SOMATOM Spirit

CT

Software
Installation of SOMARIS/5.5 VB27

This document is valid for:


SOMATOM Spirit © Siemens AG 2005
The reproduction, transmission or
use of this document or its contents
is not permitted without express
written authority. Offenders will be
liable for damages. All rights,
including rights created by patent
grant or registration of a utility
model _or_ design,_are_ reserved.

English
Print No.: CT02-025.816.03.01.02 Doc. Gen. Date: 11.05
Replaces: n.a.
0-2 Revision

Chapter Page Revision


0 to 6 all 01

Document revision level


The document corresponds to the version/revision level effective at the time of system delivery. Revi-
sions to hardcopy documentation are not distributed automatically.
Please contact your local Siemens office to order current revision levels.

Disclaimer
The installation and service of equipment described herein is to be performed by qualified personnel
who are employed by Siemens or one of its affiliates or who are otherwise authorized by Siemens or
one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated with or autho-
rized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or
one of its affiliates before attempting installation or service procedures.

SOMATOM Spirit CT02-025.816.03 Page 2 of 4 Siemens AG


Rev. 01 11.05 CSPS22 Medical Solutions
Contents 0-3

Page
1 _______General _______________________________________________________ 1 - 1

Definitions and Abbreviations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-1


Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
Software not approved by Siemens . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
Software kit VB27 for ICS and IRS . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
Backup media . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1 -4

2 _______SW Installation ICS _____________________________________________ 2 - 1

Conditions for starting the SW installation. . . . . . . . . . . . . . . . . . . . . . . .2-1


Required data carrier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-1
Required data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-1
Regional Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
Service key . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
Host name . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
Write down the host name.
System type and serial numbers . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
Preparations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 - 3
Starting the Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-4

Installation running . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 -6
Configure the keyboard language and enable Windows XP Access . . . . . . . .2-6
Restore Steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-8
Restore 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-8
Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-9
Restore 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 11
Completion of configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 12
Import Reference Images . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 12
Regional Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 12
Restart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 12
Siemens Remote Services . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 13
Functional Check of SRS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 13
SRS Basic Settings Outside NAT . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 14
Functional Check of System Management . . . . . . . . . . . . . . . . . . . . 2 - 15
Remote Access . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 15

3 _______SW Installation of IRS ___________________________________________ 3 - 1

Installation of IRS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-1


Final steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-2
Software also reinstalled on the ICS. . . . . . . . . . . . . . . . . . . . . . . . .3-2
No software reinstallation on the ICS . . . . . . . . . . . . . . . . . . . . . . . .3-2

4 _______Additional Worksteps ___________________________________________ 4 - 1

Execution of perl script "repair_config.pl" . . . . . . . . . . . . . . . . . . . . . . . . 4 - 1


Backup at the ICS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 1

Siemens AG CT02-025.816.03 Page 3 of 4 SOMATOM Spirit


Medical Solutions Rev. 01 11.05 CSPS22
0-4 Contents

Disable Windows XP Access . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 2

5 ______ Special Language Settings _______________________________________5 - 1

Chinese. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
Settings in Local Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
Settings under meduser . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
Settings under administrator . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Japanese . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5
Settings in Local Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5
Settings under meduser . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5
Settings under administrator . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-7

6 ______ Changes to previous version _____________________________________6 - 1

SOMATOM Spirit CT02-025.816.03 Page 4 of 4 Siemens AG


Rev. 01 11.05 CSPS22 Medical Solutions
1General 1-1

1.1 Definitions and Abbreviations


Abbreviation Meaning
ACR-NEMA American College of Radiology - National Electrical
Manufacturers Association
BIOS Basic Input Output System, lowest PC configuration settings
AET Application Entity Title - Name of DICOM service participating in the
communication either as user (SCU) or provider (SCP)
CD Compact Disc
CDR Compact Disc Recordable
CT Computed Tomography
DICOM Digital Image Communication in Medicine (ACR-NEMA standard)
DHCP Dynamic Host Configuration Protocol - procedure for assigning
numeric IP address to nodes in which blocks of addresses are leased
to clients on a first-come, first-serve basis. Typically, DHCP is com-
bined with WINS, meaning an IP address is both dynamically
assigned and automatically associated with a domain name. This can
be combined with DNS, resulting in a dynamically assigned numeric
IP address that is automatically associated with both NetBIOS and
Internet names.
DNS Domain Name Services - address resolution procedure that associ-
ates numeric IP addresses with Internet-style names
FTP File Transfer Protocol - a client-server application that transfers files
from one computer to another
Gateway A display that controls traffic between two networks using factors
above the network layer
ICS = Navigator Image Control System - control computer for the entire CT scanner,
with interface to the acquisition system
IMS Imaging System, consisting of ICS, IRS and console
IRS Image Reconstruction System - reconstruction computer
LAN Local Area Network (here, always Ethernet)
LMHOSTS Address resolution procedure which statically associates numeric IP
addresses with NetBIOS names. The LMHOSTS file is checked to
see if the name is being addressed in this file.
MEP Market Entrance Phase
MOD Magneto-Optical Disk
NetBIOS Network Basic Input Output System - a collection of network services
and programming interfaces. NetBIOS services run on network proto-
cols such as TCP/IP.
NIC Network Interface Card - an expansion card that provides a physical
connection to the network

