You are on page 1of 11

Brit. J. Anaesth.

(1969), 41, 731

ELECTRODES FOR THE MEASUREMENT OF OXYGEN AND


CARBON DIOXIDE TENSIONS
BY
A. CRAMPTON SMITH AND C. E. W. HAHN

Several comprehensive review articles have adapted for use in biochemistry by Prat (1926).
recently been published by anaesthetists on the Davies and Brink (1942) suggested the protection
measurement of blood-gas tensions and pH (Laver of the cathode with a membrane for measurements
and Seifea, 1965; Adams, Morgan-Hughes and in blood to eliminate poisoning due to protein
Sykes, 1967, 1968; Severinghaus, 1968). These deposition. The results of cathode protection with
papers have covered blood sampling and stor- a variety of materials remained unsatisfactory
age, the introduction of samples to electrodes, (Shepard, 1956) until the picture was altered when
the calibration of commercially available systems Clark (1956) introduced the principle of the closed
and so on. Kelman and Nunn (1966) have empha- system. In this system, both the oxygen cathode
sized the importance of applying correction and the reference electrode (the anode) are
factors for elapsed time and electrode drift, and covered by a single membrane permeable to
Linden, Ledsome and Norman (1965) have des- oxygen.
cribed methods of measuring CO2 and O2 content
using electrodes and have compared these methods Basic system.
with the manometric method of van Slyke (Peters The oxygen electrode has been developed from
and van Slyke, 1958). a simple polarographic cell involving an electro-
In the limited space available to us, we have chemical reaction and a flow of current. Figure 1
attempted a simplified description of the physical shows a container filled with electrolyte solution,
background to O2 and CO2 electrodes, which may a cathode and an anode with an external circuit
increase the anaesthetist's understanding of in- consisting of a variable voltage source and a
struments he uses frequently. This area is not galvanometer to measure any current flowing in
specifically covered in the articles mentioned the system. If the cathode is made a few tenths of
above and is difficult to extract from standard a volt negative with reference to the anode, oxygen
texts. Discussion of the practicalities of the use of molecules at the cathode surface undergo electro-
the O2 and CO2 electrodes is not included except lytic reduction, resulting in a current which can
where the authors' practice differs from that des- be measured by the galvanometer. If the voltage
cribed in the reviews—and where the differences is increased to more negative values, electrolysis
are regarded as important. It must be emphasized occurs more rapidly and the current increases in
that this paper makes no pretensions to complete- a manner which can be expressed graphically by a
ness and must be regarded only as an addendum sigmoid curve (fig. 2). The plateau of the curve
to the important articles already mentioned. occurs because the electrolytic reduction process
is capable of reducing more molecules of oxygen
THE O 2 ELECTRODE than can reach the cathode by diffusion, so that in
spite of an increase in the negativity of the voltage
The measurement of the partial pressure of oxy-
applied to the cathode, no increase in current
gen in gaseous, liquid and semiliquid media is
occurs. The shape of this sigmoid curve or polaro-
now everyday practice in most hospitals and
gram indicates the nature of a substance dissolved
physiology laboratories. The polarographic prin-
in the electrolyte, and the height of the plateau
ciples developed by Heyrovsky (1922) and his
school had obvious applications in the analysis of
many substances including oxygen, but the A. CRAMPTON SMITH, M.A., M.B., F.F.A.RC.S.; C. E. W.
HAHN, B.SC, M.SC, Nuffield Department of Anaes-
method was not used in biology until it was thetics, The University of Oxford.
732 BRITISH JOURNAL OF ANAESTHESIA

the cathode quickly becomes poisoned by protein


contaminants. The Clark electrode (fig. 3) over-
comes this problem by separating the electrolyte
in which electrolysis takes place from the medium
in which oxygen is to be measured by a mem-
brane permeable to oxygen but impermeable to
ionized contaminants. Suitable materials are
polyethylene, polypropylene, teflon or mylar.

i Anode
To measuring unit
Cathode

Ag-Ag/CI
reference Insulated container
anode

FIG. 1
Basic arrangement of a simple oxygen polarographic Electrolyte
cell.

