Professional Documents
Culture Documents
Service Manual
IVD Global Technical Support Dept
Preface
© 2017 Shenzhen Mindray Bio-Medical Electronics Co., Ltd. All rights Reserved.
For this Service Manual, the issue date is 2017-09.
This manual contains the instructions necessary to operate the product safely and in accordance with its function
and intended use. Please read this manual thoroughly before using the product. Observance of this manual is
a prerequisite for proper performance and correct operation, and it ensures patient and operator safety. All
graphics including screens and printouts in this manual are for illustration purpose only and must not be used
for any other purposes. The screens and printouts on the actual product should prevail.
This manual is used for models: BS-360E.
Manual introduction
Intellectual Property Statement
SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD. (hereinafter called Mindray) owns the
intellectual property rights to this Mindray product and this manual. This manual may refer to information
protected by copyright or patents and does not convey any license under the patent rights or copyright of
Mindray, or of others.
, , , , , , , ,RealTF,
TrackWB,TrueTCR,Q-pick,AutoOLC,iVision,DBF,DRF,RDA,DRA,DFS、SyncNavi、GQ-Ana、One-
touchIP、Holo-IS、Opt-VRA、SuperVE-Cine、NFP-DSC、iTouch、iStation、BeneView、SmarTemp are the
trademarks, registered or otherwise, of Mindray in China and other countries. All other trademarks that appear
in this manual are used only for informational or editorial purposes. They are the property of their respective
owners.
Statement
All information contained in this manual is believed to be correct. Mindray shall not be liable for errors contained
herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this
manual.
Mindray is responsible for the effects on safety, reliability and performance of this product, only if:
All installation operations, expansions, changes, modifications and repairs of this product are
conducted by Mindray authorized personnel.
All repairs involving replacement parts and associated accessories and consumables are original
(original) or approved by Mindray。
The electrical installation of the relevant room complies with the applicable national and local
requirements.
The product is used in accordance with the instructions for use.
Warning
It is important for the hospital or organization that employs this equipment to carry out a reasonable
service/maintenance plan. Neglect of this may result in machine breakdown or personal injury.
Note
This equipment must be operated by skilled/trained clinical professionals.
Warranty
THIS WARRANTY IS EXCLUSIVE AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR
IMPLIED, INCLUDING WARRANTIES OF MERCHANTABILITY OR FITNESS FOR ANY PARTICULAR
PURPOSE.
No.HSH-19007-BS-360E Version:1.0 i
IVD Global Technical Support Dept
Exemptions
Mindray's obligation or liability under this warranty does not include any transportation or other charges or
liability for direct, indirect or consequential damages or delay resulting from the improper use or application of
the product or the use of parts or accessories not approved by Mindray or repairs by people other than Mindray
authorized personnel.
This warranty shall not extend to:
Malfunction or damage caused by improper use or man-made failure.
Malfunction or damage caused by unstable or out-of-range power input.
Malfunction or damage caused by force majeure such as fire and earthquake.
Malfunction or damage caused by improper operation or repair by unqualified or unauthorized service
people.
Malfunction of the instrument or part whose serial number is not legible enough.
Others not caused by instrument or part itself.
Customer service department
Manufacturer: Shenzhen Mindray Bio-Medical Electronics Co., Ltd.
Address: Mindray Building, Keji 12th Road South, High-tech industrial
park, Nanshan, Shenzhen 518057,P.R.China
Website: www.mindray.com
E-mail Address: service@mindray.com
Tel: +86 755 81888998
Fax: +86 755 26582680
EC - Representative
EC-Representative: Shanghai International Holding Corp. GmbH(Europe)
Address: Eiffestraβe 80, 20537 Hamburg, Germany
Tel: 0049-40-2513175
Fax: 0049-40-255726
Product introduction
The instrument is a computer-controlled fully-automated chemistry analyzer, intended for quantitative
determination of clinical chemistries in serum, plasma, urine, cerebrospinal fluid (CSF), and other human body
fluids. It can fulfill auto dispensing, reaction, colorimetric measurement, process monitoring, and result
calculation. It is one of the necessary tools for laboratory automation.
Safety information
This chapter provides you with safety symbols used in this manual and their meanings, summarizes the safety
hazards and operating precautions that should be considered seriously when the instrument is being operated,
and lists the labels and silkscreens that have been applied to the instrument and their indications.
Safety symbols
Safety symbols are used in this manual in order to remind you of the instructions necessary to operate the
product safely and in accordance with its function and intended use. A safety symbol and text constitutes a
warning as shown in the table below:
Symbol Meaning
Caution
Biological risks
Summary of hazards
This section lists hazards of the instrument itself. The hazards of specific operation are included in the warning
information of each operation task.
Observe the following safety precautions when using the product. Ignoring any of them may lead to personal
injury or equipment damage.
No.HSH-19007-BS-360E Version:1.0 ii
IVD Global Technical Support Dept
WARNING
If the product is used in a manner not specified by our company, the protection provided by the product may be
impaired.
WARNING
When the MAIN POWER is turned on, users other than the servicing personnel authorized by our company
must not open the rear cover or side cover.
Spillage of reagent or sample on the product may cause equipment failure and even electric shock. Do not place
sample and reagent on the product. In case of spillage, Switch off the power immediately, remove the spillage
and contact our Customer Service Department or your local distributor.
WARNING
Do not touch such moving parts as sample/reagent carousel, reaction carousel, probe, mixer, and cuvette wash
station, when the system is in operation.
Exercise caution while using the ISE module Prevent your hair, legs or other parts of your body from being hurt
by the driving parts.
Do not put your fingers or hands into any open part when the system is in operation.
WARNING
Eye injury could occur from light emission from the photometer lamp. Do not stare into the lamp when the system
is in operation.
If you want to replace the photometer lamp, first switch off the MAIN POWER and then wait at least 5 minutes
for the lamp to cool down before touching it. Do not touch the lamp before it cools down, or you may get burned.
BIOHAZARD
Inappropriately handling samples, controls and calibrators may lead to biohazardous infection. Do not touch
samples, controls, calibrators, mixtures, or waste with your bare hands. Wear gloves and lab coat and, if
necessary, goggles.
In case your skin contacts the sample, control or calibrator, follow the standard laboratory safety procedure and
consult a doctor.
The serum samples remaining in the electrodes may contain a great number of viruses. Wear gloves to prevent
infection while operating around the electrodes.
WARNING
Reagents, diluted wash solution and concentrated wash solution are corrosive to human skins. Exercise caution
when using reagents and concentrated wash solution. In case your skin or clothes contact them, wash them off
with soap and clean water. If reagents or wash solution spills into your eyes, rinse them with much water and
consult an oculist.
Waste hazards
BIOHAZARD
Some substances contained in reagent, control, calibrator, concentrated wash solution, and waste are subject
to regulations of contamination and disposal. Dispose of the waste in accordance with your local or national rule
for biohazard waste disposal and consult the manufacturer or distributor of the reagents for details.
Wear gloves and lab coat and, if necessary, goggles.
WARNING
Materials of the analyzer are subject to contamination regulations. Dispose of the waste analyzer in accordance
with your local or national rule for waste disposal.
WARNING
Ethanol is flammable substance. Please exercise caution while using ethanol around the instrument in order to
prevent fire and explosion.
WARNING
When the analyzer is not in use, for example, in repair, transportation or disposal process, please clean and
sterilize the parts that may cause biohazards(probe, mixer, etc.) and remind the person who handles the device
of the related hazards.
CAUTION
Appropriate decontamination is carried out if hazardous material is spilled onto or into the equipment.
No decontamination or cleaning agents are used which could cause a HAZARD as a result of a reaction with
parts of the equipment or with material contained in it. Strong acid or alkaline solutions are forbidden to clean
the equipment.
If there is any doubt about the compatibility of the decontamination or cleaning agents with parts of the
equipment or with material contained in it, please contact our customer service department or the local
distributor.
Summary of precautions
This section lists precautions to be understood during instrument operation. The precautions of specific
operation are included in the warning information of each operation task.
To use the product safely and efficiently, pay attention to the following operating precautions.
Intended use
WARNING
The instrument is an automated chemistry analyzer for in vitro diagnostic use in clinical laboratories and
designed for in vitro quantitative determination of clinical chemistries in serum, plasma, urine and cerebrospinal
fluid samples.
Please consult us before you use the instrument for other purposes.
No.HSH-19007-BS-360E Version:1.0 IV
IVD Global Technical Support Dept
When drawing a clinical conclusion, please also refer to patients' clinical symptoms and other test results.
Installation Precautions
NOTE
The safety of any system incorporating the equipment is the responsibility of the assembler of the system.
Environment precautions
CAUTION
Please install and operate the system in an environment specified by this manual. Installing and operating the
system in other environment may lead to unreliable results and even equipment damage.
To relocate the system, please contact our Customer Service Department or your local distributor.
CAUTION
Electromagnetic noise may interfere with operations of the system. Do not install devices generating excessive
electromagnetic noise around the system. Do not use such devices as radio transmitters in the room housing
the system. Do not use other CRT displays around the system.
Do not use other medical instruments around the system that may generate electromagnetic noise to interfere
with their operations.
Do not use this device in close proximity to sources of strong electromagnetic radiation (e.g. mobile phones or
radio transmitters), as these may interfere with the proper operation.
The electromagnetic environment should be evaluated prior to operation of the device.
This device has been designed and tested to CISPR 11 Class A, and in a domestic environment may cause
radio interference, in which case, you may need to take measures to mitigate the interference.
NOTE
It is the manufacturer's responsibility to provide equipment electromagnetic compatibility information to the
customer or user.
It is the user's responsibility to ensure that a compatible electromagnetic environment for the equipment can be
maintained in order that it will perform as intended.
Operating precautions
CAUTION
Take the clinical symptoms or other test results of the patient into considerations when making diagnosis based
on the measuring results produced by the system.
Operate the system strictly as instructed by this manual. Inappropriate use of the system may lead to unreliable
test results or even equipment damage or personal injury.
When using the system for the first time, run calibrations and QC tests to make sure the system is in proper
state.
Be sure to run QC tests every time when you use the system, otherwise the result may be unreliable.
Start the operating software again when the analyzing unit is powered off.
Do not uncover the reagent carousel when the system is in operation. Keep the reagent carousel cover closed.
The RS-232 port on the analyzing unit is used for connection with the operation unit only. Do not use it for other
connections. Use the cables provided by our company or your local distributor for the connection.
No.HSH-19007-BS-360E Version:1.0 V
IVD Global Technical Support Dept
The operation unit is a personal computer with the operating software installed. Installing other software or
hardware on the computer may interfere with the system operation. Do not run other software when the system
is working.
Computer virus may destroy the operating software or test data. Do not use the computer for other purposes or
connect it to the Internet. If the computer is infected by virus, please install anti-virus software to check for and
clear virus.
Do not touch the display, mouse or keyboard with wet hands or hands with chemicals.
Do not place the MAIN POWER to ON again within 10 seconds after placing it to OFF; otherwise the system
may enter the protection status. If it does so, place the MAIN POWER to OFF and place it to ON again.
CAUTION
To define such parameters as sample volume, reagent volume and wavelength, follow the instructions in this
manual and the instructions of reagents.
CAUTION
To prevent ISE electrodes from being damaged due to water scarcity, if the system, when equipped with an ISE
module will be powered off for a long time, perform the electrode storage maintenance.
Sample precautions
CAUTION
Use samples that are completely free of insoluble substances like fibrin or suspended matter; otherwise the
sample probe may be clogged.
Medicines, anticoagulants or preservative in the samples may lead to unreliable results.
Hemolysis, icterus or lipemia in the samples may lead to unreliable test results; running a serum index test
therefore, is recommended.
Store the samples properly. Improper storage may change the compositions of samples and lead to unreliable
results.
Sample volatilization may lead to unreliable results. Do not leave the sample open for a long period.
Prepare sufficient sample volume before analysis.
Load samples to correct positions on the sample carousel before the analysis begins; otherwise reliable results
may not be obtained.
CAUTION
Use proper reagents, calibrators and controls on the system.
Select appropriate reagents according to the performance characteristics of the system. Consult the reagent
suppliers, our company or our authorized distributor for details, if you are not sure about your reagent choice.
Store and use the reagents, calibrators and controls strictly as instructed by the suppliers; otherwise, reliable
results or best performance of the system may not be obtained. Improper storage of reagents, calibrators and
controls may lead to unreliable results and bad performance of the system even in validity period.
Perform calibration after changing the reagents, otherwise reliable results may not be obtained.
Contamination caused by carryover among reagents may lead to unreliable test results. Consult the reagent
suppliers for details.
No.HSH-19007-BS-360E Version:1.0 VI
IVD Global Technical Support Dept
BIOHAZARD
The calibrators contain preservatives. In case your skin contacts calibrators, wash them off with soap and water.
In case the calibrators spill into your eyes, rinse them with water and consult an oculist. If you swallow them by
mistake, see a doctor.
CAUTION
Use the calibrators specified by our company. Use of other reagents or calibrators may result in unreliable
results, or damage the Hydropneumatic system, or even shorten the electrodes life span.
Prior to using the calibrators, check if they are within the expiration date.
Place them correctly; otherwise, it may cause unreliable results, or leak, or module damage.
BIOHAZARD
The ISE wash solution is sodium hypochlorite. Use the ISE wash solution carefully to prevent it from contacting
your skins or eyes. If your skins or eyes contact the ISE wash solution, rinse them off with fresh water and
consult a doctor.
NOTE
The system automatically stores the data to the built-in hard disk. Data loss, however, is still possible due to
mis-deletion or physical damage of the hard disk. You are recommended to regularly archive the data to such
medium as CDs.
To avoid the data loss caused by unexpected power failure, UPS (uninterrupted power supply) is recommended.
WARNING
For operating instructions and precautions of the computer and printer, please refer to their operation manuals.
External equipment connected to the analogue and digital interfaces must be authorized and complied with
relevant safety and EMC standards (e.g., IEC 60950 Safety of Information Technology Equipment Standard
and CISPR 22 EMC of Information Technology Equipment Standard (CLASS B)). Any person, who connects
additional equipment to the signal input or output ports and configures an IVD system, is responsible for
ensuring that the system works normally and complies with the safety and EMC requirements. If you have any
questions, consult the technical services department of your local representative.
CAUTION
Maintain the system strictly as instructed by this manual. Inappropriate maintenance may lead to unreliable
results, equipment damage or personal injury.
To wipe off dust from the system surface, use a soft, clean and wet (not too wet) cloth soaked with soap water
rather than organic solvents such as ethanol. After cleaning, wipe the surface dry with dry cloth.
Switch off all the powers and disconnect the power plug before cleaning. Take necessary measures to prevent
water ingression, otherwise equipment damage or personal injury may be caused.
Replacement of such major parts as photometer lamp, probe, mixer and syringe assembly must be followed by
a calibration.
Replacement of the photometer lamp should be done when the system power has been switched off for at least
5 minutes.
If the system fails and needs servicing, contact our Customer Service Department or the local distributor. The
system may need to be stopped or transported during servicing, which will probably cause biohazards, electric
shock hazards and moving part hazards. Exercise caution when prepare the system for servicing.
Wear gloves and lab coat and, if necessary, goggles.
NOTE
Check the safe state of the equipment after repair. Make sure the equipment is safe and then offer it to the
customer.
WARNING
When the tube or the part that contain liquid become aged or damaged, please stop its use immediately and
contact our customer service department or your local distributor to check and replace it.
For the label marked with , please consult the related documentations in order to find out the nature of the
potential HAZARDS and any actions which have to be taken to avoid them.
Check the labels regularly for cleanliness and integrity. If any of the labels becomes vague or peels off, contact
our Customer Service Department or your local distributor for replacement.
Labels and silkscreen list
Symbol Meaning
Serial Number
Date of Manufacture
Manufacturer
CE marking
Symbol Meaning
In Vitro diagnostic medical device
Biological risks
Caution
“ON” (Power)
“OFF” (Power)
Serial interface
No.HSH-19007-BS-360E Version:1.0 IX
IVD Global Technical Support Dept
Humidity range
This symbol indicates the humidity requirement under normal working condition.
Atmospheric pressure
This symbol indicates the atmospheric pressure requirement under normal working condition.
Temperature limit
This symbol indicates the temperature requirement under normal working condition.
Warning labels
No.HSH-19007-BS-360E Version:1.0 X
IVD Global Technical Support Dept
Biohazard warning
This label indicating the risk of biohazardous infection is located in the following positions:
Probe
Waste outlet
Waste tank
ISE module
This symbol and text located on the left side panel of the analyzer. Please turn off the main power before opening
the small door.
No.HSH-19007-BS-360E Version:1.0 XI
IVD Global Technical Support Dept
Chemical hazard
This label is applied on the diluted wash solution tank. Please take protective measures to prevent chemical
hazard.
Cabinet
This label is applied at the middle of the cabinet back. If you want to relocate your analyzer, contact our
Customer Service Department or your local distributor. When relocating it, do not carry it with the cabinet. Only
when the cabinet is supported by the anchors, can the analyzer be placed on the cabinet.
User Permissions
Engineer username: serviceuser
Password: #BS8A#SEU
Note: After logging into the analyzer using the engineer on a client, remember to switch back to the
account of engineer.
Revision history
Chapter History Version
1.Added revision history
Preface V1.0
2. Added description about user permission
Chapter 1 1. Rearranged V1.0
1.Rearranged
Chapter 2 V1.0
2. Changed FRU
1. Rearranged
2. Changed FRU
Chapter 3 V1.0
3. Changed wrong description
4. Aded ISE
1. Rearranged
2. Changed FRU
Chapter 4 V1.0
3. Changed PCB picture
4. Added 4.5.4
Chapter 5 1. Rearranged V1.0
1. Rearranged
Chapter 6 V1.0
2. Added 6.2
3. Added 6.3
4. Added 6.6.2
5. Added 6.7
6. Added6.8
1. Rearranged
Chapter 7 2. Changed picture. V1.0
3. Added7.9
1. Rearranged
Chapter 8 2. Added 8.1 V1.0
3. Added 8.6-8.8
Chapter 9 1. Rearranged V1.0
1. Rearranged
Chapter 10 2. Added 10.1~10.3 V1.0
3. Changed 10.4
Chapter 11 1. Added chapter V1.0
Chapter 12 1. Added chapter V1.0
Chapter 13 1. Added chapter V1.0
Chapter 14 1. Added chapter V1.0
Added :
1. Installation Acceptance Report
2. Moving Parts Position Confirmation Guide (No Alignment
Tooling)
Appendices V1.0
3. Tool list
4. Error Information Feedback Form
5. Series Recovery Checklist
6. Maintenance log sheet
Table of Contents
Preface ............................................................................................................................................................ i
Manual introduction................................................................................................................................... i
Intellectual Property Statement .......................................................................................................... i
Statement .......................................................................................................................................... i
Warranty ............................................................................................................................................ i
Exemptions ....................................................................................................................................... ii
Customer service department ........................................................................................................... ii
EC - Representative.......................................................................................................................... ii
Product introduction ......................................................................................................................... II
Safety information ............................................................................................................................ II
Summary of hazards ........................................................................................................................ II
Summary of precautions ................................................................................................................. IV
Labels and silkscreen ................................................................................................................... VIII
Instrument Labels and Silkscreen.................................................................................................... IX
User Permissions .................................................................................................................................. XII
Revision history .................................................................................................................................... XII
Table of Contents ........................................................................................................................................ XIV
1 System Description.............................................................................................................................. 1-1
1.1 Overview ................................................................................................................... 1-1
1.2 Components of Analyzing Unit ................................................................................... 1-1
1.3 Functions of Analyzing Unit ........................................................................................ 1-2
2 System Performance and Workflow ..................................................................................................... 2-1
2.1 Major Specifications ................................................................................................... 2-1
2.1.1 General Specifications ......................................................................................................... 2-1
2.2 Sample Specifications ................................................................................................ 2-1
2.2.1 Reagent Specifications ......................................................................................................... 2-3
2.2.2 Specifications of Reaction System ........................................................................................ 2-3
2.2.3 Specifications of Operation Unit............................................................................................ 2-4
2.2.4 Typical Test Procedure ......................................................................................................... 2-4
2.2.5 Startup Procedure ................................................................................................................ 2-5
2.2.6 Shutdown Procedure ............................................................................................................ 2-5
2.2.7 Workflow Descriptions .......................................................................................................... 2-5
2.2.8 Measuring Points ................................................................................................................. 2-6
3 Moudules and Units ............................................................................................................................. 3-1
3.1 Shell Assembly .......................................................................................................... 3-1
3.1.1 Module Functions ................................................................................................................. 3-1
3.1.2 Component Locations and FRU Details ................................................................................ 3-1
3.1.3 Removing and Reinstalling Shielding Cover ......................................................................... 3-2
3.1.4 Removing and Reinstalling Left and Right Panels ................................................................. 3-3
3.1.5 Removing and Reinstalling Front Panel ................................................................................ 3-3
3.1.6 Removing and Reinstalling Desk Panels .............................................................................. 3-3
3.1.7 Replacing Air Spring ............................................................................................................. 3-4
3.1.8 Remove the dust from the fans ............................................................................................. 3-4
3.2 Frame Assembly ........................................................................................................ 3-4
3.2.1 Module Functions ................................................................................................................. 3-4
3.2.2 Component Locations and FRU Details ................................................................................ 3-5
3.2.3 Replacing Dust Screens ....................................................................................................... 3-5
3.3 Reaction Carousel Assembly...................................................................................... 3-5
3.3.1 Main functions ...................................................................................................................... 3-5
No.HSH-19007-BS-360E Version:1.0 XV
IVD Global Technical Support Dept
No.HSH-19007-BS-360E Version:1.0 XX
IVD Global Technical Support Dept
1 System Description
1.1 Overview
The BS-360E is a fully automated and computer-controlled chemistry analyzer designed for the in vitro
determination of clinical chemistries in serum, plasma, urine, and cerebrospinal fluid (CSF) samples. This
product consists of Analyzing Unit and Operation Unit.
Analyzing
unit
PC(Operating software)
Main unit
Printer
The instrument is composed of the following components: one reaction carousel, one sample/reagent carousel,
one probe, one mixer, ne automatic cuvette wash station, one photometric detection system, one ISE module
(optional). The optical measurement system, which is composed of gratings and diode array, performs
photometric measurement to the reaction cuvettes that hold "sample+reagent" mixture. After testing, the cuvette
wash station performs 8-phase auto wash to the reaction cuvettes.
(5)
(1)
(4)
(2)
(3)
Name Value
Symbology Codabar, ITF, code128, code39, UPC/EAN,
Code93
Minimum bar code density 0.25mm~0.5mm
Data bits 13-30
Format and content Defined by user
Maximum width 55mm
Minimum height 10mm
Maximum inclination angle ±5°
Print quality Class A (ANSI MH10.8M)
Width and narrowness 2.5:1
Reagent probe
One probe is shared for adding sample and reagent with liquid level sense, horizontal and vertical anti-collision,
level tracking and probe clogging detection functions.
Reagent probe cleaning
Both interior and exterior of the probe are washed.
Prevention of reagent cross contamination
Users are allowed to perform carryover settings, e.g. washing interior and exterior of reagent probes and
inserting special wash between tests.
2.2.2 Specifications of Reaction System
Table 2-5 Specifications of Reaction System
Optical pathlength of reaction 5mm
cuvette
Material of reaction cuvette 5mm*5mm*29.5mm. Plastic cuvette: Cuvette segments with
10 cuvettes each segment.
Number of reaction cuvettes 80
balance of
Mixer performs mixing
reaction
Shutdown Procedure
Available
Add sample to probe
carousel
Add R1 to probe
chemistries are
Wash Cuvette
Wash Cuvette
temperature finished
Single Reagent
Chemistries
Next test
Sequential actions of the probe in each period (dispensing whole blood sample)
Wash ->rotate to sample carousel->lower into the sample tube-> aspirate sample->raise to vertical home
position->Rotate to wash well->Lower to the position for special wash -> Special wash the exterior of the probe
-> raise to “home” position vertically -> rotate to reaction carousel -> dispense sample->Raise to vertical home
position->rotate to wash well -> wash -> lower to position for special wash ->special wash the exterior of the
probe->Raise to vertical home position -> wash ->go to next period.
T1 T11
Maximum reaction time of single reagent is over 9 min 50 seconds
7 6
8
9.10
5
11
4
12
3
13
14
1
15
17
16.18
21
22
19
20
Hinge assembly
Syringe shield Sleeve of wash tubes
Shielding
cover
Air spring
Middle
panel Auto wash desk
panel
ISE port plut Reaction desk
Sample/reagent panel
carousel cover
Sample/reagentde
sk panel Front panel
Screw plug
ISE cover
Left
panel
Handle
hole panel
Optical
assembly
Cuvette
Reaction carousel temperature
control 60 fan cable M2.5*8 cross countersunk
head screws
M4*16 cross pan head
Reaction carousel temperature
screw with washer
sensor cable
M4*8 cross pan head screw MX20 hexagon socket cap
with washer screws With spring and flat
washer
Bracket of
Fan Reaction
chamber
Silicone
heating film
Thermal Guard 60C
2Lead15.6mm
Reaction
carousel motor
Coded disk sensor
Plate
Tools
Name Code Quantity
Cross screwdriver / 1
Hexagon wrench / 1
Exploded view for installation
Reaction chamber
Press plate of
lower heater M4*8 cross pan head
combination screws
Supporting rod
Figure 3-6 Replacing the cable of the heat insulation chamber of reaction carousel
How to do
1) Switch off the main power of the whole unit.
2) Open the shielding cover. Unscrew the screw on the wash probes assembly with your hand and
remove the wash probes assembly. Take caution not to collide the wash probes. Rotate the probe to
the aspirating reagent position and rotate the mixer to the wash well position. Unscrew all the M4*8
cross pan head screw with washer on the desk panels to remove the reaction carousel desk panel,
auto wash desk panel. Remove the right panel.
3) Use the cross screw to unscrew the four M4*12 cross pan-head combination screws on the reaction
carousel cover. Disconnect the cable of the heater or temperature protection switch cable to remove
the reaction carousel cover.
4) Loosen the cuvette pressing plate and remove the cuvette segments and place them properly.
Unscrew the three M3*12 hexagon socket screws used to fix the reaction carousel body to remove it
as shown below.
M3*12 hexagon
socket cap head Reaction carousel
screw body
Figure 3-7 Removing Reagent carousel body assembly
5) If you want to replace the cable of the reaction carousel temperature sensor, unscrew the M2.5*8
cross countersunk head screws and disconnect connector of the cable of the reaction carousel
temperature sensor. Replace the cable with new one and connect it.
