Professional Documents
Culture Documents
DOI:10.3233/PRM-190660
IOS Press
1 Research Article
f
3
roo
4 devices
5 Ginny Palega,∗ , Wendy Altizerb , Rachel Maloneb , Katie Ballardb and Alison Kregerc
rP
6
a Montgomery County Infants and Toddlers Program, Rockville, MD, USA
7
b MilestonesPhysical Therapy, Hurricane, WV, USA
8
c Wheeling University, Wheeling, WV, USA
9 Abstract.
tho
PURPOSE: With children who are unable to stand or walk independently in the community, therapists commonly use standing
10
Au
11 devices to assist lower-extremity weight-bearing which is important for bone and muscle health. In addition, positioning in
12 hip abduction may improve hip stability and range of motion. This is the first study to explore the effect of angle of inclination,
13 hip abduction, body orientation, and tone on weight-bearing in pediatric standing devices.
14 METHODS: This descriptive exploratory study used a convenience sample of 15 participants (2 with normal tone, 5 with
15 generalized hypotonia, and 8 with hypertonia) (mean age of 5 years and 10 months, range of 3 years 4 months to 9 years 7
d
16 months); 13 of whom used standing devices at home, as well as 2 typically developing siblings (normal tone). Each child
17 stood in 36 positions to measure the amount of weight-bearing through footplates.
cte
18 RESULTS: Weight-bearing was highest with 60 degrees of abduction and no inclination (upright) in supine positioning for
19 children with low and normal tone. Children with high muscle tone bore most weight through their feet with no abduction
20 (feet together) and no inclination (upright) in prone positioning. Overall, supine positioning resulted in more weight-bearing
21 in all positions for children with low and normal tone. Prone positioning resulted in slightly more weight-bearing in all
22 positions for children with high tone.
rre
CONCLUSIONS: Weight-bearing was affected by all three of the variables (inclination, abduction, and orientation) for
23
participants with high, normal, and low tone. To determine optimal positioning, all standers should include a system to
24
measure where and how much weight-bearing is occurring in the device.
25
27 1. Introduction and literature review mend standing programs for children who are limited 32
Un
∗ Corresponding
occurred [4, 5]. Load bearing through the legs with 38
author: Ginny Paleg, PT, DScPT, Montgomery
County Infants and Toddlers Program, 420 Hillmoor Dr, Silver
resultant muscle contractions is thought to be the 39
Spring, MD 20901, USA. Tel.: +1 3014524656; E-mail: ginny@ mechanism for stimulating bone growth and thus 40
42 decrease spasticity, assist in maintaining hip and knee neither described nor detailed. Also, the influence of 94
43 range of motion (ROM), and increase and/or maintain inclination and orientation was not explored. Clini- 95
44 hip stability [7]. cians report many reasons for choosing prone versus 96
45 Pountney and colleagues described an associa- supine positioning and/or varied inclination positions 97
46 tion between improved hip stability and positioning in standing devices [3]. There is little research to sup- 98
47 in moderate hip abduction in standing, sitting, and port this reasoning. To date, no studies have measured 99
48 lying positions [8, 9]. Three other studies [10–12] or compared weight-bearing during differing degrees 100
49 suggested standing with the legs abducted 60◦ (30◦ of hip abduction or at different inclinations while 101
f
50 away from the midline for each leg) prevents and standing. They also have not explored the influence 102
roo
51 even improves hip subluxation as well as prevents hip of body orientation or tone on weight-bearing. 103
52 ROM loss and adductor contractures in children with The purpose of this study was to explore the 104
53 cerebral palsy. However, therapists and participants effects of three differing angles of inclination (upright 105
54 report 60◦ of total abduction can be uncomfortable [0◦ ]/15◦ /30◦ ), hip abduction (feet together [0◦ ]/30◦ / 106
