Professional Documents
Culture Documents
Troubleshooting
1 SAFETY 2
2 WARNING SIGNALS 2
2.1 Acoustic signals 2
2.2 Visual signals 2
3 OVERVIEW OF ERROR MESSAGES AND WARNING MESSAGES 3
4 TROUBLESHOOTING 5
5 IGBT CHECK-UP 34
5.1 Preperation of the generator 34
5.2 IGBT check 36
5.3 Connection table 37
6 HIGH VOLTAGE TRANSFORMER CHECK-UP 37
Page 2 von 37 Troubleshooting / Error Handling EDITOR HFe
1 Safety
Only qualified and trained technicians are allowed to carry out repairs and maintenance on the X-ray
generator EDITOR HFe. All individuals authorized to use this equipment must be aware of the danger of
excessive exposure to radiation. In addition, it is important that all authorized personnel are familiar with
the regulatory standards in their country.
CAUTION
Even if the generator is switched off there are still components, inside of the generator
cabinet, which are electrically live!
Only qualified and trained service personnel may open the cabinets!
Incorrect handling can result in danger to life!
2 Warning signals
2.1 Acoustic signals
With the built-in buzzer, the user is informed about special conditions of the installation:
A short sound signalizes the end of an x-ray exposure.
A modulated sound lasting 1s in case of an error.
4 Troubleshooting
E001 tube kV max
Description: Tube kV maximal limit detection by hardware interrupt on the ZST2 board. Exposure
switched of immediately by hardware.
Contactors K1 and K2 inside of the converter switched of.
Cause 3: Wrong adjusted potentiometer R52 results in an overshoots of the kV by exposure start.
R52 on the ZST2 controls the start energy of an exposure depending of the set exposure
parameters. Turn clockwise increase start energy and turn anticlockwise reduce the energy
by exposure start.
Use service software to check the kV curve shape of an exposure. Possible overshoot of
the kV during rise time.
Reduce the start power by using the potentiometer R52 on the ZST2 (see chapter 7.3.3 in
the service manual)
Cause 5: Broken IGBT V3 of the step down converter (buck converter) inside of the converter
module.
Error is detected immediately after exposure start. In this case the error occurs also by
exposures with low set kV.
Do not repeat an exposure with broken IGBT V3.
Check the IGBT V3 and replace it (see chapter 3 IGBT check-up)
Page 6 von 37 Troubleshooting / Error Handling EDITOR HFe
Replace the defective HV cables. Clean the HV connectors and use new silicon oil or
grease.
Cause 4: Faulty or loosen connection cable between high voltage transformer KB5 and the BUS2
board.
Check the connection between high voltage transformer (X2) and the BUS2 board (X9).
Check also the shielding of this cable. The shielding must be connected on both sides
Page 8 von 37 Troubleshooting / Error Handling EDITOR HFe
Cause 4: Faulty or loosen connection cable between high voltage transformer KB5 and the BUS2
board.
Check the connection between high voltage transformer (X2) and the BUS2 board (X9).
Check also the shielding of this cable. The shielding must be connected on both sides
E009 no Tube kV
Description: The generator firmware check 1ms after start of exposure if the output high voltage level is
≥ 10kV and after 30ms if the level is ≥ 50% of the set value.
E014 watchdog
Description: No Watchdog timer reset by the reason of no responding generator software for more than
1s.
Cause 1 The EEPROM D4 on the ZST2 is damaged or not correctly placed in the socked.
Check the EEPROM D4 on the ZST2 board. If the error is still present:
Exchange the ZST2 board
Cause1: filament board HZG1 not correctly connected in the electronic rack
Check the proper seating of the HZG1 board
Page 14 von 37 Troubleshooting / Error Handling EDITOR HFe
Cause3: An offset between the set value and the present value of the filament heating current. LED
V33 (red) is active
Check the high voltage cables. (Internal cable break, broken terminals (pins))
Check fuse F1 on the HZG1
Cause 6: Defective high voltage cables or bad high voltage connection on tube or on transformer
side.
Replace the defective HV cables. Clean the HV connectors and use new silicon oil or
grease.
The actual DAP test measured value is only visible after power on.
E022 key is on
Extension: Key number
Description A button is pressed on the control desk during power on.
Cause 2: Start the control desk setup and check the functionality of the X-ray buttons.
(Connect the dongle to the control desk and start the generator without pressing any button,
except the “switch on button”. Now it is possible to control the X-ray button by pressing it.)
Cause 1: No feedback signal from the switch inside the 2 tube high voltage transformer.
Check the relays on the 2 tube high voltage transformer. Switch between both tubes several
times.
Check the cable connection from the KB6 HV-transformer to the BUS2 board.
Cause 2: Blown fuse F7 on the sub distribution. (only available by generators with 2 tube HV-
transformers)
Check the fuse F7 on the sub distribution.
E028 no Tube mA
Description This error appears, if the tube current does not reach 50% of the set value during the first
30ms of x-ray radiation.
Cause 2: This error appears sometimes during the calibration. This is normal.
Increase the heating current parameter manually. Continue the calibration
030 “Info-message”
Description: This is no error message.
Message in different languages (depends on the loaded error list)
Example: “to save press button ‘M’”
Cause 2: Broken cable connecting to the control unit (control panel, Interface Box, touch panel, IFC1).
check the cable connection to the control unit.
Cause 3: The flat interface cable between NAG2 X7 and ZST2 X10is damaged or not connected
Check this connection cable.
