Professional Documents
Culture Documents
Lab NS1
X-ray Imaging and Tomography
Daniils Ganins
Part 1.
The intensity of the X-ray beam,I, which is detected by the camera, is attenuated by any material,
with thickness, δx. Therefore all along the materials’ area the thickness should be the same. That is
why I have used these areas for the empty vial and the vial filled with water.
Picture 1
Area used for the measurements of the empty vial
Picture 2
Area used for the measurements of the vial filled with water
As one can see, these areas avoid the corners of the vial. I’ve decided to avoid them due to the fact
that the amount of glass in these regions is bigger and the actual picture looks dimmer. These are
not adequate results, therefore It’s been decided to completely avoid the areas of curvature.
Graph 1
Energy versus thickness of glass (empty)
As one can see, on both graphs there are areas (0-10 px) and (107-120 px) where graph suddenly
changes within a single pixel. This data was not taken in the account.
In order to find the absorption due to glass only and the absorption due to water and glass I will
use this equation:
I =I 0 e(−2 μ d ), where I is an intensity at any time, I0 is an intensity in time 0, and e (−2 μ d ) is an
g g
Similar equation applies for the absorption due to water and glass.
I =I 0 e−(2 μ d + μ D)
g w
, where I is an intensity at any time, I0 is an intensity in time 0, and e−(2 μ d +μ
g w D)
is an
absorption due to glass and water.
In order to obtain e−(2 μ d +μ D ) and e (−2 μ d ) I am doing to divide I/I0 for water and I/I0 for glass and
g w g
produce tables. The graph of I/I0 against Energy will fulfill the results.
0.3
0.2
0.1
0
10 15 20 25 30 35 40
Energy (kV)
Water Glass
Absorption coefficient of glass, μ g and the absorption coefficient of water, μw are presented in the
graph below.
d = 1.14 ± 0.01 cm
D = 2.052 ± 0.01 cm
𝜇g and 𝜇w vs Energy
6
5
𝜇g and 𝜇w cm-1)
4
Glass
3
Water
0
10 15 20 25 30 35 40
Energy (keV)
Questions:
1) The best beam energy to use to maximise the contrast between the glass and water is
around 20 keV, what might easily be seen if one may take a look on a graph μg and μw vs
Energy.
2) The beam energy I have selected is aligned to the peak absorption coefficient of glass. The
significance of that is that there is a maximal difference between absorption of water and
absorption of glass exactly with the beam with this energy (20 keV).
Part 2
HU Glass HU Water
1060 -510
990 -500
1180 -550
1240 -510
1300 -530
1050 -540
710 -520
1150 -460
1160 -490
1400 -480
Mean pixel intensity for water is -509 HU and for glass is 1124.
μw =0.6
μ i−μw
By using the formula p1=1000 ( )
μ w −μair
and setting μair to zero, the value μi=0.2946 ≈ 0.3 was
obtained.
It occurred that μi=0.3 is the smallest value of μi that is possible to put into the program.
By putting the value μi=0.3 into the program the following data was obtained:
Mean pixel intensity for water is now -7, which is very close to 0, for glass it is 3179 and for Perspex
it is -241.
The change in the mean pixel intensity for water is observable. It used to be -509, after calibrating it
is -7. Setting μi ¿ 0.3 means calibrating the scanner so that the intensity of water is 0 on average. It is
a medical thing.
Part 3
Data:
The diameter of the steel ball is 31.55 ± 0.05 mm.
Working out the physical dimension (cm) of one pixel.
Graph 2.
The peals on a graph of intensity correspond to the empty (air) areas in the apparatus.
Picture 3
Graph 3
Picture 4
Therefore the actual diameter of the sphere corresponds to the distance from 40 to 200
pixels. In total it is 160 pixels. Therefore by using this formula:
3.155
one may calculate the dimension (cm) of pixel. It is D= =0.024 cm
130
The example was taken of the 35 keV graph (Graph 2 – relate to Picture 3.). I have chosen
that graph due to the couple of reasons, but mostly because there are simply more photons
in the 35 keV graph so less noise than in the 15 keV graph (Graph 3 – relate to Picture 4).
The results are more adequate comparing to the 15 keV graph.
Questions:
1) The signal does not change in one pixel. This happens because the ball is thinner on
the edge, so the photon penetration is a lot less on the edge. If we a cylinder instead
it would be perfect.
2) Spatial resolution is pixels/cm. It might be obtained if this formula is used:
N pixels
Spatial resolution= Therefore
diameter of the sphere(cm)
130
Spatial resolution= =4.12± 0.05 pixels/cm
31.55
3) The plastic sphere could not be used instead of the steel one. Photons would just
penetrate that sphere and one will get less clarity on the graph.
Part 4
Derivation:
−( μ g d + (f w u w+ f 0 u 0) D )
I =I 0 e
Total fraction of water and oil is:
f w +f 0 =1
− ( μ g d+ (f w u w+ f 0 u 0) D)
J=J 0 e
− ( μ g d+ (f w uw) D )
J=J 0 e
If f o=0 , f w =1
− ( μ g d+ (uw ) D )
J=J 0 e
By dividing the two images (I and J)to get the division image (P):
( ( − μg d + f w uw +f 0 u0 D )
)
I I0 e
P= =
J − μ d +( u ) D )
J0 e ( g w
In an ideal world, nothing has changes in terms of the intensity in the photon beam between image I
and image J.
− ( μg d + ( f w uw +f 0 u0 ) D )
I I e
P= = 0
J −μ
I0e ( g
d + ( uw ) D )