Parameter report

Device tag: Promag

Device information
Device tag Promag
Module name C303
Nominal diameter DN100 / 4"
Device name Promag 300
Order code 5P3B1H-SIMULATION
Serial number SIMULATION
Firmware version 01.05.00

Protocol created 2021-10-27 23:29


Operating time (counter)

Notes

Date Operator's signature Inspector's signature

www.endress.com Page 1
Parameter report

Menu/Operation /
Web server language English
Display language English
Access status Maintenance

Menu/Operation /Display /
Format display 1 value, max. size
Contrast display 50 %
Backlight Enable

Menu/Operation /Totalizer handling/


Control Totalizer 1 Totalize
Preset value 1 0.0000 dm³
Control Totalizer 2 Totalize
Preset value 2 0.0000 dm³
Control Totalizer 3 Totalize
Preset value 3 0.0000 dm³

Menu/Operation /Custody transfer/


Custody transfer mode Off
Firmware version 01.05.00
Checksum 0x4D2
Custody transfer counter 0
Custody transfer locking Defined parameters
Timestamp last custody transfer

Menu/Setup /System units/


Volume flow unit m³/min
Volume unit m³
Conductivity unit µS/cm
Temperature unit °C
Mass flow unit kg/min
Mass unit kg
Density unit kg/l
Corrected volume flow unit Nl/h

www.endress.com Page 2
Parameter report

Menu/Setup /I/O configuration/


I/O module 1 terminal numbers 26-27 (I/O 1)
I/O module 1 information HART
I/O module 1 type Current output
I/O module 2 terminal numbers 24-25 (I/O 2)
I/O module 2 information Not configurable
I/O module 2 type Pulse/frequency/switch output
I/O module 3 terminal numbers 22-23 (I/O 3)
I/O module 3 information Not plugged

Menu/Setup /Current output 1/


Assign current output 1 Volume flow
Terminal number 26-27 (I/O 1)
Current span 4...20 mA NAMUR (3.8...20.5 mA)
Signal mode Passive
0/4 mA value 0.000 m³/min
20 mA value 1.200 m³/min
Damping output 1 3.0 s

Menu/Setup /Pulse/frequency/switch output 1/


Operating mode Pulse
Terminal number 24-25 (I/O 2)
Signal mode Passive
Assign pulse output 1 Volume flow
Pulse scaling 0.0100 m³
Pulse width 100.00 ms
Failure mode No pulses

Menu/Setup /Display /
Value 1 display Volume flow
0% bargraph value 1 0.0000 m³/min
100% bargraph value 1 1.2000 m³/min
Value 2 display None
Value 3 display None

www.endress.com Page 3
Parameter report

Menu/Setup /Low flow cut off/


Assign process variable Volume flow
On value low flow cutoff 0.0200 m³/min
Off value low flow cutoff 50 %

Menu/Setup /Advanced setup/Totalizer 1/


Assign process variable Volume flow
Unit totalizer 1 dm³
Totalizer operation mode Net flow total

Menu/Setup /Advanced setup/Totalizer 2/


Assign process variable Volume flow
Unit totalizer 2 dm³
Totalizer operation mode Forward flow total

Menu/Setup /Advanced setup/Totalizer 3/


Assign process variable Volume flow
Unit totalizer 3 dm³
Totalizer operation mode Reverse flow total

Menu/Setup /Advanced setup/Custody transfer activation/


Authorized user login Stop
Login state Logged out
Year 10
Month January
Day 1
Hour 12
Minute 0
Clear custody transfer logbook Cancel
Entry 30 of custody transfer logbook 0

www.endress.com Page 4
Parameter report

Menu/Setup /Advanced setup/SIL confirmation/


Set write protection 0
SIL preparation In progress
Character test string 0123456789+-,.

Menu/Setup /Advanced setup/Display /


Decimal places 1 x.xx
Display damping 0.0 s
Header Device tag

Menu/Setup /Advanced setup/WLAN settings/


WLAN Enable
WLAN mode WLAN access point
Network security WPA2-PSK
WLAN IP address 192.168.1.212
WLAN MAC address FF-FF-FF-FF-FF-FF
Assign SSID name User-defined

Menu/Setup /Advanced setup/Heartbeat setup/Heartbeat base


settings/
Plant operator --------------------------------
Location --------------------------------

Menu/Setup /Advanced setup/Configuration backup/


Operating time 0d00h06m39s
Last backup
Configuration management Cancel
Backup state None

Menu/Diagnostics /
Actual diagnostics C485 Measured variable simulation (Warning)0d00h03m00s Deactivate
simulation (Service ID 147)
Previous diagnostics Ok

www.endress.com Page 5
Parameter report

Menu/Diagnostics /Diagnostic list/


Diagnostics 1 C485 Measured variable simulation (Warning)0d00h03m00s Deactivate
simulation (Service ID 147)
Diagnostics 2 Ok
Diagnostics 3 Ok
Diagnostics 4 Ok