M Insulating glass
rod with platinum
wire sealed in

J Rubber " 0 " ring

Membrane
Sample
Platinum
Quantity tip of
cathode

FIG. 3
Clarke-type oxygen electrode.
Nature
FIG. 2 These are good electrical insulators and their
Current-voltage curve for the reduction of oxygen at a ability to diffuse oxygen is practically unchanged
constant oxygen concentration.
in the presence of water vapour (Myers et al.,
1962). The reference electrode is silver-silver
indicates its concentration. If the voltage applied chloride, silver-silver hydroxide, or calomel half
is constant and is such that the cell current is in cell. The electrolyte is chosen to suit the reference
the plateau region, the heights of a series of electrode, typical solutions consisting of KC1,
plateaux will be proportional to a series of con- potassium chloride-phosphate buffer or buffered
centrations of the substance dissolved in electro- KOH with pH varying between 6.7 and 10.4
lyte, in this case oxygen. (Heitmann, Buckles and Laver, 1967). The system
The system in figure 1 would be adequate for is totally enclosed in a thermostated water bath at
measurements in simple well-stirred solutions, but 37 °C, so that measurements relate to the normal
is not suitable for measurements in blood because body temperature (Laver and Seifen, 1965).
ELECTRODES FOR MEASUREMENT OF On AND CO, TENSIONS 733

Basic theory. solved, however, by many authors (Kolthoff and


Theoretical models have still to be proposed Lingane, 1952; Evans and Naylor, 1960; Birken-
which accurately describe the processes occurring bosch, 1967) to give time-dependent solutions for
at the indicating electrodes in spite of many recent both spherical and linear co-ordinate systems.
attempts (Davies, 1961; Berkenbosch, 1967; A much simplified steady state result can be
Hudson, 1967). A much simplified model will be derived by combining the Faraday law of elec-
considered which circumvents complex numerical trolysis (namely, that the amount of a product of
operations without deviating too far from the real electrolysis at an electrode is proportional to the
situation. It is assumed that the current in the current and the time for which it flows) and
system is the result of chemical processes occur- equation (1). If it is assumed that every molecule
ring at the cathode and is proportional to the reaching the cathode surface is reduced, then the
number of oxygen molecules diffusing to the current, i, due to the reduction process is given by
cathode in unit time (Kolthoff and Laitinen,
1940). It is also assumed that the membrane is i = n F (dQ/dt) •(3)
directly and tightly stretched over the cathode, where dQ/dt is transport rate of oxygen to the
and that there is only a thin layer of electrolyte cathode, F is the Faraday constant, and n is the
between the cathode surface and the membrane. net number of electrons involved in the reaction.
The primitive linear model to be described is Equations (1) and (3) can be combined to give
based on the laws of transport by diffusion (Fick, i = nFDA (dC/dx) (4)
1885). Fick's first law states that for a quantity
of substance, Q, being transported in a linear where x is measured from the cathode surface.
co-ordinate system, then For a slab of membrane (similar to fig. 4) with
concentrations Q and C2 on the surfaces, and
d Q / d t = - D A (dC/dx) (1) thickness x, equation (4) reduces to
Where D i? the diffusion coefficient, A is the i=nFDA [ ( Q - C J / x ] (5)
surface area across which Q is transported and for d is greater than C2.
dC/dx is the concentration gradient (fig. 4). The underlying assumption so far has been that
Fick's second law relates the way in which the electrolyte layer is negligibly thin, but Berken-
concentration, C, varies with distance, x, and time, bosch and Riedstra (1963) considered the effect
t, and is expressed by the second order differ- of a finite electrolyte layer for the linear diffusion
ential equation model and concluded that its effect is very small
d C / d t = D (d2C/dx2) (2) (one part in a hundred) compared to the mem-
Equation (2) is necessary only for a time- brane layer. Another important assumption is that
dependent solution for oxygen electrode current every oxygen molecule is reduced at the cathode
and is included for completeness. It has been surface, and thus C2 is very much smaller than Q
and can be ignored. Equation (5) can then be
written in its concentration form, as
i = nFDA (Q/x) (6)
| Thickness X
To express equation (6) in its final form in
terms of partial pressure of oxygen (Po2), it is
necessary to apply the law of Henry-Dalton
(Henry, 1803), which states that the gas concen-
tration in a solution is proportional to the gas
Concentrations
C,>C; pressure.
C = a Po2 (7)
The solubility coefficient a is a measure of the
quantity of gas absorbed per unit volume of
Volume element of unit surface area for an imaginary
solvent (in this case the membrane material) in
linear diffusion system. the presence of a partial pressure of 760 torr.
734 BRITISH JOURNAL OF ANAESTHESIA

6
J
5

4
1 J

2
3

• If y . ~7|
•/! . j . !.
1 .

