QUALIFICATION
FOR
TYPE: VERTICAL
MODEL: OLSC-103H-12
Date: Date:
CONTENT
System Description
Design qualification
Installation qualification
Operational qualification
Performance qualification
Cleaning
Maintenance program
SYSTEM DESCRIPTION
Vertical model.
Robust construction outer chamber is made of Mild Steel duly pretreated & finished with powder coated paint
for lasting finish OR Stainless Steel Sheet (304 Grade).
Inner chamber made of STAINLESS STEEL SHEET (316 Grade).
Fitted with durable coaxial blower for forced air circulation at triple walled back to maintain temperature
uniformity in the chamber.
Outer front double walled door is insulated and is fitted with magnetic tape for air tight closing for no
temperature loss, provided with lock and key arrangement.
Illumination lamp is fitted inside the chamber for easy visibility.
A full view inner Acrylic door enables inspection and monitoring of inner chamber specimens without
disturbing the process temperature.
High grade Glass wool between outer and inner chamber for minimal thermal losses.
Cooling is done by finned tube evaporator lie in the air circulation path by ISI marked hermetically sealed CFC
free Compressor & R-134 an eco friendly refrigerant.
Heating elements are placed in the path of moving air duly insulated from the body.
Inner chamber has ribs for placing the shelves at convenient levels. Supplied with adjustable high quality
stainless steel shelves for completely flexibility in use.
Control panel having all controls and circuitry with Temperature controller cum Indicator Microprocessor
based with digital LED display for Set Value (SV) & Process Value (PV) to select any desired temperature
provided.
Caster wheels with stopper mounted for easy portability. Supplied with cord and plug.
Operating Voltage: 230 Volts AC (50 Hz).
WHO-GMP, CE, ISO 9001, ISO 13485, ISO 14001 & OHSAS 18001 certified.
Technical specification:
Approx. Volume (Ltr) : 350
No. of Shelves (adjustable) : 3
Temperature Range : 0°C to 60°C (Adjustable)
Temperature Accuracy : ±0.5°C to ±1°C
Humidity Range : 40% to 95% RH, ± 3%
Controller : Microprocessor based safety PID controller with inbuilt thermostat
Alarm : Safety alarm for temperature overshoot
Display : Digital LED Display for Set Value (SV) and Process Value (PV)
Relay : Solid state electronic relay with protective heat sink.
Auto Tuning System : System is compatible with auto tuning system
Air Circulation : By forced convection System
Insulation : By Glass wool
Operations : Nearly silent operation with ultra-low vibration
Nominal voltage : 220-230 Volts, 50 Hz Single Phase
Printing Facility : By Built in Mini Printer to record temperature.
GSM System : Alarm sent to predefined Mobile Numbers
Data Logger : 4 point temperature data logger with sensor placed at specific points at
internal chamber with memory backup for minimum 3 months.
Regulated Voltage : By Suitable Voltage stabilizer
DQ OF STABILITY CHAMBER
PURPOSE: - The purpose of the design qualification [DQ] is to ensure that the instrument
can be operated as expected
Checked by:
IQ OF STABILITY CHAMBER
IQ: INSTALLATION QUALIFICATION FOR STABILITY CHAMBER
PURPOSE:-
1) To have visual checking of the instruments for any damage during transportation
3) To ascertain and approve that the adequate facilities shall be provided that meets the
requirement for the installation of the instruments
PARTICULARS REMARKS
Any breakage/Leakage in Stability Chamber :- No OK
Level /Balance of stability chamber :- Properly balanced OK
Power cables :- Fixed with the instruments OK
Any loose connection / fittings :- No OK
Rust if any :- No OK
Properly insulated :- Yes OK
Compressor & Fan motor properly fixed and checked :- Yes OK
Electric power supply 220-230V AC (single phase) Shall be available earthing must
OK
be proper before installation :- Yes
Inner acrylic door is fitted properly balanced :- Yes OK
Insulated double walled front door properly balanced and fitted
OK
:- Yes
Door lock with key and four castor wheels :- Yes OK
Control panel
Microprocessor based temp. controller :- Yes
Thermal Printer :- Yes
OK
USB port :- Yes
Main ON/OFF Switch :- Yes
Checked by:
OQ OF STABILITY CHAMBER
OQ: OPERATIONAL QUALIFICATION
PURPOSE: - The purpose of the operation qualification [OQ] is to ensure that the instrument
can be operated as expected and described and further to established the
standard operating procedure [SOP]
Checked by:
PQ OF STABILITY CHAMBER
PQ: PERFORMANCE QUALIFICATION
PURPOSE: - The purpose of the performance qualification [PQ] is to ensure that the instrument can be
operated as expected and describe and further to establish the standard operating
procedure [SOP]
After cleaning the chamber & front door is closed, the instrument is switched ON and
display will ON. There are soft KEYS on the keypad provided for setting the required
parameters.
PARTICULARS REMARKS
Switching ON the Mains Press SET key to Change setting in set mode by Increment
OK
& Decrement key and Press enter to start process.
Digital display to see the parameter (Temp & Humidity) :- OK OK
Temperature reading of chamber :- OK OK
Humidity reading of chamber :- OK OK
Thermal printer :- OK OK
Data logging through USB Port :- OK OK
Mains ON/OFF switches :- OK OK
Observation:
Set points: 25°C – 60% Rh, 40°C – 75% Rh, 30°C – 65% Rh, 25°C – 40% Rh
1) For Temperature
At Temperature : 250C, Accuracy :- ±10C approx
At Temperature : 300C, Accuracy :- ±0.5 to ±1.00C approx
At Temperature : 400C, Accuracy :- ±0.5 to ±0.90C approx
OK
2) For HUMIDITY
At Humidity : 60% Rh, Accuracy :- ±5%
At Humidity : 75% Rh, Accuracy :- ±3%
At Humidity : 65% Rh, Accuracy :- ±3% to ±5%
At Humidity : 40% Rh, Accuracy :- ±5%
Note:- accuracy will be maintained within 40 min after achieving the set value
All the reports are verified by
2. Cleaning Operation:
a) Wipe all the traces of organic solvent I water I moisture by tissue paper.
b) Clean all the dust from the working table by tissue paper.
c) Close the door/sash of the instrument.
3. Maintenance Program:
When the instrument is under breakdown I maintenance I not functioning, put the tag stating 'OUT
OF ORDER' and inform to the instrument manufacturer I servicing person as per the requirement.
After completion of repairing Servicing job, instrument should be checked for appropriate working
by the Q.C. Officer I Chemist.
Engineer prepares the service report and Q. C. Department retains one copy of the report.
Calibrate the instrument prior to use.
4. Records:
Adequate records shall be maintained showing the calibration observation.
Records of all maintenance shall be maintained.
Contact :