Automatic pipettes are precise laboratory tools used to transfer small volumes of liquids. They work by using air displacement and pressure to draw up and expel liquid through a disposable tip. Proper technique is required for accuracy and precision, including using fresh tips, proper immersion depth, and allowing time for liquid transfer. This document outlines the parts and operation of automatic pipettes, provides guidelines for use, and describes the step-by-step process for setting the desired volume and sample transfer using a Pipetman as an example. Calibration is recommended periodically to ensure the pipette is delivering the correct volume.
Automatic pipettes are precise laboratory tools used to transfer small volumes of liquids. They work by using air displacement and pressure to draw up and expel liquid through a disposable tip. Proper technique is required for accuracy and precision, including using fresh tips, proper immersion depth, and allowing time for liquid transfer. This document outlines the parts and operation of automatic pipettes, provides guidelines for use, and describes the step-by-step process for setting the desired volume and sample transfer using a Pipetman as an example. Calibration is recommended periodically to ensure the pipette is delivering the correct volume.
Automatic pipettes are precise laboratory tools used to transfer small volumes of liquids. They work by using air displacement and pressure to draw up and expel liquid through a disposable tip. Proper technique is required for accuracy and precision, including using fresh tips, proper immersion depth, and allowing time for liquid transfer. This document outlines the parts and operation of automatic pipettes, provides guidelines for use, and describes the step-by-step process for setting the desired volume and sample transfer using a Pipetman as an example. Calibration is recommended periodically to ensure the pipette is delivering the correct volume.
USING AN AUTOMATIC PIPETTE adjustment knob, a volume
indicator, and a plastic shaft.
Many laboratory procedures, - are available in six different sizes. especially biological and Five of them are depicted here. biochemical ones, require the precise transfer of small quantities of liquids. Glass pipettes, which are commonly used in chemistry labs, are simply not accurate when working with volumes that are less than 1 milliliter. The best method of transferring - The number on the plunger small volumes of liquids involves button indicates the maximum the use of an automatic pipette. volume in microliters that each Pipetman is capable of Automatic pipette transferring. - consists of a piston that - For example, the P20 has a displaces air to create a vacuum possible volume range of 0 to 20 when its plunger is pressed. microliters. - The tip of the automatic pipette is Reminder: Transferring volumes then placed into the sample vial at the low end of the range is not and the plunger is slowly recommended because the released. standard deviation is too high to - This action draws liquid into the ensure precision. tip. - Thus the minimum amount of - Finally, this liquid is expelled into liquid that can be transferred in the desired container by slowly microliters is 0.1 for the P2, 0.5 depressing the plunger a second for the P10, two for the P20, 10 time. for the P100, 50 for the P200, and 100 microliters for the
- As shown, a variety of automatic
pipettes exist. This video will P1000. demonstrate automatic pipetting Now that we've covered what an technique using the Pipetman. automatic pipette is and how it works, we'll take a detailed look into how the Pipetman is used in the laboratory.
STEPS IN USING PIPETMAN:
- Let's say, for example, your procedure requires you to transfer 153.5 microliters of Using good technique will ensure sample. accuracy, which is defined as 1. Selection of the Pipetman with the closeness of the dispensed an appropriate volume range. volume to the set volume, and - In this example, a couple of precision, which is defined as choices exist. The P200 has a the reproducibility of individual recommended volume range of measurements between your 50 to 200 microliters while the samples. P1000 has a recommended range of 100 to 1,000 microliters. Pipetman Reminder: The lower end of a - consists of a plunger button, a tip volume range always has a ejector button, a volume higher standard deviation. Therefore, the P200 is the better - Always use a new sterile tip for choice. each sample and never let the tip 2. Selection of the proper touch anything except sample. disposable plastic tip. a. To draw a sample into the - Three tip sizes plastic tip, depress the o Large blue tips - which fit plunger until resistance is felt. the P1000 b. Hold the Pipetman vertically o medium yellow tips - which and immerse it into the fit the P200, P100 sample. o small white tips - which fit c. After the tip is immersed, the P10 and P2. slowly release the plunger to - Because we're using the P200, draw liquid into the tip. Wait a we're going to need the yellow few seconds to ensure that tips. the full volume of sample is Reminder: Using the proper tip is drawn into the tip. essential for preventing leaks and promoting accurate delivery of Reminder: Wait one second for liquids. the P2 to P200 Pipetman and 2 3. Setting the desired transfer to 3 seconds for the P1000. volume. Reminder: The larger the volume - For the smaller Pipetman, P2 to or the more viscous the sample, P200, the volume set on the the longer you should wait. indicator translates exactly into the volume that will be delivered. d. Remove the Pipetman from - This P200 is set to deliver a the sample and carefully wipe volume of 125.9 microliters. excess liquids from the sides - For the P1000, the volume set on of the tip. the indicator must be multiplied e. To dispense the liquid in the by 10 to determine the actual Pipetman, touch its tip to the volume that will be transferred. side wall of a new container - This P1000 is set to deliver a and slowly depress the volume of 1,259 microliters. plunger to the first stop. - To