6) If you want to change the temperature control 60 fan cable of the reaction carousel, disconnect the
connector of the fan and unscrew the two M4*8 cross pan head combination screws. Remove the
fan together with its bracket. Replace new fan cable and connect it properly. Note: the side of the
fan with the label must face the bracket of the fan. The cable is extracted from the lower right corner.
7) If you want to replace the silicone heating film of the reaction carousel or Thermal Guard 60C
2Lead15.6mm, unscrew the three M5*10 hexagon socket cap head screws to remove the optical
measurement assembly. Unscrew the three M4*20 hexagon socket cap head screws with spring and
flat washers used to fix the reaction chamber. Disconnect the cables, the waste overflow tube and the
grounding wire to remove the reaction chamber.
8) If you want to replace Thermal Guard 60C 2Lead15.6mm, place the reaction chamber upside down
and unscrew the two M3*6 hexagon socket cap screws.
9) If you want to replace the silicone heating film of the reaction carousel, loosen the twelve M4*8 cross
pan head combination screws to remove the heater plate. Daub evenly a layer of thermal paste on the
silicone heating film and small amount of hot melt adhesive at the outlet of the cable to fix it.
10) Restore the components in the reversed order. Note: Align the heat insulation chamber with the drive
shaft of the reaction carousel.
Alignment and confirmation
After replacing silicone heating film, Thermal Guard 60C 2Lead15.6mm and the reaction carousel temperature
sensor cable, reconfigure the related sensor parameters on the software and perform the heat insulation
performance test. For sensor parameter configuration, see7.10.1 Configuration of Sensor Parameters; for
reaction carousel temperature test and reaction carousel temperature curve, see7.10.2 Observe Temperature
Curve.
If replacing the Optical measure assembly, refer to 7.3.2 Signal Collecting Position Adjustment to adjust
photoelectric signal collecting position; Refer to 7.3.3 Photoelectric Gain Adjustment to adjust photoelectric gain.
Mounting
plate
The sensor
installed on the
Sensor left of the
Hexagon socket cap Cross pan head
bracket bracket.
head screws M3*8 screw M3*6
Note:
Install the sensors according to the identifications on the cable connectors and the instrument connectors:
"RCD--PHO-C" for coder sensor and "RCD--PHO-O" for home position sensor.
Do not tighten the sensor screws with excessive torque force, in order to avoid damaging the sensors.
Make sure to install the home position sensor in the correct position.
How to do
Replacing coder sensor assembly
1) Switch off the main power of the whole unit.
2) Unscrew the three M4*8 cross pan-head combination screws on the right panel to remove it and then
unscrew the three M4*8 cross pan-head combination screws on the left panel to remove it. Unscrew
the four M4*8 cross pan-head combination screws on the front panel to remove it.
3) Unplug the connector of the reaction carousel coder sensor.
4) Unscrew the two M3*6 cross pan head combination screws on the sensor’s mounting plate and
remove the coder sensor cable with the sensor.
5) Loosen the two cross pan head screws on the coder sensor, and remove the coder sensor from the
mounting plate.
6) Fix the new coder sensor on the mounting plate using two M3×6 cross pan head screws.
7) Use two M3*6 cross pan head combination screws to fix the sensor mounting plate to the bracket of
the sensor. Adjust the height of the mounting plate of the sensor to keep the coder in the middle of the
coder sensor, and then tighten the retaining screws of the mounting plate.
8) Connect the connector of the reaction carousel coder sensor.
9) After aligning the sensor assembly, restore the components in the reversed order.
Replacing home position sensor assembly
1) Switch off the main power of the whole unit.
2) Unscrew the three M4*8 cross pan-head combination screws on the right panel to remove it and then
unscrew the three M4*8 cross pan-head combination screws on the left panel to remove it. Unscrew
the four M4*8 cross pan-head combination screws on the front panel to remove it.
3) Unplug the connectors of the reaction carousel home position sensor and coder sensor.
4) Loosen the three M3*8 hexagon socket cap head screws with spring and flat washers on the sensor
assembly, and then remove the sensor assembly. Pay attention not to drop the screws into the
instrument.
5) Unscrew the two M3*6 cross pan head screws on the cables of carousel home position sensor to
remove it from the bracket of the sensor.
6) Fix the new cables of carousel home position sensor onto the sensor bracket using two M3*6 cross
pan head screws.
7) Fix the sensor assembly of the reaction carousel onto the big bottom plate with three M3*8 hexagon
socket cap head screws with spring and flat washers. Manually rotate the reaction carousel and
check that the zero position stopper of the coded disk does not interfere with the zero position sensor.
8) Connect the home position sensor and the coded disk sensor.
9) After aligning the sensor assembly, restore the components in the reversed order.
Alignment and confirmation
1) Refer to 7.3.2 Signal Collecting Position Adjustment to adjust photoelectric signal collecting position.
2) Refer to 7.4 Sample/Reagent carousel unit to adjust reaction carousel circumferential position,
sample/reagent carousel circumferential position,
3) Refer to 7.5.1 Probe to Horizontal Position on Reaction Carousel and 7.6.1 Mixer to Horizontal Position
on Reaction Carousel and 7.7.1 Cuvette Wash Station Home to adjust probe and mixer horizontal
position and auto wash horizontal position.
3.3.6 Replacing motor assembly and synchronous belt
When to do
Replace the reaction carousel motor or synchronous belt when they are damaged.
Tools
Name Code Quantity
Cross screwdriver / 1
Hexagon wrench / 1
Exploded view for installation
Motor
M4*12hexagon
socket cap head
screws with spring
and flat washer
Bracket of
motor
M4*10 hexagon
socket cap head
screws with spring
and flat washer
Figure 3-9 Reaction carousel motor assembly
WARNING
During removing and installation, avoid dropping screws and washers into the analyzer.
How to do
Replacing the motor
1) Switch off the main power of the whole unit.
2) Unscrew the three M4*8 cross pan-head combination screws on the right panel to remove it and then
unscrew the three M4*8 cross pan-head combination screws on the left panel to remove it. Unscrew
the four M4*8 cross pan-head combination screws on the front panel to remove it.
3) Disconnect the motor cable, unscrew the 4 M4*10 hexagon socket head screws with spring washer
on the reaction carousel motor assembly, remove the motor assembly and avoid dropping screws into
the analyzer.
4) Fix the motor assembly to the large bottom plate using four M4*10 hexagon socket cap head screws
with spring washer. Tighten the belt and screws and connect the cable of the motor.
5) Restore the components in the reversed order.
Replacing the synchronous belt
1) Switch off the main power of the whole unit.
2) Open the shielding cover. Unscrew the screw on the wash probes assembly with your hand and
remove the wash probes assembly. Take caution not to collide the wash probes. Rotate the probe to
the aspirating reagent position and rotate the mixer to the wash well position. Unscrew all the M4*8
cross pan head screw with washer on the desk panels to remove the reaction carousel desk panel,
auto wash desk panel. Remove the left and right panel. Unscrew the four M4*8 cross pan-head
combination screws on the front panel to remove it.
3) Use the cross screw to unscrew the four M4*12 cross pan-head combination screws on the reaction
carousel cover. Disconnect the cable of the heater or temperature protection switch cable to remove
the reaction carousel cover.
4) Loosen the cuvette pressing plate and remove the cuvette segments and place them properly.
Unscrew the three M3*12 hexagon socket screws used to fix the reaction carousel body to remove it.
5) Unscrew the three M5*10 hexagon socket cap head screws to remove the optical measurement
assembly. Unscrew the three M3*12 hexagon socket cap head screws used to fix the reaction chamber.
Disconnect the cables to remove the reaction chamber.
6) Disconnect the motor cable and unscrew the four M4*10 hexagon socket head screws on the reaction
carousel motor to remove it.
7) Install a new synchronous belt.
8) Tighten the belt and screws and connect the cable of the motor.
9) Restore the components in the reversed order. Note: Align the heat insulation chamber with the drive
shaft of the reaction carousel.
Alignment and confirmation
The belt tension needs to be adjusted to ensure that the tension is the same as before the replacement.
3.3.7 Replacing Reaction Cuvette
Refer to 3.9.6 Replacing Reaction Cuvette.
3.4 Sample/Reagent carousel cover assembly
3.4.1 Module Functions
Sample/Reagent carousel assembly is located at the left front of the analyzer, including the carousel body,
refrigeration assembly, drive assembly, motor assembly, coded disk optical coupler and the reagent bar code
reader (Optional). It is used to complete various actions, such as carrying and rotating the reagent bottle or
sample tube to the specified position with sampling assembly. It provides a refrigerating environment, and the
reagents stored in such environment can be kept stable with little volatilization. For details, please refer to the
following contents.
Load sample/reagent. Place some reagent bottles with reagent and sample tubes with sample on the
carousel. Let the probe aspirate sample and reagent and dispense them into the cuvettes.
Refrigerating reagent: Provide 24-hour refrigeration and constant temperature environment (2-12°C)
to keep the reagent stable with little volatilization.
Sequential feeding: The drive assembly drives the carousel to carry certain reagent and sample in the
given order to the aspirate position for aspiration.
Input reagent and sample data automatically: Bar code reader (optional).
Reagent Temperature Sensor: It is a Negative temperature coefficient, the resistance decreases with
increasing temperature.
temperature Resistance
2°C About 11KΩ
25°C About 5KΩ
Sample/reagent carousel
cover
sample/reagent desk
panel
Bar code
assembly
Left panel
Sample/reagent
carousel
13
1
12
11
2
10
3
4
5
6
7
9
8
Figure 3-11 Sample/Reagent carousel assembly
Table 3-4 List of materials
No FRU Part Name Remarks
1 115-045517-00 Sample/Reagent carousel body
assembly
2 024-000110-00 Sensor temperature 5Kohm B3470K
with threads
3 049-000971-00 Square sealing ring
4 009-004605-00 Semi-conductive peltier (long cable)
5 049-001224-00 Square sealing ring
6 024-000920-00 Semi-conductive cooler(BA36)
7 009-002204-00 Correlative photocoupler (S)
8 115-045515-00 Sample/Reagent carousel motor
assembly
9 M6C-020007--- Synchronous cog belt, TBN180XL037
10 024-000899-00 FAN 12Vdc 92*92*38mm 110CFM
33.6dB 300mm
11 115-041761-00 Reagent carousel anti-fogging heater
assembly
12 023-001265-00 Bar Code Reader
13 BA20-20-75214 Tube holder 1
3.4.3 Replacing Sample/Reagent Carousel Body Assembly
When to do
When the sample/reagent carousel body is damaged.
Tools
Sample/reagent carousel
body assembly
M3*10 hexagon
socket cap head screw
with spring washer
Temperature sensor
cylinder
Temperature
sensor
Note
When using the sealing glue to plug up the mounting hole of the temperature sensor, make sure
sufficient glue is used to fill up the mounting hole. Wipe off the excessive glue.
Alignment and confirmation
1) Re-align Probe to outer ring, middle ring and inner ring positions,see 7.5.1 ~7.5.5 .
2) Re-align barcode scanning position,see 7.8.3 Bar Code Reader Position Adjustment.
3.4.5 Replacing cooler and sealing ring
When to do
Sample/reagent
carousel body
M3*10 hexagon socket
cap head screws with
Shaft sleeve spring and flat washers
Refrigeration board
Aluminum block
M3*6 hexagon socket M4*12cross pan head
cap head screws with combination screws
spring washer Heat insulation plate
Heat insulation plate
M4*12 cross pan head
combination screws
Square sealing ring
Square sealing ring
M4*12 cross pan head
combination screws Cooler
Cooler Radiator
Radiator
M4*12 cross pan head
M4*12 cross pan head
combination screws
combination screws
Figure 3-14 Removing cooler and sealing ring
How to do
1) Switch off the main power of the whole unit.
2) Open the shielding cover and remove the sample/reagent carousel cover. Remove the left panel, front
panel and sample/reagent desk panel. Protect the probe from being collided during the dismantling.
3) Unscrew the two spring screws to remove the sample/reagent carousel body.
4) Disconnect the drain tube from the reagent compartment. Unplug the connector of the temperature
sensor. If bar code reader is configured, remove it.
5) Remove the three M3*10 screws (hexagon socket cap head screws with spring and flat washers) and
the shaft sleeve.
6) Remove the screw plugs and remove the four M4*25 hexagon socket cap screws used to fix the
reagent refrigeration compartment with the hexagon wrench. Remove the refrigeration compartment
together with the refrigeration module.
7) Remove the four M4*12 stainless steel cross pan head combination screws of the cooler you want to
change. Clear the sealing glue at the cable and around the radiator and remove the radiator.
8) If you want to replace one cooler, unscrew the four M4*12 stainless steel cross pan head combination
screws of the cooler to remove it. Route the cable of the new cooler through the hole of the heat
insulation plate. Fix the heat insulation plate and daub heat conductive paste at the two ends of the
cooler. Install the cooler in place. Pay attention to the orientation of the cooler. Connect the cable and
daub sealing glue around the heat insulation plate and the cable routing hole. Tighten the screws of
the radiator with torque 9~10kgf and daub the sealing glue around the radiator. Restore the
instrument according to the reversed order of step 2~7.
9) If you want to replace the two coolers, remove the fixing plate of the cables of the cooler. Remove the
coolers after arranging the cables. Replace them with new coolers. Daub the heat conductive paste
at the two ends of the cooler and install them in place. Pay attention to the orientation of the cooler.
Arrange the cables and fix them with the removed cable fixing plate. Route the cable of the cooler
through the holes of the radiator. Use torque 11.5~12kgf to tighten the screws and daub sealing glue
around the radiators. Restore the instrument according to the reversed order of step 2~7.
Note
When using the sealing glue, daub sufficient sealing glue evenly.
Alignment and confirmation
1) Re-align Probe to outer ring, middle ring and inner ring positions,see 7.5.1 ~7.5.5 .
2) Re-align barcode scanning position,see7.8.3 Bar Code Reader Position Adjustment.
3.4.6 Replacing Home Position Sensor and Coder Sensor
When to do
When the Home position sensor and coded disk sensor of the reagent carousel failed.
Tools
Name Code Quantity
Cross screwdriver / 1
Hexagon wrench / 1
Medical rubber gloves / 1 pair
Exploded view for installation
M3*6 cross pan head screw M3*6cross pan
with washer head screw
Sensor mounting plate Coded disk
optical coupler
M3*6 Hexagon socket
cap screw with washer
M3*6cross
pan head
M3*6 hexagon socket cap head
screw
screw with spring and flat
Bracket of the washers
optical coupler Home position sensor
Reagent refrigeration
compartment
Air vent
How to do
1) Switch off the main power of the whole unit.
2) Open the shielding cover and remove the sample/reagent carousel cover. Remove the left panel, right
panel, front panel and sample/reagent desk panel. Protect the probe from being collided during the
dismantling.
3) Unscrew the two spring screws to remove the sample/reagent carousel body.
4) Disconnect the drain tube from the reagent compartment. Unplug the connector of the temperature
sensor and the cooler. If bar code reader is configured, remove it.
5) Remove the three M3*10 screws (hexagon socket cap head screws with spring and flat washers) and
the shaft sleeve.
6) Remove the screw plugs and remove the four M4*25 hexagon socket cap screws used to fix the
reagent refrigeration compartment with the hexagon wrench. Remove the refrigeration compartment
together with the refrigeration module.
7) Unscrew the eight M4*8 cross pan head combination screws to remove the air vent.
8) Mark a line at the place where the mounting plate of the motor aligns with big bottom plate (orientated
toward motor assembly tensioning. Remove the four M4*10 hexagon socket cap screws used to fix
the sample/reagent carousel motor assembly. Remove the sample/reagent carousel motor assembly.
9) Install new sample/reagent carousel motor and sleeve the synchronous belt.
10) Use one M3*20 hexagon socket cap screws with spring and flat washer to install into the hole of the
motor mounting plate. Tension the belt to the marked line and tighten the retaining screws. Remove
the M3*20 hexagon screw.
11) Install the air vent. Tighten the screws when the air vent aligns with the line marked before the
sample/reagent carousel motor assembly was removed.
12) Mount the reagent refrigeration compartment along with the refrigeration module onto the air vent, and
adjust the reagent refrigeration compartment’s position until its mounting holes aligns with those of the
air vent. Use the glue gun to plug up the reagent refrigeration compartment’s mounting holes with
sealing glue. Mount the four antiwear pads back onto the four M4*25 hexagon socket cap screws, fix
the reagent refrigeration compartment, and then install the screw plugs.
13) Restore the instrument according to the reversed order of Steps 2~6.
Note
Note: Do not mistaken the direction of the motor.
Alignment and confirmation
1) Align the position before fixing the air vent.
2) Re-align Probe to outer ring, middle ring and inner ring positions,see 7.5.3 7.5.5 .
3) Re-align barcode scanning position,see 7.8.3 Bar Code Reader Position Adjustment.
3.4.8 Replacing the synchronous belt
When to do
When it is damaged.
Tools
Name Code Quantity
Cross screwdriver / 1
Hexagon wrench / 1
Glue gun / 1
Medical rubber gloves / 1 pair
M3*20 hexagon / 1
socket cap head
screws with spring
and flat washers
Exploded view for installation
Screw plug
Shaft sleeve
M4*25 hexagon
socket cap head screw
Anti-wear pad
Reagent
refrigeration
chamber
Air vent
The belt tension needs to be adjusted to ensure that the tension is the same as before the replacement.
3.4.9 Replacing cooling fan of reagent refrigeration
When to do
When cooling fan failed.
Tools
Name Code Quantity
Hexagon wrench / 1
Medical rubber gloves / 1 pair
Cross screwdriver / 1
Exploded view for installation
M4*45 cross
pan head screw Φ92 Fan net
with the flat shield Cooling fan of reagent
washer refrigeration
Bar code
reader
M3*8 hexagon
socket cap head
screws with
spring and flat
washer Press plat of switch
Bar code Rubber
reader M3*6 cross pan head M3*12 cross pan head pad
Glass
screw screw window
4 7
3
9
1
10
11
12
Tools
Name Code Quantity
Cross screwdriver / 1
Medical rubber gloves / 1 pair
Exploded view for installation
M3*6cross pan
head screw
Figure 3-23 Replacing the optical coupler of horizontal and vertical movement
How to do
1) Switch off the main power of the whole unit.
2) Open the shielding cover. Unscrew the screw on the wash probes assembly with your hand and
remove the wash probes assembly. Take caution not to collide the wash probes. Unscrew all the M4*8
cross pan head screw with washer on the desk panels. Remove the cover of the sample/reagent
carousel, sample/reagent desk panel, reaction carousel desk panel and auto wash desk panel.
3) When replacing the vertical movement optical coupler, disconnect the cable and remove the M3*6
cross pan head screw to remove the optical coupler. Replace it with new optical coupler, fix it and
connect the cable.
4) When replacing the horizontal movement optical coupler, disconnect the cable, remove one M3*6
cross pan head screw to remove the bracket and then remove one M3*6 screw to remove the
horizontal movement optical coupler. Replace it with new optical coupler and fix the optical coupler
and its bracket with M3*6 screws. Connect the cable.
5) Install back the panels.
Alignment and confirmation
N/A
3.5.4 Replacing probe assembly-collision spring and powder screws
When to do
When probe assembly is damaged or anti-collision spring failed or cosmetic defect of the powder screw occurs.
Tools
Name Code Quantity
Cross screwdriver / 1
Flathead screwdriver / 1
Medical rubber gloves / 1 pair
Exploded view for installation
Spring guide
post
Probe
Reagent
Preheating
Assembly
Heat
insulation
pad
M3*6
cross pan
head
M3*8 screw
cross pan
head
screw
Hardened powder
screw(M3*6)
Level sense
board
Hardened power
screws(M3X6)
Horizontal
rotational
synchronous belt
M3*8 hexagon
socket cap screw
Horizontal
rotational
synchronous belt
M3*8 hexagon
socket cap screw
Probe movement
assembly
Synchronous
belt
Fixing plate B
M5*25 hexagon socket
cap head screws with
spring and flat washer
M4*16 hexagon
socket cap head
screws with
spring washer
Probe arm
assembly
Hardened powder
screws(M3X6)
Probe drive
assembly
M5*25 hexagon
socket cap head
screws with spring
and flat washer
10) Sleeve the cable and the tube and connect the cable with the PCBA. Connect the tube.
11) Connect the cables of the motor and the optical coupler.
12) Refer to 7.5.1 ~7.5.9 to align the positions: probe to reaction carousel, wash well, sample/reagent
carousel and ISE sample injection port. Tighten two M4*16 hexagon socket cap screws to fix the arm
assembly. Use two hardened powder screws (M3*6) to install the arm cover.
13) Install back the panels.
Alignment and confirmation
Refer to 7.5.1 ~7.5.9 to align the positions: probe to reaction carousel, wash well, sample/reagent carousel and
ISE sample injection port.
3.6 Mixer Assembly
3.6.1 Module Functions
Mixer assembly consists of drive assembly and arm assembly. The mixer can move in horizontal and vertical
direction and mix the reaction liquid after sample is dispensed and after the reagent is dispensed.
Horizontal direction: mixer can move to wash well and reaction carousel.
Vertical direction: The mixer moves in vertical direction among the positions of cuvette and wash well.
3.6.2 Component Locations and FRU Details
Mixer is located at the front of the instrument.
Mixer assembly
9
2
1
10
Diagonal pliers / 1
Medical rubber gloves / 1 pair
bracket
Figure 3-33 Replacing the optical coupler of horizontal and vertical movement
Replacement steps:
1) Switch off the main power of the whole unit.
2) Open the shielding cover. .Unscrew all the M4*8 cross pan head screw with washer on the desk panels.
Remove the cover of the sample/reagent carousel, sample/reagent desk panel, reaction carousel desk
panel and auto wash desk panel. Protect the mixer from being collided during the dismantling.
When replacing the vertical movement optical coupler, remove the left panel, right panel and front
panel.
3) When replacing the horizontal movement optical coupler, disconnect the cable, remove one M3*6
cross pan head screw to remove the bracket and then remove one M3*6 screw to remove the
horizontal movement optical coupler. Replace it with new optical coupler and fix the optical coupler
and its bracket with M3*6 screws. Connect the cable.
4) When replacing the vertical movement optical coupler, mark a line at the bracket of the wash well
before removing it. Disconnect the cables of the optical coupler, horizontal and vertical motor.
5) Draw a line at the place where the mixer assembly contacts with the big bottom plate. Remove the
three M5*25 hexagon socket cap head screws with spring and flat washers at the bottom of the mixer
assembly. Remove the mixer assembly.
6) Remove the M3*6 cross pan head screw to remove the vertical movement optical coupler. Replace it
with new one, fix it and connect the cable.
7) Use three M5*25 hexagon socket cap head screws with spring and flat washers to fix the mixer onto
the big bottom plate. The bracket must align with the marked line.
8) Install the wash well bracket on the mixer. The bracket must align with the marked line. Connect the
optical coupler and the cable.
9) Install back the panels.
Alignment and confirmation
Refer to 7.6.3 Mixer to Horizontal Wash Position to align mixer rotary to wash well.
3.6.4 Replacing horizontal rotational synchronous belt
When to do
When horizontal rotational synchronous belt failed
Tools
Name Code Quantity
Cross screwdriver / 1
Hexagon wrench / 1
Medical rubber gloves / 1 pair
Exploded view for installation
Horizontal
rotational
synchronous belt
Horizontal
rotational
synchronous belt
DC motor
Retaining nut
Mixer
Vertical
Probe movement
motor
assembly
Synchronous
belt
Motor fixing
plate B
M5*25 hexagon
socket cap head
screw
Motor fixing M4*16hexagon socket cap
plate A head screws with spring and
flat washer
Mixer arm
assembly
Hardened powder
screw(M3*6)
M3*10 hexagon
socket cap head
screws with spring
and flat washer
M5*25 hexagon
socket cap head Mixer drive
screws with spring unit
and flat washer
BA48 Wash
Wash probes
Station optical
coupler PCBA
Wipe block
Wash probe
drive assembly
OPtical coupler
wire (S)
Wash support
plate
Note:
Reinstall the tubes and cables on the wash probes assembly according to the number marking.
Alignment and confirmation
Check the wash probes' vertical position in reaction cuvettes according to 7.7.2 Cuvette Wash Station Height.
3.7.4 Replacing Wash Station
When to do
Position
adjusting washer
Position adjusting
plate
M4*12 hexagon
socket cap head screw
with washer
Support pole
041-031405-00
M4*12 hexagon
socket cap head screw
with washer
Hexagon wrench / 1
Exploded view for installation
Position
adjusting washer
Position adjusting
plate
M3*6 cross
pan head
screw
Zero position
photocoupler
Figure 3-45 Zero position photocoupler on the wash probe drive assembly
How to do
1) Switch off the main power of the whole unit.
2) Open the shielding cover. Unscrew the screw on the wash probes assembly with your hand to remove
the wash probes assembly. Take caution not to collide the wash probes. Rotate the probe to the
aspirating reagent position and rotate the mixer to the wash well position. Unscrew all the M4*8 cross
pan head screw with washer on the desk panels to remove the reaction carousel desk panel, auto
wash desk panel.
3) Unplug the connector of the zero position photocoupler.
4) Remove the M3*6 cross pan head screw to remove the zero position optical coupler. Replace it with
new one and fix it with one M3*6 cross pan head screw.
5) Restore the components in the reversed order.
Alignment and confirmation
N/A
3.8 Syringe Assembly
3.8.1 Module Functions
Syringe assembly consists of one sampling syringe and two wash syringes, of which 2.5ml syringe is used to
inject the wash solution and 10ml syringe is used to inject the cleaning fluid.
2.5ml wash
syringe
M3*12hexagon socket cap head screws with spring and 10ml wash
flat washer syringe
Figure 3-46 Location of the syringe assemblies
Table 3-8 List of materials
No FRU Part Name Remarks
1 115-045455-00 Probe syringe assembly
2 115-045444-00 2.5ml Wash syringe assembly
3 115-045917-00 10ml Wash syringe assembly
3.8.3 Replacing 2.5ml syringe and 10ml syringe
When to do
When the syringe functions abnormally.
Tools
Name Code Quantity
Cross screwdriver / 1
Hexagon wrench / 1
Exploded view for installation
See Figure 3-46 Location of the syringe assemblies.
How to do
1) Switch off the main power of the analyzer.
2) Remove the rear panel and unplug the tubes and cables. Use the hexagon wrench to remove the four
M3*12 hexagon socket cap head screws with spring and flat washer. Remove the 2.5ml or 10ml wash
syringe.
3) Install the new wash syringe.
4) Connect the tubes and cables. Install back the rear panel.
3.8.4 Replacing probe syringe assembly
When to do
When the probe syringe functions abnormally.
Tools
Name Code Quantity
Cross screwdriver / 1
Hexagon wrench / 1
entrance slit via the lens group 2. The gratings divide the light beam from the entrance slit and then converge
them to the slit array. Finally the PDA behind the slit array converts the light signals into electric signals and
then outputs them.
