Professional Documents
Culture Documents
Introduction
The B-SUGAR calibration package contains methods for the analysis of products and
intermediates that are handled in sugar industry. As the samples can be solid or liquid products
the package consists of methods for measurement in reflection as well as in transflection mode.
All methods are intended for the usage of the integrating sphere measuring channel.
Generally the reflection mode is used for solid samples whereas the transflection mode is used
for pasty samples and all liquid samples of high viscosity. For some samples (i.e. molasses) we
offer methods for reflection as well as transflection measurements. The selection of the
measurement mode depends on the viscosity of the sample. If the transflection stamp can be
pushed easily to the bottom of the vial the transflection technique should be applied. Otherwise,
i.e. when the sample has a higher viscosity the reflection technique is the appropriate way for
analysis.
Bruker offers the service package S-SUGAR in combination with this calibration package. Within
the service package you can provide samples with reference values in order to receive specific
methods that are optimized for the customer’s samples if required. Please contact your local
Bruker representative to organize the method optimization.
Page 2 of 12
B-SUGAR: Product Overview
Product Overview
Product Measurement Mode Sample Measurement
containment Conditions
Filter Cake Reflection Quartz Cup
Juices Transflection 22mm vial With Reflection
Stamp a)
Molasses Reflection 22mm vial
Sugar Reflection Quartz Cup
Bagasse Reflection Quartz Cup
Factory Products (Refl.) Reflection 22mm vial
Factory Products (Transfl.) Transflection 22mm vial With Reflection
Stamp a)
Syrups Transflection 22mm vial With Reflection
Stamp a)
Quartz Cup: Sample containment with quartz window. For ground samples you can use 51 mm
diameter cups (IN311-S/IN311-P) or 97 mm diameter cups (IN312-S/IN312-
SH/IN312-P).
The IN312 cups provide a bigger sample area and the better averaging is more
suitable for inhomogeneous samples.
22mm vial Glass vial with flat bottom, plain top with 22mm outer diameter (IN441).
a)
For transflection measurements a diffuse reflecting device made from stainless steel
(“reflection stamp”) is inserted into the sample vial (IN441). Precise spacers ensure an optical
path length of 2mm. Product number of the reflection stamp: IN481S-2.
Page 3 of 12
B-SUGAR: Sample Containments and how they should be filled
Page 4 of 12
B-SUGAR: Filter Cake
● Filter Cake
Measurement Parameters:
Resolution /cm-1 Scans Sample Rotation Comment
16 64 Enabled Apply straylight correction
Reflection measurement: Use quartz cup, see comment p. 3.
Sample Preparation:
- No sample preparation.
Filenames:
Product file: SUGAR_FilterCake.opx
Multi-Evaluation: SUGAR_FilterCake_nnnn.mev
Quant-2: Sugar_FilterCake_Q2-Parameter_nnnnnn.q2
Experimental setup: SUGAR_solid_refl_rotating.XPM
Protection mode: Change Parameter Mode
Parameter Q2- Unit n min max Comment
Parameter
Moisture Moisture % 136 65.2 79.3
Pol POL % 136 0.2 5.8
● Product Overview
Page 5 of 12
B-SUGAR: Juices
● Juices
Products: First juice, last juice, clarified juice, lime juice, filtrate.
Measurement Parameters:
Resolution /cm-1 Scans Sample Rotation Comment
16 64 Disabled Apply straylight correction
Transflection measurement: Use 22mm vial (IN441) and insert reflection stamp
(IN481S-2)!
Sample Preparation:
- No sample preparation.
Filenames:
Product file: SUGAR_Juices_Transfl.opx
Multi-Evaluation: SUGAR_Juices_Transfl_nnnn.mev
Quant-2: Sugar_Juices_Transfl_Q2-Parameter_nnnnnn.q2
Experimental setup: SUGAR_FactoryProducts_Transfl.xpm
Protection mode: Change Parameter Mode
Parameter Q2- Unit n min max Comment
Parameter
Pol POL % 2416 0.5 19.3
Brix Brix % 2467 0.6 21.7
● Product Overview
Page 6 of 12
B-SUGAR: Molasses
● Molasses
Measurement Parameters:
Resolution /cm-1 Scans Sample Rotation Comment
16 64 Disabled Apply straylight correction
Reflection measurement: Use 22mm vial (IN440 or IN441).
Sample Preparation:
- No sample preparation.
Filenames:
Product file: SUGAR_Molasses_refl.opx
Multi-Evaluation: SUGAR_Molasses_Refl_nnnn.mev
Quant-2: Molasse_refl_Q2-Parameter_nnnnnn.q2
Experimental setup: SUGAR_FactoryProd_refl_nonrotating.xpm
Protection mode: Change Parameter Mode
Parameter Q2- Unit n min max Comment
Parameter
Pol POL % 104 31.0 62.3
Brix Brix % 104 68.0 88.4
● Product Overview
Page 7 of 12
B-SUGAR: Sugar
● Sugar
Measurement Parameters:
Resolution /cm-1 Scans Sample Rotation Comment
16 64 Enabled Apply straylight correction
Reflection measurement: Use quartz cup, see comment p. 3.