55 [12]. The rationale for the influence of hip abduc- 60◦ total of bilateral hips), and body orientation 107
rP
56 tion on hip stability comes from the femoral head (supine vs prone) in individuals with different types of 108
57 applying force to the growth plate of the acetabulum, muscle tone (high, low, normal) on amount of weight- 109
58 with resultant bone response leading to more nor- bearing through the feet in two of three models of 110
tho
60 becomes more covered by the acetabulum and may
61 lead to more normal development of the neck of the
62 femur [13]. Although the research evidence is weak or 2. Methods 112
66 of hip abduction in standing programs without fully This was an exploratory descriptive study. Partic- 115
67 understanding its effects on weight-bearing. ipants were recruited from one therapist’s caseload 116
68 Standing devices, commonly referred to as as a convenience sample of all children who rou- 117
69 standers, are defined as durable medical equipment tinely used standing devices at home and received 118
which support the feet, knees, hips, and trunk, so the physical therapy at one outpatient physical therapy
d
70 119
71 user can remain upright in biomechanical alignment clinic. Participants were excluded if they had a history 120
cte
72 [16]. Standing devices are commonly used in schools, of uncontrollable seizures (>30 per day), autonomic 121
73 clinics, hospitals, homes, and other community set- dysreflexia, oxygen saturation measured by pulse 122
74 tings to positively impact body structure and function oximetry below 90 percent, or skin concerns (redness 123
75 [7]. Some models allow the user to transition from a or open lesions). 124
76 sitting position to standing and some can accommo- Typically developing siblings who attended clinic 125
rre
77 date up to 45◦ of flexion contracture at the hip, knee, on the data collection day were also recruited as a 126
78 and/or ankle. A few standing devices allow for hip comparison group. Children and their families were 127
79 abduction up to 30◦ on each side (60◦ total). Some advised of the opportunity to participate during their 128
80 standers are set in a fixed orientation (supine, upright, therapy session immediately prior to the data collec- 129
co
81 prone) while others allow for all three orientations tion day. 130
82 (multi-positional). Recently, some dynamic models Descriptive data was collected on all participants 131
83 (ones that allow for active dorsiflexion, rocking ante- at the start of the session including age, sex, diagno- 132
84 rior/posterior, and leaning in all directions) have sis, weight, and muscle tone. Typically developing 133
Un
85 been developed. There are also models that do not siblings were determined to have normal tone via 134
86 accommodate abduction but have wheels or power physical exam. Hypertonia (or high tone) was defined 135
87 for independent mobility while standing, as well as as an “increased stretch reflex” [17] in both elbows 136
88 wheelchairs that allow the person to be positioned in and knees. Hypotonia was determined using the def- 137
89 various stages of sit-to-stand, including upright. inition of “Hypotonia Syndrome” [18] in addition to 138
90 Two prior studies that directly measured weight- the participant having at least 5 out of 8 of the follow- 139
91 bearing found that subjects bore 78% [4] and 68% ing findings: decreased strength, hypermobile joints, 140
92 [5] of their body weight in their feet while using increased flexibility, delayed motor skills, leaning 141
93 standing devices. In these studies, positioning was on supports, rounded shoulders, decreased activity 142
G. Paleg et al. / Inclination, hip abduction, orientation, and tone affect weight-bearing in standing devices 3
f
roo
Fig. 1. Conditions and Design of Study.
rP
145 to data collection by placing the participant fully
146 clothed (with shoes and orthotics) supine on a scale
147 (WC Redmon Pet Scale from Amazon.com). This Fig. 2. Three Standers used in study; Zing, HLT Superstand, Stanz.
