You are on page 1of 6

USING AN AUTOMATIC PIPETTE adjustment knob, a volume

indicator, and a plastic shaft.


 Many laboratory procedures, - are available in six different sizes.
especially biological and Five of them are depicted here.
biochemical ones, require the
precise transfer of small
quantities of liquids.
 Glass pipettes, which are
commonly used in chemistry
labs, are simply not accurate
when working with volumes that
are less than 1 milliliter.
 The best method of transferring - The number on the plunger
small volumes of liquids involves button indicates the maximum
the use of an automatic pipette. volume in microliters that each
Pipetman is capable of
Automatic pipette transferring.
- consists of a piston that - For example, the P20 has a
displaces air to create a vacuum possible volume range of 0 to 20
when its plunger is pressed. microliters.
- The tip of the automatic pipette is  Reminder: Transferring volumes
then placed into the sample vial at the low end of the range is not
and the plunger is slowly recommended because the
released. standard deviation is too high to
- This action draws liquid into the ensure precision.
tip. - Thus the minimum amount of
- Finally, this liquid is expelled into liquid that can be transferred in
the desired container by slowly microliters is 0.1 for the P2, 0.5
depressing the plunger a second for the P10, two for the P20, 10
time. for the P100, 50 for the P200,
and 100 microliters for the

- As shown, a variety of automatic


pipettes exist. This video will P1000.
demonstrate automatic pipetting  Now that we've covered what an
technique using the Pipetman. automatic pipette is and how it
works, we'll take a detailed look
into how the Pipetman is used in
the laboratory.

STEPS IN USING PIPETMAN:


- Let's say, for example, your
procedure requires you to
transfer 153.5 microliters of
 Using good technique will ensure sample.
accuracy, which is defined as 1. Selection of the Pipetman with
the closeness of the dispensed an appropriate volume range.
volume to the set volume, and - In this example, a couple of
precision, which is defined as choices exist. The P200 has a
the reproducibility of individual recommended volume range of
measurements between your 50 to 200 microliters while the
samples. P1000 has a recommended
range of 100 to 1,000 microliters.
Pipetman  Reminder: The lower end of a
- consists of a plunger button, a tip volume range always has a
ejector button, a volume higher standard deviation.
Therefore, the P200 is the better - Always use a new sterile tip for
choice. each sample and never let the tip
2. Selection of the proper touch anything except sample.
disposable plastic tip. a. To draw a sample into the
- Three tip sizes plastic tip, depress the
o Large blue tips - which fit plunger until resistance is felt.
the P1000 b. Hold the Pipetman vertically
o medium yellow tips - which and immerse it into the
fit the P200, P100 sample.
o small white tips - which fit c. After the tip is immersed,
the P10 and P2. slowly release the plunger to
- Because we're using the P200, draw liquid into the tip. Wait a
we're going to need the yellow few seconds to ensure that
tips. the full volume of sample is
 Reminder: Using the proper tip is drawn into the tip.
essential for preventing leaks and
promoting accurate delivery of  Reminder: Wait one second for
liquids. the P2 to P200 Pipetman and 2
3. Setting the desired transfer to 3 seconds for the P1000.
volume.  Reminder: The larger the volume
- For the smaller Pipetman, P2 to or the more viscous the sample,
P200, the volume set on the the longer you should wait.
indicator translates exactly into
the volume that will be delivered. d. Remove the Pipetman from
- This P200 is set to deliver a the sample and carefully wipe
volume of 125.9 microliters. excess liquids from the sides
- For the P1000, the volume set on of the tip.
the indicator must be multiplied e. To dispense the liquid in the
by 10 to determine the actual Pipetman, touch its tip to the
volume that will be transferred. side wall of a new container
- This P1000 is set to deliver a and slowly depress the
volume of 1,259 microliters. plunger to the first stop.
- To correctly adjust the volume f. Wait a few seconds to ensure
setting, turn the volume complete sample flow, then
adjustment knob to one third of a press the plunger to the
revolution past the desired second step to expel any
volume. Then slowly move the remaining liquid.
dial back down to the desired g. With the plunger fully
setting. depressed, carefully remove
 Reminder: Always approach the the tip from the container by
desired volume by dialing sliding it up the container wall.
downwards to prevent The importance of sliding the
mechanical backlash and to table along the wall increases
ensure precise volume transfer. as the volumes become
- When adjusting the volume of a smaller.
Pipetman, take care not to rotate h. Finally, allow the plunger to
the dial past the maximum. You return to the up position.
cannot transfer volumes above - If you are done working with a
the range of the Pipetman. particular sample, press the tip
4. Transfering your sample. ejector button to release the tip
- You will need your appropriately into the proper waste receptacle.
sized Pipetman set to the desired
volume, the correctly sized tips, a GUIDELINES TO ENSURE
rack, the new container, a chem ACCURACY AND PRECISION WHEN
wipe, and a waste container. TRANSFERRING LIQUIDS:
- Attach a disposable tip to the 1. Use a fresh tip for each sample to
Pipetman by pressing its plastic avoid contamination.
shaft firmly into the tip with a 2. Pre-rinse each new tip. Most
twisting motion. liquids leave a film on the inside
- This action should ensure an wall of a new tip, which affects
airtight seal and prevent leaks the actual volume that can be
from occurring. transferred. By pre-rinsing a tip,
these volume variances are troubleshooting guidelines may fix your
eliminated. problem.
- To pre-rinse a tip, simply draw 1. Calibration:
solution into the tip and expel it - To ensure that your Pipetman is
into a waste container. actually delivering the desired
3. Check the immersion depth. volume, you should periodically
- If the immersion depth is too calibrate it.
deep, liquid may get into the - To calibrate a Pipetman:
plastic shaft and damage the a. determine how many grams of
mechanics of the Pipetman. your liquid equals 1 milliliter
- If the immersion depth is too from its density.
shallow, air may be drawn into - For example, if water is the
the tip, which greatly affects solvent, then 1 milliliter of water
accuracy. equals 1 gram of water.
- As a general rule, immerse the b. Tare a small container on an
weight tips to a depth of 1 to 2 analytical balance and set the
milliliters, the yellow tips to a Pipetman at a volume near
depth of 2 to 3 milliliters, and the the top of its range.
blue tips to a depth of 2 to 4 c. Transfer the set volume of
millimeters. liquid to the tared container on
- If you do find an air bubble in the the balance and note the
tip, dispense the sample into a mass of the liquid.
waste container, check your - Repeat this step five more times
immersion level, and pipette and determine the mean weight
more slowly. and standard deviation from your
- If you still see an air bubble, measurements.
discard your current tip and use a d. Repeat the entire process
new one. with the Pipetman set at a
4. Release the plunger button volume near the bottom of its
slowly. Do not let the plunger range.
snap up. - If these values deviate from what
- This violent action may force you expect, adjust the pipetman
liquid into the plastic shaft. accordingly when you transfer
5. When wiping excess liquid off the liquid.
sides of the tip, avoid touching 2. Checking for leaks:
the tip opening because some of - Occasionally, the Pipetman will
the sample may be lost. begin to leak due to improperly
6. When liquid is in its tip, never fitted tips or worn o-rings.
allow the Pipetman to tilt more a. To check for leaks, put a fresh
than 20 degrees from vertical and tip on the Pipetman.
never lay the Pipetman on its b. Pre-rinse the tip and draw up
side. liquid.
c. Keep the Pipetman vertical
BE CONSISTENT: and wait about 20 seconds.
 For optimum reproducibility, it is d. If you see a droplet, then the
important to be consistent with Pipetman is leaking.
your samples.
 Reminder: Use a consistent  Reminder: Inaccuracies of the
speed and smoothness when Pipetman results from:
pressing and releasing the push a. below 4 or
button. b. above 70 degrees Celsius
 Reminder: Use consistent and
pressure at the first step. c. using the Pipetman with acids
 Reminder: Immerse the tip to a and corrosives
consistent depth. - If you absolutely must pipette an
 Reminder: Hold the pipetman at a acid or corrosive, then you should
consistent angle that is less than disassemble the Pipetman
20 degrees from vertical. immediately to inspect and clean
the piston, shaft, seal
TROUBLESHOOTING: assemblies, and o-rings.
If you are still seeing errors with
accuracy or precision, the following SUMMARY – KEY POINTS