100 200 300 400 500 600 700 800 900 100 200 300 400 500 600
— mV Po 2 torr
Cathode polarization voltage
with respect to anode
FIG. 5
Folarogram of oxygen at a theoretically perfect membrane-covered electrode. Electrode current,
i, in arbitrary units.

The final expression is thus: commercial electrode systems may not be entirely
linear over a wide range of Po2 (Heitmann, Buckles
i = (nFAaD/x) Po2 •(8) and Laver, 1967), and care should be taken over
It is important to note that both a. and D are calibration. It is usual to calibrate with oxygen-
temperature-dependent. For the exact nature of free N 2 gas, or gas tonometered with sodium
this dependance, the reader is referred to the sulphite solution, for the instrument zero, and
publications of Rogers and others (1962) and of room air (20.93 per cent O2 v/v) for the span.
Barrer (1947). Thus, if the temperature of the For samples of high Po2, the instrument should
polarographic cell is kept constant, the term be calibrated within the range in which it is ex-
nFAaD/x can be considered as constant in equa- pected to work.
tion (8), and the current in this limiting diffusion It is difficult to quote "time-constants" for oxy-
region is directly proportional to the prevailing gen electrode systems, since the numbers obtained
Po2 outside the membrane, will depend upon the nature and thickness of the
membrane material and upon the thickness of the
i.e., electrode current i = constantxPo 2 . electrolyte layer, but it is usually of the order of
This is depicted in figure 5 for a theoretically several seconds for a 90 or 95 per cent response
perfect electrode, which shows the plateaux where to a step change in Po2. The fastest response
equation (8) holds true. Oxygen electrodes are quoted, to our knowledge, is that of Beneken
usually polarized between - 600 mV to - 800 mV Kolmer and Kreuzer (1968) who reported a res-
with respect to the reference electrode, the actual ponse time of 0.2-0.25 sec for 95 per cent deflec-
voltage depending on the particular electrode and tion with a change of O2 concentration between
manufacturer. It is as well to note, however, that 0 and 100 per cent in both directions. The latter
ELECTRODES FOR MEASUREMENT OF O, AND CO, TENSIONS 735

finding is interesting since, in our own experience, dissociation curve of oxyhaemoglobin a consider-
the time response from a high Po2 to a low Po2 able error may occur in derived values of oxygen
is much longer than in the other direction and an saturation if the proper correction is not made.
electrode may well take several minutes to reach This also applies to the use of correction factors
"true zero" after the electrode has been exposed for drift, elapsed time between sampling and
to a sample of blood or gas with a high Po2. This measurement, and for temperature differences
effect may well be due to the overall electrode between the patient and the measuring electrode
reaction taking place at the cathode, (Kelman and Nunn, 1966).
O2 + 2H,O + 2e—>H 2 O 2 + 2OH- (i) The read-out system with most commercial
units consists of a high impedance voltmeter mea-
suring the voltage drop due to a current, i, across
It is generally assumed (Davies, 1961) that re- a larger resistance. This voltage drop is com-
action (ii) proceeds fast enough for all the pensated by means of a feed-back system, and the
hydrogen peroxide to be reduced and that none output read from an analogue moving coil meter
escapes into the bulk of the electrolyte. It has scale, calibrated in torr. Hahn (1969) described a
been suggested, however, by Seaton and Lloyd system of directly measuring the electrode current
(1969, personal communication) that enough and displaying the result on a digital voltmeter.
hydrogen peroxide can escape from the cathode Such a system naturally lends itself to data pro-
area to explain the long time-constants from high cessing, but however accurate the read-out system
to low Po2. The hydrogen peroxide will tend to becomes, it must always be remembered that the
diffuse slowly back to the cathode when the electrode itself has many limitations to its absolute
oxygen supply has been discontinued, and will accuracy.
be reduced according to equation (ii) producing
a current in the absence of any oxygen. THE CO2 ELECTRODE
It has already been stated that it is convenient As the CO3 electrode is in effect a pH electrode,
to calibrate the electrode by using gas mixtures of it may be helpful briefly to consider the principles
known oxygen tension, but it is also well known behind the measurement of pH. This measure-
that the output from an electrode is lower for a ment depends on the fact that if two solutions are
sample of blood than for the gas with which the separated by a membrane of special glass, a
blood has been equilibrated (Bishop, 1960; Polgar potential difference exists across the glass which
and Forster, 1960; Severinghaus, 1962). This bears a relationship to the activity of H+ ions on
difference can be reduced to a large extent by either side of the glass. This relationship is
stirring the blood sample, but is not completely
expressed by an equation, derived from the
eliminated (Heitmann, Buckles and Laver, 1967).
original Nernst equation by substituting "activi-
The blood-gas factor in fact depends on many
ties" for concentrations, hence,
variables, including the rate of stirring, the
influence of pressure and flow changes in the E = ( R T / F In (aH'+/aH"+) (1)
sample, the area of the cathode, and the nature where In is loge, R is the gas constant, T the
and thickness of the membrane. Adams and absolute temperature, F, Faraday's constant
Morgan-Hughes (1967) quote the blood-gas factor (96,494 coulombs), aH+ is H+ ion activity and E
for two commercial unstirred electrode systems as is potential measured in volts. The "activity" of
4 + 2 per cent and describe a thermostatically con- an H+ ion is a term used to indicate the behaviour
trolled bubble tonometer to calibrate oxygen of the ion when in an electrolyte solution. West
electrode systems. The correction factor for the (1963) illustrates the term "activity" by using the
blood-gas difference should be determined for analogy of a footballer who may be relatively
each electrode-membrane system, and for the
immobilized by the presence near him of a large
range of Po2 which is to be measured. A 4 per
number of other footballers, but also may show
cent correction at a Po2 of 500-600 torr may not
more activity with a few other players on the field
be critical to the clinician, but at tensions corres-
than when he alone is present. The term
ponding to the steep section of the sigmoid
"activity" is used, with certain conditions, to
736 BRITISH JOURNAL OF ANAESTHESIA