Grating
Slit
assembly
Optical
measurement
assembly
Light source
assembly
Figure 3-48 Locations of light source assembly and optical measurement assembly
The optical measurement assembly consists of the light-splitting assembly and the pillars.
Table 3-11 List of materials
No FRU Part Name Remarks
1 115-038466-00 20W lamp assembly(FRU)
2 801-BA40-00167-00 fiber bundle
3 115-045439-00 BA36 Cuvette (Surface
processed)
4 115-041267-00 Light-splitting assembly
(BA36)
Optical fibre
801-BA40-00167-00
Nut of Cable
terminal(Manually
tightened)
M3 socket set
screw
20W
Lamp(FRU)115- Retaining screw
038466-00
Figure 3-49 Exploded view of lamp and fiber bundle for installation
Note:
Replace the lamp according to the instructions on the screen.
Please wait for at least 5 minutes (if the lamp fan works properly) after turning off the lamp to avoid
injury.
Do not touch the bulb when installing the new lamp. In case of touching, please clean the bulb with
fibre cloth or ethanol-moistened gauze.
How to do
1) Select Utility > Maintenance > Maintenance > Biochemistry Maintenance.
2) Choose Replace Lamp. The maintenance guide window pops up. Select Continue.
3) Make sure that the lamp has cooled down for 5 minutes, and then select Continue.
4) Open the small door at the right panel.
5) Loosen the nuts on the cable terminals, and then remove the O-ring connectors from the terminals.
6) Loosen the retaining screw on the left side of the lamp.
7) Remove the lamp from the lamp housing.
CAUTION
Do not hold the lamp by its bulb to prevent contamination and damage.
8) Install the new lamp, and the retaining screw, O-ring connectors, cable terminal nuts and lamp cover plate
in the reversed order.
9) Select Done. Perform the Photometer Check procedure to ensure the system power is normal.
Alignment and confirmation
Refer to 7.3.3 Photoelectric Gain Adjustment to adjust the Photoelectric Gain.
3.9.4 Replacing Fiber Bundle
When to do
Replace the fiber bundle when it is broken or the light intensity is insufficient.
Tools
Hexagon wrench
How to do
1) Switch off the power supply of the whole unit.
2) Unscrew the screws on the right panel to remove it.
3) According to the figure above, use a hexagon wrench to remove the M3 socket set screws with cup
point on the light source assembly and pull out the small end of the optical fibre.
4) Loosen the Light-splitting assembly from the pillars and push it about 50mm toward the rear panel.
5) Remove the M3 socket set screws with cup point used to fix the optical fibre bundle. Pull out the optical
fibre bundle.
6) Remove the optical fibre bundle from the heat insulation chamber of reaction carousel.
7) Replace the optical fibre bundle. Note: first insert the optical fibre bundle into the Light-splitting
assembly and then push the Light-splitting assembly forward to fix it on the three pillars.
Precautions:
1)Do not bend the fiber bundle too much (less than R60mm for the best) during removing. When leaving the
factory, the fiber bundle is fixed in the middle by a cable tie on the heat insulation chamber earthing wire. You
can also directly cutting off the cable tie and use a new one in the spare parts kit instead.
2)Protect the exposed optical part on the head of the fiber bundle from scratching and contamination.
3)There are washers on each post used to adjust the height of the optical axis. Keep them properly and do not
drop them while removing the optical assembly.
M3 socket
set screws
Adjusting
iber bundle801-
washer
BA40-00167-00
Figure 3-50 Exploded view of fiber bundle on optical measurement assembly for installation
Alignment and confirmation
Refer to 7.3.2 7.3.3 to adjust the photoelectric signal collecting position and photoelectric gain.
Light-splitting
assembly (BA36)
115-041267-00
Adjusting
washer
Pillar
WARNING
While installing the reaction cuvettes, exercise caution to avoid scratching them. Do not touch the optical
surface of the reaction cuvettes. If the optical surface is polluted, the obtained absorbance may be inaccurate.
Wear gloves free of fibre and powder to avoid polluting the optical surface of the reaction cuvettes.
Biohazards
Wear gloves and lab coat, and if necessary, goggles during the maintenance process.
Tools
Cross screw driver, Fiber-free gloves, dry cloth or gauze, reaction cuvettes, and concentrated wash solution
manufactured by our company
How to do
1) Select Utility -> Maintenance -> Maintenance -> Biochemistry Maintenance.
2) Choose Replace Cuvette.
3) Open the cuvette replacement cover.
4) When Continue is clicked, the maintenance procedure starts.
5) Input the number of the cuvette segment on the popup window. The input range is 1~8.
6) Select Replace.
7) Push open the replacement window. Loosen the screw on the press plate with screw driver or your hand
and toggle the press plate to the right side. Take out the cuvettes.
8) Install new cuvettes and install back the press plate and tighten the screw.
If you want to replace multiple cuvette segments, you can turn off the power of the instrument and rotate
the reaction carousel with your hand to replace the cuvettes.
9) After the cuvettes are replaced, close the replacement window.
10) Select Utility-Maintenance-Maintenance-Biochemistry Maintenance and then select Cuvette Check. Check
the new cuvettes. Note: if Cuvette Check is not performed, the new cuvettes may not be used due to too
large water blank or alarm "Water blank out of range (10X)" may be given.
3.9.7 Photometer Lens Maintenance
When to do
The lens of the photometer is contaminated or the maintenance time is over 1 year.
Materials required
Lens cleaning paper, cotton swabs and ethanol
Note:
When moving the sample/reagent probe and mixer, do not bend or collide with them.
Do not leave the cotton tissue on the lens.
How to do
1) Place the analyzing unit power to the OFF position.
2) Loosen the nuts on the wash station, remove it and put in a clear box.
3) Remove the desk panels of reaction carousel and wash station.
4) Remove the upper cover of the heat insulation chamber and remove the reaction carousel assembly.
5) Use cotton swabs dipped with ethanol to wipe the exposed lens of the front and rear lens assembly.
The location of the lens is indicated in the figure below.
concentration of K+, Na+ and Cl- in serum, plasma and diluted urine. The sample volume for measurement is:
70µl for serum and plasma, 140µl for urine diluted at the ratio of 1:9 (1 for urine and 9 for diluent).
Work principle
The ISE Module measures sodium, potassium, lithium and chloride in biological fluids,
using ion selective electrode technology. The flow-through of each electrode uses selective membrane
tubing, specially formulated to be sensitive to corresponding ions. The potential of each electrode is
Measured relative to a fixed, stable voltage established by the double-junction silver/silver chloride
Reference electrode. an ion selective electrode develops a voltage that varies with the concentration
of the ion to which it responds.
The relationship between the voltage and the concentration can be expressed by Nernst equation
ENa ERef
Na
Ref
Two-points Calibration
→factor S
Calibrator Calibrator
B A
Pump Pump Pump ISE
Waste tank B A W
Parameters
Table 3-12 ISE Parameters
Item Features
Sample Serum,Plasma,Urine
Sample volume Serum, plasma 70uL, diluted urine 140uL
(10-fold dilution)
Method Direct method
Time 35s/cycle
Speed 100Test/hour
Power 24VDC ,1A
Communication RS232C
Maximum ambient 32℃
temperature
composition
ISE module
There are five kinds of electrodes with Na+, K+, Cl-, reference and spacer electrodes. The sample is added by
the addition port of the ISE module, and the sample is analyzed and measured.
Spacer
Urine diluent
Cleaning solution
SP3.5*8 screw
ISE module
ISE shield
M4*8 screw
Pump bracket
Pump
M2.5*6 screw
M2.5*6 screw
WARNING
Do not spill liquid on the analyzer. Liquid ingression may cause equipment damage.
Biohazards
Wear gloves and lab coat, and if necessary, goggles during the maintenance process.
Note
Excessive bleach and DI water flushed into the ISE reagent pack waste bag may cause waste bag over
expansion and clog the Cal A & Cal B reagent flow.
To prevent this problem, connect to an old used-up reagent pack or use the connector of the used-up reagent.
How to do
1) Ensure the analyzer is in idle (standby) status. Open the ISE window on the right side panel of
the analyzer.
2) Ensure the analyzer is in idle (standby) status. Open the ISE window on the right side panel of
the analyzer.
3) Connect the waste tube fitting to a syringe and unclogging tool with 5mL of undiluted household
bleach.
4) Press the wand release button to remove the wand from the current in use ISE reagent pack and
keep it in a save place. Engage the wand to an old used-up reagent pack.
5) Inject bleach into the ISE waste tube and soak the tube for 5 minutes. Discharge the waste into
the reagent pack.
Note: When the bleach cannot be injected into the ISE pack, remove the wand and push down to open the
waste valve manually with a sharp object, and then inject again. If bleach can go through this time, the waste
bag was clogged and cannot be used. If bleach still cannot be injected, replacing the ISE wand is recommended.
6) Repeat this step with 5mL of DI water without the 5 minutes of soaking time.
7) Remove the wand from the old use-up pack and re-install it back to the current in use ISE pack.
Re-install the waste tube fitting back to the ISE electrode housing right angle adapter and waste
peri-pump tube back to the pump bracket. Re-install the housing cover.
Alignment and confirmation
Perform ISE pump calibration and if it succeeds, the tube has been successfully unclogged.
3.10.4 Replacing Pump Tube
When to do
Replace the pump tube when it is aged or leaking or has not been maintained for a half year.
Maintenance Tools
Cross screwdriver, pump tube.
Precautions:
Biohazards
Wear gloves and lab coat, and if necessary, goggles during the maintenance process.
How to do
1) Ensure the analyzer is on idle (standby) condition. Open the ISE cover on the right side analyzer
panel.
2) Select Utility-Maintenance-Maintenance-ISE-Remove Reagent Pack.
3) Select Utility -> Maintenance -> Maintenance -> ISE -> Replace Pump Tube and Calibrator Tube.
4) Select Pump Tube.
5) Replace it with new one. Make sure the connectors and tubes are correctly connected; otherwise
ISE module failure may occur.
6) Load the reagent pack and the procedure is completed.
7) Install back the right side panel.
8) Record the maintenance log.
Alignment and confirmation
Perform ISE pump calibration and if it succeeds, the tube has been successfully replaced.
3.10.5 Replacing Calibrator Tube
When to do
Replace the calibrator tube when it is aged or leaking or has not been maintained for over one year.
Maintenance Tools
Biohazards
Wear gloves and lab coat, and if necessary, goggles during the maintenance process.
How to do
1) Ensure the analyzer is on idle (standby) condition. Open the ISE cover on the right side analyzer
panel.
2) Select Utility-Maintenance-Maintenance-ISE-Remove Reagent Pack.
3) Select Utility -> Maintenance -> Maintenance -> ISE -> Replace Pump Tube and Calibrator Tube.
4) Select Calibrator Tube.
5) Replace it with new one. Make sure the connectors and tubes are correctly connected; otherwise
ISE module failure may occur.
6) Load the reagent pack and the procedure is completed.
7) Install back the right side panel.
8) Record the maintenance log.
Alignment and confirmation
Perform ISE pump calibration and if it succeeds, the tube has been successfully replaced.
3.10.6 Replacing ISE electrode
ISE electrodes are consumables and have a limited life span. When used for a long period or after measuring
a large number of samples, the ISE electrodes may have their performance degraded and should be replaced
immediately.
Purpose
To replace the ISE electrodes to ensure the optimal measurement performance.
When to do
Reference electrode: Every 6 months
Other electrodes:
When 10,000 ISE tests are performed.
When the ISE electrodes are used for 6 months since installation.
When calibration fails or quality control is abnormal as the result of degraded electrode performance.
Materials required
Reference electrode, ISE electrode
System status
Make sure that the status of the ISE module is Standby or Stopped
Precautions
BIOHAZARD
Wear gloves and lab coat, and if necessary, goggles.
NOTE
After performing this procedure, recalibrate the ISE electrodes prior to starting analysis.
How to do
1) To remove the electrodes
a) Select Utility > Maintenance > Maintenance > ISE Maintenance.
b) Choose Replace Electrode.
c) Select desired electrodes, and enter the lot number and expiration date.
d) Select Add and then select OK.
e) Select Continue.
f) Open the ISE side door and remove the cover of the shielding box.
g) Open the electrode case, take out the electrode, remove the tapes around its inside, and then
use clean tissue to wipe it.
NOTE
Take out the insert from the reference electrode, and ensure no crystallized salt exists in and around it. If
needed, clean the electrode with warm water.
Make sure the red ball of the reference electrode floats on the internal fluid. Make sure the O rings of all
electrodes remain intact.
h) Remove all electrodes from the ISE module.
2) To install the new electrodes
a) To install a new reference electrode:
Place the reference electrode at the bottom of the ISE module and make the rear part of the electrode contact
closely with the internal wall of the ISE module.
Loosen the compressor and ensure the electrodes are fixed tightly.
b) Install other electrodes in the order of Cl, K, Na, and spacer from bottom to top.
c) If the O ring is lost, install a new one. There are two more O rings in each electrode case.
d) Check if the electrode positions are correct:
The Na, K and Cl electrodes are of the same size and shape. Ensure that the electrodes are inserted in
the correct order.
If one of the electrodes cannot be easily pushed into the housing, check the electrode first and then repeat
the installation process.
Check if the 5 electrodes are relatively on the same straight line; otherwise, liquid cannot flow through the
electrode tubes smoothly.
e) Select Continue. The system primes the tubes with calibrators A and B.
f) Select Done.
g) Restore the cover of the shielding box and close the side door of the ISE module.
h) Run ISE calibration.
i) If the calibration fails, perform the following operations:
Run ISE calibration for multiple times so that the electrodes can get steady quickly.
Or, drip little serum sample in the electrode channel, and leave it for 10-30 minutes, and then run calibration
again.
NOTE
The new electrode can be calibrated successfully only after certain time period.
electrodes properly before powering off the analyzer for a long period.
Purpose
To store the electrodes separately to prevent them from being damaged due to lack of water while the analyzer
is powered off.
Materials required
Electrode cases and tapes
When to do
Perform this procedure when the analyzer is going to be powered off for over 3 days. If it will be powered off for
no more than 3 days, prime the ISE electrodes to protect them from being damaged.
System status
Make sure that the status of the ISE module is Standby or Stopped
Precautions
BIOHAZARD
Wear gloves and lab coat, and if necessary, goggles during the maintenance process
note
Make sure that the storage temperature is below 40℃.
How to do
1) Remove the reagent pack,see 3.10.7 Removing reagent pack(ISE).
2) Open the ISE side door and remove the cover of the shielding box.
3) Remove all electrodes from the ISE module.
4) To save the reference electrode:
a) Put back the insert to the cell of the reference electrode and prevent the crystallized salt
from clogging the cell.
b) Store the electrode in an electrode case at the room temperature in a sun-shielding place.
5) To save the Na, K and Cl electrodes:
a) Take out a little calibrator A from the reagent pack, inject it into the cell of the electrode and
seal it with tape. Make sure proper amount of calibrator is injected into the cell of the
electrode.
b) Store the capped electrodes in an electrode case at the room temperature in a sun-shielding
place.
6) Restore the cover of the shielding box and close the side door of the ISE module.
3.10.9 Inventory Calculation of Reagent Pack
Reagent pack as one part of the ISE module should be installed as instructed by this manual. You should
install reagent pack while the analyzer is powered off, and then check the reagent inventory and prime the ISE
module with Calibrator A and B. If the reagent pack is not properly installed, error may be produced in inventory
calculation, and the actual volume may differ from that displayed on the screen.
The following operations may result in the actual inventory of the ISE reagent pack
less than the displayed one.
1) When the analyzer is turned on and the ISE module is in failure status, the reagent pack is changed, without
recovering the module failure or checking the inventory immediately.
Operation explanation: When the ISE module is in Stopped status, the software will stop writing volume
information into the reagent pack chip. However, the analyzer will automatically execute the sip action of the
ISE module during startup to consume certain amount of reagent. Therefore, when the reagent pack is replaced,
but the ISE module is not restored or the screen display refreshed, the software will not calculate the reagent
consumption volume until reading the new reagent pack. At this moment, the actual reagent volume =
displayed volume - consumption of Sip after replacing the reagent pack, which means that the actual reagent
volume is less than the displayed volume on the screen.
2) When the analyzer is powered on but the software is closed or the computer is shut down, the reagent
pack is changed.
Operation explanation: During shutdown, the analyzer will automatically execute the sip action of the ISE
module to consume certain amount of reagent. When the analyzer is started up, the software will detect a new
reagent pack and zero the consumption that is not written into the reagent pack chip, and then display the
volume read from the reagent pack chip. At this moment, the actual reagent volume = displayed volume -
consumption of Sip after replacing the reagent pack, which means that the actual reagent volume is less than
the displayed volume on the screen.
3) When the analyzer is powered on but the software is closed or the computer is shut down, the reagent
pack is switched to another analyzer and then installed back after use.
Operation explanation: During shutdown process, the analyzer cannot detect if the reagent pack is normal.
If the reagent pack is installed on the current instrument after being used on another one, the analyzer, when
started up, will detect that the recorded reagent volume on the chip is less than the original one, and the software
will zero the reagent consumption of Sip during shutdown and display the read volume on the screen. At this
moment, the actual reagent volume = displayed volume - consumption of Sip during shutdown (with reagent
pack installed), which means that the actual reagent volume is less than the displayed volume on the screen.
1) The ISE module is de-configured, but the reagent pack is still on the instrument.
Operation explanation: Though the ISE module is de-configured, the original reagent pack information is not
cleared from the database. When the ISE module is configured again with the same reagent pack, the Sip
period without ISE module will be included in calculation of the reagent consumption.
Note:
In condition 1, restore the ISE module and check the reagent inventory immediately to minimize or eliminate
the calculation error.
In condition 2 and 3, before running the operating software to start up the analyzer, remove the reagent
pack and then run the operating software. After the startup procedure is complete, re-install the reagent
pack and inquire the reagent volume immediately to minimize or eliminate the calculation error.
In condition 4, remove the reagent pack and then de-configure the ISE module.
The following operations may result in the actual inventory of the ISE reagent pack
more than the displayed one.
1) When the analyzer is turned on and the ISE reagent pack is changed, without checking the inventory
immediately.
Operation explanation: The analyzer will automatically refresh the ISE reagent consumption, by subtracting
the consumption of calibrator A or B from the known volume (of the original reagent pack) and writing it into the
reagent pack chip, when calibrator A or B is consumed for 1%. The written volume is based on the original
reagent pack before replacement, and the volume of a used reagent pack is often less than that of a new one.
So the volume recorded on the chip of the new reagent pack is less than the actual volume, and the actual
volume is greater than the displayed volume on the screen.
2) When the analyzer is powered on but the software is closed or the computer is shut down, the reagent
pack is removed for storage, and re-installed before the operating software is run.
Operation explanation: During shutdown process, the analyzer cannot detect if the reagent pack is normal.
If the reagent pack is loaded before the analyzer is started up, the analyzer, when started up, will detect that the
reagent volume recorded on the chip is same as that recorded by the software, but the software will confirm by
mistake the reagent consumption of sip during shutdown. (Actually, no reagent is consumed.) Therefore, after
calculating the reagent consumption, the software will write the new volume information onto the chip and
display it on the screen. At this moment, the actual reagent volume = displayed volume + consumption of Sip
during shutdown (with reagent pack installed), which means that the actual reagent volume is greater than the
displayed volume on the screen.
3) The analyzer is powered off after the operating software is closed, and it is powered on again before the
operating software is run.
Operation explanation: When closed normally, the operating software cannot detect if the ISE module is
working. When powered on before the software is run, the analyzer, when started up, will detect that the reagent
volume recorded on the chip is same as that recorded by the software, but the software will confirm by mistake
the reagent consumption of sip during shutdown. (Actually, no reagent is consumed because the ISE module is
not working while the analyzer is powered off.) Therefore, after calculating the reagent consumption, the
software will write the new volume information onto the chip and display it on the screen. At this moment, the
actual reagent volume = displayed volume + consumption of Sip during shutdown (with reagent pack installed),
which means that the actual reagent volume is greater than the displayed volume on the screen.
Note:
In condition 1, inquire the ISE reagent volume immediately to minimize or eliminate the calculation error.
In condition 2 and 3, before running the operating software to start up the analyzer, remove the reagent
pack and then run the operating software. After the startup procedure is complete, re-install the reagent
pack and inquire the reagent volume immediately to minimize or eliminate the calculation error.
4 Hardware Circuits
4.1 Overview
This chapter describes the functions of the printed circuit board assemblies (PCBAs) used on theBS-360E.
4.2 Harzards
Note:
While the instrument is working, do not touch the hardware circuit boards with your hands or other objects.
Before removing a circuit board, disconnect the instrument from the (AC) power supply. Please wear a pair of
anti-static gloves or take other measures to prevent static electricity prior to removing a circuit board.
PCBA(PCB) Functions ID
Main board The main control board is the control center #1
115-048858- of the instrument. It is mainly used to fulfill
Main board(open system) the following tasks: 1) communicating with a
00
computer through the RS232 serial port to
115-048859- Main board(Close System 3 transmit data and instructions; 2)
00 User-defined) communicating with the smart modules,
including the ISE module, through the
115-048860- Main board(Close System 5 extended serial ports to transmit data and
00 User-defined) instructions; and 3) controlling digital
115-048861- Main board(Close System 0 potentiometer adjustment and photoelectric
00 User-defined) data collection of the AD collection board
and receiving the photoelectric data.
Power driver board The power drive board drives and controls #2
051-002896-00 the sample/reagent probe, mixer,
sample/reagent carousel, reaction carousel,
wash station and wash syringe and other
related components.
Integration Liquid Detecting Board The board detects the level of samples and #3
051-002479-00 reagents, and detect/convert the signals of
the vertical obstruct detection sensor.
Pre-amplifying AD board The pre-amplification board functions the #4
optics-electricity and AD conversions of the
signals of the discrete photodiode array. At
the same time TTL serial port is provided to
communicate with main control board.
DC/DC board It is used to convert the 12V power supply #5
051-002900-00 of the whole unit, outputting the A12V, A5V
digital, D12V/E12V analog, C12V power,
B24V power, and the B12V power for partial
coolers.
AC Driver Board It drives the reaction carousel, AC heater of #6
051-002872-00 the wash solution and cleaning fluid.
Sample Probe Clogging Detection Board It detects the clog and empty aspiration of #7
801-BA40-00016-00 the probe.
ISE power PCBA It is used for powering the ISE module. #8
The locations of the PCBA on the instrument.
Power driver
board
AC Driver
Board
ISE power
PCBA
Main board
Integration
Liquid Detecting
Board
Power supply
assembly
Pre-amplifying AD
board
Optical System
Photoelectric signal
Control signal
Temperature
control signal
Temperature control
PC RS232 Hardware System
system
Control signal
Control signal
Position signal
AC Drive
Board
Power supply
AC /DC terminal
DC/DC Power
Power supply
module
Power drive
Main board
board
Pre-amplifying
AD board
Bus
Drive clock/direction signal
Level floater Mixer Unit
TTL Serial
Wash pump valve signal
Port
Bus
Reaction Unit Drive clock/direction signal
PC TTL Serial
Port
Bus
Cuvette Wash Unit
TTL Serial
Wash pump valve signal
Port
Main Board
Figure 4-5 Functional diagram of main board
Description
PCB
The PCB layout of the main board is as shown below.
Connector (J39) for Pre-amplifying AD board: DB25, used for communication with the AD collection board
and providing power supply for it.
Connector (J6) for ISE module: 8-pin, RS232, used for communication with the ISE module.
Connector (J5) for bar code reader: 8-pin, RS232, used for communication with the bar code reader.
Connector for optical coupler (J12):36 PIN, used for the input of the sensors of moving parts.
Connector for optical coupler (J13):36 PIN, used for the input of the sensors of moving parts.
Connector for optical coupler (J14):10 PIN, used for the input of the liquid floater sensor.
Connector of temperature sensor (J18):4PIN, used for the input of temperature sensor of the reaction
carousel.
Connector of temperature sensor (J21):4PIN, used for the input of temperature sensor of cleaning fluid.
Connector of temperature sensor (J22):4PIN, used for the input of temperature sensor of wash solution.
Connector of temperature sensor (J23):4PIN, used for the input of temperature sensor of reagent
preheating.
Connector of the liquid level detection board (J11):4PIN, used for the signal input of liquid level detection
board and anti-collision signal.
Connectors for debugging:
JTAG connector (J8): 10-pin, used for debugging the FPGA.
Indicators
The main board contains the following indicators.
+12V power supply (D8): green It is lit when the analyzer power switch is turned on, indicating that the
+12V power supply has been connected.
+5V power supply indicator (D7): green. It is lit when the analyzer power switch is turned on, indicating that
the +5V power supply has been connected.
+3.3V power (D9): green It is lit when the analyzer power switch is turned on, indicating that the +3.3V
power supply has been connected.
D19: orange FPGA configuration indicator. On indicates FPGA is successfully configured.
D14: Green FPGA working status indicator. LED is flashing every second, indicating FPGA works
normally.
D23: green When communication with the lower layer unit is interrupted, the indicator is lit.
Test points
In the following positions of the main board can signal tests be performed.
12V: +12V power supply input. Normal range: 11.4 - 12.6V.
5V: +5V power supply input. Normal range: 4.75 - 5.25V.
VDD: +3.3V power supply. It is secondary power supply used to power the major digital circuits of the PCBA.
Normal range:2.97~3.63V
2.5V:+2.5V power supply.It is secondary power supply used to download the FPGA.Normal
range:2.25~2.75V
1.2V: +1.2V power supply.It is secondary power supply used to power the FPGA core.Normal
range:1.14~1.26V
Note
Prior to removing the PCBA, disconnect the instrument from the power supply and wear a pair of anti-static
gloves or wrist straps.
Make sure that the connectors are inserted properly into the PCBA.
Check the connectors with locks and ensure they have been locked properly.
Check other connectors and ensure that they are inserted into the end of the slots.
It requires great force to plug/unplug the J1,J12~J13 connectors. Hold the PCBA by its edge while
plugging/unplugging the connectors to prevent it from being deformed or damaged.
After connecting J39 connector (DB25), tighten the retaining screws on two sides of it.
Save the parameters before removing the PCBA in order to load the parameters after the PCBA is removed.
Bus
Reaction unit Drive clock/direction signal
PC TTLserial Wash syringe motor
port
Bus
Auto wash unit
FPGA TTLserial Wash pump/valve signal
Pump/valve Pump/valve
port drive
DC motor
mixer
drive
Rgt
Bus
Heater control signal preheating heater
Temperature drive
TTLserial control unit TTLserial port
port NTC(Reagent
Refrigeration
MCU_2 sensor)
Connectors:
The power drive board contains the following connectors.
