Professional Documents
Culture Documents
net/publication/285753008
CITATIONS READS
2 296
5 authors, including:
Some of the authors of this publication are also working on these related projects:
Development of a framework for a national pediatric pharmacotherapy collaborative practice system: Design requirements, proposed improvements, and
anticipated challenges View project
All content following this page was uploaded by Jeffrey Cies on 07 September 2016.
F
urosemide is a loop diuretic that
interferes with the chloride- Purpose. The results of a study to deter- Results. LC-MS/MS analysis showed that
binding cotransport system that mine the stability of solutions of furose- chlorothiazide typically eluted at 2.6 min-
mide and chlorothiazide over 96 hours are utes and furosemide at 4.8 minutes. Each
inhibits the reabsorption of sodium
reported. compound was degraded by exposure to
and chloride in the ascending loop Methods. Chlorothiazide and furo- strong ultraviolet light in a time-dependent
of Henley and the distal renal tu- semide were diluted in 5% dextrose manner. Both unmixed and mixed solu-
bule, causing increased excretion of USP to final concentrations of 10 and tions retained over 90% of the original
sodium and water. Chlorothiazide is 1 mg/mL, respectively, and combined. concentrations of chlorothiazide and furo-
a thiazide-type diuretic that inhibits In addition, sample solutions of chlo- semide for up to 96 hours. Furosemide and
sodium reabsorption in the distal rothiazide in dextrose, furosemide in chlorothiazide are commonly used con-
dextrose, and dextrose alone were comitantly to maximize diuresis in pediatric
tubule, also leading to increased
prepared for control purposes. The re- patients; the study findings suggest that
excretion of sodium and water. The sulting solutions were analyzed im- solutions of furosemide and chlorothiazide
combination of furosemide and a mediately after preparation and 24, 48, can be combined in the same syringe with-
thiazide diuretic (i.e., chlorothiazide 72, and 96 hours later using a liquid out loss of stability for up to 96 hours.
or hydrochlorothiazide) has been chromatography–tandem mass spectros- Conclusion. Solutions of chlorothiazide
shown to increase natriuresis to a copy (LC-MS/MS) system with an elec- (10 mg/mL) and furosemide (1 mg/mL)
greater extent than furosemide alone trospray ionization source. Mixtures and stored either separately or together in
samples were diluted 10,000-fold prior to polypropylene syringes remained stable
in animal models.1 Furthermore, the
LC-MS/MS analysis so that concentrations for up to 96 hours at room temperature and
combination of a loop diuretic and of both drugs would be within the assay’s protected from light.
a thiazide diuretic has been shown to linear range of detection. Am J Health-Syst Pharm. 2015; 72:2182-8
increase natriuresis more than furo-
semide alone, in addition to increas-
ing the amount of weight loss, in
adult patients with heart failure and
fluid-overloaded patients.2-5 Combinations of different loop di- thiazide is to remove fluid, combin-
There are data on adults suggest- uretics are also used in children to re- ing the two drugs in the same syringe
ing that continuous infusion of loop duce tolerance of or resistance to in- for continuous infusion could offer
diuretics produces greater diuresis dividual compounds.8 Since the goal an advantage by limiting the total
than does intermittent dosing. 6,7 of using either furosemide or chloro- amount of fluid administered to the
Jeffrey J. Cies, Pharm.D., M.P.H., BCPS (AQ-ID), is Pharmacist, St. Address correspondence to Dr. Cies (jeffrey.cies@gmail.com).
Christopher’s Hospital for Children, Philadelphia, PA, and Pharmacy Presented in part as an abstract (number 737) at Cardiology 2014,
Clinical Coordinator, Critical Care, and Infectious Diseases Clinical 17th Annual Update on Pediatric and Congenital Cardiovascular
Pharmacist, Alfred I. duPont Hospital for Children, Wilmington, Disease, Orlando, FL, February 20, 2014.
