New as of: 06.

2021

Teneo, Sinius
Maintenance log =

English
Teneo, Sinius
Maintenance log

Customer Dealer
Address, tel.: Address, tel.:

Seriennummern / Serial numbers / Numéros de série / Numéros de serie  Item Serial no.
⑥ ①
LEDview   Teneo
⑦ ⑫ LEDview S   Sinius ②

 Sinius CS ③
⑥ ⑩ ⑪  Sinius TS ④
Monitor ⑤
⑧ 22“  ⑥
⑪ Rückseite


rear
⑫ verso
dorso

② ④ ⑧ ⑨

SiroCam
 ⑪
AF+
⑨ SiroCam
S


 ⑬ ⑫
AF S



0

⑤③ ①

Seriennummern / Serial numbers / Numéros de série / Numéros de serie
Please enter the serial number from the label of the maintenance set!
⑭ Maintenance set  Maintenance set ⑭
Teneo, Sinius Teneo, Sinius:
62 30 192

Prior to starting maintenance, a functional test of the treatment center was performed and any defects found were
disclosed to the practice staff. These defects will not be corrected as part of this maintenance work.
Maintenance was performed according to this maintenance log using OEM parts from Dentsply Sirona.

______________________ _____________________ ______________________


Date of maintenance Name of engineer Signature

Charac-
Explanation
ter
• Is applicable or present
– Is not applicable or is not present
 Free space for documenting work performed

61 93 903 D3509
D3509.105.01.26.02
Sirona Dental Systems GmbH

Information for the service engineer:

Equipment:
General: This maintenance log is a component of the maintenance set (REF.: 62 30 192) and
is designed for treatment centers which feature the corresponding standard config-
uration. The parts contained in the maintenance set accommodate treatment cen-
ters with 3 burr drives and 2 suction handpieces. Additionally required parts must be
ordered separately for treatment centers with 4 burr drives and 3 suction hand-
pieces. Please inquire in advance whether additional spare parts / replacement parts
are required for maintenance that differ from the standard configuration!

Instruments: Please observe the operating instructions for the relevant instruments. Regarding
instruments, the maintenance accessories are designed for the following instrument
equipment:

Dentist element: Teneo Sinius

båÖäáëÜ
1 x Sprayvit 1 x Sprayvit
2 x SL electric motor or BL ISO motor 2 x BL ISO motors
1 x T1 control 1 x T1 control
1 x BL implant motor 1 x SiroSonic TL (ZEG)
1 x SiroSonic TL (ZEG)

Assistant element: 1 x Sprayvit


1 x suction hose
1 x saliva ejector hose

General information:
Tools, specified resources (e.g.: manometer), spare parts for repairs, and lubricants
are not contained in this maintenance set.
Please be sure to note that only the specified original lubricants may be used. Sirona

cêÉåÅÜ
assumes no liability for damage caused by the use of lubricants that have not been
approved.
Unless otherwise specified, all O-rings are to be greased a little with LCA 3801.
Exception: Oil the O-rings of the instruments in the dentist element with a little T1
spray.
Do not oil or grease ProSmile O-rings.

Instruction:
The work sequence specified here should serve only as an orientation. It can be
changed at one's own discretion as long as the applicable hygiene standards are
observed. For enhanced clarity, the completed work can be signed for in the last col-
umn.
Please note that the color of identical materials (e.g. O-rings) may vary in hue.
pé~åáëÜ

ESD safety note:


ATTENTION: When working with Electrostatic discharge components in connection with the treat-
ment center, be sure to observe the generally applicable ESDDirectives for han-
dling electronic components.
If you open the device: Please observe the usual precautionary measures for han-
dling printed circuit boards (ESD). Make sure you touch a ground point to discharge
yourself prior to touching the components.