express the effective concentration of an ion as bicarbonate solution. This changes the pH of the
shown by its properties in solution. bicarbonate solution and causes a change in the
It can be shown (West, 1963) that aH+ is the potential difference across the pH-sensitive glass
hydrogen ion concentration [H+] in mol/litre of what is in effect a pH electrode. The potential
multiplied by an activity coefficient. And so equa- change is measured by reference electrodes as a
tion (1) becomes voltage governed by equation (4).
E = ( R T / F ) In -]) (2) The situation illustrated in the diagram is not
This can be rewritten, practical because of the number of CO2 molecules
necessary to change the pH of a large mass of bi-
E = ( R T / F ) (In [H'+] - In [H"+]) carbonate solution and the time necessary for this
(3) to happen. The CO2 electrode in fact measures
S0rensen (1909) defined pH (the degree of the pH change in a thin film of bicarbonate solu-
acidity or alkalinity of a solution) as equal to tion in close contact with the pH-sensitive glass
- log10[H+]. Thus equation (3) can be rewritten, of the electrode (fig. 7).
converting to log10 and changing sign,
E=2.303(RT/F)(pH" - pHO (4)
This is the potential difference that is registered
on the meter of a pH measuring system and is
equal to 61.5 mV per unit pH change at 37°C.
The CO2 electrode, first described by Stow,
Baer and Randall (1957) and Gertz and Loeschcke
(1958), uses a CO 2 permeable membrane to allow
CO2 to diffuse from an unknown sample of gas
or liquid into a bicarbonate solution until equili-
brium is reached. Figure 6 is a very simplified
diagrammatic representation of the electrical
circuit. CO2 diffuses through a membrane into a CO,

Sample of blood or gas

FIG. 7
Diagram of tip of carbon dioxide electrode.

Basic theory.
When CO2 diffuses through a membrane and
comes into equilibrium with water, the equili-
brium can be expressed by the well-known equa-
tions,
Reference
-electrode K K
CO2 H+ + HCOj

0
(5)
Container
If the liquid behind the membrane is a bicar-
bonate solution, the HCO3 ion at the righthand
Bicarbonate
solution
side of equation (5) is mostly derived from the
salt, but the equation, although written for water,
still holds approximately. The equilibrium of the
CO ;
Membrane righthand side of equation (5) is governed by the
sample first ionization constant of H,CO 3 , K u and so the
FIG. 6 expression can be rewritten,
Simplified diagram of circuit of carbon dioxide
electrode. K 1= ([H+][HCO3]/[H 2 CO 3 ]) (6)
ELECTRODES FOR MEASUREMENT OF O2 AND CO2 TENSIONS 737