Siemens AG Page 1 of 4 CT02-025.816.03 SOMATOM Spirit


Medical Solutions CSPS22 Rev. 01 11.05
1-2 General

Abbreviation Meaning
NT "New Technology" Operating System - trademark of Microsoft, Inc.
OI Operating Instructions
PB Patient Browser
PPP Point-to-Point Protocol - a data link and network protocol designed for
use over serial connections (such as dial-up modems)
RAS Remote Access Services - a Windows NT service that supports dial-
in and dial-out connections using modems and ISDN.
Router A multiport computer or piece of dedicated hardware that examines
network layer addresses of packets on a network and retransmits
them on a different network if necessary
SCU Service Class User
SCP Service Class Provider
TCP/IP Transport Control Protocol/Internet Protocol - the most widely used
standard for networking
UI User Interface
USC Uptime Service Center
WINS Windows Internet Name Service - address resolution procedure that
dynamically associates numeric IP addresses with NetBIOS names.
It is possible to combine DNS and WINS.
XP "eXPerience" Operating System - trademark of Microsoft, Inc.

SOMATOM Spirit CT02-025.816.03 Page 2 of 4 Siemens AG


Rev. 01 11.05 CSPS22 Medical Solutions
General 1-3

1.2 Overview
ICS + IRS = Navigator
In the standard configuration, the ICS is the main console for interchangeable scanning
and imaging functions. The IRS is equipped with four computers as well as switchable
keyboard, mouse and monitor.

1.3 Software not approved by Siemens


Warning Unauthorized or incorrect operation or changes to software can
cause malfunctions in your system. This can result in injuries or
system defects. For this reason, the use or installation of soft-
ware not approved by Siemens is strictly prohibited.

1.4 Software kit VB27 for ICS and IRS


The software installation kit for SOMATOM Spirit systems is included in the delivery vol-
ume and comprises the following items required for a complete SW installation:

• CD OEM-Setup Windows XP(1 of 3)


• CD MUI_CHN_JPN (CD 1A/3)
• CD SOM/5.5 ICS SW VB27 (CD 2/3)
• CD Som/5.5 ICS SWFW VB27 (CD 3/3)
• 1x CD Som/5.5 IRS SW VB27

Siemens AG Page 3 of 4 CT02-025.816.03 SOMATOM Spirit


Medical Solutions CSPS22 Rev. 01 11.05
1-4 General

1.5 Backup media


The CD-Writer (drive [-R-] CD-R) is the standard backup device for ICS. Alternatively, the
MOD may be used, if available.
• Make sure a separate backup media (e.g. "CD Recordable") is on hand for the ICS.

SOMATOM Spirit CT02-025.816.03 Page 4 of 4 Siemens AG


Rev. 01 11.05 CSPS22 Medical Solutions
2 SW Installation ICS 2-1

2.1 Conditions for starting the SW installation

To ensure that all archiving, filming, copying and networking jobs are finished and that the
respective queues are empty:
• Open the patient browser.
• Click Transfer -> Select "Local Job Status" and, if necessary, delete all entries.
• Click Transfer -> Select "Network Job Status" and, if necessary, delete all entries.
• Click Patient -> Select "Film Task Status" and, if necessary, delete all entries.

2.2 Required data carrier


To perform the SW installation of the Navigator (ICS), the following data carriers are
required:

CD OEM-Setup Windows XP
CD MUI_CHN_JPN (CD 1A/3)
CD SOM/5.5 ICS SW VB27 (CD 2/3)
CD Som/5.5 ICS SWFW VB27 (CD 3/3)
Backup CD-R
CD System Management CA Unicenter

2.3 Required data


The ICS has to be switched on.
Read the following configuration data and write it down in the fields provided. These data
are required during SW Installation of the ICS and will not be restored.

