Professional Documents
Culture Documents
Service Manual
© 2008-2020 Shenzhen Mindray Bio-Medical Electronics Co., Ltd. All rights Reserved.
For this Operator’s Manual, the issue date is 2020-01.
All information contained in this manual is believed to be correct. Mindray shall not be liable
for errors contained herein or for incidental or consequential damages in connection with the
furnishing, performance, or use of this manual.
Mindray is responsible for the effects on safety, reliability and performance of this product,
only if:
the electrical installation of the relevant room complies with the applicable
national and local requirements; and
I
WARNING
It is important for the hospital or organization that employs this equipment to
carry out a reasonable service/maintenance plan. Neglect of this may result in
machine breakdown or personal injury.
Be sure to operate the analyzer under the situation specified in this manual;
otherwise, the analyzer will not work normally and the analysis results will be
unreliable, which would damage the analyzer components and cause personal
injury.
NOTE
This equipment must be operated by skilled/trained clinical professionals.
II
Warranty
Exemptions
Mindray's obligation or liability under this warranty does not include any transportation or
other charges or liability for direct, indirect or consequential damages or delay resulting
from the improper use or application of the product or the use of parts or accessories not
approved by Mindray or repairs by people other than Mindray authorized personnel.
Malfunction of the instrument or part whose serial number is not legible enough.
Service Contact
Company Address: Mindray Building, Keji 12th Road South, High-tech Industrial Park,
Nanshan, Shenzhen, 518057, P. R. China
Website: www.mindray.com
Tel: 0049-40-2513175
Fax: 0049-40-255726
III
Table of Contents
Table of Contents ......................................................................................................................1
1
Table of Contents
2
Table of Contents
3
1 Using This Manual
1.1 Scope
To use this manual effectively, you need the following capabilities:
1.2 Introduction
This manual comprises 9 chapters. Refer to the table below to find the information you need.
To CLICK the desired edit box and use the external keyboard or
to move the cursor to the character or digit that you want to delete
DELETE
by clicking the left button of the mouse or using
1-1
Using This Manual
keyboard).
to CLICK the down arrow button of the desired box to display the
SELECT from ×× pull-down list, (and DRAG SCROLL BAR) to browse and then
(for pull-down list) ([↑][↓][PageUp][PageDown]) to browse the current list and press
1.4 Symbols
personnel injury.
analysis results.
1-2
Using This Manual
You may find the following symbols on the analyzer, reagents, controls or calibrators.
BIOLOGICAL RISK
EARTH (GROUND)
ALTERNATING CURRENT
BATCH CODE
USE BY (YYYY-MM-DD)
SERIAL NUMBER
DATE OF MANUFACTURE
1-3
Using This Manual
TEMPERATURE LIMITATION
Be sure to observe the following precautions for the safety of patients and operators when
you are servicing the analyzer.
When servicing the analyzer, be sure to turn off the power. Servicing the
analyzer when it is on may bring risk of electric shock or damage to electronic
components.
Connect the analyzer to a socket having sole fuse and protective switch. Do
not use the same fuse and protective switch with other equipment (e.g. life
supporting equipment). Otherwise, the equipment failure, over current or
impulse current that occurs at the startup moment may lead to tripping.
To prevent personal injury during maintenance, keep your clothes, hairs and
hands from the moving parts, such as sample probe, clipper and piercer.
The reagents are irritating to eyes, skin and mucosa. Wear proper personal
protective equipment (e.g. gloves, lab coat, etc.) and follow safe laboratory
procedures when handling them in the laboratory.
If the reagents accidentally spill on your skin, wash them off with plenty of
water and if necessary, go see a doctor; if the reagents accidentally spill into
your eyes, wash them off with plenty of water and immediately go see a
1-4
Using This Manual
doctor.
Improper maintenance may damage the analyzer. Maintain the analyzer strictly
as instructed by the service manual and inspect the analyzer carefully after the
maintenance.
For problems not mentioned in the service manual, contact Mindray customer
service department for maintenance advice.
All the analyzer components and surfaces are potentially infectious. Take
proper protective measures for operation or maintenance.
The sample probe tip is sharp and may contain biohazardous materials.
Exercise caution to avoid contact with the probe when working around it.
1-5
2 Product Specification
2-1
Product Specification
Note: the single diluent consumable under the predilute mode includes the diluent volume for
dispensing the diluent.
2.6 Throughput
Throughput under the autoloading mode: no less than 60 samples/hour;
2-2
Product Specification
2.10 Autoloader
The autoloader shall contain no less than 30 samples and offers auto barcode reading
function.
During the counting under the autoloading mode, STAT is allowed to be inserted.
Tube specification under closed-tube whole blood mode is also applicable to the
autoloading mode.
The following two operations can be performed without considering the sequence, the
analyzer and PC are not required to be preconnected, and the analyzer and PC can
perform the following operations independently:
Only in the following condition the analyzer is powered on (Startup refers power on,
2-3
Product Specification
initialization and normal count status): power on the analyzer, startup the software, input
correct user name and password successfully and the software and analyzer can
communicate normally.
2.13 Parameter
1. Parameter classification
Measure the WBC, RBC, PLT and HGB of the blood and output 27 Parameters, 3
histogram and 1scattergram, two measurement types CBC and CBC+DIF are offered
as follows:
CBC +
Clone Name Abbreviation CBC
DIFF
White Blood Cell count WBC * *
Basophils number Bas# / *
Leukon 15 parameters)
2-4
Product Specification
CBC +
Clone Name Abbreviation CBC
DIFF
PLT-Related(4 Platelet count PLT * *
parameters)
“*” means that parameters are tested under the mode; “/” means that parameters are not
offered under the mode.
CBC mode: corresponding histograms include WBC Histogram, RBC Histogram and PLT
Histogram;
CBC+DIFF mode: Corresponding histograms include WBC/BASO Histogram, RBC
Histogram and PLT Histogram.
WBC
Lym%
Lym#
Mon%
Mon# ***.** 109/L
Bas%
Bas# ***.** 103/uL **.* %
Eos%
Eos# ****.* 102/uL .***
Neu%
Neu# ***.** /nL
ALY%
ALY#
LIC%
LIC#
2-5
Product Specification
***.* pg
***.* fL
MCV MCH **.** fmol
***.* um3
**** amol
**.* % ***.* fL
RDW-CV RDW-SD
.*** ***.* um3
**.* %
**.* fL
HCT .*** L/L MPV
**.* um3
.*** None
None
.*** %
PDW **.* ( 10 PCT
*.** mL/L
GSD)
Note: Fast setup function of the units in different country mentioned in table 8 is provided;
users can set them by passwords. Customization functions are also offered. When there are
many units to be selected, user can customize all the units by the password. The setup of
some parameters unit is correlated.
2-6
Product Specification
HGB ≤1g/L
HCT ≤ 0.5 %
PLT ≤ 10 109 / L
The Background Requirements are applicable to whole blood and predilute blood.
Background verification: Run the diluent as samples for 3 times, take the maximum value as
the background result
2-7
Product Specification
Parameter Carryover
WBC ≤0.5%
RBC ≤0.5%
HGB ≤0.6%
HCT ≤0.5%
PLT ≤1.0%
International: Take a sample within the following range, test for 11 times consecutively,
take to results of 2-11 results to calculate the (CV,%) or (d) with the formula.
2-8
Product Specification
in the formula:
s ---- Standard deviation of the sample test;
x ---- mean of the sample test;
xi
---- Test value of the sample;
2-9
Product Specification
Select a high end 5-diff analyzer to perform correlation evaluation. The compare analyzer and
the BC-5380 or BC-5180 shall be calibrated before the correlation evaluation and count the
regression equation. In the statistical analysis, pick out the samples which the accuracy will be
influenced by the principle, such as microcytes, existed nucleated red blood cell and so on. In
the statistical regression equation: Y = aX + b; count the correlation coefficient r.
2-10
Product Specification
Parameter r
WBC ≥ 0.99
RBC ≥ 0.99
HGB ≥ 0.98
MCV ≥ 0.98
PLT ≥ 0.95
Correlation analysis and regression calculation:
n is sample quantity
Regression equation: Y=aX+b,meanwhile:
n 2
( Xij X )(Yij Y )
i 1 j 1
a
n 2 2
( Xij X )
i 1 j 1
b Y a X ; n is sample quantity
Differential Parameters
2.14.7.1Correlation requirements
Select 100 normal samples and 100 abnormal samples to perform correlation test with CLSI
reference method (H20, microscope method), take the mean and analyzer results to perform
correlation analysis.
In the equation of linear regression, pick out the sample whose accuracy is restricted by
the principle. In equation of linear regression: Y = aX + b; Calculate the correlation
coefficient r, which shall meet the following requirements. Calibrate the analyzer before
performing the correlation test.
2-11
Product Specification
2.14.7.2Accuracy requirements
Select 20 samples tested with manual differentiate correlation at random,using the 99%
trustable range calculation method to gain the trastable range of manual differentiate range.
Compare the mean of results tested by the device with the trustable range,within the range of
≥99% of the trustable range lower limit or≤99% of the upeer limint are judged as qualified,
those who are out of the range are judeged as unqualified.
3.Requirement
The Lym%, Neu%, Mon%, Eos% and Bas% results tested by the analyzer must be within the
99% credibility range of the results tested by the reference method.
2-12
Product Specification
Analyzer result
Microscopic exam result positive negative
positive TP (True Positive) FN (false negative)
Negative FP (false positive) TN (True Negative)
false positive flagging rate = FP /(TN + FP) *100%
false negative flagging rate = FN /(FN + TP) *100%
Capability requirements to identify the abnormal samples: analyze the differentiate results of
the analyzer and microscope exam. results,, false negative < 15%; false positive < 20%。
2-13
3 Software System
3.1 Overview
The software system is consisted of the analyzer software and the PC operation software -
IPU. The analyzer software runs in the analyzer CF card, the operation software IPU runs in
the Windows 7 Home Basic*32, Windows 7 Ultimate*32, Windows 7 Ultimate*64, Windows
8 Professional*32, Windows 8 Professional*64, Windows 8 Standard*32 and Windows 8
Standard*64 systems. The analyzer software analyzes sequences, collects and reads data;
while the IPU saves, displays and prints results, displays analysis and QC data, and realizes
data management, parameter setup and communication.
The installation packages of open vial and closed tube models are consolidated into one.
2. Upgrade procedure
Start the upgrade tool:
3-1
Software System
Logon: only user of service access level or above may perform analyzer upgrade.
3-2
Software System
After uploading, the upgrade process starts. If the kernel is upgraded, the above steps shall
be repeated to complete upgrade.
Kernel upgrade done.
Repeat the above steps to upgrade again.
When the process completes, restart the analyzer to complete upgrade.
Upgrade duration: Complete upgrade of the analyzer software requires 5~8 minutes.
3-3
Software System
NOTE
Do not disconnect power of the analyzer during the upgrade process, even if
the process bar stays still for a while, or the upgrade may fail or the analyzer
cannot be started.
Close the IPU before upgrade. The IPU shall be closed during analyzer
upgrade process.
Power off the analyzer when prompt message instructs you so. Start up the
analyzer 2 minutes later after the shutdown to continue with the upgrade.
Ensure the completeness of the upgrade package, do not modify any file in the
package.
Keep anti-virus software and fire wall closed throughout the upgrade process.
3.4.1 PC Configuration
Hardware configuration Software configuration
CPU: Intel® 1.6GHz/AMD® 3.5GHz and Operation system:
above Windows 10 Home 64bit
RAM: ≥ 1 G Database: SQLite
3-4
Software System
2. The "User Account Control" dialog box displays, select "Yes" to install the application.
3. The installation program checks system environment, such as operation system version
and database, and then the following dialog box displays.
3-5
Software System
4. If there are modules missing for installation of the IPU, install the modules, and restart the
PC when the prompt message instructs so.
5. After all necessary modules are installed, the language selection screen displays, select
the desired language of the IPU here.
6. Click OK to enter the model selection dialog box, select the desired product model here.
3-6
Software System
7. Click "Next” to enter the installation path selection dialog box. The installation program
is normally installed in non-system disk. To make sure the "Analyzer IPU Software" can
run properly, the residual space of the installation disk shall be more than 100M.
3-7
Software System
9. Click "OK" to conclude the installation process. An shortcut icon will be created on the
desktop and in the system menu.
3-8
Software System
completed are database structure upgrade, configuration file upgrade, print template
upgrade. During the upgrade process, the print template upgrade dialog box will be
displayed.
3-9
Software System
Click "Apply". After restart the analyzer, the IP address will be refreshed.
Start the IPU and check the connection status icon on the top right; flickering
blue suggests the connection is alright.
3-10
Software System
3-11
Software System
NOTE
1. Only support network port communication.
2. If IPU as server is not selected, then the IPU serves as a client end.
3. IP address: valid in the condition of IPU client end network port communication.
When the IPU communicates as client end, fill in the IP address of LIS server; when
the IPU communicates as server, the address is neglected.
4. Port: valid in the condition of IPU network port communication When the IPU
communicates as client end, fill in the LIS server port; when the IPU communicates
as server, and fill in name of the monitoring port of the local PC.
5. IPU as server: valid in the condition of network port communication; select this item,
the IPU will communicate as TCP server, or it will communicate as TCP client end.
6. Protocol type: select protocol and coding type, "HL7+UTF8" is supported by now.
7. ACK synchronous communication: this option applies to HL7 protocol in the
condition of network port communication, or this option will be neglected.
3-12
Software System
3-13
Software System
When IPU receives analysis results of normal samples, and auto transmission is on, the
results will be auto transmitted to LIS. If transmission succeeded, the samples will be
marked as transmitted samples in the review and report screen; if failed, prompt
message will be given.
When IPU receives analysis results of QC samples, and auto transmission is on, the
results will be auto transmitted to LIS. If transmission failed, prompt message will be
given.
If "L-J QC results transmitted in the format of sample results" is selected, the L-J QC
results will be encoded as normal sample information; if it is not selected, the L-J QC
results will be encoded QC information.
Only L-J QC sample with special sample ID will be auto transmitted.
Batch transmission of normal sample results can be initiated from the review and report
screen. If transmission succeeded, the sample results will be marked as transmitted. If
transmission ACK error occurs, prompt message will be given; if network connection
breaks off, prompt message will be given and the batching transmission will be
terminated.
a. The original value and display value of X-R, X Mean QC samples will be transmitted.
b. For X-B QC, only the display value will be transmitted.
Select samples in the review screen and click the Transmit button to transmit the selected
samples.
3-14
Software System
Click the Transmit button at the QC table screen, a dialog box displays, and then click
"Start" to send data. If no data is selected or the date range is illegal, prompt message
will be given.
3-15
Software System
3.6.2.2Error indication
IPU as server
LIS off indication: no LIS connection, the LIS icon shows connection breaks off.
Error message for connection break-off during transmission: Connection breaks off! The
LIS icon shows connection breaks off.
Connection establishment failed: LIS connection failed. The LIS icon shows connection
breaks off.
Error message for connection break-off during transmission: Connection breaks off! The
LIS icon shows connection breaks off.
ACK response failed: transmission is continued, and the message "ACK response
failed!" is given. The message bubble will keep showing. The LIS icon shows
connection is on.
3-16
Software System
or pass the screen restriction check, the message "Obtained info. invalid" will
be displayed. If loading mode or sample mode is obtained, the information will
be displayed on the screen, analysis mode and valid patient information will be
displayed too.
d. If information obtained is legal, and loading mode or sample mode is obtained,
the information will be displayed on the screen, analysis mode and patient
information will be displayed too.
NOTE
The loading mode or sample mode editing can be done in the process of the
inquiry, they can be modified any time in the saving process.
3-17
Software System
Data sent from LIS will be considered invalid in the following conditions:
1) Character codes cannot be recognized.
2) String length exceeds storage limit.
3) Content is not the conventional type. For example: loading mode is not "OV", "AL" or
"CT".
For patient information, if content of a field is invalid, then the field is invalid, other patient
information fields are still valid.
Abnormal data refers to:
Communication times out;
Invalid data;
Missing field;
Not conforming to current mode.
Missing field Default mode will be used if users do not select mode
Invalid data after starting the analyzer or before turning on
bidirectional LIS. If users do select mode, the
Analysis mode
selected mode will be used. If there are samples
already analyzed, mode of the previous sample will
be used.
3-18
4 Main Parameters Description
number
Eosinophil Eos% Eos% = particle numbers of the Eos area in the
percentage DIFF channel *100%/ all particle number
except the ghost area in the DIFF channel
,
Lymphocyte Lym# Lym# = WBC*Lym%
include 4 RUO parameters
number
Lymphocyte Lym% Lym% = particle numbers of the Lym area in
percentage the DIFF channel *100%/ all particle number
except the ghost area in the DIFF channel
Monocyte Mon# Mon# = WBC*Mon%
number
Monocyte Mon% Mon% = particle numbers of the Mon area in
percentage the DIFF channel *100%/ all particle number
except the ghost area in the DIFF channel
Abnormal ALY# ALY# = WBC*ALY%
Lymphocyte
number
Abnormal ALY% ALY% = particle numbers of the ALY area in
Lymphocyte the DIFF channel *100%/ all particle number
percentage except the ghost area in the DIFF channel
4-1
Main Parameters Description
Volume
Platelet PDW Gain by the PLT histogram, which is 10GSD of
Distribution the PLT distribution
Width
Plateletcrit PCT PCT = PLT*MPV/10000
4-2
Main Parameters Description
After a predetermined volume of blood is aspirated and diluted by a certain amount of reagent,
it is injected into the flow cell. Surrounded with sheath fluid (diluent), the blood cells pass
through the center of the flow cell in a single column at a faster speed. When the blood cells
suspended in the diluent pass through the flow cell, they are exposed to a laser beam. The
intensity of scatter light reflects the blood cell size and intracellular density. The low-angle
scattered light reflects cell size, and the high-angle scattered light reflects intracellular density
(nucleus size and density). The optical detector receives this scatter light and converts it into
electrical pulses. Pulse data collected can be used to draw a 2-dimensional distribution
(scattergram). As shown in the following figure, X-axis represents the intracellular density and
Y-axis the blood cell size. Various types of analysis data can then be obtained from the
scattergrams.
4-3
Main Parameters Description
By analyzing the DIFF channel scattergram, the analyzer presents the Lym%, Mon%, Eos%
and Neu%.
4-4
Main Parameters Description
Each pulse is amplified and compared to the internal reference voltage channel, which only
accepts the pulses of a certain amplitude. If the pulse generated is above the WBC/BAS
lower threshold, it is counted as a WBC/BAS. The analyzer presents the WBC/BAS
histogram; whose x-coordinate represents the cell volume (fL) and y-coordinate represents
the number of the cells.
4.3.2 HGB
The HGB is calculated per the following equation and expressed in g/L.
4-5
Main Parameters Description
of particles that passed through the aperture. The amplitude of each pulse is proportional to
the volume of each particle.
4.5.1 Structure
4.5.2 Principle
Volumetric method of the 53 series: The sample volume of each count (WBC 500uL, RBC
300uL) are measured by tube of fixed volume and initiated by the photocoupler.
Volumetric tube is canceled of the 53 series, the volumes of the impedance channel are
guaranteed by stable count time and stable aperture flow.
Time counting method of the 53 series: The volume of each count is guaranteed by the
following formula:
Measuring volume = Volumetric time x Aperture flow rate
4-6
Main Parameters Description
4-7
Main Parameters Description
Confirmative
Lymphopenia Lym# low
flag Lym# < 0.80×10^9/L
Confirmative
Monocytosis Mon# high
flag Mon# > 1.50×10^9/L
Confirmative
Eosinophilia Eos# high
flag Eos# > 0.70×10^9/L
Confirmative
Basophils high Baso# high
flag Baso# > 0.20×10^9/L
Hemoglobin Calculate and compare
Turbidity/HGB Suspectable abnormal or there special parameters
Interference flag is HGB
interference
RBC results
Suspectable Calculate and compare
RBC possibly
flag special parameters
Agglutination inaccurate
Two or more wave Two or more wave
Dimorphic Suspectable
crests in the RBC crests in the RBC
Population flag
histogram histogram
Abnormal
Suspectable The distribution of RBC
RBC Histogram distribution of RBC
flag histogram is abnormal
Abn. histogram
RBC Suspectable Possibility of iron Calculate and compare
Iron Deficiency
flag deficiency special parameters
Confirmative RDW-CV> 22 or
Anisocytosis Anisocytosis
flag RDW-SD > 64fL
Microcytosis Confirmative MCV low
MCV < 70fL
flag
Macrocytosis Confirmative MCV high
MCV > 113fL
flag
Confirmative
erythrocytosis RBC high RBC > 6.5×10^12/L
flag
Confirmative
Anemia Anemia HGB < 90g/L
flag
Confirmative
Hypochromia Hypochromia MCHC<290
flag
Abnormal Abnormal
Suspectable The distribution of PLT
distribution of distribution of PLT
flag histogram is abnormal
PLT histogram histogram
PLT Suspectable Possibility of PLT Calculate and compare
Platelet clumps
flag clump special parameters
Confirmative PLT high
Thrombocytosis PLT>600×10^9/L
flag
4-8
Main Parameters Description
WBC Confirmative
Leucocytosis /
flag
Confirmative
Leucopenia /
flag
Confirmative
Neutrophilia /
flag
Confirmative
Neutropenia /
flag
Confirmative
Lymphocytosis /
flag
Confirmative
Lymphopenia /
flag
Confirmative
Monocytosis /
flag
Confirmative
Eosinophilia /
flag
Basophilia Confirmative /
4-9
Main Parameters Description
Channel
Flag message Property Shield relation
name
flag
4-10
Main Parameters Description
QFLA Adjustment
mechanism /Sensitivity
Suspectable QFLA Adjustment coefficient adjustment
WBC Abnormal flag mechanism mechanism
QFLA Adjustment
mechanism / Sensitivity
RBC Lyse Suspectable QFLA Adjustment coefficient adjustment
resistance flag mechanism mechanism
QFLA Adjustment
mechanism /
Sensitivity QFLA Adjustment
coefficient mechanism / Sensitivity
WBC Scattergram Suspectable adjustment coefficient adjustment
Abn. flag mechanism mechanism
WBC histogram Suspectable QFLA Adjustment QFLA Adjustment
Abn. flag mechanism mechanism
QFLA Adjustment
mechanism /Sensitivity
Suspectable QFLA Adjustment coefficient adjustment
Left Shift flag mechanism mechanism
QFLA Adjustment
mechanism / Sensitivity
Suspectable QFLA Adjustment coefficient adjustment
Immature Cell flag mechanism mechanism
QFLA Adjustment
mechanism Sensitivity
Suspectable QFLA Adjustment coefficient adjustment
Abn./Atypical Lym flag mechanism mechanism
Turbidity/HGB Suspectable QFLA Adjustment QFLA Adjustment
Interference flag mechanism mechanism
Hemo Suspectable QFLA Adjustment QFLA Adjustment
Agglutination flag mechanism mechanism
Dimorphic Suspectable QFLA Adjustment QFLA Adjustment
Population flag mechanism mechanism
RBC histogram Suspectable QFLA Adjustment QFLA Adjustment
4-11
Main Parameters Description
4-12
5 Error Information of the Analyzer
Troubleshooting and
Error ID Error Name Error Mechanism
Solution
Pressure control 1. Check if the wires
The generated
error of the between the drive
0x01000601 pressure is out of the
pressure board and main board,
set pressure range
chamber wires between the main
Pressure control board and the analog
The generated
error of the signal board are
0x01000602 pressure is out of the
vacuum connected properly;
set pressure range
chamber check if the pipe of the
Pressure value pressure sensor on the
of the pressure The pressure test is analog signal board is
0x01000605
chamber is out out of set range reliable.
of range 2. Check if the pressure
and vacuum chamber
is installed correctly.
3. Click the "Remove
error" button or check if
the status is OK in the
self-test screen, see if
the error is removed;
Pressure value
4. Restart to see if it is
of the vacuum Out of the set range of
0x01000606 normal.
chamber is out pressure test
5. Check if the valve
of range
and pump work
properly.
6. If the problem cannot
be solved, the pressure
sensor may be
ineffective, change the
Analog signal board.
5-1
Error Information of the Analyzer
Troubleshooting and
Error ID Error Name Error Mechanism
Solution
1. Check if the version
information of the drive
MCU and FPGA are
correct;
2. Check if the
temperature of the
reaction bath is within the
range of [34.5,31.5]℃;
The temperature is 0℃,
the input end of the
photocoupler is short
circuit, the sensor maybe
damaged or the wires are
short-circuit
The Reaction bath If the temperature value is
Abnormal
temperature does not 70 ℃ , confirm as the
temperature of
0x01003030 meet the design actual temperature
the Reaction
requirements in the A. If yes, replace the drive
bath
sequence design. board
B. If the input end of the
sensor is open circuit, the
sensor maybe damaged
or the wires are not well
connected
3. If the temperature is
lower than the minimum
range, confirm as the
heating light. If the light is
on constantly, confirm
that the heating wire is
OK, otherwise the heater
is damaged, replace the
components
1. Check if the version
Environment
Environment information of the drive
temperature is
temperature is out of MCU and FPGA;
0x1003031 out of the range
the range of work 2. Check if the actual
of work
environment temperature of the device
environment
is within the range of
5-2
Error Information of the Analyzer
Troubleshooting and
Error ID Error Name Error Mechanism
Solution
[15,30]℃;
3. Check if the wires of
the temperature sensor
are connected properly;
4. Replace the connecting
wires of the temperature
sensor.
1. Check if the version
information of the drive
MCU and FPGA are
correct;
2. Check if the actual
temperature of the device
is within the range of
[10,40]℃;
The temperature is 0℃,
the input end of the
photocoupler is short
circuit, the sensor maybe
damaged or the wires are
short-circuit
If the temperature value is
0x1003032 70 ℃ , confirm as the
actual temperature
A. If yes, replace the drive
board
B. If the input end of the
sensor is open circuit, the
sensor maybe damaged
or the wires are not well
connected
3. Check if the wires of
Environment the temperature sensor is
temperature is OK;
out of the Environment 4. Check if the wires and
running temperature is out of the heater are OK;
environment the running 5. Change the wires of
range environment range the temperature sensor.
