Professional Documents
Culture Documents
Under the action of oil pressure, a round diaphragm is pushed to discharge blood
from the artificial heart and then pulled back to draw blood into the heart.
Related terms:
Hydraulics, Turbine, Pistons, Hydrogen, Actuators, Lube Oil, Steam Generator, Oil
Pump, Oil Temperature, Shutdown
Automotive Instrumentation
William B. Ribbens, in Understanding Automotive Electronics (Eighth Edition), 2017
The simplest oil pressure warning system involves a spring-loaded switch connected
to a diaphragm. The switch assembly is mounted in one of the oil passageways such
that the diaphragm is exposed directly to the oil pressure. The force developed on
the diaphragm by the oil pressure is sufficient to overcome the spring and to hold
the switch open as long as the oil pressure exceeds the lower limit. Whenever the
oil pressure falls below this limit, the spring force is sufficient to close the switch.
Switch closure is used to switch on the low-oil-pressure warning message lamp.
One of the deficiencies of this simple switch-based oil pressure warning system is
that it has a single fixed low-oil-pressure limit. In fact, the threshold oil pressure for
safe operation varies with engine load. Whereas a relatively low oil pressure can pro-
tect bearing surfaces at low loads (e.g., at idle), a proportionately higher-oil-pressure
threshold is required with increasing load (i.e., increasing horsepower and RPM).
During the appropriate measurement time slot, the oil pressure sensor voltage is
sampled through the MUX switch and converted to binary numbers in the A/D
converter. The computer reads this binary number and compares it with the binary
number in memory for the allowed oil pressure limits. The oil pressure limit is
determined from load or crankshaft speed measurements that are already available
in the engine control system. These measurement data can be sent to the instrument
subsystem via a MUX system as described with respect to Fig. 8.5 and over an IVN.
These measurements serve as the address for a ROM lookup table to find the oil
pressure limit. If the oil pressure is below the allowed lower limit or above the allowed
upper limit, an output signal is generated that activates the oil pressure warning
light through the DEMUX (see Fig. 8.7).
It is also possible to use a proportional display of actual oil pressure. A digital display
can be driven directly from the computer. An analog display, such as a galvanometer,
requires a D/A converter.
1.5.7 Pressure
• Note: If safe entry to a newly started cell can be managed by a trained operator,
in addition to a quick visual inspection, their hearing and sense of smell can
often pick up insipient faults before the control system.
Oil and gas PVT correlations can usually be grouped into correlation families.
Researchers of PVT correlations usually develop a separate correlation for each PVT
property from their PVT laboratory-measured data. For selection of correlation(s) to
use for a specific fluid, the following approaches are available: (1) use of geographical
location, (2) use of the type or obvious characteristics of the fluid, (3) use of corre-
lations based on application, (4) use of data ranges and ranking, and (5) use of an
expert system or set of rules.
Al-Zahaby et al. (2012) expanded Al-Shammasi’s work and created clusters of input
data. For each cluster, they found the best correlation(s) to use among a reasonable
number of oil PVT correlations out of all oil PVT properties. They also collected
the ranges of applicability of the correlations recommended in their work. Their
approach is based on two steps: (1) the selection of a subset of correlations for every
PVT property by use of the input data and applicability ranges for each correlation
and (2) the ranking of PVT correlations within the cluster by use of a PVT data set. A
limited data set was used in this work.
The applicability ranges for gas and oil correlations can be used to reduce the
number of applicable correlations for a particular fluid. These ranges are based on
the database used to develop each correlation. The ranges are given in Appendix C,
Oil Correlations Range of Applicability and Appendix D, Gas Correlations Range of
Applicability, for oil and gas correlations, respectively.
Correlation Selection Based on Expert System or Set of Rules
Al-Gathe (2015) developed an expert system (a set of rules) for oil PVT correlation
selection. The study checked the accuracy of correlations against a large database of
PVT laboratory-measured data. From the input data for each correlation, the correla-
tions are ranked according to their accuracy. In these calculations, intermediate PVT
properties required as inputs to the following correlation calculations are assumed
to be accurate. Fig. 9.2 shows the input data required to compute each PVT property.
