Calibration Report
Client: LAMBERT NIG LTD Certificate No: WONDIlLAMBERIOO112013
EQUIPMENT: EARTH TESTER SERVICE: CALIBRATION
MANUFACTURER: KYORITSU METHOD: DIRECT COMPARISON
RANGE: 0-2000 n, 0-200 VAC TRACEABILITY: NIST TRACEABILITY
MODEL NO.: 4105A AMBIENT TEMP: 230 C
SERIAL NO: W8189871 RlH: 43% RlH
TYPE: DIGITAL TEST DATE: 10-05-2013
DUE DATE: 09-11-2013

PHYSICAL INSPECTION
DISPLAY RANGES TEST KEY PROBS RANGE SlECTOR KEY
OK OK OK OK OK
CALIBRATION DATE
APPLIED VOLTAGE MEASURED VOLTAGE APPLIED RESISTANCE MEASURED RESISTANCE STATUS
(VAC) (VAC) (0) (0) II
I
50.00 50.00 20.00 19.99 PASS
100.00 100.00 50.00 50,00 PASS
150.00 150.00 100.00 100.00 PASS
200.00 199.99 150.00 149.90 PASS
- - 200.00 199.90 PASS
- - 500.00 500.00 PASS
- - 1000.00 1000.00 PASS
- - 1500.00 1500.00 PASS
- - 2000.00 1999.90 PASS

Reference Standard Used


ASSET DESCRIPTION
FLUKE 725 MULTIPROCESS CALIBRATOR SIN: 8307038
CONSTANT CURRENT INVERTER
WONDI' COMPANY LIMITED certifies that the above listed instrument has been tested for conformity to all specifications as
published (unless otherwise noted). It has been tested using measurement standards traceable to the National Institute of
Standard and Technology (NIST) accepted intrinsic standards of measurement, or derived by the ratio type of setf-calibration
techniques. This calibration complies with MIL-STD 45662A and ANSIINCSL Z540-1-1994 and ISOIIEC 9001 . This calibration
respects the Nigerian Industrial Standard (N IS) of "Standard Organization of Nigeria (SON) and "DPR"

This report may only be reproduced with written approval of the calibration laboratory
REMARK: The data recorded are as good as accuracy and uncertainty allows.
Equipment is OK for application.

~.
==--=-:.