61 93 903 D3509
D3509.105.01.26.02 3
4
Maintenance log for Teneo, Sinius
Work to be performed Interval Teneo Sinius Spare part REF. Qty Bag no.:

1. Treatment center:
Ensure the basic functions of the entire treatment center (patient chair, 1 year • • - - - -
dentist element, assistant element, examination light, foot switch and
instrument control, patient communication, and software) before begin-
ning maintenance.
Functional test of the safety switches for the assistant element, cuspidor 1 year • - - - - -
bowl, chair lift, backrest, toeboard, lift frame cover, rear housing, left/right
manual switching strip and armrests, as described in the Maintenance
Manual.
Function check of the safety switch, assistant element, cuspidor bowl, 1 year - • - - - -
backrest, footrest, lifting frame, rear cover, as described in the mainte-
nance manual.
Check the tightness of the screw connections in the patient chair, water 1 year • • - - - -
unit, mounting bar and attachments.
SIVISION swivel arm: Check the movement function of the swivel arm, in 1 year • • - - - -
particular the fixed connection of the profile end pieces to the profile sec-
tion tube of the flexible support arm under the bellows.
Check the main valves (air/water sealing effect) and the function of the 1 year • • - - - -
main switch (see service manual for procedure).
Perform visual check of hydro pneumatic connections (air, water and 1 year • • - - - -
waste water lines) for leakage and functional check for sufficient flow rate.
Perform a visual check of the electric plug connections; in particular, 1 year • • - - - -
check the supply cable for damage.
Motorized headrest: During the procedure, check the spindle for running 2 years • • - - - -
noises. Manually move the headrest (by means of pressing the buttons
together) to check for running noises.

2. Compressor:
Check switch-on and switch-off pressure. Connection air pressure 1 year • • - - - -
between 5.5 and 7.5 bar.
Perform maintenance as specified by manufacturer, remove contamina- 1 year • • - - - -
tion, drain off condensation water, etc.
D3509.105.01.26.02

or as specified
by manufacturer
61 93 903 D3509
D3509.105.01.26.02
61 93 903 D3509
Maintenance log for Teneo, Sinius
Work to be performed Interval Teneo Sinius Spare part REF. Qty Bag no.:

3. Suction machine:
Check the intake vacuum. Volume flow min. 270 Nl/min. (Standard liter 1 year • • - - - -
per minute).
Perform maintenance as specified by manufacturer, remove contamina- 1 year • • - - - -
tion, drain off condensation water, etc. or as specified
by manufacturer

4. Examination light:
Check movement mechanisms of unit model, set brakes if necessary. 1 year • • - - - -
Caution: Lamp must hover when the vertical brake is released, otherwise
replace pneumatic spring. Pull brake only slightly.
SIROLUX FANTASTIC: Blow out the fan and rear cap; hold the fan still 1 year • - - - - -
while so doing.
SIROLUX FANTASTIC: Check firm fit of reflector and handles. 1 year • - - - - -
SIROLUX FANTASTIC: Check electrical connections and lamp contacts. 1 year • - - - - -
SIROLUX FANTASTIC serial no. 30 000 to 60 000: once • • Shadowing 62 79 629 - -
Replace shadowing upgrade kit properly. upgrade1)
SIROLUX FANTASTIC serial no. <30 000 or 1 year • • Halogen lamp1) 62 03 629 - -
> 60 000: The halogen lamp must be replaced by qualified personnel.
SIROLUX FANTASTIC: Check focusing, adjust according to operating 1 year • - - - - -
instructions if necessary
LEDview/LEDview Plus: Check composite and functions according to 1 year • • - - - -
operating instructions.

5. Tray:
Check arm movement, adjust brakes if necessary. When the vertical 1 year • • - - - -
brake is released, no noises which indicate defects may occur.
Check rotary movement of tray, adjust brake if necessary. Check height 1 year • • - - - -
adjustment.
5
6
Maintenance log for Teneo, Sinius
Work to be performed Interval Teneo Sinius Spare part REF. Qty Bag no.:

