Designation: E 915 – 96 An American National Standard

AMERICAN SOCIETY FOR TESTING AND MATERIALS


100 Barr Harbor Dr., West Conshohocken, PA 19428
Reprinted from the Annual Book of ASTM Standards. Copyright ASTM

Standard Test Method for


Verifying the Alignment of X-Ray Diffraction Instrumentation
for Residual Stress Measurement1
This standard is issued under the fixed designation E 915; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (e) indicates an editorial change since the last revision or reapproval.

1. Scope the diverging or parallel beam types.3, 4


1.1 This test method covers the preparation and use of a flat 3.2 Application of this test method requires the use of a flat
stress-free test specimen for the purpose of checking the specimen of stress-free material that produces diffraction in the
systematic error caused by instrument misalignment or sample angular region of the diffraction peak to be used for stress
positioning in X-ray diffraction residual stress measurement, or measurement. The specimen must be sufficiently fine-grained
both. so that large numbers of individual crystals contribute to the
1.2 This test method is applicable to apparatus intended for diffraction peak produced. The crystals must provide intense
X-ray diffraction macroscopic residual stress measurement in diffraction at all angles of tilt, c, which will be employed (see
polycrystalline samples employing measurement of a diffrac- Note 1).
tion peak position in the high-back reflection region, and in NOTE 1—Complete freedom from preferred orientation in the stressfree
which the u, 2u, and c rotation axes can be made to coincide specimen is, however, not critical in the application of the technique.
(see Fig. 1).
1.3 This test method describes the use of iron powder which 4. Procedure
has been investigated in round-robin studies for the purpose of 4.1 Instrument Alignment:
verifying the alignment of instrumentation intended for stress 4.1.1 Align the X-ray diffraction instrumentation to be used
measurement in ferritic or martensitic steels. To verify instru- for residual stress measurement in accordance with the instruc-
ment alignment prior to stress measurement in other alloys, tions supplied by the manufacturer. In general, this alignment
base metal powder having the same crystal structure as the must achieve the following, whether the u, 2u, and c axes are
alloy should be prepared in similar fashion and used to check variable or fixed (see Fig. 1):
instrument alignment at the appropriate diffraction angle. 4.1.1.1 The u, 2u, and c axes shall coincide.
1.4 This standard does not purport to address all of the 4.1.1.2 The incident X-ray beam shall be centered on the c
safety concerns, if any, associated with its use. It is the and 2u axes, within a focusing range, which will conform to the
responsibility of the user of this standard to establish appro- desired error and precision tolerances (see sections 5 and 6).
priate safety and health practices and determine the applica- 4.1.1.3 The X-ray tube focal spot, the c and 2u axes, and the
bility of regulatory limitations prior to use. receiving slit positioned at 2u equals zero degrees shall be on
a line in the plane of diffraction. Alternatively, for instrumen-
2. Referenced Documents tation limited to the back reflection region, the diffraction angle
2.1 ASTM Standards: 2u shall be calibrated.
E 11 Specification for Wire-Cloth Sieves for Testing Pur- 4.1.1.4 The proper sample position shall be established,
poses2 using whatever means are provided with the instrument, such
that the surface of the sample is positioned at the u and c axes
3. Significance and Use , within the focal distance range which will conform to the
3.1 This test method provides a means of verifying instru- desired error and precision tolerances (see sections 5 and 6).
ment alignment in order to quantify and minimize systematic 4.1.1.5 The angle c must be determined accurately.
experimental error in X-ray diffraction residual stress measure- 4.2 X-Ray Optics:
ment. This method is suitable for application to conventional 4.2.1 When the Ka characteristic radiation doublet is used
diffractometers or to X-ray diffraction instrumentation of either for stress measurement, it is desirable to select incident and
receiving X-ray beam optics that will produce maximum