DE. Wayne S. Moore II, Pharm.D., is Pharmacist, Alfred I. Supported in part by Nemours Biomedical Research and National
duPont Hospital for Children. Arun Chopra, M.D., is Clinician, Institutes of Health grant P20GM103464.
NYU Langone Medical Center, and Chief, Section of Critical Care The authors have declared no potential conflicts of interest.
Medicine, NYU School of Medicine, New York, NY. Guizhen Lu, B.S.,
is Research Assistant; and Robert W. Mason, Ph.D., is Head of Clini- Copyright © 2015, American Society of Health-System Pharma-
cal Biochemistry, Nemours Biomedical Research, Alfred I. duPont cists, Inc. All rights reserved. 1079-2082/15/1202-2182.
Hospital for Children. DOI 10.2146/ajhp150023
patient during therapy. Thus, it has of chlorothiazide and furosemide 100% A. The solvent flow rate was 0.4
become a practice at some institu- were prepared in triplicate each day mL/min, and the total run time per
tions to combine furosemide and using the same pharmaceutical-grade sample was 7 minutes, including col-
chlorothiazide in the same syringe compounds to account for any varia- umn equilibration. Compounds were
even though stability data are cur- tion in MS signals. Experimental val- detected using LC-MS/MS equip-
rently lacking for this combination. ues for MS signals were normalized mentj with an electrospray ionization
The purpose of the study described to a prespecified value of 100 for the source. Mass spectrometry conditions
here was to determine whether in- fresh samples. Daily variations in MS were as follows: gas temperature, 350
jectable formulations of furosemide signals were less than 5% during the °C (flow rate, 10 L/min); sheath gas
and chlorothiazide are stable either study period. temperature, 400 °C (flow rate, 12 L/
alone or when mixed together. The data reported here are ex- min); nebulizer pressure, 45 psi; and
pressed as means with S.D. values. Sta- capillary energy, 3500 volts (V) with
Methods bility was defined as not less than 90% detector in negative ion mode. The
Sample preparation. Chlorothia- and not more than 110% of furose- chlorothiazide primary ion had a peak
zidea 500 mg was reconstituted with mide and chlorothiazide remaining. mass:charge ratio (m/e) of 294, and
18 mL of bacteriostatic water for Reference standards of furosemidee the fragment ion had a peak m/e of
injection,b resulting in a final concen- and chlorothiazidef were dissolved in 214.1 with the fragmentor set at 150
tration of 28 mg/mL. Next, 3.57 mL of equimolar sodium hydroxide solution V and a collision energy of 12 electron
the chlorothiazide solution was added to attain final concentrations of 1 and volts (eV). The furosemide primary
to 1 mL of furosemide injectionc 10 mg/mL, respectively. The freshly ion had a peak m/e of 329, and the
(20 mg/2 mL), and 5% dextrose was prepared standard solutions were fragment ion had a peak m/e of 285.1
added to attain a final volume of 10 compared directly with freshly pre- with the fragmentor set at 110 V and
mL. All compounds were used prior pared pharmaceutical products. a collision energy of 12 eV. A major
to expiration. The resulting final con- To force degradation, pharma- product of UV degradation of chlo-
centrations were 1 mg/mL of furose- ceutical-grade samples were exposed rothiazide was detected (primary ion
mide and 10 mg/mL of chlorothiazide to ultraviolet (UV) light (254 nm) peak m/e, 259.9; fragment ion peak
in one sample. Three samples each from a germicidal UV lampg for up m/e, 180) with the fragmentor set at
of solutions of furosemide and chlo- to 16 hours. Samples were aliquoted 135 V and a collision energy of 25 eV.
