6.1 Introduction
The 9354A/T internal calibration use routines to confirm basic functionality, no test
equipment is required to do these test procedures. The diagnostics menu is entered by
simultaneously depressing the third and fourth menu push buttons on the right hand
side of the CRT and then by depressing the fifth.
To quickly check the performance of the digital storage oscilloscope, select the
calibration diagnostics.
Press the recalibrate completely button to do a full recalibration of the front end.
It is advisable to perform this type of check when the scope is in a stable condition.
Page 6-1
Section 6 Internal Calibration and Diagnostics
The gain and offset calibration results displayed for Channel 1, Channel 2, Channel 3
and Channel 4 are independent of the following conditions:
Time base
50 Ω or 1 MΩ input impedance
BWL on or off
Variable gain
Offset
Trigger mode and coupling
The internal calibration is checked at DC 1 MΩ, and for the six gain settings :
2 mV, 5 mV, 10 mV, 20 mV, 50 mV, .1 V/div.