Professional Documents
Culture Documents
Introduction
1
2
Meniscus adjustment with
BLAUBRAND ® Volumetric Instruments
Read at the lowest point of the meniscus. Read at the point where the two arrows
touch.
3
1. Preparation for testing
■ Volumetric flasks
☞ Trapezoidal volumetric flasks
☞ Standard volumetric flasks
☞ Amber-glass volumetric flasks
☞ Flanged-rim volumetric flasks
■ Graduated cylinders
■ Mixing cylinders
■ Density Bottles
■ Bulb pipettes
☞ One mark
☞ Two mark
■ Graduated pipettes
☞ Graduated cotton-plug pipettes, total delivery (1 and 2 ml)
☞ Graduated pipettes, total delivery
☞ Graduated Schellbach pipettes, total delivery
☞ Graduated pipettes, partial delivery
☞ Enzymatic test pipettes, partial delivery
■ Burettes
☞ Burettes, lateral needle-valve stopcock
☞ Burettes, lateral glass stopcock
☞ Burettes, straight needle-valve stopcock
☞ Burettes, straight glass stopcock
☞ Micro burettes, lateral needle-valve stopcock
☞ Micro burettes, lateral glass stopcock
☞ Micro burettes, straight needle-valve stopcock
☞ Micro burettes, straight glass stopcock
4
1.1.4 Type of instrument (continued) ⇒ Enter into Test Record
■ Automatic burettes
☞ Automatic burettes with intermediate stopcock and needle-valve stopcock
☞ Automatic burettes with intermediate stopcock and glass stopcock
☞ Automatic burettes without intermediate stopcock and with needle-valve stopcock
■ Other materials
☞ e.g., PP, PMP
5
1.2 Visual examination
■ Storage vessel
☞ Filled with testing liquid ■ Allow to adjust to room temperature.
(distilled or deionised water)
■ Weighing vessel
☞ Erlenmeyer flask 100 ml,
containing a small amount of testing
liquid (bottom should be covered)
■ Funnel
☞ To fill volumetric instrument
■ Thermometer
☞ Accuracy 0.2 °C
■ Balance
☞ Must have required accuracy
(approx. 10 times accuracy of
the instrument).
6
1.4 Testing equipment and accessories (continued)
■ Stopwatch
☞ To keep track of the waiting time,
accuracy ± 1 s.
■ Lint-free tissue
☞ For wiping
■ Pipetting aid
☞ e.g., macro pipette controller from BRAND
■ Barometer
☞ For testing the atmospheric pressure,
accuracy ± 5 hPa
2. Gravimetric test
■ Determine testing temperature (testing liquid). ⇒ Enter value into Test Record
■ Determine empty weight of the dry volumetric ⇒ Enter value into Test Record
instrument. (W 1)
7
2.1.2 Graduated pipettes, calibrated to contain (AR-Glas ®)
■ Determine testing temperature (testing liquid). ⇒ Enter value into Test Record
■ Determine empty weight of the dry volumetric ⇒ Enter value into Test Record
instrument. (W 1)
■ Determine testing temperature (testing liquid). ⇒ Enter value into Test Record
■ Determine empty weight of the dry density ⇒ Enter value into Test Record
bottle. (W 1)
■ Fill density bottle bubble-free with testing liquid. The socked should be filled to 1/3.
■ Put the stopper or the thermometer carefully onto the density bottle matching the markings of density bottle and
stopper or thermometer. The capillary fills and some testing liquid runs out.
■ Surface of stopper or capillary and the outside of the density bottle must be carefully dried.
Note: Avoid contacting the capillary with the tissue. The testing liquid must exactly level with the upper end of
the capillary.
■ Determine testing temperature (testing liquid) ⇒ Enter value into Test Record
■ Determine weight of the weighing vessel ⇒ Enter value into Test Record
again. (W 2 )
Note:
In case of pipettes graduated for partial delivery, let the water run out until approx. 10 mm above the lower mark, while
the pipette tip touches the inclined wall of the weighing vessel. After 15 seconds waiting time, adjust the meniscus
precisely to the mark.
■ Determine testing temperature (testing liquid). ⇒ Enter value into Test Record
The standard ISO 4787 describes the testing of volumetric instruments and provides the following
general equation for calculations:
1 ρ
V20 = (W 2 - W 1)( )(1 - L )(1 - γ (t - 20 ° C))
ρW - ρL ρG
Since this equation is rather cumbersome to work with, requiring a multitude of tables,
we are providing a simplified calculation in these Testing Instructions.
V20 = (W 2 - W 1) ⋅ z
The parameters are: V 20 [ml]: volume of the instrument at 20 °C
W 1 [g]: weight of the empty instrument/respectively weight of weighing vessel before delivery
W 2 [g]: weight of the filled instrument/respectively weight of weighing vessel after delivery
z [ml/g]: factor resulting from united parameters (see Tables)
To simplify the list of measuring instruments even further, we recommend to use conformity-certified
BLAUBRAND ® volumetric instruments marked with batch number or individual serial number.