rothiazide in 5% dextrose (1 and 10 into quartz cuvets that were placed A major product of UV degradation
mg/mL, respectively) were prepared approximately 2 cm away from the of furosemide was detected (primary
separately for analysis. One milliliter light source. At timed intervals, ali- ion peak m/e, 311.1; fragment ion
of the 1-mg/mL furosemide solution quots were removed and diluted for peak m/e, 202.2) with the fragmentor
was mixed with 9 mL of 5% dextrose, LC-MS/MS analysis. Assays were per- set at 120 V and a collision energy of
and the 10-mg/mL chlorothiazide formed in triplicate and with three 20 eV. Samples were diluted 10,000-
solution was prepared as described independent experiments. fold prior to LC-MS/MS unless other-
above for the two-solution mixture LC-MS/MS analysis. Prior to the wise noted; 10 mL of each sample was
except for the substitution of 1 mL stability study, LC-MS/MS methods added to 990 mL of water, the solution
of 5% dextrose for 1 mL of furose- to separate, detect, and measure was mixed, and then 10 mL of the
mide. Three separate preparations of furosemide, chlorothiazide, and deg- mixture was added to an additional
each formulation were prepared. The radation products were developed. 990 mL of water and mixed prior to
mixtures were visually examined for Chlorothiazide, furosemide, and deg- analysis by LC-MS/MS.
color change against a white back- radation products were separated by
ground and for haze, turbidity, gas chromatography on a C18 columnh Results
bubbles, and precipitation against a at 50 °C and eluted with a gradient During preparation of the samples
black background. These evaluations from 0.1% formic acid (solution and after storage, visual examination
were done immediately and after the A) to 100% methanol in 0.1% for- did not detect any color changes
samples were stored at 25 °C in the mic acid (solution B) using a high- or evidence of haze, turbidity, gas
dark for up to 96 hours to simulate performance liquid chromatography bubbles, or precipitation, indicating
storage under normal clinical use. All system.i The gradient was set up as that both drugs remained in solution
samples were stored in 15-mL poly- follows: start (0 minutes), 100% A; during the course of the experiment.
propylene tubes.d 1 minute, 100% A; 3 minutes, linear Chlorothiazide typically eluted
Prior to analysis via liquid gradient to 100% B; 3.5 minutes, hold from the chromatogram at 2.6 min-
chromatography–tandem mass spec- at 100% B; 3.6 minutes, linear gradi- utes and furosemide at 4.8 minutes
troscopy (LC-MS/MS), fresh samples ent to 100% A; and 6 minutes, hold at (Figure 1). An initial dilution range
Figure 1. Chromatographic separation of chlorothiazide and furosemide. Panel A shows superimposed typical chromatographic signals
of chlorothiazide and furosemide, as detected by liquid chromatography–tandem mass spectroscopy (LC-MS/MS). Original samples
of a mixture of the two compounds were diluted 1:10,000, and 5 mL of the diluted mixture was applied to the LC column. The peak
mass:charge ratio (m/e) values for chlorothiazide and furosemide are shown at about 2.8 and about 4.8 minutes, respectively. Panel
B shows a typical chromatographic signals of chlorothiazide and a fragment eluting at 1.4 minutes after exposure of the parent com-
pound to ultraviolet (UV) light. Panel C shows typical chromatographic signals of furosemide and a fragment eluting at 4.5 minutes after
exposure of the parent compound to UV light. In each chromatogram, the y-axis denotes ion counts for each compound normalized to
the highest peak (set at 100).
