To readers of this scientific data
Sysmex corporation
Scientific division
4-4-4 Takatsukadai, Nishi-ku, Kobe 651-2271 JAPAN
Tel : 078-991-1911
Fax : 078-991-1917
E-mail address : scientific@sysmex.co.jp
1
CA-500
1) Reproducibility - (TT)
Results
Comment
Control plasma N and Control P were measured 10 times consecutively using Test Thrombin
Reagent to test within-run reproducibility. The results were excellent and the CVs were less
than 4.50%.
2
CA-500
Results
CA-500 vs CA-6000
(sec)
50
40
CA-500
30
20 n = 26
r = 0.973
y = 0.758x + 5.53
10
10 20 30 40 50 (sec)
CA-6000
Comment
Correlation of Thrombin Time on the CA-500 and the CA-6000 using Test Thrombin Reagent
in both instruments.
Results were good and the correlation was r=0.973 and y=0.758x+5.53.
3
CA-500
Results
CA-500 vs CA-7000
(sec)
50
40
CA-500
30
20 n = 26
r = 0.975
y = 0.793x + 5.039
10
10 20 30 40 50 (sec)
CA-7000
Comment
Correlation of Thrombin Time on the CA-500 and the CA-7000 using Test Thrombin Reagent
in both instruments.
Results were good and the correlation was r=0.975 and y=0.793x+5.039.
4
CA-500
Results
CA-500 vs CA-1500
(sec)
50
40
CA-500
30
20 n = 26
r = 0.957
y = 1.040x + 0.064
10
10 20 30 40 50 (sec)
CA-1500
Comment
Correlation of Thrombin Time on the CA-500 and the CA-1500 using Test Thrombin Reagent
in both instruments.
Results were good and the correlation was r=0.957 and y=1.040x+0.064.
5
CA-500
Results
CA-500 vs CA-50
(sec)
50
40
CA-1500
30
20 n = 26
r = 0.979
y = 0.903x + 2.520
10
10 20 30 40 50 (sec)
CA-50
Comment
Correlation of Thrombin Time on the CA-500 and the CA-50 using Test Thrombin Reagent in
both instruments.
Results were good and the correlation was r=0.979 and y=0.903x+2.520.
6
CA-500
Results
Comment
Heparin was added to pooled plasma, and the heparin sensitivity on the Thrombin Time (TT)
result was examined. The reference interval was 14-21 seconds at reconstitution in normal vol
ume and 8-13 seconds at reconstitution in half normal volume.
As a result, up to 0.10 units/mL heparin at reconstitution in normal volume made no influence
on TT results. In heparin at reconstitution in half normal volume, it was influence by 0.10
units/mL heparin.
7
CA-500
Results
80
Clotting time
60
40
20 reference interval
14 - 21 seconds
60
Clotting time
40
20
reference interval
8 - 13 seconds
0
Results
(sec)
60
50
Clotting Time
40
30
20 reference interval
14 - 21 seconds
10
Comment
The fibrinogen sensitivity on the TT was examined. Low-concentration fibrinogen sample with
depleted fibrinogen and high-concentration sample with added fibrinogen were prepared.
These two samples were mixed to prepare a series of samples with various fibrinogen concen
tration. Fibrinogen concentration exceeding the reference interval(14-21 seconds) was ob
served at 141mg or less and 812mg or more.