To summarize, using proper technique 2. Rinse the electrode using distilled
with an automatic pipette is essential for water
ensuring accuracy and precision. - Various ph meter will have diff
1. Be consistent with plunger timing ways for calibration (read the
and pressure as well as manual)
immersion time and depth. - Begin the calibration using the
2. Use a fresh, pre-rinsed tip for
standard with ph 7:
each new sample.
o Once the standard has
3. Check for air bubbles and leaks
during each transfer. been calibrated, you need
4. And calibrate the automatic to remove the sample and
pipette periodically. rinse the electrode with
5. In addition, take care not to drop distilled water.
the expensive and easily o Repeat the process again
damaged Pipetman. with the ph of 4.
6. Do not allow the tip to touch o Once the standard has
anything except sample. calibrated, remove the
7. Do not introduce liquid into the
sample and rinse
plastic shaft by letting the plunger
snap up or by laying the electrode with dw.
Pipetman on its side. o Then calibrate with ph of
8. Finally, avoid using the Pipetman 10. Rinse off again the
under extreme temperature electrode with dw to
conditions or with acids and ensure the accuracy of
corrosives. measurement.
3. Measure the ph of unknown
When used with consistency and care, solution
an automatic pipette is an excellent - Place the unknown sol in
means electrode
of transferring microliter quantities of
- Sample has 5.57 and temp of
liquid.
23.9 deg C.
4. Remove the sample after reading
5. Rinse the electrode with dw
6. Replace the protective cap for
USING A PH METER
storage
 ph meters can be small or hand
held larger stationary
 video used larger stationary FUME HOOD SAFETY: Dos and
DON’Ts
PARTS:
- Fumehood safety should be
1. meter taken very seriously. You should
2. arm – allows electrode to easily regularly review the
move up and down into a beaker recommended safe practices
with the sample to be tested listed on your lab fume hoods
3. electrode corner post. Don't be like
 Party use a ph meter should be Scooter.
calibrated using color coded  When using substances that
nuffers of fixed ph (4,7,10 are produce hazardous levels of
standard) airborne chemicals such as gas
STEPS: fumes vapors and dust, make
sure the selected enclosure is
1. Remove the protective cap from appropriate for the work you
the electrode. intend to do.
- Electrode need to be stored - Consult your safety officer for a
properly reference on what type of safety
- Cup should be immersed in equipment is required.
stored solution preventing them - The safety tips in this video
to dry out applied to chemical fume hoods
specifically.
 Scooter works with dangerous  scooter keeps things that he
chemicals and in appropriate needed long ago and things he
enclosures. Don't be like scooter. might someday need in his fume
 Do not place your head inside the hood and that thing.
hood. The fume hood interior  adjust the sash to the smallest
should be considered a opening possible when in use.
contaminated space at all times. locate the sash at a comfortable
 Don't lean in to the contaminated height just above the elbows and
fume hood like scooter. look into the hood through the
 Minimize drafts and sudden glass. the sash glass should be
movements in front of the hood used as a physical barrier to
because turbulence generated in protect the user's breathing area
front of a chemical fume hood whenever possible.
can cause loss of containment.  scooter likes to keep the sash
 Locate hoods away from heavy wide open says it gives him
foot traffic. Don't be like scooter. elbow room.
 Work a minimum of 6 inches  close the sash when the hood is
inside the hood. Potential loss of unattended. Close the sash
containment increases completely protects passers-by in
significantly at distances within 6 the event of a projectile reaction
inches of the plane of the sash. inside the hood.
 Place all hazards beyond this  additionally in a variable air
threshold. Don't keep your work volume or Vav mechanical
dangerously close to the front of system this improves energy
the work surface like scooter is efficiency.
livid on the edge.  scooter shows blatant disregard
 Elevate equipment above the for his lab mates.
work surface elevated apparatus  do not remove any of the hood
is much less obstructive to the components. multiple
airflow through the hood. components of a chemical fume
Equipment left flat on the work hood are designed to manipulate
surface interferes with air flow air flow into and through the
needed below to efficiently clear hood. these components
contaminants. typically work together in
 Don't block airflow with harmony and a single
equipment like scooter. modification can result in severe
 Keep sill and baffle unobstructed. loss of containment.
The sill or airfoil of the hood is  Scooter fancies himself
critical to containment, so avoid something of a fume hood
blocking it. engineer and he likes to make
- All of the airflow through the hood changes on the fly. don't be like
eventually moves through the scooter.
baffles at the back. Obstructing  do not place flammable solvents
these baffles can cause uneven near heat flame or sparks. if
face velocities and potential loss flammable materials and heat
of containment. sources must coexist in the same
 Don't block the airfoil with notes hood, take great care and work
reminders and other obstructions practices and separate them as
like scooter. much as possible to avoid an
 Do not use the hood for storage. accidental fire.
Because hoods work more  scooter recklessly places
efficiently without obstructions flammable solvents near heating
keep unneeded objects to a devices.
minimum.  do not evaporate large amounts
 chemicals should be capped and of flammable liquids especially in
stored in the appropriate a low air volume situation such as
chemical storage cabinet with the sash closed on a Vav
mechanical system. minimize the
evaporation rate of flammables
by keeping the vessels covered
whenever possible.
 scooter is a sucker for
convenience so he leaves
flammable agent containers wide
open see you tomorrow scooter if
the lab is still here.
 wipe up spills immediately
spilling a solvent increases its
surface area evaporation rate
and probability of exposure.
 Routinely validate air flow. Proper
air flow through the fume hood is
critical to safe operation.
 Annual inspection of the face
velocity with a calibrated
instrument should be minimum
protocol. Into complete ASHRAE
110 containment assessment is
strongly recommended.
 Scooter doesn't know if anyone
has ever done that and he's not
about to start caring now.
 If the ventilation system
malfunctions or if the airflow
alarm indicates an unsafe
condition close the sash,
discontinue hood operation
immediately and call for help.
 Do not use the hood with bio
hazards or perchloric acid.
Remember your chemical fume
hood is designed for a very
specific family of hazardous
substances.
 Working with biological hazards
requires the use of an enclosure
or cabinet classified for biological
applications. The use of
perchloric acid requires a fume
hood designed and dedicated to
the use of this specific very
hazardous chemical.
 Don't be like scooter. Take
protecting your laboratory
environment seriously and pay
attention to recommended
practices when using any
chemical fume hood.

You might also like