Laser diode 1. Check if the version
Laser diode temperature is out of information of the drive
0x01003033
temp. abnormal the run environment MCU and FPGA are
range correct;
5-3
Error Information of the Analyzer
Troubleshooting and
Error ID Error Name Error Mechanism
Solution
2. Check if the optical
temperature of the
reaction bath is within the
range of [30,40]℃;
The temperature is 0℃,
the input end of the
photocoupler is short
circuit, the sensor maybe
damaged or the wires are
short-circuit
If the temperature value is
70 ℃ , confirm as the
actual temperature
A. If yes, replace the drive
board
B. If the input end of the
sensor is open circuit, the
sensor maybe damaged
or the wires are not well
connected
3. If the temperature is
lower than the minimum
range, confirm as the
heating light. If the light is
on constantly, confirm
that the heating wire is
OK, otherwise the heater
is damaged, replace the
components
1. Check if the version
information of the drive
MCU and FPGA are
correct;
2. Check if the
Abnormal temperature is within
Abnormal diluent
0x0100041D diluent the range of [25,36]℃;
temperature
temperature If the temperature is
higher than 36℃, check if
the diluent temperature is
within the range of
[25,33]℃;
The temperature is 0℃,
5-4
Error Information of the Analyzer
Troubleshooting and
Error ID Error Name Error Mechanism
Solution
the input end of the
photocoupler is short
circuit, the sensor maybe
damaged or the wires are
short-circuit
If the temperature value is
70 ℃ , confirm as the
actual temperature
A. If yes, replace the drive
board
B. If the input end of the
sensor is open circuit, the
sensor maybe damaged
or the wires are not well
connected
3. If the temperature is
lower than the minimum
range, confirm as the
heating light. If the light is
on constantly, confirm
that the heating wire is
OK, otherwise the heater
is damaged, replace the
assembly
5-5
Error Information of the Analyzer
5-6
Error Information of the Analyzer
5-7
Error Information of the Analyzer
5-8
Error Information of the Analyzer
5-9
Error Information of the Analyzer
Troubleshooting and
Error ID Error Name Error Mechanism
Solution
1. Action time conflicts in The errors mainly
Sample Syringe the sequence; occur during the R&D
0x01000300
is working 2. Transmitting time error of debugging, when error
the software occurs at the user end,
1. Action time conflict in the the main reason is time
Sampling
sequence; command conflict, click
0x01000310 syringe is
2. Transmitting time error of the "Remove error"
working
the software button, if the error is
1. Action time conflict in the not removed, please
Sheath fluid
sequence; restart the device.
0x01000320 syringe is
2. Transmitting time error of Other special
working
the software troubleshooting is not
1. Action time conflict in the required.
Hemolysin
sequence;
0x01000330 syringe is
2. Transmitting time error of
working
the software
1. Action time conflict in the
Diluent syringe sequence;
0x01000340
is working 2. Transmitting time error of
the software
Aspirate volume set by the
Aspirate volume sequence is out of range
of the sampling Use during the R&D phase
0x01000306
syringe is out of If occurs at the user end, it
range indicates software or
program error
Drain volume set by the
sequence is out of range
Drain volume of
It is use during the R&D
the sampling
0x01000307 phase
syringe is out of
If it occurs at the user end, it
range
indicates software or
program error
Sampling If the error occurs at the
0x01000308 syringe action user end, it is the software
overtime bug
Aspirate volume Aspirate volume set by the
of the Sample sequence is out of range
0x01000316
Injection Use during the R&D phase
Syringe is out of If occurs at the user end, it
5-10
Error Information of the Analyzer
Troubleshooting and
Error ID Error Name Error Mechanism
Solution
range indicates software or
program error
Drain volume set by the
Drain volume of sequence out of range
the Sample Use during the R&D phase
0x01000317
Injection Syringe If occurs at the user end, it
is out of range indicated software or
program error
Sample
If the error occurs at the
Injection
0x01000318 user end, it is the software
Syringe action
bug
overtime
Aspirate volume set by the
Aspirate volume sequence out of range
of the Sheath Use during the R&D phase
0x01000326
fluid syringe out If occurs at the user end, it
of range indicated software or
program error
Drain volume set by the
Drain volume of sequence is out of range
the sheath fluid Use during the R&D phase
0x01000327
syringe is out of If occurs at the user end, it
range indicated software or
program error
Sheath fluid If the error occurs at the
0x01000328 syringe action user end, it is the software
overtime bug
Aspirate volume set by the
sequence is out of range
Aspirate volume
It is use during the R&D
of the hemolysin
0x01000336 phase
syringe is out of
If occurs at the user end, it
range
indicates software or
program error
Drain volume set by the
sequence is out of range
Drain volume of
It is use during the R&D
the hemolysin
0x01000337 phase
syringe is out of
If occurs at the user end, it
range
indicates software or
program error
0x01000338 Hemolysin If the error occurs at the
5-11
Error Information of the Analyzer
Troubleshooting and
Error ID Error Name Error Mechanism
Solution
syringe action user end, it is the software
overtime bug
Aspirate volume set by the
Aspirate volume sequence out of range
of the Diluent Use during the R&D phase
0x01000346
syringe out of If occurs at the user end, it
range indicated software or
program error
Drain volume set by the
sequence is out of range
Drain volume of
It is use during the R&D
the diluent
0x01000347 phase
syringe is out of
If occurs at the user end, it
range
indicates software or
program error
If the error occurs at the
Diluent syringe
0x01000348 user end, it is the software
action overtime
bug
5-12
Error Information of the Analyzer
Troubleshooting and
Error ID Error Name Error Mechanism
Solution
X motor See if the LED
initialization fails (X-M1_LED2;Y-M2_LED2) of
0x01000202
to move to initial X motor failed to move relevant light flashes during
position to target position the motor action. If not,
It shall leave the initial replace the drive board;
position during the 6. Check if the photocoupler
initialization, but the is OK: check if the status of
X motor
photocoupler testing is the two photocouplers at X
initialization fails
0x01000203 still in the initial direction and one
to leave the
position, it indicates photocoupler at Y direction
initial position
the horizontal motor change when blocked or not
action error or blocked;
photocoupler error 7. Remove the bad
X motor The same as 0x201 photocoupler and check if
adjusting does error, it only occurs there is dusts to block the
0x01000204
not move to during the position luminous surface or if there is
target position adjusting process fluids on the luminous surface
If fail to move correct of the photocoupler;
position vertically, it 8. Wipe the photocoupler
indicates the vertical surface and see if the error is
motor action error or removed after the installation,
photocoupler error are if not, replace the
as follows: photocoupler;
1. The wires of 9. If it is not removed, check if
position photocoupler the motor fails to move to
Y motor failed to
or horizontal motor photocoupler position
0x01000211 move to target
are damaged because of step missing or
position
2. Position clogging.(low possibility, it
photocoupler error seldom happens).
3. Horizontal motor
error
4. Board error
5. Larger resistant of
the sampling
assembly vertically
Y motor It shall move to the
initialization fails initial position
to move to up vertically, but the
0x01000212 position photocoupler of the
initial position is not
blocked, it indicates
the vertical motor
5-13
Error Information of the Analyzer
Troubleshooting and
Error ID Error Name Error Mechanism
Solution
action error or
photocoupler error are
as follows:
Y motor It shall move from the
initialization fails initial position
to move to lower vertically, but the
position photocoupler of the
initial position is
0x01000213
blocked, it indicates
the vertical motor
action error or
photocoupler error are
as follows:
Y motor The same as 0x211
adjustment fails which appears during
0x01000214 to move to the the position
target position adjustment process
and can be merged
Y motor The same as 0x212
adjustment does and can be merged
0x01000215
not return to the
up position
If horizontal motor
moves in the following
condition, report the
error:
X motor does
1. The position
0x01000223 not permit the
photocoupler is
action
blocked
2. Vertical initial
position photocoupler
is not blocked
Y motor does If the position is
not permit the blocked, the vertical
0x01000226
action motor is working ,
report the error
Y motor does The motor cannot
not pierce to the drive the sample
lower position probe to pierce the
0x01000228
cap, when the sample
probe returns to the
up position, error is
5-14
Error Information of the Analyzer
Troubleshooting and
Error ID Error Name Error Mechanism
Solution
reported
Horizontal initial During the
position initialization, it shall
photocoupler return to the horizontal
0x01000231 error initial position, but the
photocoupler at the
horizontal initial
position is not blocked
Photocoupler on It shall return to the
the sample vertical initial position
probe error vertically, but the
0x01000235
vertical initial position
photocoupler is not
blocked
Troubleshooting and
Error ID Error Name Error Mechanism
Solution
1. It only appears at the
position adjusting process
2.Check if the process,
adjusting process and if the
When adjusting the
adjustment to the target
Y motor position, if the vertical
position are correct
0x01000218 Adjusting Steps adjusting is out of the
3.If the above items are all
out of limit limit, it shall be
correct, check if the sample
readjust
assembly are correct
4.If the above items are all
correct, replace the
sampling assembly
1. It only appears at the
position adjusting process
2.Check if the adjusting
process and if the
When adjusting the
adjustment to the target
X motor position, if the vertical
position are correct
0x01000208 Adjusting Steps adjusting is out of the
3.If the above items are all
out of limit limit, it shall be
correct, check if the sample
readjusted
assembly are correct
4.If the above items are all
correct, replace the
sampling assembly
0x0100027E RBC position At the corresponding 1. The error only occurs
5-15
Error Information of the Analyzer
Troubleshooting and
Error ID Error Name Error Mechanism
Solution
right edge of the position, the edges of after the position adjustment
gap out of limit relevant position and exit the screen
RBC position photocoupler are too 2. If 272-274, 277-278 or
0x01000272 left edge of the close 27E errors are reported,
gap out of limit check the installation of the
WBC position sampling assembly, if there
0x01000273 right edge of the is no problem, replace the
gap out of limit sampling assembly;
WBC position 3. If other errors are
0x01000274 left edge of the reported, adjust the relevant
gap out of limit position
DIFF position 4. If the error is still
0x01000275 right edge of the reported, check if the
gap out of limit adjusting process and the
DIFF position adjustment to the target
0x01000276 left edge of the position are correct
gap out of limit 5. If the above are all
Right edge of correct, check if the
the open-vial sampling assembly is
0x01000277 installed correctly
position gap out
of limit 6. If the above are all
Left edge of the correct, replace the
open-vial sampling assembly
0x01000278
position gap out
of limit
Right edge of
the autoloading
0x01000279
position gap out
of limit
Left edge of the
autoloading
0x0100027A
position gap out
of limit
Right edge of
the closed-tube
0x0100027B
position gap out
of limit
Left edge of the
closed-tube
0x0100027C
position gap out
of limit
0x01000217 Y motor During the Remove error, restart the
5-16
Error Information of the Analyzer
Troubleshooting and
Error ID Error Name Error Mechanism
Solution
adjusting adjustment, the analyzer failed
position error adjustment
transmission is not
reported at No. 0-6
position, it is
considered not to
occur at the user end
Disorder position
adjustment command
Y motor end order transmission, it Remove error, restart the
0x01000219
position error is software error, the analyzer failed
software shall be
restarted
During the
adjustment, the
adjustment
X motor
transmission is not Remove error, restart the
0x01000207 adjusting
reported at No. 0-9 analyzer failed
position error
position, it is
considered not to
occur at the user end
Disorder position
adjustment command
X motor end order transmission, it Remove error, restart the
0x01000209
position error is software error, the analyzer failed
software shall be
restarted
1. Action time conflict
in the sequence;
The sample 2. Transmitting time It happens accidentally,
0x01000220
probe is working error of the software remove the error directly
3. Drive program
count error
If the initialization is Remove error, restart the
The sample
not started, perform analyzer failed
probe does not
left or right and up and
0x01000221 permit
down adjustment and
adjustment
adjustment of the
software bug
If the error occurs at Remove error, restart the
X motor action
0x01000222 the user end, it is the analyzer failed
overtime
software bug
5-17
Error Information of the Analyzer
Troubleshooting and
Error ID Error Name Error Mechanism
Solution
If the error occurs at
Y motor action Remove error, restart the
0x01000225 the user end, it is the
overtime analyzer failed
software bug
5-18
Error Information of the Analyzer
Troubleshooting and
Error ID Error Name Error Mechanism
Solution
requirements FPGA are correct
2. Check if the wires of
the Analog signal board
J8 are connected
properly;
3. Take out the wire
connectors of the J8
Analog signal board ;
4. Test the TP48 voltage
with the multimeter, if
exceed the range of
1.206V~1.474V, replace
the Analog signal board;
5. If it does not exceed
the range, replace the J8
wires (C-009-002228-00),
connect the J8 and click
the "Remove error"
button
6. If the error cannot be
removed, replace the
main control board
1. First check if the
versions of CPU, versions
of main control board
FPGA are correct
2. Check if the wires are
connected reliably, check
if there are scratching or
damage of the wires, the
During the HGB
wires include:
measurements, the
Laser diode C-009-002226-00 control
0x01003061 current of laser board
current wires of the optical
does not meet the
system
design expectation
C-009-002228-00
low-speed wires of the
main control Analog
signal board
C-009-002229-00 digit
wire of the main control
analog signal board
3. Check if the analog
5-19
Error Information of the Analyzer
Troubleshooting and
Error ID Error Name Error Mechanism
Solution
signal board is normal
a. Check if the analog
signal board TP32 is
smaller than 0.1V, if yes,
it indicates Analog signal
board error
b. Check if the power
supply of the laser control
board is normal:
J1.1:[-12.6,-11.4]V ;
J1.4:[11.4,12.6]V, if not,
replace the wire
C-009-002226-00, if it is
still not within the range, it
indicates analog signal
board error, replace the
analog signal board;
4. Check if the laser
control board is normal
a. Check if the TPILD
voltage of the laser
control board is within the
range of 1.2V~4.5V, if not,
replace the laser control
board, test again, if the
value is still not within the
range, it indicates optical
system error, replace it
b. Check if the switch
control of the laser control
is normal:
J1.6:5V;J1.5<0.8V, if it is
within the range, replace
the laser control board;
Test again, if not, it
indicates the optical
system error, replace it
5. Check if the main
control board is normal
a. Check if the J8.9
voltage of the analog
signal board and the
5-20
Error Information of the Analyzer
Troubleshooting and
Error ID Error Name Error Mechanism
Solution
TPILD voltage of the laser
control board, if yes,
replace the wire
C-009-002228-00, if the
error still exists, it
indicates the main control
board error, replace it;if
not, replace the wire
C-009-002226-00
b. Replace the wire
C-009-002229-00, if the
error still exists, it
indicates the main control
board error, replace it
1. Check if the right-side
door can be opened, if
yes, close the door and
remove the error;
2. If the error is reported
when the door is closed,
check the installation of
the micro-switch, see if
the photocoupler can be
pressed when the door is
Open the side
Open the right side closed;
0x010030D0 door of the
door 3. Check if the wires of
analyzer
the micro switch are
reliable and see if they
are damaged;
4. See if the micro switch
is OK: see if the status
changed when pressed or
ejected the micro switch;
5. If the error is not
removed, replace the
photocoupler.
1. Check if the optical
system is locked tightly, if
The optical
The optical assembly yes, close the door and
0x010030D1 assembly is
is loosen remove the error;
opened
2. If the error is reported
when the door is closed,
5-21
Error Information of the Analyzer
Troubleshooting and
Error ID Error Name Error Mechanism
Solution
check the installation of
the micro-switch, see if
the micro switch can be
pressed when the door is
closed;
3. Check if the wires of
the micro switch are
reliable and see if they
are damaged;
4. See if the micro switch
is OK: see if the status
changed when pressed or
ejected the micro switch;
5. If the error is not
removed, replace the
micro switch.
Troubleshooting and
Error ID Error Name Error Mechanism
Solution
1. Check if the versions of
the software, sequence
Mix Mechanism and autoloader are correct
0x01005201 Software
is working 2. Remove the error
3. Restart the software and
analyzer
The motor shall not be 1. Check if each software
at the initial position, and hardware version is
Mix Mechanism but the photocoupler
correct;
0x01005202 X motor action of the initial position is
error 1 blocked, SMXM initial 2. When checking the error,
position photocoupler ensure that the relevant
or motor error wires of the error are
The motor shall be at connected properly, such
the initial position, but as if the wires of the motor
Mix Mechanism
the photocoupler of and photocoupler are
0x01005203 X motor action
the initial position is reliable, if the label of
error 2
not blocked, SMXM sensor wires match the
initial position sensor positions, if the
5-22
Error Information of the Analyzer
Troubleshooting and
Error ID Error Name Error Mechanism
Solution
photocoupler or motor labels of motor wires match
error the motor position.
Z motor or R Z motor or R motor 3. Click the remove error
0x01005204 motor not at the not at the initial button to see if it is
initial position position removed;
The motor shall not be 4. Click Mix mechanism
at the end position, self-test at the self-test
Mix Mechanism but the photocoupler screen to see if it is
0x01005207 X motor action of the end position is removed;
error 1 blocked, SMXM end 5. Check if the
position photocoupler photocoupler is OK: see if
or motor error the status changes when
blocked or not blocked;
6. Once confirm which
photocoupler is bad,
remove it and check if its
luminous surface is
blocked by the dust or
there is fluids splashed on
the surface;
The motor shall be at
7. Wipe the surface of the
the end position, but
photocoupler and install it
Mix Mechanism the photocoupler of
to see if the error is
0x01005208 X motor action the end position is not
removed, if not, replace the
error 2 blocked, SMXM end
photocoupler;
position photocoupler
8. If the error still is not
or motor error
removed, check if the
motor fail to move to the
photocoupler position
because of missing step or
fail to drive the sample
probe to pierce the cap(low
possibility, generally will not
happen).
1. When checking the error,
ensure that the relevant
X initial photocoupler
wires of the error are
Mix Mechanism and end photocoupler
connected properly, such
0x01005209 X photocoupler of the mix mechanism
as if the wires of the
error are blocked at the
photocoupler are reliable, if
same time
the label of sensor wires
match the sensor positions.
5-23
Error Information of the Analyzer
Troubleshooting and
Error ID Error Name Error Mechanism
Solution
2. Check if there is foreign
material between the two
photocouplers;
3. Click the remove error
button to see if it is
removed;
4. Click Mix mechanism
self-test at the self-test
screen to see if it is
removed;
5. Check if the
photocoupler is OK: see if
the status changes when
blocked or not blocked;
6. Once confirm which
photocoupler is bad,
remove it and check if its
luminous surface is
blocked by the dust or
there is fluids splashed on
the surface;;
5-24
Error Information of the Analyzer
Troubleshooting and
Error ID Error Name Error Mechanism
Solution
1. Check if the versions of the
software, sequence and
Mix Mechanism Software transmit autoloader are correct
0x01005401
is working command delay 2. Remove the error
3. Restart the software and
analyzer
The motor shall not be 1. Check if each software and
at the initial position, hardware version is correct;
Mix Mechanism but the photocoupler 2. When checking the error,
0x01005402 Z motor action of the initial position is ensure that the relevant wires
error 1 blocked, SMZM initial of the error are connected
position photocoupler properly, such as if the wires
or motor error of the motor and photocoupler
The motor shall be at are reliable, if the label of
the initial position, but sensor wires match the
the photocoupler of sensor positions, if the labels
Mix Mechanism
the initial position is of motor wires match the
0x01005403 Z motor action
not blocked, SMzM motor position.
error 2
initial position 3. Click the "Remove error"
photocoupler or motor button to see if it is removed;
error 4. Click "Mix mechanism
X motor is not at the self-test" at the self-test
Mix mechanism end position or the mix screen to see if it is removed;
Z motor current mechanism is not at 5. Check if the photocoupler
0x01005404
action is not the initial position, is OK: see if the status
allowed SMZM action is not changes when blocked or not
allowed blocked;
The motor shall not be 6. Once confirm which
at the end position, photocoupler is bad, remove
Mix Mechanism but the photocoupler it and check if its luminous
0x01005407 Z motor action of the end position is surface is blocked by the dust
error 1 blocked, SMZM end or there is fluids splashed on
position photocoupler the surface;
or motor error 7. Wipe the surface of the
photocoupler and install it to
The motor shall be at
see if the error is removed, if
Mix Mechanism the end position, but
not, replace the photocoupler;
0x01005408 Z motor action the photocoupler of
8. If the error still is not
error 2 the end position is not
removed, check if the motor
blocked, SMZM end
fail to move to the
5-25
Error Information of the Analyzer
Troubleshooting and
Error ID Error Name Error Mechanism
Solution
position photocoupler photocoupler position
or motor error because of missing step or
fail to drive the sample probe
to pierce the cap(low
possibility, generally will not
happen).
1. When checking the error,
ensure that the relevant wires
of the error are connected
properly, such as if the wires
of the photocoupler are
reliable, if the label of sensor
wires match the sensor
positions.
2. Check if there is foreign
material between the two
photocouplers;
3. Click the "Remove error"
button to see if it is removed;
Y initial photocoupler
4. Click Mix mechanism
Mix Mechanism and end photocoupler
self-test at the self-test screen
0x01005409 Z photocoupler of the mix mechanism
to see if it is removed;
error are blocked at the
5. Check if the photocoupler
same time
is OK: see if the status
changes when blocked or not
blocked;
6. Once confirm which
photocoupler is bad, remove
it and check if its luminous
surface is blocked by the dust
or there is fluids splashed on
the surface;;
7.Wipe the surface of the
photocoupler and install it to
see if the error is removed, if
not, replace the photocoupler;
5-26
Error Information of the Analyzer
Troubleshooting and
Error ID Error Name Error Mechanism
Solution
1. Check if the mix
mechanism Z direction is at
the end position, see if the X
Z motor is not at the
direction is at the end
Mix motor end position or X
position;
0x01005500 current action is motor is not at the end
2. Check if the wires and
not allowed position, action is not
photocoupler of the Z or X
allowed
are OK;
5-27
Error Information of the Analyzer
Troubleshooting and
Error ID Error Name Error Mechanism
Solution
of the end position is changes when blocked or
blocked, SMRM end not blocked;
position photocoupler 6. Once confirm which
or motor error photocoupler is bad,
remove it and check if its
luminous surface is blocked
by the dust or there is fluids
splashed on the surface;
7. Wipe the surface of the
photocoupler and install it to
The motor shall be at
see if the error is removed, if
the end position, but
not, replace the
The motor shall the photocoupler of
photocoupler;
0x01005508 be at the end the end position is not
8. If the error still is not
position blocked, SMRM end
removed, check if the motor
position photocoupler
fail to move to the
or motor error
photocoupler position
because of missing step or
fail to drive the sample
probe to pierce the cap(low
possibility, generally will not
happen).
1. Check if each software
and hardware version is
correct;
5-28
Error Information of the Analyzer
Troubleshooting and
Error ID Error Name Error Mechanism
Solution
removed;
5. Check if the photocoupler
is OK: see if the status
changes when blocked or
not blocked;
6. Once confirm which
photocoupler is bad,
remove it and check if its
luminous surface is blocked
by the dust or there is fluids
splashed on the surface;
7. Wipe the surface of the
photocoupler and install it to
see if the error is removed, if
not, replace the
photocoupler;
8. If the error still is not
removed, check if the motor
fail to move to the
photocoupler position
because of missing step or
fail to drive the sample
probe to pierce the cap(low
possibility, generally will not
happen).
5-29
Error Information of the Analyzer
Autoloading feeding
Troubleshooting and
Error ID Error Name Error Mechanism
Solution
1. Check if the versions of
the software, sequence
X loading motor Software transmit and autoloader are correct
0x01005700
is working command delay 2. Remove the error
3. Restart the software and
analyzer
1. Check if the
photocoupler is OK: see if
the status changes when
blocked or not blocked;
2. Once confirm which
photocoupler is bad,
remove it and check if its
luminous surface is
The photocoupler of X blocked by the dust or
loading motor initial there is fluids splashed on
X Loading motor
0x01005701 position, shall be the surface;
action error
blocked but it is not 3.Wipe the surface of the
blocked. photocoupler and install it
to see if the error is
removed, if not, replace the
photocoupler;
4. Remove the error.
5. Check if there are
interferences between the
loading and autoloading
cover
1. Check if there is tube
rack at the tube rack
position;
2. Remove the error;
3. If the error cannot be
X loading motor X Loading Motor
removed, check if the micro
0x01005702 current action is current action is not
switch is OK, see if the
not allowed allowed
status changes when
blocked or not blocked;
4. Check if the wires of the
micro switch are connected
properly;
5-30
Error Information of the Analyzer
Troubleshooting and
Error ID Error Name Error Mechanism
Solution
5. If the error is not
removed, replace the micro
switch
1. Check the current
position of the loading unit;
2. If it is between the initial
position and end position,
a. Check if there are
interferences in the loading
pathway, such as cover
The loading motor is
interference, foreign
at the initial position,
materials on the load
but the end position
platform, if the tube rack is
photocoupler is
X Loading motor placed properly; b. If there
0x01005703 blocked; The loading
action error 2 is no interference, check if
is at the end position,
the motor and the wires are
but the end position
OK;
photocoupler is not
3. If it is at the initial
blocked
position or end position,
check if the photocoupler is
OK: see if the status
changes when blocked or
not blocked; see if the
wires of the photocoupler
are connected properly.
1. When checking the error,
ensure that the relevant
wires of the error are
connected properly, such
as if the wires of the
photocoupler are reliable, if
The initial position
the label of sensor wires
photocoupler and end
X loading motor match the sensor positions.
position photocoupler
0x01005704 photocoupler 2. Check if there is foreign
at the loading platform
error material between the two
are blocked at the
photocouplers;
same time
3. Click the remove error
button to see if it is
removed;
4. Click Mix mechanism
self-test at the self-test
screen to see if it is
5-31
Error Information of the Analyzer
Troubleshooting and
Error ID Error Name Error Mechanism
Solution
removed;
5. Check if the
photocoupler is OK: see if
the status changes when
blocked or not blocked;
6. Once confirm which
photocoupler is bad,
remove it and check if its
luminous surface is
blocked by the dust or
there is fluids splashed on
the surface;;
5-32
Error Information of the Analyzer
Troubleshooting and
Error ID Error Name Error Mechanism
Solution
its wires are OK
1. During the
troubleshooting, please
ensure that the wires
relevant to the error are
connected properly, check
if the wires of the
photocoupler are reliable,
check if the labels of the
sensor match the position
of the sensor;
FX_READY micro 2. Check if there is foreign
switch error, it shall be material in the
Loading micro
0x01005708 pressed, but in the photocoupler;
switch error
real test, it is of 3. Check if the error can be
pressed status removed by clicking the
Remove error button
6. Check if the
photocoupler is OK; Check
if the status changes when
blocked or not blocked;
7. Check if the
photocoupler is bad, the
mechanism is aged, check
if it can be pressed and
ejected;
1. Check if the versions of
the software, sequence
Y loading motor Software transmit and autoloader are correct
0x01005900
is working command delay 2. Remove the error
3. Restart the software and
analyzer
There is tube rack at
Y motor the rack feeding
0x01005901 initialization direction, once initiate
forbidden the count, error is 1.Take out the rack
reported 2. Remove the error
The photocoupler of 1. Confirm that the
the unloading tray is photocoupler of the unload
Unloading
0x01005902 blocked, it indicates tray is blocked, remove the
action error
the unloading tray is tube rack on the unload
full tray;
5-33
Error Information of the Analyzer
Troubleshooting and
Error ID Error Name Error Mechanism
Solution
2. There is no tube rack on
the unload tray, check if the
photocoupler and its wires
are OK
1. Check if each software
and hardware version is
correct;
5-34
Error Information of the Analyzer
Troubleshooting and
Error ID Error Name Error Mechanism
Solution
8. If the error still is not
removed, check if the
motor fail to move to the
photocoupler position
because of missing step or
fail to drive the sample
probe to pierce the cap(low
possibility, generally will not
happen).