The usual four inputs are required in the selection process for all PVT properties. The
workflow starts with selection of the bubble point pressure correlation. The value of
the bubble point pressure is needed in the selection of the other PVT correlations
such as solution GOR, saturated oil formation volume factor, saturated oil density,
undersaturated oil compressibility, and saturated oil viscosity correlations. The value
of dead oil viscosity is also required in selection of both saturated and undersaturat-
ed oil viscosity correlations. In absence of laboratory-measured PVT data, dead oil
viscosity must be estimated from correlations. In ranking correlations for saturated
and undersaturated oil viscosity, Al-Gathe (2015) assumed that dead oil viscosity is
known. Nevertheless, testing this approach on a large data set (which was not used
in development of the expert system) demonstrated the accuracy of this approach in
the majority of cases. This result means that the expert system is capable of selecting
one (sometimes two) correlation(s) for every PVT property that provides the values
closest to laboratory-measured values.
Figure 9.2. Flow chart showing inputs needed for each correlation in PVT correlation
selection.
The horn is silenced by putting the system alarms switch on the engine room panel
to cut-out. The warning lights will continue to glow until the lube oil pressure is
restored to a level above 12 lb/in2 and when this happens the horn will sound again
until the system alarms switch is put to normal.
Reservoir Fluids
James G. Speight PhD, DSc, in Introduction to Enhanced Recovery Methods for
Heavy Oil and Tar Sands (Second Edition), 2016
The gas–oil ratio influences wax deposition in a manner that depends on the
pressure—above the bubble point, where all gases remain in solution, solution gas
helps to keep wax in solution (Luo et al., 2001; Singh et al., 2004). A high gas–oil
ratio would result in more expansion and subsequent cooling as pressure of the oil
system depletes—a situation that can increase wax deposition.
In terms of flow rate, wax deposition is influenced more by laminar flow than when
flow is in the turbulent regime and increasing the flow rate from laminar to turbulent
reduces maximum deposition rate and at the same time lowers the temperature at
which maximum deposition rate occurs (Hsu et al., 1994). Low flow rates offer the
moving oil an increased residence time in the flow channel which allows more heat
loss to the surroundings, leading to a higher chance of the bulk oil temperature
falling below the WAT and sufficient time for wax separation and deposition. In
addition, there is a difference in texture between wax deposited at high flow rates and
wax deposited at low flow rates (Jessen and Howell, 1958; Tronov, 1969). Paraffin wax
deposited at high flow rates appears harder, being more compact and more firmly
attached to the deposition surface.
The gas-oil ratio influences wax deposition in a manner that depends on the pres-
sure—above the bubble point, where all gases remain in solution, solution gas helps
to keep wax in solution (Luo et al., 2001; Singh et al., 2004). A high gas-oil ratio
would result in more expansion and subsequent cooling as pressure of the oil system
depletes—a situation that can increase wax deposition.
The low lubricating-oil pressure trip is sometimes implemented in this manner (Fig
2.45). Lubricating oil under pressure is supplied to a single-acting spring-loaded
trip cylinder. The oil is supplied via a deadweight accumulator common to front and
rear channels. This smooths out fluctuations in oil pressure and, in the event of a
sudden loss of oil pressure, will maintain the pressure to the trip cylinder for several
seconds, thus allowing possible recovery by starting of the emergency pumps.
FIG. 2.45. Mechanically-operated trips (front tripping system only)
Similarly, a direct high exhaust-pressure trip is often implemented using the same
spring-loaded trip cylinder as shown in the upper part of Fig 2.45. For the front
tripping system, there are two sets of detectors labelled 1 and 2. Lubricating oil is
supplied to the detectors via an orifice plate. The downstream pressure is controlled
by a leak-off valve, actuated by a bellows unit. The top half of the bellows is evacuated
and sealed at a reference pressure close to zero mbar absolute. The lower half
of the bellows is connected to the condenser. The leak-off valve is normally fully
seated, since the force supplied by spring A exceeds that exerted by the condenser
pressure. As the condenser pressure rises, a point is reached where the spring force is
counter-balanced and a further pressure increase causes the leak-off valve to unseat.
As lubricating oil pressure is lost within the vacuum trip unit, plunger B is moved
downwards by spring force and the pressure in the short length of pipe C is lost.
If detector 2 also trips, lubricating oil supplying the front trip cylinder is connected
to drain, thus releasing the trip latch. A similar pair of exhaust pressure detectors
operate into the trip cylinder of the rear system. On-load testing is enabled by a
mechanical interlock on the trip test lever, so that when the front protection system
is isolated, its two associated exhaust pressure detectors can be tested in turn by
slowly raising the pressure by a needle valve admitting atmospheric air. This permits
resetting of the trip point if desired.