Power supply:
Power supply input (J17): 8-pin, providing 24V, 12V analog and +5V digital for PCBAs.
Pin No. Signal Reference Value
1 +24V 23.5~26V
2 +24V 23.5~26V
3 +5V 4.75~5.25V
4 +12V 11.4~12.6V
5 GND /
6 GND /
7 GND /
8 GND /
Connectors for sending/receiving communication signals:
Connector for the control signals of main control board (J20):40PIN, TTL used to transmit the control signal.
Communication connector (J34) for the communication between the reagent refrigeration and main control
board.
Connectors for drive and control:
Connector for pump/valve drive (J6):8 PIN, used to drive sample/reagent probe interior pump and valve,
exterior wash valve and phase 2 and 3 aspirating pump.
Connector for pump/valve drive (J9):8 PIN, used to drive phase 1 and 4 waste pump, cleaning fluid injection
valve and mixer exterior valve.
Connector for pump/valve drive (J28):20 PIN, used to drive SV01 inlet valve, SV06 wash solution injection
valve and SV07 injection valve.
Reaction carousel motor drive connector (J15):4PIN used to drive reaction carousel motor.
Sample/Reagent carousel motor drive connector (J25): 4-pin, used for driving the sample/reagent carousel
motor.
Sample/Reagent horizontal motor connector (J5): 4-pin, used for driving the sample/reagent horizontal
motor.
Sample/Reagent vertical motor connector (J5): 4-pin, used for driving the sample/reagent vertical motor.
Sample/reagent syringe motor drive connector (J8): 4-pin, used for driving the sample/reagent syringe
motor.
Cleaning fluid syringe motor drive connector (J7): 4-pin, used for driving the cleaning fluid syringe motor.
Wash solution syringe motor drive connector (J12): 4-pin, used for driving the wash solution syringe motor.
Wash station motor drive connector (J4): 4-pin, used for driving the wash station motor.
Sample/reagent mixer horizontal motor connector (J10): 4-pin, used for driving the sample/reagent
horizontal motor.
Sample/reagent mixer vertical motor connector (J14): 4-pin, used for driving the sample/reagent vertical
motor.
Mixer motor drive connector (J22): 2-pin, used for driving mixer motor.
Reagent preheating drive connector (J24): 2-pin, used for driving the reagent heater.
Indicators
The power drive board contains the following indicators.
+12V power supply indicator (D23): green. It is lit when the analyzer power switch is turned on, indicating
that the +12V power supply has been connected.
+5V power supply indicator (D26): green. It is lit when the analyzer power switch is turned on, indicating
that the +5V power supply has been connected.
+24V power supply indicator (D24): green. It is lit when the analyzer power switch is turned on, indicating
that the +24V power supply has been connected.
+3.3V power supply indicator (D25): green. It is lit when the analyzer power switch is turned on, indicating
that the +3.3V power supply has been connected.
Refrigeration 5V power (D55): green it is lit when system power switch is turned on, indicating that the 5V
power supply has been connected.
Refrigeration 3.3V power (D54): green It is lit when system power switch is turned on, indicating that the
3.3V power supply has been connected.
Refrigeration unit working indicator (D62): yellow. When the power of the whole unit is turned on, the
indicator flashes indicating refrigeration unit functions normally.
Indicator (D63) when reagent compartment temperature is higher than the target range: red. When the
reagent compartment temperature is higher than the target range.
Indicator (D65) when reagent compartment temperature is within the target range: green. When the
Note
Prior to removing the PCBA, disconnect the instrument from the power supply and wear a pair of anti-static
gloves or wrist straps.
Make sure that the connectors are inserted properly into the PCBA.
Check the connectors with locks and ensure they have been locked properly.
Check other connectors and ensure that they are inserted into the end of the slots.
It requires great force to plug/unplug the J20 and J17 connectors. Hold the PCBA by its edge while
plugging/unplugging the connectors to prevent it from being deformed or damaged.
Main board
Main board
ADC
Description
PCB
The PCB layout of the pre-amplification AD board is as shown below.
D3
J3
J2
Note
Prior to removing the PCBA, disconnect the instrument from the power supply and wear a pair of anti-static
gloves or wrist straps.
Make sure that the connectors are inserted properly into the PCBA.
Check the connectors with locks and ensure they have been locked properly.
Check other connectors and ensure that they are inserted into the end of the slots.
It requires great force to plug/unplug J1 connector. Hold the PCBA by its edge while plugging/unplugging
the connectors to prevent it from being deformed or damaged.
After connecting P1 connector (DB25), tighten the retaining screws on two sides of it.
Since encapsulated inside the Optical measurement assembly, the preamplifier AD board should be maintained
together with the Optical measurement assembly rather than maintained independently.
4.4.5 Integration Liquid Detecting Board
Functions and Principles
Detecting the reagent and sample levels capable of detecting the fluid level steadily and reliably, especially
allowing the probe to correctly detect the fluid level inside reaction cuvettes.
Outputting level sense signals to the control drive board when the probe contacts the fluid level.
Capable of detecting obstruct in vertical direction and outputting the signal to the main control board.
The functional diagram of the level sense board is as shown below.
Vertical
collision
circuit
Dual
tube Capacity Connect
probe Main
detection CPU or
Board
chip circuit
RAM: vertical obstruction signal output. Normal condition: High level (about 4V) is output when no vertical
obstruction occurs, and low level (about 0V) is output when vertical obstruction happens.
Note
Prior to removing the PCBA, disconnect the instrument from the power supply and wear a pair of anti-static
gloves or wrist straps.
Make sure that the connectors are inserted properly into the end of the slots on the PCBA.
It requires great force to plug/unplug the connectors. Hold the PCBA by its edge while plugging/unplugging
the connectors to prevent it from being deformed or damaged.
AC drive board
Note
Prior to removing the PCBA, disconnect the instrument from the power supply and wear a pair of anti-static
gloves or wrist straps.
Make sure that the connectors are inserted properly into the end of the slots on the PCBA.
It requires great force to plug/unplug the connectors. Hold the PCBA by its edge while plugging/unplugging
the connectors to prevent it from being deformed or damaged.
PC(for debugging
only)
Description
PCB layout
Clog detection board control connector (J2): connected with the main board and used for communicating control
signals during clog detection process.
Pin No. Description Test Methods
1 +12V power supply The voltage lies between 11.4-12.6V when
the PCBA is working normally.
3 Confirmed clog signal The voltage at this point is less than 0.4V
when no clog signal is detected.
6 Interruption control signal The voltage at this point is less than 0.4V
when the system status is Standby.
7 Interruption enable signal The voltage at this point is less than 0.4V
when the system status is Standby.
8 Ground /
Clog detection board communication connector (J3): connected with the main board and used for serial port
communication between the clog detection board and the sample probe unit.
Pin No. Description Test Methods
1 +12V power supply The voltage lies between 11.4-12.6V when
the PCBA is working normally.
3 Serial port input signal The voltage at this point is greater than 2.4V
when the system status is Standby.
4 Serial port output signal The voltage at this point is greater than 2.4V
when the system status is Standby.
5 Online-downloading The voltage at this point is less than 0.4V
enable signal when the system status is Standby.
6 Reset signal The voltage at this point is greater than 2.4V
when the system status is Standby.
7 Ground /
8 Ground /
Clog detection board debugging connector (J4): connected with the serial port of a computer and used for
debugging the clog detection system. This connector is temporarily not used.
Pin No. Description Test Methods
1 +12V power supply /
2 -12V power supply (reserved) /
3 Serial port input signal /
4 Serial port output signal /
5 Ground /
normally.
-12V power supply (VEE): reserved test point.
Ground (GND): This test point is connected with the grounding terminal of the PCBA.
Pressure sensor output signal (ANALOG): It is an analog signal amplified by the pressure sensor. When
the pressure sensor is connected to the air, the voltage at this point lies between 0.8-1.3V.
Clog signal (CLOG_Y): It is a test point of clog signal output. When the PCBA outputs a clog signal, indicator
D3 will be lit and the voltage at this point greater than 2.4V; when the PCBA outputs no clog signal, indicator
D3 will be extinguished and the voltage at this point less than 0.4V.
Empty aspirate signal (NO_SAM):When the PCBA outputs an empty aspirate signal, indicator D4 will be lit
and the voltage at this point greater than 2.4V; when the PCBA outputs no empty aspirate signal, indicator
D4 will be extinguished and the voltage at this point less than 0.4V.
Debugging test point (TEST): used for test during debugging.
Note
Prior to removing the PCBA, disconnect the instrument from the power supply and wear a pair of anti-static
gloves or wrist straps.
Insert connectors J2 and J3 tightly into the PCBA. When inserting connector J1, you will hear a click, which
means that the connector is inserted in place into its slot.
Power supply
AC drive board
E
yellow- Shell L
green grounding
Main switch
ISE module
V+
Power drive Lamp
L V+ board
V-
AC/DC power
V-
module
A12V/2A Lamp
Pre-amplifying
AD collection
5V/4A Main board 12V board
Level sense
board
AC/DC DCDC Probe clog
12V
power Power 12V/5A detection
board
module board
Power drive
24V/6A board
Power
12V/11A
conversion ISE module
board
4 GND
5 NC
6 GND
J4 Connector for input of 1 GND
lamp control signal 2 Lamp sleep control.0 for
sleep
3 GND
4 Lamp turn-on control. 0
stands for turning on the
lamp.
5 NC
6 NC
J5 Connector for lamp 1 LAMP_V+
power output 2 LAMP_V-
J6 Connector for ISE power 1 GND
output 2 NC
3 12V
J7 Connector for power 1 24V
drive board output 2 24V
3 5V
4 12V
Open with a
flathead screwdriver
Figure 4-23 Replacing the fuse 1
2) Insert the fuse into its base. Note: The fuse for model with 220V power supply is
010-000003-00; the fuse for model with 110V power supply is 010-000082-00.
Press into
the slot
D D
AC drive
board
Power input
terminal
DC/DC AC /DC
power Power
module module
C C
pre-
amplification AD
board
B B
APPROVALS DATE
DESIGN MINDRAY
CHECK
A TITLE A
CHECK Main unit wiring diagram
File: Bytes: CONFIDENTIAL DISCLOSURE: This set of drawing(s) and all it's intellectual property rights (including copyright) subsisting CHECK
herein are property of Shenzhen Mindray Bio-medical Electronics Co.,Ltd. No use, copies or reproductions should be made of DWG NO. BA36/BA37 REV 1.0
Date: Time: this drawing or any part(s) thereof for whatever purpose nor shall any information, data, calculations, or other contents
contained in this drawing be disseminated without prior written permission of Shenzhen Mindray Bio-medical Electronics
R&D
Software & Rev: Microsoft office Visio 2003
Co.,Ltd.
CHIEF ENG. SHEET 1 OF 5 SIZE A4
1 2 3 4 5 6 7
protection switch of
protection switch of
reaction carousel
upper heater of
reaction carousel
Temperature
protection switch
upper heater of
reaction carousel
protection switch
of cleaning fluid
reaction carousel
protection switch
Upper heater of
Temperature
Lower heater of
of wash solution
of lower heater
Temperature
Cleaning fluid
Wash solution
Temperature
Temperature
of reaction
carousel
heater
heater
D 009-007706-00 D
1
009-008089-00
1 2 1 2 1 2
2
J16 J5 J7 J6
3
009-007683-00
4
5 220VCTRL1 6 blue
L 009-008112-00
brown L
E
Shell
5 yellowgr
5
6
6 220VCTRL2 4 red
8 220VCTRL3 6
AC drive board een grounding
7 33
8 34
5 black
33 VCC 1 blue 7 (051-002872-00) Light N Power
socket
Main power
switch
N blue
8
N light
blue
L brown
brow L
009-007677-00
Light
blue
9 LAMPCTRL 1 yellow
n
10
5 1
9
6 2
N
7 GND 2 black
L
V+ V+
15 GND 4 black 7 3 12Vout red
V+ AC/DC Power supply
Main board 8 4 12Vout red
module
C V+ C
9 12V- black
10 V- V- (022-000285-00 )
11 12 12V- black
V-
12V- black
V-
DC/DC Power supply
009-007678-00
J2 12GND 6
009-007680-00
6 blue
board
4 1
4 1 GND 4 4 black (051-002900-00)
12V 3 3 yellow 5 2
5 2
5V 1 1 red 6 3
6 3
12Vout
009-007679-00
GND
blue C
black
MINDRAY
+
NO
A
analyzing
Switch of
A
TITLE: Power supply wiring diagram
unit
File: Bytes:
DWG NO. BA36/BA37 REV 1
Date: Time:
PROBE
C31: Probe
GND
1black
2red
D
D
1blac
2red
k
BNC BNC BNC
Power drive
J20
1 2
Probe clog detection board
J1
Sensor of diluted wash
board
DI water sensor
Probe Power drive board
solution (051-002479-00) J2 J3
12 11
18 17
28 27
32 31
38 37
42 41
22 21
48 47
14 13
16 15
20 19
24 23
26 25
30 29
34 33
36 35
40 39
44 43
46 45
50 49
button
sensor
1
3
5
7
9
J2 J34 1 3 5 7 1 3 5 7 9
2
4
6
8
4
1 2 3 4 1 2 3 4 5 2 4 6 8 2 4 6 8 10
4 NC
5 NC
7 NC
6 NC
2 NC
3 NC
2
1
1
3
2
8
7
3
3
7
6
5
6
1
4
39 CLK_RESERVED_MOTOR3
40 DIR_ RESERVED_MOTOR3
009-007715-00
26 CLK_RESERVED_MOTOR1
30 CLK_RESERVED_MOTOR2
009-008100-00
31 DIR_ RESERVED_MOTOR2
27 DIR_RESERVED_MOTOR1
009-007684-00
009-008097-00
009-007713-00
11 VALVE_WASH_INJECT1
12 VALVE_WASH_INJECT2
CLOT_YES white 6
INTEN purple
GND black
RXD orange
GND black
PSEN brown
RST brown
INT green
38 DIR_ SYR_MOTOR2
37 CLK_SYR_MOTOR2
36 DIR_ SYR_MOTOR1
35 CLK_SYR_MOTOR1
VPP red
2 yellowLEVEL1
10 PUMP_FILL_OUT
13 PREHEAT_REAG
16 PWR_PV_WASH1
17 PWR_PV_WASH2
4 red LEVEL2
25 CLK_ WASH_UP
3
45 CLK_FILL_SYR
9 VAVLE_FILL_IN
43 CLK_FILL_UP
8 PUMP_FILL_IN
46 DIR_FILL_SYR
32 CLK_STIR_UP
41 CLK_FILL_R
33 DIR_STIR_UP
15 LAMP_CTRL
42 DIR_FILL_R
22 CLK_STIR_R
28 CLK_FILL_T
23 DIR_ STIR_R
29 DIR_FIIL_T
20 REAC-CLK
21 REAC-DIR
18 RELAY_T
14 STIR_DC
blueGND
whiteGN
19 GND
50 VCC
34 GND
47 GND
48 VCC
49 VCC
1 GND
D
12
1
3
4 NC
5 NC
7 NC
6 NC
3 NC
2 NC
11
2
5
1
4
8
9
NC
6
NC
3
2
4
8
7
5
6
J14 2 4 6 8 10 12 2 4 6 8 10 12
12 11
18 17
28 27
32 31
38 37
42 41
10 9 8 7 6 5 4 3 2 1
22 21
48 47
2 3 4
14 13
16 15
20 19
24 23
26 25
30 29
34 33
36 35
40 39
44 43
46 45
50 49
1
1
3
5
7
9
J1
C J4 1 3 5 7 9 11 C
1 3 5 7 9 11
2
4
6
8
4
J11
J3
J39
J5 J6 J7
13 12 11 10 9 8 7 6 5 4 3 2 1
2 4 6 8 10 12 14 16 18 20 22 24 26 28 30 32 34 2 4 6 8 10 12 14 16 18 20 22 24 26 28 30 32 34 2 4 6 8 J18 J23
J22 J21
2 4 6 8 2 4 6 8 10
J12 1 3 5 7 9 11 13 15 17 19 21 23 25 27 29 31 33 J13 1 3 5 7 9 11 13 15 17 19 21 23 25 27 29 31 33 1 3 5 7 25 24 23 22 21 20 19 18 17 16 15 14
1 3 5 7 1 3 5 7 9
4 3 2 1 4 3 2 1 4 3 2 1 4 3 2 1
3red
2black
3 PRTD_T1
2 PRTD_REF
1 SHIELD
1 SHIELD
009-007685-00
SENSOR
SENSOR
SHIELD
SHIELD
1red VCC 1
009-007685-00 BA10-20-78117 BA34-20-63652
+ 4white 1red 1
REF
REF
PHO_REAC_T 2 VCC
Home sensor of reaction C 2black +
009-006306-00
4white PHO_FILL_T 2
009-007706-00
carousel - GND 3 Home sensor of C
1 NC NC 1
E 3green GND 4 sample/reagent carousel - 2black GND 3 3
PC COM
1red VCC 5 3green GND 4 RXD 2 1
E
009-007688-00
+ 4white 1red
009-007688-00
6 VCC 5 2 TXD 3 NC
Coded disk sensor of reaction C 2black PHO_REAC_TCGND Coded disk sensor of + 4white PHO_FILL_TC 6 009-000880-00
2
009-007690-00
carousel - 7 C 4 NC NC 4
2black
8 AD_START 20
sample/reagent carousel GND 7
E 3green GND 8 - 1 VCC red 1 RXD232
black 2
VCC 9
+ 1red E 3green GND 8 ISE module 5 GND 5 3
red 3
1red VCC 9
1 +12V 1
C4whitePHO_FILTER_MOTOR 10
9 ISP 22
TXD232
2 GND 5
blac
7 BUSY 18
+
5 RST 21
red
3 TXD 19
4 RXD 6
Probe horizontal 2 RXD green 3 6 NC NC 6
4white PHO_WASH_UP 10
k
11
1
anti-collision sensor - 2black GND C 3 TXD blue 2 4 B
B E 3green GND 12 Idle - 2black GND 11 8 RTS 7 NC
1red
+ 4white
VCC 13 E 3green GND 12 5 GND black 4 5
1red VCC 13 7 CTS 8 Temperature
C 2black PHO_FILL_UPGND 14
preheating sensor
Home sensor of probe vertical + GND
Reagent barcode
- 15 4whitePHO_WASH_SUCKSYR 14 6 RTS white 10 9 NC sensor of
2
3
1
2
3
motor C 6
M29:reagent
E 3green GND 16 Idle - 2black GND 15
7 CTS yellow 6 NC reaction
+ 1red VCC 17
18
E 3green
1red
GND 16
17 8 TRIG orange 9 7
C 4whitePHO_FILL_R
VCC carousel
reader
Home sensor of probe rotation + ISP_RESET
Of cleaning fluid
Home sensor of probe syringe C 2black PHO_FILL_SYR + 1 2 3 4 5
of wash solution
NC 8 NC
motor - GND 23 C 4white PHO_RESERVED1 22
Home position of wash station 2black GND 23 10
E 3green GND 24 -
3green GND 24
NC 11
NC AD collection
+ VCC 25 E NC 12
Home sensor of mixer vertical C 4white PHO_STIR_UP 26 + VCC 25 SHIELD board
- 2black GND 27 4white PHO_RESERVED2 26 NC 13
motor Home position of cleaning C
E 3green GND 28 - 2black GND 27
1red VCC 29 fluid syringe 3green GND 28 NC 14
Home sensor of mixer rotation + 4white E 1red VCC 29
C 2black PHO_STIR_R 30 Home position of wash + NC 15
motor GND 31 4white PHO_RESERVED3 30
- solution syringe C 2black GND 31
E 3green GND 32 -
3green GND 32
33 NC E
34 NC 33 NC
34 NC
MINDRAY A
A
File: Bytes:
1 2 3 4 5 6 7
1P01:Interior pump
SV06:Wash solution
1
SV05:Cleaning
2
2
2
1SV02:Interior valve
SV07:Wash solution
1
21
2
Phase 3 aspirating
M19:Mixer DC
SV01:Inlet valve
preheating block
dispensing valve
dispensing valve
dispensing valve
aspirating pump
M29:Reagent
aspirating pump
P02:phase 2
SV03: Probe
exterior valve
SV04:Mixer
exterior valve
P04: phase 4
motor
P03:
pump
fluid
D - D
+
2
1
1
2
009-007693-00
1blue009-007706-00
009-007694-00
009-007695-00
1blue009-007716-00
2red
8
7
6
5
4
3
2
1
2red
8
7
5
4
2
1
5
4
2
1
009-007686-00 1 2 1 2 7 5 3 1 7 5 3 1 2 4 6 8 10 12 14 16 18 20
8 6 4 2 8 6 4 2 1 3 5 7 9 11 13 15 17 19
6 6white 4white
1
J22 J24
4 4yellow 3yellow J9 009-007715-00
1 black
J6 J28
4 d3 2
1 2
1 2
J8
3 3blue 2blue 2e 3 4
r
1 GND 1GND
1red 3 4 2 NC 2 NC
1 1red 5 6
3 NC 3 NC
5 6 4 NC 4 NC
5 NC 5 NC 7 8
C 7 8 6 NC 6 NC C
6white 4white 9 4
1
7 NC 7 NC
6 8 PUMP_FILL_IN 8
9 4
4yellow 3yellow 9 VAVLE_FILL_IN 9 11 12
2
4
J13
10 PUMP_FILL_OUT 10
11 12 11 VALVE_WASH_INJECT1 11
12 13 14
3 3blue 2blue 12 VALVE_WASH_INJECT2
3
13 14 13 PREHEAT_REAG 13
1red 14 STIR_DC 14 15 16
1 1red 15 16 15 LAMP_CTRL 15
4
16 CLK_WASH_SYR 37 17 18
17 DIR_WASH_SYR 38
17 18
J20 23 24 24 CLK_RESERVED_MOTOR1 24
J7
25 DIR_RESERVED_MOTOR1 25 25 26
3 2blue 2blue 26 CLK_WASH_UP 26
3
25 26 27 DIR_WASH_UP 27
28 CLK_FILL_T 28 27 28
1 1red 1red 27 28 29 DIR_FIIL_T 29
4
30 CLK_RESERVED_MOTOR2 30 29 30
29 30 31 DIR_ RESERVED_MOTOR2 31
32 CLK_STIR_UP 32 31 32
6white 4white 31 32 33 DIR_STIR_UP 33
1
6 34 GND 34 33 34
35 CLK_SYR_MOTOR1 35
4yellow 3yellow 33 34 36 DIR_ SYR_MOTOR1 36
4
2
37 CLK_RESERVED_MOTO4 16 35 36
J5
37 38 40 DIR_ RESERVED_MOTOR3 40
B 1 1red 1red 41 CLK_FILL_R 41 39 40 B
4
42 DIR_FILL_R 42
39 40 43 CLK_FILL_UP 43
44 DIR_ FILL _UP 44 41 42
41 42 45 CLK_FILL_SYR 45
46 DIR_FILL_SYR 46 43 44
1
45 46
J11
4 50 VCC 47 48
2blue 47 48
3
3 3blue 49 50
1 1red 1red 49 50
4
3yellow
4white
3yellow
1red
1red
4white
2blue
1red
4white
3yellow
4white
3yellow
2blue
3yellow
1red
2blue
4white
1red
1red
2blue
2blue
2blue
MINDRAY
4yellow
6white
4yellow
6white
4yellow
4yellow
3blue
4 4yellow
4yellow
6white
6white
6white
6white
3 3blue
3 3blue
3blue
3 3blue
3blue
1 1red
1 1red
1 1red
1red
1 1red
1 1red
A A
6
4
3
6
4
6
4
6
4
3
1
6
TITLE:
6
4
3
Power drive board wiring I
File: Bytes:
DWG NO. BA36 REV 1
Date: Time:
1 2 3 4 5 6 7 8
Anti-fogging
Anti-fogging temperature
Semi conductor Semi conductor Semi conductor switch
heater
cooler cooler cooler
Thermistor
BA30-10-06633 BA30-10-06633 BA30-10-06633
BA40- M07-00062S-
3001-21-07100 --
D D
2 green
1 black
3 white
2 black
2 black
1 red
4 red
2 black
2 black
1 red
1 red
Air butt
1 red
009-007709-00
009-007707-00
009-007708-00
GND
GND
VCC
VCC
Control
control
control
VCC
VCC
VCC
1 black
1 black
3 black
2 red
2 red
1 black
2black
2 black
6red
7red
1 red
5red
5 6 7 8 1 2 3 4
1 2
1 2 3 4 J39
J38
J32
C C
J35 J36
J37
1 2 3 4 5 6 1 2 3 4 5 6 1 2 3 4 5 6
B B
3 black
1 black
5 black
1 black
3 black
1 black
4 red
2 red
6 red
2 red
4 red
2 red
009-007710-00
009-007712-00
GND
GND
GND
GND
GND
VCC
VCC
VCC
VCC
VCC
009-007711-00
GND 1 red
GND 1 red
GND 1 red
VCC2 black
VCC2 black
VCC2 black
2 black
2 black
2 black
2 black
2 black
1 red
1 red
1 red
1 red
1 red
Reagent refrigeration fan Whole unit cooling fan Reaction carousel fan Reaction carousel fan Reaction carousel fan Lamp fan
Reagent refrigeration fan Reaction carousel fan
A
MINDRAY A
5 Fluidic system
5.1 Overview
The major function of the fluidic system is:
To provide deionized water for washing interior/exterior of probe and mixer.
To provide deionized water and wash solution for washing cuvettes.
The DI water tank provides DI water for the system and the wash solution tank provides wash solution.
The high concentration waste tank holds the high concentration waste liquid.
This chapter describes the working principles and repairing methods of the hydropneumatic system.
5.2 Fluidic Diagram
See A.7Fluidic Diagram.
FL02 FL03
FL01
DI water tank Diluted wash solution Low conc. waste tank High conc. waste tank
Reaction cuvette
Reaction cuvette
Waste drainage
Mixer cleaning
Probe Wash
dispensing
aspiration
Inlet
Figure 5-2 Principles of Hydropneumatic System
5.3.1 Probe wash module
The probe/mixer wash module consists of one probe, one mixer, two syringes and one diaphragm pump, of
which probe syringe SY03 is 500ul;probe exterior wash syringe SY01 is 10ml;probe interior wash and mixer
wash shares one diaphragm pump P01.
This module is used for fixed-quantity sampling and probe/mixer cleaning.
Fixed-quantity sampling is realized through the syringe SY03, wash valve SV02 and probe NS with the aim
of delivering fixed amount of reagent and sample.
Probe interior wash and mixer wash shares one diaphragm pump P01 with the cooperations of the probe
interior wash valve SV02 and mixer wash valve SV04.
Probe exterior wash uses 10ml syringe SY01.Valve SV03 is turned on or off to control wash operations.