Sample Preparation:
- No sample preparation.
Filenames:
Product file: SUGAR_Sugar.opx
Multi-Evaluation: SUGAR_Sugar_nnnn.mev
Quant-2: AllSugar_Q2-Parameter_nnnnnn.q2
Experimental setup: SUGAR_solid_refl_rotating.XPM
Protection mode: Change Parameter Mode
Parameter Q2- Unit n min max Comment
Parameter
Ash Ash % 739 0.00 0.26
Ash Ash % 456 0.00 0.04 Low range mev)
Moisture Moisture % 2892 0.0 0.9
Pol POL % 703 97.0 100.0
Color Color 813 1 4204
Reducing Sugars RS % 625 0.0 0.8
mev)
The low range calibration is implemented as sub-method of the method for the complete
concentration range within the Multi-Evaluation-file (.mev).
● Product Overview
Page 8 of 12
B-SUGAR: Bagasse
● Bagasse
Products: Bagasse.
Measurement Parameters:
Resolution /cm-1 Scans Sample Rotation Comment
16 64 Enabled Apply straylight correction
Reflection measurement: Use quartz cup, see comment p. 3.
Sample Preparation:
- No sample preparation.
Filenames:
Product file: SUGAR_Bagasse.opx
Multi-Evaluation: SUGAR_Bagasse_nnnn.mev
Quant-2: SUGAR_Bagasse_066a_Q2-Parameter_nnnnnn.q2
Experimental setup: SUGAR_solid_refl_rotating.XPM
Protection mode: Change Parameter Mode
Parameter Q2- Unit n min max Comment
Parameter
Pol POL % 220 1.1 9.9
● Product Overview
Page 9 of 12
B-SUGAR: Factory Products - Reflection
Calibrations for the analysis of highly viscous factory products and intermediates in reflection
mode.
Measurement Parameters:
Resolution /cm-1 Scans Sample Rotation Comment
16 64 Disabled Apply straylight correction
Reflection measurement: Use 22mm vial (IN440 or IN441).
Sample Preparation:
- No sample preparation.
Filenames:
Product file: SUGAR_FactoryProd_Refl.opx
Multi-Evaluation: SUGAR_FactoryProd_Refl_nnnn.mev
Quant-2: SugarFactoryProd_refl_Q2-Parameter_nnnnnn.q2
Experimental setup: SUGAR_FactoryProd_refl_nonrotating.xpm
Protection mode: Change Parameter Mode
Parameter Q2- Unit n min max Comment
Parameter
Pol POL % 362 47.4 92.9
Brix Brix % 364 70.8 98.4
● Product Overview
Page 10 of 12
B-SUGAR: Factory Products - Transflection
Calibrations for the analysis of low and mid viscous factory products in transflection mode.
Products: Juices (first juice, last juice, clarified juice, lime juice, molasses, pan feed, syrups
(raw syrup, fine syrup, golden syrup, invert syrup), remelt, liquors (carbonate
liquor).
Measurement Parameters:
Resolution /cm-1 Scans Sample Rotation Comment
16 64 Disabled Apply straylight correction
Transflection measurement: Use 22mm vial (IN441) and insert reflection stamp
(IN481S-2)!
Sample Preparation:
- No sample preparation.
Filenames:
Product file: SUGAR_FactoryProd_Transfl.opx
Multi-Evaluation: SUGAR_FactoryProd_Transfl_nnnn.mev
Quant-2: SugarFactoryProd_Transfl_Q2-Parameter_nnnnnn.q2
Experimental setup: SUGAR_FactoryProducts_Transfl.xpm
Protection mode: Change Parameter Mode
Parameter Q2- Unit n min max Comment
Parameter
Pol POL % 4100 0.5 70.5
Brix Brix % 4639 0.6 88.4
● Product Overview
Page 11 of 12
B-SUGAR: Syrups - Transflection
● Syrups - Transflection
Measurement Parameters:
Resolution /cm-1 Scans Sample Rotation Comment
16 64 Disabled Apply straylight correction
Transflection measurement: Use 22mm vial (IN441) and insert reflection stamp
(IN481S-2)!
Sample Preparation:
- No sample preparation.
Filenames:
Product file: SUGAR_Syrups_Transfl.opx
Multi-Evaluation: SUGAR_Syrups_Transfl_nnnn.mev
Quant-2: Syrups_Transfl_Q2-Parameter_nnnnnn.q2
Experimental setup: SUGAR_FactoryProducts_Transfl.xpm
Protection mode: Change Parameter Mode
Parameter Q2- Unit n min max Comment
Parameter
Pol POL % 1206 14.2 65.2
Brix Brix % 1173 16.7 75.2
● Product Overview
Page 12 of 12