148 scale was calibrated with a five-pound weight before
tho
149 each child was weighed. Study risks and benefits were weight at the beginning of each measurement series. 183
150 reviewed with each family and child. Parent consent The child was placed in the stander by their regular 184
151 and child assent to participate was obtained prior to therapist, who positioned feet, ankles, knees, hips, 185
152 start of data collection. and pelvis in alignment. If the participant typically 186
153 Each participant chose index cards from a tray used orthotics during their standing program, these 187
Au
154 which randomized their order of conditions including were used during data collection. The Face, Legs, 188
155 brand of standing device, inclination (0◦ /15◦ /30◦ ), Activity, Cry, and Consolability Scale (FLACC) scale 189
156 amount of abduction (0◦ /30◦ /60◦ ), and orientation was administered to ensure that children were com- 190
157 (prone/supine) (see Fig. 1 for study design and fortable and not experiencing pain (particularly for 191
158 conditions). Total weight-bearing was measured by those children not able to express this verbally). The 192
two digital postal scales mounted on the stand- FLACC scale has been shown to be valid and reliable
d
159 193
160 ing device footplate, one beneath each foot (Smart in measuring pain and non-pain distress in children 194
cte
161 Weigh Digital Shipping and Postal Weight Scale from [19]. 195
162 Amazon.com) on the standing devices. Weight mea- After achieving body alignment in the standing 196
163 surements from both scales were added together and device, the randomized conditions of inclination and 197
164 divided by participant’s total weight and represented abduction were applied. Leg internal/external rota- 198
165 as a percentage. tion was not controlled. The sole of the shoe always 199
rre
166 This study used three standing devices: 1) Jenx maintained contact with the scale. The positioning 200
167 Standz, 2) EasyStand Zing, and 3) Prime Engineering therapist conducted the FLACC at each position to 201
168 HLT Superstand with Abduction System (see Fig. 2). ensure the child was comfortable and willing to con- 202
169 They were all multi-positional standers allowing up tinue with the study. Inclination was measured at each 203
171 inclination. Each participant was positioned in two of iPhone). Angle of abduction was determined by mea- 205
172 the three standers. Devices were chosen out of con- suring the inseam length and ensuring this equaled the 206
173 venience, as they were loaned free-of-charge by the distance between the feet. This configuration resulted 207
Un
174 manufacturers and their footplates accommodated the in an equilateral triangle, in which each angle is 60◦ . 208
175 mounted scales. The investigators removed the heel When the distance between the feet was half the mea- 209
176 cups and straps from all three models and mounted the surement of the inseam, the angle between the two 210
177 digital scales on the footplates (See Figs. 3A and 3B). femurs (abduction) was determined to be 30◦ total. 211
178 Five therapists were involved in data collection. This methodology was developed by Doug Nunn, 212
179 One recorded all data while the other four positioned PT at the Perlman Center at Cincinnati Children’s 213
180 subjects and called out weight values. Inter-rater reli- Hospital [20]. 214
181 ability was established during a practice session. Each Each subject was randomized to be placed 215
182 scale was set to “zero” and calibrated with a 5-pound in prone or supine positioning and then moved 216
4 G. Paleg et al. / Inclination, hip abduction, orientation, and tone affect weight-bearing in standing devices
f
228
roo
229
2020. 232
rP
234
tho
therefore we have used 10% change in weight- 239
3. Results 241
Au
Fifteen participants (13 who used standing devices 242
246
217 passively in randomly assigned order of inclination explained. This may result when the child “pushes” 265
218 (0◦ /15◦ /30◦ ) and abduction (0◦ /15◦ /30◦ ). This pro- on the body supports and down through their feet, as 266
219 cess was repeated for the second randomly selected there would need to be some force production to get a 267
220 stander for a total of 36 position combinations per reading over 100%. Readings over 100% were seen in 268
221 stander (72 total per child). Each participant session two similar studies [11, 12] and were not unexpected. 269
G. Paleg et al. / Inclination, hip abduction, orientation, and tone affect weight-bearing in standing devices 5
f
roo
rP
Fig. 4. Weight-Bearing in All Positions.