It may be useful to store the following file to floppy in case of problems with the backup
files or during installation. This file contains the complete configuration of the system.

C:\Somaris\Service\html\Reports\report.htm

Siemens AG Page 1 of 16 CT02-025.816.03 SOMATOM Spirit


Medical Solutions CSPS22 Rev. 01 11.05
2-2 SW Installation ICS
2.3.1 Regional Settings
• From the syngo user interface, select -> Options -> Configuration.
• In the configuration panel, double-click the Regional and language options icon.
• Write down the following data:

Regional Options Language


Customized Time Time separator:
Date Date separator:
Languages Input locales

2.3.2 Service key


• From the syngo user interface, select Options -> Local Service.
• Write down the service key.

Service key (level=7)


Service key = 20 characters ---------------------- 14 characters ---------------- --- last 6 characters ---

2.3.3 Host name


• In the main menu of the service software, select Configuration.
• Click Next.
• Select Local Host -> TCP/IP LAN.

2.3.4 Write down the host name.

Computer name
IP address

NOTE Setting at Local Host- TCP/IP LAN a DNS-Domain e.g "med.sie-


mens.de" will not be in the backup. Please note the configuration.

System type and serial numbers

Name Identifier for Material No. Serial No.


backup files
SOMATOM Spirit P15A 100 45 692 31xxx

SOMATOM Spirit CT02-025.816.03 Page 2 of 16 Siemens AG


Rev. 01 11.05 CSPS22 Medical Solutions
SW Installation ICS 2-3

2.4 Preparations
• Insert the CD OEM-Setup Windows XP in the Navigator.
• From the ICS, select System/End and Shut down the system.
• Switch off the system at the UPS-Rack (system -off).
• Disconnect the network cable of the ICS from the Ethernet switch.

Siemens AG Page 3 of 16 CT02-025.816.03 SOMATOM Spirit


Medical Solutions CSPS22 Rev. 01 11.05
2-4 SW Installation ICS

2.5 Starting the Installation


• Switch on the system at the UPS-RACK (system-on).
¹ The computer boots from the CD-ROM.

Starting Windows 98
......

After a short while, the following window appears:

Siemens Medical Solutions syngo CT 2005A installation


SELECTION MENU INFORMATION WINDOW

Turn off the computer to interrupt

or press any key to continue Creating a new primary partition on


first hard drive. This partition will
be converted to NTFS during Windows XP
installation.
CONTINUE INSTALLATION

HARDWARE DETECTION IN PROGRESS

Checking first hard drive

• Press <Enter> to continue installation (the ICS is beeping). This reboots the ICS.
• Select distribution type: CUSTOMER (default).
• Enter the machine name (host name) as noted by replacing the default entry "MedPC"
and press Enter.

NOTE Please note that the computer name must not contain blanks or
underscores. Use the following naming convention for the ICS
computer name: ct<serial number> (e.g. ct123456).

• Select Keep data on second hard drive or


select Clear data on second hard drive (deletes patient database).
Use ENTER to change the selection.

NOTE Do not use Clear all harddrives 1-2. This option will format all hard
drives only (removes all software). Reinstallation of SW will not be
started.

SOMATOM Spirit CT02-025.816.03 Page 4 of 16 Siemens AG


Rev. 01 11.05 CSPS22 Medical Solutions
SW Installation ICS 2-5

NOTE Using Keep Data on other Hard Drive will extend the software
installation time by up to 3 hours! The duration depends on the
number of images stored in the database.

• Use CONTINUE INSTALLATION to confirm the start of installation (ICS is beeping if


"Clear data" is selected).
• Press any key.
¹ All necessary files are copied from CD OEM-Setup Windows XP to the hard
drive.
Chinese and Japanese language users: please read the note at the bottom before
the CD is ejected!

After approximately 12 minutes, the CD OEM-Setup Windows XP is ejected and has to


be removed.

For Chinese and Japanese language users only:


• Insert the additional CD MUI_CHN_JPN (CD 1A/3). All other users proceed with the next
step.

NOTE CD MUI_CHN_JPN (CD 1A/3) has to be inserted directly after the


OEM CD was ejected.
If you missed to insert this CD in time, the computer asks for CD SOM/
5.5 ICS SW VB27 (CD 2/3).
You have to start the installation at the beginning.
For CD MUI_CHN_JPN (CD 1A/3), no warning is generated as it is for
the other CD’s, if you did not insert it in time.

• Insert the CD SOM/5.5 ICS SW VB27 (CD 2/3).


The installation continues automatically.