Menu/Diagnostics /Event logbook/


Filter options All

Menu/Diagnostics /Custody transfer logbook/


Logbook entry 1
Event number 0
Event logbook --------

Menu/Diagnostics /Device information /


Device tag Promag
Serial number SIMULATION
Device name Promag 300
Manufacturer Endress+Hauser
Order code 5P3B1H-SIMULATION
Extended order code 1 5P3B1H-###BAEAG#####
Extended order code 2 #G#A2
Extended order code 3 #G#A2
ENP version 2.02.00
Device revision 0x6
Device ID 0x820E5A
Device type 0x113C

Menu/Diagnostics /Measured values /Process variables/


Volume flow 0.0000 m³/min
Mass flow 0.0000 kg/min
Corrected volume flow 0.0000 Nl/h
Flow velocity 0.0000 m/s

www.endress.com Page 6
Parameter report

Menu/Diagnostics /Measured values /Output values/Current output


1/
Output current 1 4.00 mA

Menu/Diagnostics /Measured values /Totalizer /


Totalizer value 1 0.0000 dm³
Totalizer overflow 1 0
Totalizer value 2 0.0000 dm³
Totalizer overflow 2 0
Totalizer value 3 0.0000 dm³

Menu/Diagnostics /Heartbeat/Performing verification /


Year 10
Month January
Day 1
Hour 12
Minute 0
Start verification Cancel
Progress 100
Status Not done

Menu/Diagnostics /Simulation /
Assign simulation process variable Volume flow
Process variable value 0.0000 m³/min
Current output 1 simulation Off
Pulse output simulation 1 Off
Device alarm simulation Off
Diagnostic event simulation 65533
Diagnostic event category Process

www.endress.com Page 7
Parameter report

Menu/Expert/System/Diagnostic handling/Diagnostic behavior/


Assign behavior of diagnostic no. 043 Warning
Assign behavior of diagnostic no. 302 Warning
Assign behavior of diagnostic no. 376 Warning
Assign behavior of diagnostic no. 377 Warning
Assign behavior of diagnostic no. 441 Warning
Assign behavior of diagnostic no. 442 Warning
Assign behavior of diagnostic no. 443 Warning
Assign behavior of diagnostic no. 531 Warning
Assign behavior of diagnostic no. 832 Logbook entry only
Assign behavior of diagnostic no. 833 Logbook entry only
Assign behavior of diagnostic no. 834 Warning
Assign behavior of diagnostic no. 835 Warning
Assign behavior of diagnostic no. 842 Off
Assign behavior of diagnostic no. 962 Warning
Assign behavior of diagnostic no. 937 Warning
Assign behavior of diagnostic no. 938 Alarm

Menu/Expert/System/Administration/
Transmitter identifier 300
Activate SW option 36131071

Menu/Expert/Sensor/Process parameters/
Filter options Binomial
Flow damping 4
Median filter depth 1
Noise suppression Off
Response time 95 % 0.5 s
Flow override Off
Conductivity measurement Off
Conductivity damping 0.00 s
Conductivity temperature coefficient 2.10 %/K

Menu/Expert/Sensor/External compensation/
Density source Fixed density
Fixed density 1.0000 kg/l

www.endress.com Page 8
Parameter report

Menu/Expert/Sensor/Sensor adjustment/
Integration time 16.7 ms

Menu/Expert/Sensor/Sensor adjustment/Process variable


adjustment/
Volume flow offset 0.0000 m³/s
Volume flow factor 1.0000
Mass flow offset 0.0000 kg/s
Mass flow factor 1.0000
Conductivity offset 0.0000 S/m
Conductivity factor 1.0000
Corrected volume flow offset 0.0000 Nm³/s
Corrected volume flow factor 1.0000
Corrected conductivity offset 0.0000 S/m
Corrected conductivity factor 1.0000
Flow velocity offset 0.0000 m/s

Menu/Expert/Sensor/Calibration/
Nominal diameter DN100 / 4"
Calibration factor 0.8200
Zero point 0.0
Conductivity calibration factor 1.0000