Table 1
Participant Characteristics
Subject weight GMFCS Age tone Dx Standers
tho
1. 74.8 IV 6 yrs 8 months hypotonia CP Zing 2 Standz 3
2. 78.8 n/a 8 yrs 1 months hypotonia SCI Zing 2 Standz 3
3. 71 IV 9 yrs 7 months hypertonia CP Zing 2 Standz 3
4. 73.2 n/a 8 yrs 9 months typical control Zing 2 Standz 3
5. 32.1 III 4 yrs 11 months hypertonia CP Standz 2 Zing 3
6. 28.1 V 6 yrs 7 months hypotonia
Au CP HLT Superstand 2 Standz 3
7. 31.4 IV 4 yrs 2 months hypertonia CMV Zing 2 HLT Superstand 1
8. 26.1 IV 3 yrs 4 months hypertonia CP HLT Superstand 1 Standz 3
9. 21.0 IV 7 yrs 1 months hypertonia CP Standz 3 Zing 2
10. 34.0 V 3 yrs 11 months hypertonia HIE Standz 3 HLT Superstand 1
11. 28.0 V 3 yrs 11 months hypotonia Agenesis corpus callosum HLT Superstand 1 Zing 2
12. 32.1 V 7 yrs 11 months hypertonia CP Standz 3 HLT Superstand 1
d
Table 2
Number of participants weight-bearing over 100% in each position
Prone Supine
rre
270 However, these data points may have skewed results when upright with 60◦ of abduction in supine posi- 280
Un
271 to show more weight-bearing than occurred. Since tion and the least (65%) with 30◦ of inclination, feet 281
272 these data points occurred in all groups, it should not together in prone position (see Fig. 6). The children 282
273 have affected the overall findings. with typical tone demonstrated the most weight- 283
274 The high tone group demonstrated the most bearing (108%) when upright with 60◦ of abduction 284
275 weight-bearing (91%) when upright with feet to- in supine position and the least (69%) with 30◦ incli- 285
276 gether in prone position, and the least weight-bearing nation, feet together in prone position (see Fig. 7). 286
277 (67%) with 30◦ of inclination and 60◦ of abduction Supine positioning resulted in more weight bearing 287
278 in supine position (see Fig. 5). The low tone group in all positions for the children with low and nor- 288
279 showed the highest amount of weight-bearing (93%) mal tone. Children with high tone bore slightly more 289
6 G. Paleg et al. / Inclination, hip abduction, orientation, and tone affect weight-bearing in standing devices
f
roo
rP
Fig. 5. Weight-bearing for Participants with High Tone (%±standard deviation (range)).
tho
d Au
cte
Fig. 6. Weight-bearing for Participants with Low Tone (%±standard deviation (range)).
rre
co
Un
Fig. 7. Weight-bearing for Participants with Normal Tone (%±standard deviation (range)).
G. Paleg et al. / Inclination, hip abduction, orientation, and tone affect weight-bearing in standing devices 7
f
roo
rP
tho
Fig. 8. Weight-bearing in Supine vs. Prone Orientation.
d Au
cte
rre
co
Un
290 weight in all positions when in prone position (see with normal and low tone behaved similarly in 295
291 Figs. 8 and 9). comparison to those with high tone. As predicted, 296
292 4. Discussion decreased weight bearing for children with high tone, 298
293 Tone affected weight-bearing through the feet in abduction. Conversely, maximal abduction resulted 300
294 varied positions in this exploratory study. Children in increased weight-bearing for the group with low 301
8 G. Paleg et al. / Inclination, hip abduction, orientation, and tone affect weight-bearing in standing devices
302 and normal tone. These results suggest children with 5. Conclusion 351
305 for Clinical recommendations hip abduction, and orientation affected weight- 353
306 These results demonstrate that one should avoid bearing through the feet for children in standing 354
307 extremes in positioning (inclination and abduction) devices. Clinicians should be mindful of these fac- 355
308 for children with high tone if weight-bearing is tors when developing a standing program for children 356
309 the main goal of the standing program. For chil- and understand the cost-benefit relationship between 357
f
