Tune Report: Count Check Count Unit Category Mass
Tune Report: Count Check Count Unit Category Mass
[No Gas]
Integration Time [sec] 0.1 Acquisition Time [sec] 22.74 Y Axis Linear
Tune Parameters
## Plasma Paramters ##
ParameterName Value Unit ParameterName Value Unit
RF Power 1550 W Nebulizer Pump 0.10 rps
RF Matching 1.70 V S/C Temp 2 ℃
Smpl Depth 8.7 mm Gas Switch Makeup Gas
Carrier Gas 1.05 L/min Makeup/Dilution Gas 0.00 L/min
Option Gas 0.0 %
## Lenses Parameters ##
ParameterName Value Unit ParameterName Value Unit
Extract 1 0.0 V Cell Entrance -38 V
Extract 2 -185.0 V Cell Exit -58 V
Omega Bias -85 V Deflect 16.8 V
Omega Lens 11.6 V Plate Bias -50 V
## Cell Parameters ##
ParameterName Value Unit ParameterName Value Unit
Use Gas false OctP Bias -8.0 V
He Flow 0.0 mL/min OctP RF 200 V
H2 Flow 0.0 mL/min Energy Discrimination 5.0 V
3rd Gas Flow 0 %
[He]
Mass Range Count Response Response Response
(Actual) (Actual) (Required) (Flag)
[cps/ug/l] [cps/ug/l]
59 2000 977 9769.54 5000.00
89 1000 819 8190.31 5000.00
205 5000 4385 43848.29 5000.00
Tune Parameters
## Plasma Paramters ##
ParameterName Value Unit ParameterName Value Unit
RF Power 1550 W Nebulizer Pump 0.10 rps
RF Matching 1.70 V S/C Temp 2 ℃
Smpl Depth 8.7 mm Gas Switch Makeup Gas
Carrier Gas 1.05 L/min Makeup/Dilution Gas 0.00 L/min
Option Gas 0.0 %
## Lenses Parameters ##
ParameterName Value Unit ParameterName Value Unit
Extract 1 0.0 V Cell Entrance -22 V
Extract 2 -190.0 V Cell Exit -70 V
Omega Bias -95 V Deflect 3.0 V
Omega Lens 12.5 V Plate Bias -60 V
## Cell Parameters ##
ParameterName Value Unit ParameterName Value Unit
Use Gas true OctP Bias -18.0 V
He Flow 5.0 mL/min OctP RF 200 V
H2 Flow 0.0 mL/min Energy Discrimination 5.0 V
3rd Gas Flow 0 %