1. Check if each software
and hardware version is
correct;
5-35
Error Information of the Analyzer
Troubleshooting and
Error ID Error Name Error Mechanism
Solution
luminous surface is
blocked by the dust or
there is fluids splashed on
the surface;
8.Wipe the surface of the
photocoupler and install it
to see if the error is
removed, if not, replace the
photocoupler;
Y right photocoupler 1. Check if each software
shall be blocked but and hardware version is
Y right
instead not blocked;
0x01005905 photocoupler correct;
or if it shall not be
error
blocked but instead it 2. When checking the error,
is blocked ensure that the relevant
wires of the error are
connected properly, such
as if the wires of the motor
and photocoupler are
reliable, if the label of
sensor wires match the
sensor positions, if the
labels of motor wires match
the motor position.
3. Click the remove error
button to see if it is
Y left photocoupler
removed;
shall be blocked but
Y left 4. Click Feeding Unit
instead not blocked;
0x01005906 photocoupler self-test at the self-test
or if it shall not be
error screen to see if it is
blocked but instead it
removed;
is blocked
5. Check if the
photocoupler is OK: see if
the status changes when
blocked or not blocked;
6. Check which
photocoupler is bad,
remove it and check if its
luminous surface is
blocked by the dust or
there is fluids splashed on
the surface;
5-36
Error Information of the Analyzer
Troubleshooting and
Error ID Error Name Error Mechanism
Solution
7. Wipe the surface of the
photocoupler and install it
to see if the error is
removed, if not, replace the
photocoupler;
8. If the error still is not
removed, check if the
motor fail to move to the
photocoupler position
because of missing step or
fail to drive the sample
probe to pierce the cap(low
possibility, generally will not
happen).
1. Check if each software
and hardware version is
correct;
5-37
Error Information of the Analyzer
Troubleshooting and
Error ID Error Name Error Mechanism
Solution
remove it and check if its
luminous surface is
blocked by the dust or
there is fluids splashed on
the surface;
7. Wipe the surface of the
photocoupler and install it
to see if the error is
removed, if not, replace the
photocoupler;
8. If the error still is not
removed, check if the
motor fail to move to the
photocoupler position
because of missing step or
fail to drive the sample
probe to pierce the cap(low
possibility, generally will not
happen).
If the following error
occurs during the
unload process, report 1. Confirm the Y feed initial
the error photocoupler and unload
1. Y feed initial initial position status, check
Y feed motor position photocoupler if the two photocouplers
0x01005909
unload error is not blocked and their wires are OK;
2. Unload initial 2. Check if there is
position photocoupler inference when the tube
is not blocked rack is feeding, which
3. Left photocoupler cause the rack fails to pass
counter is not 0 the left photocoupler totally
1. Check if the tube rack is
the 3106 specified rack
2. Check if the tube rack
During the rack
pathway is smooth;
Y left and right feeding process, the
3. Check if the left and right
photocoupler left and right
0x0100590D photocoupler and its wires
cooperation photocouplers jump
error abnormally, which is are OK;
illogical.
4. Check if the installation
of the photocoupler barrier
is OK
5-38
Error Information of the Analyzer
Troubleshooting and
Error ID Error Name Error Mechanism
Solution
1. Check if the tube racks
with tube are placed at the
tube testing position
manually;
2. If not, check if the tube
testing photocoupler is OK:
place tube at the tube
testing position, see if the
After the
After the initialization, photocoupler changes
initialization, the
the tube testing normally when blocked or
0x0100590F tube testing
photocoupler is not blocked;
photocoupler is
blocked 3. Check if there is foreign
blocked
materials of the L
photocoupler testing
pathway;
4. Wipe the photocoupler
and see if it is normal;
5. If not, replace the
reagent testing
photocoupler
1. Check if the tube rack
feed channel is smooth, or
if there is foreign material
impede the rack to move;
2. Check if the software
and hardware version is
correct;
The action is
3. Check if the wires of the
performed for 5 times
The right right photocoupler are
but the right
photocoupler is connected reliably and
photocoupler does not
damaged or the there are no damaged;
0x01005910 jump, the right
tube rack 3. Check if the right
photocoupler may be
cannot be photocoupler is OK; press
damaged or the tube
moved the photocoupler blade,
rack cannot be
check if the photocoupler
dragged.
status is normal;
4. Check if the assembly of
the photocoupler barrier is
OK, check if the
photocoupler is block when
pressed;
5. Replace the wires, check
5-39
Error Information of the Analyzer
Troubleshooting and
Error ID Error Name Error Mechanism
Solution
if the photocoupler is
normal;
6. Replace the
photocoupler
Width of the 1. Check whether any
pulse generated foreign matter sticks to the
During tube rack
by the right bottom of the tube rack.
feeding, the width of
photocoupler for 2. Check whether the
0x01005911 the right counter
detecting tube barcode label stuck to the
photocoupler pulse
rack movement tube is easily to stretch, this
exceeds 250 ms.
exceeds the could cause stagnation of
upper threshold the tube rack.
1. Check whether the right
counter is stagnated.
No pulse of the The right counter 2. Press and then loose the
right photocoupler does not right counter, and check
0x01005912 photocoupler for generate any pulse whether the status of the
detecting tube during the tube rack right counter photocoupler
rack movement feeding. changes. If the status does
not change, the counter
photocoupler is faulty.
1. Check whether this
happens because
someone manually
dragged or touched the
tube rack.
Abnormal 2. Check whether any
excessive foreign matter sticks to the
pulses The number of pluses bottom of the tube rack.
generated by generated by the right 3. Check whether the
0x01005913 the right photocoupler is more barcode label stuck to the
photocoupler than one during the tube is easily to stretch,
counter for tube rack feeding. this could cause stagnation
detecting tube of the tube rack.
rack movement 4. If the fault occurs
accidentally and cannot be
cleared, replace the right
counter photocoupler to
check whether this fault is
cleared.
Width of the During tube rack 1. Check whether any
0x01005914
pulse generated feeding, the width of foreign matter sticks to the
5-40
Error Information of the Analyzer
Troubleshooting and
Error ID Error Name Error Mechanism
Solution
by the left the pulse generated bottom of the tube rack.
photocoupler for by the left 2. Check whether the
detecting tube photocoupler counter barcode label stuck to the
rack movement exceeds 250 ms. tube is easily to stretch,
exceeds the this could cause stagnation
upper threshold of the tube rack.
1. Check whether the left
counter is stagnated.
No pulse is
The left photocoupler 2. Press and then loose the
generated by
counter does not left counter, and check
the left
0x01005915 generated any pulse whether the status of the
photocoupler for
during the tube rack left counter photocoupler
detecting tube
feeding. changes. If the status does
rack movement
not change, the counter
photocoupler is faulty.
1. Check whether this
happens because
someone manually
Abnormal dragged or touched the
excessive tube rack.
number of 2. Check whether any
The number of pluses
pulses foreign matter sticks to the
generated by the left
generated by bottom of the tube rack.
photocoupler counter
0x01005916 the left 3. Check whether the
is more than one
photocoupler barcode label stuck to the
during the tube rack
counter for tube is easily to stretch,
feeding.
detecting tube this could cause stagnation
rack movement of the tube rack.
excessive 4. If the fault occurs
accidentally and cannot be
cleared, replace the left
counter photocoupler.
The system detects 1. Check whether the
that the photocoupler barcode label stuck to the
at the initial feeding tube is easily to stretch, this
position has already could cause stagnation of
Feeding motor been triggered when the tube rack.
0x01005917
loses steps the feeding assembly 2. Check whether the
is returning to the height of the tube fixing
initial position after plate over the tube clamp is
feeding the tube rack higher than the vertical
one position ahead. back plate.
5-41
Error Information of the Analyzer
Troubleshooting and
Error ID Error Name Error Mechanism
Solution
This indicates that the
feeding motor loses
steps when feeding
the the tube rack one
position ahead.
1. The user may have
placed two tubes with the
same barcode in two
The system detects adjacent positions in the
the same sample tube rack. If the user
barcode on two needs to place two tubes
Barcodes of two adjacent tubes in the with the same barcode in
0x01005918 adjacent tubes tube rack. This fault one row of tube rack, place
are the same. could be reported only these two tubes in two
by equipment that is positions that are not
installed with a adjacent.
scanner. 2. The tube rack may be
stagnated during tube
feeding. Please
troubleshoot accordingly.
1. Check whether this
The left or right happens because
counter photocoupler someone manually draged
generates a pulse or touched the tube rack.
The tube rack when the feeding 2. Check whether the two
0x01005919 moves assembly is returning pawls on the feeding
unexpectedly. to its initial position assembly rotate smoothly.
after feeding the tube If the pawls get rusty
rack one position because of liquid leakage,
ahead. they fail to rotate smoothly,
the fault occurs.
5-42
Error Information of the Analyzer
Troubleshooting and
Error ID Error Name Error Mechanism
Solution
1. . Check if the software,
autoloader board MCU and
FPGA version are correct,
check if the 4V, 12V and 5V
voltage of the autoloader
power are correct;
2. Check if the compartment
door is opened, if yes,
checks if the compartment
door tests photocoupler and
Open the
Fail to open the its wires are OK;
0x01005a01 compartment
compartment door 3. If the door is not opened,
door failed
check if the wires of the
electromagnet are
connected reliably;
4. Open the compartment
door manually, see if there is
interference or resistance
Note, ensure that the sample
probe is not in the
compartment when open or
close the door
adjustment The error only appears
During the debug, the
parameter during the production or
0x01005a04 adjustment is out of
setting out of position adjustment by the
range, report the error
range customer service, readjust
1. Check if the software,
autoloader board MCU and
FPGA version are correct,
check if the 4V, 12V and 5V
Scanner reading error, voltage of the autoloader
Scanner reading there is no information power are correct;
0x01005a05
error return value within 2. Check if the wires of the
500ms scanner are reliable;
3. Restart the device, see if
the error still exists;
4. If the error is still reported,
replace the scanner
5-43
Error Information of the Analyzer
5-44
Error Information of the Analyzer
5-45
Error Information of the Analyzer
Troubleshooting and
Error ID Error Name Error Mechanism
Solution
amount;2.The reagent replacing
hardware reports
there is no reagent,
the software records
the remain reagent is
less than 3%
0x01003099 DIFF2 lyse 1. The software
insufficient records the remain
reagent is less than
Click the Remove error
-20% of the total
button, load the new valid
amount;2.The
reagent in the prompt
hardware reports
reagent box and perform
there is no reagent,
reagent replacing
the software records
the remain reagent is
less than 3%
1. Check if the waste
container is full;
2. Check if the float
sensor is at reliable
floating status.
3. Check if the wires and
0x01003085 Waste full Waste full connectors of the waste
sensor is OK.
4. Check if the connector
of the waste sensor on
the drive board is OK;
5. Replace the waste
sensor
1. Check if the diluent
container is full;
2. Check if the float
sensor is at reliable
floating status.
3. Check if the wires and
0x01003080 No diluent No diluent connectors of the diluent
sensor is OK.
4. Check if the connector
of the diluent sensor on
the drive board is OK;
5. Replace the diluent
float sensor.
5-46
Error Information of the Analyzer
Troubleshooting and
Error ID Error Name Error Mechanism
Solution
6. Check if the wires
between the reagent
testing board and the
driver board are
connecting reliably and
properly;
7. Check if the reagent
sensor is OK;
8. Replace the reagent
testing board
0x01003081 No HGB lyse No HGB lyse 1. Check if the reagents
0x01003082 No DIFF1 lyse No DIFF1 lyse are used up;
2. Check if the wires
between the reagent
testing board and the
driver board are
connecting reliably and
properly;
3. Check if the reagent
sensor is OK;
4. Replace the reagent
0x01003083 No DIFF2 lyse No DIFF2 lyse testing board
5-47
Error Information of the Analyzer
Troubleshooting and
Error ID Error Name Error Mechanism
Solution
valve is blocked,
1.b.1 If yes, disconnect the
pipe at both ends of the
relieve valve
1.b.1.1 If it is still blocked,
it indicates the relieve
valve error, please replace
it
1.b.1.2 If not, the relieve
valve is normal, there may
be error in other places
1.b.2 If not, it indicates the
circuit error, it maybe
photocoupler or wires
problems
1.c Check if there are
errors at the pinch valve
PV28 and its pinch pipes,
move the pinch valve
manually and see if the
pinch pipe is folded.
1.d Check if the fluids
adding inlet of the DIFF
bath is clogged. If error is
reported after continuous
testing, this step can be
skipped, if error is reported
after startup, or there is
maintenance in the
previous period or there is
error untreated before
shutting down, which need
to be confirmed. Perform
probe cleaner soak for 5
min
1.e Check if the valve V15
can be opened and closed
normally, or the tube TT39
and T40 are folded
2. If the error can be
removed, but "Flow cell
clog" error is reported in
5-48
Error Information of the Analyzer
Troubleshooting and
Error ID Error Name Error Mechanism
Solution
the next count
2.a It is recommended to
perform flow cell
unclogging and DIFF
channel Probe Cleanser
Maintenance, if the error
still exists, it is
recommended to perform
per the order of the
suggestions.
2.b The valve V18 cannot
be opened normally or it is
clogged or the pipes
around the V18 is folded,
pay special attention to the
T24 and T22;
2.b The valve V17 cannot
be opened normally or it is
clogged or the pipes
around the V17 is folded,
pay special attention to the
T26;
2.c The flow cell is
clogged by the foreign
materials, draw out the
flow cell manually with the
syringe, if the device is
installed recently or it is
place still for a long time, it
is suggested to check this
item.
2.c The reset of the relieve
valve is too small, the
cover is damaged and the
spring is rusted, if the
device is installed recently
or it is place still for a long
time, it is suggested to
check this item.
5-49
Error Information of the Analyzer
5.1.12 Clog
Troubleshooting and
Error ID Error Name Error Mechanism
Solution
0x01002010 WBC Clog During the counting 1. Perform "Zap
process, all or part of Apertures", "Flush
the apertures are Apertures" and "Unclog"
clogged, which results operation for several
in the aperture flow times.
0x01002030 RBC Clog
become slower and 2. If the clog error cannot
aperture voltage be removed, perform
higher which report "Probe Cleanser
the clog error. Maintenance".
5-50
Error Information of the Analyzer
Troubleshooting and
Error ID Error Name Error Mechanism
Solution
3. If the clog error cannot
be removed, add the
probe cleanser into the
bath directly and soak for
10 minutes.
4. If the clog error cannot
be removed, then take off
the aperture and soak it in
the probe cleanser and
clean the aperture
manually.
5-51
6 Fluidic System
The system flow chart of predilute CD mode is as follows: (there is no DIFF or optical channel
of the CBC mode)
6-1
Fluidic System
M-53LH lyse: dissolves RBC, PLT, and differentiates basophils from other WBCs by
size.
Diluent: cleans the system, and provides environment for reaction and analysis.
Dilution ratio:1:500*1
1Dilution ratio in this chapter refers to the dilution ratio of whole blood mode, if not otherwise
stated.
6-2
Fluidic System
Function description: 6μl of blood, 2.5ml of diluent and 0.5ml of LH lyse are mixed
and reacted in the counting bath, and then the compound is aspirated into the back
bath via the aperture by vacuum of the vacuum chamber, the cells are analyzed
when they passes the aperture.
HGB analysis
Reagents used:
Analysis parameter:HGB
Dilution ratio:1:500
The analysis principle of HGB channel is colorimetric method. By comparing the scatter light
intensity of the diluent and the sample, the HGB concentration can be obtained.
Dilution ratio:1:138
Measuring volume: the sample flow is stable, by controlling the analysis duration,
the measuring volume can be calculated.
Function description: 0.6ml of Leo(I) lyse is dispensed to wash the DIFF bath, then
1.1 ml of Leo(I)lyse is added before 9μl of blood is dispensed. 0.135ml of Leo(II) lyse
is added after the blood and Leo(I) react for a while. After a while again, the sample
will be delivered to below the flow cell, the sheath fluid syringe is started to form
sheath fluid, and the sample syringe is started to push the sample into the flow cell
6-3
Fluidic System
for analysis.
Dilution ratio:1:20000
Function description: the sample probe aspirates 52.08μl of sample (dilution rate:
1: 417.7) from the WBC bath; the sample probe moves to RBC bath to mix the
sample with 2.5ml of diluent, the dilution rate of the sample is1:20000. The sample
is then aspirated into the bath back through the aperture by vacuum of the vacuum
chamber, the cells are analyzed when they pass the aperture.
6-4
Fluidic System
Preheating Reactio
Item Optical system
bath n bath
Target temperature ℃ 30 35 36
Normal startup 83 2 2 2 0
6-5
Fluidic System
Appearance:
Pin lift
Function:
Two-way valve: connects and disconnects channels. When the electromagnetic valve is not
electrified, the input and output ports are blocked; when the electromagnetic valve is electrified,
the input and output ports are connected.
Three-way valve: switches channels. When the electromagnetic valve is not electrified, the
common port and normally open port are connected; when the electromagnetic valve is
electrified, the common port and normally closed port are connected.
Note:
The working voltage of the Mindray valve is 12V, and the maximum pressure resistance is
200KPa. The Mindray valve works via the electromagnet, and restores by using spring, it shall
not be electrified for long term. When the electromagnetic valve is electrified, the pin-lift of the
valve goes down; when the valve is not electrified, the pin-lift restores. By touching the pin-lift,
you can feel its movement to judge its status.
6-6
Fluidic System
Appearance:
Function:
Pressure resistance of the two-way pressureproof Mindray valve is greater than that of the
ordinary two-way Mindray valve, their operating principle is the same.
Note: Be sure to distinguish the ordinary two-way valve and the pressureproof two-way valve
when replacing the valves.
Appearance:
Function:
Same as the Mindray valves. The pressure resistance of this valve is greater than that of the
two-way pressureproof Mindray valve, so it can adapt to greater temperature and pressure
change.
Note:
6-7
Fluidic System
PV28
Appearance:
Function:
Clamp type, on/off controlled by electromagnetic power. The valve controls the flow and break
of fluid.
6-8
Fluidic System
Appearance:
LF1 LF2
6.5.6 Syringes
Symbols:
A A
A
Appearance: N/A.
Function:
There are 5 types of syringes, there parameters and functions are as follows:
6-9
Fluidic System
Sample Full range The syringe pushes sample into the flow cell
Sp-Syringe
syringe 250μl for measurement.
6-10
Fluidic System
6.5.7 Pump
6.5.7.1Pressure pump
Symbols:
GP
Appearance:
6.5.7.2Vacuum pump
Symbols:
LP2
LP3
Appearance:
Function:
LP2: the pump drains the DIFF bath and vacuum chamber, and generates vacuum of the
vacuum chamber.
LP3: the pump drains the probe wipe, WBC bath and RBC bath.
6-11
Fluidic System
Appearance:
Function:
Provides a corrosion resistant cavity, which is able to aspirate and dispense blood and probe
cleanser.
Note:
The closed tube sample probe has piercing function, its probe tip is conical.
6-12
Fluidic System
Appearance:
Function:
Provides a cavity to clean closed tube sample probe under the effect of liquid flow, and to
collect the waste produced.
Note:
6-13
Fluidic System
RV
Appearance:
Block plate
Photocouple
r
Function:
Detects the pressure in the sheath fluid channel, when the pressure goes beyond specified
range, the photocoulper will be blocked, and alarm signal will be given.
6-14
Fluidic System
6.5.11 Baths
WBC bath: the WBC bath is formed by the front bath, back bath and aperture. The
bath provides space for WBC sample mixing and reaction, HGB and WBC
measurement.
RBC bath: the RBC bath is formed by the front bath, back bath and aperture. The
bath provide space for RBC sample mixing and reaction, RBC/PLT measurement.
DIFF reaction bath: provides space for DIFF sample mixing and reaction, and
provides prepared DIFF sample.
Vacuum chamber: generates and stores stable vacuum for WBC and RBC
impedance counting, cleans the back bath, and drains the flow cell when is it
clogged to remove impurities.
Pressure chamber: generates and stores stable pressure for bubble generation of
the baths and aperture flushing.
RBC isolating chamber: provides room to block interference signals from outside
6-15
Fluidic System
SV18
C84 (T23)
C15 J19-T24-J20
J44-T174-J45 T175 J1-T11-J2 C81 C80 C45 C49 DILUENT
T170 T169 T4
A C14
DIL
T25 C46 C50 LEO(I) LYSE
T7 T3
DIFF bath LEO(I)
LEO(II) LYSE
J39-T14 -J11
C83 C1 C47 C51
J40-T15-J12
T6 T2
SV17
J17-T22-J18
LEO(II)
LF2 C2 C48 C52 LH LYSE
(T21)
J3-T12 -J4 T5 T1
C12 C13 C77
T164
LH
T131
J25-T26-J26
T173
T30 T31 T28 T29
T27
Flow cell
T49
T136
T70
PV28
T171
T172
J7-T39-J8 J21-T40-J22 T16 T73 -T132
T8
C37
T10
T134
C26 T43 J16-T17-J41 C82 T133 T71 T72 PC
T9
T33
J5-T18 -J6
T32
J9-T13-J10
J15
B T36 C27 GF2
T35
T37
J42-T165-J43
T34 C3
C29 C28 SV10
T75-T76
C4
T19 J13-T20-J14
T41
T45
SP(250uL)
C30
T77
SH(10mL)
Isolating chamber 1
T47
C68
LYSE(2.5mLX3)
T140
T155
SV08
C5 SV09
J31-T137-J32
J33-T138-J34
J35-T139-J36
C78 T68
T156
C64 T166
T149
T78
C42 Transducer C69 C24
SV13
C18
T79
T62
T168
T82
T81
C65
T146
C67 C19
T64
T83
T48
T69
T152
T63
T153
DH 名称:
SV07
T61
VC
T162 ASP(100uL)
C66 T65
T151 T58
T59 SV19
T54
SV01 SV03 SV02 TS
T66
C75 C76 SV22
T154
T163
T53
T67
T55
(T148)
J37-T52-J38
T51
T84
T108
T56
T96
C9
T85
C20 T104 C33
T92
C32 T102
T160
T90 T94 T106 C74
(T147)
T161
T95 T107
SV27 SV26 C53 C55 C59
DIL(10mL)
T103 C54
T91 C58
T127
LEO(II) reagent
T93
LEO(I) reagent
C21 C23
T144
T87
LH reagent
container
container
T143
container
T141
T142
J29-T101-J30
J27-T89-J28
(T100)
(T88)
T113
T99
T159
SV23
T158
SV25
SV24
T130
T117
T122
C63
C10
T128 T120 T119 T118 T115
C39 C40
LP2
Waste C43 C11
T129 T121 T116
container
LP3
F 保密:此图及其全部知识产权(含著作权)归深圳迈瑞生物医疗电子股份有限公司所有。未经深圳迈瑞生物医疗电子股份有限公司预先书面许可,严禁出于任何目的,对此图的全部或部分内容(包括但不限于图中信息、数据、运算结果等)泄露、使用、拷贝或复制。
Classified documents, This set of drawing(s) and all it's intellectual property rights (including copyright) subsisting herein are property of Shenzhen Mindray Bio-medical Electronics Co.,Ltd. No disclosure,use,copies or reproductions
should be made of this drawing or any part(s) thereof for whatever purpose nor shall any information, data, calculations, or other contents contained in this drawing be disseminated without prior written permission of Shenzhen Mindray
Bio-medical Electronics Co.,Ltd
6-16
Fluidic System
T45
T46
SV14 Diluent
SV13
T57
C64 SV01 SV03 SV02
T150
C65
T149
T163
C76 T53
T34 T55
T48
T47
C44 T50
T56
C27
J37-T52-
J38
T35 T51
T162 ASP(100uL)
C28
T36
DIL(10mL)
C29
T37
T127
C75
SPB
SH(10mL)
T123
T126
LF1 Waste
Probe pump
wipe
Major functions:
Sample aspiration and dispensing. The ASP syringe and sample probe SPB works together to
aspirate 20uL of blood2, and dispenses 9uL, 6uL and 52.08uL of blood sample respectively.
Cleaning inside and outside of the sample probe. Cleaning inside of the sample probe can be
done by DIL syringe, SV01, 02, and 03 valves working together, it can also be done by SH
syringe, SV13 and 14 valves working together. Cleaning outside of the sample probe is done
by DIL syringe, SV02 and 03 valves working together. Waste is collected by the probe wipe
and waste pump.
Aspiration and dispensing probe cleanser. When aspirating probe cleanser, SV13 and SV14 is
electrified, the SH syringe aspirates 3.6mL of probe cleanser from the sample probe SPB. The
probe cleanser is stored in the reservoir T47 (marked in red in the figure above). When
dispensing probe cleanser, the sample aspirating assembly delivers the sample probe to the
reaction baths (DIFF, WBC and RBC bath), SV13 and SV14 is electrified, the SH syringe
dispenses certain amount of probe cleanser to the counting bath.
refers to whole blood CD mode. The sample aspiration volume of whole blood CBC mode is 15uL,
and for predilute CD and CBC mode, 80uL and 40uL of diluted sample shall be aspirated respectively. If
not otherwise stated, it refers to whole blood CD mode in the following text.
6-17
Fluidic System
T153
T152
J9-T13-
J10
DH T58
SV07 C2
C66
T151
T54
SV01 SV03 SV02
T5 C52 C48
T53 T1
T55
T49
SV04
WBC
J31-T137-
C20 SV22
T56
T92
C32 diluent
J32
T90 Reagent
T10
T96
T61
T94
detection
T57
T91 SV21Vacuum
Diluent T93 T95
C21
T62chamber C53
DIL(10mL)
C56
Bath shielding
J27-T89-
Pressur
(T88)
LH reagent
SV12 e
J28
cover
container
T141
chamber
T114
Waste
T97 SV23
Isolating pump
T157
T117
T99
T98 LYSE(2.5mLX3)
chamber 2 C70
6.6.2.1WBC analysis
The diluent syringe dispenses diluent into the WBC bath along the blue paths, the sample
probe dispenses blood sample into the WBC bath, and the lyse syringe dispenses LH lyse into
the WBC bath along the peach paths. The sample, diluent and lyse are mixed by bubbles
generated by the pressure chamber via SV12, and aspirated into the back bath by vacuum via
the aperture (wine lines). The calls are measured when they passes the aperture, and the
sample volume is calculated from the analysis duration. After analysis, the WBC bath is
drained by SV23 and the waste pump.
6.6.2.2HGB analysis
The analysis principle of the HGB channel is colorimetric method. By comparing the intensity
of the transmitted light through the background and the blood, the HGB concentration can be
obtained. The intensity of the transmitted light through the diluent is detected first, and during
WBC analysis, the intensity of the transmitted light through the blood sample is detected, the
ratio of the two intensity values is the HGB result.
