changed
J Med Syst (2018) 42: 28 Page 11 of 12 28
accordingly in order to describe real working conditions. In the conclusion, implementation of the total labora-
Every laboratory should define their own KPIs and our tory automation improved many processes in our labo-
KPIs may only be template for that. ratory. TLA reduced waiting time for samples registra-
To define their own KPIs before implementation of new tion, wasting time of personnel for manual sample han-
system, every laboratory can follow some recommendations dling and analyzers maintenance as well as contact of
that arose from our study and are given in flowchart (Fig. 6). personnel with contaminated biological samples.
Fig. 6 Flowchart of
recommended steps for
implementation of changes in
laboratory with explained steps
on the left side of the chart
28 Page 12 of 12 J Med Syst (2018) 42: 28
Funding This research received no specific grant from any funding agen- 15. Simundic, A.M., and Lippi, G., Preanalytical phase – a continuous
cy in the public, commercial, or not-for-profit sectors. challenge for laboratory professionals. Biochem. Med. (Zagreb).
22(2):145–149, 2012.
Compliance with Ethical Standards 16. Kopcinovic, L.M., Trifunović, J., Pavosevic, T., and Nikolac, N.,
Croatian survey on critical results reporting. Biochem. Med.
(Zagreb). 25(2):193–202, 2015.
Conflict of Interest Author Marijana Miler declares that she has no
17. Favaloro, E.J., Lippi, G., and Adcock, D.M., Preanalytical and
conflict of interest. Author Nora Nikolac declares that she has no conflict
postanalytical variables: the leading causes of diagnostic error in
of interest. Author Lora Dukic declares that she has no conflict of interest.
hemostasis? Semin. Thromb. Hemost. 34(7):612–634, 2008.
Author Ana-Maria Simundic declares that she has no conflict of interest.
18. Katz, C., McNicholas, K., Bounds, R., Figurelle, T., Jones, C.,
Farley, H., Witkin, G., McLane, M.A., and Johnson, S.R.,
Ethical Approval Not applicable. Improving patient safety through enhanced communication be-
tween emergency department clinicians and medical laboratory
Human and Animal Rights and Informed Consent This article does not staff. J. Clin. Outcomes Manag. 20(10):455–462, 2013.
contain any studies with human participants or animals performed by any 19. Flegar-Meštrić, Z., Perkov, S., and Radeljak, A., Standardization in
of the authors. laboratory medicine: Adoption of common reference intervals to
the Croatian population. World J. Method. 6(1):93–100, 2016.
20. Cuhadar, S., Atay, A., Koseoglu, M., Dirican, A., and Hur, A.,
Stability studies of common biochemical analytes in serum separa-
References
tor tubes with or without gel barrier subjected to various storage
conditions. Biochem. Med. (Zagreb). 22(2):202–214, 2012.
1. Hawker, C.D., Laboratory automation: total and subtotal. Clin. Lab. 21. Lam, C.W., and Jacob, E., Implementing a laboratory automation
Med. 27(4):749–770, 2007. system: experience of a large clinical laboratory. J. Lab. Autom.
2. Zaninotto, M., and Plebani, M., The Bhospital central laboratory^: 17(1):16–23, 2012.
automation, integration and clinical usefulness. Clin. Chem. Lab. 22. Martin, H., Metcalfe, S., and Whichello, R., Specimen labeling
Med. 48(7):911–917, 2010. errors: a retrospective study. Online J. Nurs. Inform. 19(2), 2015.
3. Cameron, P.A., Schull, M.J., and Cooke, M.W., A framework for Available at http://www.himss.org/ojni. Accessed 22 May 2017.
measuring quality in the emergency department. Emerg. Med. J. 28: 23. Ialongo, C., Pieri, M., and Bernardini, S., Artificial neural network
735–740, 2011. for total laboratory automation to improve the management of sam-
4. Harvey, H.B., Hassanzadeh, E., Aran, S., Rosenthal, D.I., Thrall, ple dilution. LAS Technol. 22(1):44–49, 2017.
J.H., and Abujudeh, H.H., Key performance indicators in radiolo-
24. Dukic, K., Zoric, M., Starcic, J., Culjak, M., Saracevic, A., and
gy: you can’t manage what you can’t measure. Curr. Probl. Diagn.