Menu/Expert/Sensor/Calibration/Recalibration/
Calibration factor 1.0000

Menu/Expert/Sensor/Calibration/Logbook/
Calibration counter 0
Select calibration log entry 1
Timestamp --------

www.endress.com Page 9
Parameter report

Menu/Expert/Sensor/Testpoints/
Status 0x9
Flow velocity 0.0000 m/s
Raw volume flow 0.0000 m³/min
Filtered volume flow 0.0000 m³/min
Medium temperature -nan °C
Electronic temperature 30.0000 °C
Current difference potential 0.0000 V
Current potential electrode 1 0.0000 V
Current potential electrode 2 0.0000 V
Current potential electrode Pipe GND 0.0000 V
Impedance electrode EPD 0 Ohm
Impedance electrode E1 0 Ohm
Impedance electrode E2 0 Ohm
Current gain mode High
Current measuring period 50.00 ms
Current rise time 6.0850 ms
Current integration time 16.70 ms
Split time coil current 8.00 ms
Coil peak current 91.8850 mA
Current overvoltage 59.9987 V
Current hold voltage 7.7576 V
Coil impedance 85.9035 Ohm
Coil inductance 2.06 H
Power line current synchronisation Asynchronous
Ground connection switch Open
Assign test point 1 0
Test point 1 -nan
Assign test point 2 0
Test point 2 -nan
Assign test point 3 0

Menu/Expert/Sensor/Properties/
Nominal diameter 0.0500 m
Tube inner diameter 0.0500 m
Temperature sensor type No
EPD electrode existing Yes

www.endress.com Page 10
Parameter report

Menu/Expert/Output /Current output 1/


Measuring mode Forward flow
Trim Off

Menu/Expert/Communication/HART input/Input /
Value 0.0000

Menu/Expert/Communication/HART output /Configuration/


HART short tag PROMAG
HART address 0
No. of preambles 5

Menu/Expert/Communication/HART output /Burst configuration /


Burst configuration 1/
Burst mode 1 Off
Burst command 1 Command 2
Burst trigger mode Continuous

Menu/Expert/Communication/HART output /Burst configuration /


Burst configuration 2/
Burst mode 2 Off
Burst command 2 Command 2
Burst trigger mode Continuous

Menu/Expert/Communication/HART output /Burst configuration /


Burst configuration 3/
Burst mode 3 Off
Burst command 3 Command 2
Burst trigger mode Continuous

www.endress.com Page 11
Parameter report

Menu/Expert/Communication/HART output /Information /


HART revision 7
HART descriptor PROMAG300/500
HART message PROMAG300/500
Hardware revision 1
Software revision 6

Menu/Expert/Communication/HART output /Output /


Assign PV Volume flow
Primary variable (PV) 0.000 m³/min
Assign SV Totalizer 1
Secondary variable (SV) 0.000 dm³
Assign TV Totalizer 2
Tertiary variable (TV) 0.000 dm³
Assign QV Totalizer 3

Menu/Expert/Communication/Web server/
MAC address 00-07-05-10-00-00
DHCP client Off
IP address 192.168.1.212
Subnet mask 255.255.255.0
Default gateway 0.0.0.0
Web server functionality On

Menu/Expert/Communication/WLAN settings/
WLAN subnet mask 255.255.255.0
2.4 GHz WLAN channel 6

Menu/Expert/Communication/OPC-UA configuration/
Activate OPC-UA function No
Application URI urn:dev:mac:000705100000

www.endress.com Page 12
Parameter report

Menu/Expert/Communication/OPC-UA configuration/OPC-UA
settings/
Application name Promag
Port 4840
Minimum publishing interval 1000

Menu/Expert/Communication/OPC-UA configuration/OPC-UA
security/
Security policy None

Menu/Expert/Communication/Diagnostic configuration/
Event category 043 Out of specification (S)
Event category 376 Out of specification (S)
Event category 377 Out of specification (S)
Event category 441 Out of specification (S)
Event category 442 Out of specification (S)
Event category 443 Out of specification (S)
Event category 531 Out of specification (S)
Event category 832 Out of specification (S)
Event category 833 Out of specification (S)
Event category 834 Out of specification (S)
Event category 835 Out of specification (S)
Event category 861 Out of specification (S)
Event category 842 Out of specification (S)
Event category 962 Out of specification (S)
Event category 937 Out of specification (S)

Menu/Expert/Application/Inventory counter/
Totalizer value 0.0000 l
Totalizer overflow 0

Menu/Expert/Diagnostics /Sensor module/


Hardware revision V1.00.00
Module name Promag P

www.endress.com Page 13
Parameter report

Menu/Expert/Diagnostics /Main electronic module + I/O module 1/


Software revision 10500
Build no. software 26
Hardware revision
Module name C303
Serial number 88888888

Menu/Expert/Diagnostics /Sensor electronic module (ISEM)/


Software revision 10702
Build no. software 22
Hardware revision V1.00.00
Module name V313-01
Serial number 66666666

Menu/Expert/Diagnostics /I/O module 2/


Software revision 10000
Build no. software 0
Hardware revision 1.00.00
Module name U300_01
Serial number 12345678901

Menu/Expert/Diagnostics /Display module /


Software revision 10101
Build no. software 2
Hardware revision ÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿ
Module name ÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿ
Serial number ÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿ

Menu/Expert/Diagnostics /Min/max values /Main electronic


temperature/
Minimum value 30.0 °C

www.endress.com Page 14