SV01 controls the aspirating and dispensing of the syringe and the wash syringe controls the amount of
the wash solution.
The schematic diagram of the probe wash module is as shown below.
Wash syringe SY01 (10ml) provides DI water for phase 2-3 wash probes. No liquid is injected for phase 4 wash
probe.
The wash solution for phase 1 wash probe is obtained through manual dilution with ratio 1:9(one portion of
concentrated wash solution and nine portions of DI water.) The dispensing and aspirating operations are
controlled by solenoid valve. The dispensing volumes for phase 1-3 are same.
Aspirating module: A diaphragm pump is used to aspirate the waste fluid and wipe the cuvettes after cleaning.
High-concentration waste is generated during phase 1 and low-concentration waste during phase 2-4. Wipe
blocks are provided to absorb the remaining wash solution inside cuvettes during phase 4.
The wash station lowers to the cuvettes and the waste pump is turned on to aspirate the waste in the cuvette
and then the dispensing module injects the diluted wash solution and DI water. The cuvettes are washed
repeatedly for two times.
The schematic diagram of the module is as shown below.
No. Name
1. DI water
2. Wash solution
5 6 7 8 3.
4.
High conc. waste
Low conc. waste 1
5. Liquid level sensor of DI
water
6. Sensor of diluted wash
1 2 3 4 7.
solution
Sensor of high conc. waste
8 Low conc. Waste 2
Name Quantity
Cross screwdriver 1
Hexagon screwdriver 1
Flathead screwdriver 1
Cable tie Several
Diagonal pliers 1
Tube cutter 1
5.5.2 Probe and mixer wash module
Cuvette wash
Syringe
station
Fluidic
interface
Figure 5-7 Assembly drawing of whole unit -- Probe and mixer wash module-front view
Figure 5-8 Assembly drawing of whole unit -- Probe wash module-rear view
Table 5-3 List of materials
No. FRU Part Name Remarks
1 082-000430-00 Syringe 500μl (ZDV) Probe Syringe
2 009-000448-00 Probe interior wash return valve cable probe interior wash
valve
3 082-002273-00 Valve LVMK21-6J cable probe exterior wash
valve
4 009-000448-00 Probe interior wash return valve cable mixer wash valve
5 082-002373-00 LVMK three-way valve cleaning fluid inlet
valve
6 BA30-21-15311 KNF diaphragm pump assembly probe interior wash
pump(mixer wash
pump)
7 115-045917-00 10ml Wash syringe assembly wash syringe
8 M6G-020049--- Tube.PTFEID1.5mmXOD2.5mmX100M sampling tube
9 0040-10-32301 Tube.PTFE,0.040"IDX0.066"OD sampling tube
10 M90-000026--- Tube.PTFE,1/32"IDX1/16"OD mixer wash
restriction tube
11 082-002374-00 Tube.3.2*6.4mm TPU Polyether Wash Tube
12 082-002482-00 3/32"X7/32",tube Wash Tube
13 043-000556-00 Wash well shell Probe wash well
14 BA10-20-77752 Wash well (mould MR77752) mixer wash well
15 115-037085-00 BA24 Sample probe assembly for probe
HbA1c test
16 BA31-20-41651 Mixer FRU/Mixer
Connector of wash
Connector of cleaning
solution syringe tube
fluid syringe tube
Screw
3) Make sure that the power cord is connected to the correct positive and negative ends.
Verification steps
Refer to 7.9 Hydropneumatic Unit.
Screw of bracket
Note
The tubes must be connected in correct order otherwise the analyzer may not work normally. Do not
confuse the inlet tube with the outlet tube.
Take caution not to spill the liquid onto other parts.
The positive and negative of the cable of the solenoid valve must be correctly connected.
After replacement or reparation, conduct fluidic prime in case the air bubbles in the tube affect the
performance of the analyzer.
Note
The tubes must be connected in correct order otherwise the analyzer may not work normally. Take caution
not to spill the liquid onto other parts.
Make sure that the power cord of the pump is connected to the correct positive and negative ends. After
replacement or reparation, conduct fluidic prime in case the air bubbles in the tube affect the performance
of the analyzer.
bracket
screw
Phase 4 waste
pump
Phase 2/3 waste
pump
Phase 1 waste
pump
Probe interior
wash/mixer wash
pump
6) Remove the tubes and apply straps on their openings or place a container under it to prevent liquid
from entering the instrument.
Installation Procedure
1) Connect the tubes according to the marks and then tighten the tube clamps or cable tie. For the
severely twisted tube, cut about 5mm off its end.
2) Install the pump on its bracket and fix the bracket.
3) Make sure that the power cord is connected correctly.
Verification steps
Refer to 7.9 Hydropneumatic Unit.
Note
The tubes must be connected in correct order otherwise the analyzer may not work normally. Take caution
not to spill the liquid onto other parts.
Make sure that the power cord of the pump is connected to the correct positive and negative ends.
After replacement or reparation, conduct fluidic prime in case the air bubbles in the tube affect the
performance of the analyzer.
Note
Cleaning fluid temperature Sensor and Wash solution temperature Sensor:Negative temperature
coefficient, the resistance decreases with increasing temperature.
temperature Resistance
2°C About 11KΩ
25°C About 5KΩ
Take caution not to spill the liquid onto other parts.
Connect the connectors of the tubes and cables correctly otherwise the components or the instrument may
be damaged.
After the tubes are connected, make sure the tubes are tightly connected to the deep of the connector.
After the washing preheater is replaced, prime it first otherwise the instrument may not work normally.
Heating
temperature
switch Of
cleaning fluid
Removing/Reinstalling Filter
When to do
The filter used for more than 1 year.
The filter damage.
Removing Steps
1) There are one filter respectively on the cleaning fluid and wash solution inlet tubes. The filters are
installed on the tubes between the fluidic interfaces and the DI water tank and wash solution tank.
2) Disconnect the tubes at the two ends of the filter and remove the filter.
Installation Procedure
1) Install the new filter and fix it with clamp or cable tie. For the severely twisted tube, cut about 5mm off
its end. Make sure the tubes are tightly connected to the deep of the connector.
2) After the filter is replaced, prime it first otherwise the instrument may not work normally.
Verification steps
Refer to 7.9 Hydropneumatic Unit.
1) Manually loosen the retaining screws on the wash station.See Figure 3-41 Wash probe.
2) Remove the wash station with tubes not connected, put the wash probes in a container.
3) Inject cleaning fluid and wash solution to find the place where the probe is clogged.
4) Use a needle to unclog the probe.
Installation Procedure
1) Align the locating holes on the wash station with the stop studs on the bracket, and then slightly tighten
the retaining screws.
2) Lower the wash probes assembly, use an alignment tool or visually check the phase 4 wipe block in
the cuvette. The wipe block must not contact the interior of the cuvette.
3) After finishing the second step, completely tighten the retaining screws on the wash station.
Verification steps
Refer to 7.9 Hydropneumatic Unit.
Note
Operate carefully to prevent liquid from dropping into the reaction carousel and cuvettes.
6 Software
6.1 Software Installation
6.1.1 Introduction of Installation Package
The installation package of the operating software contains three folders: Setup, SetupGuide, and Thirdparty.
during software installation, find the configuration file in the new directory.
3) The serial port in the configuration file should be set as the one used by the instrument.
Step 1: Open the configuration file COMMUNICATION_USE.INI in the default directory:
D:\Mindray\ BS360\OperationSoft\COMMUNICATION_USE.INI
Step 2: Check if the configurations in the COMMUNICATION_USE.INI file are the same with the following
examples. If they are different, change them according to relevant instructions.
[INSTRUMENTS]
HasTestInstrumentsNum=1
; Note: Number of Instrument
HasSDM=0
; Note: Is SDM configured
IsUseInstrument1=1
IsUseInstrument2=1
IsUseInstrument3=1
IsUseInstrument4=1
NameOfInstrument1 = 36
NameOfInstrument2 = 36
NameOfInstrument3 = 36
NameOfInstrument4 =36
[SAMPRACK_COMM_USE]
SAMPRACK_COMM_TYPE=SERIALPORT
SAMPRACK_COMM_NAME=SIM
; Note: Serial port of SDM
[BAUDRATE_SEC]
Baudrate=38400
; Note: Baud rate of SDM
[Demo]
ComPort=COM1
Step 3: Open the configuration file InstrServerConfig.ini in the default directory:
D:\Mindray\ BS360 \OperationSoft\Instrument_1\ InstrServerConfig.ini
Step 4: Check if the configurations in the InstrServerConfig.ini file are the same with the following examples. If
they are different, change them according to relevant instructions.
[INSTR]
INSTRNO=1
SERVER_IP=127.0.0.1
SERVER_PORT=7001
CLIENT_IP=127.0.0.1
CLIENT_PORT=8001
[SERIALPORT]
COMPORT=COM1
; Note: Serial port
BAUDRATE=38400
; Note: Baud rate
6.4 Software Upgrading
The software upgrade is divided into two parts, operating software upgrades and control system upgrades.
Please refer to the “UpgradeGuide_EN” in the “SetupGuide” folder of the installation package,see Figure 6-1
Folders in operating software installation package ,which will be updated with the software update.
Note:After the upgrade is complete, be sure to turn off the analyzer's power and restart to configure the
instrument parameters
6.4.1 Preparation before Upgrading
Check the software version
Enter the current operating software system on the operating computer to confirm the current software version
information, as follows:
Steps:
Enter the main screen of the operating software, select Utility > System Setup > Instrument Setup > Software
Version, and confirm the current software version information.
Data Backup
Upgrading the software may result in database loss, so please make a backup before upgrading. Please refer
to 6.7.1 Data backup.
2) Sqlserver is started, and it will exit if an error indicating database initialization failure occurs.
3) The software process is started.
4) Enter the username and password.
5) The software shakes hands with the middle-/lower-layer units. (If handshake fails, an alarm is given
prompting the unit is abnormal.)
6.5.2 Software antostart settings
Normally, with the startup of Windows, the software will start automatically. If there are special circumstances
that must prohibit the program from running automatically, you can modify the computer startup items:
How to do:
1) Switch to the desktop using the service user account, Right click the task bar and select Task Manager.
2) Select Startup and the CL-1000i(Here is the CL-1000i as an example. The actual display is workstation)
icon. The screen is shown as follows:
Then the setting parameters are out of range and new setting parameters need to be replaced.
3) If you need to recover by backup parameters, take the following steps: (taking the reaction unit
as an example)Select reaction unit in the unit parameter. 1→Click Search 2→Click Load
3→Select the backup parameter path in the pop-up box→Select the reaction unit parameter
→Click to open . At this time, the parameters have been imported to the current interface,click
7 Alignment
7.1 Basic Operation
7.1.1 Utility - Maintenance
Table 7-1 Utility - Maintenance
No Description
1, 2 Maintenance and Diagnostics: used to maintain and diagnose the
system at user end.
3 Alignment: used to align all mechanical positions, and the
hydropneumatic, pyrology and ISE units.
4 Parameters: used to inquire and configure parameters, and to
export parameters of each unit.
5 SPT: provides basic performance tests for the system, including
photometer test, precision test (dye), precision test (weight),
cuvette residue test, and carryover test.
6 Export parameters: Export alignment parameters of the
mechanical positions.
Note
Before replacing the main control board, you must export the alignment parameters to a local folder on the
Parameter Configuration window.After finishing the replacement, import the parameters to the software.
Otherwise, the system, when started up, will give an alarm indicating key parameters not configured and cannot
be restored. Also change the parameter of fluidic prime status.
2 Access the Alignment window and select Mech Reset to reset the mechanical parts of
the system.
Install operating
Preparations Power on Log on
software
Alignment of photoelectric
Photoelectric
photoelectric signal collecting
Gain
unit position
Preparations
Alignment of Floater/valve Fluidic
before Syringe Check
fluidic unit check Prime
Alignment
Reaction
Alignment of Reaction Temp Wash Solution Reagent
Carousel
Pyrology Unit Sensors Tempe Curve refrigeration test
Heating
End
Reaction carousel
position sensor
Note
Ensure that the screws on the bracket of the position sensor of the reaction
carousel have been tightened.
Alignment parameters can be saved only after each procedure is complete.
Another alignment is required if a procedure is terminated. If the parameter is not
successfully configured or out of range, the parameter range will be displayed by
the alarm message and re-perform the alignment.
If the water blank exceeds the upper limit(Light intensity is too weak), adjust the
position of the fiber Bunch as shown in the figure below and then align again.
BA24-J04
BA60-J07
Probe's orbit
Note
The alignment tools will be used in following alignment. Do not remove them.
The left arrow button is equivalent to rotating clockwise and the right arrow button
rotating counterclockwise.
When selecting Large Step, you can set up the concrete steps according to the
reference range.
Once the probe arm length is determined, it must not be adjusted during alignment
of probes and carousels.
If the probe arm length is changed, the probe and carousel units should be aligned
again.
Alignment parameters can be saved only after each procedure is complete.
Another alignment is required if a procedure is terminated. If the parameter is not
successfully configured or out of range, the parameter range will be displayed by
the alarm message and re-perform the alignment.
If you loosen the two fixing hex socket screws of the probe rocker arm, you need
to adjust the horizontal position of the probe to the reaction carousel, the wash well
and the reagent carousel. If the debugging is out of range, it is possible to rotate
when the rocker arm is fixed. Over the rocker arm, you need to re-adjust and fix
the rocker arm.
BA48-J10
Note
When the alignment is complete, there is no need to take out the alignment tool.
Alignment parameters can be saved only after each procedure is complete.
Another alignment is required if a procedure is terminated. If the parameter is not
successfully configured or out of range, the parameter range will be displayed by
the alarm message and re-perform the alignment.
BA24-J04
The probe tip can enter the central hole of the alignment fixture (BA48-J10) in 1#, 19# and 37# positions of the
sample/reagent carousel middle ring.
Alignment methods and procedure:
1) Select Probe Rotary to Sample/Reagent Carousel Middle Ring.
2) Place a sample position alignment tool (BA10-J16) in position #1, #19 and #37 on the middle ring of
the sample/reagent carousel, and then select Continue.
3) Select the up/down and left/right arrow buttons to move the probe so that it can enter into the central
hole on the alignment tool.
4) Select Continue to save parameters and exit the window.
7.5.5 Probe to Sample/Reagent Carousel Inner Ring
Alignment index:
The probe tip can enter the central hole of the alignment fixture (BA48-J16) in 51#, 69# and 87# positions of the
sample/reagent carousel inner ring.
Alignment methods and procedure:
1) Select Probe Rotary to Sample/Reagent Carousel Inner Ring.
2) Place a reagent position alignment tool (BA10-J16) in position #51, #69 and #87 on the inner ring of
the sample/reagent carousel, and then select Continue.
3) Select the up/down and left/right arrow buttons to move the probe so that it can enter into the central
hole on the alignment tool.
4) Select Continue to save parameters and exit the window.
7.5.6 Probe to Vertical Limit Position on Reaction Carousel
Alignment index:
When reaching the cuvette bottom, the probe is lifted for 0.15mm, which means that the clearance gauge can
pass with 0.15mm through the gap between the sensor plate and the arm base and it cannot pass with 0.20mm.
Alignment methods and procedure:
When reaching the bottom of the reagent bottle, the probe is lifted for 0.15mm, which means that the clearance
gauge can pass with 0.15mm through the gap between the sensor plate and the arm base and it cannot pass
with 0.20mm.
Alignment methods and procedure:
1) Select Probe to Vertical Limit Position in Reagent Bottle.
2) Load the reagent bottle in 1#, 19# and 37# position on the middle ring of the sample/reagent carousel.
3) Adjust the probe over the 3 reagent positions, and ensure that the lifting height meets the requirement.
4) Select Continue to save parameters and exit the window.
7.5.8 Probe to Horizontal Postion on ISE Part (Optional)
Alignment index:
The probe has its tip above the ISE sample injection port but deviates slightly from the central hole. The probe
tip should not contact with the inwall of the sample inject port.
Alignment methods and procedure:
Note
Alignment of each position is verified the next step.
Alignment parameters can be saved only after each procedure is complete.
Another alignment is required if a procedure is terminated.
If the parameter is not successfully configured or out of range, the parameter range
will be displayed by the alarm message and re-perform the alignment.
BA48-J16:
Alignment
lever
(inserted to
the end)
BA24-J05
BA40-J76
Bar
Sample tube
code
area
9mm
2) Use cross screw driver to remove the tube holder in 14# and 15# position on the outer ring and place
the bar code alignment tool BA24-J03(The slot facing the outside of the carousel. )Click Continue.
Note
Check that the sponge cushion seals the bar code reader completely without
blocking the scanning light. Otherwise, adjust the bar code reader again.
After alignment, check if the screws on the bar code reader and conversion board
have been tightened.
inserted with the same depth and not twisted, and all tubes are smooth without crossover or protrusion.
3) Check if the solenoid valves have been connected to the correct cables.
4) The diluted wash solution tank holds sufficient diluted wash solution.
Cuvette Type Dilution ratio of diluted wash solution 1
Plastic Cuvette 9 portions of DI water:1 portion of concentrated
wash solution
5) Fill the water tank with sufficient water.
6) Connect the water supply tube to the analyzer, and connect the high-/low-concentration waste tanks.
7) Check if the inlet filter assembly has been installed.
8) Remove the pseudo probe from the probe arm, and install the real probe instead.
Installation requirements:
1) Gap between the probe plate and the arm surface is not allowed. Otherwise, check the retaining bolt
and tube, because the tube's elasticity will also affect the probe's bouncing;
2) A washer has been applied on the retaining bolt, and the nut has been tightened;
3) After fixing the probe with bolt and spring, manually lift the probe and release it, and check that the
probe can bounce back freely through the spring;
4) Adjust the level sense board and the relative positions of the probe plate and photocoupler so that the
probe plate is in the middle of the photocoupler.
7.9.2 Checking Floater and Valve Status
5) Click Continue. According to software screen, perform mixer exterior wash until the mixer wash well
can spray water continuously. Click Stop and OK to move to the next step.
b) The mixer outer wall wash tube D22 can be properly corrected and retested until the requirements are
met. The D22 tube correction method is as follows:
Flow(mL/min) Tube length(cm)
>120 60
90~100 30
80~90 25
c) If there is no problem with the piping, replace P01.
7.9.6 Probe interior Wash Flow Test
Alignment index:>40ml/min,no liquid sputtering.
Tools: measuring bottle,empty container
Alignment methods and procedure:
1) Select Utility—>Maintenance—>Alignment—>Hydro Unit—> Probe interior Wash Flow Test.
Place a container below the probe.
7.9.7 the volume of the reaction liquid during cuvette washing process
Alignment index: Afte the Auto Wash is cleaned, the amount of liquid remaining in the reaction cup, the
liquid level is 5-10 mm from the reaction cup opening.
Alignment methods and procedure:
1) Select Utility -> Maintenance -> Alignment -> Pyrology Unit -> Reaction Carousel Temperature
Curve. The window is as shown below.
2) According to the sequence number on the temperature sensor, contact the headquarters to find the
resistance of rated temperature points (0, 37 and 100), select sensor 1 and input the resistance for
each temperature point, and then select Calculate.
3) When a dialog box pops up indicating qualified sensor, select Configure to configure the parameters
to the relevant sensor.
4) Observe the △AD value of PCB, and input it in the edit boxes at lower-right corner of the window for
sensor 1. Select Configure, and then select Search to verify them.
5) When configuration is complete, select Exit.
d) If the current data display area returns calibrated data as <ERC PMC x000000>. For x, see Appendix.
The next line displays "specific error" (marked with red text). For error codes, see Table 7-3 ISE Error
Codes.
Wash:
Prepara sufficient wash solution in the sample carousel (usage in each wash operation is 100 ul), and then input
the level of the wash solution in the cuvette in "Set liquid level in cuvette", and click "Clean". Wait until the data
area returns <ISE!> (wait for about 160 s). The next line displays "Clean completed", including the wash is
complete.
Two-point calibration:
a) Click "Start" in the "Two-point calibration" area to display the confirmation dialog for fluidic tube wash.
If it is possible to check that the last wash operation has been waited for more than 15 min, directly
click OK; otherwise, wait until the countdown time ends.
b) After the countdown time ends, automatically execute the two-point calibration testing. After the testing
is complete, the current data display area returns the calibration result "< Na xx.xx K xx.xx Cl xx.xx
xxxxxxxx>". ;
c) If an error code displays during calibration, repeat several times of two-point calibration. If the problem
persists, replace the corresponding electrode reporting error as instructed by error message. For error
codes, see Table 7-3 ISE Error Codes.
Cl-,K+ 0 A A A A A A
Cl-,K+,Na+ 0 C C C C C C
The error types in the table above are described as below:
1) - Air bubble/hardware;
2) - Electrode voltage overflowed during calibrator B calibration or blood measurement;
3) - Electrode voltage overflowed during two-point calibration or calibrator A calibration in blood
measurement mode, or
4) calibrator B calibration in urine measurement mode;
5) - Electrode voltage noise during calibrator B calibration or blood measurement;
6) - Electrode voltage noise during two-point calibration or calibrator A calibration in blood measurement
mode, or
7) calibrator B calibration in urine measurement mode;
8) - Voltage drift during calibrator A calibration in blood measurement mode, or slope drift
9) during two-point calibration;
10) - Exceeds the slope or module measurement range;
8 Installation
8.1 Installation Procedure
8.1.1 Installation Procedure
The approximate installation procedure of the analyzer is as follows: for your reference:
Table 8-1 Analyzer installation procedure
Classification Main Content Remarks
Before Check if the analyzer and reagents are arrived by telephone, The LIS and water unit
installation check whether the water unit, UPS, LIS, etc. are prepared in are not required
advance, confirm the trainees, and confirm that the arrival
reagents have been refrigerated.
Prepare tools and alignment fixtures
Download relevant information on TDP: latest software (if Please search for the
new software is released, and the version is higher than the specific file path on the
factory analyzer); download the maintenance manual for TDP
backup
Once again, confirm that the above is correct.
During Check the quantity of the goods and confirm whether the In case of any missing,
installation installed reagents are refrigerated. timely give feedback
Checking Packing List
Unpacking Keep boxes and screws
for three months
Transfer the analyzer to the department installation location
Remove the analyzer fixed parts and foam tape Please save the fixed
parts being removed.
Install accessories
Connect the pipeline
Install the PC
Software Installation
Start the analyzer
Adjust the positions
Prime the fluidic system
Set the analyzer, and initialize maintenance records
Apply for testing for performance verification
Record data, and fill in the installation report
Assist the LIS in connection (if any)
Train the customer for operations and simple troubleshooting
Print template settings (refer to the corresponding chapter of
the instruction manual)
Confirm that the analyzer works normally, and the pipeline
and wire are neat and beautiful.
After confirming the above items are correct, accept them
and sign your name
After Regularly visit the customer after installation (not more than The customers whose
installation one week), and take the initiative to care analyzers are newly
installed have more
operational problems
Wall
Min. 500
Operation Unit
660
Analyzing Unit
Front
860
Min. 500
WARNING
Make sure the power socket is grounded correctly. Incorrect grounding may lead
to electric shock or equipment damage.
Measurement of ground voltage:
First confirm that the neutral and ground wires cannot be shorted,set the
multimeter to the AC 250V scale, connect the black probe to the earth wire and the
red probe to the live wire and neutral wire. If grounding is proper, the voltage
between earth wire and neutral wire should be less than 5V, and the voltage
between earth wire and live wire should be similar to that between live wire and
neutral wire.
Check if the power socket outputs voltage meeting the specified requirements and
has a proper fuse installed.
Biohazards
Dispose of liquid waste according to the local regulations.
CAUTION
The supplied water must meet the requirements of water quality; otherwise, unqualified water may result in
incorrect test results.
Figure 8-6 The leveling feet of the main unit fit the grooves in the bracket
If you cannot
loosen the
fixing screw
of the
sample
probe, you
can use a
flat-blade
screwdriver.
Figure 8-10 Remove the arm shell; loosen the screw that secure the sample probe
2) Insert the sample probe from the top and bottom into the probe hole on the arm of the sample
probe, and align the hole on the shutter of the sample probe to the column in the arm cavity.
3) Check again that the washer is inside the metal connector of the sample probe. If so, align the
tube connector with the connector of the sample probe and tighten it clockwise. If not, find it and
install it in the metal connector before connection.
Figure 8-11 Confirm the shutter of the sample probe in the middle of the sensor and connect the plug
5) Insert the spring into the column in the arm cavity and tighten the screw compression spring.
8) Pinch the sample probe by the part near the probe arm. Push the sample probe upwards and
then release it to check if the spring works well. If it does, proceed to the next step. It not, it
indicates that the arm cover is not installed correctly. Reinstall the arm cover and check the spring
until it can move freely.
Keep the
sections flush.
Rotate the retaining screw clockwise.
If there are no fixtures on site, refer to Appendix A.2BS-360E Moving Parts Position Confirmation Guide (No
Alignment Tooling) for alignment.
8.4.10 Aligning the ISE (Optional)
If the analyzer is equipped with an ISE module, please align the ISE by referring to 14.2.3 Alignment of the ISE
module.
8.4.11 Prime the fluidic system
Prime the fluidic system for the newly installed analyzer by referring to 7.9.4 Fluidic Prime.
8.4.12 Initial Maintenance Record
Select Utility -> Maintenance -> Maintenance to display the maintenance window. Confirm all scheduled
maintenance procedures and set up start time for them. The maintenance frequencies include Daily, Weekly,
Two-Week, Monthly, Three-Month, Six-Month, and Other. Select the Select All button on each maintenance
frequency tab page, and then select OK. The current date and time appears in the Date Performed column of
each maintenance procedure as shown in the figure below.
Figure 8-26 Load Wash D and Saline W
3) Load the Water reagent
In the sequence shown below, enter the Reagent/Calibration screen, select the reagent
position 1#, click Load, select the chemistry Water in the pop-up screen, and click Load.
Then click Exit, and click Load Complete. At the same time, fill the water reagent pack with
deionized water (not less than 10 ml) and place it in the 1# position of the reagent carousel.
(This manual takes 1# position as an example, and other positions can be selected in
practice)
2) The vertical movement of the sample probe syringe is normal, without abnormal sound; the
position of the auto wash station down to the cuvette should be not rubbed, and should be
located in the center of the cuvette.
3) If any of the units is positioned abnormally, select UtilityMaintenanceAlignment, and
select the corresponding unit module to align the position parameters. The alignment result
is recorded in the BS-360E_Installation Acceptance Report.
8.5.3 Testing Optical Performance
1) Select UtilityMaintenanceBiochemistry MaintenancePhotometer Check. The
absorbance of each wavelength of the newly installed analyzer should be in the range of
5000 - 8000. If it exceeds 8000, please check whether the installation of the light source
lamp is in place.
