Reader 720: User's Manual
Reader 720: User's Manual
User`s Manual
Page 1 of 18
Orgenics Ltd MediScreen Reader II
Contents
2. System Installation
2.1 Unpacking
2.2 Interfacing System
3. Functions of Components
3.1 Keyboard
3.2 LCD
3.3 Printer
3.4 Waste Box
3.5 Strip Bed
6. Maintenance of instrument
6.1 Daily check up
6.2 Cautions for keeping
6.3 Error, errors!
Page 2 of 18
Orgenics Ltd MediScreen Reader II
Page 3 of 18
Orgenics Ltd MediScreen Reader II
Page 4 of 18
Orgenics Ltd MediScreen Reader II
On/Off
2. System Installation
2.1 Unpacking
Check that all the items in the list below are present in the package.
Item(Quantity)
1. Reader 720(1)
2. Thermal Printing Paper(1)
3. Power Cable(1)
4. RS 232c Cable(1)
5. User’s manual(1)
Page 5 of 18
Orgenics Ltd MediScreen Reader II
MediScreen
Reader II
Orgenics Ltd.
[Figure 3] Keyboard
1) ENT Key
It is used for starting test. It also completes all entries of patient/sample’s ID numbers
and is used to confirm the data.
2) ESC Key
It is used for termination of each procedure.
Page 6 of 18
Orgenics Ltd MediScreen Reader II
3) ◀▶ ▲▼ (Direction) Keys
Direction keys are used for entering patient’s ID No., date, and time....etc.
They are also used for moving to the left, the right, up and down respectively in menu.
4) 0-9 (Numeric) Keys
Numeric keys are used for the execution of each procedure or in-put data.
3.2 LCD
It displays data which show the operation status of the system and the test
results.
The display size is 21 characters x 8 lines.
Reader 720
Orgenics Ltd.
14:51:35
[Fig.3.1] LCD
3.3 Printer
This analyzer prints test results of urine samples. The printer can be on or off. When the
printer is out of paper, a red line will appear. This is when new thermal paper should be
fed to the printer.
Page 7 of 18
Orgenics Ltd MediScreen Reader II
Page 8 of 18
Orgenics Ltd MediScreen Reader II
"Urine reader"
Ver TC-1.25R
MODEL 720P
ALL RIGHTS RESERVED
Nov 2003 20:52:20
Page 9 of 18
Orgenics Ltd MediScreen Reader II
Page 10 of 18
Orgenics Ltd MediScreen Reader II
[Fig 5.5]
If more changes are required, press ENT key and you will
go back to System Control mode.
Page 11 of 18
Orgenics Ltd MediScreen Reader II
Page 12 of 18
Orgenics Ltd MediScreen Reader II
1) Enter operator’s ID in part ⓐ and strip lot number in part ⓑ by using numeric keys.
If you don’t need any record about test operator’s ID and strip lot number, you may
skip and leave it blank.
2) Enter patient/sample’s ID in part ⓒ using either the numeric keys or the barcode
reader .
- Press ENT key or UP key. Then you can enter another patient/sample’s ID.
- If you want change an ID, you can go to the ID using up, down key and enter ID
number again.
3) Partⓓ is the serial number of urine samples you are going to test. As you register
patient/sample’s ID numbers, number in part ⓓ increases in order.
4) Part ⓔ is data’s number which will be stored in data base after measurement.
- If IDs are serial numbers, you may register only the first sample’s ID. Then the others
will be numbered in order automatically. In that case next ID will be increasing one by
one from the first number.
- In case you don’t register patient/sample’s ID before test it will be blank in ID.
5) Press ESC key and you will return to System Control mode.
5.8 Measurement
You should start measurement in Standby Mode. you can return to Standby mode
anywhere you are in by pressing ESC key once or twice .
1) Dip a strip into fresh urine sample and take it out immediately.
2) Remove excessive urine by contacting the side of strip on
the material like soft tissue
3) Lay it at the start zone on the strip bed carefully as in [Fig.3.3] .
4) Press ENT key in standby mode (start page).
5) Test will be carried out automatically.
6) Test result will be printed out , shown on the LCD and
automatically saved in the memory. ( Refer to [Fig.5.9]Result Sheet )
- If you have more than 2 urine samples repeat the step 1), 2), 3) continuously.
- Test result comes out in 2 minutes for the first sample and 5 seconds for the next ones
as long as you lay the strip in time.
7) The instrument will stop automatically after finishing test.
8) Press ESC key twice to return to standby mode.
Page 13 of 18
Orgenics Ltd MediScreen Reader II
Tested Time
Data’s NO
Instrument Serial No
Patient’s NO
Operator’s NO
Strip’s Lot NO
[Fig.5.9] Result Sheet
Note:
1) Use only strips manufactured by Orgenics Ltd.
2) Do not touch on the pad strip by hand.
3) Do not use bent strips.
4) Always use a fresh urine sample.
5) Lay only one strip on the strip bed at a time.
6) As with all laboratory tests, definitive diagnostic or therapeutic decisions
should not be based on any single test result or method.
Page 14 of 18
Orgenics Ltd MediScreen Reader II
16:50:14 MediScreen 10
Page 15 of 18
Orgenics Ltd MediScreen Reader II
Page 16 of 18
Orgenics Ltd MediScreen Reader II
3) The beeper will sound if strips more than 200 strips and the instrument stopped if
over 250 strips are detected in the waste container.
4) If the instrument didn’t sense strip on bed, press the “0” key it and will move
forward.
5) Some times the instrument has stopped the measuring cause it has detected 28
empty strips. In order to prevent this situation (In case of no strip on bed, But
instrument is move work.) Press “0” key or “ENT” key when you put on strip in
bed.
Manufacturing Facility:
Industrial Zone
Yavne 70650
Israel
Tel. +972-8-9429201
Fax. +972-8-9438758
Page 17 of 18
Orgenics Ltd MediScreen Reader II
IMPORTANT
1. In order to prevent electric shock ensure the connect earth ground Power Cable.
2. If a fuse is being replaced in the power supply, disconnect the power cable prior to
replacement.
3. In the case of sensed gas smell, unusual noise or mailfunction of the system, stop the
instrument and disconnect the power cable immediately
4. If you think need repair the system, Please contact our skilled technical manpower.