NOTE If you missed inserting the requested CD in time, a warning will come
up:
No package has been found. Please insert the CD "........" in the CD
drive to continue the installation.
Insert the requested CD and press any key to continue.

Siemens AG Page 5 of 16 CT02-025.816.03 SOMATOM Spirit


Medical Solutions CSPS22 Rev. 01 11.05
2-6 SW Installation ICS

2.6
Installation running
From this point on, the ICS installation runs automatically.

After approximately 45 minutes, CD SOM/5.5 ICS SW VB27 (CD 2/3) is ejected and has
to be removed.

¹ Remove CD SOM/5.5 ICS SW VB27 (CD 2/3) and insert


CD Som/5.5 ICS SWFW VB27 (CD 3/3) and close the CD drive. The system
reboots automatically after finishing the copy job.

NOTE While the installation is running, you may start the installation of
the IRS, if necessary (refer to chapter 3).

Completion of the ICS installation (approx. 75 minutes) is indicated by a pop-up window


displaying the message Please enter license.
• Click ’Click here to start the Local Service Configuration’ to start the service software
(complete service license required).

2.6.1 Configure the keyboard language and enable Windows XP Access

Perform the following steps to configure keyboard languages other than English.

• Select Start -> Settings -> Control Panel


• Search for and click Date, Time, Language and Regional and Options
• Click Regional and Language Options.
• Select the Languages -> Details tab card
• Click Add and select the correct Input Language and confirm with OK.
• Select the English keyboard language, click Remove and confirm with OK.
¹ A message window appears informing you that removal is only possible after
reboot. Confirm with OK
• Click OK in the remaining windows or close remaining windows.

Perform the following steps to enable Windows XP access at the time of update.

• Enter the Service Key and open the Local Service main menu.
• From the home menu of the service software , select Utilities.
• Under Source, select Escape to OS and at Command NT Command Interpreter.
• Under Parameters, type c:\Somaris\Tools\bin\KeyFilter.bat disable.

SOMATOM Spirit CT02-025.816.03 Page 6 of 16 Siemens AG


Rev. 01 11.05 CSPS22 Medical Solutions
SW Installation ICS 2-7
• Activate with Return or GO.
¹ Only a reboot makes the change effective. This reboot will be performed after
Restore 1!
• Click Home.

Siemens AG Page 7 of 16 CT02-025.816.03 SOMATOM Spirit


Medical Solutions CSPS22 Rev. 01 11.05
2-8 SW Installation ICS

2.7 Restore Steps

NOTE If you install the IRS software, wait until the IRS installation fin-
ishes and ensure that the IRS is ready before performing the sub-
sequent steps.
Reason: after Restore 1, the ICS tries to transfer the IRS "default
tables" to the IRS. If the IRS is not ready at that time, the transfer
will fail and the IRS "default tables" will have to be transferred
manually.

2.7.1 Restore 1
• In the main menu of the service software, select the Backup & Restore function.
• Insert the Navigator backup CD.
• From Command, select Restore.
• From Drives, select the corresponding drive.
• From Archive, select <hostname>_SW-Settings02-VB27_date_time_ok.ar.
(This is only an example!)

Command Drives Groups


Restore [-R-] CD-R UserSettings
Service\LocalHost\Users
Archive Service\TCPIP\LAN-Settings
<hostname>_SW-Settings02-VB27_date&time.ar Service\TCPIP\RAS-Settings
Service\Database/Satellite Console
Service\Database\DBOffset
Service\Applications\kWorklistResults

• From Groups, select all except for User Settings and Service\Database\DBOffset and
click Go.
¹ BackupRestore Running ...; wait until Ready is displayed.

SOMATOM Spirit CT02-025.816.03 Page 8 of 16 Siemens AG


Rev. 01 11.05 CSPS22 Medical Solutions
SW Installation ICS 2-9
• From Archive, select <hostname>_CT-Settings-VB27_date_time_ok.ar.
(This is only an example!)

Command Drives Groups


Restore [-R-] CD-R
Mode
Archive Tube
<hostname>_CT-Settingsr-VB27_date&time.ar Service
IRS
General
Firmware

• From Groups, select all except IRS and click Go.


¹ BackupRestore Running ...; wait until Ready is displayed.

• From Archive, select <hostname>_SecuritySettings-VB27_date_time_ok.ar.


• From Groups, select all and click Go.
• Wait until "Ready".

• Click Home.

NOTE At this point the SW installation of IRS has to be completed.


• If not already done, remove the CDs from the IRS computer.