correctly adjust the volume f. Wait a few seconds to ensure setting, turn the volume complete sample flow, then adjustment knob to one third of a press the plunger to the revolution past the desired second step to expel any volume. Then slowly move the remaining liquid. dial back down to the desired g. With the plunger fully setting. depressed, carefully remove Reminder: Always approach the the tip from the container by desired volume by dialing sliding it up the container wall. downwards to prevent The importance of sliding the mechanical backlash and to table along the wall increases ensure precise volume transfer. as the volumes become - When adjusting the volume of a smaller. Pipetman, take care not to rotate h. Finally, allow the plunger to the dial past the maximum. You return to the up position. cannot transfer volumes above - If you are done working with a the range of the Pipetman. particular sample, press the tip 4. Transfering your sample. ejector button to release the tip - You will need your appropriately into the proper waste receptacle. sized Pipetman set to the desired volume, the correctly sized tips, a GUIDELINES TO ENSURE rack, the new container, a chem ACCURACY AND PRECISION WHEN wipe, and a waste container. TRANSFERRING LIQUIDS: - Attach a disposable tip to the 1. Use a fresh tip for each sample to Pipetman by pressing its plastic avoid contamination. shaft firmly into the tip with a 2. Pre-rinse each new tip. Most twisting motion. liquids leave a film on the inside - This action should ensure an wall of a new tip, which affects airtight seal and prevent leaks the actual volume that can be from occurring. transferred. By pre-rinsing a tip, these volume variances are troubleshooting guidelines may fix your eliminated. problem. - To pre-rinse a tip, simply draw 1. Calibration: solution into the tip and expel it - To ensure that your Pipetman is into a waste container. actually delivering the desired 3. Check the immersion depth. volume, you should periodically - If the immersion depth is too calibrate it. deep, liquid may get into the - To calibrate a Pipetman: plastic shaft and damage the a. determine how many grams of mechanics of the Pipetman. your liquid equals 1 milliliter - If the immersion depth is too from its density. shallow, air may be drawn into - For example, if water is the the tip, which greatly affects solvent, then 1 milliliter of water accuracy. equals 1 gram of water. - As a general rule, immerse the b. Tare a small container on an weight tips to a depth of 1 to 2 analytical balance and set the milliliters, the yellow tips to a Pipetman at a volume near depth of 2 to 3 milliliters, and the the top of its range. blue tips to a depth of 2 to 4 c. Transfer the set volume of millimeters. liquid to the tared container on - If you do find an air bubble in the the balance and note the tip, dispense the sample into a mass of the liquid. waste container, check your - Repeat this step five more times immersion level, and pipette and determine the mean weight more slowly. and standard deviation from your - If you still see an air bubble, measurements. discard your current tip and use a d. Repeat the entire process new one. with the Pipetman set at a 4. Release the plunger button volume near the bottom of its slowly. Do not let the plunger range. snap up. - If these values deviate from what - This violent action may force you expect, adjust the pipetman liquid into the plastic shaft. accordingly when you transfer 5. When wiping excess liquid off the liquid. sides of the tip, avoid touching 2. Checking for leaks: the tip opening because some of - Occasionally, the Pipetman will the sample may be lost. begin to leak due to improperly 6. When liquid is in its tip, never fitted tips or worn o-rings. allow the Pipetman to tilt more a. To check for leaks, put a fresh than 20 degrees from vertical and tip on the Pipetman. never lay the Pipetman on its b. Pre-rinse the tip and draw up side. liquid. c. Keep the Pipetman vertical BE CONSISTENT: and wait about 20 seconds. For optimum reproducibility, it is d. If you see a droplet, then the important to be consistent with Pipetman is leaking. your samples. Reminder: Use a consistent Reminder: Inaccuracies of the speed and smoothness when Pipetman results from: pressing and releasing the push a. below 4 or button. b. above 70 degrees Celsius Reminder: Use consistent and pressure at the first step. c. using the Pipetman with acids Reminder: Immerse the tip to a and corrosives consistent depth. - If you absolutely must pipette an Reminder: Hold the pipetman at a acid or corrosive, then you should consistent angle that is less than disassemble the Pipetman 20 degrees from vertical. immediately to inspect and clean the piston, shaft, seal TROUBLESHOOTING: assemblies, and o-rings. If you are still seeing errors with accuracy or precision, the following SUMMARY – KEY POINTS To summarize, using proper technique 2. Rinse the electrode using distilled with an automatic pipette is essential for water ensuring accuracy and precision. - Various ph meter will have diff 1. Be consistent with plunger timing ways for calibration (read the and pressure as well as manual) immersion time and depth. - Begin the calibration using the 2. Use a fresh, pre-rinsed tip for standard with ph 7: each new sample. o Once the standard has 3. Check for air bubbles and leaks during each transfer. been calibrated, you need 4. And calibrate the automatic to remove the sample and pipette periodically. rinse the electrode with 5. In addition, take care not to drop distilled water. the expensive and easily o Repeat the process again damaged Pipetman. with the ph of 4. 6. Do not allow the tip to touch o Once the standard has anything except sample. calibrated, remove the 7. Do not introduce liquid into the sample and rinse plastic shaft by letting the plunger snap up or by laying the electrode with dw. Pipetman on its side. o Then calibrate with ph of 8. Finally, avoid using the Pipetman 10. Rinse off again the under extreme temperature electrode with dw to conditions or with acids and ensure the accuracy of corrosives. measurement. 