6. Patient chair/connection box:


When maintenance work is carried out on a dental treatment unit with 1 year - • Double-jointed 64 75 417 1 -
built-in double-jointed headrest (REF. 6460211) up to serial number headrest1)
25,000, the double-jointed headrest must be replaced once due to wear
and tear.
The headrest is not part of the maintenance kit and must therefore be
ordered separately (Spare part: REF 6475417).
If you have any further questions please contact the customer service
center.
Patient chair: Check the backrest and seat padding for a correct fit. 1 year • • - - - -
Patient chair: If a MultiMotion headrest is installed, check the locking 1 year • - - - - -
mechanisms of the headrest.
Patient chair: If double-jointed headrest, check the locking mechanisms 1 year - • - - - -
of the headrest.
Patient chair: If a motorized headrest is installed, replace the drive belt for 2 years • • Round belt 58 91 051 4 6.1
the headrest and sword extension in the backrest.
Patient chair: Check the hard-soft connection of the head support cover 2 years • • Head plate 61 38 106 1 6.1
for surface damage and replace it if necessary.
Patient chair: Visually check bearing shaft drive for firm fit (only optically, 1 year • • - - - -
i.e. without tools).
Patient chair: Check drive spindles of the lifting, tilting and backrest 1 year • - - - - -
motors to ensure these are tight, and lubricate with a thin coat of PD2
grease.
Patient chair: Check the spindles of the lift and backrest motor drives to 1 year - • - - - -
ensure they are tight.
Patient chair: Lubricate the spindles in the backrest and the headrest with 1 year • • - - - -
a thin coat of grease E.
Patient chair: Check the play in the guideways of the backrest (ergomo- 1 year • • Teneo roller, coupl.1) 54 33 870 1 -
tion movement), replace the outer rollers if necessary and grease with Sinius roller, coupl.1) 64 89 376 1
PD2 (including the guideway).
Armrest option: Replace the armrest fixing screw with a coated M5x10 1 year - • Screw M5x10 67 62 624 2 6.1
D3509.105.01.26.02

screw. If present, use both armrests (left and right version). Tighten the Tuflok-KLF
61 93 903 D3509

M5x10 screw as described on the instruction sheet REF 67 78 075 .


Check the function of the armrest.
Connection box: Read service codes via Ethernet connection on NSA/ 1 year • • - - - -
HSA board and evaluate according to service manual.
D3509.105.01.26.02
61 93 903 D3509
Maintenance log for Teneo, Sinius
Work to be performed Interval Teneo Sinius Spare part REF. Qty Bag no.:

6. Patient chair/connection box:


Compare the current treatment center software release and, if necessary, 1 year • • - - - -
carry out an update to the current software version.

7. Water unit:
Service block: Check air pressure at MA1, if necessary set to 4.7 bar flow- 1 year • • - - - -
ing with AD1.
Service block: Check air pressure at MA2, if necessary set to 2.3 bar flow- 1 year • • - - - -
ing with AD2.
Service block: Replace air and water filters AF1, AF2, and AF6 (only if dis- 1 year • • Filter element 14 38 360 3 7.1
infectant system is present). Molded seal filter 65 38 065 3
Tumbler outlet: Replace outlet nozzle and O-rings. Grease O-rings with 1 year • • Outlet nozzle 14 45 407 1 7.2
LCA 3801. O-ring 6.07 x 1.78 70 36 353 1
O-ring 8.73 x 1.78 70 53 721 1
Automatic tumbler filling: Remove any deposits of dissolved matter 1 year • • - - - -
between the two water outlet nozzles.
Undo the quick-release fastener to remove and clean the cuspidor bowl, 1 year • • Funnel 33 19 592 1 7.2
replace funnel.
Visual check for leakage. 1 year • • - - - -
If a disinfection system is installed: Check inlet screen of disinfectant tank 1 year • • - - - -
for contamination, clean if necessary.
If disinfection system is installed: Replace cover of mixing tank and 1 year • • Cover mixing tank, 67 62 491 1 7.3
remount solenoid valve and level sensor with cable according to instruc- preassembled
tion sheet REF 6762517. Replace both O-rings on the solenoid valve. O-ring 7,0 x 1,8 18 72 030 2
If a disinfection system is installed: Replace the mixing tanks, 2 x plugs, 1 year • • Tank 62 24 120 2 7.3
O-rings and flushing tanks. Grease O-rings with LCA 3801. Mount insert Plugs 14 43 287 2
plate in flushing tank. O-ring 9,25 x 1,78 70 08 097 4
Insert plate 61 80 553 1
Cuspidor bowl tumbler holder: Replace O-rings. Grease O-rings with LCA 1 year • • O-ring 60 38 587 1 7.4
3801. O-ring 60 38 579 1
Cuspidor bowl tumbler holder: Replace tumbler holder. 1 year • • Tumbler holder 65 71 009 1 7.4
If an amalgam separator is installed: Remove and clean the water detec- 1 year • • - - - -
tor.
If an amalgam separator is installed: Perform visual check for leaks. 1 year • • - - - -
7
8
Maintenance log for Teneo, Sinius
Work to be performed Interval Teneo Sinius Spare part REF. Qty Bag no.:

7. Water unit:
If an amalgam separator or automatic separator is installed: Remove and 1 year • • Sealing ring 29 59 005 1 7.5
clean lower part. The return suction line must be free! Replace the sealing
ring and grease with LCA 3801.
If an amalgam separator or automatic separator is installed: Replace the 1 year • • Gasket 14 34 245 1 7.5
check valve ("gasket") and thrust piece in the automatic separator pump. Thrust piece 14 34 278 1
If an amalgam separator or automatic separator is installed: Replace the 1 year • • Cuff 29 59 443 1 7.5
cuff between the separator tank and the pump.
If an amalgam or automatic separator is installed: 1 year • • - - - -
Remove and clean the rubber siphon.
If an amalgam separator or automatic separator is installed: 1 year • • - - - -
Remove and clean separator tank depending on contamination.
If a wet suction system is installed: Remove the yellow filter screen and 1 year • • - - - -
clean it.
Check drain hose in the entire treatment center for deposits. Ensure that 1 year • • - - - -
the on-site waste water pipe has good drainage. If necessary, clean the
waste water line.
Clean selecting valve seat MV33. Replace membrane. 1 year • • Membranes 59 29 807 1 7.5
Suction hose cleaning option: The hose connection is to be replaced from 1 year • • Hose connection 66 58 673 2 7.5
hose W53, and the O-ring on the chemical tank outlet and the inlet of the O-ring 3.68x1.78 65 02 061 2
MV51.
Check suction hose cleaning function.
Remove connection hose from the WU to the ASE head in the assistant 1 year - • O-ring 20.3 x 2.4 70 52 269 1 8.6
element (Sinius)/on the water unit. Replace the O-rings and grease with O-ring 24 x 2.5 70 52 251 1
LCA 3801.
If option CDS60: 1 year • • - - - -
Visual check for leakage on the metering valve.
While the tank for cleaning the suction hose is being filled, Orotol leaks
from W57 for approx. 0.5 s.
Perform an inspection and, if necessary, cleaning of the water collector 1 year • • - - - -
and the cleaning adapter.
D3509.105.01.26.02
61 93 903 D3509

8. Assistant element:
Visual check for leakage. 1 year • • - - - -
D3509.105.01.26.02
61 93 903 D3509
Maintenance log for Teneo, Sinius
Work to be performed Interval Teneo Sinius Spare part REF. Qty Bag no.:

8. Assistant element:
Remove suction and saliva ejector hose. Replace upper filter part and O- 1 year • • Upper filter part 18 94 851 2 8.6
ring. Grease O-ring with LCA 3801. O-ring 16.0 x 2.5 70 38 391 2
Replace the grip piece of the saliva ejector. 1 year • • Grip piece 59 11 578 1 8.6
If a 3rd suction connection is installed: Replace filter and O-ring. Grease 1 year • - Filter1) 18 94 851 1 -
O-ring with LCA 3801. O-Ring 16.0 x 2.51) 70 38 391 1
Remove the lower connecting hose running from the support arm to the 1 year • - O-ring 20.3 x 2.4 70 52 269 2 8.6
ASE head in the assistant element. Replace the O-rings and grease with O-ring 24 x 2.5 70 52 251 2
LCA 3801.
Remove connection hose from the WU to the ASE head in the assistant 1 year - • O-ring 20.3 x 2.4 70 52 269 1 8.6
element (Sinius)/on the water unit. Replace the O-rings and grease with O-ring 24 x 2.5 70 52 251 1
LCA 3801.
Clean suction hose handpiece. Replace the O-ring and grease with LCA 1 year • • O-ring 15.0 x 2.0 70 47 160 1 8.6
3801.
Clean selection valve seat. Replace membranes. 1 year • • Membranes 59 29 807 2 8.6