6-18
Fluidic System
J39-T14 -
T31 T28 T29
J11
J40-T15-
LF2 DIFF bath C1 C47 C51
T6
J12
J25-T26- T2
T131
(T21)
J26 J3-T12 -
SV0 SV0
J35-T139-
C12 C48 C52
T27
J4
6
J36
C77 Flow 5
T164
T30
J17-T22-
cell
J33-T138-
Reagent
J18
T8
T9
J34
RV SV15 detection
T32
J7-T39-
J42-T165- PV28
J8 J21-T40- T16 C54 C55
SV14
J43
J22 T44 C57 C58
T34 J16-T17-
LEO(I) reagent
T33
C26 T43
J41 C25
T45
container
C29
container
reagent
T143
LEO(II)
T142
diluent
J23-T42-
J5-T18 -
T35
T38 J24
J6
C28
T41
T36
C27 SV16
T37
T47
LYSE(2.5mLX3)
C64
SV1
T149
Pressur
SP(250uL)
0
SV08
C4 J13-T20- e
T69
T19
SH(10mL)
J14 chambe
T68
SV13 Waste
r
T160
Isolating
T87
chamber 1 pump
SV27
T166
Vacuum
chamber
Figure 6-1
Leo(I) lyse is dispensed into the DIFF bath by lyse syringe via T11, Leo(II) lyse is dispensed
into the DIFF bath by lyse syringe via T12. Bubbles are dispensed from the pressure chamber
to DIFF bath via SV10, after reaction, the sample is transmitted under the flow cell. The sheath
fluid syringe is started to form sheath fluid along the blue lines in the figure. The sample is
driven to the flow cell for analysis by the sample syringe. After the analysis, the sheath fluid
syringe cleans the flow cell and sample preparing tubes along the blue lines in the figure. The
DIFF bath is drained by waste pump and SV27.
C67
T153
T152
DH T58
SV07
C66
T151
T54
SV01 SV03 SV02
T53
T55
T92
C32 diluent
T90
T96
T61
T94
T57
T91 SV20Vacuum
diluent T93 T95
C21
T62chamber
DIL(10mL)
Bath shielding
J27-T89-
Pressur
(T88)
SV11 e
J28
cover
chamber
T114
Waste
T97 SV24
Isolating pump
T157
T117
T99
T98
chamber 2 C70
Figure 6-2
6-19
Fluidic System
The diluent syringe dispenses diluent into the RBC bath along the blue paths, the sample
probe dispenses diluted blood sample into the RBC bath. The sample and diluent are mixed by
bubbles generated by the pressure chamber via SV11, and aspirated into the back bath by
vacuum via the aperture (wine lines). The calls are measured when they passes the aperture,
and the sample volume is calculated from the analysis duration. After analysis, the RBC bath is
drained by SV24 and the waste pump.
6-20
Fluidic System
0 5 0 5 10 15 20 25 30 35 40 45 50 55 60
1.24
SNH-MOTOR SNH-MOTOR
1.33 1.9 2.9 5.556.15
SNV-MOTOR SNV-MOTOR
5.9
15.2 15.8 18.3 47.1
Sample probe
7.65 7.6 20.2 12.5 22.95
ASP- +2 5 +2 14 +3 5
+20 5 -9 11 -0.5 3 -6 5 -9 14 +52.08 ASP-
14 INIT 14
SYRINGE SYRINGE
V01 1.65 1.92 22.85 25.45 V01
V03 14.5 20.1 25.8
2.95 5.6 9.5 2.8 4.98 6.6 9.4 10.8 V03
16.9 21.9 27.6
V02 47.5
1.65 1.92 2.95 5.6 9.5 2.8 4.98 6.6 9.4 27.6 43.8 45.6 V02
51.1
V13 4.45 5.2 15 16.8 V13
V07 17.4 19 49.4 51.1 V07
V25 1.2
9.5 3.3 4.98 6.6 9.1 11.3 14.1 17.4 19.7 22.4
25.4 V25
1.3 1.68 7.355.25 5.7 0.0 5.0 6.3 6.65 14.6 20.2
27.5 27.2
25.9 38 43.8 46.2 47.6
3 +200 9.5 11.6 16.1 17.5 22.95 +200 49.6
+2001.95
8 4 -2300 7 8 5 -1300
+50 3 +4500 -1647.92 -800 2 -700 6 8 INIT 13
DIL-SYRINGE +1550 15
-1600 6 8 +7650 13 -850 6+100 8 -2500 -600 5 -450 6 10 7
-4100 13 DIL-SYRINGE
-200 6 13 +2300
-100 8 -800 6 10 +8550 10 -4100 13
15. 15. 16. 17. 19.7 13
3.5 4.1 4.5 6.7 43.7 45.3 46.8 49 50.7
-300 5 3.7 1 7 2 2 5
SH-SYRINGE +450 9
-150 5 -300 -450 -7758 2
+1900 10
-1290 7
-1900 +1650 10 INIT 7 SH-SYRINGE
+1200 3 8
+8500 15 7 +250 7
-150 8 +150 8
7
V16 0 3.6 20.6 21.2 46.7 47.4 V16
V15 43.6 V15
17.15 19.7 46.7 48.6 50.6 53.2
45.2
V14 4.45 5.2 15 16.7 17.15 19.7
20.5
46.7 48.6 50.6 53.2 V14
45.2
V17 20.05 48.6 50.6 53.2 V17
V18 20.3 V18
42.7
V28 0 3.8
19.6 V28
0.3 42
4 51.1
DIFF 20.5 21.8
-140 2
INIT 5
SP-SYRINGE +190 12 SP-SYRINGE
+20 14 -60 14
0 2.8 4.2 6.7 12 15 17 20.7 40 49.6 50.7
LYSE- -145 14 INIT 5
+2000 14
-1100 14 +600 11 +50 5 +20 1 LYSE-
-600 -10 11 -50 9 -520 11
SYRINGE SYRINGE
14
V06 2.8 3.8 6.6 8.7 V06
V05 4.1 4.7 11.9 13 49.6 50.6 V05
V27 0 1.5 4.5 6
25.4
23.5 35.2 40.5 46.7 48.6 49.6 V27
27.7
V10 8.1 V10
13.9
DIFF_COUNT 34 DIFF_COUNT
50
V23 6.1 9.1
42.2 45.4 V23
43.8 47.6
V12 13.5 V12
4.6 5.1 20.5 23.6 44.9 45.4
16.1
V04 20.6 21.53 V04
V22 53.65 56.5 V22
WBC V21 26.2
53.65 56.5 V21
42.1
WBC_COUNT 28.9 WBC_COUNT
40.9
HGB_BLANK 2.5 HGB_BLANK
HGB_PARA 41 HGB_PARA
V24 12 48.1
V24
14 49.4
V11 25.8 56.2 V11
26.25 57.1
RBC V19 54 56 V19
V20 26.2
54 56 V20
48
RBC_COUNT 32.1 RBC_COUNT
46.1
V09 57.2 59.5 V09
V26 55.8 V26
58.2
PUMP_PRES PUMP_PRES
S 57.2 59.3 S
BUILD_PRES BUILD_PRES
Public SPUMP_VA
0
25.3
44.5 45.5 54.5 57 S
40.5 55.5
PUMP_VA
C 0 1.5 4.5 6 25.2 27.5 33.5 35 48.6 49.6 C
46.7 58.5
BUILD_VAC 16.5 22.1 49.6 BUILD_VAC
22 25 55.3
PUMP_WAST 1.2 9.4 25.4 41.7 PUMP_WAST
E 19.7 22.4
7.35 17.4 27.5 51 E
CONST 19 CONST
49
SELECT_WBC 18 SELECT_WBC
SELECT_RB SELECT_RB
Zap C 18 C
ZAP_WBC 50 ZAP_WBC
53
ZAP_RBC 51.2 ZAP_RBC
54
6-21
Fluidic System
6-22
Fluidic System
SV18
(T23)
J19-T24- J1-T11-
C15
J2
C45 C49 DILUENT
J20 T4
T25 C14 C46 C50 LEO( I( LYSE
T7 T3
LEO(I)
DIFF bath J3-T12 - C1 C47 C51 LEO( II( LYSE
T6
J39-T14 -
T2
J40-T15-
SV17 J4
LEO(II)
J17-T22-
(T21)
J11
J12
LF2 C2 C48 C52 LH LYSE
J18
C12 C13 T5 T1
J25-T26- C77 Flow LH
T164
T30 T31 T28 T29
T131
J26 cell
T27
Reagent
T136
T70
J21-T40- PV28
T10
T8
J7-T39-J8
T9
J22 T16 C37 T73 -T132
T44 T134
C26 J16-T17- T72 PC
T43 T133 T71
T33
T38 J41
T32
C25
J5-T18 -
T135C41 C16
SV14 SV16 GP T74
J6
J23-T42-
J9-T13-
J24
J15
J10
T36 C27 GF2
T35 SV1
T37
T34 C3
J42-T165-
C29 C28
0
J43
T75-T76
C4 J13-T20-
T19
T41
J14
T45
SP(250uL)
C30
Isolating LYSE(2.5mLX3)
T77
SH(10mL)
chamber 1
T47
C68
T140
T155
SV08
T68
C5
C78
SV09
J31-T137-
J33-T138-
J35-T139-
T156
C64 T166
J32
J34
J36
T149
Transducer
T78
C42 C69 C24
SV13
C18
T79
T62
T168
C65
T82
T81
T146
C67 C19
T64
T83
T48
T69
T152
T153
DH
T66
SV07
T61
VC
T162 ASP(100uL)
C66 T65
T151 T58
T59 SV19
T54
SV01 SV03 SV02 TS
C75 C76 SV22
T154
T53
T163
T67
T55
J37-T52-
(T148)
T51
J38
T84
T108
T56
T96
C9
RBC C22 SV20 C73
WBC C20 SV21
T85
T92 T104 C33
C32 T102
T90 T106
T160
T94 C74
(T147)
T161
T95 T107
SV27 SV26 C53 C55 C59
C54
DIL(10mL)
T91 T103
C56 C57 C58
T127
LEO(I) reagent
C21 C23
LH reagent
T144
container
container
T87
container
reagent
T143
LEO(II)
T141
T142
J29-T101-
J27-T89-
(T100)
Bath shielding
(T88)
J30
C31 T86
cover SV12 cover SV11
LF1
C6
Probe T126 T125 C34
Isolating T97 Isolating T109 T111
T124
wipe
T98 chamber 3 T110
chamber 2 T112
T113
T99
C7
T114 C35
T123
C62 C61
C70
C72 C71
T157
T145
T159
SV23
T158
SV25
SV24
T130
T117
T122
C63
C10
T128 T120 T119 T118 T115
C39 C40
Waste LP2
C43 C11
containe T129 T121 T116
r LP3
6-23
Fluidic System
SV18
(T23)
J19-T24- J1-T11-
C15
J2
C45 C49 DILUENT
J20 T4
T25 C14 C46 C50 LEO( I( LYSE
T7 T3
LEO(I)
DIFF bath J3-T12 - C1 C47 C51 LEO( II( LYSE
T6
J39-T14 -
T2
J40-T15-
SV17 J4
LEO(II)
J17-T22-
(T21)
J11
J12
LF2 C2 C48 C52 LH LYSE
J18
C12 C13 T5 T1
J25-T26- C77 Flow LH
T164
T30 T31 T28 T29
T131
J26 cell
T27
Reagent
T136
T70
J21-T40- PV28
T10
T8
J7-T39-J8
T9
J22 T16 C37 T73 -T132
T44 T134
C26 J16-T17- T72 PC
T43 T133 T71
T33
T38 J41
T32
C25
J5-T18 -
T135C41 C16
SV14 SV16 GP T74
J6
J23-T42-
J9-T13-
J24
J15
J10
T36 C27 GF2
T35 SV1
T37
T34 J42-T165- C3
C29 C28
J43 0
T75-T76
C4 J13-T20-
T19
T41
J14
T45
SP(250uL)
C30
Isolating LYSE(2.5mLX3)
T77
SH(10mL)
chamber 1
T47
C68
T140
T155
SV08
T68
C5
C78
SV09
J31-T137-
J33-T138-
J35-T139-
T156
C64 T166
J32
J34
J36
T149
Transducer
T78
C42 C69 C24
SV13
C18
T79
T62
T168
C65
T82
T81
T146
C67 C19
T64
T83
T48
T69
T152
T153
DH
T66
SV07
T61
VC
T162 ASP(100uL)
C66 T65
T151 T58
T59 SV19
T54
SV01 SV03 SV02 TS
C75 C76 SV22
T154
T53
T163
T67
T55
J37-T52-
(T148)
T51
J38
T84
T108
T56
T96
C9
RBC C22 SV20 C73
WBC C20 SV21
T85
T92 T104 C33
C32 T102
T90 T106
T160
T94 C74
(T147)
T161
T95 T107
SV27 SV26 C53 C55 C59
C54
DIL(10mL)
T91 T103
C56 C57 C58
T127
LEO(I) reagent
C21 C23
LH reagent
T144
container
container
T87
container
reagent
T143
LEO(II)
T141
T142
J29-T101-
J27-T89-
(T100)
Bath shielding
(T88)
J30
C31 T86
cover SV12 cover SV11
LF1
C6
Probe T126 T125 C34
Isolating T97 Isolating T109 T111
T124
wipe
T98 chamber 3 T110
chamber 2 T112
T113
T99
C7
T114 C35
T123
C62 C61
C70
C72 C71
T157
T145
T159
SV23
T158
SV25
SV24
T130
T117
T122
C63
C10
T128 T120 T119 T118 T115
C39 C40
Waste LP2
C43 C11
containe T129 T121 T116
r LP3
6-24
Fluidic System
SV18
(T23)
J19-T24- J1-T11-
C15
J2
C45 C49 DILUENT
J20 T4
T25 C14 C46 C50 LEO( I( LYSE
T7 T3
LEO(I)
DIFF bath J3-T12 - C1 C47 C51 LEO( II( LYSE
T6
J39-T14 -
T2
J40-T15-
SV17 J4
LEO(II)
J17-T22-
(T21)
J11
J12
LF2 C2 C48 C52 LH LYSE
J18
C12 C13 T5 T1
J25-T26- C77 Flow LH
T164
T30 T31 T28 T29
T131
J26 cell
T27
Reagent
T136
T70
J21-T40- PV28
T10
T8
J7-T39-J8
T9
J22 T16 C37 T73 -T132
T44 T134
C26 J16-T17- T72 PC
T43 T133 T71
T33
T38 J41
T32
C25
J5-T18 -
T135C41 C16
SV14 SV16 GP T74
J6
J23-T42-
J9-T13-
J24
J15
J10
T36 C27 GF2
T34 T35
T37 SV1 C3
J42-T165-
C29 C28
0
J43
T75-T76
C4 J13-T20-
T19
T41
J14
T45
SP(250uL)
C30
Isolating LYSE(2.5mLX3)
T77
SH(10mL)
T47
chamber 1
C68
T140
T155
SV08
T68
C5
C78
SV09
J31-T137-
J33-T138-
J35-T139-
T156
C64 T166
J32
J34
J36
T149
Transducer
T78
C42 C69 C24
SV13
C18
T79
T62
T168
C65
T82
T81
T146
C67 C19
T64
T83
T48
T69
T152
T153
DH
T66
SV07
T61
VC
T162 ASP(100uL)
C66 T65
T151 T58
T59 SV19
T54
SV01 SV03 SV02 TS
C75 C76 SV22
T154
T53
T163
T67
T55
J37-T52-
(T148)
T51
J38
T84
T108
T56
T96
C9
RBC C22 SV20 C73
WBC C20 SV21
T85
T92 T104 C33
C32 T102
T90 T106
T160
T94 C74
(T147)
T161
T95 T107
SV27 SV26 C53 C55 C59
C54
DIL(10mL)
T91 T103
C56 C57 C58
T127
LEO(I) reagent
C21 C23
LH reagent
T144
container
container
T87
container
reagent
T143
LEO(II)
T141
T142
J29-T101-
J27-T89-
(T100)
Bath shielding
(T88)
J30
C31 T86
cover SV12 cover SV11
LF1
C6
Probe T126 T125 C34
Isolating T97 Isolating T109 T111
T124
wipe
T98 chamber 3 T110
chamber 2 T112
T113
T99
C7
T114 C35
T123
C62 C61
C70
C72 C71
T157
T145
T159
SV23
T158
SV25
SV24
T130
T117
T122
C63
C10
T128 T120 T119 T118 T115
C39 C40
Waste LP2
C43 C11
containe T129 T121 T116
r LP3
6-25
Fluidic System
5.9~6.1sThe ASP syringe generates 2uL of isolating bubbles in the sample probe
6.2~7.5sThe pierce probe moves to the aspirate position.
7.6~10sThe ASP syringe aspirates 20uL of blood sample from the sample probe.
9.4sStart waste pump LP2
SV18
(T23)
J19-T24- J1-T11-
C15
J2
C45 C49 DILUENT
J20 T4
T25 C14 C46 C50 LEO( I( LYSE
T7 T3
LEO(I)
DIFF bath J3-T12 - C1 C47 C51 LEO( II( LYSE
T6
J39-T14 -
T2
J40-T15-
SV17 J4
LEO(II)
J17-T22-
(T21)
J11
J12
LF2 C2 C48 C52 LH LYSE
J18
C12 C13 T5 T1
J25-T26- C77 Flow LH
T164
T30 T31 T28 T29
T131
J26 cell
T27
Reagent
T136
T70
J21-T40- PV28
T10
T8
J7-T39-J8
T9
J22 T16 C37 T73 -T132
T44 T134
C26 J16-T17- T72 PC
T43 T133 T71
T33
T38 J41
T32
C25
J5-T18 -
T135C41 C16
SV14 SV16 GP T74
J6
J23-T42-
J9-T13-
J24
J15
J10
T36 C27 GF2
T35 SV1
T37
T34 C3
J42-T165-
C29 C28
0
J43
T75-T76
C4 J13-T20-
T19
T41
J14
T45
SP(250uL)
C30
Isolating LYSE(2.5mLX3)
T77
SH(10mL)
chamber 1
T47
C68
T140
T155
SV08
T68
C5
C78
SV09
J31-T137-
J33-T138-
J35-T139-
T156
C64 T166
J32
J34
J36
T149
Transducer
T78
C42 C69 C24
SV13
C18
T79
T62
T168
C65
T82
T81
T146
C67 C19
T64
T83
T48
T69
T152
T153
DH
T66
SV07
T61
VC
T162 ASP(100uL)
C66 T65
T151 T58
T59 SV19
T54
SV01 SV03 SV02 TS
C75 C76 SV22
T154
T53
T163
T67
T55
J37-T52-
(T148)
T51
J38
T84
T108
T56
T96
C9
RBC C22 SV20 C73
WBC C20 SV21
T85
T92 T104 C33
C32 T102
T90 T106
T160
T94 C74
(T147)
T161
T95 T107
SV27 SV26 C53 C55 C59
C54
DIL(10mL)
T91 T103
C56 C57 C58
T127
LEO(I) reagent
C21 C23
LH reagent
T144
container
container
T87
container
reagent
T143
LEO(II)
T141
T142
J29-T101-
J27-T89-
(T100)
Bath shielding
(T88)
J30
C31 T86
cover SV12 cover SV11
LF1
C6
Probe T126 T125 C34
Isolating T97 Isolating T109 T111
T124
wipe
T98 chamber 3 T110
chamber 2 T112
T113
T99
C7
T114 C35
T123
C62 C61
C70
C72 C71
T157
T145
T159
SV23
T158
SV25
SV24
T130
T117
T122
C63
C10
T128 T120 T119 T118 T115
C39 C40
Waste LP2
C43 C11
containe T129 T121 T116
r LP3
6-26
Fluidic System
6-27
Fluidic System
SV18
(T23)
J19-T24- J1-T11-
C15
J2
C45 C49 DILUENT
J20 T4
T25 C14 C46 C50 LEO( I( LYSE
T7 T3
LEO(I)
DIFF bath J3-T12 - C1 C47 C51 LEO( II( LYSE
T6
J39-T14 -
T2
J40-T15-
SV17 J4
LEO(II)
J17-T22-
(T21)
J11
J12
LF2 C2 C48 C52 LH LYSE
J18
C12 C13 T5 T1
J25-T26- C77 Flow LH
T164
T30 T31 T28 T29
T131
J26 cell
T27
Reagent
T136
T70
J21-T40- PV28
T10
T8
J7-T39-J8
T9
J22 T16 C37 T73 -T132
T44 T134
C26 J16-T17- T72 PC
T43 T133 T71
T33
T38 J41
T32
C25
J5-T18 -
T135C41 C16
SV14 SV16 GP T74
J6
J23-T42-
J9-T13-
J24
J15
J10
T36 C27 GF2
T35 SV1
T37
T34 C3
J42-T165-
C29 C28
0
J43
T75-T76
C4 J13-T20-
T19
T41
J14
T45
SP(250uL)
C30
LYSE(2.5mLX3)
Isolating
T77
SH(10mL)
T47
chamber 1 C68
T140
T155
SV08
T68
C5
C78
SV09
J31-T137-
J33-T138-
J35-T139-
T156
C64 T166
J32
J34
J36
T149
Transducer
T78
C42 C69 C24
SV13
C18
T79
T62
T168
C65
T82
T81
T146
C67 C19
T64
T83
T48
T69
T152
T153
DH
T66
SV07
T61
VC
T162 ASP(100uL)
C66 T65
T151 T58
T59 SV19
T54
SV01 SV03 SV02 TS
C75 C76 SV22
T154
T53
T163
T67
T55
J37-T52-
(T148)
T51
J38
T84
T108
T56
T96
C9
RBC C22 SV20 C73
WBC C20 SV21
T85
T92 T104 C33
C32 T102
T90 T106
T160
T94 C74
(T147)
T161
T95 T107
SV27 SV26 C53 C55 C59
C54
DIL(10mL)
T91 T103
C56 C57 C58
T127
LEO(I) reagent
C21 C23
LH reagent
T144
container
container
T87
container
reagent
T143
LEO(II)
T141
T142
J29-T101-
J27-T89-
(T100)
Bath shielding
(T88)
J30
C31 T86
cover SV12 cover SV11
LF1
C6
Probe T126 T125 C34
Isolating T97 Isolating T109 T111
T124
wipe
T98 chamber 3 T110
chamber 2 T112
T113
T99
C7
T114 C35
T123
C62 C61
C70
C72 C71
T157
T145
T159
SV23
T158
SV25
SV24
T130
T117
T122
C63
C10
T128 T120 T119 T118 T115
C39 C40
LP2
Waste C43 C11
T129 T121 T116
container
LP3
6-28
Fluidic System
SV18
(T23)
J19-T24- J1-T11-
C15
J2
C45 C49 DILUENT
J20 T4
T25 C14 C46 C50 LEO( I( LYSE
T7 T3
LEO(I)
DIFF bath J3-T12 - C1 C47 C51 LEO( II( LYSE
T6
J39-T14 -
T2
J40-T15-
SV17 J4
LEO(II)
J17-T22-
(T21)
J11
J12
LF2 C2 C48 C52 LH LYSE
J18
C12 C13 T5 T1
J25-T26- C77 Flow LH
T164
T30 T31 T28 T29
T131
J26 cell
T27
Reagent
T136
T70
J21-T40- PV28
T10
T8
J7-T39-J8
T9
J22 T16 C37 T73 -T132
T44 T134
C26 J16-T17- T72 PC
T43 T133 T71
T33
T38 J41
T32
C25
J5-T18 -
T135C41 C16
SV14 SV16 GP T74
J6
J23-T42-
J9-T13-
J24
J15
J10
T36 C27 GF2
T34 T35 T37 SV1 C3
J42-T165-
C29 C28
0
J43
T75-T76
C4 J13-T20-
T19
T41
J14
T45
SP(250uL)
C30
Isolating LYSE(2.5mLX3)
T77
SH(10mL)
chamber 1
T47
C68
T140
T155
SV08
T68
C5
C78
SV09
J31-T137-
J33-T138-
J35-T139-
T156
C64 T166
J32
J34
J36
T149
Transducer
T78
C42 C69 C24
SV13
C18
T79
T62
T168
C65
T82
T81
T146
C67 C19
T64
T83
T48
T69
T152
T153
DH
T66
SV07
T61
VC
T162 ASP(100uL)
C66 T65
T151 T58
T59 SV19
T54
SV01 SV03 SV02 TS
C75 C76 SV22
T154
T53
T163
T67
T55
J37-T52-
(T148)
T51
J38
T84
T108
T56
T96
C9
RBC C22 SV20 C73
WBC C20 SV21
T85
T92 T104 C33
C32 T102
T90 T106
T160
T94 C74
(T147)
T161
T95 T107
SV27 SV26 C53 C55 C59
C54
DIL(10mL)
T91 T103
C56 C57 C58
T127
LEO(I) reagent
C21 C23
LH reagent
T144
container
container
T87
container
reagent
T143
LEO(II)
T141
T142
J29-T101-
J27-T89-
(T100)
Bath shielding
(T88)
J30
C31 T86
cover SV12 cover SV11
LF1
C6
Probe T126 T125 C34
Isolating T97 Isolating T109 T111
T124
wipe
T98 chamber 3 T110
chamber 2 T112
T113
T99
C7
T114 C35
T123
C62 C61
C70
C72 C71
T157
T145
T159
SV23
T158
SV25
SV24
T130
T117
T122
C63
C10
T128 T120 T119 T118 T115
C39 C40
LP2
Waste C43 C11
T129 T121 T116
container
LP3
6-29
Fluidic System
10.8~11.3The sample aspirating assembly moves to upper position of the WBC bath
11.4~7.6sThe sample probe goes down to the WBC bath
12.5~14.2sThe sample probe wiggles
12.5~13.5sDispense 6uL of blood sample into the WBC bath with the ASP syringe
13.5~16.1sTurn on and off V12 continuously to generate bubbles to mix the sample in WBC bath
11.6~13.1sDispense 2500uL of diluent into the WBC bath with the DIL syringe
12~13sDispense 135uL of LEO(II) lyse into the DIFF bath with the lyse syringe
6-30
Fluidic System
SV18
(T23)
C15
J19-T24-J20 J1-T11-J2 C45 C49 DILUENT
T4
T25 C14 C46 C50 LEO(I) LYSE
T7 T3
LEO(I)
LEO(II) LYSE
J39-T14 -J11
DIFF bath C1 C47 C51
J40-T15-J12
J3-T12 -J4 T6 T2
SV17
J17-T22-J18
(T21)
LEO(II)
LF2 C2 C48 C52 LH LYSE
C12 C13 T5 T1
C77 Flow
T164
J25-T26-J26 LH
T131
T30 T31 T28 T29
cell
T27
T49
Reagent
T136
T70
PV28
T10
T8
J21-T40-J22
T9
J7-T39-J8 T16 C37 T73 -T132
T44 T134
C26 T43 J16-T17-J41 T133 T71 T72 PC
T33
J5-T18 -J6
T32
J23-T42-J24
SV14 GP
J9-T13-J10
SV16 T74
J15
T35 T36 C27 GF2
T37
J42-T165-J43
T34 C3
C29 C28 SV10
T75-T76
C4
T19 J13-T20-J14
T41
T45
SP(250uL)
C30
Isolating LYSE(2.5mLX3)
T77
SH(10mL)
chamber 1
T47
C68
T140
T155
SV08
T68
C5 SV09
J31-T137-J32
J33-T138-J34
J35-T139-J36
C78
T156
C64 T166
T149
T78
C42 Transducer C69 C24
SV13
C18
T79
T62
T168
T82
T81
C65
T146
C67 C19
T64
T83
T48
T69
T152
T153
T63
T66
DH
SV07
T61
VC
T162 ASP(100uL)
C66 T65
T151 T58
T59 SV19
T54
SV01 SV03 SV02 TS
C75 C76 SV22
T154
T163
T53
T67
T55
(T148)
J37-T52-J38
T51
T84
T108
T56
T96
C9
RBC C22 SV20 C73
WBC SV21
T85
C20 T104 C33
T92
C32 T102
T160
T90 T94 T106 C74
(T147)
T161
T95 T107
SV27 SV26 C53 C55 C59
DIL(10mL)
T103 C54
T91 C58
T127
LEO(II) reagent
T93 T105
LEO(I) reagent
C21 C23
LH reagent
T144
T87
container
container
T143
container
T141
T142
J29-T101-J30
J27-T89-J28
(T100)
Bath shielding
(T88)
wipe
T98 chamber 3 T110
chamber 2 T112
T113
T99
C7
T114 C35
T123
C62 C61
C70
C72 C71
T157
T145
T159
SV23
T158
SV25
SV24
T130
T117
T122
C63
C10
T128 T120 T119 T118 T115
C39 C40
LP2
Waste C43 C11
T129 T121 T116
container
LP3
6-31
Fluidic System
6-32
Fluidic System
SV18
(T23)
J19-T24- J1-T11-
C15
J2
C45 C49 DILUENT
J20 T4
T25 C14 C46 C50 LEO( I( LYSE
T7 T3
LEO(I)
DIFF bath J3-T12 - C1 C47 C51 LEO( II( LYSE
T6
J39-T14 -
T2
J40-T15-
SV17 J4
LEO(II)
J17-T22-
(T21)
J11
J12
LF2 C2 C48 C52 LH LYSE
J18
C12 C13 T5 T1
J25-T26- C77 Flow LH
T164
T30 T31 T28 T29
T131
J26 cell
T27
Reagent
T136
T70
J21-T40- PV28
T10
T8
J7-T39-J8
T9
J22 T16 C37 T73 -T132
T44 T134
C26 J16-T17- T72 PC
T33 T43 T133 T71
T38 J41
T32
C25
J5-T18 -
T135C41 C16
SV14 SV16 GP T74
J6
J23-T42-
J9-T13-
J24
J15
J10
T36 C27 GF2
T35 SV1
T37
T34 C3
J42-T165-
C29 C28
0
J43
T75-T76
C4 J13-T20-
T19
T41
J14
T45
SP(250uL)
C30
Isolating LYSE(2.5mLX3)
T77
SH(10mL)
chamber 1
T47
C68
T140
T155
SV08
T68
C5
C78
SV09
J31-T137-
J33-T138-
J35-T139-
T156
C64 T166
J32
J34
J36
T149
Transducer
T78
C42 C69 C24
SV13
C18
T79
T62
T168
C65
T82
T81
T146
C67 C19
T64
T83
T48
T69
T152
T153
DH
T66
SV07
T61
VC
T162 ASP(100uL)
C66 T65
T151 T58
T59 SV19
T54
SV01 SV03 SV02 TS
C75 C76 SV22
T154
T53
T163
T67
T55
J37-T52-
(T148)
T51
J38
T84
T108
T56
T96
C9
RBC C22 SV20 C73
WBC C20 SV21
T85
T92 T104 C33
C32 T102
T90 T106
T160
T94 C74
(T147)
T161
T95 T107
SV27 SV26 C53 C55 C59
C54
DIL(10mL)
T91 T103
C56 C57 C58
T127
LEO(I) reagent
C21 C23
LH reagent
T144
container
container
T87
container