The initial testing of certified volumetric instruments can be omitted since the testing results are
already confirmed in the certificate.
Example:
Factor "z" from Table 1, medium atmospheric pressure range, since capacity of volumetric flask ≤ 250 ml:
z 23 °C, DURAN = 1.00348 ml/g
®
■ Table 1
☞ Table 1 gives Factor "z" for the temperatures from
15 °C to 30 °C and atmospheric pressures from 980 hPa
to 1040 hPa for glass types AR-Glas ® and DURAN ®.
■ Table 2
☞ If volumetric instruments made of plastic are also to be tested,
Table 2 provides Factor "z" for PP and PMP.
11
Testing of volumetric instruments
Factor "z" [ml/g]
Table 1
Note:
Intermediate values can be linearly interpolated easily.
Tables based on other atmospheric pressure ranges are available on request.
12
Testing of volumetric instruments
Factor "z" [ml/g]
Table 2
Note:
Tables based on other atmospheric pressure ranges are available on request.
13
Notes:
14
Test Record for Volumetric Instruments
Trademark: ❏ BLAUBRAND ®
❏ BLAUBRAND ®-ETERNA
❏l l
2. Damages: ❏ None
❏ Type of damage l
L
4. Calculation: V 20 = (W 2 - W 1) ⋅ z
5. Evaluation:
Weighing Value W 2 Value W 1 Factor "z" Volume V 20
values [g] [g] [ml/g] [ml]
X1
X2
X3
Mean value:
Cut out and copy!
Date: Tester: 15
✂
4. EASYCAL ™ 3.0 Software – advanced calibration technology
4.1 For liquid handling instruments and glass or plastic volumetric instruments
☞ Connect PC and balance (optional), then start the EASYCAL™ 3.0 software.
☞ 40 common balances are preprogrammed for ease of installation.
The test record contains all important test data on one page, including a
graphic representation of test results.
16
5. DKD-Calibration from BRAND for Volumetric Instruments
The DKD-Calibration Certificate with the DKD-Calibration Symbol documents officially on a high level the traceability of
measuring results to national and international standards and to international SI-units, as required by standards as DIN EN
ISO 9001, DIN EN ISO 10012, DIN EN ISO/IEC 17025 and DIN EN 45000 for monitoring of measuring devices.
The major difference between traditional works calibration services and PTB-accredited calibration laboratories is the accu-
rate determination of uncertainty of each measurement, guaranteed by the accredited laboratory and supervised by the PTB.
DKD-Calibration Certificates are issued when calibrations of an accredited laboratory are requested, when high level
calibrations are necessary, when national and international standards are demanded and when reference instruments have
to be calibrated.
5.3 Internationally recognized
TThe DKD is member of the European Cooperation for Accreditation (EA). A multilateral agreement assures obligatory recognition
of the DKD Calibration Certificate in a variety of countries. In addition, a „Mutual Recognition Arrangement“ (MRA) has been signed
in Washington, DC, on 2 November 2000. This new arrangement, which involves 36 member bodies from 28 economies
represented at the General Assembly of the International Laboratory Accreditation Cooperation (ILAC, www.ilac.org), has enabled
accredited laboratories, as such accredited by PTB in Germany, to achieve a form of international recognition, and allowed test
data accompanying exported goods to be more readily accepted on overseas markets. This is a major step towards reducing or
eliminating the need for re-testing of the goods by the importing country.
5.4 DKD-K-20701
In 1998 a calibration laboratory for volumetric instruments at BRAND has been accredited by PTB according to DIN EN
45001, registration number DKD-K-20701. Our calibration laboratory is therefore authorized to issue DKD Calibration
Certificates for the volumetric instruments listed below. Calibration is available for new instruments directly from the factory
as well as for those already in use – regardless of their make. Furthermore we offer adjustment and – for BRAND-products
– repair. For ordering information on DKD Calibration Certificates for volumetric instruments please consult our General
Catalog 600.
5.5 Volumetric instruments for which you can obtain DKD Calibration Certificates
from BRAND
BRAND calibrates the following volumetric instruments (new or already in use and regardless of their make):
■ Piston-operated pipettes, from 0.1 µl to 10 ml
■ Multichannel piston-operated pipettes, from 0.1 µl to 300 µl
■ Piston-operated burettes, from 5 µl to 200 ml
■ Dispensers, Dilutors, from 5 µl to 200 ml
■ Volumetric instruments of glass, adjusted to contain (TC, In) from 1 µl to 10 000 ml
■ Volumetric instruments of glass, adjusted to deliver (TD, Ex) from 100 µl to 100 ml
■ Volumetric instruments of plastic, adjusted to contain (TC, In) from 1 ml to 2000 ml
■ Volumetric instruments of plastic, adjusted to deliver (TD, Ex) from 1 ml to 100 ml
■ Density bottles of glass, from 1 cm 3 to 100 cm 3 17
18
19
9976 36 · Printed in Germany · 5/0900
20