Taking logarithms of both sides of (6) and If the CO2 electrode responded perfectly, the
defining pK t as - log Kx and pH as - log [H+] plot of this relationship on semilog paper would
(S0rensen, 1909) yields be a straight line with a slope of - 1 (fig. 8).
The relationship between the change of pH
P H = p K 1 + log ([HCO3]/[H2CO3])
(A pH) in the bicarbonate solution in a CO2 elec-
(7)
trode caused by a change, A Pco2, in Pco2 of the
This equation expresses the pH of the film of bicarbonate solution has been defined by Sever -
bicarbonate solution in a CO2 electrode if there inghaus and Bradley (1958) as the sensitivity
is a change of H 2 C0 3 in the film. The effect of factor " S " for an electrode.
CO, on H,CO 3 is expressed by the left side of
the equilibrium equation (5). [H2CO3] is, by A p H = - S A log Pco,
Henry's law, linearly proportional to the partial ••• S = - ( A p H / A l o g P c o 2 ) .(10)
pressure of CO2 in the sample outside the mem- This would follow immediately from equation
brane and therefore can be replaced in equation (9) if it is assumed that S = - 1 and, as has been
(7) by aPco,, where a is the solubility coefficient said before, this would define the relationship
of CO2. So equation (7) can be rewritten as between pH and log Pco2 in a perfect electrode.
Equation (9) is, however, a gross over-simplifica-
PH=pK1 + log ([HCO3]/«Pco2 (8)
tion of a complex group of chemical reactions and
The addition of the HCO3 in equation (8) to more exact dissociation equations for a CO 3
the total of HCO3 ions from NaHCO3 in the electrode (Lunn and Mapleson, 1963; Severing-
bicarbonate solution of a CO2 electrode (see haus and Bradley, 1958) must be used to define S.
above) has a negligible effect and hence [HCOi] These take into account the dissociation of
is essentially constant. So equation (8) becomes NaHCO 3 and the laws of electroneutrality as well
pH = pKj + log [HCO3] - log a - log Pco2 as those governing the solution of CO2 in the
= constant - log Pco2 (9) electrolyte. The value of S changes with different

8.500

8.000

7.500

pH FIG.8
pH/log Pco2 plot.
7.000

6.500

6.000
100 1000
Pco, torr
738 BRITISH JOURNAL OF ANAESTHESIA

bicarbonate concentrations in the electrode, but in To


voltage
certain conditions of concentration of bicarbonate amplifier
S can be regarded as constant. So equation (9)
can be rewritten, to acknowledge imperfections in
the electrode, as
pH=constant - S log Pco2 (11)
where S is the sensitivity factor of the electrode
or, in other words, the slope of the plot of pH
against log Pco,. Reference electrode
In summary, the measurement of a change in
Pco2, A Pco2, of a sample introduced into a CO2
Glass rod
electrode, depends on the measurement of the
change in pH of a film of bicarbonate solution
separated from the sample by a CO2 permeable
membrane. The Pco2 of the sample changes the
Pco2 of the bicarbonate solution in the electrode
and hence the H+ activity of the solution. The pH sensitive glass