9. Dentist element:
Visual check for leakage. 1 year • • - - - -
Replace filter on heater. 1 year • • Filter 22 57 897 1 9.7
Replace O-ring. O-ring 13.0x1.5 70 47 251 1
Module, instrument holder 2 - 4: 1 year • • 9.8
Replace spray water membranes Membranes 46 95 347 3
Module, instrument holder 5: 1 year • - -
Replace spray water membranes Membranes1) 46 95 347 1
Module, instrument holder 2 - 4: Remove module. 1 year • • 9.8
Replace membranes. Membranes 14 45 696 12
Replace O-ring. O-ring 4.0x1.0 70 36 882 3
Module, instrument holder 5: Remove module. 1 year • - -
Replace membranes. Membranes1) 14 45 696 4
Replace O-ring. O-ring 4.0x1.01) 70 36 882 1
If necessary: Replace water outlet. 2 years • • Water outlet 14 45 670 1 9.8
CS dentist element: Replace oil collector 1 year - • Oil collector 63 31 180 1 9.8
Clean the oil pot and replace the O-ring. 1 year • • O-ring 11.5x1.0 70 47 236 1 9.8
9
10
Maintenance log for Teneo, Sinius
Work to be performed Interval Teneo Sinius Spare part REF. Qty Bag no.:

10. Instruments:
SPRAYVIT: Disconnect handpiece from hose. 1 year • • O-ring 2.2x1.0 hose 18 91 303 4 10.9
Replace O-rings on hose. O-ring 1.5x1.0 heat- 70 41 734 4
Replace O-rings of heating cartridges. ing cartridge
Replace O-rings of nozzle (Caution: Do not confuse with the O-rings of O-ring 5.5x1.03 18 91 840 6
the highspeed handpiece quick coupling). nozzle
Oil all O-rings with a little T1 spray.
Clean nozzle with cleaning wire.
Check device end and hose end contacts for wear.
High-speed handpiece: Disconnect handpiece from hose. Check spray 1 year • • Compression ring 33 27 694 1 10.10
cartridge. Replace blue compression ring on quick coupling. Replace the O-ring 5.5x1.03 18 91 840 4
O-rings of the quick coupling and grease with a little T1 spray.
Check device end and hose end contacts for wear.
SL motor: Disconnect motor cartridge from hose Check carbon brushes 1 year • - O-ring 7.0x1.0 hose 70 41 403 6 10.11
for wear. Replace O-rings of motor hose. Lubricate all O-rings with a little
T1 spray. Check spray cartridge.
Check device end and hose end contacts for wear.
Motor BL ISO: Check latching spring washer and grease with T1 spray. 1 year • • Latching spring 59 42 669
If latching spring washer shows evidence of rust, replace with tool washer1)
64 59 536
BL motor, BL ISO motor, BL ISO C/S motor: Replace O-rings of hand- 1 year • - O-ring 8.4 x 0.7 ISO 58 60 390 8 10.13
piece coupling. adaptor
Motor BL implant: Replace O-rings of handpiece coupling. 1 year • - O-Ring 8x1 70 36 189 1 10.14
O-Ring 15x0.7 62 42 742 1
SIROSONIC TL: Disconnect handpiece from hose. Replace the sealing 1 year • • Sealing ring for the 61 97 797 1 10.15
ring for the screw cap (check that the sealing ring is correctly seated - fit screw cap
the flat face to the front). Use a tip gauge to check the existing instrument
tips and replace them as required after consulting with the customer.
CAUTION: Laser radiation, do not look directly into the beam.

11. Dentist element/assistant element:


Dentist element: Check the propellant, cooling and spray air pressure 1 year • • - - - -
setting using the four-way measuring head. When setting the operating
D3509.105.01.26.02

pressures, observe the manufacturer's instructions for the instruments.


61 93 903 D3509

Adjustment on module with choke RD3 or RD5.