310 dren with high adductor tone, it is possible that postural alignment and weight-bearing in standing 358
roo
311 increased abduction resulted in increased pressure devices. If the goal is to increase weight-bearing in 359
312 or weight-bearing through the pommel or knee supported standing, it is recommended that children 360
313 supports, although this was not measured. It is with hypertonia be placed in prone upright position- 361
314 unknown if the reduction in weight-bearing through ing with their hips in neutral (feet together) or slight 362
315 the feet that was measured is also accompanied abduction, and children with hypotonia should be 363
rP
316 by a similar reduction of forces through the aceta- placed in upright supine positioning with a moderate 364
318 For children with low tone, it may be even more research include scheduling more time for each ses- 366
319 important to measure weight-bearing to ensure it is sion, recording of stander use (e.g., journal of usage) 367
tho
320 maximized. They have been described as “leaning to record whether weight-bearing changes over time, 368
321 into supports” [17] which may account for the differ- including a larger number of subjects, using only one 369
322 ences. If the children with low tone “hung” or leaned brand of stander, and including whole body pressure 370
323 on the lateral supports and surface of the stander, mapping to document all weight-bearing surfaces. 371
329 a solid posterior support similar to a tilt table. The this study from Prime Engineering. Dr. Paleg is also 374
331 port surfaces with open space between the supports. She claims no bias in this study. 376
cte
332 Sit-to-stand devices were not included because a No other authors received funding or have any con- 377
333 previous study concluded that these devices demon- flicts. 378
334 strated the lowest amount of weight bearing [5] All products were provided free of charge from the 379
335 and these models currently do not allow for abduc- manufacturers. 380
336 tion.
rre
References 381
337 4.1. Limitations
[1] Uddenfeldt Wort U, Nordmark E, Wagner P, Düppe H, West- 382
co
338 This was a small exploratory descriptive study bom L. Fractures in children with cerebral palsy: a total 383
339 that measured weight-bearing through the footplates population study. Dev Med Child Neurol. 2013;55(9):821-6. 384
340 only. The convenience sample limits generalizability doi: 10.1111/dmcn.12178. 385
[2] Rodby-Bousquet E, Czuba T, Hägglund G, Westbom L. 386
341 and the small heterogenous sample limits conclu- Postural asymmetries in young adults with cerebral palsy.
Un
387
342 sions. It is possible that different results may have Dev Med Child Neurol. 2013;55(11):1009-15. doi: 10.1111/ 388
343 been found with larger samples. Forces through the dmcn.12199. 389
344 head of the femur into the acetabulum may differ at [3] Taylor K. Factors affecting prescription and implementation 390
of standing-frame programs by school-based physical thera- 391
345 the various angles and inclinations in comparison to pists for children with impaired mobility. Pediatr Phys Ther. 392
346 weight-bearing through the feet, and weight may have 2013;21(3):282-8. doi: 10/1097/PEP.0b013e3181b175cd. 393
347 transferred to other stander surfaces (such as the pom- [4] Kecsemethy HH, Herman D, May R, Paul K, Bachrach SJ, 394
Henderson RC. Quantifying weight bearing while in passive 395
348 mel or knee blocks). However, neither was measured
standing devices and a comparison of standing devices. Dev 396
349 in this pilot study and these factors may influence Med Child Neuro. 2008;50(7):520-3. doi: 10.1111/j.1469- 397
350 results. 8749.2008.03021.x. 398
G. Paleg et al. / Inclination, hip abduction, orientation, and tone affect weight-bearing in standing devices 9
399 [5] Herman D, May R, Vogel L, Johnson J, Henderson RC. [14] Gmelig Meyling C, Ketelaar M, Kuijper M-A, Voorman J, 437