3. Measure the ph of unknown When used with consistency and care, solution an automatic pipette is an excellent - Place the unknown sol in means electrode of transferring microliter quantities of - Sample has 5.57 and temp of liquid. 23.9 deg C. 4. Remove the sample after reading 5. Rinse the electrode with dw 6. Replace the protective cap for USING A PH METER storage ph meters can be small or hand held larger stationary video used larger stationary FUME HOOD SAFETY: Dos and DON’Ts PARTS: - Fumehood safety should be 1. meter taken very seriously. You should 2. arm – allows electrode to easily regularly review the move up and down into a beaker recommended safe practices with the sample to be tested listed on your lab fume hoods 3. electrode corner post. Don't be like Party use a ph meter should be Scooter. calibrated using color coded When using substances that nuffers of fixed ph (4,7,10 are produce hazardous levels of standard) airborne chemicals such as gas STEPS: fumes vapors and dust, make sure the selected enclosure is 1. Remove the protective cap from appropriate for the work you the electrode. intend to do. - Electrode need to be stored - Consult your safety officer for a properly reference on what type of safety - Cup should be immersed in equipment is required. stored solution preventing them - The safety tips in this video to dry out applied to chemical fume hoods specifically. Scooter works with dangerous scooter keeps things that he chemicals and in appropriate needed long ago and things he enclosures. Don't be like scooter. might someday need in his fume Do not place your head inside the hood and that thing. hood. The fume hood interior adjust the sash to the smallest should be considered a opening possible when in use. contaminated space at all times. locate the sash at a comfortable Don't lean in to the contaminated height just above the elbows and fume hood like scooter. look into the hood through the Minimize drafts and sudden glass. the sash glass should be movements in front of the hood used as a physical barrier to because turbulence generated in protect the user's breathing area front of a chemical fume hood whenever possible. can cause loss of containment. scooter likes to keep the sash Locate hoods away from heavy wide open says it gives him foot traffic. Don't be like scooter. elbow room. Work a minimum of 6 inches close the sash when the hood is inside the hood. Potential loss of unattended. Close the sash containment increases completely protects passers-by in significantly at distances within 6 the event of a projectile reaction inches of the plane of the sash. inside the hood. Place all hazards beyond this additionally in a variable air threshold. Don't keep your work volume or Vav mechanical dangerously close to the front of system this improves energy the work surface like scooter is efficiency. livid on the edge. scooter shows blatant disregard Elevate equipment above the for his lab mates. work surface elevated apparatus do not remove any of the hood is much less obstructive to the components. multiple airflow through the hood. components of a chemical fume Equipment left flat on the work hood are designed to manipulate surface interferes with air flow air flow into and through the needed below to efficiently clear hood. these components contaminants. typically work together in Don't block airflow with harmony and a single equipment like scooter. modification can result in severe Keep sill and baffle unobstructed. loss of containment. The sill or airfoil of the hood is Scooter fancies himself critical to containment, so avoid something of a fume hood blocking it. engineer and he likes to make - All of the airflow through the hood changes on the fly. don't be like eventually moves through the scooter. baffles at the back. Obstructing do not place flammable solvents these baffles can cause uneven near heat flame or sparks. if face velocities and potential loss flammable materials and heat of containment. sources must coexist in the same Don't block the airfoil with notes hood, take great care and work reminders and other obstructions practices and separate them as like scooter. much as possible to avoid an Do not use the hood for storage. accidental fire. Because hoods work more scooter recklessly places efficiently without obstructions flammable solvents near heating keep unneeded objects to a devices. minimum. do not evaporate large amounts chemicals should be capped and of flammable liquids especially in stored in the appropriate a low air volume situation such as chemical storage cabinet with the sash closed on a Vav mechanical system. minimize the evaporation rate of flammables by keeping the vessels covered whenever possible. scooter is a sucker for convenience so he leaves flammable agent containers wide open see you tomorrow scooter if the lab is still here. wipe up spills immediately spilling a solvent increases its surface area evaporation rate and probability of exposure. Routinely validate air flow. Proper air flow through the fume hood is critical to safe operation. Annual inspection of the face velocity with a calibrated instrument should be minimum protocol. Into complete ASHRAE 110 containment assessment is strongly recommended. Scooter doesn't know if anyone has ever done that and he's not about to start caring now. If the ventilation system malfunctions or if the airflow alarm indicates an unsafe condition close the sash, discontinue hood operation immediately and call for help. Do not use the hood with bio hazards or perchloric acid. Remember your chemical fume hood is designed for a very specific family of hazardous substances. Working with biological hazards requires the use of an enclosure or cabinet classified for biological applications. The use of perchloric acid requires a fume hood designed and dedicated to the use of this specific very hazardous chemical. Don't be like scooter. Take protecting your laboratory environment seriously and pay attention to recommended practices when using any chemical fume hood.