reagent
T143
LEO(II)
T141
T142
J29-T101-
J27-T89-
(T100)
Bath shielding
(T88)
J30
C31 T86
cover SV12 cover SV11
LF1
C6
Probe T126 T125 C34
Isolating T97 Isolating T109 T111
T124
wipe
T98 chamber 3 T110
chamber 2 T112
T113
T99
C7
T114 C35
T123
C62 C61
C70
C72 C71
T157
T145
T159
SV23
T158
SV25
SV24
T130
T117
T122
C63
C10
T128 T120 T119 T118 T115
C39 C40
Waste LP2
C43 C11
containe T129 T121 T116
r LP3
6-33
Fluidic System
6-34
Fluidic System
SV18
(T23)
J19-T24- J1-T11-
C15
J2
C45 C49 DILUENT
J20 T4
T25 C14 C46 C50 LEO( I( LYSE
T7 T3
LEO(I)
DIFF bath J3-T12 - C1 C47 C51 LEO( II( LYSE
T6
J39-T14 -
T2
J40-T15-
SV17 J4
LEO(II)
J17-T22-
(T21)
J11
J12
LF2 C2 C48 C52 LH LYSE
J18
C12 C13 T5 T1
J25-T26- C77 Flow LH
T164
T30 T31 T28 T29
T131
J26 cell
T27
Reagent
T136
T70
J21-T40- PV28
T10
T8
J7-T39-J8
T9
J22 T16 C37 T73 -T132
T44 T134
C26 J16-T17- T72 PC
T33 T43 T133 T71
T38 J41
T32
C25
J5-T18 -
T135C41 C16
SV14 SV16 GP T74
J6
J23-T42-
J9-T13-
J24
J15
J10
T36 C27 GF2
T35 SV1
T37
T34 C3
J42-T165-
C29 C28
0
J43
T75-T76
C4 J13-T20-
T19
T41
J14
T45
SP(250uL)
C30
Isolating LYSE(2.5mLX3)
T77
SH(10mL)
chamber 1
T47
C68
T140
T155
SV08
T68
C5
C78
SV09
J31-T137-
J33-T138-
J35-T139-
T156
C64 T166
J32
J34
J36
T149
Transducer
T78
C42 C69 C24
SV13
C18
T79
T62
T168
C65
T82
T81
T146
C67 C19
T64
T83
T48
T69
T152
T153
DH
T66
SV07
T61
VC
T162 ASP(100uL)
C66 T65
T151 T58
T59 SV19
T54
SV01 SV03 SV02 TS
C75 C76 SV22
T154
T53
T163
T67
T55
J37-T52-
(T148)
T51
J38
T84
T108
T56
T96
C9
RBC C22 SV20 C73
WBC C20 SV21
T85
T92 T104 C33
C32 T102
T90 T106
T160
T94 C74
(T147)
T161
T95 T107
SV27 SV26 C53 C55 C59
C54
DIL(10mL)
T91 T103
C56 C57 C58
T127
LEO(I) reagent
C21 C23
LH reagent
T144
container
container
T87
container
reagent
T143
LEO(II)
T141
T142
J29-T101-
J27-T89-
(T100)
Bath shielding
(T88)
J30
C31 T86
cover SV12 cover SV11
LF1
C6
Probe T126 T125 C34
Isolating T97 Bath shielding T109 T111
T124
wipe
T98 cover 3 T110
chamber 2 T112
T113
T99
C7
T114 C35
T123
C62 C61
C70
C72 C71
T157
T145
T159
SV23
T158
SV25
SV24
T130
T117
T122
C63
C10
T128 T120 T119 T118 T115
C39 C40
Waste LP2
C43 C11
containe T129 T121 T116
r LP3
6-35
Fluidic System
6-36
Fluidic System
SV18
(T23)
J19-T24- J1-T11-
C15
J2
C45 C49 DILUENT
J20 T4
T25 C14 C46 C50 LEO( I( LYSE
T7 T3
LEO(I)
DIFF bath J3-T12 - C1 C47 C51 LEO( II( LYSE
T6
J39-T14 -
T2
J40-T15-
SV17 J4
LEO(II)
J17-T22-
(T21)
J11
J12
LF2 C2 C48 C52 LH LYSE
J18
C12 C13 T5 T1
J25-T26- C77 Flow LH
T164
T30 T31 T28 T29
T131
J26 cell
T27
Reagent
T136
T70
J21-T40- PV28
T10
T8
J7-T39-J8
T9
J22 T16 C37 T73 -T132
T44 T134
C26 J16-T17- T72 PC
T33 T43 T133 T71
T38 J41
T32
C25
J5-T18 -
T135C41 C16
SV14 SV16 GP T74
J6
J23-T42-
J9-T13-
J24
J15
J10
T36 C27 GF2
T35 SV1
T37
T34 C3
J42-T165-
C29 C28
0
J43
T75-T76
C4 J13-T20-
T19
T41
J14
T45
SP(250uL)
C30
Isolating LYSE(2.5mLX3)
T77
SH(10mL)
chamber 1
T47
C68
T140
T155
SV08
T68
C5
C78
SV09
J31-T137-
J33-T138-
J35-T139-
T156
C64 T166
J32
J34
J36
T149
Transducer
T78
C42 C69 C24
SV13
C18
T79
T62
T168
C65
T82
T81
T146
C67 C19
T64
T83
T48
T69
T152
T153
DH
T66
SV07
T61
VC
T162 ASP(100uL)
C66 T65
T151 T58
T59 SV19
T54
SV01 SV03 SV02 TS
C75 C76 SV22
T154
T53
T163
T67
T55
J37-T52-
(T148)
T51
J38
T84
T108
T56
T96
C9
RBC C22 SV20 C73
WBC C20 SV21
T85
T92 T104 C33
C32 T102
T90 T106
T160
T94 C74
(T147)
T161
T95 T107
SV27 SV26 C53 C55 C59
C54
DIL(10mL)
T91 T103
C56 C57 C58
T127
LEO(I) reagent
C21 C23
LH reagent
T144
container
container
T87
container
reagent
T143
LEO(II)
T141
T142
J29-T101-
J27-T89-
(T100)
Bath shielding
(T88)
J30
C31 T86
cover SV12 cover SV11
LF1
C6
Probe T126 T125 C34
Isolating T97 Isolating T109 T111
T124
wipe
T98 chamber 3 T110
chamber 2 T112
T113
T99
C7
T114 C35
T123
C62 C61
C70
C72 C71
T157
T145
T159
SV23
T158
SV25
SV24
T130
T117
T122
C63
C10
T128 T120 T119 T118 T115
C39 C40
Waste LP2
C43 C11
containe T129 T121 T116
r LP3
6-37
Fluidic System
6-38
Fluidic System
SV18
(T23)
J19-T24- J1-T11-
C15
J2
C45 C49 DILUENT
J20 T4
T25 C14 C46 C50 LEO( I( LYSE
T7 T3
LEO(I)
DIFF bath J3-T12 - C1 C47 C51 LEO( II( LYSE
T6
J39-T14 -
T2
J40-T15-
SV17 J4
LEO(II)
J17-T22-
(T21)
J11
J12
LF2 C2 C48 C52 LH LYSE
J18
C12 C13 T5 T1
J25-T26- C77 Flow LH
T164
T30 T31 T28 T29
T131
J26 cell
T27
Reagent
T136
T70
J21-T40- PV28
T10
T8
J7-T39-J8
T9
J22 T16 C37 T73 -T132
T44 T134
C26 J16-T17- T72 PC
T33 T43 T133 T71
T38 J41
T32
C25
J5-T18 -
T135C41 C16
SV14 SV16 GP T74
J6
J23-T42-
J9-T13-
J24
J15
J10
T36 C27 GF2
T35 SV1
T37
T34 C3
J42-T165-
C29 C28
0
J43
T75-T76
C4 J13-T20-
T19
T41
J14
T45
SP(250uL)
C30
Isolating LYSE(2.5mLX3)
T77
SH(10mL)
chamber 1
T47
C68
T140
T155
SV08
T68
C5
C78
SV09
J31-T137-
J33-T138-
J35-T139-
T156
C64 T166
J32
J34
J36
T149
Transducer
T78
C42 C69 C24
SV13
C18
T79
T62
T168
C65
T82
T81
T146
C67 C19
T64
T83
T48
T69
T152
T153
DH
T66
SV07
T61
VC
T162 ASP(100uL)
C66 T65
T151 T58
T59 SV19
T54
SV01 SV03 SV02 TS
C75 C76 SV22
T154
T53
T163
T67
T55
J37-T52-
(T148)
T51
J38
T84
T108
T56
T96
C9
RBC C22 SV20 C73
WBC C20 SV21
T85
T92 T104 C33
C32 T102
T90 T106
T160
T94 C74
(T147)
T161
T95 T107
SV27 SV26 C53 C55 C59
C54
DIL(10mL)
T91 T103
C56 C57 C58
T127
LEO(I) reagent
C21 C23
LH reagent
T144
container
container
T87
container
reagent
T143
LEO(II)
T141
T142
J29-T101-
J27-T89-
(T100)
Bath shielding
(T88)
J30
C31 T86
cover SV12 cover SV11
LF1
C6
Probe T126 T125 C34
Isolating T97 Isolating T109 T111
T124
wipe
T98 chamber 3 T110
chamber 2 T112
T113
T99
C7
T114 C35
T123
C62 C61
C70
C72 C71
T157
T145
T159
SV23
T158
SV25
SV24
T130
T117
T122
C63
C10
T128 T120 T119 T118 T115
C39 C40
Waste LP2
C43 C11
containe T129 T121 T116
r LP3
6-39
Fluidic System
6-40
Fluidic System
SV18
(T23)
J19-T24- J1-T11-
C15
J2
C45 C49 DILUENT
J20 T4
T25 C14 C46 C50 LEO( I( LYSE
T7 T3
LEO(I)
DIFF bath J3-T12 - C1 C47 C51 LEO( II( LYSE
T6
J39-T14 -
T2
J40-T15-
SV17 J4
LEO(II)
J17-T22-
(T21)
J11
J12
LF2 C2 C48 C52 LH LYSE
J18
C12 C13 T5 T1
J25-T26- C77 Flow LH
T164
T30 T31 T28 T29
T131
J26 cell
T27
Reagent
T136
T70
J21-T40- PV28
T10
T8
J7-T39-J8
T9
J22 T16 C37 T73 -T132
T44 T134
C26 J16-T17- T72 PC
T33 T43 T133 T71
T38 J41
T32
C25
J5-T18 -
T135C41 C16
SV14 SV16 GP T74
J6
J23-T42-
J9-T13-
J24
J15
J10
T36 C27 GF2
T35 SV1
T37
T34 C3
J42-T165-
C29 C28
0
J43
T75-T76
C4 J13-T20-
T19
T41
J14
T45
SP(250uL)
C30
Isolating LYSE(2.5mLX3)
T77
SH(10mL)
chamber 1
T47
C68
T140
T155
SV08
T68
C5
C78
SV09
J31-T137-
J33-T138-
J35-T139-
T156
C64 T166
J32
J34
J36
T149
Transducer
T78
C42 C69 C24
SV13
C18
T79
T62
T168
C65
T82
T81
T146
C67 C19
T64
T83
T48
T69
T152
T153
DH
T66
SV07
T61
VC
T162 ASP(100uL)
C66 T65
T151 T58
T59 SV19
T54
SV01 SV03 SV02 TS
C75 C76 SV22
T154
T53
T163
T67
T55
J37-T52-
(T148)
T51
J38
T84
T108
T56
T96
C9
RBC C22 SV20 C73
WBC C20 SV21
T85
T92 T104 C33
C32 T102
T90 T106
T160
T94 C74
(T147)
T161
T95 T107
SV27 SV26 C53 C55 C59
C54
DIL(10mL)
T91 T103
C56 C57 C58
T127
LEO(I) reagent
C21 C23
LH reagent
T144
container
container
T87
container
reagent
T143
LEO(II)
T141
T142
J29-T101-
J27-T89-
(T100)
Bath shielding
(T88)
J30
C31 T86
cover SV12 cover SV11
LF1
C6
Probe T126 T125 C34
Isolating T97 Isolating T109 T111
T124
wipe
T98 chamber 3 T110
chamber 2 T112
T113
T99
C7
T114 C35
T123
C62 C61
C70
C72 C71
T157
T145
T159
SV23
T158
SV25
SV24
T130
T117
T122
C63
C10
T128 T120 T119 T118 T115
C39 C40
Waste LP2
C43 C11
containe T129 T121 T116
r LP3
6-41
Fluidic System
6-42
Fluidic System
SV18
(T23)
J19-T24- J1-T11-
C15
J2
C45 C49 DILUENT
J20 T4
T25 C14 C46 C50 LEO( I( LYSE
T7 T3
LEO(I)
DIFF bath J3-T12 - C1 C47 C51 LEO( II( LYSE
T6
J39-T14 -
T2
J40-T15-
SV17 J4
LEO(II)
J17-T22-
(T21)
J11
J12
LF2 C2 C48 C52 LH LYSE
J18
C12 C13 T5 T1
J25-T26- C77 Flow LH
T164
T30 T31 T28 T29
T131
J26 cell
T27
Reagent
T136
T70
J21-T40- PV28
T10
T8
J7-T39-J8
T9
J22 T16 C37 T73 -T132
T44 T134
C26 J16-T17- T72 PC
T33 T43 T133 T71
T38 J41
T32
C25
J5-T18 -
T135C41 C16
SV14 SV16 GP T74
J6
J23-T42-
J9-T13-
J24
J15
J10
T36 C27 GF2
T35 SV1
T37
T34 C3
J42-T165-
C29 C28
0
J43
T75-T76
C4 J13-T20-
T19
T41
J14
T45
SP(250uL)
C30
Isolating LYSE(2.5mLX3)
T77
SH(10mL)
chamber 1
T47
C68
T140
T155
SV08
T68
C5
C78
SV09
J31-T137-
J33-T138-
J35-T139-
T156
C64 T166
J32
J34
J36
T149
Transducer
T78
C42 C69 C24
SV13
C18
T79
T62
T168
C65
T82
T81
T146
C67 C19
T64
T83
T48
T69
T152
T153
DH
T66
SV07
T61
VC
T162 ASP(100uL)
C66 T65
T151 T58
T59 SV19
T54
SV01 SV03 SV02 TS
C75 C76 SV22
T154
T53
T163
T67
T55
J37-T52-
(T148)
T51
J38
T84
T108
T56
T96
C9
RBC C22 SV20 C73
WBC C20 SV21
T85
T92 T104 C33
C32 T102
T90 T106
T160
T94 C74
(T147)
T161
T95 T107
SV27 SV26 C53 C55 C59
C54
DIL(10mL)
T91 T103
C56 C57 C58
T127
LEO(I) reagent
C21 C23
LH reagent
T144
container
container
T87
container
reagent
T143
LEO(II)
T141
T142
J29-T101-
J27-T89-
(T100)
Bath shielding
(T88)
J30
C31 T86
cover SV12 cover SV11
LF1
C6
Probe T126 T125 C34
Isolating T97 Isolating T109 T111
T124
wipe
T98 chamber 3 T110
chamber 2 T112
T113
T99
C7
T114 C35
T123
C62 C61
C70
C72 C71
T157
T145
T159
SV23
T158
SV25
SV24
T130
T117
T122
C63
C10
T128 T120 T119 T118 T115
C39 C40
Waste LP2
C43 C11
containe T129 T121 T116
r LP3
6-43
Fluidic System
43.8~45.6sDispense 4000uL of diluent into the WBC bath with the DIL syringe
44.9~45.4sTurn on and off V12 continuously to generate bubbles in WBC bath
43.7~45.3sClean sample preparing tubes and DIFF bath with the SH syringe
45.4~47.6sStart waste pump LP3 to drain the DIFF bath
6-44
Fluidic System
SV18
(T23)
J19-T24- J1-T11-
C15
J2
C45 C49 DILUENT
J20 T4
T25 C14 C46 C50 LEO( I( LYSE
T7 T3
LEO(I)
DIFF bath J3-T12 - C1 C47 C51 LEO( II( LYSE
T6
J39-T14 -
T2
J40-T15-
SV17 J4
LEO(II)
J17-T22-
(T21)
J11
J12
LF2 C2 C48 C52 LH LYSE
J18
C12 C13 T5 T1
J25-T26- C77 Flow LH
T164
T30 T31 T28 T29
T131
J26 cell
T27
Reagent
T136
T70
J21-T40- PV28
T10
T8
J7-T39-J8
T9
J22 T16 C37 T73 -T132
T44 T134
C26 J16-T17- T72 PC
T43 T133 T71
T33
T38 J41
T32
C25
J5-T18 -
T135C41 C16
SV14 SV16 GP T74
J6
J23-T42-
J9-T13-
J24
J15
J10
T36 C27 GF2
T35 SV1
T37
T34 C3
J42-T165-
C29 C28
0
J43
T75-T76
C4 J13-T20-
T19
T41
J14
T45
SP(250uL)
C30
Isolating LYSE(2.5mLX3)
T77
SH(10mL)
chamber 1
T47
C68
T140
T155
SV08
T68
C5
C78
SV09
J31-T137-
J33-T138-
J35-T139-
T156
C64 T166
J32
J34
J36
T149
Transducer
T78
C42 C69 C24
SV13
C18
T79
T62
T168
C65
T82
T81
T146
C67 C19
T64
T83
T48
T69
T152
T153
DH
T66
SV07
T61
VC
T162 ASP(100uL)
C66 T65
T151 T58
T59 SV19
T54
SV01 SV03 SV02 TS
C75 C76 SV22
T154
T53
T163
T67
T55
J37-T52-
(T148)
T51
J38
T84
T108
T56
T96
C9
RBC C22 SV20 C73
WBC C20 SV21
T85
T92 T104 C33
C32 T102
T90 T106
T160
T94 C74
(T147)
T161
T95 T107
SV27 SV26 C53 C55 C59
C54
DIL(10mL)
T91 T103
C56 C57 C58
T127
LEO(I) reagent
C21 C23
LH reagent
T144
container
container
T87
container
reagent
T143
LEO(II)
T141
T142
J29-T101-
J27-T89-
(T100)
Bath shielding
(T88)
J30
C31 T86
cover SV12 cover SV11
LF1
C6
Probe T126 T125 C34
Isolating T97 Isolating T109 T111
T124
wipe
T98 chamber 3 T110
chamber 2 T112
T113
T99
C7
T114 C35
T123
C62 C61
C70
C72 C71
T157
T145
T159
SV23
T158
SV25
SV24
T130
T117
T122
C63
C10
T128 T120 T119 T118 T115
C39 C40
Waste LP2
C43 C11
containe T129 T121 T116
r LP3
6-45
Fluidic System
SV18
(T23)
J19-T24- J1-T11-
C15
J2
C45 C49 DILUENT
J20 T4
T25 C14 C46 C50 LEO( I( LYSE
T7 T3
LEO(I)
DIFF bath J3-T12 - C1 C47 C51 LEO( II( LYSE
T6
J39-T14 -
T2
J40-T15-
SV17 J4
LEO(II)
J17-T22-
(T21)
J11
J12
LF2 C2 C48 C52 LH LYSE
J18
C12 C13 T5 T1
J25-T26- C77 Flow LH
T164
T30 T31 T28 T29
T131
J26 cell
T27
Reagent
T136
T70
J21-T40- PV28
T10
T8
J7-T39-J8
T9
J22 T16 C37 T73 -T132
T44 T134
C26 J16-T17- T72 PC
T43 T133 T71
T33
T38 J41
T32
C25
J5-T18 -
T135C41 C16
SV14 SV16 GP T74
J6
J23-T42-
J9-T13-
J24
J15
J10
T36 C27 GF2
T35 SV1
T37
T34 C3
J42-T165-
C29 C28
0
J43
T75-T76
C4 J13-T20-
T19
T41
J14
T45
SP(250uL)
C30
Isolating LYSE(2.5mLX3)
T77
SH(10mL)
chamber 1
T47
C68
T140
T155
SV08
T68
C5
C78
SV09
J31-T137-
J33-T138-
J35-T139-
T156
C64 T166
J32
J34
J36
T149
Transducer
T78
C42 C69 C24
SV13
C18
T79
T62
T168
C65
T82
T81
T146
C67 C19
T64
T83
T48
T69
T152
T153
DH
T66
SV07
T61
VC
T162 ASP(100uL)
C66 T65
T151 T58
T59 SV19
T54
SV01 SV03 SV02 TS
C75 C76 SV22
T154
T53
T163
T67
T55
J37-T52-
(T148)
T51
J38
T84
T108
T56
T96
C9
RBC C22 SV20 C73
WBC C20 SV21
T85
T92 T104 C33
C32 T102
T90 T106
T160
T94 C74
(T147)
T161
T95 T107
SV27 SV26 C53 C55 C59
C54
DIL(10mL)
T91 T103
C56 C57 C58
T127
LEO(I) reagent
C21 C23
LH reagent
T144
container
container
T87
container
reagent
T143
LEO(II)
T141
T142
J29-T101-
J27-T89-
(T100)
Bath shielding
(T88)
J30
C31 T86
cover SV12 cover SV11
LF1
C6
Probe T126 T125 C34
Isolating T97 Isolating T109 T111
T124
wipe
T98 chamber 3 T110
chamber 2 T112
T113
T99
C7
T114 C35
T123
C62 C61
C70
C72 C71
T157
T145
T159
SV23
T158
SV25
SV24
T130
T117
T122
C63
C10
T128 T120 T119 T118 T115
C39 C40
Waste LP2
C43 C11
containe T129 T121 T116
r LP3
6-46
Fluidic System
47.6~49.6sDispense 4000uL of diluent into the WBC bath with the DIL syringe
46.8~48.7sClean sample preparing tubes and DIFF bath with the SH syringe
48.6~49.6sStart waste pump LP3 to drain the DIFF bath
48.1~49.4Turn on waste pump LP2 to drain RBC bath
SV18
(T23)
J19-T24- J1-T11-
C15
J2
C45 C49 DILUENT
J20 T4
T25 C14 C46 C50 LEO( I( LYSE
T7 T3
LEO(I)
DIFF bath J3-T12 - C1 C47 C51 LEO( II( LYSE
T6
J39-T14 -
T2
J40-T15-
SV17 J4
LEO(II)
J17-T22-
(T21)
J11
J12
LF2 C2 C48 C52 LH LYSE
J18
C12 C13 T5 T1
J25-T26- C77 Flow LH
T164
T30 T31 T28 T29
T131
J26 cell
T27
Reagent
T136
T70
J21-T40- PV28
T10
T8
J7-T39-J8
T9
J22 T16 C37 T73 -T132
T44 T134
C26 J16-T17- T72 PC
T43 T133 T71
T33
T38 J41
T32
C25
J5-T18 -
T135C41 C16
SV14 SV16 GP T74
J6
J23-T42-
J9-T13-
J24
J15
J10
T36 C27 GF2
T35 SV1
T37
T34 C3
J42-T165-
C29 C28
0
J43
T75-T76
C4 J13-T20-
T19
T41
J14
T45
SP(250uL)
C30
Bath shielding LYSE(2.5mLX3)
T77
SH(10mL)
cover
T47
C68
T140
T155
SV08
T68
C5
C78
SV09
J31-T137-
J33-T138-
J35-T139-
T156
C64 T166
J32
J34
J36
T149
Transducer
T78
C42 C69 C24
SV13
C18
T79
T62
T168
C65
T82
T81
T146
C67 C19
T64
T83
T48
T69
T152
T153
DH
T66
SV07
T61
VC
T162 ASP(100uL)
C66 T65
T151 T58
T59 SV19
T54
SV01 SV03 SV02 TS
C75 C76 SV22
T154
T53
T163
T67
T55
J37-T52-
(T148)
T51
J38
T84
T108
T56
T96
C9
RBC C22 SV20 C73
WBC C20 SV21
T85
T92 T104 C33
C32 T102
T90 T106
T160
T94 C74
(T147)
T161
T95 T107
SV27 SV26 C53 C55 C59
C54
DIL(10mL)
T91 T103
C56 C57 C58
T127
LEO(I) reagent
C21 C23
LH reagent
T144
container
container
T87
container
reagent
T143
LEO(II)
T141
T142
J29-T101-
J27-T89-
(T100)
Bath shieding
(T88)
J30
C31 T86
cover SV12 cover SV11
LF1
C6
Probe T126 T125 C34
Isolating T97 Isolating T109 T111
T124
wipe
T98 chamber 3 T110
chamber 2 T112
T113
T99
C7
T114 C35
T123
C62 C61
C70
C72 C71
T157
T145
T159
SV23
T158
SV25
SV24
T130
T117
T122
C63
C10
T128 T120 T119 T118 T115
C39 C40
Waste LP2
C43 C11
containe T129 T121 T116
r LP3
6-47
Fluidic System
SV18
(T23)
J19-T24- J1-T11-
C15
J2
C45 C49 DILUENT
J20 T4
T25 C14 C46 C50 LEO( I( LYSE
T7 T3
LEO(I)
DIFF bath J3-T12 - C1 C47 C51 LEO( II( LYSE
T6
J39-T14 -
T2
J40-T15-
SV17 J4
LEO(II)
J17-T22-
(T21)
J11
J12
LF2 C2 C48 C52 LH LYSE
J18
C12 C13 T5 T1
J25-T26- C77 Flow LH
T164
T30 T31 T28 T29
T131
J26 cell
T27
Reagent
T136
T70
J21-T40- PV28
T10
T8
J7-T39-J8
T9
J22 T16 C37 T73 -T132
T44 T134
C26 J16-T17- T72 PC
T43 T133 T71
T33
T38 J41
T32
C25
J5-T18 -
T135C41 C16
SV14 SV16 GP T74
J6
J23-T42-
J9-T13-
J24
J15
J10
T36 C27 GF2
T34 T35
T37 SV1 C3
J42-T165-
C29 C28
0
J43
T75-T76
C4 J13-T20-
T19
T41
J14
T45
SP(250uL)
C30
Isolating LYSE(2.5mLX3)
T77
SH(10mL)
chamber 1
T47
C68
T140
T155
SV08
T68
C5
C78
SV09
J31-T137-
J33-T138-
J35-T139-
T156
C64 T166
J32
J34
J36
T149
Transducer
T78
C42 C69 C24
SV13
C18
T79
T62
T168
C65
T82
T81
T146
C67 C19
T64
T83
T48
T69
T152
T153
DH
T66
SV07
T61
VC
T162 ASP(100uL)
C66 T65
T151 T58
T59 SV19
T54
SV01 SV03 SV02 TS
C75 C76 SV22
T154
T53
T163
T67
T55
J37-T52-
(T148)
T51
J38
T84
T108
T56
T96
C9
RBC C22 SV20 C73
WBC C20 SV21
T85
T92 T104 C33
C32 T102
T90 T106
T160
T94 C74
(T147)
T161
T95 T107
SV27 SV26 C53 C55 C59
C54
DIL(10mL)
T91 T103
C56 C57 C58
T127
LEO(I) reagent
C21 C23
LH reagent
T144
container
container
T87
container
reagent
T143
LEO(II)
T141
T142
J29-T101-
J27-T89-
(T100)
Bath shielding
(T88)
J30
C31 T86
cover SV12 cover SV11
LF1
C6
Probe T126 T125 C34
Isolating T97 Isolating T109 T111
T124
wipe
T98 chamber 3 T110
chamber 2 T112
T113
T99
C7
T114 C35
T123
C62 C61
C70
C72 C71
T157
T145
T159
SV23
T158
SV25
SV24
T130
T117
T122
C63
C10
T128 T120 T119 T118 T115
C39 C40
Waste LP2
C43 C11
containe T129 T121 T116
r LP3
6-48
Fluidic System
50.7~52.8sThe SH syringe dispenses 1500uL of diluent into the DIFF bath through sample preparing tubes
51.1~523.1sSP syringe initialization
50.7~52.7sLYSE syringe initialization
53.65~56.5sClean back bath of the WBC bath
54~56sClean back bath of the RBC bath
56.2~57.1sTurn on and off V11 continuously to generate bubbles in RBC bath
6-49
Fluidic System
SV18
(T23)
J19-T24- J1-T11-
C15
J2
C45 C49 DILUENT
J20 T4
T25 C14 C46 C50 LEO( I( LYSE
T7 T3
LEO(I)
DIFF bath J3-T12 - C1 C47 C51 LEO( II( LYSE
T6
J39-T14 -
T2
J40-T15-
SV17 J4
LEO(II)
J17-T22-
(T21)
J11
J12
LF2 C2 C48 C52 LH LYSE
J18
C12 C13 T5 T1
J25-T26- C77 Flow LH
T164
T30 T31 T28 T29
T131
J26 cell
T27
Reagent
T136
T70
J21-T40- PV28
T10
T8
J7-T39-J8
T9
J22 T16 C37 T73 -T132
T44 T134
C26 J16-T17- T72 PC
T43 T133 T71
T33
T38 J41
T32
C25
J5-T18 -
T135C41 C16
SV14 SV16 GP T74
J6
J23-T42-
J9-T13-
J24
J15
J10
T36 C27 GF2
T35 SV1
T37
T34 C3
J42-T165-
C29 C28
0
J43
T75-T76
C4 J13-T20-
T19
T41
J14
T45
SP(250uL)
C30
Isolating LYSE(2.5mLX3)
T77
SH(10mL)
chamber 1
T47
C68
T140
T155
SV08
T68
C5
C78
SV09
J31-T137-
J33-T138-
J35-T139-
T156
C64 T166
J32
J34
J36
T149
Transducer
T78
C42 C69 C24
SV13
C18
T79
T62
T168
C65
T82
T81
T146
C67 C19
T64
T83
T48
T69
T152
T153
DH
T66
SV07
T61
VC
T162 ASP(100uL)
C66 T65
T151 T58
T59 SV19
T54
SV01 SV03 SV02 TS
C75 C76 SV22
T154
T53
T163
T67
T55
J37-T52-
(T148)
T51
J38
T84
T108
T56
T96
C9
RBC C22 SV20 C73
WBC C20 SV21
T85
T92 T104 C33
C32 T102
T90 T106
T160
T94 C74
(T147)
T161
T95 T107
SV27 SV26 C53 C55 C59
C54
DIL(10mL)
T91 T103
C56 C57 C58
T127
LEO(I) reagent
C21 C23
LH reagent
T144
container
container
T87
container
reagent
T143
LEO(II)
T141
T142
J29-T101-
J27-T89-
(T100)
Bath shielding
(T88)
J30
C31 T86
cover SV12 cover SV11
LF1
C6
Probe T126 T125 C34
Isolating T97 Isolating T109 T111
T124
wipe
T98 chamber 3 T110
chamber 2 T112
T113
T99
C7
T114 C35
T123
C62 C61
C70
C72 C71
T157
T145
T159
SV23
T158
SV25
SV24
T130
T117
T122
C63
C10
T128 T120 T119 T118 T115
C39 C40
Waste LP2
C43 C11
containe T129 T121 T116
r LP3
6-50
Fluidic System
55.8~58.2sOpen V26 and waste pump LP3 to drain the vacuum chamber
57.2~59.5sOpen V09 to release pressure in the pressure and vacuum chamber.