2.7.2 Configuration
• In the main menu of the service software, select the Configuration function.
• Select Site Info and click >. Check/correct the Time zone
Note: do not activate the NTP Time server or select automatic Daylight Time
• If changes were made, click Save. Confirm the message displayed with OK.

NOTE During the next steps, host names will be used. Host names
always have to be written in lower case characters.

• Select Local Host -> Technical.


- At "Country for Quality" select USA or other.
- Click Save and Ok.

• Click Finish and Home.


• Select Start -> Shutdown -> System Restart
¹ Shutdown and Reboot may take approx. 15 min.

Siemens AG Page 9 of 16 CT02-025.816.03 SOMATOM Spirit


Medical Solutions CSPS22 Rev. 01 11.05
2 - 10 SW Installation ICS
• Switch on the gantry if it is not already.
• Connect the ICS network cable to the Ethernet switch.

• Press Cancel at the upcoming Checkup Platform

SOMATOM Spirit CT02-025.816.03 Page 10 of 16 Siemens AG


Rev. 01 11.05 CSPS22 Medical Solutions
SW Installation ICS 2 - 11
2.7.3 Restore 2
• In the main menu of the service software, select the Backup & Restore function.
• From Command, select Restore.
• From Drives, select the corresponding drive.
• From Archive, select <hostname>_SW-Settings02-VB27_date_time_ok.ar.

(This is only an example!)

Command Drives Groups


Restore [-R-] CD-R UserSettings
Service\LocalHost\Users
Archive Service\TCPIP\LAN-Settings
<hostname>_SW-Settings02-VB27_date&time.ar Service\TCPIP\RAS-Settings
Service\Database/Satellite Console
Service\Database\DBOffset
Service\Applications\kWorklistResults

NOTE The following red error message during the restore of SW-Set-
tings can be ignored
cmd /C C:\Somaris\Config\BaReUp\CapPb\update.bat

• From Groups, select User Settings and click Go.


¹ BackupRestore Running ...; wait until Ready is displayed.

• From Archive, select <hostname>_SecuritySettings-VB27_date_time_ok.ar.


• From Groups, select All and click Go.
¹ BackupRestore Running ...; wait until Ready is displayed.

• From Archive, select <hostname>_CT-Settings-VB27_date&time.


(This is only an example!)

Command Drives Groups


Restore [-R-] CD-R Mode
Tube
Archive Service
<hostname>_CT-Scanner-VB27_date&time.ar General
IRS
Firmware

• From Groups, select IRS and click Go.


¹ BackupRestore Running ...; wait until Ready is displayed.
• Take out the backup CD.
• Click Home.

Siemens AG Page 11 of 16 CT02-025.816.03 SOMATOM Spirit


Medical Solutions CSPS22 Rev. 01 11.05
2 - 12 SW Installation ICS

2.8 Completion of configuration


2.8.1 Import Reference Images
This is mandatory if "Clear data on second hard drive" was selected in the installation
mask.
• From the patient browser, select Transfer -> Import from Offline.
• From the drop-down list, select the path H:\SiteData\ReferenceImages.
• Select all images and click OK.
¹ After the import, the images are available in the patient browser.

2.8.2 Regional Settings


• From the syngo user interface, select -> Options -> Configuration.
• In the configuration panel, double-click the Regional and language options icon.
• In the Regional Options tab card, select a preferred country and click Customize.
• Change the date, time and numbers as noted on page 2-2.
• In the Languages tab card, click Details and select the default input language type as
noted on page 2-2.

2.8.3 Restart
• Perform System/End -> Restart the System
(A restart is necessary to enable options and make the option tab cards available in the
customer UI (user interface)).

SOMATOM Spirit CT02-025.816.03 Page 12 of 16 Siemens AG


Rev. 01 11.05 CSPS22 Medical Solutions
SW Installation ICS 2 - 13

2.9 Siemens Remote Services


Navigator
2.9.1 Functional Check of SRS
Fig. 1
• Select Service -> SRS
• On page SRS Basic Settings:
Click " > " to perform a connection test to
the SRS access server.
¹ The connection test is successful if
both, the connection to the SRS server
and the connection to the Somatom
Life Extranet are OK.
If the check is not ok, perform the SRS
Configuration. Fill in the site-specific
entries on SRS Basic Settings page.
• 1 Outside NAT: normally outside NAT is
deselected. If outside NAT is used,
continue with 2.9.2
1 • 2 SRS Target Host (TCP/IP Address):
2 - Fürth for sites in Europe/Africa
- Newark for sites in America
3 - Singapore for sites in Asia/Australia
• 3 SRS Gateway (static route):
If a static route was configured before,
the IP address of the SRS gateway is
displayed under the TCP/IP address
(192.168.217.253 is only an example)
Press <Save> to save the Site Data.
Press ">" to perform a connection test