2) Select UtilityMaintenanceBiochemistry MaintenanceCuvette Check. The new
analyzer should have no red marked cuvette, if there is, please replace the red marked
cuvette.
8.5.4 Testing Bar Code Performance
Select UtilityMaintenanceAlignmentSample/Reagent Bar CodeBar Code
Scanning Stability Test. Put a valid barcode (the installation reagent is carried with a
barcode, which can be placed in the sample/reagent carousel for scanning). The analyzer
can accurately read the barcode. Please refer to 7.8.4 Bar Code Stability Test.
9 Preventive maintenance
9.1 Overview
9.1.1 Preventive maintenance
Preventive maintenance is a preventive measure taken by service personnel of Mindray or authorized by
Mindray with the aim of eliminating hidden troubles to ensure system reliability and achieve best performance
during operation. The Preventive maintenance cycle for the chemistry analyzer is one year and for ISE module,
half a year. The maintenance includes changing, cleaning or checking the parts as shown in the table below.
Table 9-1 Maintenance items to be performed by service personnel
Category Procedure When to do
Replacement Item Replacing Lamp 2000 hours,or more than 6
months
Replacing cuvettes More than 3 months
Replace Water Inlet Filter More than 1 year
Replace Diluted Wash Solution More than 1 year
Filter
Replace Syringe over 100,000 times
Replace Calibrator Tube(ISE More than 1 year
configured)
Replace Pump Tube(ISE More than half year
configured)
Replace the ISE electrode(ISE More than 6 months or
configured) over 10,000 tests
Replace Probe Washer More than 1 year
Clean Item Clean Dust Screens When needed
Clean the dust of the heatsink When needed
fan
Clean Wash Well When needed
Clean Cuvette Wash Station When needed
and Tubes
Clean ISE Injection Port (ISE When needed
Configured)
Clean DI Water Tank When needed
Clean Diluted Wash Solution When needed
Tank
Clean Sample/Reagent When needed
Compartment
Clean Analyzer Panels When needed
Special Wash When needed
Clean Probe Exterior When needed
Clean Mixers When needed
Photometer Lens Maintenance More than 1 year
Check Item Pump Calibration (ISE When needed
Configured)
Air Bubble Detector When needed
Calibration(ISE configured)
Cuvette Check When needed
Photometer Check When needed
Check Waste Pump When needed
Replace Cleaning Fluid Syringe When needed
Replace Wash Solution Syringe When needed
9.1.2 Biochemistry maintenance commands
Table 9-2 Biochemistry maintenance commands
Replace Syringe
1) Check the using count of the syringe on the status screen. When the syringe has been
used for 100,000times, change it.
2) Prepare a new syringe plunger assembly and washer, put the plunger head in the deionized
water beaker to remove air from the syringe, and then moisten the washer in the deionized
water.
3) Place the analyzing unit power to the OFF position.
4) Loosen the screw on the shield of the syringe.
5) Loosen counterclockwise the retaining screw at the bottom of the syringe and then remove
it.
6) Loosen counterclockwise the four retaining screws on top of the syringe, and then remove
the screws and the fixing blocks
7) Hold the T piece with one hand and the syringe connector with the other hand. Loosen the
syringe counterclockwise. And then remove the washer.
8) Loosen the plunger guide cap counterclockwise, hold the plunger head and pull it slightly to
remove the plunger assembly from the syringe.
9) Insert the plunger head of the new plunger assembly into the bottom of the syringe, and
then tighten the retaining screw to fix the plunger head. Tighten the plunger nut clockwise
10) Soak the new syringe connector in the deionized water beak, pull gently the plunger head
to aspirate half syringe of deionized water, and then push the plunger head to remove the
air.
11) If no washer is found in the T piece, put the new washer in the T piece. Hold the T piece
with one hand and the syringe connector with the other hand, and then screw the T piece
clockwise.
Clean Mixers
1) Place the analyzing unit power to the OFF position.
2) Lift gently the mixer to the utmost height and move it to the position convenient for
maintenance.
3) Use gauze soaked with ethanol to gently wipe the mixer exterior. Clean it until it becomes
clear without stain. Use gauze moistened with deionized water to clear the ethanol on the
mixer. Do not pull the mixer vertically to prevent probe damage.
4) After the procedure, restore the upper protective shield of the analyzer. Power on the
analyzing unit and select Home.
5) After maintenance, open the logs and record related information.
Special Wash
1) Open the upper protective shield of the analyzer.
2) Place 40ml concentrated wash solution on the No.49 position of the reagent carousel.
3) Select Utility-Maintenance-Maintenance- Biochemistry Maintenance and select special
wash.
4) Confirm if cuvette check is needed after the concentrated wash. If it is, mark the checkbox
in front of Check Cuvettes.
5) Select Continue. The system starts cleaning the probe, mixer, cuvettes and wash station.
When special wash is finished, the system performs cuvette check automatically.
6) Select Done.
7) Restore the upper protective shield of the analyzer.
8) After maintenance, open the logs and record related information.
Cuvette Check
1) Select Utility-Maintenance-Maintenance- Biochemistry Maintenance and then select
Cuvette check.
2) After check, check the cuvette check results for yellow indication. Record the cuvettes
highlighted in yellow and perform the Clean Cuvettes or Replace Cuvettes procedure.
3) After maintenance, open the logs and record related information.
Photometer Check
1) Select Utility-Maintenance-Maintenance- Biochemistry Maintenance and then select
Photometer Check.
2) If the result is Normal, the system will give no alarms, otherwise the following alarm will be
given.
a) If the alarm is “Lamp is not turned on”,refer to 10.2.10 Lamp Is Not Turned On C07005
b) If the alarm is “Light intensity is too strong”,refer to 10.2.11 Light Intensity Is Too Strong
C07006.
c) If the alarm is “Light intensity is too weak”,refer to10.2.8 Light Intensity Is Too Weak
C07003 .
10 Troubleshooting
This chapter describes how to locate failure and determine relevant corrective actions.
10.1 Overview
10.1.1 Classification of logs
The fault log records all types of faults that occur with various parts of the instrument. All faults are classified by
part as shown in the following table.
Error logs
Each error has a unit code used for identification and locating probable causes and solutions. An error code
consists of 6 letters and numbers, such as "C01001", in which "C" indicates that the error occurs on the operation
unit, "01" is the error description of instrument connection, and "001" is the serial number of the error. Therefore,
"C01001" is described as "the first error of instrument connection on the operation unit".
The following tables provide a summary of error codes for the operation unit and analyzing unit.
Table 10-2 Error code of the operation unit
11-Wash unit
12-Temperature unit
21-Probe interior wash unit
22-Home process or ISE unit
000~999 Serial number of the error
Recalling logs
Error logs can be recalled by all users in any system status. Error logs can be recalled by date, event ID,
symptom and event class.
Perform the following steps to recall desired event logs:
1) Select Alarm > Error Log .
2) Select Search F1.
3) Enter one or more of the following conditions:
Date
Event ID (available for error logs only)
Error indications
Errors may occur on hardware, software and the entire system. When an error occurs, it will be indicated in
many ways to help identify it and determine the possible causes and solutions. Errors can be indicated by
alarm tone, alarm message, color, alarm message box, result flag and error log, through which you will obtain
detailed information about errors and find the relevant solutions.
Alarm tone
When an error occurs, the buzzer gives alarm tone reminding you to notice the error and take corrective actions.
Alarm tone can be adjusted manually or silenced.
Perform the following steps to adjust the alarm tone:
1) Select Utility > System Setup.
2) Adjust the alarm tone in the Alarm Volume field.
3) Test the alarm tone until it is satisfied.
4) To silence the alarm tone, drag the slider to the leftmost position of the scale.
5) Select Save F8 to save the adjustment.
Alarm message
When an error occurs, the system gives an alarm and displays the alarm message in the second line of the
prompt message area.
Color highlight
An error will be indicated by highlighting relevant buttons and screen texts with different colors. Yellow indicates
a warning, and red indicates a serious warning or error.
Reagent button
Utility button
Alarm button
Select a button to access relevant function page, check for abnormities and take corrective actions. When the
problem is solved, the alarm indication disappears.
Alarm message box
An error can also be shown in an alarm message box, which contains the date/time, event ID, time(s) and help
icon.
Errors that are indicated through an alarm message box are divided into the following types:
Common error: including those that are indicated by warning the user, and by invalidating tests,
reagents and samples. When such error occurs, the alarm message box shows with the title bar
highlighted in yellow.
Serious error: including those except for the common error. When such error occurs, the alarm
message box shows with the title bar highlighted in red, and you are only allowed to reboot or exit the
system.
When an alarm message box appears, select the Alarm button to view the new error logs, analyze the possible
causes and determine relevant corrective actions.
Flag
Flag is also called data alarm. When calibration error or failure, or sample result error occurs due to the sample,
reagent or system failure, a flag will appear near the corresponding calibration result or sample results.
Error log
All alarms are recorded in the error logs. By recalling the error logs you are enabled to master the current status
of the system and troubleshoot errors.
Identifying errors
To identify errors, understand the error indication thoroughly, check the error logs and system status, and then
determine relevant solutions.
The table below shows the error types that may occur on the system. Find relevant corrective actions according
to the description.
Table 10-4 Error types
1)Confirm that the current account of the system is the administrator account by referring to 6.2.8
Confirmation of Administrator Permission and UAC.
2)Select the software startup program, right-click Properties→Compatibility, and confirm that
Open in Compatibility View is not ticked.
3)Back up the database file Database. Please refer to 6.7.1 Data backup.
4)Enter the Database folder in the software installation directory. The default path:
D:\Mindray\BS360\OperationSoft\DataBase. Only keep the Backup folder. After the rest of the files are
deleted, start the software. If the operating software can be entered normally, the current database file
is wrong, if not, proceed to the next step.
5)Uninstall the SQL Database. Please refer to 6.6.2 SQL Database Uninstalling .
6)Reinstall the software. The software will automatically install the database during the installation
procedure. Refer to 6.3 Installation of Operating Software.
7)Reinstall the operating system. For domestic customers, please use the backup GHOST file in
the E drive for recovery.
8)Repeat step 6.
10.2.2 Database backup failed
Symptom: Analyzer alarm database backup failed
Fault Mechanism: this alarm is generated if the database cannot be backed up when the analyzer
is turned on.
Fault Scenario: this alarm is generated when the software is turned on
7.5.6 Probe to Vertical Limit Position on Reaction Carousel in the alignment guide, if the positions are
correct, please proceed to the next step.
3) Check the verticality of the sample probe, visually measure it, or use a vertical reference for
comparison. If the sample probe is not straight, please replace the sample probe.
Refer to
4) Check if the analyzer is falsely alarmed due to interference:
a) Confirm that the analyzer has a grounding wire and the zero ground voltage is <5V. The
measuring method is: set the multimeter to AC 250V voltage, and connect the black test lead to
the ground wire jack, and the red test lead to the live wire/zero wire jack, respectively. When the
grounding is good, the voltage between the ground wire and the zero wire is less than 5V, and
the voltage between the ground wire and the live wire is similar to the voltage between the live
wire and the zero wire.
b) Check whether the connection line of the level detection board is damaged, or is tied too tightly
by the cable tie. Check if the board connector is loose. The specific operation method is
Power off the whole analyzer, open the protective cover, remove 2 hardened dusting screws (M3X6), remove
the arm cover of the sample probe (DS193), refer to A.2BS-360E Moving Parts Position Confirmation Guide
(No Alignment Tooling)Judging and Calibrating Verticality of Probe:
disconnect the J1. J2. and J3 connectors of the level detection board, and re-connect them. If the connection
wire of the J2 connector to the J11 connector of the main control board is damaged, wind it with an electrical
tape. If the cable tie is too tight, loosen the cable tie and re-plug the J11 connector.
c) If other level detection alarms appear at the same time, the level detection board 051-002479-00
is suspected and should be replaced.
5) Check the residual water source of the cuvette. There are two general conditions for the overflow of
the wash station: one is that the liquid absorption is not smooth, and the other is excessive liquid
dispensing. The first case is more common, please refer to the section 10.2.9 for specific steps3) in
section 10.2.9..
10.2.6 Bar Code Scanning Faults
The barcode scanning problems are generally divided into two types:
1) Software alarm, with specific alarm information. For example: A09011, the reagent barcode device is
not working properly; C04001, the sample barcode is repeated. Sample ID/bar code: %s Repeated
position 1: %s Repeated position 2: %s. For such alarms, please refer to Table 10-6 Error messages
and corrective actions for the cause description and procedure to process them.
2) There is no alarm in the software, but the reagent or sample barcode placed in the sample/reagent
carousel is not recognized.
The procedure below is for the second case.
Procedure:
Scanning reagent bar code failed.
1) Confirm the client environment and usage habits:
If the client environment is under high humidity condition, with too much condensed water, it will affect the
scanning of the reagent barcode. The fault is that some reagent barcodes cannot be scanned occasionally.
Customers are advised to install air conditioners to ensure air humidity.
The client usually puts the reagent carousel in the refrigerator and then shuts down the analyzer. This operation
may cause the reagent barcode to be unscanned. There are two ways to improve it: Method 1: the customer is
recommended not to take the reagent carousel out. Leave it in the analyzer, keep the power of the analyzer on
at night, and plug the reagent carousel cover. Method 2: Remove the reagent carousel at night and put it in the
refrigerator. The analyzer can be powered off. The next morning, power on the analyzer first to start the Reagent
Refrigeration Unit, and after the Reagent Refrigeration Unit is stable (about 5° under normal temperature
conditions), take the reagent carousel out from the refrigerator, and place it in the reagent compartment as
quickly as possible to reduce the contact time with the outside air.
2) Check if there is too much condensed water in the sample/reagent carousel, if the barcode window is
dirty or if there is condensed water, please clean it.
3) View the reagent scanning situations, which are divided into the following types:
a) All reagents cannot be scanned:
Check whether the barcode scanner has red light. If not, check whether the J5 connection line
form the reagent barcode scanner to the main control board is normal, re-plug and check if the
connection line is damaged or is tied too tightly by the cable tie. To replace the barcode scanner,
please refer to 3.4.10 Replacing Bar Code Reader. If the replacement of the main control board
is invalid, replace the barcode scanner (this is rare).
Note: If all reagents cannot be scanned, the chemistries may not be imported, or the chemistries
are wrong.
b) The chemistries cannot be scanned.
Try to manually load the reagent position. Operation method: enter the software, Reagent
Reagent Calibration, select the position where the reagent is placed, click Reagent Load,
input the barcode on the reagent pack at the barcode field in the pop-up screen. Note: Please
keep in mind that the barcord is case sensitive.
If the reagent can be loaded successfully, the barcode may have water mist or scratches or poor
adhesion.
If the reagent cannot be loaded successfully, it may be due to a chemistry problem. Re-import the
chemistry.
c) The barcode occasionally cannot be scanned; the position is not fixed.
Enter the software, align the barcode scanning position by referring to 7.8 Bar Code Unit
(Optional).
If it still cannot be solved after alignment, please replace the barcode scanner.
10.2.7 Cuvette Blank Out of Range C07004
Error Code: C07004
Fault Information: cuvette blank out of range. Cuvette No.: XX
Fault Mechanism: the 6th phase water blank is less than the lower limit of the allowable light intensity
alarm
Note: the 6th phase here is actually the 2nd phase of the wash station (facing the analyzer, from left to right),
and the 360E wash station has only four phases.
Possible Cause:
The cuvette is polluted.
The lamp is aging.
The lamp is not installed correctly.
The wash station dispenses liquid incorrectly.
The wash station is overflowed.
The optical lens is dirty.
The light splitter lens assembly is abnormal.
Procedure:
1) Check the alarm information and determine the alarm cuvette No. If the interior of the cuvette is dirty,
please perform the cuvette detection and special wash by referring to Cuvette Check and Special
Wash. If the interior of the cuvette cannot be cleaned or the exterior is dirty, follow the steps in 3.3.7
Replacing Reaction Cuvette to replace the abnormal cuvette.
2) If the lamp is used for more than 2000 hours or 6 months, replace it. Refer to3.9.3 Replacing Lamp.
3) Check if the wash station overflows. If overflow occurs, refer to 3) in section 10.2.9.
4) Check the liquid dispensing volume in the 6th phase of the wash station and check the liquid volume
left in the cuvette according to7.9.7 the volume of the reaction liquid during cuvette washing process.
If the volume of liquid stored in the 6th phase is too small or there are bubbles, focus on the tubes and
connectors between SY01 and the wash probe. There may be leakage or blockage, please refer to
A.7Fluidic Diagram.
10.2.8 Light Intensity Is Too Weak C07003
Error Code: C07003
Fault Information: light intensity is too weak
Fault Mechanism: this alarm is triggered after Cuvette blank out of range occurs successively
Possible Cause: refer to 10.2.7 .
Procedure: refer to 10.2.7 .
10.2.9 Water Blank Out of Range (10X) C07009
Error Code: C07009
Fault Information: water blank out of range (10X)
Fault Mechanism: when the lamp fluctuates or the wash station overflows, the 340 nm water blank (light
intensity) of the 6th phase in the same cuvette will change significantly compared with the previous water blank.
If the changes in the 340 nm water blank (light intensity) of the 6th phase in 10 consecutive cuvettes exceed
the alarm threshold compared with the previous water blank, the alarm Water blank out of range (10X) is
generated.
Note: the 6th phase here is actually the 2nd phase of the wash station (facing the analyzer, from left to right),
and the 360E wash station has only four phases.
Possible Cause:
e) If the 2nd and 3rd-phase overflows occur, focus on whether the external low-concentration waste
tube is blocked, check whether the W17, W16, W14, and W12 tubes are loosened or blocked,
and then use the replacement method to check whether the P03 is damaged. The W09 and W10
tubes are seldom blocked at the same time as and the 2nd and 3rd phase wash probe, but
investigation is required.
ZQ01
W09
W10
W11
T36
W12
SI1 SI2 SI3 SI4
T39 T38 T37
ZZ10
SV07 SV05
W15 W16 ZQ03
D27 D28
ZZ04 Wash
solution
preheating Low- High-
concentrat concentration
From Diluted Wash ion waste waste
Solution Tank
4) Check the liquid dispensing volume in the 6th phase of the wash station and check the liquid volume
left in the cuvette according to 7.9.7 the volume of the reaction liquid during cuvette washing process..
If the volume of liquid stored in the 6th order is too small or there are bubbles, focus on the tubes and
connectors between SY01 and the wash probe. There may be leakage or blockage, please refer
toA.7Fluidic Diagram.
10.2.10 Lamp Is Not Turned On C07005
Error Code: C07005
Fault Information: lamp is not turned on
Fault Mechanism: after the lamp is tuned on, the detected light intensity is less than the lower alarm limit.
Fault Scenario: the error is reported when the analyzer is running, or is turned on.
Possible Cause(s)
The lamp is damaged.
The lamp post is not tightened.
The power board of the lamp is not connected properly.
The power supply of the analyzing unit is disconnected.
The wash station overflows.
The light splitter lens assembly is faulty.
Procedure:
1) Open the reaction carousel cover and check if the lamp is turned on. If it is not, rerun the operating
software.
2) At the same time, check if the wash station overflows. If overflow occurs, refer to 3) in section 10.2.9.
3) Open the upper panel of the lamp to confirm that the post nut is tightened. Refer to 3.9.3 Replacing
Lamp.
4) Replace the lamp.
5) Re-plug the J5 connector of the DCDC power board and check the wire of the J5 connector to the
lamp for damage.
6) Enter the Utility screen —> Maintenance—>Alignment —>Photometric Unit. Click Turn On Lamp,
turn the multimeter to the DC gear, and measure the J5 connector on the DCDC power board. If it is
not in the range of 11.4 -12.6 V, replace the board.
7) Check if the fiber is damaged or unable to conduct light. When one end of the fiber is illuminated at
the lamp, the light will be seen at the other end of the fiber.
If the fiber is damaged, replace it. Refer to 3.9.4 Replacing Fiber Bundle.
8) Check if there is any foreign matter in the lens of the photometer in the reaction carousel. If there is,
remove it according to3.9.7 Photometer Lens Maintenance.
9) Replace the light splitter lens assembly 115-041267-00 by referring to 3.9.5 Replacing Light-splitting
Assembly..
10.2.11 Light Intensity Is Too Strong C07006
Error Code: C07006
Fault Information: light intensity is too strong
Fault Mechanism: After the lamp is tuned on, the detected light intensity is greater than the upper alarm
limit
Fault Scenario: The error is reported when the analyzer is running or is turned on
Possible Cause:
Fluctuation of light source or lamp problem
Cuvette uninstalled
The wash station is overflowed.
Photoelectric gain adjustment too high
Incorrect fiber splice position
Faulty light splitter lens assembly
Procedure:
1) For lamp problems, if the lamp is used for more than 2000 hours or 6 months or a non-original bulb is
used, replace it. Refer to 3.9.3 Replacing Lamp.
2) Check if the cuvette is installed, or if the wash station overflows. Refer to 3) in section 10.2.9.
3) Enter the Utility->Maintenance->Parameter Search and Configuration screen of the operating
software, select Reaction Carousel Unit, and view
the photoelectric gain value of each wavelength, which is normally 31 to 255. The larger the value, the smaller
the measured AD value and the higher the absorbance.
There are two ways to adjust the value. Method a) is recommended.
a)Refer to 7.3.3 Photoelectric Gain Adjustment for auto adjustment of photoelectric gain. The value
will change automatically after adjustment.
b)Re-configure the gain value by referring to 6.7.4 Parameter Configuration.
4) If the gain value has been adjusted to 255 and the alarm still exists, try to adjust the fiber position,
loosen the fiber splice fixing screw at the reaction carousel, and move the fiber along the lamp direction.
Please refer to Figure 7-8 Adjust the fiber bunch position.
5) Replace the light splitter lens assembly 115-041267-00 by referring to 3.9.5 Replacing Light-splitting
Assembly.
10.2.12 Equipment cannot be connected
There are several types of faults when the analyzer cannot be connected:
Software alarm: Fault code: C01001 Fault Information: the analyzer cannot be connected
An error occurred while you enter the software. The prompt: Inquiring instrument configuration failed.
The following information is displayed on the screen:
Possible Cause:
The analyzer main power is off
Faulty power switch
PC serial port abnormal
Serial cable abnormal
Online configuration file error
Main control board abnormal
Procedure:
1) Check that the main switch of the analyzer and the switch of the analyzing unit are turned on. Normally,
the power switch indicator should be always on. If the switch is not on, please check and reconnect
the power cord of the analyzer, and confirm that the fuse at the power outlet is intact. Please refer to
Replacing the fuse. If it is confirmed that there is no problem with the power supply, replace the
switch.Refer to 4.5.4 Replacing Power swich and fuse
2) Check that the analyzer is connected to the computer, re-plug the serial cable, and check whether the
serial cable is normal.
Serial cable inspection mode: The serial cable is the RS232 direct connection, connect to 2, 3, 4, 5,
and 7 directly. Use the multimeter to adjust to the buzzer gear and measure whether the 2, 3, 4, 5,
and 7 at both ends of the serial cable are connected. If not, replace the serial cable.
3) View the serial port status of the computer, exit to the desktop of the computer, right click on the
computer, select the device manager, and view the port. Confirm that there is a COM port on the
computer, and the status is not ! or ?. If it is abnormal, please reinstall the serial port driver or replace
the serial port card.
4) Check the configuration file information and view if the COM port is set the same as the actual COM
port. If it is different, please change it. Please refer to 6.3.3 Confirm the configuration file.
5) Download the latest version of the software and upgrade it. Refer to 6.4 Software Upgrading.
6) If the above cannot be solved, please check the wiring of the serial port of the analyzer to the main
control board, and apply for the main control board to replace it. Refer to Installation Methods and
Precautions of the main control board.
10.2.13 Sample Probe Fails to Detect the Level of the Wash Well A01028
Error Code: A01028
Fault Information: the sample probe fails to detect the level of the wash well
Fault Mechanism: When the sample probe is cleaned in the wash well, the upper computer does not
receive the level detection signal or the level detection signal does not last for a long time.
Procedure:
1)First confirm that whether there is water in the wash well, and whether the water quantity meets
the requirements when the analyzer generates the alarm. Please follow the Check Probes/Mixers
Exterior Wash Flow method for inspection.
2)If the wash well is dry, check the hydropneumatic system. First check the water in the mixer
wash well.
a)If there is water in the mixer, focus on the exterior cleaning tube of the sample probe,
especially whether the restricting capillary of the SV03 valve outlet is blocked, and then check
whether the tubes and connectors of D21, D19, D18 and D16 are blocked.
Check if SY01 is blocked or loosened. If you need to replace it, please refer to
Removing/Reinstalling Cuvette Wash Preheating Assembly.
Check the SV01 valve. Refer to 7.9.2 Checking Floater and Valve Status.
Then consider the SV03 valve failure, you can open the valve with reference to the
content of 7.9.2 Checking Floater and Valve Status and check if there is a crisp sound.
If not, the valve may be damaged, replace it. In addition, the replacement method can
be used, SV03 and SV04 are interchanged. If the fault is transferred, the SV03 is faulty.
If it is not transferred, check whether the J6 connector from SV03 to the power drive
board is damaged or loosened. If not, apply for the power drive board for replacement.
Refer to Installation Methods and Precautions of the power drive board.
b)If the mixer is also waterless, check the shared tube and check it in the following order:
If the tank does not have enough deionized water, add the deionized water.
If the inlet filter is blocked or damaged, try to replace it. Please refer to
Removing/Reinstalling Filter
If the tubes (D01 - D04) related to the exterior cleaning are blocked, try to dredge the
tubes.
3)If there is little water in the wash well, focus on the exterior cleaning tube of the sample probe,
especially whether the restricting capillary of the SV03 valve outlet is blocked.
4)If there is water in the wash well and the flow rate is normal, check if the sample probe is bent.
Check the verticality of the sample probe, visually measure it, or use a vertical reference for comparison.
If the sample probe is not straight, please replace the sample probe. Refer to A.2BS-360E Moving Parts
Position Confirmation Guide (No Alignment Tooling)Judging and Calibrating Verticality of Probe.
5)If the wash well is not well drained, causing the DI water retained, check whether the waste
tube of the wash well is blocked. If so, dredge the tube and check it according to the fluidics diagram
of the whole machine.
6)Check if the horizontal and vertical positions of the sample probe to the wash well are correct.
Refer to 7.5.2 Probe to Horizontal and Vertical Positions on Wash Well for alignment.