A
1.1
Normalized Ion Count × 102
1.0
0.9
0.8
0.7
0.6
0.5
0.4
0.3
0.2
0.1
0
–0.1
0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6 1.8 2.0 2.2 2.4 2.6 2.8 3.0 3.2 3.4 3.6 3.8 4.0 4.2 4.4 4.6 4.8 5.0 5.2 5.4 5.6 5.8
Time (min)
B
1.1
Normalized Ion Count × 102
1.0
0.9
0.8
0.7
0.6
0.5
0.4
0.3
0.2
0.1
0
–0.1
0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6 1.8 2.0 2.2 2.4 2.6 2.8 3.0 3.2 3.4 3.6 3.8 4.0 4.2 4.4 4.6 4.8 5.0 5.2 5.4 5.6 5.8
Time (min)
C
1.1
Normalized Ion Count × 102
1.0
0.9
0.8
0.7
0.6
0.5
0.4
0.3
0.2
0.1
0
–0.1
0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6 1.8 2.0 2.2 2.4 2.6 2.8 3.0 3.2 3.4 3.6 3.8 4.0 4.2 4.4 4.6 4.8 5.0 5.2 5.4 5.6 5.8
Time (min)
was prepared to determine the linear LC-MS/MS analysis (to attain 1,000- those of the USP standards. Analysis
range of the LC-MS/MS assay. The ng/mL chlorothiazide and 100-ng/ of freshly prepared pharmaceutical
assay for chlorothiazide was linear mL furosemide samples) ensured formulations of chlorothiazide and
from 8 to 1,250 ng/mL (Figure 2); the that both compounds were measured furosemide indicated peak area un-
assay for furosemide was linear from in the linear range of the assay. The der the curve values that were within
8 to 1,000 ng/mL (Figure 2). Above LC-MS/MS system used in this study 95% of values obtained using the
1,250 ng/mL, the signal intensities gave consistent results that did not USP standards.
of chlorothiazide samples were very vary by more than 1% for repeat To force the loss of compound
high and no longer linear, so these injections of the same sample. The and validate the assay method, both
points were not used to generate the two commercially sourced drugs compounds were exposed to UV light.
calibration curve. Thus, the 10,000- were found to have chromatographic Mass spectrometry scans of exposed
fold dilution of samples prior to and molecular properties identical to chlorothiazide samples revealed the
Figure 2. Chlorothiazide and furosemide calibration curves. Serial twofold dilutions of chlorothiazide were prepared in order to gener-
ate 10 levels of concentration, and 5 mL of each sample was analyzed via liquid chromatography–tandem mass spectroscopy (LC-MS/
MS); the results are shown in panel A. The three highest data points for chlorothiazide (open circles) were beyond the linear phase and
consequently were not used in the calibration curve. Serial twofold dilutions of furosemide were prepared in order to generate 10 con-
centration levels, with 5 mL of each sample applied to the LC column and quantified by MS/MS (panel B). Three separate dilutions were
prepared in order to obtain 30 separate LC-MS/MS data points to show consistency of measurement.
1.0
Mean ± S.D. Ion Count × 106
y = 214.083854x + 5628.923089
0.9
R 2 = 0.99565547
0.8
0.7
0.6
0.5
0.4
0.3
0.2
0.1
0
0 1,000 2,000 3,000 4,000 5,000 6,000 7,000 8,000 9,000 10,000
Concentration (ng/mL)
1.4 R 2 = 0.99489463
1.2
1.0
0.8
0.6
0.4
0.2
0
0 100 200 300 400 500 600 700 800 900 1,000
Concentration (ng/mL)
Figure 3. Time-dependent ultraviolet (UV) degradation of chlorothiazide and furosemide. Chlorothiazide and furosemide were exposed
to UV light, and samples were analyzed at timed intervals by liquid chromatography–tandem mass spectroscopy to quantify the remain-
ing concentrations of parent compounds and new fragments. Peak area under the curve values for chlorothiazide are shown after UV
exposure for up to 960 minutes (squares, panel A). Loss of parent compound was accompanied by the appearance of a fragment (dia-
monds, panel A). Area under the curve values corresponding to furosemide are shown after UV exposure for up to 180 minutes (squares,
panel B); the diamonds in panel B denote the appearance of a fragment. The UV degradation product was diluted 1:100 to enable detec-
tion of the fragment at early time points.