57.2~59.3sStart the pressure pump to help release vacuum of the vacuum chamber
SV18
(T23)
J19-T24- J1-T11-
C15
J2
C45 C49 DILUENT
J20 T4
T25 C14 C46 C50 LEO( I( LYSE
T7 T3
LEO(I)
DIFF bath J3-T12 - C1 C47 C51 LEO( II( LYSE
T6
J39-T14 -
T2
J40-T15-
SV17 J4
LEO(II)
J17-T22-
(T21)
J11
J12
LF2 C2 C48 C52 LH LYSE
J18
C12 C13 T5 T1
J25-T26- C77 Flow LH
T164
T30 T31 T28 T29
T131
J26 cell
T27
Reagent
T136
T70
J21-T40- PV28
T10
T8
J7-T39-J8
T9
J22 T16 C37 T73 -T132
T44 T134
C26 J16-T17- T72 PC
T43 T133 T71
T33
T38 J41
T32
C25
J5-T18 -
T135C41 C16
SV14 SV16 GP T74
J6
J23-T42-
J9-T13-
J24
J15
J10
T36 C27 GF2
T35 SV1
T37
T34 C3
J42-T165-
C29 C28
0
J43
T75-T76
C4 J13-T20-
T19
T41
J14
T45
SP(250uL)
C30
Isolating LYSE(2.5mLX3)
T77
SH(10mL)
chamber 1
T47
C68
T140
T155
SV08
T68
C5
C78
SV09
J31-T137-
J33-T138-
J35-T139-
T156
C64 T166
J32
J34
J36
T149
Transducer
T78
C42 C69 C24
SV13
C18
T79
T62
T168
C65
T82
T81
T146
C67 C19
T64
T83
T48
T69
T152
T153
DH
T66
SV07
T61
VC
T162 ASP(100uL)
C66 T65
T151 T58
T59 SV19
T54
SV01 SV03 SV02 TS
C75 C76 SV22
T154
T53
T163
T67
T55
J37-T52-
(T148)
T51
J38
T84
T108
T56
T96
C9
RBC C22 SV20 C73
WBC C20 SV21
T85
T92 T104 C33
C32 T102
T90 T106
T160
T94 C74
(T147)
T161
T95 T107
SV27 SV26 C53 C55 C59
C54
DIL(10mL)
T91 T103
C56 C57 C58
T127
LEO(I) reagent
C21 C23
LH reagent
T144
container
container
T87
container
reagent
T143
LEO(II)
T141
T142
J29-T101-
J27-T89-
(T100)
Bath shielding
(T88)
J30
C31 T86
cover SV12 cover SV11
LF1
C6
Probe T126 T125 C34
Isolating T97 Isolating T109 T111
T124
wipe
T98 chamber 3 T110
chamber 2 T112
T113
T99
C7
T114 C35
T123
C62 C61
C70
C72 C71
T157
T145
T159
SV23
T158
SV25
SV24
T130
T117
T122
C63
C10
T128 T120 T119 T118 T115
C39 C40
Waste LP2
C43 C11
containe T129 T121 T116
r LP3
6-51
Fluidic System
6-52
Fluidic System
6-53
7Optical System
7.1 Overview of Optical System Principle
The optical system principle is based on that of the flow cytometer: when the stained cells pass
the facula detection area under the influence of "Hydrodynamic focusing", light scatter is
produced after exposure of the a cell under light beam, where the forward scatter reflects the
volume of the cell, and the side scatter reflects the complexity of the nucleus. Therefore,
BC-5380 or BC-5180 is able to differentiate different cells by collecting and analyzing the
forward scatter and side scatter, and presents the result using 2D scattergram. The optical
system is composed of the laser radiation module and scatter sensors.
13
12
11
1 2 3 4 5 6 7 8 9 10
The optical system uses semiconductor laser (1) as the source of light, which emits a red laser
beam of 670nm. The laser beam passes through the collimating lens and gets collimated, and
then adjusted by cylindrical lens A (3) and B (4) which are placed orthogonally, to form a flat
elliptic laser beam of 15um*300um (@13.5%) going through the sample flow (6) in the center
of the flow cell (5).
7-1
Optical System
7-2
Optical System
7-3
Optical System
Select the venous whole blood mode (open-vial) for sample analysis, and run the 7um
standard particle. After the analysis is completed, select the analysis data in the "Review"
screen, and then click the "Optical" button on top right to go to the optical adjustment screen,
as shown in Figure 7-5.
7-4
Optical System
Figure 7-6 Brightened dot on the exterior wall of the flow cell
Troubleshooting: if evident brightened dot is found, wipe the place with dust-free cloth
dipped with anhydrous alcohol until the dot disappears.
Note:
1. Avoid direct eye contact with the laser beam in the maintenance process;
2. Do not touch the radiation-proof diaphragm or damage the flow cell assembly while
wiping;
3. Check if there is any evident brightened dot on the exterior wall of the flow cell after
wiping.
7-5
Optical System
Figure 7-7 Scattergram of normal blood sample when the flow cell interior wall needs
cleaning
Figure 7-8 Scattergram of standard particle when the flow cell interior wall needs
cleaning
Error confirmation: open the top cover and right door using a cross-headed screwdriver,
loosen the 6 screws fixing the optical system shielding cover to remove the cover. Check
if there is brightened dot on the interior wall of the flow cell. At the meantime, place a white
paper before the side scatter PD assembly, and check if there is a ring in the focal facula,
shown as follows:
Figure 7-9 Brightened dot on the interior wall of the flow cell
7-6
Optical System
Figure 7-10 The optical ring before the side scatter PD assembly
Troubleshooting: click Menu>Service>Maintenance, and then select "Probe Cleanser
Soak" to perform automatic soaking and maintenance to the flow cell. If the optical ring in
the focal facula before the side scatter PD assembly still exists, manual probe cleanser is
needed to be performed as instructed below:
1. Take out 2 clean S-50 tubes or 3350 silicone tubes (length: about 10cm). Connect
one tube to the newly unpacked syringe, and the other one to the waste outlet
connector of the flow cell. Snip the plastic cable tie using a pair of diagonal pliers, and
then remove the flow cell tray using a screwdriver. Unplug the sheath fluid inlet tube,
and plug the tube which is connected to the syringe;
2. Draw the diluent from the flow cell assembly with the syringe, and then put the other
end of the tube connecting to the waste outlet into a small beaker. Inject a certain
volume of probe cleanser with the syringe, and make sure it is visible that the probe
cleanser is flowing into the flow cell.
7-7
Optical System
Note:
1. Generally speaking, if the scattergram distribution is similar to Figure 7 or it is judged
that the flow cell interior wall needs cleaning, perform automatic maintenance of the
flow cell first, and do not remove the cover for check until the automatic maintenance
fails or does not take effect;
2. Avoid direct eye contact with the laser beam; wear proper personal protective
equipment during manual cleaning, and wear the PVC gloves properly; do not drop
the probe cleanser on the instrument or on the clothes/skin of any people.
Figure 7-12 Removal of the top cover and the right door
3) Snip the plastic cable tie fixing the flow cell tray with a pair of diagonal pliers, and then
unscrew the 2 M3x8 cross-recessed panhead screws (with washers) fixing the flow cell
tray with a cross-headed screwdriver to remove the tray;
4) Wear a pair of disposal PVC gloves, and then remove the T connector of the commutation
assembly and the tubing of SV18 valve in turn (marked by the arrows in the figure below).
If there is any fluid leakage, wipe it with tissue or wet cloth immediately to avoid corrosion
7-8
Optical System
to the instrument;
Figure 7-13 Removing the flow cell tray and tubing of the T connector
5) Get a clean S-50 tube or 3350 silicone tube of about 10cm, and connect one end to the
sheath fluid inlet, and the other end to the sample inlet. Connect the tube that is removed
from the SV18 valve to the sample outlet (marked by red arrows in the figure);
6) Disconnect the optical system signal transmission cable and control wire from the data
board;
7) Unscrew the 6 inner hexagon screws fixing the optical system shielding cover with a
screwdriver, and then remove the 4 M4x8 cross-recessed panhead screws (with washers)
from the base plate;
8) Remove the whole optical system from the instrument, and install the shielding cover
back to the removed system, and put them into the packing box;
9) Install the new optical system based on the reversed procedure above.
7-9
Optical System
7-10
8 Hardware System
8.1 Overview
This chapter introduces functions of the hardware system and its interfaces. The hardware
system includes data flow, user interface, control flow, power system, structure and
connections by logic.
Hardware System
Optical Product
detection application
principle interlink
Counting bath
Impedance Signal sorting
Laser tube
detection Signal collection Direct user Network cable
HGB sensor
principle Digital processing input
Photocell
Main control Key switch
Colorimetry
detection embedded
system Direct user Indicator
principle
Data stream design platform output Buzzer
Barcode
External scanner
interfaces
Motor Structure
Moving
Electromagne
mechanism and interlink User interaction design
t
drive and design
Valve detection
Pump Fluid
Float switch component
drive and
Heater detection
Fan Thermal Auxiliary
component control Power
Temperature drive and platform monitor
sensor detection
Pressure Power switch
sensor System status Grid power
monitor
Power Grid power
Barcode input
Fluid conversion
scanner
detection
Photocouplers Control stream design Power system design
J15 P24V
sensor and J8 preamplification
Diluent/waste board
wires
board
sensor- +- LaserHGB
control board
float&switch
Sample J9
PowerA12V Analog assembly
collection board WBC bath
assembly/relieve J10 AC120 board
valve V RBC bath
Syringe
photocoupler Pressure sensor
J11
photocouple J6 D5V/
Door
r P12V
Fan P12V/2 electromagnet
J5-12V Mix motor
4V(
photocoupler
Diff/laser J14-
D5V
Autoloading Autoloading
heater
Sample
P24V board photocoupler
2 autoloading
J16-J19 motors
collection Mix motor
assembly/syring Up/down/forwar
e motor d/backward
motors
Figure 8-2
8-1
Hardware System
8.3.2 Function
The analog board can be divided into 7 modules: analog power adjusting module, voltage
monitoring module, HGB channel module, impedance channel module, optical channel
module, drive buffering module, and pressure monitoring module.
Analog power adjusting module:
This module filters the A±12V voltage and transforms it into the 5V/-5V analog voltage
required by the analog section; meanwhile it doubling rectifies the A+12V voltage, and
stabilizes it into 56V direct current.
Voltage monitoring module:
This module converts level of the voltages to be monitored on the board to the level that
can be received by AD (0~5V), and makes the impedance fulfill requirement of AD
conversion.
The monitored signals include: WBC/RBC aperture voltage, +/-12V, 56V zapping power
and FS direct current background.
HGB channel module:
8-2
Hardware System
This module drives HGB sensor, amplifies and adjusts HGB signals.
Impedance channel module:
This module drives the RBC/PLT and WBC sensors, amplifies and adjusts RBC/PLT and
WBC impedance signals. Besides, it provides counting bath zapping function.
Optical channel module:
This module amplifies and adjusts FS/SS and SF signals input from the preamplification
board.
Pressure monitoring module:
This module drives the pressure sensor, coverts pressure/vacuum change into voltage
change, and amplifies the voltage.
Drive buffering module:
This module converts TTL control signals input from external boards into actuating signals
with certain driving capability.
8.3.3 Structure
The PCBA structure of the analog board is as follows.
8-3
Hardware System
P3
P6
P4
P5 P7
P2
P1
8-4
Hardware System
8.3.4 Interfaces
Interface layout and description
J2 J4 J3 J6
J5
4
J
HGB 8
drive
and Optical signal
Power adjusting amplifi amplification
module cation
J
9
Control
Monitor signal adjusting signal drive
circuit circuit
Shielding
region
J2
8-5
Hardware System
Indicator Function
D47 Indicator of analog +12V, normal status: on
D48 Indicator of analog -12V, normal status: on
D49 Indicator of analog +5V, normal status: on
D50 Indicator of analog -5V, normal status: on
The functions of main test points in the board are listed in the following table.
Table 8-4 Test points in the analog board
Test
Printed label Function Note
point
TP16 2.5V Reference voltage in the Normal voltage range:
pressure monitoring circuit 2.4V~2.6V
TP17 2.5V Reference voltage in the Normal voltage range:
pressure monitoring circuit 2.4V~2.6V
TP18 1.25V Reference voltage in the Normal voltage range:
pressure monitoring circuit 1.2V~1.3V
TP19 2.5V Reference voltage in the Normal voltage range:
pressure monitoring circuit 2.4V~2.6V
TP22 12V Test point of 12V power Normal voltage range:
11.4V~12.6V
TP26 N12V Test point of -12V power Normal voltage range:
-11.4V~-12.6V
TP23 5V Test point of 5V power Normal voltage range:
4.75V~5.25V
TP27 N5V Test point of -5V power Normal voltage range:
-4.75V~-5.25V
TP1 56V Voltage test point of impedance Normal voltage range:
channel constant-current source 47V~60V
TP48 VCONST_MON 56V monitoring voltage Normal voltage range:
1.8V~2.2V
TP44 12VM 12V monitoring voltage Normal voltage range:
2.2V~2.6V
TP47 N12VM -12V monitoring voltage Normal voltage range:
1.2V~1.5V
TP37 HGB Output signal of HGB channel
TP2 RBC Output signal of RBC channel
TP9 WBC Output signal of WBC channel
Vacuum monitoring output
TP20 NP
signal
Pressure monitoring output
TP21 PP
signal
TP49 FS Output signal of FS channel
TP61 SS Output signal of SS channel
8.3.6 Troubleshooting
Table 5 lists the frequent errors and solutions of the analog board to guide the troubleshooting
actions. The error here refers to hardware error only, the same error situation caused by the
fluidic, optical, reagent or software systems (for which you can refer to the related chapters for
solution) are excluded.
Before troubleshooting analog board errors, the following checks shall be performed.
1. Check if the power output is normal;
8-6
Hardware System
2. Check if the wires are firmly connected to the analog board; if the wire No. and
socket No. match; and if the wires are damaged.
3. Check if the power input of the board is normal; and check if the indicators are
normal according to Table 3.
4. Restart the analyzer to see if the error is removed.
If all the above actions are done, and the error still exists, you may go on to troubleshoot the
error as instructed in the following table.
WARNING
Be sure to wear antistatic gloves when assembling and disassembling the
board, and always hold on the edge of board.
When using multimeter or other equipments to test a component, make sure
not to touch other components or wires to avoid board damage caused by
short circuit.
Error
No. Cause Error Diagnosis
Consequence
The following situations can all be
diagnosed as analog error:
"+12V analog
D47 indicator turns off
voltage
1 Circuit error The voltage of test point TP22 (12V) is
abnormal" is
outside 11.4~12.6V
reported
The voltage of test point TP44 (12VM)
is outside 2.2V~2.6V
The following situations can all be
diagnosed as analog error:
"-12V analog
D48 indicator turns off
voltage
2 Circuit error The voltage of test point TP26 (N12V)
abnormal" is
is outside -11.4~-12.6V
reported
The voltage of test point TP47
(N12VM) is outside 1.2V~1.5V
The following situations can all be
diagnosed as luminotron error:
The luminotron is off, the level of
TP72(HGBLED_N) is low (about 0V)
The luminotron is on, the HGB voltage
can be adjustable along with the gain, but
HGB luminotron
3 when the gain is adjusted to its upper limit,
error
the HGB voltage is still lower than 3.2V.
The situation suggests the aging of the
"HGB Blank luminotron (when fluidic problems are
Voltage excluded)
abnormal" is The error is removed after replacing
reported the HGB assembly
The following situations can all be
diagnosed as analog board drive circuit
error:
The luminotron is off, and the
HGB luminotron
4 TP31(HGB_ON_N) is of high level (3.3V)
drive circuit error
when running analysis sequence
After replacing the HGB assembly, the
luminotron is still off
The HGB luminotron is off even after it
8-7
Hardware System
Error
No. Cause Error Diagnosis
Consequence
has been replaced
HGB
photoelectric cell
HGB voltage is 0; the voltage of test point
error (usually the
TP38 (HGB1) is positive and greater than
5 anode and
0.6V. The error is removed after replacing
cathode are
HGB assembly.
incorrectly
connected)
The following situations can all be
diagnosed as circuit error:
HGB voltage is higher than 4.8V when the
gain is at its minimum
The voltage of test point TP37 (HGB) is
3.2~4.8V, but the HGB voltage displayed
6 HGB circuit error on the screen is not within the range
HGB voltage is lower than 3.2V when the
gain is at its maximum, and replacing the
HGB bath fails to solve the problem
D49 (+5V) indicator is off
The voltage of test point
TP40(VCONST_HGB) is outside 2.4~2.9V
Error of the digital
HGB gain The gain changes from 50 to 200, but the
7 potentiometer on
calibration fails HGB tested is unchanged
the board
The following situations can all be
diagnosed as analog error:
The voltage of 56V constant current
source is outside 47~60V, or the voltage of
TP48 (VCONST_MON) is outside
1.8V~2.2V
Constant current
The voltage difference between pin 2
"Aperture source
and pin 3 of J1 is 12~18V when running
voltage error/monitoring
8 RBC analysis, but error is reported by the
abnormal" is circuit
analyzer. This situation suggests
reported error//power
monitoring circuit error.
source error
D49 (+5V) indicator is off
Replace the counting bath with 20K
resistor, if the voltage tested by the two
ends of the resistor is outside 11.4~13.2V,
that means error occurs to the constant
current source.
The following situations can all be
"Forward diagnosed as circuit error:
scatter blank The scattergram is normal while "Forward
9 voltage Circuit error scatter blank voltage abnormal" is reported
abnormal" is D49 (+5V) indicator is off
reported The voltage of test point TP49(FS) is not
1.5~1.9V when analysis is not running
Confirm as per the following procedure:
1) Disconnect all wires connected to the
analog board and start up the analyzer, if
the error is removed, then short circuit of
Auto power-off Short circuit of
10 analog board power source can be
at startup power source
excluded; check other wires and
components connected to the analog
board.
2) Disconnect all wires connected to the
8-8
Hardware System
Error
No. Cause Error Diagnosis
Consequence
analog board and start up the analyzer, if
the error persists, then short circuit of
analog board power source can be
concluded, and the analog board should be
replaced.
8.4.2 Function
Basic functions of the digital control circuit are:
1. Data processing: receiving data collected by the analog circuit, providing operation
system and software running platform for data processing,
2. System control: providing operation system and software running platform for
scheduling and monitoring of the analyzer operations.
3. User interacting: providing hardware platform for user interaction.
8-9
Hardware System
8.4.3 Structure
Optical system
RS232
RJ45 DB9 serial
J85 DC/DC circuit
-J port
J65
EEPROM
6cm circuit RTC
circuit
fpga-LCD display LCD back Touch
cpu-LCD display socket screen
socket light socket port
8.4.4 Interfaces
Table 8-6 Hardware interfaces of the Pinaster main control board
8-10
Hardware System
Test point
Board Position No. Function Description
CPU module TP1 Test point of 1.8V power Normal voltage range: 1.8V~±5%
CPU module TP2 Test point of 3.3V power Normal voltage range: 3.3V~±5%
CPU module TP3 Test point of 1.3V power Normal voltage range: 1.3V~±5%
Bottom plate TP1 Test point of 1.8V power Normal voltage range: 1.8V~±5%
Bottom plate TP2 Test point of 3.3V power Normal voltage range: 3.3V~±5%
Bottom plate TP3 Test point of 1.2V power Normal voltage range: 1.2V~±5%
Bottom plate TP36 Test point of 5V power Normal voltage range: 5V~±5%
Bottom plate TP39-43 GND Normal voltage range: 0V~±5%
8.4.6 Troubleshooting
Table 8-7 Errors and troubleshooting list of the digital control board
Error
No. Cause Error Diagnosis
Consequence
Confirm as per the following procedure:
1) Disconnect all wires connected to the digital control
board and start up the analyzer, if the error is removed,
then short circuit of digital control board power source
Short
can be excluded; check other wires and components
Auto power-off circuit of
1 connected to the digital control board.
at startup power
2) Disconnect all wires connected to the digital control
source
board and start up the analyzer, if the error persists,
then short circuit of digital control board power source
can be concluded, and the digital control board should
be replaced.