Siemens AG Page 13 of 16 CT02-025.816.03 SOMATOM Spirit


Medical Solutions CSPS22 Rev. 01 11.05
2 - 14 SW Installation ICS

Navigator
2.9.2 SRS Basic Settings Outside NAT
In an outside NAT environment, the customer network administrator only allows
certain IP addresses in his network. If a network node, e.g. the CT system, wants to
communicate with the "outside world" e.g. the SRS Access Server in Fürth, it can-
not use the real IP address 194.138.39.18. Instead it has to use a NAT address given
by the network administrator. For SRS, the adminstrator provides a NAT address
for the SRS Access Server and the 3 CA Unicenter Server. The NAT addresses have
to be entered in the SRS Basic Settings (Fig.12).

NOTE Outside NAT is used very rarily, except in countries such as Swe-
den, where the complete SRS environment is built up on Outside
NAT functionality.

Fig. 2
• Select Service -> SRS
• Selected 1 Outside NAT.
- SRS Target Host:
select the SRS server where your sys-
tem is connected to (Fürth for Europe/
Africa, Newark for America, etc.)
- for configuring the remaining IP ad-
dresses, contact the hospital network
administrator.
In our example the NAT-address is
10.10.10.18 - 22

• Press <Save> to save the Site Data.


• Click to next page via ">" to perform a
SRS connection check.

¹ The connection test is successful if


both, the connection to the SRS server
and the connection to the Somatom
Life Extranet are OK.

Fig. 3

SOMATOM Spirit CT02-025.816.03 Page 14 of 16 Siemens AG


Rev. 01 11.05 CSPS22 Medical Solutions
SW Installation ICS 2 - 15

Navigator
2.9.3 Functional Check of System Management
Fig. 4
• Select System Mgmt -> Agent_Control
• Check the status of System Management
as displayed in Fig.4. CA Unicenter is
correctly installed if both check boxes,
"System Management Base - Activated"
and "Installed" are enabled.
¹ If the check box "System Manage-
ment Base - Activated" is disabled,
mark the check box and press GO
to start the final CA Unicenter
Installation.

2.9.4 Remote Access


• Set up Remote Access as it was prior to the installation.

Siemens AG Page 15 of 16 CT02-025.816.03 SOMATOM Spirit


Medical Solutions CSPS22 Rev. 01 11.05
2 - 16 SW Installation ICS
This page intentionally left blank.

SOMATOM Spirit CT02-025.816.03 Page 16 of 16 Siemens AG


Rev. 01 11.05 CSPS22 Medical Solutions
3 SW Installation of IRS 3-1

3.1 Installation of IRS


• Power on the IRS.

Insert the IRS SW Installation CD into the CD drive.

• To reboot the IRS, press and hold the ON/OFF switch about 4 seconds. After the IRS is
switched off, switch it back on again.

NOTE On booting to the IRS CD, the automatic and unattended installa-
tion of Windows XP and of SOMARIS/5.5 image reconstruction
software starts up (approx. 30 min.). The IRS CD is ejected after
approx. 5 minutes.

• Remove the IRS-SW CD from the CD drive immediately after it is ejected.


• Press the ON/OFF switch twice again to reboot the IRS. The installation will now run
automatically.
• When the IRS software installation is finished (approx. 30 min.), continue with the final
steps

Siemens AG Page 1 of 2 CT02-025.816.03 SOMATOM Spirit


Medical Solutions CSPS22 Rev. 01 11.05
3-2 SW Installation of IRS

3.2 Final steps


3.2.1 Software also reinstalled on the ICS

Notice After Restore 1 (refer to chapter 2), the ICS and IRS synchronize
and the IRS "default tables" are transferred automatically from the
ICS to the IRS. With Restore 2 (refer to chapter 2), the "site-spe-
cific tune-up tables" are restored to the IRS. No further steps are
necessary.

3.2.2 No software reinstallation on the ICS

Notice During the IRS software installation, the IRS "default tables" are
deleted. The IRS "default tables" have to be transferred manually
to the IRS. As a result, the batch file IrsTablesInstall.bat has to be
executed. Afterwards, the "site-specific tune-up tables" have to
be restored.

• From the service software home menu, select Utilities.