A
450,000
400,000
350,000 R 2 = 0.9895
300,000
Ion Count
250,000
200,000
150,000
100,000
R 2 = 0.9861
50,000
0
0 200 400 600 800 1,000
Time (min)
B
30,000 350,000
300,000
25,000
250,000
Ion Count (Furosemide)
20,000
200,000
15,000
R 2 = 0.9238
150,000
10,000
100,000
5,000
50,000
0 0
0 50 100 150
Time (min)
appearance of a new compound with results clearly showed that the assay those for the pure USP standards,
a peak m/e of 259.9; an ion fragment used to measure the area under the indicating that the excipients did
(peak m/e, 180) was generated in the curve values corresponding to chlor- not affect the measurement of drug
collision cell of the mass spectrometer othiazide and furosemide accurately concentration. For chlorothiazide,
(as described in the methods) to quan- indicated levels of each compound UV-dependent loss of the parent
tify the appearance of this UV degra- and could be used to indicate sample compound was directly proportional
dation product. The product eluted at stability. to the appearance of a degradation
1.4 minutes and was separated from Using the developed LC-MS/MS product with a peak m/e of 259.9,
chlorothiazide (Figure 1). Similarly, method, the test solutions prepared which was consistent with photode-
the major degradation product gen- from commercially sourced chlor- halogenation; similar (but slower)
erated by exposure of furosemide to othiazide and furosemide, when degradation has been reported after
UV light was detected at a peak m/e of stored unmixed or mixed at 25 °C chlorothiazide exposure to UV light
311.1, and a fragment ion (peak m/e, and protected from light, were found (313 nm) in methanol solutions.9
202.2) eluted at 4.5 minutes, just be- to retain at least 90% of the initial Furosemide was more sensitive to
fore the elution of furosemide (Figure quantities of active compounds for UV exposure, with more than 50% of
1). The peak corresponding to chlo- the 96 hours of the study (Table 1). the compound degraded after three
rothiazide decayed in a linear fashion Levels of chlorothiazide were notably hours of treatment. Among a range
over 4 hours, and this decay continued lower in both formulations at 48 and of degradation products, a major
such that only 21% of the parent com- 96 hours after preparation but did fragment with a peak m/e of 311.1
pound remained after 16 hours of UV not change appreciably between 72 appeared at the first analytical time
exposure (Figure 3). The mass of the and 96 hours, indicating that mixing point, peaking after two hours of UV
degradation product was consistent the two drugs did not affect stability exposure. A recent study detected at
with the loss of a chloride ion from and that there was an initial small least 12 novel UV-absorbing peaks
the parent compound. The degrada- reduction in levels of chlorothiazide. during the degradation of furose-
tion product increased linearly over The degradation products produced mide, although their molecular char-
the first 4 hours of UV exposure, an by forced degradation were not de- acteristics were not determined.10
indication that this product was stable tected in the 96-hour samples. While our studies show that chlo-
relative to the parent compound. Fu- rothiazide and furosemide, as for-
rosemide was less stable during UV Discussion mulated in our hospital, are stable
exposure, with approximately 25% Both chlorothiazide and furose- for up to 96 hours at 25 °C and
of the original compound remaining mide can be degraded by exposure protected from light, these same
after 3 hours of exposure to UV light to strong UV light (254 nm). Loss drugs may not be stable in different
(Figure 3). A product with a reduced of parent compound was time de- formulations. For example, a related
m/e (311.1) appeared as the original pendent and corresponded to the compound, hydrochlorothiazide, is
compound was lost, but this product appearance of degradation products. stable in acid or in suspension but
was also unstable during UV expo- The commercial drugs contained degrades at a pH above 6.5.11 The USP
sure. After 16 hours of UV exposure, excipients, including mannitol and standard of chlorothiazide is the free
neither the parent compound nor the dextrose. MS/MS signals from the acid; when dissolved in equimolar
fragment could be detected. These commercial drugs were identical to sodium hydroxide for our study, it
Table 1.
Stability of Unmixed and Mixed Furosemide and Chlorothiazide Solutions Stored in Syringes