The network Confirm as per the following procedure:
cannot be 1.Replace the PC to see if the error is removed
Hardware
2 connected 2.Observe indicator D6 to see if it is off
error
with upper 3.Check if the network cable is properly connected
device 4.Replace the board to see if the error is removed
Confirm as per the following procedure:
1. Observe the FPGA indicator D2, if it does not flicker,
Fail to collect FPGA
3 then the FPGA is not working normally.
analog data error
2. Check the connection wire to the analog board.
3. Check if the analog board is OK.
8-11
Hardware System
the information to the upper modules as necessary. Its functional diagram is as follows:
Upper
control
system
Command and
data transmission
Thermal
Moving parts
parts drive
drive and Step motor
and
detection
detection
Heater
Valve
Pump
Control Sensor
platform (pressure, temperature,
etc)
Photocoupler
Float
Switch
System
Fluid parts
status Others
drive
monitor
8.5.2 Function
The functions of power drive board are:
1. Main control module: this module is consisted of the ARM+FPGA structure, it analyzers
upper layer commands, drives the execution parts, collects and sends reference data of the
sensors to the digital control board, so that it may control the power of the analyzer effectively.
2. Driving motor: it drives the sample collection mechanism and the syringe assembly via the
drive motor.
3. Moving mechanism detection (photocoupler detection): detects mechanism position via
sensors.
4. Driving valves and pumps: it drives valves and pumps so that samples or reagents may flow
in the fluidic system.
5. Driving heater: it forms a closed-loop control system with the temperature sensor to provide
proper temperature for operation of the components or reaction of the reagents.
6. Driving the fan: it drives the fan to eliminate heat.
7. Temperature detection: it detects the temperature of components, reagents and the
environment via temperature sensors.
8. Pressure detection: it detects output pressure of the pressure sensor to control pressure
generation of the gas system and monitor the pressure.
9. Fluid detection: it detects the electrical level signal of the fluid detection board to monitor if
there is any reagent.
10. Float switch detection: it detects the on/off status of the float switch to monitor the
full/empty status of waste container and diluent container.
11. Voltage detection: monitors the voltage of the board itself.
12. Parameter storage: it stores all types of calibration parameters so that necessary
information can be saved when power-fail occurs.
8-12
Hardware System
13. Power module :it realizes the conversion from 12V to 5V or 3.3V.
8.5.3 Structure
See Table 8-8 and Figure 8-9 for the modules of power drive board.
Table 8-8 Modules of power drive board
No. Module
P1 Motor drive module
8-13
Hardware System
8.5.4 Interfaces
Table 8-9 Interfaces of power drive board
Position
Interface
No.
J2
Valve drive interface
J3
J17
Motor drive interface
J18
J19
8-14
Hardware System
P24V_LED D98 /
VDD_LED(3.3V) D99 /
HT1 D81 /
When the heating
Heating drive drive is working, the
HT2 D82 /
indicator LED is on; and vice
versa.
HT3 D83 /
When FPGA is
running normally,
FPGA indicator FPGA_LED D94 / the indicator
flickers; and vice
versa.
8-15
Hardware System
Test point
power drive board has a complicated circuit, so there are many test points. Here we only list
some critical test points.
Table 8-11 Critical test points of power drive board
Test
Printed label Function Note
point
8.5.6 Troubleshooting
This chapter introduces the troubleshooting process of the drive board based on its modules
and structure. The following preparations must be done before troubleshooting.
Check if the power supply of the analyzer is OK.
Check if the peripheral parts (motors, valves, and pumps), sensors (photocouplers,
temperature sensors), wires and other accessories related to the drive board are OK.
Check if the wires are firmly connected to the ports.
Restart the analyzer to see if the error is removed.
If the cause of the error still cannot be located, continue with error troubleshooting as
instructed in the following table.
8-16
Hardware System
Error
No. Cause Error Diagnosis Note
Type
Before
troubleshooting
The following situations can be errors of other
concluded as power drive board errors: power modules
12V
Circuit 1. Indicator D97 turns off; that are related
1 power
error 2. Voltage of TP7 is outside 11.4V-12.6V; to the drive
error
3. Electrical components connected to board, be sure
12V power module are burnt out. to check if the
power supply is
OK.
The
troubleshooting
method above
The following situations can be
can be used for
Error of concluded as drive board errors:
modules with
the 1. When executing motor running
indicators, like
sample command, indicators D51 and D61 turn
the main control
collection Circuit off.
2 module, heating
assembly error 2. Components of the drive circuit are
module,
X burnt out.
photocoupler
direction 3. The components are not burnt out, but
detection
motor after replacing motor and wire, the motor
module, motor
still fails to run as expected.
module and fluid
detection
module.
The
troubleshooting
method above
can be used for
The following situations can be modules without
concluded as drive board errors: indicators, like
1. Components of the drive circuit are the float switch
Pump Circuit
3 burnt out. detection
errors error
2. The components are not burnt out, but module, fan
after replacing motor and wire, the motor drive module,
still fails to run as expected. temperature
detection
module and
pump drive
module.
8.6.2 Function
Driving step motors
The autoloading board drives the following 5 step motors:
Mix mechanism X motor
Mix mechanism Z motor
8-17
Hardware System
Mix motor
X loading motor
Y feeding motor
The step motor drive is bipolar constant current half-step drive, the position of each step
motor is detected by the corresponding position sensor.
Electromagnet control
The electromagnet controls the opening of the sample compartment (manual closing).
Built-in barcode scanner control
The built-in barcode scanner reads tube barcodes and tube rack barcodes. The
autoloading board controls the built-in barcode scanner via MCU. The MCU
communicates with the built-in barcode scanner (UART) via serial port 1.
Position sensor and micro-switch status detection
There are 13 position sensors and 1 micro-switch in the autoloading board. Status
detection of position sensors and micro-switch is done by FPGA, the MCU inquires about
the storage status of the FPGA when necessary.
Parameter storage
Due to the production and installation deviation of the mix mechanism, initial position of
the X motor and mix motor of the mix mechanism need to be adjusted, the position
parameters after the adjustment are stored in the autoloading board.
8-18
Hardware System
8.6.3 Structure
8.6.4 Interfaces
Table 8-13 List of autoloading board interfaces
8-19
Hardware System
Indicator Description
D3 12V power indicator, on when the 12V power is normal
D4 5V power indicator, on when the 5V power is normal
D5 3.3V power indicator, on when the 3.3V power is normal
D25 MCU working indicator, flickers every 0.5s when MCU works as normal
D26 FPGA working indicator, flickers every 0.5s when FPGA works as normal
Table 8-15 List of autoloading board test points
8.6.6 Troubleshooting
Error possibility of the autoloading board is low, and the errors can be troubleshooted with
reference to troubleshooting methods of the power drive board. The frequent errors during
autoloading process are introduced as follows.
8-20
Hardware System
8-21
Hardware System
8.7.2 Function
The functions of the laser control board are: power adjusting, laser drive current monitoring
and laser power control.
Power adjusting: this module filters the ±12V power from the analog board, the ripple after the
adjustment is lower than 100mV.
Laser drive current monitoring: measures the working current of laser, and sends the result to
analog board for monitoring.
Laser power control: laser control board controls the laser using constant power control
method, the photoelectric detector inside the laser monitors its output power real-time, and the
monitoring result forms a closed-loop system via negative feedback to realize constant power
output. The power is controlled by adjusting the potentiometer VR1; it shall be within the range
3mW~5 mW.
Control
Laser constant power unit Laser
signal
(closed loop control unit) (HL6714G)
8.7.3 Structure
The PCBA structure of the laser control board is as follows.
Table 8-17 Modules of the laser control board
8-22
Hardware System
P3 P2
P1
Figure 8-12 The PCBA structure of the laser control board (top)
8.7.4 Interfaces
The laser control board has two external interfaces, J1 to analog board and J2 to the laser.
See the following figure.
8-23
Hardware System
8-24
Hardware System
8.7.6 Troubleshooting
See the following table to troubleshoot errors of the laser control board:
8-25
Hardware System
8.8.2 Function
The functions of the preamplification board are: power adjustment, photoelectric conversion
and signal adjustment.
Power adjusting: this module filters the ±12V power from the analog board, the ripple after the
adjustment is lower than 50mV.
Photoelectric conversion: the photoelectric conversion unit converts optical signal to current
signal using photodiode.
Signal adjusting: converts current signal to voltage signal via I/V, and then sends the signal to
the amplification and adjustment unit to process it into signal that meets the input requirement
of the analog board.
Data
Signal adjusting unit
board
Photoelectric Power
FS( SS adjusting
conversion
(π filtering)
FS/SS
A±12V
preamplification
board
8.8.3 Structure
The PCBA structure of the preamplification board is as follows.
Table 8-22 Modules of the preamplification board
8-26
Hardware System
P1
P2
P3
(a) (b)
Figure 8-15 PCBA structure of the preamplification board (a) top (b) bottom
8.8.4 Interfaces
The FS/SS preamplification board has 1 external interface, which is J1 to the motherboard.
The layout of the board is as follows:
8-27
Hardware System
8.8.6 Troubleshooting
See the following table to troubleshoot errors of the preamplification board:
Table 8-25 Troubleshooting errors of the preamplification board
Error
No. Criterion Cause Troubleshooting
Name
The connection with the
Reconnect or replace
analog board is loose, or the
The voltage of the wire.
wire is damaged.
test point +12V
Check if the power
is outside the The 12V power sent by the
12V supplied by the analog
range analog board is incorrect.
1 power board is correct.
12V±0.6V, or the
error Dry joint or damage of the
ripple noise is
components related to test
higher than Reweld or replace the
point +12V (such as inductor
50mV. components.
L1, capacitor C24, C27, C20
and C23).
The connection with the
Reconnect or replace
analog board is loose, or the
The voltage of the wire.
wire is damaged.
test point -12V is
Check if the power
outside the The -12V power sent by the
-12V supplied by the analog
range analog board is incorrect.
2 power board is correct.
-12V±0.6V, or
error Dry joint or damage of the
the ripple noise
components related to test
is higher than Reweld or replace the
point -12V (such as inductor
50mV. components.
L2, capacitor C22, C26, C25
and C21).
The FS board is installed to the
SS output signal position of the SS board by Replace the boards.
too weak mistake.
(assuming it is
Output caused by Incorrect resistance, dry joint Reweld or replace the
3
error circuit problem) or damage of R4, R5 or R6. components.
FS output signal The SS board is installed to
too strong the position of the FS board by Replace the boards.
(assuming it is mistake.
8-28
Hardware System
Error
No. Criterion Cause Troubleshooting
Name
caused by
Incorrect resistance, dry joint Reweld or replace the
circuit problem)
or damage of R4, R5 or R6. components.
No output signal
of the Dry joint or damage of Reweld or replace the
preamplification photodiode D1 or R1. components.
board (FF/SS),
or the signal is Dry joint or damage of U2 or Reweld or replace the
too weak. U3. components.
Check the related resistors
Bandwidth error
and capacitors, such as R4,
of the Reweld or replace the
R5, R6, C1, C5 and C4, to see
preamplification components.
if there is any dry joint or
board (FS/SS)
damage.
Linear
390V circuit Flyback Forward
protectiv protectiv
VDD e circuit e circuit
A+12V
A-12V
Flyback circuit
AC130V
P12V
Forward circuit
P24V
8.9.2 Function
The power board works under the voltage of (50~60Hz) input by AC 90V~264V.
Once the power switch is on, all circuits start to work, the D5V, A+12V, A-12V, AC130V, P12V
and P24V all have voltage output.
8-29
Hardware System
Voltage Load
Minimum Rated Output voltage
Voltage adjustment adjustment Sound
current current range
rate rate
D5V 2A 5A 4.85/5.25V ±1% ±5% 100mV
+A12V 0mA 1A 11. 5/12. 5V ±1% ±5% 100mV
-A12V 0mA 650mA -11.5/-12.5V ±1% ±5% 100mV
P12V 0.3A 5.5A 11.5/12.5V ±1% ±5% 150mV
P24V 0A 4A 22/27V ±1% ±10% 150mV
AC120V 0mA 60mA 115/145V(RMS) ±1% ±10% /
The highest current generated by the P24V power is 7.8A, whose duration is 2S, cycle is 60S.
8-30
Hardware System
8.9.3 Structure
8.9.4 Interfaces
The power board has 7 external interfaces, 2 of them are sockets (J1 and J2), and the others
are connected to external components with wires (the PCB numbers are TP1~TP20). See the
following figure for position of the interfaces.
8-31
Hardware System
Welding
Interfaces Function Pin number Description
disk number
J1 AC input 3 / /
J2 Fan power output 6 / /
Analog section Output via wire
A-J12 power output 6 4 welded to PCB
Output via wire
A-J13 AC output part 3 2 welded to PCB
Power section power Output via wire
A-J35 output 10 10 welded to PCB
Digital section power Output via wire
A-J37 output 4 4 welded to PCB
8-32
Hardware System
Definition of J1
Table 8-28 Input AC socket
Pin Definition
1 L, connect to the live wire of the grid power.
2 PG, connect to ground wire
3 N, connect to the null wire of the grid power.
Definition of J2:
Table 8-29 Fan socket
Pin Definition
1/3/5 P12V, connect to positive end of P12V output voltage
2/4/6 PGND, connect to negative end of P12V output voltage
8-33
Hardware System
8-34
Hardware System
8.9.6 Troubleshooting
The errors of power board can be troubleshooted as per the following procedure.
Power error
A+12V/A-12V/AC130 abnormal
Are the A+12V/A- N indicates flyback converter error(
12V/AC130V( P12V/P24V
P12V/P24V abnormal indicates
normal?
forward converter error.
Other errors
8-35
Hardware System
Figure 8-21 The connections of fluid detection board and other boards
8.10.2 Function
The functional diagram of fluid detection board is as follows:
Constant
current drive luminotron
Photoelectri Retarding OUT1
c receiver comparator
circuit
Constant
Switch of the current drive luminotron Photoelectri Retarding OUT2
Power constant circuit c receiver comparator
Constant
current drive luminotron Photoelectri Retarding OUT4
c receiver comparator
circuit
8.10.3 Structure
The fluid detection board PCB:
Top
Bottom
Figure 8-23 Fluid detection board
8-36
Hardware System
8.10.4 Interfaces
Table 8-35 Interfaces of the fluid detection board to the motherboard
8.10.6 Troubleshooting
When error occurs to the fluid detection board, the analyzer will report "insufficient reagent".
Troubleshoot the fluid detection board error as per the following procedure:
8-37
Hardware System
YES
YES
NO
Is the voltage between
Other errors Fluid detection
J1.9 and J1.10 about 0V? Is the on/off status of D1~D4(green) in the NO board error
fluid detection board consistent with the
fluid status detected by the photocouplers?
YES
YES
Other errors
8.11.2 Function
The indicator board has 2 basic functions:
Status indicating: the indicating system of the indicator board consists of 3 colors, which are
red, yellow and green, indicating different status of the analyzer.
Sound alarming: the buzzer in the indicator board provides sound alarm function when error
occurs.
8-38
Hardware System
8.11.3 Structure
Top
Bottom
Figure 8-25 The indicator board
8.11.4 Interfaces
The indicator board has 1 interface, J1. The definition of J1 is listed in the following table.
Table 8-37 Definition of J1
Test Printed
Function Note
point label
Normal voltage range
TP1 VCC Test point of 5V power
4.75V-5.25V
Normal voltage range
TP2 GND Ground test point
4.75V-5.25V
Test point of green
TP3 / 3.3V TTL level
indicator control signal
8-39
Hardware System
Test Printed
Function Note
point label
Test point of yellow
TP4 / 3.3V TTL level
indicator control signal
Test point of red indicator
TP5 / 3.3V TTL level
control signal
When the indicator is on, the
Test point of green
TP6 / electric level is low; when it is
indicator working status
off, the electric level is high (5V).
When the indicator is on, the
Test point of yellow
TP7 / electric level is low; when it is
indicator working status
off, the electric level is high (5V).
When the indicator is on, the
Test point of red indicator
TP8 / electric level is low; when it is
working status
off, the electric level is high (5V).
8.11.6 Troubleshooting
There are 2 types of frequent indicator board error:
The indicator fails to work as intended
Possible cause: the wire is not firmly connected to J1, or the indicator is damaged.
Solution: reconnect or replace the wire, or replace the LED indicator.
The buzzer fails to work as intended
Possible cause: the wire is not firmly connected to J1, or the buzzer is damaged.
Solution: reconnect or replace the wire, or replace the buzzer.
8.12.2 Function
The key board has 2 basic functions:
Human machine interface by the [ASPIRATE] key: users press the [ASPIRATE] key to start
autoloading analysis.
Human machine interface by the [OPEN] key: users press the [OPEN] key to open the sample
compartment door.
8-40
Hardware System
8.12.3 Structure
8.12.4 Interfaces
The key board has only 1 external interface, J1.
8.12.6 Troubleshooting
Frequent error: the keys do not respond to pressing.
Possible cause: the wire is not firmly connected to J1, or the key is damaged.
Solution: reconnect the wire; if the error still exist, replace the key board.
8.13.2 Function
The mini network board provides direct connection to the 2 RJ45 connector in and outside
the analyzer. See the following figure:
8-41
Hardware System
8.13.3 Structure
The PCB of the mini network board (top and bottom) is as follows:
8.13.4 Interfaces
The pin definitions of J1 and J2 interfaces are the same, see the following table:
Table 8-40 Description of mini network board interfaces and pins
8-42
Hardware System
8.13.6 Troubleshooting
Table 8-41 Troubleshooting the mini network board
Error
Cause of error Troubleshooting method
phenomenon
Step 1: Check if the network cable connection is
loose. If yes, reconnect the cable and see if the
1. The network
problem is solved; if no, go to step 2.
cable is not
Step 2: Take out the mini network board, and
The PC cannot properly
test the conducting state of the pins of J1 and
communicate connected;
J2. See the figure above for pin definition of J1
with BC-5380 or 2. The cable is
and J2. If pin 1-8 of J1 are conducted with pin
BC-5180 damaged or poor
1-8 of J2, the mini network board is normal, the
contact of the
error may be caused by some other part of the
interface
analyzer; otherwise the mini network is
damaged and must be replaced.
YES
Reconnect the
Does the cable connected to mini
cable network board get loose?
NO
YES
Can pin 1~8 of J1 be
Other errors
conducted with pin
1~8 of J2
NO
Mini network
board error
Socket
No. Board No. Board Name Unit Labeling
Prefix
1 051-001146-00 Power board PCBA 1 A A-J1…
Pinaster main control
2 051-000985-00 1 B B-J1...
board PCBA
3 051-000982-00 Analog board PCBA 1 C C-J1…
8-43
Hardware System
Socket
No. Board No. Board Name Unit Labeling
Prefix
4 051-000981-00 Drive board PCBA 1 D D-J1…
Autoloading Board
5 051-000393-00 1 E E-J1…
PCBA
6 051-001062-00 Indicator board PCBA 1 G G-J1…
Position
Board Name Line Name Part No. Connected to
No.
P24V,P12V and D5V
D5V, P12V and P24V
J11 009-002601-00 power patching line
power extension line
(009-002286-00)
8-44
Hardware System
Position
Board Name Line Name Part No. Connected to
No.
Data board and
Autoloading board
autoloading board
J8 009-002271-00 PCBA (051-000393-00)
serial port connection
-J11, J12
line
FS preamplification
board
Optical system signal (3101-30-68515)-J1
J3 009-002225-00
line and SS preamplification
board
(3101-30-68517)-J1
8-45
Hardware System
Position
Board Name Line Name Part No. Connected to
No.
Connection line of
J3 009-002284-00 Valve 1-12
drive board valve 1-12
Temperature sensor
J8 009-002276-00 Temperature sensor
connection line
Float switch
J9 009-002278-00 Float switch
connection line
Sample collection
assembly Sample collection
J10 009-002288-00
photocoupler assembly photocoupler
connection line
Syringe photocoupler
J11 009-002289-00 Syringe photocoupler
connection line
SH syringe assembly
J18 009-002457-00 SH motor
motor connection line
LYSE_DIL syringe
LYSE and DIL syringe
J19 assembly motor 009-002283-00
motor
connection line
Autoloading
Electromagnet Sample compartment
board PCBA J3 3102-20-69103-04
connection line electromagnet
051-000393-00
8-46
Hardware System
Position
Board Name Line Name Part No. Connected to
No.
Power board PCBA
P24V,P12V and D5V
J5 009-002286-00 (051-001146-00) -A-J37
power patching line
and A-J35
Mix mechanism
Mix mechanism
J6 photocoupler 009-002516-00
photocoupler
connection line
Autoloading
mechanism position Autoloading sensor
J7 3102-20-69103-01
sensor connection line group 1
1
Autoloading
mechanism position Autoloading sensor
J8 3102-20-69103-02
sensor connection line group 2
2
Data board and
Pinaster main control
autoloading board
J11 009-002271-00 board PCBA
serial port connection
(051-000985-00) -J8
line
Data board and
Pinaster main control
autoloading board
J12 009-002271-00 board PCBA
serial port connection
(051-000985-00) -J8
line
Autoloading
Loading motor lateral
J14 mechanism motor 3102-20-69103-03
and longitudinal motors
connection line
FS
preamplification Optical system signal Analog board PCBA
J1 009-002225-00
board line (051-000982-00) -J3
3101-30-68515
SS
preamplification Optical system signal Analog board PCBA
J1 009-002225-00
board line (051-000982-00) -J3
3101-30-68517
Power board
Switch and power
PCBA J1 3101-20-68589 Power switch
board connection line
051-001146-00
8-47
Hardware System
Position
Board Name Line Name Part No. Connected to
No.
A±12V and A±12V and AC120V
A-J12 AC120Vpower output 009-002287-00 power extension line
line (009-002603-00)
A±12V and A±12V and AC120V
A-J13 AC120Vpower output 009-002287-00 power extension line
line (009-002603-00)
D5V, P12V and P24V
P24V,P12V and D5V
009-002286-00 power extension line
power patching line
(009-002601-00)
A-J35
D5V and P12V power
D5V and P12V power
009-002514-00 extension line
output line
(009-002602-00)
D5V, P12V and P24V
P24V,P12V and D5V
A-J37 009-002286-00 power extension line
power patching line
(009-002601-00)
Fluid detection
Fluid detection board Drive board PCBA
board J1 009-002280-00
connection line (051-000981-00) -J7
051-000983-00
Network cable
J1 (connecting the 009-000043-00 Analyzer PC
Mini network analyzer to PC)
board
051-001122-00 Pinaster main control
J2 Network patching line 009-002561-00 board PCBA
(051-000985-00) -J1
P24V,P12V and D5V
D5V, P12V and P24V
24V-to-13V J1 009-002601-00 power patching line
power extension line
power module (009-002286-00)
PCBA Connection line of
051-001130-00 J2 drive board valve 009-002272-00 Valve 28
13-29
Pinaster main control
Indicator/key board
009-002245-00 board PCBA
connection line
Indicator board (051-000985-00) -J78
J1
051-001062-00 Open vial model Pinaster main control
indicator board 009-002500-00 board PCBA
connection line (051-000985-00) -J78
Pinaster main control
Key board Indicator/key board
J1 009-002245-00 board PCBA
3102-30-69197 connection line
(051-000985-00) -J78
8-48
Hardware System
Table 8-45 Connections of wires and components of the power drive board
8-49
Hardware System
8-50
Hardware System
M4-SP SP motor
M5-SH SH motor
Analyzer working status indicator: the indicator is in the front cover of the analyzer, it
indicated the analyzer status using 3 different colors.
Power-off Off
8-51
Hardware System
8-52
9 Mechanical System
9.1 Analyzer Structure
The instrument is composed of the main unit (including the autoloader) and the PC, as well as
the 3 lyses and diluent connected to it.
9.2 Appearance
The front of the analyzer is shown as Figure9-1.
9-1
Mechanical System
9-2
Mechanical System
9-3
Mechanical System
The side view of the analyzer with the sample compartment door open is shown in Figure 9-4.
Figure 9-4 Side view of the analyzer (sample compartment door open)
9-4
Mechanical System
The layout of the components on the right plate are shown in Figure 9-5.
9-5
Mechanical System
The layout of the components on the left plate are shown in Figure 9-6.
9-6
Mechanical System
3
2 Y X
8 7 6
Figure 9-8 Autoloader (top cover not included to show the sensor position)
1 --- X-direction loading start sensor 2 --- X-direction loading end sensor
3 --- Tube detection position counting sensor 4 --- Tube piercing position counting sensor
5 --- Y-direction feeding start sensor 6 --- Unloading sensor
9-7
Mechanical System
9-8
Mechanical System
9-9
Mechanical System
9-10
Mechanical System
Trigger of
Error Code Error Name Troubleshooting/Solution
Error Report
the loading interference along the loading pathway
mechanism is (e.g. interference with the cover,
at the start obstacle on the loading tray, tube rack
position; or the improperly placed, etc.); b. if there is no
end position interference, check if the motor is OK,
sensors are not and the cables are properly connected
blocked when to it.
the loading 3. If it is at the start or end position,
mechanism is check if the sensors are OK: the
at the end sensors are in different state when
position blocked and not blocked, and the
cables are all properly connected to
them.
0x01005704 X-direction 1. Before troubleshooting, make sure all
loading cables of the related parts are properly
motor sensor connected, e.g. if the cables are well
error connected to the sensors, if the sensor
cables are connected to the correct
positions according to the labels.
2. Check if there is any obstacle
blocking the sensors.
3. Click "Remove Error" to see if the
error can be removed.
4. Check if the sensors are OK: see if
the they are in different state when
The start blocked and not blocked.
position 5. Find the sensor with error and
sensors and remove it. Check if there is dust or
end position splashed fluid covering the glittering
sensors of the side.
loading tray are 7. Wipe the sensor and mount it back to
blocked at the see if the error can be removed. If not,
same time replace with a new one.
0x01005705 X-direction 1. If there is no tube rack on the loading
loading tray: could be X-direction loading end
motor position sensor error. Check if the end
loading error position sensors are OK.
2. If there is/are tube rack(s) on the
loading tray: a. the micro-switch is not
After the pressed by any tube rack: check for
loading is obstacle or interference on the loading
completed, if tray; check if all tube racks are correctly
both the loading placed (make it unable to press the
end position micro-switch); b. the micro-switch is
sensors and pressed by tube rack: check if the
tube detection micro-switch works properly.
micro-switch 3. If all the items above are checked to
are not blocked, be OK, check if the loading motor works
the error will be properly and all the cables are properly
reported connected to it.
0x01005708 Loading 1. Before troubleshooting, make sure all
micro-switch cables of the related parts are properly
error connected, e.g. if the cables are well
Micro-switch connected to the sensors, if the sensor
error: it is cables are connected to the correct
detected to be positions according to the labels.
pressed when it 2. Check if there is any obstacle
is not pressed blocking the sensors.
9-11
Mechanical System
Trigger of
Error Code Error Name Troubleshooting/Solution
Error Report
3. Click "Remove Error" to see if the
error can be removed.
