You are on page 1of 7

MDR500A Mobile X-ray machine

Installation manual

Edition:A2

Document No:

Release date:

Please read the user manual and pay attention to all safety precautions before operation;

Please be sure to keep this user manual for future reference;

Please make full use of the content on the basis of full understanding。
Document No -
iRay 奕瑞 Edition A2
Revision date 2020.1.30
MDR500 Installation manual Page number 2 / 7

Version record
Version Organization Review Approval Date Explain

0 宋钊林 彭冲 First edition

Precautions for this SOP


1. The tool used in this assembly SOP is metric tool
Document No -
iRay 奕瑞 Edition A2
Revision date 2020.1.30
MDR500 Installation manual Page number 3 / 7

Step 1 :overall introduction

NO. Name Photo

1 Air spring lock

2 Computer support lock

3 Cantilever buckle

4 Air spring

5 Cantilever lock

6 Front wheel support lock

7 Front wheel support

8 Cantilever
Document No -
iRay 奕瑞 Edition A2
Revision date 2020.1.30
MDR500 Installation manual Page number 4 / 7

9 Protable fixing screw

10 Cantilever handrail

11 Protable lock

12 Protable x-ray

13 Front wheel support lock

14 Push the armrest

15 PC support

16 Storage box

17 Foot brake

Step 2 :Unpack the wooden box Tool list


Document No -
iRay 奕瑞 Edition A2
Revision date 2020.1.30
MDR500 Installation manual Page number 5 / 7

1. Open end wrench

Operation steps:
1. Remove the screws on the top cover of the packing case and remove the top cover of the packing
case.
2. Remove block 1 and block 2.
3. 2-3 people lift the machine out of the packing box.

Safety
Step 3 : Front wheel bracket landing
considerations
1. Unlock 12. Press down the
front wheel support lock 7.
When the front wheel
supports, the machine will
dump forward about 10 °,
please pull 14. Push the
armrest.

Operation steps:
1. Unlock 6. Front wheel support lock.
2. Unlock with one hand (pull towards the frame) 13. Front wheel support lock, press down with
one hand 7. Front wheel support.

Step 4 : Lock brake Considerations


Document No -
iRay 奕瑞 Edition A2
Revision date 2020.1.30
MDR500 Installation manual Page number 6 / 7

1. Remember to slightly push


forward and backward when
locking the foot brake to
lock the foot brake. If the
brute force is locked, the
foot brake will be damaged

Operation steps:
1. Step on 17. Push the foot brake forward and backward slightly to step on the brake.
2. Lock 17. Foot brake

Step 5 :Cantilever pull up 45 ° Safety


considerations

1. In the case of no combined


head, when the cantilever is
pulled out, press the cantilever
with 150n force, otherwise
personal safety accident may be
caused.

Operation steps:
1. Unlock 3. Cantilever buckle.
2. Unlock (rotate counterclockwise) 1. Air spring lock.
3. Press both hands firmly on 10. Cantilever handrail, unlock (press down) 5. Cantilever lock,
pull out the cantilever to 45 ° angle.
4. Lock (clockwise rotation) 1. Air spring lock.
Document No -
iRay 奕瑞 Edition A2
Revision date 2020.1.30
MDR500 Installation manual Page number 7 / 7

Step 5 :Installation of combined head Safety


considerations

1. Remember to lock the


combined head after installation
11. Fixed lock of combined head.
2. When the combined head is
taken out, remember to lock
(rotate clockwise) 1. Air spring
lock. Otherwise, it may cause
personal safety accidents.

Operation steps:
1. Installation 12. Protable x-ray.
2. Lock 11. Protable lock.
3. Use the fixing screws on both sides of 9. Combined head pan to fix the head pan.
4. Now the mechanical part of the combined head has been installed.

Step 7 : connection Safety


considerations
1. After the wiring is
completed, fix the wire to
avoid the interference of the
harness when turning the
combined head again.

Operation steps:
1. Connect the hand brake switch to exposure.
2. Connect the power plug to power.
3. Connect the communication line to PC.
4. The wiring is now completed