1
This test method is under the jurisdiction of ASTM Committee E-28 on
3
Mechanical Testing and is the direct responsibility of Subcommittee E28.13 on Hilley, M. E., Larson, J. A., Jatczak, C. F., and Ricklefs, R. E., eds., Residual
Residual Stress Measurement. Stress Measurement by X-ray Diffraction, SAE J784a, Society of Automotive
Current edition approved Apr. 10, 1996. Published June 1996. Originally Engrs., Inc., Warrendale, PA (1971).
4
published as E 915 – 83. Last previous edition E 915 – 90. “Standard Method for X-Ray Stress Measurement,” Committee on Mechanical
2
Annual Book of ASTM Standards, Vol 14.02. Behavior of Materials, The Society of Materials Science, Japan, (20 April 1973).

1
E 915

FIG. 1 X-Ray Diffraction Stress Measurement Geometry and Angles Defined

separation of the Ka1 − Ka2 doublet. Perform stress measure- work on a diffractometer) using a 10 % solution of nitrocellu-
ments on the stress-free specimen employing the Ka1 diffrac- lose cement5 diluted with acetone as a suitable amorphous
tion peak at all c angles investigated. Because resolution of the binder. Place several drops of the solution on a clean micro-
Ka doublet may vary with the angle c, and because some scope slide or in a sample tray, and sprinkle the powder into the
instrumentation may be incapable (due to fixed X-ray optics) of binder. The powder may be spread and leveled with a second
obtaining resolution of the doublet, care must be taken not to microscope slide. When a uniform flat surface has been
resolve the doublet at some c angles while blending the produced by alternately wetting with the binder solution and
doublet into a single peak at other c angles. wiping with a second slide, set the specimen aside and allow it
4.3 Selection of Powder for a Stress-Free Iron Specimen: to dry for several hours. Excess amounts of the binder may
4.3.1 Use iron powder with a particle size greater than 1 µm cause it to peel away from the surface of the microscope slide.
(4 3 10− 5 in.) (See Note 2.) Rewetting of the surface with acetone and redrying may
NOTE 2—Annealed armco iron powder of <45 µm (325 mesh) has been
eliminate this difficulty. Make the surface of the specimen as
found suitable. flat as possible so that the specimen surface is clearly defined.
4.4.2 A temporary specimen may be rapidly prepared using
4.3.2 Annealing of the powder in vacuum reduces diffrac- petroleum jelly as an amorphous binder. Place a small quantity
tion peak width, thereby increasing diffraction peak resolution. of petroleum jelly on the face of one microscope slide and
This is generally desirable (see Note 3). Powders in the form of press it against a second slide to extrude the petroleum jelly
plastically deformed filings may be used, but will produce into a uniform flat film. Remove the second microscope slide
broader diffraction peaks. In the event that an instrument with a wiping action taking care to keep the surface layer of
incapable of resolution of the Ka1 − Ka2 doublet is being petroleum jelly thin and flat. Holding the petroleum jelly-
employed, it may be desirable to deliberately obtain plastically coated slide at a steep angle to a vertical line, sprinkle the iron
deformed powders which insure that partial resolution of the powder from a sufficient height above the slide so that the
Ka doublet does not occur. Extremely fine powders have also powder strikes the coated surface and either adheres or is
been shown to produce line broadening, sufficient to suppress deflected away. Do not allow the powder to pack and build up
resolution of the Ka doublet. on the surface.
NOTE 3—It may be advantageous to anneal an oxide-forming powder in 4.4.3 The surface area of the powder must be of sufficient
a reducing atmosphere rather than in vacuum to avoid problems from size to intersect the entire incident X-ray beam at all c angles
surface contamination. It is not necessary to anneal ceramic powders since to be used during stress measurement.
these materials do not tend to show line broadening from plastic 4.5 Instrument Alignment Check:
deformation.
4.5.1 Position the stress-free specimen on the X-ray diffrac-
4.4 Stress-Free Specimen Preparation: tion apparatus (see 4.1.1.4). In the event that a mechanical gage
4.4.1 A permanent stress-free specimen may be prepared by
mounting the powder on the face of a microscope slide or in a 5
Duco Cement available from DuPont Co., Wilmington, DE 19898 has been
shallow powder tray (of the type used for powder diffraction found suitable. Paraffin wax may also be used.

