Tune Report

Batch Folder C:\Users\Admin\AppData\Roaming\Microsoft\Windows\Network Shortcuts\AGUAS.b


Acq. Date-Time 4/28/2021 17:02
Report Comment
Instrument Name G8403A JP15150793

[No Gas]

Mass Range Count Response Response Response


(Actual) (Actual) (Required) (Flag)
[cps/ug/l] [cps/ug/l]
7 10000 5375 53752.42 20000.00
89 10000 9179 91794.77 80000.00
205 10000 8337 83374.10 40000.00

Mass Resp Ratio Resp Ratio Resp Ratio


(Actual) (Required) (Flag)
7 -
89 -
205 -

Mass RSD% RSD% RSD%


(Actual) (Required) (Flag)
7 1.89 5.00
89 2.43 5.00
205 2.46 5.00

Mass Background Background Background


(Actual) (Required) (Flag)
7 7.40
89 1.50
205 9.20
Category Mass Count Unit Count Check
Ratio (oxide) 156/140 1.627 %
Ratio (2+) 70/140 39.419 %

Integration Time [sec] 0.1 Sampling Period [sec] 0.311

Page 1 of 3 Generated at: 5:03 PM on:4/28/2021


Tune Report

Axis Axis W-10% W-10% W-10%


Mass Peak Height (Actual) Axis (Required) (Flag) W-50% (Actual) (Required) (Flag)
7 5420.59 7.00 6.9 - 7.1 0.65 0.767 0.800
89 9136.51 89.00 88.9 - 89.1 0.49 0.724 0.800
205 8123.08 205.00 204.9 - 205.1 0.46 0.778 0.800

Integration Time [sec] 0.1 Acquisition Time [sec] 22.74 Y Axis Linear

Tune Parameters
## Plasma Paramters ##
ParameterName Value Unit ParameterName Value Unit
RF Power 1600 W Nebulizer Pump 0.12 rps
RF Matching 1.70 V S/C Temp 2 ℃
Smpl Depth 8.0 mm Gas Switch Makeup Gas
Carrier Gas 1.05 L/min Makeup/Dilution Gas 0.00 L/min
Option Gas 0.0 %

## Lenses Parameters ##
ParameterName Value Unit ParameterName Value Unit
Extract 1 0.0 V Cell Entrance -38 V
Extract 2 -165.0 V Cell Exit -52 V
Omega Bias -85 V Deflect 16.0 V
Omega Lens 10.8 V Plate Bias -50 V

## Cell Parameters ##
ParameterName Value Unit ParameterName Value Unit
Use Gas false OctP Bias -8.0 V
He Flow 0.0 mL/min OctP RF 200 V
H2 Flow 0.0 mL/min Energy Discrimination 5.0 V
3rd Gas Flow 0 %

[He]
Mass Range Count Response Response Response
(Actual) (Actual) (Required) (Flag)
[cps/ug/l] [cps/ug/l]
59 1000 858 8576.86 5000.00
89 1000 658 6582.76 5000.00
205 5000 4528 45283.27 5000.00

Mass Resp Ratio Resp Ratio Resp Ratio


(Actual) (Required) (Flag)
59 -
89 -

Page 2 of 3 Generated at: 5:03 PM on:4/28/2021


Tune Report
205 -

Mass RSD% RSD% RSD%


(Actual) (Required) (Flag)
59 4.07 5.00
89 4.93 5.00
205 2.82 5.00

Mass Background Background Background


(Actual) (Required) (Flag)
59 0.30
89 0.10
205 0.30
Category Mass Count Unit Count Check
Ratio (oxide) 156/140 0.944 %
Ratio (2+) 70/140 0.891 %

Integration Time [sec] 0.1 Sampling Period [sec] 0.31

Tune Parameters
## Plasma Paramters ##
ParameterName Value Unit ParameterName Value Unit
RF Power 1600 W Nebulizer Pump 0.12 rps
RF Matching 1.70 V S/C Temp 2 ℃
Smpl Depth 8.0 mm Gas Switch Makeup Gas
Carrier Gas 1.05 L/min Makeup/Dilution Gas 0.00 L/min
Option Gas 0.0 %

## Lenses Parameters ##
ParameterName Value Unit ParameterName Value Unit
Extract 1 0.0 V Cell Entrance -38 V
Extract 2 -200.0 V Cell Exit -70 V
Omega Bias -100 V Deflect 3.2 V
Omega Lens 11.2 V Plate Bias -60 V

## Cell Parameters ##
ParameterName Value Unit ParameterName Value Unit
Use Gas true OctP Bias -18.0 V
He Flow 5.0 mL/min OctP RF 200 V
H2 Flow 0.0 mL/min Energy Discrimination 5.0 V
3rd Gas Flow 0 %

Page 3 of 3 Generated at: 5:03 PM on:4/28/2021