potential across the pH-sensitive glass of the


electrode increases or decreases, and this increase
or decrease, measured in mV, is registered by the Matrix
meter in the system. The change is governed by
equation (11). Thus if the S of an electrode is - 1, " O " ring
for each change in Pco2 by a factor of 10, e.g.
from 1 torr to 10 torr, or from 10 torr to 100 torr, Membrane Bicarbonate solution
there is a unit change in the pH of the bicarbonate FIG. 9
solution and hence a potential change registered Diagram of carbon dioxide electrode.
on the meter of 61.5 mV.
As previously mentioned, however, for many or cellophane, which holds a thin film of solution
electrodes S is not equal to - 1 but is almost containing 0.1 M NaHCO3 with some NaCl and
always nearly so. The manufacturers of measuring sometimes silver chloride. Another electrode is
systems allow for this by providing controls to needed to complete the cell circuit for the glass
match the measuring instrument to the S of any electrode, and this is again a silver-silver chloride
given electrode between two calibrating gases of electrode in physical contact with the bicarbonate
different Pco2's. This assumes that S is constant solution. These two reference electrodes produce
between two calibrating gases, but our experience, a constant potential with regard to the chlorine
and apparently that of Severinghaus and Bradley solution in which they find themselves (Eisenman
(1958), would suggest that this is often not so. et ah, 1966) and therefore a constant potential
The errors introduced by this assumption are difference reading with respect to each other.
probably insignificant in clinical practice, but The special glass of the CO2 electrode is
might be of importance in clinical research, and separated from a test solution by a teflon mem-
their magnitude is at present under study. brane permeable to CO2 molecules. When the
The modern CO2 electrode, of which that electrode is in use, the Pco2 of an unknown solu-
manufactured by Radiometer* is a good example, tion comes into equilibrium with the Pco2 of the
consists of a silver-silver chloride electrode in a bicarbonate solution in the matrix immediately
closed container filled with a phosphate buffer of adjacent to the special H+ ion sensitive glass (fig.
pH about 6. One end of this container is made of 9). This changes the pH or the H+ ion concentra-
special H+ ion sensitive glass (fig. 9) and is tion in the bicarbonate solution in the immediate
covered by a matrix of nylon mesh, Joseph paper neighbourhood of the special glass and therefore
changes the electrical potential existing across the
* Radiometer A.S., Copenhagen, Denmark. glass between the two solutions. This change is
ELECTRODES FOR MEASUREMENT OF O, AND CO, TENSIONS 739

amplified and read on a sensitive voltmeter. Both tration in per cent CO, v/v. This is multiplied by
electrodes and the special glass membrane must the barometric pressure of the day to give Pco2.
be kept at the same constant temperature. It has The hinge gas reading is, of course, checked
been agreed (New York Academy of Sciences, before and after each sample measurement.
1966) that the temperature used should be 37 "C This method has two advantages. The semilog
to be comparable with the normal physiological graph paper chosen gives considerably better dis-
temperature in man. crimination than the best of several commercial
Calibration of a CO* electrode. meters available (Trubuhovich, 1968). A range of
Calibration of CO, electrodes is simplified by 10-100 torr is expressed on 250 mm of log scale
the fact that there is no blood-gas difference for on the paper used, whereas 10-100 torr is ex-
CO, (Adams and Morgan-Hughes, 1967). The pressed on 176 mm on the longest and 114.3 mm
electrode can therefore be calibrated with gas and on the shortest commercial scales, and the twice-
used to measure the Pco, of blood. Manufacturers daily check with three gases is virtually certain to
of electrode systems find it convenient to calibrate disclose a faulty electrode.
electrodes with two gases of fairly widely differing The use of linearizing circuits enable Pco2 to
CO, contents, e.g. 3 and 7 per cent v/v. One gas be read directly and digitally from a change in
is given its correct value on the scale of the instru- electrical output from the electrode and should
ment and the other calibrating gas is matched to result in increased accuracy. Any linearizing
the scale of the measuring instrument by means circuit must, however, by definition assume a
of a control. This allows for the fact that the S linear relationship between electrical output or pH
of an electrode may be less than 1 but assumes and log Pco,, at least for the range of Pco2 under
that the pH/log Pco2 response of the electrode is study. The extent to which circuits can be modi-
linear. fied to reduce inaccuracies from this assumption
We see advantages of constructing a semilog is under study.
plot of pH or mV/per cent CO2 (v/v) for the Trubuhovich (1968), in a study of the perform-
electrode in use. It would be logical to construct ance of CO, electrodes, made observations which
a plot using 100 and 0 per cent CO, v/v but this will be more fully described elsewhere, and in
is not practicable. CO, electrodes do not respond respect of calibration again drew attention to two
to 0 per cent CO, v/v, and because of the log- potential sources of error. He showed first of all
arithmic relationship between mV or pH and that if humidified gas is passed into an electrode
Pco,, at 100 per cent CO2 v/v it becomes very at a rate just fast enough to compensate for leaks
difficult to inscribe a point with any accuracy. We and diffusion (Lunn and Mapleson, 1963), a very
investigate the characteristics of any new stable and reproducible equilibration pattern can
electrode with a Wosthoff pump* giving 1 per be achieved. In the example shown, the output of
cent step changes of CO2 content from 1 to 10 the electrode changes by less than 0.001 of a pH
per cent v/v, and with a cylinder of nominal 5 per unit over periods of 10-20 minutes. If the elec-
cent CO2 (v/v) which has been carefully analyzed trode "dries out", however, the pH reading
with a Haldane apparatus. The 5 per cent gas is changes some 10 minutes later and may subse-
called the "hinge" gas and the electrode is exposed quently be constant at its new value or irregular
to it after each exposure to another gas. In this (fig. 10). This source of error may be diminished
way, an accurate calibration line is drawn for each or avoided, either by using warmed and humidi-
electrode using a relatively large number of points. fied calibrating gases or by introducing very small
In normal use, the electrode is exposed to the quantities of water into the equilibration chamber
"hinge" gas, a low gas and a high gas twice a day. from time to time. The ideal situation is when
If the pH values observed agree with the values the calibrating gas bubbles through a small
on which the calibration line is based to within amount of water remaining in the equilibration
2 or 3 thousandths of a pH unit, the original line chamber.
is used to give the value of an unknown concen- Trubuhovich also showed that if gases are in-
* H. Wosthoff. O. H. G., 463. Bochum, Hagenstrasse, troduced into the equilibration chamber at fast
30, Germany. flow rates, the output from the CO, electrode
740 BRITISH JOURNAL OF ANAESTHESIA