Cause 4: The cable to the external door contact is defective or the door contact themselv.
Check the external cable and door contact.
Cause 1: Wrong exposure parameter selection or wrong measuring field selection by the operator.
Direct X-ray on selected measuring field.
Check the used exposure settings and check the chamber settings in the service software.
Check the presets in the chosen anatomic program.
Cause 1: Wrong exposure parameter settings in 2-point technique and in 3-point technique.
mAs-Product exposure preset to low in 2-point technique,
ms preset exposure preset to short in 3 point technique.
Page 23 von 37 Troubleshooting / Error Handling EDITOR HFe
Cause 2: Wrong beam direction of the X-Ray tube. Wall stand / Bucky-table
X-ray beam does not cover selected AEC chamber. Fail save function of the AEC.
Cause 1: Operator presses the preparation button for a long than the set maximum preparation time
without pressing exposure. Default value is 5s.
Check the workflow of the operator, to adjust the maximum preparation time correctly.
Choose the exposure device and start an exposure. The exposure should start immediately
after the preparation is ready and the second stage of the exposure button is pressed.
Faulty external device
Cause 2: Broken fuse F2 on the NAG2 board if the low speed starter is used.
Check the fuse F2 on the NAG2 board.
Cause 3: If high speed starter option HSS1 is used: Error indication on the HSS1 device.
If a high speed starter is installed check the error LEDs on the HSS1 board (Chapter 9.3.9
in the service manual).
Page 25 von 37 Troubleshooting / Error Handling EDITOR HFe
Cause 5: Anode boost time (HSS2 settings) is set higher than the anode start time in the service
setting of the generator.
Set the anode start time of the tube (service software) minimum 1sec. higher than the boost
time of the HSS2 setting.
042 grid is on
Description The external moving scatter grid in the bucky device is active during generator start.
The input signal Bucky must be inactive during start of the generator.
Cause 1: Active bucky input signal on the interface board NAG2 during start of the generator.
Check the cable connections of the bucky on the NAG2 board (X3/X4 and X5/X6).
Remove the connected cables of the bucky and restart the generator
If no error the failure is in the external bucky device.
- Check the power supply cable (rubber PC connector) to the converter on the backside
of the converter module.
- Check the cable connections on the sub distribution to the circuit breakers F1, F2, F3,
and F4.
- Check the cable connection from the ANS2 to the contactor K1 and K2 inside off the
converter.
Cause 1: Setup via the standard control panel is not finished properly.
The service data are saved in the EEPROM during the generator service mode via the
control panel. At service mode must be finished with the command to store the data to flash.
Start the service mode via control panel again and finish this mode properly. The changed
data are not lost if the generator is switched of.
Cause 1: External digital image system not working. Initialization command from the image system is
missing in between 1 min after generator power on.
Check the Serial communication cables between generator fluoroscopy interface IFF1 and
the external image system.
Reboot the hole system (Generator and image system)
Cause 2: Wrong parameter set in the service software (cooling date anode).
Check the set values in the service software (cooling data anode).
Cause 1: Too many request commands from external device. Serial communication has physical
limits.
Cause 1: Communication error control panel (Error code with extension “desk intern”)
Defective CAN chip inside of the control panel
check the cable connection between control panel and generator.
Page 30 von 37 Troubleshooting / Error Handling EDITOR HFe
Cause 3: Communication error to an external device (Error code with extension “COM1”)
Check the connection of external devices and the devices itself
Cause 4: Communication error to the connected DAP duo chamber (Error code with extension
“COM2”).
Check the connection to the DAP duo chamber and the DAP chamber itself.
E064 CAN-System
Extension (CAN061 – CAN067)
Description Internal communication failure on the CAN bus. The CAN bus controller set this error if the
transmission failure rate is too high. Each communication device on the CAN bus can set
this error.
List of possible generator devices which communicate on the CAN bus:
CAN1.061 Generator device 26, High speed starter HSS1
CAN1.063 Generator device 28, Control panel 2
CAN1.064 Generator device 29, Patient system
CAN1.066 Generator device 31, AEC (BA6/ BA7)
CAN1.067 Generator device 32, Control device
- Standard control panel
- Interface box
- Interface IFC1
E065 BUS-System
Description: CAN bus controller is faulty. Hardware defect.
Cause 1: Timeout of the Storz-Bus-System. Only used with Fluoroscopy systems using Storz SCB
bus.
If error appears on a RAD system check the activation of the SCB Bus in the control desk
settings. This functionality must be enabled if the generator is connected to the SCB Bus.
See control panel settings in the service software or in the service mode of the control panel
to check this function. (Chapter Fehler! Verweisquelle konnte nicht gefunden werden.)
Cause 1: The wrong or non SCB software version is detected during initialization phase.
Check SCB connection and SCB Kit (inside control panel)
Cause 1: No power supply on the HSS2 board (LED DS1-Middle is not on):
Check the rubber connector on the back of the inverter module
Check the power connection cable of the HSS2 module
5 IGBT check-up
5.1 Preperation of the generator
Make sure that the system is free of electrical voltage. Measure the voltage on the input lines with a
multimeter.
3 phases, L1, L2, L3, neutral N and the protective earth PE are required.
Open the front cover of the converter module. Disconnect the cable and unscrew the thirteen screws.
ANS2
Plug X1 IGBT V3
Plug X2 IGBT V1
Plug X3 IGBT V2
Plug X4 IGBT V1
Plug X5 IGBT V2