2
E 915
which contacts the surface of the specimen is used for 5.2.1 Experience has shown that the standard deviation of
specimen positioning, a thin metal shim may be placed in front the five measurements should be within approximately 6.9
of the powder surface to protect it. Place this gage against the MPa (1.0 ksi). In the event that the standard deviation of the
face of the metal shim, and adjust the positioning to account for five measurements exceeds 14 MPa (2.0 ksi), the stress-
the inclusion of the shim in front of the gage such that the measurement technique employed and the instrumentation
surface of the powder is at the correct distance from the should be investigated for sources of random error affecting the
reference point of the gage for stress measurement. measurement precision. Random error due to counting statis-
tics may result from failure to take sufficient time during the
NOTE 4—Failure to place the powder surface directly over the center of
rotation of the c and 2u axes induces a systematic specimen displacement measurement to obtain accurate intensity information, and thus
error. to accurately determine the diffraction peak positions. Methods
are available3 for estimating the standard deviation of the
4.5.2 Without adjusting the specimen position, perform five
measured stress due to the errors involved in counting and
successive stress measurements using the method and correc-
curve fitting to determine peak positions. Mechanical sources
tion procedures normally employed for the instrument. To
of error such as loose bearings and ways in the apparatus may
avoid systematic error in the verification process when Ka
result in significant random error.
radiation is being used, care must be taken to either completely
split or blend the Ka1 − Ka2 doublet (see 4.2). 6. Precision and Bias
NOTE 5—Values for accuracy and precision of the various angles and 6.1 The precision of this method will be dependent upon the
displacements are not specified herein. These may be considered to be met type of X-ray diffraction instrumentation employed and the
collectively when overall measurement errors and tolerances are within methods of data reduction used in stress measurement. The
those specified in Sections 5 and 6. preliminary results of round-robin investigations using this
method indicate that instrument alignment within 14 MPa (2.0
5. Calculations and Interpretation of Results
ksi) (see 5.1) can be achieved for both standard diffractometers
5.1 Systematic Error: and two types of X-ray diffraction instrumentation designed for
5.1.1 Reduce the X-ray diffraction data obtained from the stress measurement in the back reflection region only. Instru-
five measurements in whatever manner is normally employed mental precision measured by this method (see 5.2) has been
for the X-ray diffraction instrumentation in use, and include all found to be less than 66.9 MPa (1.0 ksi).
corrections normally applied to raw X-ray diffraction data. 6.2 The accuracy of this method is considered to be absolute
Calculate the simple arithmetic mean and standard deviation because the specimen is stress-free. Deviation of results
about the mean for the five measurements. If the mean value is obtained in performing this method, provided the specimen has
within 14 MPa (2.0 ksi) of zero, the instrument and specimen- been properly prepared and maintained, can be attributed to the
positioning gage can be considered to be properly aligned. In instrumentation under investigation.
the event that the mean differs from zero by more than 14 MPa
(2.0 ksi), repeat 4.1 and 4.5. 7. Keywords
5.2 Random Error: 7.1 alignment; residual stress; x-ray diffraction

The American Society for Testing and Materials takes no position respecting the validity of any patent rights asserted in connection
with any item mentioned in this standard. Users of this standard are expressly advised that determination of the validity of any such
patent rights, and the risk of infringement of such rights, are entirely their own responsibility.

This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM Headquarters. Your comments will receive careful consideration at a meeting of the responsible
technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should make your
views known to the ASTM Committee on Standards, 100 Barr Harbor Drive, West Conshohocken, PA 19428.