• Under Source, select Escape to OS and at Command, NT Command Interpreter.
• Under Parameters, type c:\Somaris\utils\IrsTablesInstall.bat
• Activate with Return or GO.
¹ The "default tables" are transferred to the IRS.

• In the service software main menu, select the function Backup & Restore.
• Insert the Navigator backup CD.
• From Command, select Restore.
• From Drives, select the corresponding drive.
• From Archive, select CT-Settings-VB27.
(This is only an example!)

Command Drives Groups


Restore [-R-] CD-R Bolus
Tube
Archive Service
CT-Settings-VB27 date&time.ar IRS
General
Firmware

• From Groups, select IRS and click Go.


¹ The "site-specific tune-up tables" are transferred to the IRS.

• When "Ready" is displayed, remove the backup CD Navigator and restart the complete
system.

SOMATOM Spirit CT02-025.816.03 Page 2 of 2 Siemens AG


Rev. 01 11.05 CSPS22 Medical Solutions
4 Additional Worksteps 4-1

ICS

4.1 Execution of perl script "repair_config.pl"


• Log in as Administrator at the Navigator
- Select System -> End -> Restart Application
- Press Shift until the login window appears.
- Change user to administrator and login with the administrator password.
• In the open command window, type in perl C:\somaris\config\ repair_config.pl
• Press Return.
¹ This script will execute some checks and repairs.
• Close the command window.

ICS

4.2 Backup at the ICS


• Start the service software and select the Backup & Restore function in the main menu.
• Insert the backup CD-R for the ICS.
• From Command, select Backup.
• From Drives, select the CD drive, e.g., [-R-] CD-R.
• From Packages, select CT-Settings-VB27 and click Go.
¹ BackupRestore Running ...

Command Drives
Backup [-R-] CD-R

Packages
CT-Settings-VB27

¹ Wait until Ready is displayed.


• Close the CD tray.

Siemens AG Page 1 of 2 CT02-025.816.03 SOMATOM Spirit


Medical Solutions CSPS22 Rev. 01 11.05
4-2 Additional Worksteps
• From Packages, select SW-Settings02--VB27 and click Go.
¹ BackupRestore Running ...

Command Drives
Backup [-R-] CD-R

Packages
SW-Settings02-VB27

¹ Wait until Ready is displayed.


• Close the CD tray.
• From Packages, select Security-Settings-VB27 and click Go.
¹ BackupRestore Running ...

Command Drives
Backup [-R-] CD-R

Packages
Security-Settings-VB27

¹ Wait until Ready is displayed.

• When "Ready" is displayed, remove the backup media.


• Click Home and terminate the service software by closing the patient browser window.

NOTE The backup files have been saved correctly when ok is added at
the end of the backup filename.
e.g. <hostname>-SW-Settings02-VB27-date_time_ok.ar

4.2.1 Disable Windows XP Access


Perform the following steps to disable Windows XP access for the customer. Administrator
privileges are required; you must use the command shell of the service software.
• From the service software home menu, select Utilities.
• Under Source, select Escape to OS and at Command NT Command Interpreter.
• Under Parameters, type c:\Somaris\Tools\bin\KeyFilter.bat enable.
• Activate with Return or GO.
• Reboot to make the change effective.

To enable Windows XP access, type in


c:\Somaris\Tools\bin\KeyFilter.bat disable

SOMATOM Spirit CT02-025.816.03 Page 2 of 2 Siemens AG


Rev. 01 11.05 CSPS22 Medical Solutions
Special Language Settings 5 5-1

5.3 Chinese
5.3.1 Settings in Local Service
• In the main menu of the service software, select the Configuration function.
• Go to DICOM -> Character Set
• Set allow only one character set
• Select ISO_IR_100
• To save the settings, click Save.

Fig. 1

• Click Finish and Home.


¹ Starting Report generation ..., wait until ready
• Do not accept the reboot; click Cancel.

5.3.2 Settings under meduser

NOTE In some cases, the dialog window disappears. Do not try to


restart the "Regional Language Options" dialog . You can move
the dialog into foreground by pressing Alt + TAB.

• From the syngo user interface, select -> Options -> Configuration.

Siemens AG CT02-025.816.03 Page 1 of 8 SOMATOM Spirit


Medical Solutions Rev. 01 11.05 CSPS22
5-2 Special Language Settings
• In the configuration panel, click the Regional and Language Options icon.
Note: The window disappears. Click with the left mouse button to move the window into
foreground.
• In the Regional Options tab card, select Chinese (PRC)
• Under Location, select the country where you are.
• Click Apply.
• If necessary, press Alt + Tab to move the dialog window into foreground.