7)Check if the analyzer is falsely alarmed due to interference:
a) Confirm that the analyzer has a grounding wire and the zero ground voltage is <5V. The
measuring method is: set the multimeter to AC 250V voltage, and connect the black test lead to
the ground wire jack, and the red test lead to the live wire/zero wire jack, respectively. When the
grounding is good, the voltage between the ground wire and the zero wire is less than 5V, and
the voltage between the ground wire and the live wire is similar to the voltage between the live
wire and the zero wire.
b)Check whether the connection cable of the level detection board is damaged, or the cable
is bound too tightly by the cable tie, or the board connector is loose. The specific operation method
is: power off the whole analyzer, open the protective cover, remove 2 hardened dusting screws
(M3X6), remove the arm cover of the sample probe (DS193), refer to Figure 3-26 Removing
level sense board, disconnect the J1. J2. and J3 connectors of the level detection board, and re-
connect them. If the connection wire of the J2 connector to the J11 connector of the main control
board is damaged, wind it with an electrical tape. If the cable tie is too tight, loosen the cable tie
and re-plug the J11 connector.
8)If the level detection board 051-002479-00 is faulty replace it by referring to3.5.6 Replacing
level sense board.
9)If the main control board is faulty, refer to Installation Methods and Precautions of the main
control board for replacement.
MF01
T05 T06 D08 T07 T08 D09
T09 S ample
SV02 probe Mi xer
SY03
Wash well
of sa mple Wash well
pr obe of m ixer
T14
D18 D19 T17 D21
D07 SV03
T15 T16
T46 D20 T18 D22
D33
SV04
D17
T04
D06
P01
D16
To Cuvette
Wash
T13
D05
T12 D11
T10
D10
D15 T11 D14 D12 D13
T03 SV01
D04
SY01
D01 FS01
, V01
10.2.14 Sample Probe Fails to Detect the Level on the Reaction Carousel
when Dispensing A01033
Error Code: A01033
Fault Information: the sample probe fails to detect the liquid level on the reaction carousel when
dispensing
Alarm Mechanism: When the sample probe is dispensing in the reaction carousel, firstly, detect the R1
volume added before, and determine the number of steps of the vertical movement of the sample probe adding
the sample according to the R1 volume (volume tracking technology). This alarm is triggered if no level is
detected within the specified number of steps.
Troubleshoot as follows:
When analyzing the problem, follow this sequence: level signal generation (sample probe) → signal analysis
(level detection board) → transmission (connection line) → signal processing (main control board).
Procedure:
1)First check the alarm log of the analyzer and check if the analyer still has an alarm that the
sample probe does not detect the liquid level in the sample carousel or during cleaning. If there is a
similar alarm, judge that the level detection function is faulty. At this time, check if the sample probe,
the level detection board, and the power supply and wire of the level detection board are normal.
2)If there is only this alarm, according to the alarm mechanism, first check whether the reagent
in the reagent pack is sufficient, if not, add the reagent in time.
Note: There should be no air bubbles in the reagent pack; if it is an open reagent, do not pour too
much reagent.
3)Observe the sample probe. If the reagent probe lowers into the reagent pack during the whole
procedure of aspirating the reagent, focus on whether the sample probe is blocked, and whether the
reagent volume added to the cuvette is insufficient due to the leakage of the sample syringe and the
pipeline.
If the sample probe is blocked, please dredge the probe by referring to Cleaning probe interior.
If the syringe leaks, please replace it by referring to Replace Syringe.
Check the Teflon tubing between the sample probe and the syringe. There should be no wear
and tear, if any, please replace it directly.
4)If the sample probe does not lower to aspirate the reagent but return directly during the test,
judge that the function of the level detection board may be damaged or interfered. At this time, check
the power supply and wire of the level detection board. Replace the level detection board.
5)Check the vertical reagent dispense position from the sample probe to the reaction carousel,
and align it according to 7.5.6 Probe to Vertical Limit Position on Reaction Carousel.
10.2.15 Reagent Refrigeration Temperature Is Out of Range
Fault Mechanism: the temperature of reagent carousel is not within 2-12°.
Key note: The refrigeration module is cooled by three Peltiers.
Error Code:
A14007: reagent refrigeration temperature is out of range.
TDISP temperature:
A14008: reagent refrigeration temperature is out of range.
TDISP: ?
Fault Analysis:
Does the ambient
Reage nt refrigera tion Is ambient temperature of the
temperature temperature Yes department meet the
overran ge abn ormal standard? Is air
conditioning required?
No
No
No
No
Replace th e p ower
drive boar d
Figure 10-7 Troubleshooting flowchart for reagent refrigeration temperature out of range
Procedure:
1) View alarm information and temperature display status.
2) Confirm that the ambient temperature of the department is within 15 to 30 degrees. If it is out of range,
please install an air conditioner.
3) Enter the software Utility - Status - Power, and view the Peltier status. if it is out of range, please
check the Peltier and wiring. To replace the Peltier, please refer to 3.4.5 Replacing cooler and sealing
ring.
4) Check the air filter and the cooling fan for serious dust accumulation. If so, clean it. If the fan is
abnormal, replace it. Refer to 3.4.9 Replacing cooling fan of reagent refrigeration.
5) Check the wiring from the temperature sensor to the power drive board, measure the resistance value
of the temperature sensor for reagent refrigeration. The sensor has a negative temperature coefficient.
Under normal conditions, the resistance value is 5 KΩ at 25 Celsius degrees, and the resistance value
decreases with increasing temperature. If the sensor is abnormal, replace it. Refer to3.4.4 Replacing
Reagent Temperature Sensor.
6) If the above steps cannot solve the problem, please replace the power drive board by referring to
Installation Methods and Precautions of the power drive board.
10.2.16 Clinical Result Problems
Troubleshoot as follows:
Result inaccurate
Confirm Fluctuated
Reage Check the start Check the Normal
that the reaction curve
nt position of the use time of reaction
parameters
photoelectric the curve
invalid are set Check the start
acquisition and reagent, correctly, Check the position of
replace the light and and probe, photoelectric
splitter lens replace the perform syringe, acquisition,
assembly if reagent for cross- cuvette, and cuvette, lamp,
necessary recalibratio contaminat optics and mechanical
n. ion check. position
In the calibration rule, the calibrator is incorrectly selected, and there may be multiple calibrators
for one chemistry.
3) QC setup error
The control target value and SD value are set incorrectly, or the wrong control is taken.
The QC rule is set incorrectly, and the correct control is not selected.
The control displayed by the software expired and was not modified in time, and the QC could
not be applied.
4) Redissolution error
Dry powder calibrator and QC solution should be redissolved before use. An error occurs when
redissolution with water, for example, only 3 ml of water is required, but 5 ml is added.
The redissolved water is used incorrectly, for example, use the distilled water placed for a long
time, or directly use the mineral water for redissolution. For redissolution, the distilled water for
injection is recommended.
Pipettes are recommended for redissolution when syringe or an uncalibrated sample gun is used.
After redissolution, the storage is incorrect. After normal redissolution, the solution should be split
charged and stored frozen, and the freezing and thawing cannot be repeated. The storage is
recommended not to exceed 1 month.
5) Reagent problem
Reagent is expired
Reagent is not placed in the refrigerator
Air bubbles in the reagent pack.
Rgt load error
Reagent tilt
6) Calibrator and control problem
Matching calibrator and control are not used
Calibrator and control used is not assigned
7) Analyzer problem
Micro-blocking of the sample probe may result in an assignment or a low result
The sample syringe leaks, which may result in poor repeatability
The Teflon tubing from the sample probe to the sample syringe is folded or damaged, which may
lead to poor repeatability of the results.
If the lamp life exceeds 2000 hours or 6 months, the stability of the results may be poor, and the
reaction curve may fluctuate. In particular, the enzymatic chemistry will frequently send an alarm
of linearity limit out of range.
If the photoelectric acquisition position is wrong, all the results will be unstable.
If the cuvette is contaminated, the result will be unstable.
If the mixer is dirty, the result will be unstable.
Position shift will affect loading and mixing, resulting in unstable results.
8) Water quality problem
Poor water quality will affect the results of chemistries such as ions.
The TBA chemistry is affected by the pH value of water
The water quality will affect the cleaning effect. There will be cleaning residue, affecting the results
9) Sample problem
Sample centrifugation problem. The sample containing fibrinogen may block the probe.
Sample trait problems, hemolysis, jaundice, chyle blood, etc.
Blood collection tube problems. Separation of glue, etc. may be easy to result in probe blockage.
Some patients have taken certain drugs and caused abnormal detection in some chemistries.
A01028 Sample Error Sample probe fails to / 1. No DI water or hydro error; Refer to 10.2.13 .
probe unit detect liquid level 2. Probe level detection error.
during cleaning
A01029 Sample Error Sample is insufficient / 1. Sample contains clots or insufficient 1. Check if sample has been pre-processed
probe unit or contains fibrins sample; 2. Probe hydro error; 3. Clog properly; check if sample contains foreign matters
and clots or Probe detection system error. such as clot; check if there is sufficient sample.
clog detection failed.
2. Clean the probe with diluted wash solution. If the
Position:
Sample ID/bar code: problem remains, remove the probe and unclog it.
3. Check if hydro unit is normal, especially if there
is any air bubbles in clog detection pressure
sensor.
4. Check clog detection board and clog detection
pressure sensor, if previous two possible causes
have been eliminated.
A02023 Probe unit Warning Insufficient reagent / 1.There is no reagent or insufficient 1. Check if there is sufficient sample, and run the
Or reagent on the designated position; test again.
Probe R2 level
2. Probe level detection error. 2. If there are level detection error, check the
detection failed.
probe, level detection board and connecting
cable; replace if any abnormality is identified.
A02025 Probe unit Warning Probe dispenses / 1. Sample probe aspirates nothing; 1. Check if the reagent meets the requirement and
insufficient reagent Syringe assembly error. is sufficient in volume, and then try again.
2. Check if there is any abnormity of syringe
assembly.
A02027 Probe unit Warning Water residues exist / Refer to 10.2.5 . Refer to 10.2.5 .
in the cuvette or
Probe level detection
failed
A02032 Probe unit Warning Reagent is / 1. The reagent probe aspirates nothing. 1. Check if the reagent meets the requirements,
insufficient or 2. The probe syringe assembly goes and is sufficient in volume and free of air bubbles.
contains air bubbles Then try again.
wrong.
Chemistry:
3. There is an error with the vertical 2. Check if the probe syringe assembly works
Position:
extreme position of the sample probe properly.
above the reagent carousel. 3. Check if the vertical extreme position of the
sample probe above the reagent carousel is
correct.
A05006 Mixer unit Error Mixer vertical / Mixer vertical movement error Mixer vertical movement error
movement error
1. Sensor status error 1. Sensor status error: Check if belt or belt wheel is
Mixer horizontal
Belt or belt pulley is loose. loose, tighten them if so.
movement error
2. Failed to find the zero position. 2. Failed to find the zero position: 1) Remove
1) Assembly is jammed; obstacle or replace with a new assembly. 2)
Replace PCB; 3) Replace motor. 4) Replace
2) Driving circuit error;
sensor. 5) Tighten properly.
3) Vertical motor error;
3. Vertical movement is not allowed in current
4) Vertical sensor error; horizontal position: Home instrument or unit.
A22001 ISE unit Error Slope out of range, SLEX 1. Electrode is not installed correctly. 1. Remove and reinstall the electrode.
electrode:
2. Calibrator expired. 2. Replace the calibrator.
3. Electrode degenerated 3. Replace the failed electrode and run tests again.
4. Bubbles in reference electrode 4. Remove the electrode and clap on it to eliminate
5. Reference electrode has been used for bubbles. Reinstall the electrode and run
a long time. calibration.
6. Electrodes interfere with each other. 5. Replace the reference electrode and run tests
7. Module or fluidic temperature exceeds again.
32℃. 6. Replace electrode by different combinations to
troubleshoot the failed one.
7. Monitor temperature, if too high, relocate
equipment.
8. Contact our customer service department or
your local distributor.
A22002 ISE unit Error Air in sample. SMPA 1. Sample is insufficient, or contains 1. Increase the sample volume. At least 90μl
Position: bubbles after being dispensed. sample should be prepared.
2. Electrode is not installed correctly. Reinstall it.
2. No or insufficient sample has been
3. Check the waste tube, and if necessary,
dispensed into the sample injection port.
replace it.
3. The electrodes are not properly
installed, causing leakage.
4. The waste pump tube is aging or
broken.
A22004 ISE unit Error ISE unit cannot be / 1. Power supply of ISE module goes 1. Replace the 24V power supply board.
connected. wrong. 2. Replace the ISE communication cable.
3. Replace the interface or pins.
2. Communication cable between ISE
4. Replace the ISE main control board.
module and middle-layer unit goes wrong.
3. Communication interface or pins failure
4. Main control board of the ISE module
goes wrong.
A22008 ISE unit Error Voltage overflow, VOUT 1. Electrode or reagent pack is invalid. 1. Replace the problematic electrode and reagent
electrode: pack.
2. Electrode is unsteady.
2. New electrode will become steady after 15
3. New reagent pack is unstable. minutes since installed.
4. Reference electrode has been used for 3. Run a couple of calibrations after installing new
over 6 months. reagent pack.
5. ISE main control board failure. 4. Replace the reference electrode.
5. Replace the ISE main control board.
A22009 ISE unit Error Electrode slope drift. VDRF/ 1. Electrode or reagent pack is invalid. 1. Replace the failed electrode or reagent pack.
(during calibration) SLDR
2. Electrode is unsteady. 2. New electrode will become steady after 15
Or
3. New reagent pack is unstable. minutes since installed.
Electrode voltage
drift. (during sample 4. Reference electrode has been used for 3. Run a couple of calibrations after installing new
reagent pack.
A22016 ISE unit Error Reagent pack chip / 1. Reagent pack is not installed. 1. Install reagent pack.
writing error. Unload 2. Reagent pack wand is failed. 2. Replace the wand.
the reagent pack and
load it again.
A22017 ISE unit Error Air in ISE wash / 1. ISE wash solution is insufficient. 1, Place sufficient ISE wash solution.
solution
2. Waste pump tube is aging, blocked, or 2. Replace the pump tube
broken: 3. Clean the sample injection port and reinstall
3. Sample injection port and fluidic path are electrodes.
blocked or leaking. 4. Replace the bubble detector.
4. Air bubble detector failed
A22018 ISE unit Error No fluid in tubing CALF 1. ISE wash solution is insufficient. 1, Place sufficient ISE wash solution.
2. Waste pump tube is aging, blocked, or 2. Replace the pump tube.
3. Clean the sample injection port and reinstall
broken:
electrodes.
3. Sample injection port and fluidic path are 4. Replace the air bubble detector.
blocked or leaking.
4. Air bubble detector failed
A22019 ISE unit Error Saving calibration DEP 1. Communication cable is disconnected. 1. Replace the ISE communication cable.
result error 2. Replace the interface or pins.
2. The interface or pin goes wrong.
3. Replace the ISE main control board.
3. ISE main control board failure. 4. Upgrade the operating software or reinstall it.
4. Wrong Software Version.
A22021 ISE unit Error Command format or RESP 1. Communication cable is disconnected. 1. Replace the ISE communication cable.
execution error 2. Replace the interface or pins.
2. The interface or pin goes wrong.
3. Replace the ISE main control board.
3. ISE main control board failure. 4. Upgrade the operating software or reinstall it.
4. Wrong Software Version.
A22024 ISE unit Error ISE response check RESP 1. Communication cable is disconnected. 1. Replace the communication cable.
code error
2. The interface or pin goes wrong. 2. Replace the interface or pins.
3. ISE main control board failure. 3. Replace the main control board of the ISE
4. Wrong Software Version. module.
4. Upgrade the operating software or reinstall it.
A22027 / Error Fluidic prime failed. / Fluidic prime is not performed. 1.If fluidic prime is not performed,refer to 7.9.4
for fluidic prime.
2. Enter the Parameter Configuration screen,
select Wash Unit, find the Is Fluidic Prime
Performed parameter, check whether the value of
this parameter is 1.Refer to Figure 13-9
Parameter Configuration -Change the Is Fluidic
Prime Performed parameter
A22036 / Error Initializing sample / Due to the system error, the sample bar Refer to 10.2.6 .
bar code reader code reader cannot work normally. If the
failed. error repeats and cannot be removed,
then the bar code reader has not
performed the initialization normally or the
communication cables are disconnected.
A22038 / Error Scanning reagent / Scanning reagent bar code failed. Refer to 10.2.6 .
bar code failed.
A22039 / Error Unmatched software / 1. Executing version query instruction Refer to 10.2.4 .
version. failed.
2. Returned control software version and
A22042 Temperature Warning Temperature of / Warning No. 1:Sensor shortcut or 1. Reproduce failure.
control unit cuvette cleaning fluid temperature above 75℃. 2. Measure ambient temperature to check if
is out of range: %s\r\n Warning No. 2:Sensor disconnected or temperature is out of range.
Temperature: %s temperature below 0℃.
3. Temperature sensor error: check for any signs of
Warning No.3:Heater is heating for a long
short circuit or broken circuit; check if connections
time
are secure.
Warning No.4: Heater control is locked.
4. Temperature protection switch error: check if
protection switch is shut and connections are
secure.
5. Heater error: check if resistance under normal
temperature is correct, and connections are
secure.
6. PCB failure.
7. Parameters are lost. Check if the temperature
control parameters are correct.
8. Electromagnetic interference exists.
A22043 Temperature Warning Temperature of / Warning No. 1:Sensor shortcut or 1. Reproduce failure.
control unit cuvette wash solution temperature above 75℃. 2. Measure ambient temperature to check if
is out of range: %s\r\n Warning No. 2:Sensor disconnected or temperature is out of range.
Temperature: %s temperature below 0℃.
3. Temperature sensor error: check for any signs of
Warning No.3:Heater is heating for a long
short circuit or broken circuit; check if connections
time
Warning No.4: Heater control is locked. are secure.
4. Temperature protection switch error: check if
protection switch is shut and connections are
secure.
5. Heater error: check if resistance under normal
temperature is correct, and connections are
secure.
6. PCB failure.
7. Parameters are lost. Check if the temperature
C00007 Operating Error CPU performance / The CPU is too busy. Reboot the computer and operating software. If
system low this message appears for 3 times, contact our
R&D department.
C00011 Operating Error The last abnormal / The operating software is abnormal or the After rebooting the software, perform the special
system exit may cause analyzer is powered off accidentally. wash procedure.
carryover not
handled. Execute the
Special Wash
maintenance
command before
starting analysis to
ensure accurate
results.
C00012 Operating Warning Sound card failure / No sound card is installed; Sound card Reinstall sound card or sound card driver
system failure; Incorrect sound card driver
C00013 Operating Error The cuvette status / The operating software is not exited Take out and check the standby cuvettes. If they
system may not be refreshed normally due to some reasons.(BS-230) are used, replace them. When you are uncertain if
due to the last they are used, replace them.
abnormal exit of the
system. To ensure
the correct test
results, please check
the cuvettes and
replace them.
C01001 Instrument Error Equipment cannot be / Database software is not installed; Refer to 10.2.12 for processing.
connection connected database is not established; database
system or database file is damaged or
lost; database is opened exclusively
C02001 Database Error Database initialing / Database upgrade failed Refer to 10.2.1 for processing.
failed
C02004 Database Warning Database backup / Cannot read or write necessary Refer to10.2.2 for processing.
failed information from or into database
C02005 Database Warning Reading/Writing / Database software is not installed; Refer to 10.2.1 for processing.
database failed database is not established; database
system or database file is damaged or
lost; database is opened exclusively
C03001 Result Warning Result cannot be RCE Absorbance data for calculation is Investigate the cause of photoelectric data loss;
calculation calculated incomplete, or the dividend is 0. contact our R&D department
Sample ID/bar code:
Position:
Chemistry:
C03002 Result Warning Absorbance out of ABS The absorbance measured at the primary Check the reaction curve. If the absorbance of R1
calculation range and secondary wavelength is greater than is too high, check if reagent has air bubbles or
Sample ID/bar code: 3.3A. syringe leak.
Position:
2. Check if the reagent is placed in the correct
Chemistry:
position.
3. Rerun the test after dilution.
4. Contact our R&D department.
C03003 Result Warning R1 blank absorbance RBK The reagent goes wrong; the cuvette is not Check if cuvette is clean or overflowed; check if
calculation out of range clear; the reaction cuvette is overflowed; or reagent is sufficient, without air bubbles; check if
insufficient reagent is dispensed. chemistry parameters are reasonable; if previous
problems do not exist, replace reagent and rerun
the test.
C03004 Result Warning Substrate depletion BOE The sample concentration is too high, and Check the reaction curve and the substrate
calculation Sample ID/bar code: substrate depletion occurs during fixed- depletion limit. Rerun the test with diluted sample.
Position: time measurements.
Chemistry:
C03019 Result Warning Chemistry: 1-3s The current QC result is greater than ±3 Check if the reagent is qualified, control sample is
calculation Control: 1-3s out of standard deviations from the assigned normal, and the instrument is working correctly.
control mean concentration.
C03020 Result Warning Chemistry: 2-2s Results of two controls or two results of Check if the reagent is qualified, control sample is
calculation Control: 2-2s out of one control within a run are normal, and the instrument is working correctly.
control simultaneously greater than +2 or -2
standard deviations from the assigned
C03024 Result Error Biochemistry test / Software error Rerun the test, or reboot the operating software
calculation period time out. Operating system error and analyzing unit. If the error remains, contact
Cannot continue our customer service department or your local
distributor.
C03026 Result Warning Photoelectric data is / Communication error. If the error persists, contact our R&D department.
calculation lost
C03027 Result Warning Chemistry: 2.7s Multiple QC data and threshold values or Check if the reagent is qualified and control is
calculation Control: 1.0-2.7 out cumulative sum exceed ±2.7SD. normal.Refer to 10.2.16 for processing.
of control
C03028 Result Warning Chemistry: 3.0s Multiple QC data and threshold values or Check if the reagent is qualified and control is
calculation Control: 1.0-3.0 out cumulative sum exceed ±3.0SD. normal.Refer to 10.2.16 for processing.
of control
C03029 Result Warning Chemistry: 5.1s Multiple QC data and threshold values or Check if the reagent is qualified and control is
calculation Control: 0.5-5.1 out cumulative sum exceed ±5.1SD. normal.Refer to 10.2.16 for processing.
of control
C05010 Reagent bar Error The pretreatment / The pretreatment reagent position is Place the pretreatment reagent in correct position.
code reagent position is occupied by other reagent.
occupied by other
reagent. Position:
C05011 Reagent bar Error The ISE wash / The ISE wash solution position is Place the pretreatment reagent in correct position.
code solution position is occupied by another reagent.
occupied by another
reagent.
C06001 Host Error LIS initialization error / Host file is damaged or does not exist. Reboot operating software; initialize LIS module
communicati
on
C06002 Host Error LIS communication / Host parameters error Check system configuration parameters; re-
communicati parameter error connect LIS
on
C06003 Host Error LIS communication / Communication failure If the error occurs accidentally, send or receive
communicati error the instruction again. If the error occurs for several
on times, contact LIS vendor. If problem remain,
contact our R&D department.
C06004 Host Error LIS host cannot be / Abnormal network connection, or the LIS Check if LIS host and LIS station is started
communicati connected host is not started. normally.
on
C06005 Host Warning Sending sample / Communication failure If the error occurs accidentally, send or receive
communicati results failed. the instruction again. If the error occurs for several
on Sample ID/bar code: times, contact LIS vendor. If problem remain,
Position: contact our R&D department.
C06006 Host Warning Sending sample / Communication failure If the error occurs accidentally, send or receive
communicati information failed. the instruction again. If the error occurs for several
C07005 Light source Error Lamp is not turned / 1. The lamp is damaged. Refer to 10.2.10 for processing.
on 2. The lamp cable is not connected
properly.
3. The power board of the lamp is not
connected properly.
4. The power supply of the analyzing unit
is disconnected.
5. The photoelectric collection board goes
wrong.
C07013 Other error of Warning Storage device error. / No floppy disk or U disk is inserted. Check if U disk or floppy disk is inserted or full;
operation Cannot export data Insufficient disk space. The floppy disk or Check if storage device is damaged
unit U disk is locked or damaged.
C07014 Other error of Warning Reagent exhausted / All reagents in this kind of bottles do not Refill or replace the reagent.
operation Chemistry: reach minimum limit; All reagents of this
unit Position: kind cannot be detected
C07016 Other error of Warning Insufficient wash / Insufficient wash solution on the Refill the wash solution on the sample/reagent
operation solution sample/reagent carousel. carousel.
unit Position:
C07017 Other error of Warning Wash solution is / Wash solution is exhausted on the Refill the wash solution on the sample/reagent
operation exhausted sample/reagent carousel. carousel.
unit Position:
C07040 Other Warning Reagent exhausted / Reagent exhausted Refill or replace the reagent.
Chemistry: Reagent probe does not detect liquid level
C07041 Other Error ISE reagent is less / Inventory of the ISE reagent is lower than Check the inventory, and refill ISE reagent if
than %s the alarm limit. necessary.
C07042 Other Warning %s, lot number: %s, / One or more special reagents have Replace the reagent.
position: %s, has expired.
been expired
C07043 Other Warning Pretreatment reagent / Pretreatment reagent is used up. The test for whole blood needs pretreatment
is exhausted. reagent. Fill the pretreatment reagent.
Position:
C07044 Other Warning Pretreatment reagent / Pretreatment reagent is insufficient. The test for whole blood needs pretreatment
is insufficient. reagent. Fill the pretreatment reagent.
Position:
C07045 Other Warning ISE wash solution is / ISE Wash solution is exhausted. Add the ISE wash solution.
exhausted.
C07046 Other Warning ISE wash solution is / ISE wash solution is insufficient. Add the ISE wash solution.
insufficient.