I I!
I -I I! i

ilLiiJ: .j.
1 . 4 - .1

FIG. 10
Carbon dioxide 5 per cent v/v dry gas at 5 ml/min through water in equilibration chamber.
Upper record pressure in equilibration chamber, lower record pH 0.1 pH unit full scale. Time
scale in minutes. Pressure drops when water dries out and bubbling spikes almost cease. pH
reading becomes acid (Trubuhovich, 1968).

FIG. 11
Upper record pressure, lower record pH. 5 ml/min humidified gas. Time scale minutes. At A,
gas flow turned off. Alkaline "decay" drift starts 20 sec later. At B, gas flow increased to 25 ml/
min. giving more acid reading by 0.010 pH units. At C, gas flow turned off. Recovery from
pressure change overtaken by "decay" drift, showing no recognizable end point (Trubuhovich,
1968).
ELECTRODES FOR MEASUREMENT OF O2 AND CO, TENSIONS 741

changes, no doubt because the Pco2 is increased Heitmann, H., Buckles, R. G., and Laver, M. B. (1967).
in the chamber by increased ambient pressure. Blood Po2 measurements: performance of micro-
electrodes. Resp. Physiol., 3, 380.
When the gas has been flowing at a fast rate and Henry, W. (1803). Experiments on the quantity of gases
is stopped, the electrode reading falls, first of all absorbed by water at different temperatures and
due to pressure changes, but this fall is later over- under different pressures. Phil. Trans, roy. Soc,
93, 29.
taken by the fall due to leaks and diffusion. There Heyrovsky, J. (1922). Electrolysis with the dropping
is no recognizable end point such as can be mercury electrode. Chemicke Listy, 16, 256.
achieved with a slow—5 ml per minute—flow of Hudson, T. A. (1967). Measurement of oxygen tension
by the oxygen cathode. Med. biol. Engng., 5, 207.
humidified gas into a wet chamber (fig. 11). Kelman, G. R., and Nunn, J. F. (1966). Nomograms
Trubuhovich's (1968) re-examination of the for correction of blood Po2, Pco2, pH and base
performance of CO2 electrodes has suggested to excess for time and temperature. J. appl. Physiol.,
21, 1484.
us that their reproducibility is high, and if allow- Kolthoff, I. M., and Laitinen, H. (1940). The volta-
ance can be made for the fact that the S of an metric determination of oxygen. Science, 92, 152.
electrode is a perfectly straight line over only a Lingane, J. J. (1952). Polarography, Vol. I and II.
New York: Interscience.
limited range, the potential accuracy of CO2 Laver, M. B., and Seifen, A. (1965). Measurement of
electrodes is quite considerable. This potential blood oxygen tension in anesthesia. Anesthesiology,
accuracy is still under study. 26, 73.
Linden, R. J., Ledsome, J. R., and Norman, J. (1965).
REFERENCES Simple methods for the determination of the con-
Adams, A. P., and Morgan-Hughes, J. O. (1967). Deter- centrations of carbon dioxide and oxygen in blood.
mination of the blood-gas factor of the oxygen Brit. J. Anaesth., 37, 77.
electrode using a new tonometer. Brit. J. Anaesth., Lunn, J. N., and Mapleson, W. W. (1963). The Sever-
39, 107. inghaus Pco2 electrode: a theoretical and experi-
Sykes, M. K. (1967). pH and blood-gas mental assessment. Brit. J. Anaesih., 35, 666.
analysis: Part I. Anaesthesia, 22, 575. Myers, A. W., Meyer, J. A., Rogers, C. E., Stannett,
(1968). pH and blood-gas analysis: V., and Szwarc, M. (1962). The Permeation of
