System Check Report


Checked Date : 2020/10/27 14:02:22
User Name : Admin

Total Judgment: Fail


<Consumables> Judgment Ratio Actual/Standard
LCD Back Light Pass [ ] 1% 446/46380 Hour
Fan Motor Pass [ ] 1% 425/61320 Hour
Injection Unit Septum
SPL1 Fail [**********] 121% 121/100 counts
DINJ1-L. Pass [ ] 0% 0/100 counts
Injection Unit Insert
SPL1 Fail [**********] 121% 121/100 counts
DINJ1-L. Pass [ ] 0% 0/100 counts
CRG coolant Used Time

<GC Check> Judgment Ratio Actual/Standard


Heat Sensor in over 300C
Column Oven 17 Hour
SPL1 0 Hour
DFID1 135 Hour
DINJ1 0 Hour
PECD1 0 Hour
Heat Sensor Check
Column Oven Pass
SPL1 Pass
DFID1 Pass
DC Voltage
DC5V Pass 4.99 V
DC24V Pass 24.25 V
DC-15V Pass -15.07 V
Environment
Room Temperature Pass 23.9 C
Atmosphere Pass 94.6 kPa
Primary Pressure
SPL1 Pass 378/150 kPa
CPU
CPU Register Pass
Real Time Clock Pass
Detector ROM
DFID1 Pass
PECD1 Pass
Detector ADC Register
DFID1 Pass
PECD1 Pass
Detector High Voltage
DFID1 Pass 179 V
Detector Ignite Pulse
DFID1 Pass
Detector Ignite Action
DFID1-R. Pass
ECD Cell Frequency
PECD1 Pass [ ] 0% 1/0 kHz
Flow Unit Board ROM
SPL1(CAR1) Pass
DINJ1(CAR2) Pass
Flow Unit Board ADC
SPL1(CAR1) Pass
DINJ1(CAR2) Pass
Flow Control
SPL1 Carrier Pass
SPL1 Purge Pass

<System Configration>
Instrument Name : Instrument1
1 / 2
Print Out : 10/27/2020 2:02:44 PM
Instrument Type : GC-2014
[UNIT]
GC-2014
Serial #
ROM Version 2.2010
Unit ID
[Analytical Line 1]
Injection Port : SPL1
Column : ZB-5
Detector : DFID1