Dentist element: Check spray water flow rate on highspeed handpiece or 1 year • • - - - -
motor: Flow rate > 50 ml/min
D3509.105.01.26.02
61 93 903 D3509
Maintenance log for Teneo, Sinius
Work to be performed Interval Teneo Sinius Spare part REF. Qty Bag no.:

11. Dentist element/assistant element:


Dentist element: Check water flow rate on SPRAYVIT: 1 year • • - - - -
Flow rate > 130 ml/min.
Assistant element: Check water flow rate on SPRAYVIT: 1 year • • - - - -
Flow rate > 130 ml/min.

12. Water unit:


Service block: Check input pressure of air and water supply if necessary. 1 year • • - - - -
Air: 5.5 - 7.5 bar; water 2.5 - 6 bar.

13. Motorized headrest:


During maintenance work on a TENEO treatment center with backrest - • - motorized 62 29 137 1 -
serial number <1937, on this unit, the motorized headrest Ref: 6229137 headrest 1)
must be changed as a one-off due to wear. The headrest is not part of the
maintenance kit and must therefore be ordered separately.

14. Treatment center:


the following work must be performed on completion of
the maintenance work:
Functional test of the safety switches for the assistant element, cuspidor 1 year • - - - - -
bowl, chair lift, backrest, toeboard, lift frame cover, rear housing, left/right
manual switching strip and armrests, as described in the Maintenance
Manual.
Function check of the safety switch, assistant element, cuspidor bowl, 1 year - • - - - -
backrest, footrest, lifting frame, rear cover, as described in the mainte-
nance manual.
If a high-frequency surgical device is installed: Perform a safety check 1 year • • - - - -
according to the Ordinance on the Installation, Operation and Use of
Medical Devices and record the results in the medical product log.
(Required by law in Germany)
If an amalgam separator is installed: Check the optical and acoustic dis- 1 year • • - - - -
play elements for functioning and record the results according to the
operator's log. (Required by law in Germany)
If an amalgam separator is installed: Check and record the function of the 5 years • • - - - -
amalgam separator according to the operator's log. (Required by law in
Germany)
If a disinfection system is installed: Check functionality. 1 year • • - - - -
11
12
Maintenance log for Teneo, Sinius
Work to be performed Interval Teneo Sinius Spare part REF. Qty Bag no.:

14. Treatment center:


the following work must be performed on completion of
the maintenance work:
After all components have been installed, the media lines must be 1 year • • - - - -
checked for functioning and leaks, especially the media and drain paths.
Visual check of all electrical connections, if visible.
Reset the maintenance display: Reset the maintenance display under 1 year • • - - - -
"Service" in the setup dialog.

15. Maintenance log:


If the work specified above has been performed and no defects have 1 year • • Maintenance log for 61 93 895 1 -
occurred, the treatment center must be handed over to the customer Teneo, Sinius
together with the fully completed maintenance log.

16. Required lubricants:


Grease the spindles in the backrest and in the headrest. - • • Grease E 67 77 382 10 g 10.16
Grease the drive spindles of the lifting, tilting and backrest motors, the ball - • • PD2 grease 34 91 271 20g 10.16
bearing/spherical liner running surface and the guideway/roller. 73 95 445 1000g
Greasing of instrument O-rings. - • • T1 spray 18 94 299 50 ml -
Greasing other O-rings with grease LCA 3801. - • • Grease LCA 3801 65 01 261 7g 10.16
Caution: Do not use this grease for the O-rings of the instruments!

17. Required tools:


Tip gauge for checking the instrument tips (ZEG). - • • Tip gauge 54 55 667 - 10.16
Spring hook for changing the funnel in the flushing. - • • Spring hook 35 86 369 - -
Measuring cup for determining the spray water/Sprayvit water flow rate. - • • - - - -
Four-way measuring head for setting the spray air pressure. - - • Four-way measur- 36 66 336 - -
ing head

1) Spare part not included in maintenance set.


D3509.105.01.26.02
61 93 903 D3509
We reserve the right to make any alterations which may be required due to technical improvements.

© Sirona Dental Systems GmbH Sprache: englisch Printed in Germany


D3509.105.01.26.02 06.2021 Ä.-Nr.: 130 912 Imprimé en Allemagne

Sirona Dental Systems GmbH

Fabrikstraße 31 Order No 61 93 903 D3509


64625 Bensheim
Germany
www.dentsplysirona.com

You might also like