• To modify the appearance of the date,


time or numbers, click Customize.

Note: change the formats according to


the notes you made before the update.
There is only one restriction: A
maximum of 2 characters can be entered
for the month in the date format,
e.g YYYY-MM-DD

- Confirm any change with Apply.


- Click OK

Fig. 2

• Next, go to the Languages tab card.

SOMATOM Spirit CT02-025.816.03 Page 2 of 8 Siemens AG


Rev. 01 11.05 CSPS22 Medical Solutions
Special Language Settings 5-3

• Select Chinese as the "Language used in


menus and dialogs".
• Click Details....

Fig. 3

• Set the default input language to English


• Select the "Advanced" card and turn off
the "advanced test services" combo box
• "Apply" these changes and exit from the
options panel

Fig. 4

5.3.3 Settings under administrator


• Log in as administrator
- Select System -> End -> Restart Application
- Press Shift until the login window appears and log in as administrator
• Go to Start-> Settings -> Control Panel -> Regional and language options

Siemens AG CT02-025.816.03 Page 3 of 8 SOMATOM Spirit


Medical Solutions Rev. 01 11.05 CSPS22
5-4 Special Language Settings

• Select the "Advanced" card


• Under "Language for non unicode
programs" select "Chinese (PRC)"
• Apply and click OK to exit

Fig. 5

SOMATOM Spirit CT02-025.816.03 Page 4 of 8 Siemens AG


Rev. 01 11.05 CSPS22 Medical Solutions
Special Language Settings 5-5
5.4 Japanese
5.4.1 Settings in Local Service
• In the main menu of the service software, select the Configuration function.
• Go to DICOM -> Character Set
• Set allow only one character set
• Select ISO IR_100 under Default Repertoire
• To save the settings, click Save.

Fig. 6

• Click Finish and Home.


¹ Starting Report generation .. , wait until Ready is displayed.
• Do not accept the reboot; click Cancel.

5.4.2 Settings under meduser

NOTE In some cases, the dialog window disappears. Do not try to


restart the "Regional Language Options" dialog . You can move
the dialog into foreground by pressing Alt + TAB.

• From the syngo user interface, select -> Options -> Configuration.

Siemens AG CT02-025.816.03 Page 5 of 8 SOMATOM Spirit


Medical Solutions Rev. 01 11.05 CSPS22
5-6 Special Language Settings
• In the configuration panel, click the Regional and Language Options icon.
Note: The window disappears. Click with the left mouse button to move the window into
foreground.
• In the Regional Options tab card, select Japanese
• Under Location, select the country where you are.
• Click Apply.
• If necessary, press Alt + Tab to move the dialog window into foreground.

• To modify the appearance of the date,


time or numbers, click Customize.

Note: change the formats according to


the notes you made before the update.
There is only one restriction: A
maximum of 2 characters can be entered
for the month in the date format,
e.g YYYY-MM-DD

- Confirm any change with Apply.


- Click OK

Fig. 7

• Next, go to the Languages tab card.

SOMATOM Spirit CT02-025.816.03 Page 6 of 8 Siemens AG


Rev. 01 11.05 CSPS22 Medical Solutions
Special Language Settings 5-7

• Select Japanese as the "Language used


in menus and dialogs".
• Click Details....

Fig. 8

• Set the default input language to English


• select advanced card and turn off the
"advanced test services" combo box.
• "Apply" these changes and exit from the
options panel

Fig. 9

5.4.3 Settings under administrator


• Log in as administrator
- Select System -> End -> Restart Application
- Press Shift until the login window appears and log in as administrator
• Go to Start-> Settings -> Control Panel -> Regional and language options

Siemens AG CT02-025.816.03 Page 7 of 8 SOMATOM Spirit


Medical Solutions Rev. 01 11.05 CSPS22
5-8 Special Language Settings

• Select the "Advanced" Card


• Under "Language for non unicode
programs" select "Japanese"
• Apply and click OK to exit

Fig. 10

SOMATOM Spirit CT02-025.816.03 Page 8 of 8 Siemens AG


Rev. 01 11.05 CSPS22 Medical Solutions
Changes to previous version 6 6-1
No changes; initial publication.

Siemens AG Page 1 of 2 CT02-025.816.03 SOMATOM Spirit


Medical Solutions CSPS22 Rev. 01 11.05
6-2 Changes to previous version
This page intentionally left blank.

SOMATOM Spirit CT02-025.816.03 Page 2 of 2 Siemens AG


Rev. 01 11.05 CSPS22 Medical Solutions