400 Quantifying weight-bearing by children with cerebral palsy Buizer AI. Effects of Postural Management on Hip Migra- 438
401 while in passive standing devices. Pediatr Phys Ther. 2007; tion in Children With Cerebral Palsy: A Systematic Review. 439
402 19(4):283-7. doi: 10/1097/PEP.0b013e318156cc4d. Pediatr Phys Ther. 2018;30(2):82-91. doi: 10.1097/PEP. 440
403 [6] Caulton JM, Ward KA, Alsop CW, Dunn G, Adams JE, 0000000000000488. 441
404 Mughal MZ. A randomised controlled trial of standing pro- [15] Miller SD, Juricic M, Hesketh K, Mclean L, Magnuson S, 442
405 gramme on bone mineral density in non-ambulant children Gasior S, et al. Prevention of hip displacement in children 443
406 with cerebral palsy. Arch Dis Child. 2004;89(2):131-5. doi: with cerebral palsy: a systematic review. Dev Med Child 444
407 10.1136/adc.2002.009316. Neurol. 2018;59(11):1130-8. doi: 10.1111/dmcn.13480. 445
408 [7] Paleg G, Smith BS, Glickman LB. Systematic review and [16] Paleg G, Livingstone R. Systematic Review and Evidence- 446
f
409 evidence-based clinical recommendations for dosing of Based Clinical Recommendations for Dosage of Supported 447
roo
410 pediatric supported-standing programs. Pediatr Phys Ther. Standing Programs for Adults with Neuromotor Conditions. 448
411 2013;25(3):232-47. doi: 10.1097/PEP.0b013e318299d5e7. BMC Musculoskelet Disord. 2015;16(1):358. doi: 10.1186/ 449
412 [8] Pountney TE, Mandy A, Green E, Gard PR. Hip subluxation s12891-015-0813-x. 450
413 and dislocation in cerebral palsy—a prospective study on the [17] Bar-On L, Molenaers G, Aertbeliën E, Van Campenhout A, 451
414 effectiveness of postural management programmes. Physio- Feys, H, Nuttin B, et al. Spasticity and Its Contribution to 452
415 ther Res Int. 2009;14(2):116-27. doi: 10.1002/pri.434. Hypertonia in Cerebral Palsy. Biomed Res Int. 2015;2015: 453
rP
416 [9] Pountney T, Mandy A, Green E, Gard P. Management 317047. doi: 10.1155/2015/317047 454
417 of hip dislocation with postural management. Child Care [18] Martin KS, Westcott S, Wrotniak BH. Diagnosis dialog 455
418 Health Dev. 2002;28(2):179-85. doi: 10.1046.j.1365-2214. for pediatric physical therapists: hypotonia, developmental 456
419 2002.00254.x. coordination disorder, and pediatric obesity as examples. 457
420 [10] Macias-Merlo L, Bagur-Calafat C, Girabent-Farrés M, Stu- Pediatr Phys Ther. 2013;25(4):431-43. doi: 10.1097/PEP. 458
421 berg WA. Standing Programs to Promote Hip Flexibility in 0b013e31829ec53f. 459
tho
422 Children With Spastic Diplegic Cerebral Palsy. Pediatr Phys [19] Crellin DJ, Harrison D, Santamaria N, Huque H, Babl FE. 460
423 Ther. 2015;27(3):243-9. doi: 10.1097/PEP.0000000000 The Psychometric Properties of the FLACC Scale Used 461
424 000150. to Assess Procedural Pain. J Pain. 2018;19(8):862-72. doi: 462
425 [11] Macias-Merlo L, Bagur-Calafat C, Girabent-Farrés M, Stu- 10.1016/j.jpain.2018.02.013. 463
426 berg WA. Effects of the standing program with hip abduction [20] Nunn D, Clark K. Pediatric Stander Evaluation & Appli- 464
427 on hip acetabular development in children with spastic diple- cations for Fun & FUNction! IC20. International Seating 465
Au
428 gia cerebral palsy. Disabil Rehabil. 2016;38(11):1075-81. Symposium Vancouver, BC, Canada 2019. 466
429 doi: 10.3109/09638288.2015.1100221. [21] Hustedt JW, Blizzard DW, Baumgaertner MR, Leslie 467
430 [12] Martinsson C, Himmelmann K. Effect of weight-bearing in MP, Grauer JN. Current advances in training orthopaedic 468
431 abduction and extension on hip stability in children with patients to comply with partial weight-bearing instructions. 469
432 cerebral palsy. Pediatr Phys Ther. 2011;23(2):150-7. doi: Yale J Biol Med. 2012;85(1):119-25. 470
433 10.1097/PEP.0b013e318218efc3.
434 [13] Miller F. Natural History and Surveillance of Hip Dysplasia
d