4. Check if the sensors are OK: see if
the they are in different state when
blocked and not blocked.
5. Check if the sensors work improperly,
or the micro-switch is aged that could
be not able to go up and down properly.
0x01005900 Y-direction 1. Check if the software, sequence, and
feeding autoloader versions are correct.
motor in Software 2. Click "Remove Error".
action command 3. Restart the software and the
sending delay analyzer.
0x01005901 Y-direction when there is
motor tube rack along
initialization the feeding
forbidden pathway, the
error will be
reported when
the autoloading 1. Remove the tube rack.
count is started 2. Click "Remove Error".
0x01005902 Unloading The unloading 1. If the unloading tray sensors are
action error tray sensors confirmed to be blocked, remove the
are blocked, tube racks on the tray.
which means 2. If there is no tube rack on the
that the unloading tray. check if the sensors
unloading tray work properly and all cables are well
is full connected.
0x01005903 Y-direction 1. Check if all software and hardware
feeding versions are correct.
motor action 2. Before troubleshooting, make sure all
error cables of the related parts are properly
connected, e.g. if the cables are well
connected to the motor and sensors, if
the sensor cables are connected to the
correct positions according to the
labels, if the motor cables are
connected to the correct positions
according to the labels.
3. Click "Remove Error" to see if the
error can be removed.
4. Perform feeding mechanism self-test
at the "Self-Test" screen, and see if the
error can be removed.
5. Check if the sensors are OK: see if
the they are in different state when
The Y-direction blocked and not blocked.
feeding start 6. For sensor error, find the sensor with
position error and remove it. Check if there is
sensors are dust or splashed fluid covering the
blocked when glittering side.
they are 7. Wipe the sensor and mount it back to
supposed to be see if the error can be removed. If not,
not blocked, or replace with a new one.
not blocked 8. If the error still exists, check the
when they are motor to see if it does not reach the
supposed to be sensor position because of step loss or
blocked. operation obstruction (less likely to
9-12
Mechanical System
Trigger of
Error Code Error Name Troubleshooting/Solution
Error Report
happen).
0x01005904 Unloading 1. Check if all software and hardware
position versions are correct.
sensor error 2. Before troubleshooting, make sure all
cables of the related parts are properly
connected, e.g. if the cables are well
connected to the sensors, if the sensor
cables are connected to the correct
positions according to the labels.
3. Click "Remove Error" to see if the
error can be removed.
4. Perform feeding mechanism self-test
at the "Self-Test" screen, and see if the
error can be removed;
5. Check if the unloading mechanism
works properly, and if the gear rack can
run out properly. If the gear rack cannot
run out or cannot reach the sensor
The unloading detection area, check the mechanism;
start position 6. Check if the sensors are OK: see if
sensors are the they are in different state when
blocked when blocked and not blocked;
they are 7. For sensor error, find the sensor with
supposed to be error and remove it. Check if there is
not blocked, or dust or splashed fluid covering the
not blocked glittering side.
when they are 8. Wipe the sensor and mount it back
supposed to be to see if the error can be removed. If
blocked. not, replace with a new one.
0x01005905 Y-direction The Y-direction 1. Check if all software and hardware
right sensor feeding right versions are correct.
error sensors are 2. Before troubleshooting, make sure all
blocked when cables of the related parts are properly
they are connected, e.g. if the cables are well
supposed to be connected to the motor and sensors, if
not blocked, or the sensor cables are connected to the
not blocked correct positions according to the
when they are labels, if the motor cables are
supposed to be connected to the correct positions
blocked. according to the labels.
0x01005906 Y-direction 3. Click "Remove Error" to see if the
left sensor error can be removed;
error 4. Perform feeding mechanism self-test
at the "Self-Test" screen, and see if the
error can be removed;
5. Check if the sensors are OK: see if
the they are in different state when
The Y-direction blocked and not blocked;
feeding left 6. For sensor error, find the sensor with
sensors are error and remove it. Check if there is
blocked when dust or splashed fluid covering the
they are glittering side;
supposed to be 7 Wipe the sensor and mount it back to
not blocked, or see if the error can be removed. If not,
not blocked replace with a new one;
when they are 8. If the error still exists, check the
supposed to be motor to see if it does not reach the
blocked. sensor position because of step loss or
9-13
Mechanical System
Trigger of
Error Code Error Name Troubleshooting/Solution
Error Report
operation obstruction (less likely to
happen).
0x01005908 Y-direction 1. Check if all software and hardware
feeding versions are correct.
motor 2. Before troubleshooting, make sure all
feeding error cables of the related parts are properly
connected, e.g. if the cables are well
connected to the motor and sensors, if
the sensor cables are connected to the
correct positions according to the
labels, if the motor cables are
connected to the correct positions
according to the labels.
3. Click "Remove Error" to see if the
error can be removed;
4. Perform feeding mechanism self-test
at the "Self-Test" screen, and see if the
error can be removed;
5. Check if the sensors are OK: see if
the they are in different state when
blocked and not blocked;
6. For sensor error, find the sensor with
error and remove it. Check if there is
Y-direction dust or splashed fluid covering the
feeding start glittering side;
sensor not 7 Wipe the sensor and mount it back to
blocked, which see if the error can be removed. If not,
means the replace with a new one;
motor is not at 8. If the error still exists, check the
the start motor to see if it does not reach the
position, and sensor position because of step loss or
the feeding is operation obstruction (less likely to
forbidden happen).
0x01005909 Y-direction Error reported
feeding in the following
motor cases:
unloading 1. Y-direction
error feeding start
position
sensors not 1. Check the state of the Y-direction
blocked feeding start position sensors and that
2. Unloading of the unloading start position sensors,
start position and make sure they are OK and the
sensors not cables are properly connected;
blocked 2. Check if there is any interference in
3. Left sensor the feeding process, which make the
counter not 0 tube rack not fully pass the left sensor
0x0100590D Y-direction 1. Make sure the tube racks are used;
left and right 2. Make sure there is no obstacle along
sensors Unexpected the feeding pathway;
cooperation hop of the left 3. Make sure the left and right sensors,
error and right as well as all cables connected to them
sensors while are OK;
pushing the 4. Make sure the sensor barriers are
tube rack well mounted
0x0100590F Tube Tube detection 1. Check if there is a tube rack with
detection sensor blocked tube(s) at the tube detection position;
sensor after 2. If no, check if the tube detection
9-14
Mechanical System
Trigger of
Error Code Error Name Troubleshooting/Solution
Error Report
blocked after initialization sensors are OK; place a tube at the
initialization. tube detection position, and check if the
sensors are in different state when
blocked and not blocked;
3. Check if there is an obstacle along
the pathway of L sensor detection;
4. Wipe the sensors and check if they
work properly;
5. If the error still exists, replace the
tube detection sensors with a new pair
0x01005910 Right sensor 1. Check if there is any obstacle along
damaged or the tube rack feeding pathway which
unable to prevent the tube rack from moving;
drag the 2. Check if all software and hardware
tube rack Right sensor versions are correct;
does not 3. Check if all cables connected to the
change after right sensor are properly connected,
trying the and if there is any damage;
feeding action 4. Check if the right sensor is OK: press
for 5 times, the spring leaf, and check if the sensor
which is is in normal state;
probably 5. Check if the sensor barrier is properly
because the mounted, and if the sensor is blocked
right sensor is when pressing the barrier;
damaged or 6. Replace the cables, and check if the
unable to drag sensor works properly;
the tube rack 7. Replace the sensor
0x01005a01 Opening Unable to open 1. Check if the software, autoloading
sample the sample board MCU and FPGA versions are
compartment compartment correct; check if the autoloading board
door failed door 24V, 12V, 5V powers are normal;
/ / 2. Check if the sample compartment
door is open; if it is open, check if the
compartment door detection sensors
and the cables connected to them are
OK;
3. If the door is not open, check if the
electromagnet cable is well connected;
4. Pull the tail of the electromagnet to
open and close the compartment door,
and see if there is interference or
obstruction. Check if the jump ring of
the electromagnet core falls off;
5. If the jump ring falls off, install a new
one;
6. If there is obstruction, remove the
electromagnet assembly from the
sample compartment assembly, and
then remove the guide bushing.
Remount the guide bushing and make
sure there is no obstruction between the
bushing and the protruded axle of the
magnet core (rotate the core in different
directions, and there should be no
obstruction);
Closing sample 7. Mount the electromagnet assembly
compartment back to the sample compartment
door failed assembly, and check if the
9-15
Mechanical System
Trigger of
Error Code Error Name Troubleshooting/Solution
Error Report
compartment door can be opened.
Note: make sure the sample probe is
not inside the sample compartment
while opening and closing the
compartment door.
0x01005a04 Adjustment This error only occurs in the position
parameter adjustment performed by manufacture
out of range or service engineers
Value out of Re-adjust when this error is reported
range while (according to F-3102-CTO-01
adjusting Instrument Adjustment Procedure
0x01005a05 Read 1 Check if the software, autoloading
scanner board MCU and FPGA versions are
error correct; check if the autoloading board
24V, 12V, 5V powers are normal;
2. Check if the cables to the scanner
are well connected;
Read scanner 3. Restart the instrument, and see if the
error. No error still exists;
message sent 4. If the error still exists, replace the
back in 500ms scanner
9-16
Mechanical System
9-17
Mechanical System
9-18
Mechanical System
X-direction mechanism consists of the horizontal bracket, X-direction guide bar, guide bar
fixer, synchronous belt and pulley, X-direction start position sensor, and motor. There are
9-19
Mechanical System
notches on the horizontal bracket, each of which matches a position detection sensor to
detect the position
Figure 9-16 The structure of the X-direction mechanism of the sampling assembly
For the closed-tube model, there are 5 positions on the X direction for the sample probe,
which are (from back to front): RBC bath position, WBC bath position, DIFF bath position,
autoloading sampling position, closed-tube sampling position.
For the open-vial model, there are 4 positions on the X direction for the sample probe,
which are (from back to front): RBC bath position, WBC bath position, DIFF bath position,
and open-vial sampling position.
9-20
Mechanical System
Figure 9-17 The structure of the Y-direction mechanism of the sampling assembly
The sample probe is driven by the piercing slide. The rotation of the motor is transferred
into the linear motion by the lead screw nut set, which is the Y-direction motion guided by
the guiding axle. The Y-direction start position sensor is installed on the piercing bracket,
and the barrier of the piercing slide is used as the barrier of the sensor, in order to locate
the Y-direction positions of the sample probe.
9-21
Mechanical System
Check if the X-direction motor is damaged. Replace with a new one if damaged.
The following sections introduce the replacing procedures of motors and sensors (taking
closed-tube model as an example, the replacing procedures of which is similar to that of
the open-vial model).
Figure 9-18 Replacing the X-direction start position sensors of the sampling assembly
Remove the 2 M3x8 screws fixing the X-direction start position sensor bracket using a
cross-headed screwdriver, and then remove the X-direction start position sensors together
with the bracket from the analyzer. Remove the 2 M3x6 screws fixing the sensors using a
cross-headed screwdriver, and then remove the sensors. Install the new sensors in the
reversed order of the steps above.
You can also replace the sensors after removing the sampling assembly.
After the replacement, pull the Y-direction movement module, and check if the X-direction
start position sensor barrier are between the 2 sensors.
9-22
Mechanical System
Figure 9-19 Replacing the X-direction detection sensors of the sampling assembly
Remove the 2 M3x5 screws fixing the X-direction detection sensor bracket using an inner
hexagon spanner, and then remove the X-direction detection sensors together with the
bracket from the analyzer. Remove the 2 M3x6 screws fixing the sensors using a
cross-headed screwdriver, and then remove the sensors. Install the new sensors in the
reversed order of the steps above.
You can also replace the sensors after removing the sampling assembly.
After the replacement, pull the X-direction movement module, and check if the fold of the
horizontal bracket (used as the barrier) is between the 2 sensors. Make adjustment if
necessary.
9-23
Mechanical System
9-24
Mechanical System
Remove the 4 M3x8 screws (with washers) fixing the X-direction motor with a
cross-headed screwdriver, and then remove the motor with the driving band wheel fixed
on it.
Remove or loosen the 2 M3x5 set screws fixing the driving band wheel with an inner
hexagon spanner, and then remove the driving band wheel from the X-direction motor
shaft.
Install the X-direction motor in the reversed order of the steps above after the replacement.
Pay attention to the notes below while installing:
1. The distance from the driving band wheel to the end of the X-direction motor shaft is
2.5mm;
2. One of the set screws of the driving band wheel should be pressed to the flat of the
X-direction motor shaft;
3. While installing the X-direction motor on the sampling assembly, the cable coming out
of the motor should stand upwards (pointing to the top of the instrument);
4. While installing the X-direction motor on the sampling assembly, adjust the lengthwise
position, making the synchronous belt properly stretched; after the installation of the
sampling assembly, lean it to the left and then right, to see if the Y-direction movement
module can slide down properly;
5. While installing the sampling assembly on the analyzer, make sure all cables and tubes
are well protected.
9-25
Mechanical System
As shown in the figure above, remove the 2 M3x10 inner hexagon screws fixing the press
bar of the synchronous belt using an inner hexagon spanner, remove the M4x8 screw
fixing the horizontal guide bar and guide bar fixing pedestal using a cross-headed
screwdriver, remove the 2 M4x20 inner hexagon screws fixing the driven wheel assembly
using an inner hexagon spanner, and then demount the Y-direction movement mechanism
from the sampling assembly.
As shown in the figure above, remove or loosen the 2 M3x5 set screws fixing the
Y-direction guide bar using an inner hexagon spanner, remove the 4 M3x10 inner hexagon
screws (with spring and flat washers) fixing the Y-direction motor using an inner hexagon
spanner, and then remove the Y-direction motor.
Install the Y-direction motor in the reversed order of the steps above after the replacement.
Pay attention to the notes below while installing:
1. One of the set screws of the Y-direction guide bar should be pressed to the flat of the
Y-direction motor shaft;
2. While installing the Y-direction motor on the Y-direction movement mechanism, the
9-26
Mechanical System
2
8
9-27
Mechanical System
10
2
3 9
8
4
7
5
1
11
10
3
4 9
5
8
9-28
Mechanical System
9
1
8
2
7
3
9-29
Mechanical System
Troubleshooting/Solut
Error Code Error Name Trigger of Error Report
ion
3. Should not be at the start 1. Check if the software
position after resetting, but version and hardware
the sensors are blocked. version are correct, and
if the 24V, 12V and 5V
power are proper;
2. Check if the cables of
the sensors and motors
Sampling are well connected and
syringe Sensor not blocked when make sure there is no
0x01000302 aspiration/dispe the syringe is supposed to bad connection;
nsation action be at the start position; 3. Click "Remove Error"
failure 1 to see if it can be
Sampling removed;
syringe Sensor blocked when the 4. Perform self-test at
0x01000303 aspiration/dispe syringe is supposed to be the "Self-test" screen;
nsation action out of the start position 5. When the motor is
failure 2 operating, check if the
Sample syringe When the syringe starts LED indicator of the
aspiration/dispe moving, it is supposed to be corresponding channel
0x01000304 (ASP-M3_LED2;SP-M4
nsation action at the start position, but the
not allowed 1 sensor is not blocked; _LED2, SH-M5_LED2,
Sample syringe When the syringe starts DIL-M6_LED2,
aspiration/dispe moving, it is supposed to be LYSE-M7_LED2) is
0x01000305 flickering. If not, replace
nsation action out of the start position, but
not allowed 2 the sensor is blocked; the driver board;
1. Sensors are still blocked 6. Check if the sensor
when the syringe is out of states are different
the detection area in when it is blocked and
initialization; unblocked;
Sample 7. Find the sensor with
2. Sensors are not blocked
injection error and remove it.
0x01000311 when the syringe is inside
syringe sensor Check if there is dust or
sensor detection area in
error splashed fluid covering
initialization;
3. Should not be at the start the glittering side;
position after resetting, but 8. Wipe the sensor and
the sensors are blocked. mount it back to see if
Sample the error can be
injection removed. If not, replace
Sensor not blocked when with a new one;
syringe
0x01000312 the syringe is supposed to 9. Check if the syringe
aspiration/dispe
be at the start position; assembly is well
nsation action
failure 1 installed; remove the
Sample shielding cover, wipe
injection the lead screw and the
Sensor blocked when the guide bar, and add
syringe
0x01000313 syringe is supposed to be lubricant;
aspiration/dispe
out of the start position 10. If the error still
nsation action
failure 2 exists, replace the
Sample corresponding syringe
injection assembly with a new
Sensor not blocked when one.
syringe
0x01000314 the syringe is supposed to
aspiration/dispe
be at the start position;
nsation action
not allowed 1
Sample
Sensor blocked when the
injection
0x01000315 syringe is supposed to be
syringe
out of the start position
aspiration/dispe
9-30
Mechanical System
Troubleshooting/Solut
Error Code Error Name Trigger of Error Report
ion
nsation action
not allowed 2
1. Sensors are still blocked
when the syringe is out of
the detection area in
initialization;
Sheath fluid 2. Sensors are not blocked
0x01000321 syringe sensor when the syringe is inside
error sensor detection area in
initialization;
3. Should not be at the start
position after resetting, but
the sensors are blocked.
Sheath fluid
syringe Sensor not blocked when
0x01000322 aspiration/dispe the syringe is supposed to
nsation action be at the start position;
failure 1
Sheath fluid
syringe Sensor blocked when the
0x01000323 aspiration/dispe syringe is supposed to be
nsation action out of the start position
failure 2
Sheath fluid
syringe Sensor not blocked when
0x01000324 aspiration/dispe the syringe is supposed to
nsation action be at the start position;
not allowed 1
Sheath fluid
syringe Sensor blocked when the
0x01000325 aspiration/dispe syringe is supposed to be
nsation action out of the start position
not allowed 2
1. Sensors are still blocked
when the syringe is out of
the detection area in
initialization;
2. Sensors are not blocked
Lyse syringe
0x01000331 when the syringe is inside
sensor error
sensor detection area in
initialization;
3. Should not be at the start
position after resetting, but
the sensors are blocked.
Lyse syringe
Sensor not blocked when
aspiration/dispe
0x01000332 the syringe is supposed to
nsation action
be at the start position;
failure 1
Lyse syringe
Sensor blocked when the
aspiration/dispe
0x01000333 syringe is supposed to be
nsation action
out of the start position
failure 2
Lyse syringe
Sensor not blocked when
aspiration/dispe
0x01000334 the syringe is supposed to
nsation action
be at the start position;
not allowed 1
9-31
Mechanical System
Troubleshooting/Solut
Error Code Error Name Trigger of Error Report
ion
Lyse syringe
Sensor blocked when the
aspiration/dispe
0x01000335 syringe is supposed to be
nsation action
out of the start position
not allowed 2
1. Sensors are still blocked
when the syringe is out of
the detection area in
initialization;
2. Sensors are not blocked
Diluent syringe
0x01000341 when the syringe is inside
sensor error
sensor detection area in
initialization;
3. Should not be at the start
position after resetting, but
the sensors are blocked.
Diluent syringe
Sensor not blocked when
aspiration/dispe
0x01000342 the syringe is supposed to
nsation action
be at the start position;
failure 1
Diluent syringe
Sensor blocked when the
aspiration/dispe
0x01000343 syringe is supposed to be
nsation action
out of the start position
failure 2
Diluent syringe
Sensor not blocked when
aspiration/dispe
0x01000344 the syringe is supposed to
nsation action
be at the start position;
not allowed 1
Diluent syringe
Sensor blocked when the
aspiration/dispe
0x01000345 syringe is supposed to be
nsation action
out of the start position
not allowed 2
Replace a new Mindray
syringe. The
replacement should be
performed as follows:
10ml syringe remove the tailored
/ /
leakage screw, and then the 4
MX25 screws fixing the
syringe. Unplug the
tubes, and replace the
syringe with a new one.
Replace a new Mindray
syringe. The
replacement should be
performed as follows:
100ul/250ul/2.5
remove the tailored
/ ml syringe /
screw, and then the 4
leakage
MX25 syringing fixing
plate. Unplug the tubes
and remove the syringe
for replacement.
9-32
Mechanical System
9-33
Mechanical System
Note: the "Sample Probe Up Position" variation is included in the formula of autoloading and
closed-tube piercing position step calculation. Therefore, sample probe up position adjustment
is of the first priority; if you adjust the autoloading/closed-tube piercing position first, further
adjustment is needed after the sample probe up position adjustment, since the change of the
sample probe up position will lead to the change of the autoloading and closed-tube piercing
positions.
Tools needed for the adjustment include:
Cross-headed screwdriver
Ruler
Fitting pin or sample probe up position fixture
If the requirements are not met, adjust again using the "Up" and "Down" buttons.
If there is no fixture, use a ruler and the fitting pin. For open-vial models, it is required that the
sample probe is inside the probe wipe, and the probe tip is 5.3±0.2mm up from the bottom of
the probe wipe; for closed-tube models, the probe tip should be 7.2±0.2mm up from the bottom
of the probe wipe.
9-34
Mechanical System
sampling position.
Exit the "Debug screen" and run a blank count cycle. Make sure that the sample probe do not
touch the bath during DIFF bath mixing, WBC bath Mixing, and RBC bath mixing.
Check if the pierced position is at the center of the tube cap (you can cover the tube cap with
adhesive tape for easier check) without obvious deviation. If it does not meet the requirement,
click "Return", and the sample probe will withdraw from the tube. Adjust as instructed below:
① For deviation to left or right, remove the 3 screws fixing the autoloader, and make the
autoloader appressed to the front plate. Move the autoloader to left/right, and secure the
screws after the adjustment.
② For deviation to front or back, check if the autoloader is appressed to the front plate
(mechanical method), and adjust the position of the sample probe using the "Forward" and
"Backward" buttons at the "Debug" screen.
③ After the adjustment, click "Check" to check if the piercing position is at the center.
3)Remove the tube cap of the tube and put it back into the tube rack. At the "Debug" screen of
the PC software, click the "Check" button of the "AL Piercing Position" area, and the sample
probe start aspirating from the collection tube. Put the collection tube upward manually, and
make the probe tip touch the bottom of the tube. Check if the tube is 3-5mm higher (you can
mark on the exterior wall of the tube)
9-35
Mechanical System
Check if the pierced position is at the center of the tube cap (you can cover the tube cap with
adhesive tape for easier check) without obvious deviation. If it does not meet the requirement,
click "Return", and the sample probe will withdraw from the tube. Adjust as instructed below:
① For deviation to left or right, remove the 3 screws fixing the sample compartment, and
check if the distance between the center of the compartment and the front plate is 117.5mm,
move the compartment to left/right, and secure the screws after the adjustment
2 screws on
1 screw
the right
on the left
② For deviation to front or back, check if the distance between the center of the compartment
and the front plate is 117.5mm, and adjust the position of the sample probe using the
"Forward" and "Backward" buttons at the "Debug" screen.
③ After the adjustment, click "Check" to check if the piercing position is at the center.
Remove the tube cap of the tube and put it back into the adapter. At the "Debug" screen of the
PC software, click the "Check" button of the "CT Piercing Position" area, and the sample probe
start aspirating from the collection tube. Put the collection tube upward manually, and make the
probe tip touch the bottom of the tube. Check if the tube is 3-5mm higher (you can mark on the
exterior wall of the tube)
9-36
Mechanical System
3)If the left/right position is incorrect, loosen the 4 M4X8 cross-recessed panhead screws, and
then move the mixing assembly to left/right until the manipulator is aligned with the center of
the tube position, and then secure the screws.
4 M4X8 cross-
recessed panhead
screws
Pin
Note 1: prevent from deformation while moving the mixing assembly, and adjust the pin in the
assembly if necessary.
Note 2: after the screws are secured, the position of the mixing assembly may change, and
you need to check the clamp position again.
9-37
Mechanical System
threaded hole in the front) and the old one (shorter, a threaded hole in the front, and two
sides of it are made flat) will use the same adjustment method. Click the “Up” button in the
“Manipulator” section, loosen the two M3X10 stainless-steel inner hexagon screws used
to fix the tube clamp, then move the clamp forward/backward until the manipulator is
aligned with the center of the tube position, and then secure the screws.
Loosen the 2 inner
The old rotation
hexagon screws.
axis, a threaded
The new rotation
hole in the front,
axis, no threaded
and two sides of it
hole in the front.
are made flat.
b) For the old tube clamp (115-006303-00) and old rotation axis, use a retaining screw on the
top of the old tube clamp and a M3X10 stainless-steel inner hexagon screw in the front of
the rotation axis to fix them together. And the manipulator location can’t be adjusted
forward or backward.
An inner
hexagon screw
and a retaining
screw
c) Note: the old tube clamp can’t be installed on the new rotation axis. Because the new
rotation axis is fixed with only one retaining screw, then the old tube clamp can’t be locked
tightly.
9-38
Mechanical System
5)Select "X Backward", and place a 12X75 tube into the tube position the clamp aims at.
Check the relative state of the tube after it is placed back using the X direction, Z direction, and
mix buttons. It is required there should be no interference in the tube pinching and mixing
processes, and all actions should be completed smoothly.
No interference, and
all actions are
completed smoothly. If
the manipulator
position is not correct,
adjust using the
buttons in the
"Manipulator Position"
area.
If the requirement is not met, adjust the manipulator position using the buttons in the
"Manipulator Position" area. After the adjustment, go back to the "Autoloader" screen and
check.
Ensure that:
The height of the underside of the tube fixing plate over the tube clamp is slightly lower than
the vertical back plate for the tube.
The underside of the tube clamp is at least 1.5 mm higher than the topside of the tube rack.
Tube fixing plate over the Vertical back plate
tube clamp for the tube
≥1.5mm
9-39
Mechanical System
3)Place a piece of paper right before the tube rack, and observe the reflection of the beam on
the paper. The beam should be aligned with the center of the tube position notch.
If the requirement is not met, loosen the 2 M3X8 panhead screws, and manually adjust the
position of the beam. After the adjustment is successful, re-secure the screws. After the screws
are secured, re-check the position of the beam.
Loosen the
2 screws to
adjust the
position of
the built-in
barcode
scanner.
9-40
Mechanical System
PN:051-000982-00
New New HGB assembly with Applicable
circuit “new 新 ” label (applicable to EJ295F
board to EJ295F and later version) and later
version
New label
location
PN:051-000982-01
115-065236-00 HGB
Assembly (New LED)
115-065237-00 WBC
Dial switch + label Assembly(New LED)
Old HGB assembly
(applicable to version
before EJ295F)
801-3102-00050-00 HGB
Unit(BC-5380)
801-3102-00012-00 WBC
bath assembly
9.12.1.2Solution
Maintenance scenario 1: new circuit board + new HGB assembly
1. Replace a new HGB circuit board.
2. Adjust the dial switch on the circuit board according to the following method.
9-41
Mechanical System
3. Verify whether the adjustment has achieved satisfactory results referring to “Verification
after Replacement and Adjustment (Applicable to Maintenance Scenario 1&2)”.
3. Verify whether the adjustment has achieved satisfactory results referring to “Verification
after Replacement and Adjustment (Applicable to Maintenance Scenario 1&2)”.
NO
Finish
9-42
P/N: 046-000142-00(8.0)