Miler, M., How compliant are technicians with universal safety
Radiol. 45(2):115–121, 2016.
measures in medical laboratories in Croatia? Biochem. Med.
5. Bovend’Eerdt, T.J.H., Writing SMART rehabilitation goals and
25(3):386–392, 2015.
achieving goal attainment scaling: a practical guide. Clin.
25. Miler, M., and Nikolac, N., Patient safety is not compromised by
Rehabil. 23:4352–4361, 2009.
excluding microscopic examination of negative urine dipstick. Ann.
6. Armbruster, D.A., Overcash, D.R., and Reyes, J., Clinical
Clin. Biochem., 2017. https://doi.org/10.1177/0004563216687589.
Chemistry Laboratory Automation in the 21st Century - Amat
Victoria curam (Victory loves careful preparation). Clin. Biochem. 26. Sarkozi, L., Simson, E., and Ramanathan, L., The effects of total
Rev. 35(3):143–153, 2014. laboratory automation on the management of a clinical chemistry
7. https://www.isixsigma.com/process-sigma-calculator/. Accessed 8 laboratory. Retrospective analysis of 36 years. Clin. Chim. Acta.
Apr 2017. 329(1–2):89–94, 2003.
8. https://www.easycalculation.com/statistics/learn-process-sigma. 27. Goswani, B., Singh, B., Chawla, R., Gupta, V.K., and Mallika, V.,
php. Accessed 8 Apr 2017. Turn around Time (TAT) as a Benchmark of Laboratory
9. Rizk, M.M., Zaki, A., Hossam, N., and Aboul-Ela, Y., Evaluating Performance. Indian J. Clin. Biochem. 25:376–379, 2010.
laboratory key performance using quality indicators in Alexandria 28. Angeletti, S., De Cesaris, M., Hart, J.G., Urbano, M., Vitali, M.A.,
University Hospital Clinical Chemistry Laboratories. J. Egypt Publ. Fragliasso, F., and Dicuonzo, G., Laboratory automation and intra-
Health Assoc. 89(3):105–113, 2014. laboratory turnaround time: experience at the University Hospital
10. Salinas, M., López-Garrigós, M., Gutiérrez, M., Lugo, J., Sirvent, Campus Bio-Medico of Rome. J. Lab. Autom. 20(6):652–658,
J.V., and Uris, J., Achieving continuous improvement in laboratory 2015.
organization through performance measurements: a seven-year ex- 29. Lou, A.H., Elnenaei, M.O., Sadek, I., Thompson, S., Crocker, B.D.,
perience. Clin. Chem. Lab. Med. 48(1):57–61, 2010. and Nassar, B., Evaluation of the impact of a total automation sys-
11. Gannon, B., Jones, C., McCabe, A., O’Sullivan, R., and Wakai, A., tem in a large core laboratory on turnaround time. Clin. Biochem.
An economic cost analysis of emergency department key perfor- 49(16–17):1254–1258, 2016.
mance indicators in Ireland. Eur J Emerg Med. 24(3):196–201, 2017. 30. Fei, Y., Zeng, R., Wang, W., He, F., Zhong, K., and Wang, Z.,
12. Santos, M.A., Moraes, R.M., and Passos, S.R., Performance indi- National survey on intra-laboratory turnaround time for some most
cators and decision making for outsourcing public health laboratory common routine and stat laboratory analyses in 479 laboratories in
services. Rev. Saude Publica. 46(3):456–465, 2012. China. Biochem. Med. (Zagreb). 25(2):213–221, 2015.
13. Markin, R.S., and Whalen, S.A., Laboratory automation: trajectory, 31. Pati, H.P., and Singh, G., Turnaround Time (TAT): difference in
technology, and tactics. Clin. Chem. 46(5):764–771, 2000. concept for laboratory and clinician. Indian J. Hematol. Blood.
14. Plebani, M., Errors in clinical laboratories or errors in laboratory Transfus. 30(2):81–84, 2014.
medicine? Clin. Chem. Lab. Med. 44:750–759, 2006.