11 Parts List
Table 11-1 Parts list
Order Number 801 Number Part Number Part Description Part Name in PI&PO Applicable Range Location & Reference
801-BA40-00016-00 801-BA40-00016-00 051-000218-00 Sample Probe Clogging Detection Board Sample Probe Clogging Detection Board / /
051-002479-00 / 051-002479-00 Integration Liquid Detecting Board Integration Liquid Detecting Board PCBA / /
051-002872-00 / 051-002872-00 BA36 AC Driver Board BA36 AC Driver Board PCBA / /
051-002896-00 / 051-002896-00 Power driver board Power driver board PCBA / /
051-002900-00 / 051-002900-00 BA36DC/DC board BA36DC/DC board PCBA / /
115-056343-00 / 115-056343-00 Amplification assembly amplification assembly / /
115-048858-00 / 115-048858-00 Main board(open system) Main board(opend system) BS-360E /
115-048859-00 / 115-048859-00 Main board(Close System 3 User-defined) Main board(Close System 3 User-defined) BS-360E /
115-048860-00 / 115-048860-00 Main board(Close System 5 User-defined) Main board(Close System 5 User-defined) BS-360E /
115-048861-00 / 115-048861-00 Main board(Close System 0 User-defined) Main board(Close System 0 User-defined) BS-360E /
041-001143-00 / 041-001143-00 Lamp base retaining screw Lamp base retaining screw / /
115-038466-00 / 115-038466-00 20W Lamp Module (FRU) 20W Lamp Module (FRU) / /
115-041267-00 / 115-041267-00 Optical measure assembly optical measure assembly / /
115-041273-00 / 115-041273-00 Lamp Len Cone Assembly Len Assembly / /
801-BA40-00167-00 801-BA40-00167-00 BA40-20-61352 Fiber Bunch Fiber Bunch / /
082-002426-00 / 082-002426-00 Diaphragm pump of NMP830 diaphragm pump of NMP830 / Auto washing, step 1-8 waste pumps, P02-P
801-BA10-00025-00 801-BA10-00025-00 BA30-21-15311 Probe Interior-washing Pump Probe Interior-washing Pump / P01
024-000908-00 / 024-000908-00 Motor Step 1.8°1.2A250mNm42mm Motor Step 1.8°1.2A250mNm42mm / Auto washing assembly vertical motion moto
024-000909-00 / 024-000909-00 Motor Step 1.8°2A 0.54Nm 42*50mm Motor Step 1.8°2A500mNm42mm / Washing syringe motor
115-045515-00 / 115-045515-00 Sample reagent Disk Motor Assembly Sample reagent Disk Motor Assembly / /
115-045534-00 / 115-045534-00 reaction disk motor assembly reaction disk motor assembly / /
115-045521-00 / 115-045521-00 vertical motor assembly vertical motor assembly / Probe/mixer vertical motion motor
115-045522-00 / 115-045522-00 horizontal motor assembly horizontal motor assembly / Probe/mixer horizontal motion motor
115-045899-00 / 115-045899-00 motor module of syringe motor module of syringe / /
801-BA10-00023-00 801-BA10-00023-00 BA30-10-15052 DC Motor(6V) DC Motor(6V) / Mixer DC motor
009-000448-00 / 009-000448-00 Inner wash return valve The wire of inner wash return valve / SV02, SV04
082-002720-00 / 082-002720-00 2-way Solenoid Valve WIRE FOR Solenoid Valve / SV03, SV05, SV07
BA38-21-88190 / BA38-21-88190 3-way solenoid valve WTB-3R-N4E 3-way solenoid valve wire / SV01, SV06
801-BA10-00139-00 801-BA10-00139-00 2100-20-08144 Fan D08T-12PH Fan D08T-12PH / Cooler fan for analyzer
801-BA80-00136-00 801-BA80-00136-00 BA10-20-78211 Fan of lamp (FRU) FAN(FRU) / /
801-BA30-00075-00 801-BA30-00075-00 BA30-21-06530 Fan of Reaction Disk Fan 0.5W / Fan for reaction disk
024-000899-00 / 024-000899-00 FAN 12Vdc 92*92*38mm 110CFM 33.6dB 300mm FAN 12Vdc 92*92*38mm 110CFM 33.6dB 300mm / Fan for reagent cooling
009-004605-00 / 009-004605-00 Long Peltier Wire Long Peltier Wire / /
009-006306-00 / 009-006306-00 Temprature Sensor Wire of Reaction Disk Temprature Sensor Wire of Reaction Disk / /
801-BA20-00146-00 801-BA20-00146-00 024-000086-00 AC Heater 110V 125W AC Heater 110V 125W / /
801-BA20-00147-00 801-BA20-00147-00 024-000085-00 AC Heater 220V 125W AC Heater 220V 125W / /
801-BA30-00088-00 801-BA30-00088-00 024-000110-00 Reagent Temperature Sensor Reagent Temp Sensor / /
12.2.1 Querying State of RS232 Serial Port Card and Network Card
Identifying the installation state of RS232 serial port and network card driver is an important part for detecting
the network state. If the driver is not installed, or any exception occurs to the installed driver, even if the
hardware including serial cable or network cable works normally, the network communication may be faulty.
Select [Computer] - [Property] - [Device Manager] - [Port] to check the installation of serial port card driver to
Note: Regardless of bidirectional LIS communication or unidirectional LIS communication, the LIS shall always
make response to the information sent from the machine.
Messages in LIS communication use fixed format. Each sample message contains control characters which
are invisible to the naked eye but they have to exist. The machine will convert the start character to <SB> and
stop character to <EB><CR> in order to completely express received messages.
Regardless of receiving the sample application information or response message from the LIS system, the
machine will always detect both the start and stop characters in the message frame. An LIS engineer shall not
treat the start character and stop character as a character in the LIS interface, nor add or delete control
characters.
12.5 Common Problems and Corrective Measures
12.5.1 LIS cannot be connected
First, check whether the port number set on the workstation computer matches that of the LIS interface. Do
not use port number 80/8080. Next, check whether the IP address of the workstation matches that of the LIS
host and check that the workstation is properly connected to the LIS host.
The machine is used as the client and the LIS is used as the server side by default. The machine can be
automatically connected to the LIS.
11,09:57:56:467,LinkLayer Log:=>
<SB>MSH|^~\&|||||20180811095756||QRY^Q02|1|P|2.3.1||||||ASCII|||<CR>
QRD|20180811095756|R|D|1|||RD|002100418080230050|OTH|||T|<CR>
QRF||||||RCT|COR|ALL||<CR>
<EB><CR>
11,09:57:56:558,LinkLayerLog: <=<SB>MSH|^~\&|||||20180811095756||QCK^Q02|1|P|2.3.1||||||ASCII|||<CR>
MSA|AA|1|Message accepted|||0|<CR>
ERR|0|<CR>
QAK|SR|OK|<CR>
<EB><CR>
MSH|^~\&|||||20180811095756||DSR^Q03|1|P|2.3.1||||||ASCII|||<CR>
MSA|AA|1|Message accepted|||0|<CR>
ERR|0|<CR>
QAK|SR|OK|<CR>
QRD|20180811095756|R|D|2|||RD||OTH|||T|<CR>
QRF||||||RCT|COR|ALL||<CR>
DSP|1|||||<CR>
DSP|2|||||<CR>
DSP|3|||||<CR>
DSP|4|||||<CR>
DSP|5|||||<CR>
DSP|6|||||<CR>
DSP|7|||||<CR>
DSP|8|||||<CR>
DSP|9|||||<CR>
DSP|10|||||<CR>
DSP|11|||||<CR>
DSP|12|||||<CR>
DSP|13|||||<CR>
DSP|14|||||<CR>
DSP|15|||||<CR>
DSP|16|||||<CR>
DSP|17|||||<CR>
DSP|18|||||<CR>
DSP|19|||||<CR>
DSP|20|||||<CR>
DSP|21|002100418080230050||||<CR>
DSP|22|||||<CR>
DSP|23||20180811095756|||<CR>
DSP|24||N|||<CR>
DSP|25|||||<CR>
DSP|26||serum|||<CR>
DSP|27|||||<CR>
DSP|28|||||<CR>
DSP|29||2^^^|||<CR>
DSP|30||13^^^|||<CR>
DSP|31||6^^^|||<CR>
DSP|32||7^^^|||<CR>
DSP|33||8^^^|||<CR>
DSP|34||9^^^|||<CR>
DSP|35||10^^^|||<CR>
DSP|36||11^^^|||<CR>
DSP|37||12^^^|||<CR>
DSP|38||1^^^|||<CR>
DSC||<CR>
<EB><CR>
11,09:57:56:732,LinkLayer
Log: =><SB>MSH|^~\&|||||20180811095756||ACK^Q03|1|P|2.3.1||||||ASCII|||<CR>
MSA|AA|1|Message accepted|||0|<CR>
ERR|0|<CR>
<EB><CR>
Role of Analysis Tool
A data analysis tool allows copy and paste of raw data of the machine for analysis.
08,23:03:18:947,LinkLayer
Log: =><SB>MSH|^~\&|||||20180708230318||ORU^R01|2800|P|2.3.1||||0||ASCII|||<CR>
PID|1472|||||||O|||||||||||||||||||||||<CR>
OBR|1472||9015|^|N|20180708224731|20180708224703|20180708224703||1^1||||20180708224703|serum|||
|||||||5|||||||||||||||||||||||<CR>
OBX|1|NM|4|calcium|2.252133|mmol/L|-|N|||F||2.252133|20180708225609|||0|<CR>
OBX|2|NM|5|magnesium|0.569389|mmol/L|-|N|||F||0.569389|20180708225829|||0|<CR>
OBX|3|NM|6|inorganic phosphorus|1.578690|mmol/L|-|N|||F||1.578690|20180708230304|||0|<CR>
OBX|4|NM|10|total bilirubin (vanadate oxidation method)|8.779144|μmol/L|-
|N|||F||8.779144|20180708230315|||0|<CR>
OBX|5|NM|11|direct bilirubin (vanadate oxidation method)|3.441980|μmol/L|-
|N|||F||3.441980|20180708230318|||0|<CR>
OBX|6|NM|16|adenosine deaminase|7.739112|U/L|-|N|||F||7.739112|20180708230311|||0|<CR>
OBX|7|NM|17|prealbumin|230.296762|mg/L|-|N|||F||230.296762|20180708230300|||0|<CR>
OBX|8|NM|18|total bile acid|1.787443|μmol/L|-|N|||F||1.787443|20180708230130|||0|<CR>
OBX|9|NM|19|alanine aminotransferase|19.820950|U/L|-|N|||F||19.820950|20180708230134|||0|<CR>
OBX|10|NM|20|aspartate amino transferase |39.088345|U/L|-|N|||F||39.088345|20180708230137|||0|<CR>
OBX|11|NM|21|alkaline phosphatase|130.115058|U/L|-|N|||F||130.115058|20180708230130|||0|<CR>
OBX|12|NM|22|γ-glutamyltransferase|67.893341|U/L|-|N|||F||67.893341|20180708230134|||0|<CR>
OBX|13|NM|23|lipoprotein (a)|71.263917|mg/L|-|N|||F||71.263917|20180708230217|||0|<CR>
OBX|14|NM|24|total protein|48.843538|g/L|-|N|||F||48.843538|20180708230224|||0|<CR>
OBX|15|NM|25|cholinesterase|4114.479156|U/L|-|N|||F||4114.479156|20180708225949|||0|<CR>
OBX|16|NM|26|albumin|29.379028|g/L|-|N|||F||29.379028|20180708225721|||0|<CR>
OBX|17|NM|27|lipase|10.279494|U/L|-|N|||F||10.279494|20180708230130|||0|<CR>
OBX|18|NM|28|α-amylase|63.568402|U/L|-|N|||F||63.568402|20180708230039|||0|<CR>
This tool can display how the LIS interface process raw data from the machine and how the raw data from the
machine is analyzed and extracted, and generates a log in a specified format
Sample ID 9015
Channel ID -------------4--------|| item name --------calcium:--------:result 2.252133
Channel ID -------------5--------|| item name --------magnesium:--------:result 0.569389
Channel ID -------------6--------|| item name --------inorganic phosphorus:--------:result 1.578690
Channel ID-------------10--------||item name--------total bilirubin (vanadate oxidation method):--------:result
8.779144
Channel ID-------------11--------||item name--------direct bilirubin (vanadate oxidation method):--------:result
3.441980
Channel ID -------------16--------|| item name --------adenosine deaminase:--------:result 7.739112
Channel ID -------------17--------|| item name --------prealbumin:--------:result 230.296762
Channel ID -------------18--------|| item name --------total bile acid:--------:result 1.787443
Channel ID -------------19--------|| item name --------alanine aminotransferase:--------:result 19.820950
Channel ID -------------20--------|| item name --------aspartate amino transferase:--------:result 39.088345
Channel ID -------------21--------|| item name --------alkaline phosphatase:--------:result 130.115058
Channel ID -------------22--------|| item name --------γ-glutamyltransferase:--------:result 67.893341
Channel ID -------------23--------|| item name --------lipoprotein (a):--------:result 71.263917
Channel ID -------------24--------|| item name --------total protein:--------:result 48.843538
Channel ID -------------25--------|| item name --------cholinesterase:--------:result 4114.479156
Channel ID -------------26--------|| item name --------albumin:--------:result 29.379028
Channel ID -------------27--------|| item name --------lipase:--------:result 10.279494
Channel ID -------------28--------|| item name --------α-amylase:--------:result 63.568402
Channel ID -------------29--------|| item name --------apolipoprotein A1:--------:result 1.740667
Channel ID -------------30--------|| item name --------apolipoprotein B:--------:result 1.241503
Channel ID -------------31--------|| item name --------triglyceride:--------:result 5.302102
Channel ID -------------33--------|| item name --------low density lipoprotein cholesterin:--------:result 2.838998
Channel ID -------------34--------|| item name --------high density lipoprotein cholesterin:--------:result 1.156541
Channel ID -------------35--------|| item name --------total cholesterol:--------:result 5.675223
Channel ID -------------36--------|| item name --------creatinine (sarcosine oxidase method):--------:result
4) Remove all electrodes except the reference electrode. Dispense the aspirated calibrator A into
electrodes K+, Cl- and Na+, and use adhesive tape to seal both ends of the electrodes.
5) Install 4 spacers, use an emptying device to connect the A/B ports on the wand to deionized water,
and then execute Purge A/B periodically for 40 times.
6) Disconnect deionized water from A/B ports and continue with the purging.
Remarks: The number of purging operations can be set according to the emptying condition until the residual
deionized water in the A and B tubes is completely drained, and no liquid flows out from the waste outlet.
7) Remove all electrodes and the emptying device, place the wand in the reagent pack, remove the
peristaltic pump tube, install the shielding box door, and tighten the screws on the shielding box.
Figure 13-4 Remove the pump tube from the peristaltic pump
Manually remove the diluted wash solution tank, empty it, and perform 1) Click Continue. A dialog box will pop up. Confirm that the prompt
Auto Wash and Drain. content is consistent with the real object, and click OK.
2) Set the number of emptying to 50, click Start to perform Auto
Wash and Drain.
3) If there is still liquid out at the inlet of the wash solution tube at the
end of the procedure, click Start, and execute emptying again.
Clean the wash solution tube: 1) Operate according to the prompts on the screen. There is no
1) Remove the diluted wash solution tank and clean it, fill it with no visible wash solution foam in the pack after cleaning. Click
less than 200ml of deionized water, and put it back in place. Continue. A prompt dialog box will pop up. Confirm that the
2 2) Remove the auto wash probe and place it in a vacant container of prompt content is consistent with the real object, and click OK.
not less than 200 mL. 2) Set the number of emptying to 60, click Start to perform Prime
3) Perform Prime Wash Station. Wash Station.
Install the auto wash probe back to the place. Operate according to the prompts on the screen. Select Continue to
3
proceed to the next step.
Wash the cuvette and auto wash tube. 1) Click Continue to set the number of washing to 80, and perform
Cuvette Wash.
4 2) If there is still colored liquid out from the waste tube at the end of
the cleaning procedure, click Start again and execute cleaning
again.
Manually empty the DI water tank and the diluted wash solution tank and Operate according to the prompts on the screen. Click Continue. A dialog
5 put it back to the place. Put the DI water and the wash solution inlet pipes box will pop up. Confirm that the prompt content is consistent with the real
respectively back into the tanks. object, and click OK.
Empty the liquid in the clean tubing until no water flows out from the inlet of 1) Disconnect the water inlet tubing, empty the water in the diluted
the DI water and wash solution tubes, otherwise increase the number of wash solution tank, click Continue, set the number of emptying,
priming. and click Start to perform Empty Cuvette Wash Tube.
6
2) If there is still liquid out at the inlet of the DI water and the wash
solution tube at the end of the procedure, click Start, and execute
emptying again.
1) Empty the DI water tank, wash it and put it back into place. 1) Click Continue, set the number of emptying, and click Start to
2) Empty the water in the exterior and interior wash tubing of the perform Empty Probe Interior and Exterior Wash Tubes.
probe. Observe if no water is sprayed out from the sample probe 2) If there is still water out from the probe tip or from the outlet of the
tip and the wash well. Otherwise, increase the priming times. probe exterior wash well at the end of the procedure, click Start,
The residual water in the syringe can be drained by hand as needed: loosen and execute emptying again.
3) Operate according to the prompts on the screen, empty the water
the syringe, push and pull syringe plunger repeatedly until residual water is
7 in the loading syringe (Note: When loosening the fixing bolt of the
removed completely from syringe and T pieces, and then install the syringe.
syringe, the washer as shown below should be kept properly, and
reinstall the washer when reinstalling the syringe. Otherwise the
reliability of the syringe will be affected) Then install the syringe
so that the upper edge of the V-groove is aligned with the 7.5th
scale line of the sample syringe (the 7.5th scale line from top to
bottom) and reinstall the washer.
Washer
Piston locking
screw
Empty the water in the exterior wash tubing of the mixer. Observe if no 1) Click Continue, set the number of emptying, and click Start to
water is sprayed out from the wash well of the mixer. Otherwise, increase perform Empty Mixer Exterior Wash Tube.
8 the priming times. 2) If there is still water from the outlet of the mixer exterior wash well
at the end of the procedure, click Start, and execute emptying
again.
9 This procedure is complete. Select Continue to save it and exit the screen. Select Continue to exit the alignment.
3) Put the auto wash probe back in place, click Start to perform Empty Waste Pump Tubes, and
observe that no water flows out from the outlet of the waste tube. Click Stop to end the process.
Tap "OK".
2) Enter the Parameter Configuration screen, select Wash Unit, find the Is Fluidic Prime Performed
parameter, check whether the value of this parameter is 0. If yes, click Done to exit the screen;
otherwise confirm whether emptying has been performed as required, and change the parameter to
0, and then click Configure to configure the parameter.
Figure 13-9 Parameter Configuration -Change the Is Fluidic Prime Performed parameter
13.4 Packing
If the analyzer should be packed, please confirm whether the contents in the table below are
completed and whether the relevant materials have been prepared.
No. Check Item
1 The sample/reagent carousel has been emptied and cleaned
2 The reagents, cleansers, etc. in the analyzer have been taken out
3 The fluidic tubes have been emptied. Refer to 13.2 .
4 The materials list is as below:
Tape. TESA4298 tape ivory white, 18 mmX50 (material ID: 095-000039-00)
Stretch film width 450 mmX thickness 0.02 mm (material ID: A90-000095---)
Roll foam. 53 M*1 M*8 mm (Material ID: 0000-10-10765)
Bubble film Width 1 m Double flat Bubble Diameter 10 mm (material ID: A90-000077---)
Heat shrinkable sleeve. Φ2.0 mm 105 Celsius degrees 600 V transparent (material ID: A20-
000018---), 1 for protecting the probe
Cable tie (common tools, self-provided)
Transparent sealed bag (self-provided)
Sealing transparent tape (self-provided)
Computer box (package left when installed)
Accessory box (package left when installed)
Main box and fixing screws (retained items when installed)
After confirming that the above items are correct, follow the installation guide in reverse. The specific steps are
as follows:
1) Separate the main unit from the cabinet and wrap them with stretch film separately, first from the lower
side to the upper side, and then wrap the protective cover diagonally (the whole machine is wrapped
around tightly with 3-4 layers), and expose the carrying handle.
2) Pack the main unit and the cabinet according to the reverse order of section 8.4.1 Unpacking.
1) Attach warning labels to the hand holes on both sides of the box of the computer main unit, and seal the
hand holes.
2) After attaching the serial number label on the box, bind the box with four straps as shown.
14 Optional Modules
14.1 Overview
The optional modules include the ISE module, clog detection module and barcode scanning module.
When the client wants to add the above functions, you should perform the upgrade installation and alignment
work as follows to ensure that the modules function properly.
Shielding box
Four screws for fixing
4)Insert the connector of the waste tube into the ISE module, which should be tightly fit. Pay
attention to the installation direction, and the red seal should not fall off.
Check the
connector of the
ISE waste tube.
5)Install the adapter cable of the ISE module onto the ISE module, and install the ISE module on the
mainframe bottom plate with four M4X8 pan head combination screws. The screws are not tightened
at the moment, and then tighten them after confirming the position of the ISE module. Be careful not
to press on the wire.
6) Remove the six M3X6 screws from the three peristaltic pumps, add one M3 spring washer to each
screen, and then re-tighten the screws. Use twelve M2.5X6 Phillips pan head screws with flat
washers and spring washers to fix the three peristaltic pumps onto the lower bracket hole of the
syringe. Note that the wiring directions are upwards as shown in the figure.
7) Use four M3X6 pan head screws with washers to mount the ISE power board onto the board
bracket:
For the position of the ISE power board, see Figure 4-1 Locations of the PCBA on the instrument
1.
a) Connect the power cord of the ISE module to the corresponding ISE power board and
DC-DC board.
ISE module
ISE module
power cord
cable
From the
elliptical hole
8) Insert the cables from the main control board to the ISE module onto the main control board and
bundle the cables onto the cable seat.
Cables
Bundle from the
the main
cables control
onto the board to
cable the ISE
seat module
Figure 14-7 Cables between the main control board and the ISE module
9) Connect the cables between the ISE module and the three peristaltic pumps.
Note: Bundle the cables at the peristaltic pumps during the assembly procedure, and prevent the cables from
coming into contact with the shafts of the three peristaltic pumps.
Bundle the power cord of the ISE module and the cables between the main control board and the clog detection
board to prevent the cables from coming into contact with the fan of the main control board.
10) Connect the ISE tubes as shown in the following figure. Use the small cable ties on the ISE module
interface to attach the A and B tubes to the shielding box. Install a retaining ring SB-1316 in the
outlet hole of the shielding box of the ISE waste tube.
Wire harness
retainer (small)
11) After the installation is complete, follow the above steps to check again.
12) Open the operating software with the engineer user name and password and click UtilitySystem
SetupFactory Setup.
3) Connect the reagent pack to the reagent pack wand: as shown in the figure below, install the reagent
pack wand in place until you hear a "click" sound. Therefore, the "ISE communication error" appears.
4) Align the probe postion. (Refer to7.5.8 Probe to Horizontal Postion on ISE Part (Optional) and 7.5.9
Probe to Vertical Postion on ISE Part (Optional))
5) Module Intialization,refer to 7.11.1 Module Intialization.
6) Put away the panels including the ISE shielding shell.
Note: After the ISE module is installed, do not cut off the mains power (the power switch for
the analyzing unit can be cut off). If the mains power must be cut off and the downtime is more than 2
hours, empty the ISE and remove the electrodes. For details, refer to rob in 13.2.1 Empty ISE Tube
(optional). To remove electrodes, the ISE module shall be cleaned and emptied as well.
Pressure change
during aspiration
Aspiration Aspiration
Pressure
starts ends
Normal
Aspiration aspiration
waveform during waveform
probe clog
Time
For the clog detection board, please refer to 4.4.7 Clog Detection Board(Optional).
The upgrade code of the clog detection module is 115-046239-00.
Procedure
1) Power off the whole machine, and remove the back plate of the analyzer.
2) Use four M3×6 pan head screws with washers to mount the clog detection board onto the board
bracket.
1 - Installation position of clog detection tee; 2 - Mounting bracket of clog detection board
3) Then connect the cables of the main control board and the clog detection board to the corresponding
clog detection board (051-000218-00) (J2 and J3 interfaces) and the main control board (J3 interface).
4) Use the M3×8 Phillips pan head screws with washers to secure the clog detection tee to the frame.
The clog detection tee assembly with the mark should be faced up.
5) Cut the original D08 tube by 50 mm, and screw the D08 connector into the upper threaded hole of the
clog detection tee (the threaded hole at the end closer to the sensor wire). Then assemble the
connector of the D34 tube, screw one end of the D34 into the lower threaded hole of the clog detection
tee, and screw one end into the SV02 threaded hole.
Note: Be careful of the Teflon tubing, do not squeeze or bend it.
3) Install the reagent carousel anti-fog heating unit onto the reagent chamber with four M3X8 Phillips pan
head screws.
4) Attach the dust cover with the sponge adjustment pad to the fixing plate with four M3X6 Phillips pan
head screws.
5) Use two M3X8 hex socket head caps with spring washers and large flat pads to secure the scanner
bracket to the barcode mounting plate. Use two M3X10 hex socket screws with spring washers and
large flat pads to mount the barcode scanner onto the scanner bracket. In this step, do not tighten the
two sets of screws. Align the position of the scanner first and then tighten the screws. Use two M3X8
hexagon socket screws with spring washers and large flat pads to secure the barcode mounting plate
to the underside of the frame beam and tighten the screws.
Scanner bracket
Barcode scanner
6) Connect the cables between the anti-fog heater and the protection switch and between the main
control board and the barcode module to the corresponding connectors as shown in the figure.
The cable between the anti-fog heater and protection switch passes through the
elliptical hole in the bracket and is pulled to the back of the bracket
The barcode cable is routed along the path shown in the figure and is bundled on the
bracket hole and then pulled to the back of the bracket
7) For barcode alignment, please refer to 7.8 Bar Code Unit (Optional).
8) After alignment, open the operating software with the engineer user name and password and click
UtilitySystem SetupFactory Setup. See Figure 14-11 Utility-System Setup-Factory Setup
9) Click Optional Modules. See Figure 14-12 Factory Setup-Optional Modules.
10) Tick the Enable/Disable Sample Carousel Bar Code and Enable/Disable Reagent Bar Code
checkboxes, and click Save.
Appendices
A.1 BS-360E Installation Acceptance Report
BS-360E
Installation Acceptance Report _V1.0_EH.doc
BS-360E Moving
Parts Position Confirmation Guide (No Alignment Tooling)_V1.0_EH.doc
Tool list.xlsx
BS-360E_Error
Information Feedback Form
BS-360E Series
Recovery Checklist_V1.0_EH.docx
Maintenance log
sheet.xlsx
MF01
T05 T06 D08 T07 T08 D09
Reagent Reaction
T09 Carousel Carousel ZQ01
SV02 NS MIX W09
SY03 W10
PWCS PWCM W11
T36
T14 T17 D21
D18 D19
W12
SI1 SI2 SI3 SI4
D07 SV03 T39 T38 T37
T46 D20 T18 D22
T15 T16
D33 D29 T33 D31 W13 W14 ZQ02
D27 D28
T04 T34 T35
W05 W06
D06 T28 T27 T26
T31
1 2 3
P01 D25 D26 ZZ06 ZZ09
T40 T41
D16 SV06
T25
ZZ07
T13
D05
T21 T23 SY02
T12 D11
T10
D10
D15 T11 D14 D12 D13 CAN01
T03 SV01
W07
D04
SY01
ZZ04
D03 ZZ03
FL01 FL02 W08 W18 ZQ04
D02 ZZ02
T01 T45
T19
FS03
D01 FS01 High conc.
ZZ01 FS02 Diluted wash Low conc.
V01 DI water tank V02 solution tank Waste tank Waste tank
V04