Extrinsic Factors (II, V, VII, X) Intrinsic Factors (VIII, IX, XI, XII)
Detector Clot (Other) Detector Clot (Other)
End Point 50 % End Point 50 %
Maximum Time 180 sec Maximum Time 180 sec
Sensitivity Low Gain Sensitivity Low Gain
WARNING WARNING
Protein S Deficient Plasma set at Diluent Holder Protein C Deficient Plasma set at Diluent Holder
CLICK CLICK
Test Protocol for CA-500 -(5) System software version 00-09
WARNING
AT III Reagent dilute 1:2 with Owren's veronal buffer
before use.
WARNING
Urokinase Reagent set at Diluent Holder
CLICK