Part II. Anaesthesia, 23, 47. Water Vapour. Tappi Monograph Series 23.
Barrer, R. M. (1947). The solubility of gases in elas- New York Academy of Sciences (1966). Report of ad
tomers. Trans. Faraday Soc, 43, 3. hoc committee on the methodology of the New
Beneken Kolmer, H. H., and Kreuzer, F. (1968). York Academy of Sciences Conference; current
Continuous polarographic recording of oxygen concepts of acid-base measurement. Ann. N.Y.
pressure in respiratory air. Resp. Physiol., 4, 109. Acad. Sci., 133, 259.
Berkenbosch, A. (1967). Time course of response of Peters, J. P., and van Slyke, D. D. (1958). Quantitative
the membrane-covered oxygen electrode. Ada Clinical Chemistry: Methods, Vol. 2, 5th ed., p.
physiol. pharmacol. NeerL, 14, 300. 244. Baltimore: Williams and Wikins.
Riedstra, J. W. (1963). Temperature effects in Polgar, G., and Forster, R. E. (1960). Measurement of
amperometric oxygen determinations with the oxygen tension in unstirred blood with a platinum
Clark electrode. Acta physiol. pharmacol. Neerl., electrode. J. appl. Physiol., 15, 706.
12, 144. Prat, S. (1926). Die Anwendung der polarographischen
Bishop, J. M. (1960). Measurement of blood oxygen methodic in der biologic Biochem. Z., 175, 268.
tension. Proc. toy. Soc. Med., 53, 177. Rogers, C. E., Meyer, J. A., Stannett, V., and Szwarc,
Clark, L. C. (1956). Monitor and control of blood and M. (1962). Some factors affecting the Permeability
tissue oxygen tensions. Trans. Amer. Soc. art. Constant. Tappi Monograph Series, 23, 12.
intern. Org., 2, 41. Severinghaus, J. W. (1962). Electrodes for blood and
Davies, P. W. (1961). The oxygen cathode. Phys. gas Pco2, Po2 and blood pH. Acta anaesth. scand.,
Techn. Biol. Res., 4, 137 Suppl. II, 207.
Brink, F. (1942). Micro-electrodes for measuring (1968). Measurement of blood gases. Ann. N.Y.
local oxygen tension in animal tissues. Rev. sci. Acad. Sci., 148, 115.
Instrum., 13, 524. Bradley, A. F. (1958). Electrodes for blood Po2
Eisenman, G., Bates, R., Mattock, G., and Friedman, and Pco2 determination. J. appl. Physiol, 13, 515.
S. M. (1966). The Glass Electrode. New York: Shepard, R. H. (1956). Polarographic oxygen tension
Interscience. determination: factors concerning calibration
Evans, N. T. S., and Naylor, P. F. D. (1960). The change in blood. Fed. Proc, 15, 169.
measurement of the partial pressure of oxygen in S0rensen, S. P. L. (1909). Enzymstudien II. tjber die
vivo. J. Polarog. Soc, 2, 26. messung und dei bedeutung der wasserstofnonen-
Fick, A. (1885). Ueber die diffusion. An. Phys. Lpz., konzentration bei enzymatischen prozessen. Bio-
94, 59. chem. Z., 21, 131 (see p. 159).
Gertz, K. H., and Loeschcke, H. H. (1958). Elektrode Stow, R. W., Baer, R. F., and Randall, B. F. (1957).
zur bestimmung des CO2-drucks. Naturwiss, 45, Rapid measurement of the tension of carbon
160. dioxide in blood. Arch. phys. Med., 38, 646.
Hahn, C. E. W. (1969). The measurement of oxygen Trubuhovich, R. F. (1968). B.Sc. Thesis, Oxford
microcathode currents by means of a field-effect University.
transistor operational amplifier system with digital West, E. S. (1963). Textbook of Biophysical Chemistry,
display. J sci. Instr. (Jfhys.E.), 2, 48. 3rd ed., p. 76. New York: Macmillan.

You might also like